Skip to main content

REVIEW article

Front. Physiol., 02 December 2021
Sec. Cardiac Electrophysiology
This article is part of the Research Topic Advances in Cardiac Pacing and Neural Control Strategies: Basic, Translational and Clinical Research View all 5 articles

Advances in Cardiac Pacing: Arrhythmia Prediction, Prevention and Control Strategies

\r\nMehrie Harshad PatelMehrie Harshad Patel1Shrikanth SampathShrikanth Sampath1Anoushka KapoorAnoushka Kapoor1Devanshi Narendra DamaniDevanshi Narendra Damani1Nikitha ChellapuramNikitha Chellapuram2Apurva Bhavana ChallaApurva Bhavana Challa1Manmeet Pal KaurManmeet Pal Kaur3Richard D. Walton,,Richard D. Walton4,5,6Stavros StavrakisStavros Stavrakis7Shivaram P. Arunachalam,,,*Shivaram P. Arunachalam1,2,3,8*Kanchan Kulkarni,,*Kanchan Kulkarni4,5,6*
  • 1Department of Cardiovascular Diseases, Mayo Clinic, Rochester, MN, United States
  • 2Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, MN, United States
  • 3Department of Medicine, GAIL, Mayo Clinic, Rochester, MN, United States
  • 4IHU LIRYC, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, Bordeaux, France
  • 5Centre de Recherche Cardio-Thoracique de Bordeaux, University of Bordeaux, Bordeaux, France
  • 6INSERM, Centre de Recherche Cardio-Thoracique de Bordeaux, Bordeaux, France
  • 7Heart Rhythm Institute, University of Oklahoma Health Sciences Center, Oklahoma City, OK, United States
  • 8Department of Radiology, Mayo Clinic, Rochester, MN, United States

Cardiac arrhythmias constitute a tremendous burden on healthcare and are the leading cause of mortality worldwide. An alarming number of people have been reported to manifest sudden cardiac death as the first symptom of cardiac arrhythmias, accounting for about 20% of all deaths annually. Furthermore, patients prone to atrial tachyarrhythmias such as atrial flutter and fibrillation often have associated comorbidities including hypertension, ischemic heart disease, valvular cardiomyopathy and increased risk of stroke. Technological advances in electrical stimulation and sensing modalities have led to the proliferation of medical devices including pacemakers and implantable defibrillators, aiming to restore normal cardiac rhythm. However, given the complex spatiotemporal dynamics and non-linearity of the human heart, predicting the onset of arrhythmias and preventing the transition from steady state to unstable rhythms has been an extremely challenging task. Defibrillatory shocks still remain the primary clinical intervention for lethal ventricular arrhythmias, yet patients with implantable cardioverter defibrillators often suffer from inappropriate shocks due to false positives and reduced quality of life. Here, we aim to present a comprehensive review of the current advances in cardiac arrhythmia prediction, prevention and control strategies. We provide an overview of traditional clinical arrhythmia management methods and describe promising potential pacing techniques for predicting the onset of abnormal rhythms and effectively suppressing cardiac arrhythmias. We also offer a clinical perspective on bridging the gap between basic and clinical science that would aid in the assimilation of promising anti-arrhythmic pacing strategies.

Introduction

Sudden cardiac death (SCD) instigated by cardiac arrhythmias is one of the leading causes of mortality worldwide, accounting for about 300,000 deaths annually in the United States alone (Srinivasan and Schilling, 2018). Atrial tachyarrhythmias (abnormally faster atrial activation rates) such as atrial flutter and atrial fibrillation (AF), although non-fatal are associated with substantial complications, most commonly thromboembolic (cerebral, extremities, or visceral) leading to increased risk of strokes, limb ischemia, organ infarctions and hence, hospitalizations (Petersen, 1990). Bradyarrhythmias (abnormally slower heart rates) originating due to disorders of the atrial conduction pathways may either lead to morbidity or progress to SCD by instigating ventricular tachycardia (VT) or asystole (Khurshid et al., 2018). Ventricular tachyarrhythmias (VT/VF) on the other hand can be lethal and are the primary cause of SCD, mandating immediate clinical intervention. About 1.5–5% of the general population is currently afflicted by arrhythmias, either atrial or ventricular (Desai and Hajouli, 2021).

The past decade has seen an abundance of medical devices aiming to restore normal cardiac rhythm with new pacing strategies being designed to apply electrical stimulations for the prevention and suppression of arrhythmias. From anti-tachycardia pacing and novel pacemaker algorithms, to cardiac resynchronization therapy and implantable defibrillators, multiple electrical stimulation modalities have been proposed and implemented for arrhythmia management (Trohman et al., 2004). Conventional non-physiological pacing from the right ventricular apex has shown reliable results in the treatment of bradyarrhythmias, however, its adverse hemodynamics of dyssynchronous ventricular contraction predisposes the patients to develop AF, cardiomyopathy, heart failure (HF) or death (Vijayaraman et al., 2017; Arora and Suri, 2021; Perla et al., 2021). An alternative approach is physiological cardiac pacing, which transmits electrical impulses via the normal conduction pathway (Arora and Suri, 2021) and has shown a relative risk reduction of 27% for new onset chronic AF compared to ventricular pacing (Skanes et al., 2001; Trohman et al., 2004), yet technical challenges hinder its application (Arora and Suri, 2021).

Technological advances in pacing have enabled rate responsive or adaptive pacing which successfully tackle chronotropic incompetence wherein the sino-atrial node is unable to increase the heart rate to meet the changing metabolic demands (Trohman et al., 2004, 2020). Most modern devices with adaptive pacing have sensors to detect physiological or physical indices of activity and simulate normal sino-atrial nodal response (Trohman et al., 2004, 2020). Yet, despite the advances in cardiac arrhythmia treatment modalities, given the spatiotemporal and structural complexity of the human heart, designing algorithms to effectively control the heart rate and prevent fatal rhythms has been challenging. In addition, although implantable cardioverter defibrillator (ICD) shocks and anti-tachycardia pacing techniques can protect individuals from life threatening arrhythmias, excessive and inappropriate pacing or shocks have detrimental effect, ranging from impaired myocardial contraction leading to systolic dysfunction, progressive HF, unnecessary drug treatment, proarrhythmogenicity, reduced quality of life and psychological impacts on the patient (Germano et al., 2006; Bhavnani et al., 2010).

Thus, there is a pressing need to better understand cardiac dynamics, the initiation of unstable rhythms in the heart, and subsequently, their prevention to minimize unnecessary shock and drug treatments. Here, we present a systematic review of the state-of-the-art cardiac pacing methodologies aimed at predicting and controlling cardiac arrhythmias. We present a clinical perspective on the current arrhythmia prevention strategies and discuss the efforts to bridge the gap between translational and clinical science that would enable the assimilation of novel pacing modalities.

Search Strategy

Multiple databases including Google Scholar and PubMed, were comprehensively searched without language and time restriction. For prediction techniques, the literature search was focused on traditionally used clinical methods as well as novel techniques incorporating non-linear dynamical and artificial intelligence (AI) models. The following keywords were used as search criteria: “(clinical markers OR deep learning OR artificial intelligence OR machine learning OR EP study OR non-linear dynamics OR clinical diagnosis OR prediction) AND cardiac arrhythmias.” For arrhythmia prevention and control methods, the literature search was focused on clinical and translational pacing modalities for suppressing abnormal cardiac rhythms. The following keywords were used as search criteria: “(prevention OR control OR suppression OR overdrive pacing OR biventricular pacing OR autonomic control OR stochastic pacing OR diastolic interval pacing OR novel pacing method) AND cardiac arrhythmias”.

Management of Atrial Arrhythmias

Prediction Modalities

Traditional Clinical Methods

Various clinical tests and devices are traditionally used to diagnose/predict arrhythmias including ambulatory electrocardiogram (ECG) monitors and event recorders, trans-telephonic transmitters, stress tests, echocardiography, electrophysiology (EP) studies, or tilt table tests (Clinic, 2018). ECGs remain the predominant clinical tool for monitoring cardiac electrical activity and predicting the onset of cardiac arrhythmias (Castro-Torres et al., 2015).

Holter monitoring is a prominent diagnostic method used to monitor the heart continuously for 24–72 h (Clinic, 2021). This technique was introduced in the 1960’s and has been used since as an integral tool in cardiac arrhythmia prediction (De Asmundis et al., 2014). There are limitations of Holter monitoring if the arrhythmias only occur sporadically; however, it remains the most frequently used method to monitor individuals for cardiac arrhythmias (De Asmundis et al., 2014). An event monitor, similar to a Holter, is routinely recommended to be worn for a longer period of time and used for rarer arrhythmias (Clinic, 2019). These are user friendly devices that start recording cardiac electrical activity when the patient has onset of arrhythmic symptoms (Clinic, 2019). Multiple ambulatory devices are now in practice that enable monitoring vital signs both in-hospital and in remote populations (Kulkarni et al., 2021a). Trans-telephonic transmitters on the other hand are a slight variation from other ambulatory devices and are not worn throughout the day but rather only during phone monitoring sessions (Clinic, 2019).

Non-invasive cardiac mapping is now extensively used in clinical practice (Dubois et al., 2015) to identify the sources of electrical disorders and guide catheter ablation of atrial arrhythmias (premature atrial beat, atrial tachycardia, AF), ventricular arrhythmias and ventricular pre-excitation (Wolff–Parkinson–White syndrome) (Dubois et al., 2015). In conjunction, echocardiography is often used when managing patients with congenital heart disease and supraventricular arrhythmias (Papadopoulos et al., 2018). It aids clinicians in making decisions about therapies for rhythm/rate control, cardiac ablations, and left atrial appendage closures (Papadopoulos et al., 2018).

Tilt table tests are another useful clinical modality for arrhythmia diagnosis in specific subgroups of patients presenting with syncope. Studies have demonstrated that patients with a positive head-up tilt table test are likely to exhibit arrhythmias (Hermosillo et al., 1999; Lj et al., 2010; Teodorovich and Swissa, 2016) and in particular sinus bradycardia, junctional escape rhythm or sinus arrest (Lj et al., 2010). A retrospective study of patients with a history of syncope, who had undergone head-up tilt table test with nitroglycerin or isoprenaline demonstrated a higher yield of a positive response, lower incidence of unwanted arrhythmias and better tolerability with nitroglycerin provocation compared to isoprenaline in patients with suspected neuro cardiogenic syncope (Prabhu et al., 2017). Yet, in recent years, the clinical applicability of this test has been questioned given the high false positive rates and the failure to establish an explicit cause of syncope making it incapable of guiding treatment (García-Civera et al., 2005; Kulkarni et al., 2020).

Novel Non-linear Dynamical Methods

Advances in data science techniques and the use of machine learning (ML) and deep learning (DL) have enabled the development of accurate models that can analyze large amounts of clinical data to classify cardiac rhythms and predict the onset of cardiac arrhythmias (Table 1). AI has shown its presence in cardiac electrophysiology since the 1970s in the form of automated ECGs (Krittanawong et al., 2020). Nonetheless, recent developments of expertly labeled data in large electronic databases have increased the value of AI in cardiac imaging and electrophysiology and has been put to use for enhanced prediction, response to therapy and characterization of health and diseases.

