Front. Neuroimaging, 20 February 2023
Sec. Brain Imaging Methods

DSC MRI in the human brain using deoxyhemoglobin and gadolinium—Simulations and validations at 3T

  • 1Department of Medical Biophysics, University of Toronto, Toronto, ON, Canada
  • 2Krembil Brain Institute, University Health Network, Toronto, ON, Canada
  • 3Department of Physiology, University of Toronto, Toronto, ON, Canada
  • 4Joint Department of Medical Imaging and the Functional Neuroimaging Lab, University Health Network, Toronto, ON, Canada
  • 5Department of Anaesthesia and Pain Management, University Health Network, University of Toronto, Toronto, ON, Canada
  • 6Toronto General Hospital Research Institute, Toronto General Hospital, Toronto, ON, Canada
  • 7The Joint Department of Medical Imaging, The Toronto Western Hospital, Toronto, ON, Canada
  • 8Center for Neuroscience Imaging Research, Institute for Basic Science & Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Republic of Korea

Introduction: Dynamic susceptibility contrast (DSC) MRI allows clinicians to determine perfusion parameters in the brain, such as cerebral blood flow, cerebral blood volume, and mean transit time. To enable quantification, susceptibility changes can be induced using gadolinium (Gd) or deoxyhemoglobin (dOHb), the latter just recently introduced as a contrast agent in DSC. Previous investigations found that experimental parameters and analysis choices, such as the susceptibility amplitude and partial volume, affect perfusion quantification. However, the accuracy and precision of DSC MRI has not been systematically investigated, particularly in the lower susceptibility range.

Methods: In this study, we compared perfusion values determined using Gd with values determined using a contrast agent with a lower susceptibility—dOHb—under different physiological conditions, such as varying the baseline blood oxygenation and/or magnitude of hypoxic bolus, by utilizing numerical simulations and conducting experiments on healthy subjects at 3T. The simulation framework we developed for DSC incorporates MRI signal contributions from intravascular and extravascular proton spins in arterial, venous, and cerebral tissue voxels. This framework allowed us to model the MRI signal in response to both Gd and dOHb.

Results and discussion: We found, both in the experimental results and simulations, that a reduced intravascular volume of the selected arterial voxel, reduced baseline oxygen saturation, greater susceptibility of applied contrast agent (Gd vs. dOHb), and/or larger magnitude of applied hypoxic bolus reduces the overestimation and increases precision of cerebral blood volume and flow. As well, we found that normalizing tissue to venous rather than arterial signal increases the accuracy of perfusion quantification across experimental paradigms. Furthermore, we found that shortening the bolus duration increases the accuracy and reduces the calculated values of mean transit time. In summary, we experimentally uncovered an array of perfusion quantification dependencies, which agreed with the simulation framework predictions, using a wider range of susceptibility values than previously investigated. We argue for caution when comparing absolute and relative perfusion values within and across subjects obtained from a standard DSC MRI analysis, particularly when employing different experimental paradigms and contrast agents.

Introduction

Cerebral perfusion imaging is used to investigate diseases characterized by vascular abnormality, such as atherosclerotic disease, vasculitis, Moyamoya disease, and neoplasms (Harris et al., 1998; Law et al., 2008; Copen et al., 2011; Qiao et al., 2017). In dynamic susceptibility contrast (DSC) MRI, a bolus of paramagnetic contrast agent is tracked as it passes through the vasculature, enabling perfusion quantification based on the principles of indicator dilution theory (Zierler, 1962; Rosen et al., 1990). The paramagnetic contrast agent creates inhomogeneities in the local magnetic field, reducing signal by diminishing the phase coherence and increasing the relaxation rate (R2*) of extravascular and intravascular water proton spins (Rosen et al., 1990). The signal time courses resulting from the passage of contrast agent are then used to calculate cerebral blood volume (CBV), cerebral blood flow (CBF) and mean transit time (MTT), all of which possessing clinical and research utility (Harris et al., 1998; Law et al., 2008; Copen et al., 2011; Tanaka et al., 2011; Wang et al., 2013; Qiao et al., 2017).

Gadolinium (Gd), a paramagnetic metal administered intravenously, is the standard contrast agent used in DSC MRI (Jahng et al., 2014). It has long been known that deoxyhemoglobin (dOHb), the deoxygenated form of oxygen's carrier protein in our blood, hemoglobin, is paramagnetic as well (Pauling and Coryell, 1936). While being used to localize neuronal activity in functional MRI (fMRI) since the early 1990s (Ogawa et al., 1990, 1992; Bandettini et al., 1992; Kwong et al., 1992), the paramagnetic properties of dOHb were only recently exploited in humans to generate a bolus for DSC MRI analysis by transiently reducing the oxygen concentration in inhaled gas (Poublanc et al., 2021; Vu et al., 2021; Lee et al., 2022). Although not using a standard DSC analysis, it should be noted that hypoxia was used to estimate CBV in cats as early as 1998 (van Zijl et al., 1998).

A critical area of DSC MRI research pertains to modeling the changes in R2* that result from the passage of contrast agent through simulated vasculature (Weisskoff et al., 1994; Boxerman et al., 1995; van Zijl et al., 1998). By incorporating the effects of contrast agent susceptibility and tissue structure, researchers have been able to simulate and understand how various properties, such as vessel heterogeneity, tissue composition, and contrast agent extraversion, affect DSC MRI signal and perfusion quantification (Kiselev, 2001; Kjølby et al., 2009; Quarles et al., 2009; Semmineh et al., 2014; Digernes et al., 2017). These simulation studies have so far been limited to DSC MRI applications when using Gd.

From a simulation standpoint, Gd and dOHb can be modeled with the same framework for DSC MRI, with the only major differences between them being bolus shape, vascular compartmentalization, and molar susceptibility: Gd induces a higher susceptibility change than dOHb; Gd is limited to the plasma fraction whereas dOHb is limited to the red blood cell fraction; currently, the dOHb bolus can only be generated to be longer in duration than the Gd bolus. While researchers have investigated the accuracy and precision of perfusion quantification in the higher susceptibility range (references in Calamante, 2013), this has not been investigated in the lower susceptibility range. Some of the parameters relevant to perfusion quantification include the dosage of contrast agent [in the case of dOHb, the magnitude of the hypoxic bolus (ΔSaO2)], baseline oxygen saturation (base-SaO2), bolus duration, and choice of reference voxel (i.e., artery or vein).

To examine these questions, we developed a DSC MRI signal model, which incorporates separate vascular compartments with both linear extravascular and quadratic intravascular relaxation rate contributions, as used in a previous fMRI model (Uludag et al., 2009). This enabled us to model tissue, arterial, and venous signal in response to generic contrast agents with arbitrary susceptibility values and bolus characteristics, such as duration and shape. We use this model to examine the effects of contrast agent susceptibility, ΔSaO2, base-SaO2, bolus duration, tissue properties, and the choice of reference voxel on perfusion quantification. This framework is theoretically applicable to any contrast agent if its molar susceptibility is known. Experimentally, bolus properties such as ΔSaO2, base-SaO2, and duration were easily controllable with the gas control system used to create dOHb contrast, allowing us to study the effects of these parameters. We validated our model by comparing the simulation predictions to experimental results obtained from six healthy subjects at 3T when using dOHb or Gd as contrast.

DSC MRI signal theory

Section 1: Forward model

The forward model is based on work by Uludag et al., 2009, originally developed to model gradient- (GRE) and spin-echo (SE) fMRI signals for magnetic field strengths between 1.5 and 16.4T (Uludag et al., 2009). As the fMRI signal change results from susceptibility variations, caused by dOHb in the case of neuronal activation or vascular challenges, the model can also be applied to DSC MRI utilizing any contrast agent that induces magnetic susceptibility changes. Here, this framework is generalized and applied to model 3T GRE signal changes induced by boluses of dOHb and Gd (please see the Discussion section for other DSC signal models and how they relate to the proposed model).

Signal intensity

T2*-weighted MRI signal (Stot) is composed of extravascular (EES) and intravascular (IVS) signal contributions, weighted by their respective volumes (Uludag et al., 2009):1

Stot= (1-CBV)·SEES+iCBVi·SIVS, i    (1)

SEES and SIVS are the extravascular and intravascular signal contributions2, respectively. The index i denotes the specific vascular compartment within the voxel being simulated (artery, vein, capillary, venule, or arteriole). Therefore, the signal is a summation of extravascular signal along with various intravascular signal contributions, each scaled by their volume proportion within the voxel. Of note, many DSC signal models from previous clinical research studies do not distinguish SIVS and SEES components (Calamante et al., 2000; Patil and Johnson, 2011; Patil et al., 2013; Chappell et al., 2015). The separate signal components can be written as:

SEES=e-TE·(R2,0,EES*+R2,con,EES*)    (2)
SIVS=e-TE·(R2,0,IVS*+R2,con,IVS*)    (3)

SEES and SIVS (Eqs. 2 and 3) are modeled as mono exponentials to the power of the negative echo time (here, TE is set to 30 ms) multiplied by the transverse relaxation rate (R2*), the inverse of the transverse relaxation time (Yablonskiy and Haacke, 1994). R2,con,EES* and R2,con,IVS* are the magnetic relaxation rates induced by applied contrast agent in the extravascular and intravascular space of a voxel, respectively. R2,0,EES* and R2,0,IVS* are the baseline magnetic relaxation rates without contrast agent (R2,0*) for extravascular and intravascular space, respectively.

As shown in Eq. 1, separate signal exponentials are added together for each vascular compartment within a voxel for SIVS, however, relaxation rates are added from each vascular compartment within a single exponential for SEES, as will be described in Eq. 5. As a reminder, it is intravascularly contained contrast agent that gives rise to T2* signal change in both the intravascular and extravascular spaces (please see the Discussion section for leakage of contrast agent into the extravascular space).

Extravascular and intravascular relaxation rates

At 3T, R2,0,EES* and R2,0,IVS* have been experimentally approximated to be 20.99 s−1 and 13.8 s−1 for fully oxygenated blood, respectively (Zhao et al., 2007; Uludag et al., 2009). Polynomial equations that model R2,con,EES* for various vessel radii have been determined using Monte-Carlo simulations; for a large range of susceptibility values at 3T, the equations (expressed as per one percent of CBV) can be simplified by a linear relationship (see Table 1 and Uludag et al., 2009):

ΔvdOHb=0.264·Hct·(1-Y0-ΔYdOHb)·γ·B0    (4)
TABLE 1
www.frontiersin.org

Table 1. Coefficients for extravascular relaxation rates in various vessel radii.

ΔvdOHb is the frequency shift induced by dOHb, Hct is the hematocrit fraction (simulated as 0.4), ΔYdOHb is the percentage change in the oxygenation as a decimal (which corresponds to changes in dOHb), γ is the gyromagnetic ratio (2π·42.6 MHz/T), and B0 is the magnetic field strength (3T). Y0 represents the baseline oxygen saturation as a decimal, dependent on the vascular compartment (refer to Table 2). R2,con,EES* can then be determined by summating R2,con,EES,i* for each vessel compartment within the voxel:

R2,con,EES*=i(pCBVi·R2,con,EES,i*)    (5)
TABLE 2
www.frontiersin.org

Table 2. Summary of simulated voxel properties for arterial, tissue, and venous voxels.

Here, i denotes an individual vessel compartment and pCBVi represents the percentage blood volume of the vessel contributing to the signal. Note that pCBVi must be a percentage and not a decimal (i.e., for a blood volume of 5%, a value of 5 is used as opposed to 0.05), as the equations for R2,con,EES,i* in Table 1 are modeled for one percent of CBV.

Finally, the R2,con,IVS* dependence on dOHb at 3T (Hct ≈ 0.4) was found to be quadratic from previous blood phantom work (Zhao et al., 2007):

R2,con,IVS*=181·(1-Y0-ΔYdOHb)2    (6)

Quantifying the relationship between dOHb and Gd

The above equations quantify MRI signal in response to changes in dOHb. To calculate the signal change induced by Gd, the effect of Gd on R2,con,IVS* and R2,con,EES* must be determined. One way to determine this is by finding a quantitative relationship between Gd and dOHb; given that dOHb and Gd are both paramagnetic contrast agents, a certain increase in [dOHb] will induce the same susceptibility effect as 1 mM of Gd.

The frequency shifts on the surface of the blood vessel caused by changes in dOHb and Gd are known. The frequency shift induced by dOHb is described in Eq. 4 while the frequency shift induced by Gd (ΔvGd), when Hct ≈ 0.4, is described by the following equation, derived from Kjølby et al. (2006):

ΔvGd=γ·B0·0.026·[Gd]    (7)

[Gd] represents the concentration of Gd (mM) in blood, and 0.026 ppm/mM represents the molar susceptibility of Gd, specifically Gd-DTPA (van Osch et al., 2003). Setting Eq. 7 equal to Eq. 4 allows for the interconversion of percentage dOHb saturation and Gd concentration (mM). This interconversion is an important step in being able to simulate and compare dOHb- and Gd-induced signal changes with the same DSC MRI framework:

0.264·0.4·(1-Y0-ΔYdOHb)·γ·B0=γ·B0·0.026·[Gd]                                        + 0.264·0.4·(1-Y0)·γ·B0    (8.1)
  YdOHb=-0.246·[Gd]    (8.2)

By setting ΔvdOHb equal to vGd at 3T, a certain concentration of Gd associated with magnetic susceptibility can be converted to a value of blood oxygenation and vice versa. For example, a 1 mM increase in Gd is equivalent in susceptibility to a 24.6% increase in dOHb at 3T. This equivalency is referred to as the pseudo-oxygenation (the oxygenation change that has the same susceptibility effect as 1 mM of Gd on the magnetic field at 3T). Eqs. 1–6 were then used to simulate either a Gd- or dOHb-induced bolus by substituting ΔYdOHb with −0.246·[Gd] or using ΔYdOHb, respectively. Thus, it is important to note that Gd and dOHb were modeled with the same signal framework, with differences only in the input bolus properties (i.e., susceptibility, concentration, shape, and duration) and compartmentalization in the vasculature. This idea can be further understood in Supplementary Figure 1.