TABLE 1
www.frontiersin.org

Table 1. Prediction of cardiac arrhythmias: summary of recent non-linear dynamical and artificial intelligence (AI) algorithms for arrhythmia classification and prediction.

Many studies have shown ML methods to be superior over statistical-based models in predicting cardiovascular events (Patel and Sengupta, 2020; Bazoukis et al., 2021). Shakibfar et al. (2019) used a random forest ML model to analyze daily summaries of ICDs monitoring to predict electrical storms (arrhythmic syndrome). Using data from 19,935 patients, they demonstrated that short risk of arrhythmic syndrome could be successfully predicted using a ML model with an accuracy of 96%, significantly outperforming traditional logistic regression analysis. Another recent study found that after training a deep neural network with >1 million 12-lead resting EKGs, the model was able to predict new-onset AF within 1 year in patients with no history of AF with an area under the receiver operating characteristic curve of 0.85 (Raghunath et al., 2021). The DL model significantly outperformed both the extreme gradient boosting (XGBoost) (Chen and Guestrin, 2016) model using only age and sex as inputs and the CHARGE-AF (Cohorts for Aging and Research in Genomic Epidemiology) 5-year risk prediction model (Alonso et al., 2013). These promising studies demonstrate the utility of AI models to predict cardiac arrhythmias, while also highlighting the importance of labeling clinical data and outcomes, and harmonizing data across institutions for improved diagnostics.

Prevention and Control of Atrial Arrhythmias

Traditional Clinical Methods

Clinical management of arrhythmia begins by formulating a definitive diagnosis. A goal directed treatment approach for the specific type of arrhythmia is mandatory. Management strategies have been structured based on the use of antiarrhythmic drugs or ICDs or both. Even with recent advances, pharmacological therapy has remained as mainstay unless contraindicated. Traditionally, many studies conducted to study the risks and benefits of the currently available antiarrhythmic drugs have shown that except for Amiodarone and Sotalol, other classes of antiarrhythmic medications are not sufficiently effective (Auer et al., 2002). Amiodarone and Sotalol are primarily classified as class III antiarrhythmic drugs which act by blocking the outward potassium channels responsible for the repolarization of the heart. Both are known to reduce the risk of ICD shock from ventricular as well as atrial arrhythmia (Auer et al., 2002; John et al., 2012). Though Amiodarone is known to be better tolerated in comparison to the other drugs, its use and overall efficacy is limited due to its known side effects affecting the cardiac, ophthalmologic, thyroid, and hepatic systems (Auer et al., 2002; Florek and Girzadas, 2018; Lantz and Agarwal, 2021). Thus, ongoing pharmacological interventional studies are now focused on the selective ion channels [e.g., atrial I(Na), I(Kur) and I(K, ACh)], pathology-selective ion channels [constitutively active I(K, ACh), TRP channels], ischemia-uncoupled gap junctions, proteins related to malfunctioning intracellular Ca(2+) homeostasis [e.g., “leaky” ryanodine receptors, overactive Na(+), Ca(2+) exchanger] or risk factors for arrhythmias (“upstream” therapies) (Ravens, 2010).

In addition, catheter ablation is considered the mainstay therapy for drug-refractory AF and flutter (Hindricks et al., 2021). It has been an important part of the treatment protocol for a wide spectrum of atrial arrhythmias (Lee et al., 2012). Focal atrial tachycardia and typical atrial flutter is treated with catheter ablation as a first line with an efficacy of over 90%. Furthermore, ablation is sought to provide highly effective palliation and improved quality of life in patients who have had previous atrial surgery or complex congenital heart conditions and are hence, prone to developing late flutter recurrence or AF (Lee et al., 2012). In patients with persistent AF, the role of catheter ablation continues to evolve, as does our understanding of the crucial underlying mechanisms (Lee et al., 2012).

Atrial arrhythmias are also often controlled traditionally by overdrive pacing. The Pacemaker Atrial Fibrillation Suppression study performed a randomized, multicentre investigation of the effects of the third generation anti-atrial fibrillation pacemaker algorithms in patients with paroxysmal AF and demonstrated that the rate-soothing algorithm by atrial overdrive pacing decreased premature atrial contractions-initiated AF (Sulke et al., 2007). Similarly the ADOPT (Atrial Dynamic Overdrive Pacing Trial) study, conducted in patients with symptomatic paroxysmal or persistent AF and sinus node dysfunction demonstrated that overdrive atrial pacing with the AF Suppression Algorithm (St. Jude Medical Cardiac Rhythm Management Division, Sylmar, CA, United States) decreased symptomatic AF burden significantly in patients with sick sinus syndrome and AF (Carlson et al., 2003).

Implantable pacemakers have been a breakthrough in preventing atrial arrhythmias by using specific atrial pacing algorithms (Dimmer et al., 2003). Various studies have been done in the past which have shown that majority of the paroxysmal AF episodes are initiated by bradycardias, premature atrial complexes (PAC) or immediate reinitiation of AF, and multiple pacing modalities have been developed to prevent AF recurrences (Hoffmann et al., 2000; Attuel et al., 2003; Yang et al., 2003). These algorithms primarily rely on the classification of atrial events by the pacemakers and work either by responding to triggers or by continuous overdrive atrial pacing.

Commonly utilized arrhythmia control algorithms currently in clinical practice include pace conditioning, PAC suppression, post-PAC response, post-exercise response, post-AF response and rate soothing (Mitchell and Sulke, 2004). The post AF response algorithm is shown to prevent tachyarrhythmias which occur soon after termination of the previous episode due to immediate reinitiation of AF. After confirming the termination of an atrial tachycardia episode, the algorithm increases the pacing rate to the programmed response rate of 70–100 bpms for the next 600 beats and decays slowly until sinus rhythm is detected or a lower rate limit is achieved. Pace conditioning works similar to the dynamic atrial overdrive algorithm in St. Jude Medical (Sylmar, CA, United States) pacemakers, atrial preference pacing in Guidant (St. Paul, MN, United States) pacemakers, the DDD+ algorithm in Biotronik (Berlin, Germany) pacemakers and atrial preference pacing in Medtronic (Minneapolis, MN, United States) pacemakers. It works by adjusting the atrial pacing rate to just above the underlying intrinsic rhythm in a way that the atrium is paced for almost 95% of time. PAC suppression works by increasing the heart rate on detection of PACs in order to reduce incidence of PACs. The pacing rate is increased by 15 bpm above the physiological rate of 600 beats on detection of PACs. This increased rate is kept stable for a period of time and any PAC detected in this stable phase does not lead to further increase in the rate. After the stable period, the atrial rate decays by 1 bpm every 16 beats.

Increased vagal tone and consequent bradycardia in athletes can precipitate AF. The post-exercise pacing algorithm works by preventing rapid drop in heart-rate after exercise by enabling a post-exercise rate. During exercise, the post-exercise pacing rate slowly rises to 90% of the physiological rate which is dependent on the difference between the two rates. When the heart rate suddenly decreases after exercise, the pacemaker paces at the post-exercise pacing rate. The Medtronic “atrial rate stabilization” works using the principle of post PAC response. Post PAC response prevents pauses after PACs by controlling the atrial rate in the two beats after a PAC. Finally, the rate soothing technique works similar to pace conditioning except that the heart rate is not substantially increased but rather atrial tachycardias are prevented by overdrive pacing the atrium at a rate that is only just above the sinus rate.

Despite these promising advances in arrhythmia control algorithms, the clinical efficacy of the current conventional methods including overdrive pacing remain questionable due to false positives and adverse effects on device battery life (Camm et al., 2007; de Voogt et al., 2007; Hohnloser et al., 2012), warranting the need for novel approaches to suppress abnormal cardiac rhythms. Furthermore, though pharmacological treatment along with interventional techniques is the mainstay of cardiac arrhythmia management, studies conducted consistently demonstrated a strong association between lifestyle interventions and arrhythmia prevention. Obesity (defined as BMI ≥ 30 kg/m2) is documented to be a strong risk factor and associated with the incidence of AF (Chung et al., 2020). It has been demonstrated that a patient who lost and maintained the loss of ≥10% of their body weight enjoyed a sixfold arrhythmia-free likelihood compared with those who lost <3% or gained weight (Chung et al., 2020). Regular aerobic exercise as well as moderate intensity of physical activity is highly recommended (Chung et al., 2020). A randomized control trial of 28 patients with persistent AF compared a non-exercise control with 2 months of moderate-intensity continuous training (Chung et al., 2020). The study demonstrated the latter subjects showed reduced AF-related symptoms along with improved quality of life and exercise capacity.

Autonomic Modulation of Cardiac Arrhythmias

The cardiovascular system is innervated by the central nervous system, the extrinsic intrathoracic ganglia and the intrinsic cardiac autonomic nervous system (CANS). Autonomic and neurohormonal reflexes regulate the CANS. CANS balance is maintained via baroreflex which maintains the blood pressure and after load on the heart achieved through sympathetic-parasympathetic balance (Benarroch, 2008) while the neurohormonal feedback is achieved by the renin-angiotensin aldosterone system (Goldsmith, 2004). For over 20 years autonomic modulation has been used for neurological diseases like epilepsy (Ben-Menachem, 2002) and psychiatric diseases like depression (Bajbouj et al., 2010). In recent years, studies have shown that autonomic modulation can play a role in treatment of cardiovascular diseases like heart failure and arrhythmias (Florea and Cohn, 2014; Goldberger et al., 2019; Stavrakis et al., 2020a; Kulkarni et al., 2021b).

Autonomic modulation can be used for the treatment of atrial as well as ventricular arrhythmias and can be achieved by vagal nerve stimulation, tragus stimulation, renal denervation, spinal cord stimulation, baroreceptor stimulation or left cardiac sympathetic denervation (Stavrakis et al., 2020a). Even though treatment of cardiac arrhythmias through automatic modulation has proven results preclinically, large randomized clinical trials have shown mixed results (Premchand et al., 2014; Zannad et al., 2015; Gold et al., 2016; Stavrakis et al., 2020b), highlighting the need for optimized biomarkers enabling appropriate patient selection. Yet, recent studies have demonstrated that vagal stimulation can modulate cardiac electrophysiological properties (Kulkarni et al., 2018b; Lee et al., 2018) including both atrial and ventricular alternans (Kulkarni et al., 2021b,c) as well as reduce AF burden (Qin and Singh, 2020; Stavrakis et al., 2020b).