Section 2: Model inversion

Relaxation curves

The first step in standard DSC analysis, following preprocessing, is the conversion of a signal time course (S(t)) into a relaxation rate time course (ΔR2*(t)):

ΔR2*(t)=-(1TE)·ln(S(t)S0)    (9)

This step requires the specification of a baseline signal (S0), along with the TE. Of note, as intravascular relaxation is quadratically related to contrast agent, some studies have opted for a quadratic equation as opposed to Eq. 9 to solve the concentration time course in artery or vein (see references in Calamante, 2013). While we tested this method, we do not apply it here as it assumes that the selected arterial voxel is entirely intravascular, and due to partial volume errors, this is almost never the case.

Calculations of cerebral blood volume, cerebral blood flow, and mean transit time

From this point forward, the term “relative” will be used to describe tissue signal normalized by signal from a reference voxel (arterial or venous). Once tissue, arterial, and venous signal time courses are converted into ΔR2*(t), the relative CBV (rCBV) can be calculated (Knutsson et al., 2010). Of note, normalizing tissue to venous as opposed to arterial signal is of clinical interest, as venous voxels tend to have reduced partial volume effects (Calamante, 2013). However, we only normalize to venous signal for the purposes of calculating rCBV, as doing so for CBF and MTT would violate principles of indicator dilution theory in our method (Meier and Zierler, 1954); this is because we normalize tissue time courses directly to the venous time courses, and do not first scale the arterial time course by the area under the venous curve. Thus, performing a deconvolution of tissue data from the venous time course would not accurately reflect CBF and MTT.

rCBV is typically scaled by κ, the hematocrit correction factor [(1−Hct)/(1 − 0.69Hct)] when using Gd contrast (Tudorica et al., 2002). However, when using dOHb contrast, we concluded that the κ value should be set to Hct/0.69Hct, as contrast is limited to the red blood cell fraction and not the plasma.

The relative CBF (rCBF) and MTT are calculated based on indicator dilution theory and the central volume principle, respectively (Meier and Zierler, 1954). These values were estimated using discretized singular value decomposition (Østergaard et al., 1996; Wirestam et al., 2000) with a noise threshold of 20%, which was determined to be most effective in previous work (Bjørnerud and Emblem, 2010).

The simulation framework and analysis pipeline are visually summarized in Supplementary Figure 2.

Methods

Part 1: Simulation methods

Arterial, tissue, and venous concentration input functions

The forward model, as described above, can be used to model any shape, duration, or concentration of Gd and dOHb input time courses. For the examples in this study, we generally assume parameters closely resembling experimental data.

The Gd input time course [Gd(t)] in an artery is well-modeled as a gamma variate function (Thompson et al., 1964; Davenport, 1983).

Gd(t)=a·(tb)c·e(c·(1-tb))    (10)

While it is difficult to attain the exact concentration profile of Gd in an artery (Patil et al., 2013; Kellner et al., 2018; Lind et al., 2020), the shape parameters, b (time to peak) and c, are set to 2.5 s and 3 (unitless), respectively, to model the shape observed experimentally. a (the peak concentration) is set to approximately 2 mM—see Discussion. The dOHb input, as informed by our experimental validations and previous work (Poublanc et al., 2021), was modeled as a rectangular hypoxic bolus of roughly 30 s duration, convolved with a decaying mono-exponential as the bolus takes time to reach the target oxygenation and enter the cerebral vasculature.

To attain the tissue concentration profile, the arterial input was convolved with a well-defined biexponential residue function [R(t)] which designates fast and slow flowing vascular compartments (Mehndiratta et al., 2014):

R(t)=f·e-t·t1+(1-f)·e-t·t2    (11.1)

Here t1 (fast-flowing compartment transit time), t2 (slow-flowing compartment transit time), and f (fraction of flow in the fast-flowing compartment), are vascular flow parameters which yield the MTT in the following equation (Mehndiratta et al., 2014):

MTT=ft1+1-ft2    (11.2)

As we expect the venous voxel to be generalizable to a single vascular compartment, we convolve the arterial input with a mono-exponential residue function to obtain the venous concentration profile (Mehndiratta et al., 2014):

R(t)=e-tMTT    (12)

The flow parameters, informed by clinical data and literature yield a tissue voxel MTT of roughly 3 s (f = 0.92, t1 = 0.68, t2 = 0.05) and a venous voxel MTT of 4 s (MTT = 4) (Ibaraki et al., 2007). As well, a delay of 1 s and 2 s were applied to the tissue and venous voxels, respectively (Ibaraki et al., 2007). These parameters are easily adjustable to model other values.

Simulated voxel parameters

The voxel parameters used for the simulations were based off previously reported physiological estimates and are summarized in Table 2.

Part 2: Experimental methods

Subjects

This study was approved by the Research Ethics Board of the University Health Network, Toronto, Canada, and conforms to the Declaration of Helsinki. Written informed consent was obtained in all 6 healthy volunteers (age range 22–60, 1 woman). All subjects were non-smokers and not taking regular medication. While our sample size is rather small, the results were consistent across all subjects.

Scan protocols and MRI sequences

The RespirAct™ RA-MR (Thornhill Medical, Toronto, Canada) was used to target lung SaO2 while maintaining isocapnia (Sobczyk et al., 2016). The maintenance of isocapnia ensures that vasodilation and CBF changes are mitigated. All gases and sensors were calibrated prior to use. GRE-EPI and anatomical T1 scans were acquired on a 3T MRI system (Signa HDx—GE Healthcare, Milwaukee) with echo-planar (EPI) acquisition.