Management of Ventricular Arrhythmias

Prediction Modalities

Traditional Clinical Methods

Various ECG ventricular repolarization markers (Tse and Yan, 2017) such as QT interval, the time interval from the start of the QRS complex to the end of the T wave (Castro-Torres et al., 2015), QTc, corrected QT interval normalized to the underlying heart rate, QTd, the dispersion of the QT interval, Tp-e, the time duration from the peak to the end of the T wave (Gürdal et al., 2017), Tp-ed, the dispersion of the Tp-e interval (Castro-Torres et al., 2015) and Tp-e/QT, the ratio between the Tp-e and QT intervals, have been used to predict the prevalence of cardiac arrhythmias. QTd is commonly used to denote the variability of ventricular repolarization. It has been proposed that the higher the QTd, the higher the risk for an individual to be predisposed to ventricular arrhythmias (Castro-Torres et al., 2015). Similarly, the Tp-e interval marks the transmural cardiac repolarization and is an important risk indicator for ventricular arrhythmias (Castro-Torres et al., 2015; Tse et al., 2017). It has been shown to be a more accurate marker to predict cardiac arrhythmias than QTc or QTd (Castro-Torres et al., 2015). Using the Tp-e marker, individuals with Brugada syndrome were found to be at a greater risk for ventricular arrhythmias than the healthy control group (Castro-Torres et al., 2015; Maury et al., 2015), while a meta-analysis of 155,856 patients demonstrated that prolonged Tp-e interval was a significant predictor of arrhythmic or mortality outcomes (Tse et al., 2017). The ratio between Tp-e and QT is a novel method to detect cardiac arrhythmias and is suggested to have greater accuracy than other markers as it has very few to no variations between 60 and 100 beats/min and hence, relatively independent of heart rate (Castro-Torres et al., 2015). Studies have shown that reduced coronary flow is positively correlated to the prolonged Tp-e interval and Tp-e/QT ratio (Zehir et al., 2015). Furthermore, prolonged Tp–e interval and increased Tp-e/QTc ratio have been shown to be independently associated with ventricular arrhythmic events in hypertrophic cardiomyopathy patients (Akboğa et al., 2017). Yet, while ventricular repolarization markers have been found to be useful in predicting malignant cardiac arrhythmias in many medical conditions, comprehensive investigative studies focused on their interpretation, reproducibility, robustness against interpatient variability and clinical utility are still warranted (Castro-Torres et al., 2015; Tse and Yan, 2017).

Another common clinical arrhythmia diagnostic method are stress tests, typically performed on individuals whose arrhythmias are adrenergically-driven and thus present or worsen with exercise (Clinic, 2018). Exercise is a known arrhythmogenic that may trigger ventricular premature beats, VT, VF or extrasystoles in patients with coronary heart disease (Jelinek and Lown, 1974). Similarly, they are an extremely important modality to detect occult arrhythmias (Jelinek and Lown, 1974). However, there are many limitations that make stress tests an obsolete testing modality (Jelinek and Lown, 1974). Many studies have shown that stress tests may present false positive results in asymptomatic patients (Coplan and Fuster, 1990). Furthermore, individuals with coronary artery disease are less likely to complete the stress test due to stress angina, thus making it challenging to detect the arrhythmia (Jelinek and Lown, 1974) and limiting the use of stress tests in clinical diagnosis (Coplan and Fuster, 1990). EP study, is a minimally invasive procedure that uses catheters to record the electrical activity of the heart (Stouffer et al., 2001) and is sometimes used to stratify arrhythmic risk and propensity for SCD, especially in patients with a left ventricular ejection fraction greater or equal to 35% (Hilfiker et al., 2015). More recently, non-invasive cardiac mapping has been shown to be useful in identifying sources of premature ventricular complexes and VT (Dubois et al., 2015) and successfully guiding catheter ablation, reducing the need for invasive mapping (Cakulev et al., 2013; Jamil-Copley et al., 2014).

Novel Non-linear Dynamical Methods

Non-linear dynamical approaches have long been utilized for predicting the onset of cardiac arrhythmias and the transition from steady state to abnormal electrical activity. Wenckebach phenomenon and parasystole were the first couple of arrhythmias where non-linear dynamics were used for systematic investigation (Krogh-Madsen and Christini, 2012). Recently, using preclinical studies, it has been shown that when the heart is perturbed from its steady state by electrical stimulation, only a small number of dominant eigenvalues are associated with it and principal component analysis enables evaluating the dominant eigenvalues for prediction of alternans (Jackson, 2005; Petrie and Zhao, 2012), a beat-to-beat alternation in the action potential duration of cardiac myocytes which manifests as oscillations in the T-wave morphology or duration on the ECG. During alternans, the heart transitions from a 1:1 stable rhythm to a 2:2 alternating pattern by means of a period doubling bifurcation, which occurs when a negative eigenvalue approaches – 1 (Strogatz, 1994; Li and Otani, 2003; Otani et al., 2005). Using ex vivo whole heart experiments, it was demonstrated that the onset of localized spatial alternans can be successfully predicted by applying principal component analysis to 2-dimentional cardiac optical mapping data (Kakade et al., 2013). Furthermore, the application of classical non-linear dynamical theory to investigate the bifurcation to abnormal cardiac rhythms, has enabled the classification of the transition from steady state to alternans. Preclinical optical mapping experiments on isolated whole rabbit hearts has demonstrated that the onset of alternans predominantly occurs through a smooth bifurcation (Kulkarni et al., 2015), which provides insights into a systemic understanding of cardiac dynamics and offers potential modalities for the prevention and control of the cardiac system based on non-linear approaches.

Another important biomarker that has shown to aid detection and prediction of VT/VF is heart rate variability (HRV), an inherent variation in the time interval between consecutive heart beats. Clinically, low HRV is a prognostic marker of an underlying cardiac functional abnormality (Jeong et al., 2021). Application of non-linear systems theory based methods to HRV markers have provided additional prognostic information compared to traditional HRV measures (Huikuri et al., 2003). In a novel study combining the benefits of AI with clinical HRV measures for cardiac arrhythmia diagnosis, Jeong et al. (2021) predicted the occurrence of VF with high accuracies of 88.18% and 88.68% using an artificial neural network which was trained using features from HRV signals with optimal data lengths of 10 s and 20 s and the forecast time period being 0 s.

Another study by Taye et al. (2019) compared the accuracy of applying DL techniques and predicting cardiac arrhythmias using HRV vs. QRS complex features. They trained two artificial neural networks and tested one each with features derived from either HRV measures or the QRS complex and found the former had 72% prediction accuracy while the latter had 98.6%, hence highlighting the QRS complex’s important role in accuracy improvement of VF prediction. They were able to predict onset of VF 30 s before occurrence using features extracted from 120 s long ECG and HRV signals. Prediction models dependent on QRS complexes are based on the principle of reentrant circuits. Electrical activity in the heart starts to change when the ventricles begin making reentrant waves due to ectopic focus hence changing the QRS complex and prompting the hypothesis that QRS shape features alone could be efficient VF predictors (Taye et al., 2019). In alignment with this theory, Bertsimas et al. (2021) designed a prediction system based on the QRS complex with the potential to be integrated in wearables that could predict cardiac arrhythmias and distinguish between AF, tachycardia, bradycardia or noise within 30 ms, from an ECG signal shorter than a minute.

Prevention and Control of Ventricular Arrhythmias

Traditional Clinical Methods

Overactivity of the sympathetic nervous system has been associated to ventricular tachyarrhythmias and sudden death (Singh, 2005). Beta-blockers can effectively suppress ventricular ectopic beats and arrhythmias, can be used to control rapid ventricular activation due to rapid and irregular atrial firing during AF as well as to prevent SCD in a wide range of cardiac diseases (Haverkamp et al., 1990; Poole-Wilson et al., 2003; Grandi and Ripplinger, 2019). Beta-blockers, as a class of compounds, have the essential pharmacologic function of counteracting the actions of competitive adrenergic receptors (Singh, 2005; Grandi and Ripplinger, 2019). The net clinical effects of various beta-receptor blockers, however, may differ quantitatively due to differences in related intrinsic sympathomimetic agonism and intrinsic potency for binding to beta-receptors (Grandi and Ripplinger, 2019). Evidence supports the idea that the antiarrhythmic effects of certain beta-receptor blockers, such as Carvedilol and Metoprolol, extend beyond the ventricular tissue to encompass atrial cells and help maintain sinus rhythm in patients with AF, particularly when combined with potent antifibrillatory agents like Amiodarone (Haverkamp et al., 1990). Side effects are typically dose-related; optimal and lowest effective beta-blocker doses are critical in limiting unwanted effects (Haverkamp et al., 1990). A larger use of beta-blockers as antiarrhythmic medicines in the future, possibly in combination with class I or III therapy, is feasible.

However, with the evolving advancements, the management of cardiac arrhythmias has been changing significantly over the years. ICDs are designed to monitor the heart constantly, identify malignant ventricular tachyarrhythmias and deliver anti-tachycardia pacing and/or an electrical countershock to restore normal rhythm (Mirowski, 1985). A meta-analysis study conducted showed that defibrillators when used as a primary or secondary prevention strategy is highly effective (Lee et al., 2003). The study demonstrated a relative risk reduction of 50% and 45% for arrhythmic death among the secondary and primary prevention trials, respectively. In addition ICDs have been shown to successfully terminate lethal arrhythmias and enable substantial improvement in survival in high-risk patients (Mirowski, 1985). Furthermore, ICDs have enabled a 20% relative risk reduction in all-cause mortality and a 33% reduction in arrhythmic mortality in comparison to pharmacological treatment with Amiodarone (Connolly et al., 2000).

Low energy shocks have been proposed to be painless and shown efficacy in terminating ventricular tachyarrhythmias in large animal models (Jackman and Zipes, 1982; Moreno et al., 2021). Furthermore, it has been postulated that autodecremental overdrive pacing and low-energy cardioversion have similar efficacy and acceleration rates (Bardy et al., 1993). In this study comparing anti-tachycardia pacing with low-energy cardioversion in 24 patients, VT was repeatedly terminated in 18 out of 24 patients by low-energy cardioversion, while 5 out of 24 patients saw VT accelerated to faster VT or VF, overall demonstrating similar response as with over drive pacing (Bardy et al., 1993). However, more recent human studies have shown low-energy shocks to be less effective than and more arrhythmogenic than high energy shocks (Strik et al., 2020), warranting further investigation of the clinical utility of this approach.

For non-ischemic VTs, it has been evidenced that catheter ablation is a reasonable first-line therapy as there has been a recorded 80–90% success rate in an experienced clinical center (Hall and Todd, 2006). With the advent of the modern non-fluoroscopic mapping system, catheter ablation is now possible in patients with or without structural heart disease. With the use of catheter ablation becoming widely available, a study demonstrated a curable rate of 90% for atrial flutter patients with severe symptoms (Hall and Todd, 2006). Radiofrequency ablation has evolved as a treatment of choice for idiopathic VT arising from the right ventricular outflow tract, demonstrating success rates between 75% and 100%, and for idiopathic left ventricular (fascicular) VT success varies from 50% to 90% (Nakagawa et al., 1993; Wellens and Smeets, 1993; O’donnell et al., 2003).

Another treatment modality for VT suppression is ventricular overdrive pacing, wherein stimuli are applied from the right ventricular apex during tachycardia at a rate that is little faster than the tachycardia cycle length until the paced QRS stabilizes (Kneller, 2015). The long-term overdrive suppression of refractory ventricular arrhythmias is reported to be frequently successful, more so if the heart rates are slow, but also in patients with normal heart rates (Johnson et al., 1974).

Novel Pacing Modalities

Besides the traditional methods of arrhythmia prevention, more recently, novel pacing techniques have reported promising anti-arrhythmic results in in silico and preclinical experimental studies (Table 2). Given the intrinsic variability in cardiac beat-to-beat intervals (HRV), many studies have investigated the effect of incorporating stochasticity in pacing on cardiac arrhythmogenesis (Dvir and Zlochiver, 2013; Kulkarni et al., 2016; Prudat et al., 2016; Wilson and Ermentrout, 2017). Cardiac alternans, which is suggested to be a precursor to fatal cardiac arrhythmias such as VT and VF, is a complex dynamical phenomenon originating at the cellular level as a result of instabilities in ion current dynamics (voltage-driven alternans) or intracellular calcium [Ca(2+)] cycling (Ca(2+)-driven alternans) (Prudat et al., 2016; Kulkarni et al., 2019). In a recent study, Prudat et al. (2016) demonstrated the utility of stochastic pacing in discriminating between voltage-gated and Ca(2+)-driven alternans, providing insights into the onset of cardiac rhythm instabilities. The influence of stochastic pacing on cardiac arrhythmogenicity was further illustrated by Dvir and Zlochiver (2013) using one-dimensional and two-dimensional ventricular models. They demonstrated that increased stochasticity in pacing promoted anti-arrhythmic effects and eliminated discordant action potential duration alternans (Dvir and Zlochiver, 2013). Their results suggest that the stabilization of cardiac electrical conduction can be improved by adding specific stochasticity to the programmed pacing sequence which inhibits the initiation of the arrhythmic events instigated by conduction blocks or wave breaks (Dvir and Zlochiver, 2013).

TABLE 2
www.frontiersin.org

Table 2. Prevention and control of cardiac arrhythmias: summary of novel pacing techniques for arrhythmia suppression.

Similar anti-arrhythmic effects of stochastic pacing were demonstrated by Wilson and Ermentrout (2017) wherein the formation of discordant alternans was inhibited in both one and two dimensional models. However, contradictory effects of stochastic pacing were reported in other studies wherein the onset of alternans was still present despite incorporating variability in pacing intervals (McIntyre et al., 2014; Kulkarni et al., 2016, 2018a). These recent studies suggested that pacing, with or without stochasticity can be pro-arrhythmic under certain conditions especially when the mean duration between consecutive stimuli is maintained constant, namely during periodic pacing.

A novel pacing technique that eliminated the constraints of periodic pacing by controlling the diastolic interval, was investigated by Jordan and Christini (2004). Their adaptive diastolic interval pacing technique demonstrated control of alternans in a one dimensional Purkinje fiber in silico study, but exhibited limited spatial control (Jordan and Christini, 2004). However, more recently the anti-arrhythmic effects of maintaining a constant diastolic interval on an every beat basis was demonstrated using ex vivo optical mapping experiments on isolated whole rabbit hearts (Kulkarni et al., 2018a). Kulkarni et al. (2018a) developed a novel closed loop system capable of detecting T-waves from ECG in real-time, enabling the successful implementation of constant diastolic interval pacing, which was shown to be anti-arrhythmic and prevented the spatio-temporal onset of alternans in the heart. While the utility of controlling the diastolic interval on a beat-by-beat basis in preventing voltage driven restitution dependent alternans has been effectively demonstrated, its efficacy in suppressing arrhythmias originating from metabolic and Ca(2+) cycling instabilities remains to be investigated (Kulkarni et al., 2018a,2021d).

Another anti-arrhythmic pacing technique, namely the alternating-period-feedback protocol, has been proposed which demonstrably reduced the presence of alternans in isolated whole heart experiments (Sridhar et al., 2013). The protocol aimed at suppressing alternans by applying small perturbations to the pacing cycle length and was proposed to be more robust than previous alternans reduction techniques offering promising implications for experimental investigations into its clinical utility for arrhythmia prevention.

To treat patients with ventricular dysfunction, biventricular pacing (BV) has been an effective modality even though electrophysiologic effects of BV pacing remain poorly understood (Zagrodzky et al., 2001). A study by Zagrodzky et al. (2001) assessed the effect of BV pacing on the inducibility of sustained monomorphic VT in patients with coronary artery disease and demonstrated a 60% reduction in inducibility with BV pacing as compared to single-site right ventricular pacing. Similar findings were reported by other studies over the years supporting the hypothesis that BV pacing reduces the frequency of ventricular arrhythmias, however, the mechanism of arrhythmia suppression remains unclear (Kowal et al., 2004).

Dynamic Control of Arrhythmias

As the development of arrhythmias is a dynamic instability shift from sinus rhythm, many dynamical approaches are being investigated to control these instabilities and prevent the occurrence of arrhythmias. Dynamic R-wave triggered pacing during the absolute refractory period has been shown to suppress spontaneous repolarization alternans, a pattern of ventricular repolarization that repeats on every other beat basis, which has been seen to be closely associated with VT/VF (Armoundas et al., 2013; Merchant et al., 2020). This novel pacing modality has been shown to suppress cardiac alternans in vivo and under diseased conditions by delivering subthreshold stimuli during the absolute refractory period to modulate ventricular repolarization, thus offering a potential technique to improve the current ICDs by preventing the onset of ventricular arrhythmias. Cardiac alternans has also been shown to be controlled by mechanical perturbation, which can alter the electrical activity of tissue by mechano-electric coupling, in in silico studies (Dubljevic and Christofides, 2008; Hazim et al., 2015).

Chaos control based non-linear dynamical approaches have shown promising results in controlling arrhythmias in in vitro and in silico studies (Christini et al., 1999, 2001; Kulkarni and Tolkacheva, 2021). Christini et al. (1999, 2001) demonstrated the successful termination of pacing-induced period-2 atrioventricular-nodal conduction alternans using non-linear dynamical control by stabilizing the underlying unstable steady-state conduction in 52/54 control attempts in five patients. Chaos control techniques have been shown to modulate human cardiac electrophysiological dynamics and successfully prevent cardiac arrhythmias by stabilizing unstable rhythms through small time continuous perturbations aimed at restoring the heart to a steady state, which has potential to be incorporated into pacemakers (Garfinkel et al., 1992, 1995; Christini et al., 2001; Ferreira et al., 2011).

Future Perspectives: Bridging the Gap Between Basic and Clinical Science

As we embark into 60 years of cardiac pacing, the field has seen significant technological advancements and continued progress. Cardiac stimulation strategies have registered significant benefits with reductions in HF hospitalization, AF and mortality. However, as summarized in this review, there are several shortcomings in the existing treatment strategies and many current pacing devices face challenges of false positives, suboptimal accuracy of predicting the onset of abnormal rhythms and low efficacy in suppressing ongoing arrhythmic events. Novel pacing strategies that are under research and development although show promise, there exist several challenges to bring these technologies into clinical practice that warrants further research for testing and validating the safety and efficacy of these newer modalities.

Innovative approaches to predict the onset of cardiac arrhythmias as well as control abnormal rhythms offer huge promise in reducing the arrhythmic burden in preclinical and in silico experiments. Yet focused studies investigating their robustness and accuracy are required that can show the pathway toward clinical translation. Firstly, non-linear dynamical approaches and AI based strategies offer computational challenges in pacing devices which directly demands longer battery life. Limited battery life, warranting device replacements, lead extractions and related complications have challenged the pacing field significantly. Researchers should be wary of this important limitation and consider concurrent investigations of both cardiac pacing and battery technology together. Battery-less, solar-powered cardiac pacing offers significant promise, however, it has fundamental challenges to be addressed in terms of climate, energy harvesting and early warning of low battery levels.

Secondly, the advent of AI in cardiac dynamics is showing a steady increase in both prediction and control strategies of cardiac pacing. However, to date, there are no ML models that have been highly robust and accurate when dealing with interpatient and inter-institutional variability, hence questioning their translatability into a usable universal diagnostic tool. Several researchers have used millions of ECG datasets to develop prediction models, which are mostly black box applications with no novel insights into the physiological mechanisms of arrhythmia etiology that significantly hamper clinical translation. The gap between basic and clinical science is inadvertently increasing due to this approach and more interpretable ML models with physiological significance are required to bridge this gap. Therefore, a strong synergistic effort between basic scientists for developing and validating novel pacing strategies, cardiologists as well as pacemaker design industries is imperative to advance this important clinical need that can play a crucial role in cardiac patient healthcare.

Conclusion

Novel pacing modalities for control and suppression of cardiac arrhythmias have shown promising results in translational preclinical studies. Applying non-linear dynamical approaches combined with advances in AI has opened the doors for investigating volumes of patient data and extracting relevant information for aiding clinical diagnostics and decision making. Predicting the onset of arrhythmias and the transition from steady to unstable cardiac rhythms has been a long standing challenge which newer non-linear methods have shown promising efficacy in addressing. However, the clinical adaptation of AI and dynamical pacing strategies would require a comprehensive framework that accounts for a reliable, accountable and secure diagnostic system, encouraging the active participation by researchers, engineers, technicians and physicians alike. While the novel pacing modalities and predictive algorithms discussed here are a promising first step for advancing cardiac arrhythmia management strategies, a thorough clinical investigation and physician assessment are necessary for the successful translation of these techniques into a diagnostic approach. Such an integrative approach combining medicine and engineering principles holds promise in pushing traditional boundaries and translating preclinical findings into clinically viable devices for cardiac arrhythmia prediction, prevention and control.

Author Contributions

All authors listed have made a substantial, direct, and intellectual contribution to the work, and approved it for publication.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Akboğa, M. K., Balcı, K. G., Yılmaz, S., Aydın, S., Yayla, Ç, Ertem, A. G., et al. (2017). Tp-e interval and tp-e/qtc ratio as novel surrogate markers for prediction of ventricular arrhythmic events in hypertrophic cardiomyopathy. Anatolian J. Cardiol. 18:48. doi: 10.14744/AnatolJCardiol.2017.7581

PubMed Abstract | CrossRef Full Text | Google Scholar

Alonso, A., Krijthe, B. P., Aspelund, T., Stepas, K. A., Pencina, M. J., Moser, C. B., et al. (2013). Simple risk model predicts incidence of atrial fibrillation in a racially and geographically diverse population: the charge-af consortium. J. Am. Heart Assoc. 2:e000102. doi: 10.1161/JAHA.112.000102

PubMed Abstract | CrossRef Full Text | Google Scholar

Armoundas, A. A., Weiss, E. H., Sayadi, O., Laferriere, S., Sajja, N., Mela, T., et al. (2013). A novel pacing method to suppress repolarization alternans in vivo: implications for arrhythmia prevention. Heart Rhythm. 10, 564–572.