MR parameters for dOHb and Gd GRE scans were TR/TE = 1,500/30 ms, flip angle = 73°, 3 mm isotropic voxels, 29 contiguous slices—no interslice gap. The coverage is lower as the scanner does not have multi-band capabilities, although the coverage is not a relevant parameter for our results. The dOHb scan duration was approximately 210 s per paradigm; the Gd scan duration was approximately 90 s. A high-resolution T1-weighted 3D spoiled GRE sequence was acquired for each subject as well (TI = 450 ms, TR = 7.88 ms, TE = 3 ms, flip angle = 12°, voxel size = 0.859 x 0.859 x 1 mm, matrix size = 256 x 256, 146 slices, field of view = 24 x 24 cm, no interslice gap).

In the dOHb experiments, the hypoxic bolus magnitude (ΔSaO2) and/or baseline oxygen saturation (base-SaO2) were varied (Figure 1). To determine the effect of ΔSaO2 and base-SaO2 for perfusion quantification, SaO2 values were modified to generate four hypoxic paradigms with three ~30 s hypoxic boluses as follows: 98–90%, 98–84%, 98–75%, and 88–80%. Each hypoxic drop was acquired over multiple breaths, depending on the rate and volume of ventilation, as attaining the hypoxic state required wash-out of oxygen from the functional residual capacity of the lungs (air remaining after exhalation). Due to this physiological limitation, the dOHb bolus is longer in duration that the Gd bolus, which depends on the speed of injection and dispersion of the bolus in its trajectory from peripheral vein to the brain. The hypoxic target was maintained for approximately 30 s before returning to baseline in a single breath. The rapidity of the return to baseline was due to the higher available oxygen partial pressure target (713 mmHg) relative to the hypoxic target (<100 mmHg). The baseline before and after each bolus was 30 s in duration.

FIGURE 1
www.frontiersin.org

Figure 1. Summary of the dOHb Experimental Method. Subjects (n = 6) were imaged in a 3T MRI scanner with a mask attached to the RespirAct®, where SaO2 was altered to attain desired levels of base-SaO2 and ΔSaO2. In sum, the following paradigms were generated: small ΔSaO2 with high base-SaO2 (red: 98–90% SaO2), medium ΔSaO2 with high base-SaO2 (purple: 98–84% SaO2), large ΔSaO2 with high base-SaO2 (blue: 98–75% SaO2), and small ΔSaO2 with low base-SaO2 (green: 88–80% SaO2).

The dOHb paradigms were acquired prior to the Gd paradigm such that Gd would not remain in the system as a steady-state and influence the dOHb data. Here, 5 mL of 1 M gadobutrol (Gadovist®) was injected at 5 mL/s followed by 30 mL of saline at 5 mL/s. This corresponds to a dose that is slightly below the standard dose of 0.1 mmol/kg for the average subject.

One subject reported a mild change in respiratory sensation during the dOHb protocol. Two subjects reported discomfort and slight nausea during Gd acquisitions. Only data from one paradigm of one subject was removed due to excessive motion during the scan.

Preprocessing

FSL (version 6.0.4) was used for image preprocessing (Woolrich et al., 2009). Brain extraction was performed on T1 anatomical images; brain extraction, motion correction, and interleaved slice time correction were performed on dOHb and Gd scans. Gaussian temporal filtering was only performed on the dOHb scan by averaging each signal time point by a 1 × 5 Gaussian kernel; doing so for the Gd bolus was not required due to high signal-to-noise. For all dynamic scans, the first 2–3 timepoints were removed to allow for signal equilibration. Percentage maximal signal change (ΔS) and the signal to noise ratio (SNR) were calculated:

ΔS=(S0-SPeak)S0·100    (13)
SNR=ΔSεt     (14)

In Eq. 13, S0 is the average signal of ten temporal volumes before and after the selected bolus, and SPeak is the maximum signal drop. The SNR is computed as shown in Eq. 14, where εt is the standard deviation of signal for ten temporal volumes before and after the selected bolus divided by S0. S(t) was then converted to ΔR2 *(t) using Eq. 9. The above equations were applied to dOHb and Gd signal data from all subjects using Python 3.7.6.

Selection of arterial and venous voxels

Typically, in a DSC MRI analysis, the T2*-weighted DSC data is registered to high-resolution T1-weighted anatomy. In this study, to reduce registration errors and the misidentification of tissue classes, we sought to identify arterial and venous voxels in the functional rather than anatomical space. However, the main results are not affected by either analysis choice. As arterial voxels are difficult to clearly identify on functional images, we used the following conditions for voxel selection: Arterial voxels were defined based on their proximity to the M1 segment of the middle cerebral artery (MCA) using both functional and registered T1 anatomical data, ΔS within the top 10% of the ΔS values, and short delay (~0–1 s). Venous voxels were defined based on their proximity to the superior sagittal sinus (SSS) using registered T1 anatomical data, ΔS within the top 10% of the ΔS values, long delay on the Gd and dOHb maps (~2 s), and low S0 (as vein contains more dOHb than artery, reducing the baseline signal). Once the voxels were selected on the Gd map, they were linearly registered to the dOHb paradigms to avoid displacement due to motion between the runs.

We recognize that for the purposes of maintaining better inter-subject agreement, it is often better to take an average of arterial or venous candidate voxels. However, for the questions that we were trying to address in our study, we wanted to ensure that the voxel selected was mostly intravascular, and thus sought to avoid averaging between multiple candidate voxels by selecting one voxel.

Perfusion quantification

After defining an arterial input function (AIF) from the M1 segment of the MCA or venous output function (VOF) from the dorsal section of the SSS, each time series was truncated to begin at the start of the AIF bolus and finish at the end of the VOF bolus. Then, rCBV, rCBF, and MTT for each voxel were calculated (see Theory section, above). Any voxel yielding a negative rCBV was excluded, observed in voxels with high noise and low blood volume (roughly 10% of the voxels for Gd and 20% of the voxels for dOHb).

Gray and white matter segmentation

As mentioned above, while DSC MRI studies typically employ T1 anatomical segmentation to DSC maps due to higher anatomical contrast in T1 images relative to baseline T2* images, this form of segmentation requires image registration. In the current study, we use a segmentation method in the native space to avoid registration errors: GM exhibits larger signal changes in T2* images than WM and can therefore be easily distinguished. Nevertheless, applying these different segmentation strategies does not affect the main findings of this study. The ΔS functional maps for Gd were segmented into gray matter (GM) and white matter (WM) using a threshold method. WM voxels were defined as those containing ΔS values within the lowest 15–20% of the ΔSGd data range, while GM was defined as voxels containing the highest 75% ΔS values. The maps were then binarized and linearly registered to the functional space of each dOHb paradigm. The GM and WM regions were further masked with a thresholded εt map (voxels with values in the lowest 10% of the εt range were kept in the mask) to remove noisy voxels in peripheral brain tissue and skull base. The exclusion of voxels based on εt also removes large arterial and venous voxels from the GM masks due to high levels of physiological noise in these regions. The GM and WM were finally masked with a thresholded Gd MTT map (voxels within the lowest 70% of robust range of MTT were kept in the mask) to remove voxels in the CSF (Supplementary Figure 4).

Statistical testing

Statistical testing was conducted for some of the experimental data by applying a two-tailed student's t-test, with an alpha level of 0.05. As we only compare mean values of various parameters with simulation-guided hypotheses, we believe this statistical method is sufficient. Linear regressions were also conducted for voxel-wise comparisons.

Data and code availability

Anonymized data will be shared by request from any qualified investigator for purposes such as replicating procedures and results presented in the article, so long as the data transfer agrees with the University Health Network and Health Canada legislation on the general data protection regulation.

Code used for the simulations can be found on GitHub through the following link: https://github.com/jschulman-1998/NeuroImage2022.

Results

Figure 2 displays S(t) and ΔR2 *(t) for simulated tissue, arterial, and venous voxels in response to a bolus of either Gd or dOHb. The baseline signal is highest for artery and lowest for vein due to differences in baseline oxygen saturation. In addition, the dOHb boluses are longer than Gd boluses, mimicking experimentally observed bolus durations (roughly 30–40 s for hypoxic boluses). According to this simulation, ΔR2,max*, the peak of ΔR2 *(t), attained with dOHb is roughly 4 times larger in vein relative to artery even though the same ΔSaO2 is simulated for artery and vein, indicating a substantial effect of the baseline oxygen saturation on ΔR2 *(t). Remarkably, for Gd, ΔR2,max* is only roughly 1.1 times larger in vein relative to artery, indicating a reduced effect of the baseline oxygen saturation on ΔR2 *(t) for contrast agents with a higher susceptibility. As well, the ratio of arterial to tissue ΔR2,max* is much higher when using Gd in comparison with dOHb. Conversely, the ratio of venous to tissue R2,max* is much more consistent between Gd and dOHb.

FIGURE 2
www.frontiersin.org

Figure 2. Simulated S(t) and R2*(t) for Gd and dOHb Contrast. (A) S(t) in artery (CBV = 100%), tissue (CBV = 4%, MTT = 3 s), and vein (CBV = 100%, MTT = 4 s) following contrast agent administration of dOHb (top) and Gd (bottom). (B) Corresponding R2*(t). Note that the simulated dOHb profile follows a 98–75% SaO2 and Gd profile has a peak input concentration of 2 mM.

Figure 3 provides an analysis of the maximal relaxation rate (ΔR2,max*) in artery, tissue, and vein as a function of the simulated CBV, mimicking partial volume effects. For tissue, linear changes in the simulated CBV result in linear increases in the ΔR2,max*, in response to either contrast agent. This is not the case for arterial and venous responses. For the arterial response to dOHb, increasing CBV increases the ΔR2,max* up to a certain CBV (roughly 30%), such that any further increase in CBV leads to decreases in ΔR2,max*. The simulated arterial CBV to ΔR2,max* relationship in response to Gd also has a maximum for an intermediate CBV value. For venous voxels, ΔR2,max* saturates at high CBV values in response to either Gd or dOHb. These results point to a complex relationship between relaxation rate changes and voxel composition, which is further elaborated upon in Supplementary Figure 5.

FIGURE 3
www.frontiersin.org

Figure 3. ΔR2,max* as a Function of Simulated CBV (%). (A) ΔR2,max* as a function of the simulated CBV for artery and vein in response to Gd and dOHb. (B) ΔR2,max* as a function of the simulated CBV for tissue (CBVWM ≈ 2%; CBVGM ≈ 4%) in response to Gd and dOHb. Note that the simulated dOHb bolus follows a 98–75% SaO2 and Gd bolus has a peak input concentration of 2 mM.

Figure 4 shows ΔS maps, simulation data, and collective experimental results in response to dOHb and Gd boluses. For simplicity here and elsewhere, ΔS represents ΔSmax and is displayed as an absolute signal change. ΔSGd is ~10x higher than ΔSdOHb. When the ΔSdOHb values are divided by the ΔSGd values in corresponding voxels, hereafter termed the ΔS ratio, differences between artery, vein, CSF, GM, and WM are visually highlighted (see Supplementary Figure 4 for masks of these regions). The ΔS ratio is significantly different in each vascular/tissue compartment, as evident from the box and whisker plot in Figure 4, except for artery vs. WM. The experimental finding of the ΔS ratio being significantly higher in venous voxels in comparison to tissue voxels (GM and WM) is corroborated by simulation findings, which show that the venous ΔS ratio is larger than either the GM or WM ΔS ratio. The experimental finding of the ΔS ratio being higher in arterial voxels in comparison to tissue voxels (GM and WM) is not corroborated by simulation findings when the artery is simulated at a CBV of 90%, as is shown here, but is when artery is simulated at a lower blood volume. The WM ΔS ratio being higher than the GM ΔS ratio, although present in the experimental data, is not observed in the simulations—refer to Discussion.

FIGURE 4
www.frontiersin.org

Figure 4. ΔS for Gd and dOHb in Artery, Vein, GM, and WM. (A) ΔS (%) maps (representing ΔSmax) for Gd and dOHb in a representative subject. Note that the dOHb map here is an average of the following paradigms: small ΔSaO2 with high base-SaO2, medium ΔSaO2 with high base-SaO2, large ΔSaO2 with high base-SaO2. (B) Six representative axial slices of ΔSdOHb map divided by ΔSGd map. (C) Box and whisker plot (average from all subjects) of ΔSdOHb/ΔSGd for various tissue compartments, including artery (red), vein (blue), GM (gray), and WM (white). Significant differences are denoted with asterisks. (D) ΔSdOHb/ΔSGd for simulated artery (red), vein (blue), GM (gray), and WM (white) voxels. Simulated dOHb bolus followed a 98–84% SaO2 profile. Simulated Gd bolus has a peak concentration of 6 mM. Simulated CBV values were as follows: artery (90%), vein (90%), GM (4%), and WM (2%).