Google Scholar

Arora, V., and Suri, P. (2021). Physiological pacing: a new road to future. Indian J. Clin. Cardiol. 2, 32–43. doi: 10.1093/milmed/usy171

PubMed Abstract | CrossRef Full Text | Google Scholar

Attuel, P., Danilovic, D., Konz, K. H., Brachmann, J., Allaf, D. E., Löscher, S., et al. (2003). Relationship between selected overdrive parameters and the therapeutic outcome and tolerance of atrial overdrive pacing. Pacing Clin. Electrophysiol. 26, 257–263. doi: 10.1046/j.1460-9592.2003.00028.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Auer, J., Berent, R., and Eber, B. (2002). Amiodarone in the prevention and treatment of arrhythmia. Curr. Opin. Investig. Drugs 3, 1037–1044.

Google Scholar

Bajbouj, M., Merkl, A., Schlaepfer, T. E., Frick, C., Zobel, A., Maier, W., et al. (2010). Two-year outcome of vagus nerve stimulation in treatment-resistant depression. J. Clin. Psychopharmacol. 30, 273–281.

Google Scholar

Bardy, G. H., Poole, J. E., Kudenchuk, P. J., Dolack, G. L., Kelso, D., and Mitchell, R. (1993). A prospective randomized repeat-crossover comparison of antitachycardia pacing with low-energy cardioversion. Circulation 87, 1889–1896. doi: 10.1161/01.cir.87.6.1889

CrossRef Full Text | Google Scholar

Bazoukis, G., Stavrakis, S., Zhou, J., Bollepalli, S. C., Tse, G., Zhang, Q., et al. (2021). Machine learning versus conventional clinical methods in guiding management of heart failure patients—a systematic review. Heart Fail. Rev. 26, 23–34.

Google Scholar

Benarroch, E. E. (2008). The arterial baroreflex: functional organization and involvement in neurologic disease. Neurology 71, 1733–1738. doi: 10.1212/01.wnl.0000335246.93495.92

PubMed Abstract | CrossRef Full Text | Google Scholar

Ben-Menachem, E. (2002). Vagus-nerve stimulation for the treatment of epilepsy. Neurology 1, 477–482.

Google Scholar

Bertsimas, D., Mingardi, L., and Stellato, B. (2021). Machine learning for real-time heart disease prediction. IEEE J. Biomed. Health Inform. 25, 3627–3637. doi: 10.1109/jbhi.2021.3066347

PubMed Abstract | CrossRef Full Text | Google Scholar

Bhavnani, S. P., Kluger, J., Coleman, C. I., White, C. M., Guertin, D., Shafi, N. A., et al. (2010). The prognostic impact of shocks for clinical and induced arrhythmias on morbidity and mortality among patients with implantable cardioverter-defibrillators. Heart Rhythm. 7, 755–760. doi: 10.1016/j.hrthm.2010.02.039

PubMed Abstract | CrossRef Full Text | Google Scholar

Boon, K. H., Khalil-Hani, M., Malarvili, M., and Sia, C. W. (2016). Paroxysmal atrial fibrillation prediction method with shorter hrv sequences. Comput. Methods Prog. Biomed. 134, 187–196. doi: 10.1016/j.cmpb.2016.07.016

PubMed Abstract | CrossRef Full Text | Google Scholar

Bundy, J. D., Heckbert, S. R., Chen, L. Y., Lloyd-Jones, D. M., and Greenland, P. (2020). Evaluation of risk prediction models of atrial fibrillation (from the multi-ethnic study of atherosclerosis [mesa]). Am. J. Cardiol. 125, 55–62. doi: 10.1016/j.amjcard.2019.09.032

PubMed Abstract | CrossRef Full Text | Google Scholar

Cakulev, I., Sahadevan, J., Arruda, M., Goldstein, R. N., Hong, M., Intini, A., et al. (2013). Confirmation of novel noninvasive high-density electrocardiographic mapping with electrophysiology study: implications for therapy. Circulation 6, 68–75. doi: 10.1161/CIRCEP.112.975813

PubMed Abstract | CrossRef Full Text | Google Scholar

Camm, A., Sulke, N., Edvardsson, N., Ritter, P., Albers, B., Ruiter, J., et al. (2007). Conventional and dedicated atrial overdrive pacing for the prevention of paroxysmal atrial fibrillation: the aftherapy study. Europace 9, 1110–1118. doi: 10.1093/europace/eum253

PubMed Abstract | CrossRef Full Text | Google Scholar

Carlson, M. D., Ip, J., Messenger, J., Beau, S., Kalbfleisch, S., Gervais, P., et al. (2003). A new pacemaker algorithm for the treatment of atrial fibrillation: results of the atrial dynamic overdrive pacing trial (adopt). J. Am. College Cardiol. 42, 627–633.

Google Scholar

Castro-Torres, Y., Carmona-Puerta, R., and Katholi, R. E. (2015). Ventricular repolarization markers for predicting malignant arrhythmias in clinical practice. World J. Clin. Cases 3:705. doi: 10.12998/wjcc.v3.i8.705

PubMed Abstract | CrossRef Full Text | Google Scholar

Chen, T., and Guestrin, C. (2016). “Xgboost: a scalable tree boosting system,” in Proceedings of the 22nd Acm Sigkdd International Conference on Knowledge Discovery and Data Mining, (New York, NY: Association for Computing Machinery), 785–794.

Google Scholar

Christini, D. J., Stein, K. M., Markowitz, S. M., Mittal, S., Slotwiner, D., and Lerman, B. (1999). The role of nonlinear dynamics in cardiac arrhythmia control. Heart Dis. 1, 190–200.

Google Scholar

Christini, D. J., Stein, K. M., Markowitz, S. M., Mittal, S., Slotwiner, D. J., Scheiner, M. A., et al. (2001). Nonlinear-dynamical arrhythmia control in humans. Proc. Natl. Acad. Sci. U.S.A. 98, 5827–5832. doi: 10.1073/pnas.091553398

PubMed Abstract | CrossRef Full Text | Google Scholar

Chung, M. K., Eckhardt, L. L., Chen, L. Y., Ahmed, H. M., Gopinathannair, R., Joglar, J. A., et al. (2020). Lifestyle and risk factor modification for reduction of atrial fibrillation: a scientific statement from the american heart association. Circulation 141, e750–e772.

Google Scholar

Clinic, C. (2018). Arrhythmia. Treasure Island, FL: StatPearls.

Google Scholar

Clinic, C. (2019). Ambulatory Monitors. Rochester: Mayo Clinic.

Google Scholar

Clinic, M. (2021). Holter Monitor. Rochester: Mayo Clinic.

Google Scholar

Connolly, S. J., Gent, M., Roberts, R. S., Dorian, P., Roy, D., Sheldon, R. S., et al. (2000). Canadian implantable defibrillator study (cids) a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation 101, 1297–1302.

Google Scholar

Coplan, N. L., and Fuster, V. (1990). Limitations of the exercise test as a screen for acute cardiac events in asymptomatic patients. Am. Heart J. 119, 987–990. doi: 10.1016/s0002-8703(05)80352-7

CrossRef Full Text | Google Scholar

Costin, H., Rotariu, C., and Pãsãricã, A. (2013). “Atrial fibrillation onset prediction using variability of ecg signals,” in Proceedings of the 2013 8th International symposium on advanced topics in electrical engineering (ATEE), Bucharest, 1–4.

Google Scholar

De Asmundis, C., Conte, G., Sieira, J., Chierchia, G.-B., Rodriguez-Manero, M., Di Giovanni, G., et al. (2014). Comparison of the patient-activated event recording system vs. Traditional 24 h holter electrocardiography in individuals with paroxysmal palpitations or dizziness. Europace 16, 1231–1235. doi: 10.1093/europace/eut411

PubMed Abstract | CrossRef Full Text | Google Scholar

de Voogt, W., van Hemel, N., de Vusser, P., Mairesse, G. H., van Mechelen, R., Koistinen, J., et al. (2007). No evidence of automatic atrial overdrive pacing efficacy on reduction of paroxysmal atrial fibrillation. Europace 9, 798–804. doi: 10.1093/europace/eum149

PubMed Abstract | CrossRef Full Text | Google Scholar

Desai, D. S., and Hajouli, S. (2021). Arrhythmias. Treasure Island, FL: StatPearls.

Google Scholar

Dimmer, C., Szili-Torok, T., Tavernier, R., Verstraten, T., and Jordaens, L. (2003). Initiating mechanisms of paroxysmal atrial fibrillation. Europace 5, 1–9. doi: 10.1053/eupc.2002.0273

PubMed Abstract | CrossRef Full Text | Google Scholar

Dubljevic, S., and Christofides, P. D. (2008). Optimal mechano-electric stabilization of cardiac alternans. Chem. Eng. Sci. 63, 5425–5433.

Google Scholar

Dubois, R., Shah, A. J., Hocini, M., Denis, A., Derval, N., Cochet, H., et al. (2015). Non-invasive cardiac mapping in clinical practice: application to the ablation of cardiac arrhythmias. J. Electrocardiol. 48, 966–974. doi: 10.1016/j.jelectrocard.2015.08.028

PubMed Abstract | CrossRef Full Text | Google Scholar

Dvir, H., and Zlochiver, S. (2013). Stochastic cardiac pacing increases ventricular electrical stability—a computational study. Biophys. J. 105, 533–542. doi: 10.1016/j.bpj.2013.06.012

PubMed Abstract | CrossRef Full Text | Google Scholar

Ferreira, B. B., de Paula, A. S., and Savi, M. A. (2011). Chaos control applied to heart rhythm dynamics. Chaos Solitons Fractals 44, 587–599.

Google Scholar

Florea, V. G., and Cohn, J. N. (2014). The autonomic nervous system and heart failure. Circulation Res. 114, 1815–1826.

Google Scholar

Florek, J. B., and Girzadas, D. (2018). Amiodarone. Treasure Island, FL: StatPearls.

Google Scholar

García-Civera, R., Ruiz-Granell, R., Morell-Cabedo, S., Sanjuan-Mañez, R., Ferrero, A., Martínez-Brotons, A., et al. (2005). Significance of tilt table testing in patients with suspected arrhythmic syncope and negative electrophysiologic study. J. Cardiovasc. Electrophysiol. 16, 938–942. doi: 10.1111/j.1540-8167.2005.50029.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Garfinkel, A., Spano, M. L., Ditto, W. L., and Weiss, J. N. (1992). Controlling cardiac chaos. Science 257, 1230–1235. doi: 10.1126/science.1519060

PubMed Abstract | CrossRef Full Text | Google Scholar

Garfinkel, A., Weiss, J. N., Ditto, W. L., and Spano, M. L. (1995). Chaos control of cardiac arrhythmias. Trends Cardiovasc. Med. 5, 76–80. doi: 10.1016/1050-1738(94)00083-2

CrossRef Full Text | Google Scholar

Germano, J. J., Reynolds, M., Essebag, V., and Josephson, M. E. (2006). Frequency and causes of implantable cardioverter-defibrillator therapies: is device therapy proarrhythmic? Am. J. Cardiol. 97, 1255–1261. doi: 10.1016/j.amjcard.2005.11.048

PubMed Abstract | CrossRef Full Text | Google Scholar

Gold, M. R., Van Veldhuisen, D. J., Hauptman, P. J., Borggrefe, M., Kubo, S. H., Lieberman, R. A., et al. (2016). Vagus nerve stimulation for the treatment of heart failure: the inovate-hf trial. J. Am. Coll. Cardiol. 68, 149–158.