Figure 5 shows calculated rCBV maps from the Gd and dOHb paradigms in a representative subject when normalizing the tissue relaxation rate time courses to that of a selected arterial (top) or venous voxel (bottom). This is elaborated upon in Figure 6, where cumulative rCBV results from all subjects in addition to simulation results are presented for various paradigms. Both figures show that the calculated rCBV depends on the molar susceptibility of contrast agent (Gd vs. dOHb), amount of contrast agent applied, baseline susceptibility (base-SaO2), and choice of normalizing voxel (artery or vein). In the experimental validations and simulations, the order of rCBV estimation for the dOHb paradigms, from lowest to highest, is as follows: small ΔSaO2 with high base-SaO2, medium ΔSaO2 with high base-SaO2, large ΔSaO2 with high base-SaO2, and small ΔSaO2 with low base-SaO2. There is a significant difference between the small ΔSaO2 with high base-SaO2 paradigm and all the others, except for the medium ΔSaO2 with high base-SaO2 paradigm. The pattern of rCBV estimations when normalizing tissue to vein is the same as in artery, but the effects of ΔSaO2, base-SaO2, and choice of contrast agent are no longer significant, as observed in the experimental data and corroborated by the simulations. The same pattern of the aforementioned parameters when normalizing to artery was also observed in the calculation of rCBF (Figure 7 and Supplementary Table 1). Furthermore, as shown in the experimental validations, inter-subject variability (defined as the interquartile range in the box and whisker plot) appears to be reduced as the ΔSaO2 increases and base-SaO2 decreases (Figure 6). Ultimately, the experimental and simulation data agree and show that rCBV and rCBF quantifications depend heavily on ΔSaO2 and base-SaO2, and that normalizing tissue to vein as opposed to artery results in more accuracy and consistency in the quantification of rCBV across different dOHb paradigms.

FIGURE 5
www.frontiersin.org

Figure 5. rCBV Dependency on ΔSaO2 and Base-SaO2. Top (Arterial Normalization). rCBV (mL/100 g) maps from representative subject axial slice where tissue is normalized to an arterial voxel in the MCA for five subsequent paradigms (left to right): standard Gd, large ΔSaO2 with high base-SaO2, medium ΔSaO2 with high base-SaO2, small ΔSaO2 with high base-SaO2, small ΔSaO2 with low base-SaO2. Bottom (Venous Normalization). rCBV (mL/100g) maps from representative subject axial slice where tissue is normalized to a venous voxel in the SSS for the same paradigms.

FIGURE 6
www.frontiersin.org

Figure 6. rCBV Dependency on Arterial vs. Venous Normalization. (A) Validations. Box and whisker plot of calculated rCBV when normalizing tissue to arterial voxel in the MCA vs. venous voxel in the SSS (n = 6) for Gd and dOHb at varying ΔSaO2 and Base-SaO2. Significant differences are denoted with asterisks. (B) Simulations. Calculated rCBV when normalizing tissue to simulated artery or vein for the same paradigms as in the experiments. Calculated values for all paradigms are further normalized to the rCBV attained with Gd (values above one are an overestimation relative to Gd). Simulated tissue CBV = 4%; Simulated arterial CBV = 100%; Simulated venous CBV = 100%.

FIGURE 7
www.frontiersin.org

Figure 7. rCBF Dependency on ΔSaO2 and Base-SaO2. rCBF (mL/100g/min) maps from representative subject axial slice where tissue is normalized to an arterial voxel in the MCA for five subsequent paradigms (left to right): standard Gd, large ΔSaO2 with high base-SaO2, medium ΔSaO2 with high base-SaO2, small ΔSaO2 with high base-SaO2, small ΔSaO2 with low base-SaO2.

Of note, the pattern of rCBV and rCBF estimation for different paradigms when normalizing tissue to artery was consistent in all subjects. When normalizing to vein, the pattern was consistent in 5 out of 6 subjects; in the sixth subject, the ΔSaO2 and base-SaO2 pattern appeared to reverse when normalizing to vein, although the quantifications were not significantly different.

Figure 8 elaborates further on the results described in Figures 5, 6 by showing voxel-wise regressions for each of the dOHb paradigms in a representative subject. The maps show all voxels within the GM and WM masks (Supplementary Figure 4). For these figures, the CBV maps from each dOHb paradigm were registered to the average dOHb paradigm, and then spatially smoothed with a 3 × 3 gaussian kernel. The results show that the slope of the fitted regression decreases as ΔSaO2 increases and base-SaO2 decreases, in agreement with the patterns in Figure 6. The calculated R2-score is very low in the paradigms with lower SNR but increases as the ΔSaO2 increases and base-SaO2 decreases, which is expected.

FIGURE 8
www.frontiersin.org

Figure 8. Voxel-wise Comparison of Experimentally Calculated rCBV (mL/100g) From dOHb Paradigms. (A) Large ΔSaO2 dOHb vs. average dOHb. (B) Medium ΔSaO2 dOHb vs. average dOHb. (C) Small ΔSaO2 dOHb vs. average dOHb. (D) Low base-SaO2 dOHb vs. average dOHb. Each plot displays data from all GM and WM voxels in a single subject. b is the regression slope and R2 is the coefficient of determination. Note that for this analysis specifically, rCBV maps were spatially smoothed with a 3x3 gaussian filter. The average dOHb paradigm is simply an average of the small, medium, and large ΔSaO2 paradigms.

Figure 9 displays simulation and experimental results for MTT calculated with dOHb and Gd. Surprisingly, for the experimental data, the estimated MTT values were much greater when using dOHb than when using Gd contrast. This was observed in all dOHb paradigms for all subjects, as displayed in Supplementary Table 1. While absolute values of calculated MTT differed in GM and WM between Gd and dOHb, the ratio of GM to WM was more consistent, regardless of the applied contrast agent or paradigm. The simulations show the same pattern as the experimental data, with dOHb contrast resulting in higher calculated MTT than Gd. We noted that this is especially the case when the singular values are thresholded in deconvolution, a necessary step in removing noise prior to rCBF and MTT estimation as described in the Methods. When thresholded, a dOHb bolus of longer duration results in a higher MTT estimation than a dOHb or Gd bolus of shorter duration. In sum, thresholding the singular values for deconvolution, a necessary step in the experimental data analysis, and applying a bolus of longer duration both result in a greater overestimation of the MTT.

FIGURE 9
www.frontiersin.org

Figure 9. MTT Dependency on Bolus Duration. (A) Validations. MTT maps from a representative subject when exposed to a hypoxic bolus of roughly 30 s duration vs. a Gd bolus of roughly 10 s duration. (B) Simulations. Calculated MTT with either no thresholding (solid) or standard deconvolution thresholding (dotted) for dOHb and Gd contrast (no noise). Simulated dOHb bolus is either longer (dark gray) or shorter (light gray) to match the duration of a standard Gd bolus (black). The ground truth MTT is represented by the black dotted line (3 s)—anything above is an overestimation of the true MTT.

Discussion

DSC MRI is widely used to determine perfusion maps in cerebral tissue. In the current study, we extend previous investigations on the accuracy and precision of perfusion quantification using Gd (references in Calamante, 2013) to the lower susceptibility range, using dOHb as contrast agent. Specifically, we investigate the influence of the following parameters on perfusion quantification: the susceptibility of contrast agent, the amount of contrast agent, baseline oxygen saturation, duration of the bolus, and choice of reference voxel (i.e., artery or vein). To address these questions, we developed a signal framework for DSC MRI that incorporates signal contributions from intravascular and extravascular water proton spins at 3T for arterial, venous, and cerebral tissue voxels. We compared the simulated predictions to experimental data obtained at 3T on six healthy human subjects administered Gd and dOHb boluses in separate acquisitions. Our major findings in the simulations and experimental validations are as follows:

• Reducing the simulated arterial CBV from a higher value (80-100%) to a moderate value (40-60%) reduces the estimated values of tissue rCBV and rCBF.

• Reducing the baseline SaO2, increasing the susceptibility of the applied contrast agent (Gd vs. dOHb), and/or increasing the ΔSaO2 increases inter-subject precision and reduces the estimated values of tissue rCBV and rCBF.

• Normalizing tissue to a venous rather than arterial relaxation rate time course increases the accuracy of the calculated rCBV and decreases quantification dependency on ΔSaO2, base-SaO2, and normalizing voxel CBV.

• Shortening the bolus duration increases the accuracy and reduces the estimated values of tissue MTT.

Please note that the aim of this work was not to optimize imaging protocols in DSC MRI; it was similarly not our goal to determine the preferred contrast agent for imaging. Rather, we used our model and experimental validations to study how different susceptibility parameters affect the precision and accuracy of perfusion quantification.

Signal modeling framework

The signal model applied in this work was previously used to simulate fMRI signal changes in response to neuronally-driven blood oxygenation changes (Uludag et al., 2009). Given that the dOHb bolus used in this study is an externally driven susceptibility change of the same contrast agent as in fMRI, we applied the same biophysical model to DSC MRI. We generalized the framework to work for Gd contrast as well. To do so, the relationship between Gd concentration and susceptibility changes had to be determined. In doing so, we found that at 3T, a 1 mM increase of Gd is approximately equivalent in susceptibility to a 24.6% increase in dOHb (Eq. 8.2). Thus, Gd was modeled in the same way as dOHb, with differences only in the input bolus shape, duration, and susceptibility value (Supplementary Figures S1, S6). Consequently, any difference in calculated physiological parameters (e.g., CBV, CBF and MTT) can be attributed to differences in susceptibility and bolus properties.

Our model calculates the DSC MRI signal response to contrast agent in tissue, which results from both intravascular and extravascular relaxation rate changes in separate vascular compartments: artery, vein, capillary, arteriole, and venule. This has theoretical advantages over previous in vivo and simulation work which models DSC MRI through a “bulk model”, wherein there is no theoretical separation between extravascular and intravascular compartments (e.g., see Calamante et al., 2000; Patil and Johnson, 2011; Chappell et al., 2015). The contrast agent's influence on the proton spins in these compartments is not identical and should be treated as separate signal contributions (Zhao et al., 2007; Blockley et al., 2008; Uludag et al., 2009). The model described in Kiselev et al. incorporates extravascular and intravascular signal contributions and notes differences in the calculated and ground truth CBV and CBF of simulated voxels (Kiselev, 2001). A major difference between our work and Kiselev's is that the latter assumes an intravascular relaxation rate that is linearly dependent and an extravascular relaxation rate that is quadratically dependent on contrast concentration—our model assumes a linear extravascular and quadratic intravascular relationship.

Numerous experimental and theoretical papers report quadratic dependencies between contrast agent (dOHb or Gd) and the intravascular relaxation rate (Jensen and Chandra, 2000; Silvennoinen et al., 2003; van Osch et al., 2003; Zhao et al., 2007; Blockley et al., 2008). As a linear intravascular relaxation rate in our model would result in little to no dependency of the perfusion metrics on the susceptibility value, our experimental results are also a strong argument for a non-linear relationship between the contrast agent dose and intravascular relaxation rate. This is because the effects of contrast dosage, for example, would mostly cancel out in a linear relationship, resulting in little dependency. Regarding the extravascular relaxation rate, there is strong agreement based on rigorous Monte Carlo simulations that it is linearly related to the concentration of contrast agent (Boxerman et al., 1995; Kjølby et al., 2006; Uludag et al., 2009).

To the best of our knowledge, two studies conducted by Kjølby et al. (2006, 2009) are the only ones to account for linear extravascular and quadratic intravascular relaxation rates in DSC MRI simulations. The paper in 2006 (Kjølby et al., 2006) focused on deriving tissue relaxivity in response to Gd and did not quantify perfusion. The paper in 2009 (Kjølby et al., 2009) set out to quantify perfusion in response to changes in AIF partial volume. However, this work neglected separate vascular compartments in tissue. Therefore, even though components of our model have been proposed in some studies, the integration of all these components is novel, that is, incorporating quadratic intravascular and linear extravascular components from separate vascular compartments. In addition, the current work has investigated perfusion quantification more systematically and for a wider range of properties than previous studies.

Furthermore, it was not our aim to investigate perfusion quantification dependent on specific tissue properties, such as vessel branching, cellular packing, or vessel permeability; we instead focused on properties related to the applied contrast agents and intravascular volume fraction. Thus, we refer the reader to other DSC MRI signal models which have been developed to understand the effects of various tissue parameters (i.e., vessel heterogeneity, contrast agent extraversion) on signal and relaxation rate (Quarles et al., 2009; Semmineh et al., 2014; Digernes et al., 2017).

Signal scaling differs between Gd and dOHb

While Gd-induced signal changes are larger than those induced by dOHb, signal scaling is not homogeneous in all areas of the brain (Figure 4). When the ΔSdOHb map is divided by the ΔSGd map, termed the ΔS ratio, we found a significant scaling difference between arterial, venous, GM, and WM regions. Specifically, large vessel voxels, on average, show a larger ΔS ratio than tissue voxels in both the simulations and experiments. The likely reason for this is due to non-linear signal scaling. When signal in the large vessels approaches zero, also known as saturation (Ellinger et al., 2000), signal change becomes non-linear with respect to dosage (doubling the dosage or CBV results in much smaller signal changes). Given that signal in response to Gd contrast can approach the noise floor, something that does not occur with a typical hypoxic bolus (Poublanc et al., 2021; Vu et al., 2021), there is an inherent difference in how signal is scaled between contrast agents with high vs. low susceptibility in large vessels. The scaling is more linear for dOHb and more non-linear for Gd, with respect to the amount of contrast agent.