Google Scholar

Goldberger, J. J., Arora, R., Buckley, U., and Shivkumar, K. (2019). Autonomic nervous system dysfunction: Jacc focus seminar. J. Am. College Cardiol. 73, 1189–1206. doi: 10.1016/j.jacc.2018.12.064

PubMed Abstract | CrossRef Full Text | Google Scholar

Goldsmith, S. R. (2004). Interactions between the sympathetic nervous system and the raas in heart failure. Curr. Heart Fail. Rep. 1, 45–50.

Google Scholar

Golińska, A. K., Lesiński, W., Przybylski, A., and Rudnicki, W. R. (2020). Towards prediction of heart arrhythmia onset using machine learning. Int. Conf. Comput. Sci. 12140, 376–389. doi: 10.1016/j.cmpb.2018.07.014

PubMed Abstract | CrossRef Full Text | Google Scholar

Grandi, E., and Ripplinger, C. M. (2019). Antiarrhythmic mechanisms of beta blocker therapy. Pharmacol. Res. 146:104274. doi: 10.1016/j.phrs.2019.104274

PubMed Abstract | CrossRef Full Text | Google Scholar

Gürdal, A., Eroğlu, H., Helvaci, F., Sümerkan, M. Ç, Kasali, K., Çetin, Ş, et al. (2017). Evaluation of tp-e interval, tp-e/qt ratio and tp-e/qtc ratio in patients with subclinical hypothyroidism. Ther. Adv. Endocrinol. Metab. 8, 25–32.

Google Scholar

Hall, M., and Todd, D. (2006). Modern management of arrhythmias. Postgraduate Med. J. 82, 117–125. doi: 10.1136/pgmj.2005.033654

PubMed Abstract | CrossRef Full Text | Google Scholar

Haverkamp, W., Hindricks, G., and Gülker, H. (1990). Antiarrhythmic properties of beta-blockers. J. Cardiovasc. Pharmacol. 16, S29–S32.

Google Scholar

Hazim, A., Belhamadia, Y., and Dubljevic, S. (2015). Control of cardiac alternans in an electromechanical model of cardiac tissue. Comput. Biol. Med. 63, 108–117. doi: 10.1016/j.compbiomed.2015.05.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Heo, J., Yoon, J. G., Park, H., Kim, Y. D., Nam, H. S., and Heo, J. H. (2019). Machine learning–based model for prediction of outcomes in acute stroke. Stroke 50, 1263–1265. doi: 10.1161/strokeaha.118.024293

PubMed Abstract | CrossRef Full Text | Google Scholar

Hermosillo, A., Falcon, J., Marquez, M., Arteaga, D., and Cardenas, M. (1999). Positive head-up tilt table test in patients with the long qt syndrome. Europace 1, 213–217. doi: 10.1053/eupc.1999.0056

PubMed Abstract | CrossRef Full Text | Google Scholar

Hilfiker, G., Schoenenberger, A. W., Erne, P., and Kobza, R. (2015). Utility of electrophysiological studies to predict arrhythmic events. World J. Cardiol. 7:344. doi: 10.4330/wjc.v7.i6.344

PubMed Abstract | CrossRef Full Text | Google Scholar

Hill, N. R., Sandler, B., Mokgokong, R., Lister, S., Ward, T., Boyce, R., et al. (2020). Cost-effectiveness of targeted screening for the identification of patients with atrial fibrillation: evaluation of a machine learning risk prediction algorithm. J. Med. Econ. 23, 386–393. doi: 10.1080/13696998.2019.1706543

PubMed Abstract | CrossRef Full Text | Google Scholar

Hindricks, G., Potpara, T., Dagres, N., Arbelo, E., Bax, J. J., Blomström-Lundqvist, C., et al. (2021). 2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the european association for cardio-thoracic surgery (eacts) the task force for the diagnosis and management of atrial fibrillation of the european society of cardiology (esc) developed with the special contribution of the european heart rhythm association (ehra) of the esc. Eur. Heart J. 42, 373–498.

Google Scholar

Hoffmann, E., Janko, S., Steinbeck, G., Edvardsson, N., and Camm, J. (2000). Analysis of onset mechanisms of atrial fibrillation in pacemaker patients. Heart 83, 656–669.

Google Scholar

Hohnloser, S. H., Healey, J. S., Gold, M. R., Israel, C. W., Yang, S., van Gelder, I., et al. (2012). Atrial overdrive pacing to prevent atrial fibrillation: insights from assert. Heart Rhythm 9, 1667–1673. doi: 10.1016/j.hrthm.2012.06.012

PubMed Abstract | CrossRef Full Text | Google Scholar

Huikuri, H. V., Mäkikallio, T. H., and Perkiömäki, J. (2003). Measurement of heart rate variability by methods based on nonlinear dynamics. J. Electrocardiol. 36, 95–99.

Google Scholar

Jackman, W. M., and Zipes, D. P. (1982). Low-energy synchronous cardioversion of ventricular tachycardia using a catheter electrode in a canine model of subacute myocardial infarction. Circulation 66, 187–195. doi: 10.1161/01.cir.66.1.187

CrossRef Full Text | Google Scholar

Jackson, J. E. (2005). A User’s Guide to Principal Components. Hoboken, NJ: John Wiley & Sons.

Google Scholar

Jamil-Copley, S., Bokan, R., Kojodjojo, P., Qureshi, N., Koa-Wing, M., Hayat, S., et al. (2014). Noninvasive electrocardiographic mapping to guide ablation of outflow tract ventricular arrhythmias. Heart Rhythm. 11, 587–594. doi: 10.1016/j.hrthm.2014.01.013

PubMed Abstract | CrossRef Full Text | Google Scholar

Jelinek, M. V., and Lown, B. (1974). Exercise stress testing for exposure of cardiac arrhythmia. Prog. Cardiovasc. Dis. 16, 497–522.

Google Scholar

Jeong, D. U., Taye, G. T., Hwang, H.-J., and Lim, K. M. (2021). Optimal length of heart rate variability data and forecasting time for ventricular fibrillation prediction using machine learning. Comput. Math. Methods Med. 2021:6663996.

Google Scholar

Jin, Z., Sun, Y., and Cheng, A. C. (2009). “Predicting cardiovascular disease from real-time electrocardiographic monitoring: an adaptive machine learning approach on a cell phone,” in Proceedings of the 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, (Piscataway, NJ: IEEE), 6889–6892. doi: 10.1109/IEMBS.2009.5333610

PubMed Abstract | CrossRef Full Text | Google Scholar

John, R. M., Tedrow, U. B., Koplan, B. A., Albert, C. M., Epstein, L. M., Sweeney, M. O., et al. (2012). Ventricular arrhythmias and sudden cardiac death. Lancet 380, 1520–1529.

Google Scholar

Johnson, R. A., Hutter, J. R. A. M., Desanctis, R. W., Yurchak, P. M., Leinbach, R. C., and Harthorne, J. W. (1974). Chronic overdrive pacing in the control of refractory ventricular arrhythmias. Ann. Intern. Med. 80, 380–383. doi: 10.7326/0003-4819-80-3-380

PubMed Abstract | CrossRef Full Text | Google Scholar

Jordan, P. N., and Christini, D. J. (2004). Adaptive diastolic interval control of cardiac action potential duration alternans. J. Cardiovasc. Electrophysiol. 15, 1177–1185. doi: 10.1046/j.1540-8167.2004.04098.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Kakade, V., Zhao, X., and Tolkacheva, E. G. (2013). Using dominant eigenvalue analysis to predict formation of alternans in the heart. Phys. Rev. E 88:e052716.

Google Scholar

Khurshid, S., Choi, S. H., Weng, L.-C., Wang, E. Y., Trinquart, L., Benjamin, E. J., et al. (2018). Frequency of cardiac rhythm abnormalities in a half million adults. Circulation 11:e006273. doi: 10.1161/CIRCEP.118.006273

PubMed Abstract | CrossRef Full Text | Google Scholar

Kneller, J. (2015). Ventricular overdrive pacing during SVT: an opportunity for EP techs and allied professionals. EP Lab Digest 15.

Google Scholar

Kowal, R. C., Wasmund, S. L., Smith, M. L., Sharma, N., Carayannopoulos, G. N., Le, B., et al. (2004). Biventricular pacing reduces the induction of monomorphic ventricular tachycardia: a potential mechanism for arrhythmia suppression. Heart Rhythm. 1, 295–300. doi: 10.1016/j.hrthm.2004.05.008

PubMed Abstract | CrossRef Full Text | Google Scholar

Krittanawong, C., Virk, H. U. H., Bangalore, S., Wang, Z., Johnson, K. W., Pinotti, R., et al. (2020). Machine learning prediction in cardiovascular diseases: a meta-analysis. Sci. Rep. 10, 1–11.

Google Scholar

Krogh-Madsen, T., and Christini, D. J. (2012). Nonlinear dynamics in cardiology. Annu. Rev. Biomed. Eng. 14, 179–203.

Google Scholar

Kulkarni, K., Lee, S. W., Kluck, R., and Tolkacheva, E. G. (2018a). Real-time closed loop diastolic interval control prevents cardiac alternans in isolated whole rabbit hearts. Ann. Biomed. Eng. 46, 555–566. doi: 10.1007/s10439-018-1981-2

PubMed Abstract | CrossRef Full Text | Google Scholar

Kulkarni, K., Xie, X., Fernandez de Velasco, E. M., Anderson, A., Martemyanov, K. A., Wickman, K., et al. (2018b). The influences of the m2r-girk4-rgs6 dependent parasympathetic pathway on electrophysiological properties of the mouse heart. PLoS One 13:e0193798. doi: 10.1371/journal.pone.0193798

PubMed Abstract | CrossRef Full Text | Google Scholar

Kulkarni, K., Lee, S. W., and Tolkacheva, E. G. (2016). “Pro-arrhythmic effect of heart rate variability during periodic pacing,” in Proceedings of the 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Orlando, FL, 149–152. doi: 10.1109/EMBC.2016.7590662

PubMed Abstract | CrossRef Full Text | Google Scholar

Kulkarni, K., Merchant, F. M., Kassab, M. B., Sana, F., Moazzami, K., Sayadi, O., et al. (2019). Cardiac alternans: mechanisms and clinical utility in arrhythmia prevention. J. Am. Heart Assoc. 8:e013750.

Google Scholar

Kulkarni, K., Sevakula, R. K., Kassab, M. B., Nichols, J., Roberts, J. D., Isselbacher, E. M., et al. (2021a). Ambulatory monitoring promises equitable personalized healthcare delivery in underrepresented patients. Eur. Heart J. Digital Health 2, 494–510. doi: 10.1093/ehjdh/ztab047

PubMed Abstract | CrossRef Full Text | Google Scholar

Kulkarni, K., Singh, J. P., Parks, K. A., Katritsis, D. G., Stavrakis, S., and Armoundas, A. A. (2021b). Low-level tragus stimulation modulates atrial alternans and fibrillation burden in patients with paroxysmal atrial fibrillation. J. Am. Heart Assoc. 10:e020865. doi: 10.1161/JAHA.120.020865

PubMed Abstract | CrossRef Full Text | Google Scholar

Kulkarni, K., Stavrakis, S., Elkholey, K., Singh, J. P., Parks, K. A., and Armoundas, A. A. (2021c). Microvolt t-wave alternans is modulated by acute low-level tragus stimulation in patients with ischemic cardiomyopathy and heart failure. Front. Physiol. 12:707724. doi: 10.3389/fphys.2021.707724

PubMed Abstract | CrossRef Full Text | Google Scholar

Kulkarni, K., Walton, R. D., Armoundas, A. A., and Tolkacheva, E. G. (2021d). Clinical potential of beat-to-beat diastolic interval control in preventing cardiac arrhythmias. J. Am. Heart Assoc. 10:e020750.

Google Scholar

Kulkarni, K., and Tolkacheva, E. G. (2021). Nonlinear analytical approaches for prediction of alternans mediated cardiac arrhythmias. Highlights Med. Med. Sci. 5, 35–47. doi: 10.9734/bpi/hmms/v5/8498d

CrossRef Full Text | Google Scholar

Kulkarni, K., Visweswaran, R., Zhao, X., and Tolkacheva, E. G. (2015). Characterizing spatial dynamics of bifurcation to alternans in isolated whole rabbit hearts based on alternate pacing. Biomed. Res. Int. 2015:170768. doi: 10.1155/2015/170768

PubMed Abstract | CrossRef Full Text | Google Scholar

Kulkarni, N., Mody, P., and Levine, B. D. (2020). Abolish the tilt table test for the workup of syncope! Circulation 141, 335–337. doi: 10.1161/circulationaha.119.043259

PubMed Abstract | CrossRef Full Text | Google Scholar

Lai, D., Zhang, Y., Zhang, X., Su, Y., and Heyat, M. B. B. (2019). An automated strategy for early risk identification of sudden cardiac death by using machine learning approach on measurable arrhythmic risk markers. IEEE Access. 7, 94701–94716. doi: 10.1109/access.2019.2925847

CrossRef Full Text | Google Scholar

Lantz, R., and Agarwal, A. (2021). A review of amiodarone side effects in a cohort of patients: a quality improvement project. Chest 160:A1414. doi: 10.1161/CIRCOUTCOMES.116.003122

PubMed Abstract | CrossRef Full Text | Google Scholar

Lee, D. S., Green, L. D., Liu, P. P., Dorian, P., Newman, D. M., Grant, F. C., et al. (2003). Effectiveness of implantable defibrillators for preventing arrhythmic events and death: a meta-analysis. J. Am. College Cardiol. 41, 1573–1582. doi: 10.1016/s0735-1097(03)00253-5

CrossRef Full Text | Google Scholar

Lee, G., Sanders, P., and Kalman, J. M. (2012). Catheter ablation of atrial arrhythmias: state of the art. Lancet 380, 1509–1519. doi: 10.1016/s0140-6736(12)61463-9

CrossRef Full Text | Google Scholar

Lee, S. W., Kulkarni, K., Annoni, E. M., Libbus, I., KenKnight, B. H., and Tolkacheva, E. G. (2018). Stochastic vagus nerve stimulation affects acute heart rate dynamics in rats. PLoS One 13:e0194910. doi: 10.1371/journal.pone.0194910

PubMed Abstract | CrossRef Full Text | Google Scholar

Li, M., and Otani, N. F. (2003). Ion channel basis for alternans and memory in cardiac myocytes. Ann. Biomed. Eng. 31, 1213–1230. doi: 10.1114/1.1616930

CrossRef Full Text | Google Scholar

Lj, W., Cy, H., Mx, L., and Hy, L. (2010). Arrhythmia after a positive head-up tilt table test. Zhonghua Xin Xue Guan Bing Za Zhi 38, 805–808.

Google Scholar

Maury, P., Sacher, F., Gourraud, J.-B., Pasquié, J.-L., Raczka, F., Bongard, V., et al. (2015). Increased tpeak-tend interval is highly and independently related to arrhythmic events in brugada syndrome. Heart Rhythm. 12, 2469–2476. doi: 10.1016/j.hrthm.2015.07.029

PubMed Abstract | CrossRef Full Text | Google Scholar

McIntyre, S. D., Kakade, V., Mori, Y., and Tolkacheva, E. G. (2014). Heart rate variability and alternans formation in the heart: the role of feedback in cardiac dynamics. J. Theor. Biol. 350, 90–97. doi: 10.1016/j.jtbi.2014.02.015

PubMed Abstract | CrossRef Full Text | Google Scholar

Merchant, F. M., Sayadi, O., Sohn, K., Weiss, E. H., Puppala, D., Doddamani, R., et al. (2020). Real-time closed-loop suppression of repolarization alternans reduces arrhythmia susceptibility in vivo. Circ. Arrhythm. Electrophysiol. 13:e008186. doi: 10.1161/CIRCEP.119.008186

PubMed Abstract | CrossRef Full Text | Google Scholar

Mirowski, M. (1985). The automatic implantable cardioverter-defibrillator: an overview. J. Am. College Cardiol. 6, 461–466.

Google Scholar

Mitchell, A. R., and Sulke, N. (2004). How do atrial pacing algorithms prevent atrial arrhythmias? EP Europace 6, 351–362. doi: 10.1016/j.eupc.2004.03.005

PubMed Abstract | CrossRef Full Text | Google Scholar

Mohebbi, M., and Ghassemian, H. (2012). Prediction of paroxysmal atrial fibrillation based on non-linear analysis and spectrum and bispectrum features of the heart rate variability signal. Comput. Methods Prog. Biomed. 105, 40–49. doi: 10.1016/j.cmpb.2010.07.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Moreno, A., Walton, R. D., Bernus, O., Vigmond, E. J., and Bayer, J. D. (2021). Low-energy single-pulse surface stimulation defibrillates large mammalian ventricles. Heart Rhythm. [Epub ahead of print]. doi: 10.1016/j.hrthm.2021.10.006

PubMed Abstract | CrossRef Full Text | Google Scholar

Nakagawa, H., Beckman, K. J., McClelland, J. H., Wang, X., Arruda, M., Santoro, I., et al. (1993). Radiofrequency catheter ablation of idiopathic left ventricular tachycardia guided by a purkinje potential. Circulation 88, 2607–2617. doi: 10.1161/01.cir.88.6.2607

CrossRef Full Text | Google Scholar

Namarvar, H. H., and Shahidi, A. V. (2004). “Cardiac arrhythmias predictive detection methods with wavelet-svd analysis and support vector machines,” in Proceedings of the The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Lyon, 365–368. doi: 10.1109/IEMBS.2004.1403168

PubMed Abstract | CrossRef Full Text | Google Scholar

O’donnell, D., Cox, D., Bourke, J., Mitchell, L., and Furniss, S. (2003). Clinical and electrophysiological differences between patients with arrhythmogenic right ventricular dysplasia and right ventricular outflow tract tachycardia. Eur. Heart J. 24, 801–810. doi: 10.1016/s0195-668x(02)00654-1

CrossRef Full Text | Google Scholar

Ong, M. E. H., Ng, C. H. L., Goh, K., Liu, N., Koh, Z. X., Shahidah, N., et al. (2012). Prediction of cardiac arrest in critically ill patients presenting to the emergency department using a machine learning score incorporating heart rate variability compared with the modified early warning score. Crit. Care 16, 1–12. doi: 10.1186/cc11396

PubMed Abstract | CrossRef Full Text | Google Scholar

Otani, N. F., Li, M., and Gilmour, R. F. (2005). What can nonlinear dynamics teach us about the development of ventricular tachycardia/ventricular fibrillation? Heart Rhythm. 2, 1261–1263. doi: 10.1016/j.hrthm.2005.07.015

PubMed Abstract | CrossRef Full Text | Google Scholar

Papadopoulos, C. H., Oikonomidis, D., Lazaris, E., and Nihoyannopoulos, P. (2018). Echocardiography and cardiac arrhythmias. Hellenic J. Cardiol. 59, 140–149.

Google Scholar

Patel, B., and Sengupta, P. (2020). Machine learning for predicting cardiac events: what does the future hold? Expert Rev. Cardiovasc. Ther. 18, 77–84. doi: 10.1080/14779072.2020.1732208

PubMed Abstract | CrossRef Full Text | Google Scholar

Perla, H. T., Patloori, S. C. S., Manickavasagam, A., Chase, D., and Roshan, J. (2021). Do the predictors of right ventricular pacing-induced cardiomyopathy add up? Indian Heart J. 73, 582–587. doi: 10.1016/j.ihj.2021.07.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Petersen, P. (1990). Thromboembolic complications in atrial fibrillation. Stroke 21, 4–13. doi: 10.1161/01.str.21.1.4

CrossRef Full Text | Google Scholar

Petrie, A., and Zhao, X. (2012). Estimating eigenvalues of dynamical systems from time series with applications to predicting cardiac alternans. Proc. R. Soc. A Math. Phys. Eng. Sci. 468, 3649–3666. doi: 10.1371/journal.pcbi.1002399

PubMed Abstract | CrossRef Full Text | Google Scholar

Poole-Wilson, P. A., Swedberg, K., Cleland, J. G., Di Lenarda, A., Hanrath, P., Komajda, M., et al. (2003). Comparison of carvedilol and metoprolol on clinical outcomes in patients with chronic heart failure in the carvedilol or metoprolol european trial (comet): randomised controlled trial. Lancet 362, 7–13.

Google Scholar

Prabhu, M. A., Pillai, V., and Shenthar, J. (2017). Comparison of efficacy, pattern of response, occurrence of arrhythmias, and the tolerability of nitroglycerine and isoprenaline as provocative drugs during head-up tilt test. Heart Lung Circ. 26, 586–592. doi: 10.1016/j.hlc.2016.10.006

PubMed Abstract | CrossRef Full Text | Google Scholar

Premchand, R. K., Sharma, K., Mittal, S., Monteiro, R., Dixit, S., Libbus, I., et al. (2014). Autonomic regulation therapy via left or right cervical vagus nerve stimulation in patients with chronic heart failure: results of the anthem-hf trial. J. Card. Fail. 20, 808–816. doi: 10.1016/j.cardfail.2014.08.009

PubMed Abstract | CrossRef Full Text | Google Scholar

Prudat, Y., Madhvani, R. V., Angelini, M., Borgstom, N. P., Garfinkel, A., Karagueuzian, H. S., et al. (2016). Stochastic pacing reveals the propensity to cardiac action potential alternans and uncovers its underlying dynamics. J. Physiol. 594, 2537–2553. doi: 10.1113/JP271573

PubMed Abstract | CrossRef Full Text | Google Scholar

Qin, D., and Singh, J. P. (2020). Low-level tragus stimulation for atrial fibrillation: a glimpse of hope for neuromodulation? JACC Clin. Electrophysiol. 6, 292–294. doi: 10.1016/j.jacep.2020.01.003

PubMed Abstract | CrossRef Full Text | Google Scholar

Raghunath, S., Pfeifer, J. M., Ulloa-Cerna, A. E., Nemani, A., Carbonati, T., Jing, L., et al. (2021). Deep neural networks can predict new-onset atrial fibrillation from the 12-lead ecg and help identify those at risk of atrial fibrillation–related stroke. Circulation 143, 1287–1298. doi: 10.1161/CIRCULATIONAHA.120.047829

PubMed Abstract | CrossRef Full Text | Google Scholar

Ravens, U. (2010). Novel pharmacological approaches for antiarrhythmic therapy. Naunyn Schmiedebergs Arch. Pharmacol. 381, 187–193.

Google Scholar

Shakibfar, S., Krause, O., Lund-Andersen, C., Aranda, A., Moll, J., Andersen, T. O., et al. (2019). Predicting electrical storms by remote monitoring of implantable cardioverter-defibrillator patients using machine learning. Ep Europace 21, 268–274. doi: 10.1093/europace/euy257

PubMed Abstract | CrossRef Full Text | Google Scholar

Singh, B. N. (2005). B -adrenergic blockers as antiarrhythmic and antifibrillatory compounds: an overview. J. Cardiovasc. Pharmacol. Ther. 10, S3–S14. doi: 10.1177/10742484050100i402

PubMed Abstract | CrossRef Full Text | Google Scholar

Skanes, A. C., Krahn, A. D., Yee, R., Klein, G. J., Connolly, S. J., Kerr, C. R., et al. (2001). Progression to chronic atrial fibrillation after pacing: the canadian trial of physiologic pacing. J. Am. Coll. Cardiol. 38, 167–172. doi: 10.1016/s0735-1097(01)01326-2

CrossRef Full Text | Google Scholar

Sridhar, S., Le, D.-M., Mi, Y.-C., Sinha, S., Lai, P.-Y., and Chan, C. (2013). Suppression of cardiac alternans by alternating-period-feedback stimulations. Phys. Rev. E 87:e042712. doi: 10.1103/PhysRevE.87.042712

PubMed Abstract | CrossRef Full Text | Google Scholar

Srinivasan, N. T., and Schilling, R. J. (2018). Sudden cardiac death and arrhythmias. Arrhythm Electrophysiol. Rev. 7, 111–117.

Google Scholar

Stavrakis, S., Kulkarni, K., Singh, J. P., Katritsis, D. G., and Armoundas, A. A. (2020a). Autonomic modulation of cardiac arrhythmias: methods to assess treatment and outcomes. JACC Clin. Electrophysiol. 6, 467–483.

Google Scholar

Stavrakis, S., Stoner, J. A., Humphrey, M. B., Morris, L., Filiberti, A., Reynolds, J. C., et al. (2020b). Treat af (transcutaneous electrical vagus nerve stimulation to suppress atrial fibrillation): a randomized clinical trial. JACC Clin. Electrophysiol. 6, 282–291. doi: 10.1016/j.jacep.2019.11.008

PubMed Abstract | CrossRef Full Text | Google Scholar

Stouffer, G. A., Sheahan, R. G., and Lenihan, D. J. (2001). Syncope and arrhythmias: role of the electrophysiological study. Am. J. Med. Sci. 322, 37–43.

Google Scholar

Strik, M., Ramirez, F. D., Welte, N., Bonnin, T., Abu-Alrub, S., Eschalier, R., et al. (2020). Progressive implantable cardioverter-defibrillator therapies for ventricular tachycardia: the efficacy and safety of multiple bursts, ramps, and low-energy shocks. Heart Rhythm. 17, 2072–2077. doi: 10.1016/j.hrthm.2020.07.032

PubMed Abstract | CrossRef Full Text | Google Scholar

Strogatz, S. (1994). Nonlinear Dynamics and Chaos. Boulder, CO: Avalon Publishing.

Google Scholar

Sulke, N., Silberbauer, J., Boodhoo, L., Freemantle, N., Kamalvand, K., O’Nunain, S., et al. (2007). The use of atrial overdrive and ventricular rate stabilization pacing algorithms for the prevention and treatment of paroxysmal atrial fibrillation: the pacemaker atrial fibrillation suppression (pafs) study. Europace 9, 790–797. doi: 10.1093/europace/eum111

PubMed Abstract | CrossRef Full Text | Google Scholar

Taye, G. T., Shim, E. B., Hwang, H.-J., and Lim, K. M. (2019). Machine learning approach to predict ventricular fibrillation based on qrs complex shape. Front. Physiol. 10:1193. doi: 10.3389/fphys.2019.01193

PubMed Abstract | CrossRef Full Text | Google Scholar

Teodorovich, N., and Swissa, M. (2016). Tilt table test today-state of the art. World J. Cardiol. 8:277. doi: 10.4330/wjc.v8.i3.277

PubMed Abstract | CrossRef Full Text | Google Scholar

Tiwari, P., Colborn, K. L., Smith, D. E., Xing, F., Ghosh, D., and Rosenberg, M. A. (2020). Assessment of a machine learning model applied to harmonized electronic health record data for the prediction of incident atrial fibrillation. JAMA Netw. Open 3:e1919396. doi: 10.1001/jamanetworkopen.2019.19396

PubMed Abstract | CrossRef Full Text | Google Scholar

Trohman, R. G., Huang, H. D., Larsen, T., Krishnan, K., and Sharma, P. S. (2020). Sensors for rate-adaptive pacing: how they work, strengths, and limitations. J. Cardiovasc. Electrophysiol. 31, 3009–3027. doi: 10.1111/jce.14733

PubMed Abstract | CrossRef Full Text | Google Scholar

Trohman, R. G., Kim, M. H., and Pinski, S. L. (2004). Cardiac pacing: the state of the art. Lancet 364, 1701–1719. doi: 10.1016/s0140-6736(04)17358-3

CrossRef Full Text | Google Scholar

Tse, G., Gong, M., Wong, W. T., Georgopoulos, S., Letsas, K. P., Vassiliou, V. S., et al. (2017). The tpeak- tend interval as an electrocardiographic risk marker of arrhythmic and mortality outcomes: a systematic review and meta-analysis. Heart Rhythm. 14, 1131–1137.

Google Scholar

Tse, G., and Yan, B. P. (2017). Traditional and novel electrocardiographic conduction and repolarization markers of sudden cardiac death. Ep Europace 19, 712–721. doi: 10.1093/europace/euw280

PubMed Abstract | CrossRef Full Text | Google Scholar

Vassilikos, V., Dakos, G., Chatzizisis, Y. S., Chouvarda, I., Karvounis, C., Maynard, C., et al. (2011). Novel non-invasive p wave analysis for the prediction of paroxysmal atrial fibrillation recurrences in patients without structural heart disease: a prospective pilot study. Int. J. Cardiol. 153, 165–172. doi: 10.1016/j.ijcard.2010.08.029

PubMed Abstract | CrossRef Full Text | Google Scholar

Vijayaraman, P., Bordachar, P., and Ellenbogen, K. A. (2017). The continued search for physiological pacing: where are we now? J. Am. Coll. Cardiol. 69, 3099–3114. doi: 10.1016/j.jacc.2017.05.005

PubMed Abstract | CrossRef Full Text | Google Scholar

Wellens, H., and Smeets, J. (1993). Idiopathic left ventricular tachycardia. Cure by radiofrequency ablation. Circulation 88, 2978–2979. doi: 10.1161/01.cir.88.6.2978

CrossRef Full Text | Google Scholar

Wilson, D., and Ermentrout, B. (2017). Stochastic pacing inhibits spatially discordant cardiac alternans. Biophys. J. 113, 2552–2572. doi: 10.1016/j.bpj.2017.10.001

PubMed Abstract | CrossRef Full Text | Google Scholar

Yang, A., Hochhäusler, M., Schrickel, J., Bielik, H., Shlevkov, N., Schimpf, R., et al. (2003). Advanced pacemaker diagnostic features in the characterization of atrial fibrillation: impact on preventive pacing algorithms. Pacing Clin. Electrophysiol. 26, 310–313. doi: 10.1046/j.1460-9592.2003.00039.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Zagrodzky, J. D., Ramaswamy, K., Page, R. L., Joglar, J. A., Sheehan, C. J., Smith, M. L., et al. (2001). Biventricular pacing decreases the inducibility of ventricular tachycardia in patients with ischemic cardiomyopathy. Am. J. Cardiol. 87, 1208–1210. doi: 10.1016/s0002-9149(01)01498-9

CrossRef Full Text | Google Scholar

Zannad, F., De Ferrari, G. M., Tuinenburg, A. E., Wright, D., Brugada, J., Butter, C., et al. (2015). Chronic vagal stimulation for the treatment of low ejection fraction heart failure: results of the neural cardiac therapy for heart failure (nectar-hf) randomized controlled trial. Eur. Heart J. 36, 425–433. doi: 10.1093/eurheartj/ehu345

PubMed Abstract | CrossRef Full Text | Google Scholar

Zehir, R., Karabay, C. Y., Kalaycı, A., Akgün, T., Kılıçgedik, A., and Kırma, C. (2015). Evaluation of tpe interval and tpe/qt ratio in patients with slow coronary flow. Anatolian J. Cardiol. 15:463. doi: 10.5152/akd.2014.5503

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: arrhythmias, pacing, alternans, control, non-linear dynamics, prediction

Citation: Patel MH, Sampath S, Kapoor A, Damani DN, Chellapuram N, Challa AB, Kaur MP, Walton RD, Stavrakis S, Arunachalam SP and Kulkarni K (2021) Advances in Cardiac Pacing: Arrhythmia Prediction, Prevention and Control Strategies. Front. Physiol. 12:783241. doi: 10.3389/fphys.2021.783241

Received: 25 September 2021; Accepted: 08 November 2021;
Published: 02 December 2021.

Edited by:

Ademuyiwa S. Aromolaran, The University of Utah, United States

Reviewed by:

Brian P. Delisle, University of Kentucky, United States
John Pearce Morrow, Columbia University, United States

Copyright © 2021 Patel, Sampath, Kapoor, Damani, Chellapuram, Challa, Kaur, Walton, Stavrakis, Arunachalam and Kulkarni. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Shivaram P. Arunachalam, PoigaiArunachalam.Shivaram@mayo.edu; Kanchan Kulkarni, kanchan.kulkarni@ihu-liryc.fr

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.