Although to a much smaller degree, there is also a scaling difference between the GM and WM voxels, wherein the GM ΔS ratio between Gd and dOHb induced changes is lower than the WM ΔS ratio. This effect is observed in the experiments but not in the simulations (Figure 4). One potential reason for the observed difference between GM and WM might be macrovascular contamination (Chappell et al., 2015), particularly the so-called blooming effect (reviewed by Willats and Calamante, 2013). It is known that the administration of Gd results in signal contamination in the cortical GM, where magnetic field disturbances from large vessels extends to the surrounding tissue (reviewed by Willats and Calamante, 2013). Assuming that these vessels are large arteries, this effect would be largely scaled down when using dOHb contrast due to the far lower signal arising from the arterial vessels on the cortical surface (refer to Figures 2, 3). Further work would be required to isolate for and identify the magnitude of the blooming effect when using Gd and dOHb. In summary, the signal change induced by Gd is larger than that induced by dOHb, but the scaling is not uniform throughout the brain, confirmed both in the experiments and simulations. This is a novel finding, as no previous work has compared signal change maps obtained from contrast agents with vastly different molar susceptibilities.

Finally, the ΔS ratio in arterial voxels is lower than that in venous voxels for both the simulations and experimental data. The reason for this is more complicated than the prior findings and can best be understood by referring to the simulation findings in Figure 3A. Here, although results are displayed for ΔR2,max* as opposed to ΔSmax, the same rationale applies. The difference between Gd and dOHb is magnified in arterial voxels and is much less apparent in venous voxels—the division of ΔSmax, dOHb by ΔSmax, Gd therefore yields much lower values in arterial voxels. The mathematical reasoning for the difference between artery and vein will be discussed further in the following sections.

Calculated rCBV and rCBF depend on simulated CBV of normalizing voxel, ΔSaO2, and base-SaO2

The simulations and experiments show a dependence of rCBV, rCBF, and MTT on user-controlled parameters, such as the type of contrast agent, choice of normalizing voxel, ΔSaO2, base-SaO2, and bolus duration. Figure 3 and Supplementary Figure 5 show that the area and magnitude of the arterial bolus relaxation curve at 3T do not increase linearly as the simulated blood volume increases, but rather increase and then decrease at higher simulated CBV occupancy. This counterintuitive pattern is also observed in previous work with Gd, where simulated arterial voxels with 80% CBV showed a greater signal change than those with 100% CBV (Kjølby et al., 2009). It is the predominant intravascular component of simulated arterial and venous voxels that results in a non-linearity in CBV dependence (Supplementary Figure 7); for simulated tissue voxels, this non-linearity is not observed due to the predominant linear extravascular relaxation rate. As such, the model predicts that normalizing tissue voxels to an arterial voxel with high blood volume results in higher calculated rCBV values than when normalized to an arterial voxel with an intermediate blood volume (Supplementary Figure 7).

As shown in Figures 58 and Supplementary Table 1, rCBV and rCBF attained with dOHb becomes more accurate and decreases as the bolus's ΔSaO2 increases and base-SaO2 decreases. Experimentally, this pattern should be more predominant when tissue is normalized to an arterial voxel with a high CBV, as from the theory and simulations, it is primarily the intravascular quadratic relaxation rate that is responsible for these dependencies (Supplementary Figure 7). The concept of the intravascular space contributing to perfusion overestimation is also acknowledged in previous works (Kjølby et al., 2009; Wirestam et al., 2010). As mentioned above, we show that increasing the susceptibility of contrast agent decreases the calculated rCBV and rCBF, without the confounds of saturation or partial volume errors in the case of our simulations. This pattern is more evident when using a lower susceptibility contrast agent, such as dOHb, which likely explains why previous researchers investigating the dosage of Gd found only a small effect of dosage on quantification (Alger et al., 2009; Wirestam et al., 2010).

Ultimately, the large deviations in the estimation of rCBV and rCBF between different dOHb and Gd paradigms are expected in accordance with the simulations. Of note, the rCBF values obtained in this work overestimate those from previous work using dOHb, which in turn overestimate physiological values determined by PET (Knutsson et al., 2010; Vu et al., 2021). This is similarly not unexpected as previous work applied a different experimental paradigm, which we now know from our simulations and validations will lead to a differing calculation of rCBF.

As an additional note, Bleeker et al. showed that it is beneficial to select an AIF adjacent to the intravascular space of an artery (Bleeker et al., 2009). Our work supports this claim for two reasons: firstly, our simulations show that normalizing tissue to an AIF with a low to intermediate simulated CBV results in higher accuracy of the calculated rCBV; secondly, our simulations reveal a lower dependence of the calculated rCBV on ΔSaO2 and base-SaO2 when tissue is normalized to a less intravascular AIF (Supplementary Figure 7).

Calculated rCBV depends on the choice of artery or vein for tissue normalization

From a clinical perspective, it might be beneficial to normalize tissue signal to a venous as opposed to arterial voxel, as the larger radius of the venous blood vessel results in less partial volume errors (reviewed by Willats and Calamante, 2013). Although there are still limitations to this type of analysis in DSC MRI (Knutsson et al., 2007), we sought to investigate how perfusion quantification depends on arterial vs. venous normalization.

In simulations and experiments, normalizing ΔR2,tissue*(t) to ΔR2,vein*(t) results in lower overestimations of rCBV, along with a reduced ΔSaO2 and base-SaO2 dependency in comparison to when ΔR2,tissue*(t) is normalized to ΔR2,artery*(t) (Figures 5, 6, and Supplementary Figure 5). As previously described, a lower base-SaO2 results in larger relaxation rate changes on account of the quadratic intravascular dependence on [Gd] and SO2, which then influences perfusion estimation. Thus, if a vein and an artery are provided the same ΔSaO2 or [Gd] and are simulated with the same CBV, it is clear from the previous results that the voxel with a lower base-SaO2, as is the case in veins, will have a larger relaxation rate change (specifically when the voxel is mostly intravascular and when the contrast agent has a lower susceptibility or concentration). This theoretical discovery likely explains the high venous to arterial relaxation rate ratio observed in previous studies using dOHb in DSC MRI (Poublanc et al., 2021; Vu et al., 2021). When tissue signal is then normalized to the venous voxel with a larger relaxation rate change relative to artery, and thus, larger area under curve, a lower and more accurate estimate of rCBV is obtained. When normalizing to a vein, we found that varying the ΔSaO2 and base-SaO2 through different paradigms had less of an effect on perfusion quantification in both the simulations and experimental validations. A reduced contrast dosage dependency and lower calculated rCBV when normalizing to vein as opposed to artery is also observed in previous experimental results using Gd (Wirestam et al., 2010).

Calculated MTT and rCBF depend on bolus duration

The experimental data reveal an overestimation of the calculated MTT in every participant when using dOHb in comparison to Gd for quantification (Figure 9). Vu et al. also reported high calculated MTT values when using dOHb as a contrast agent in a standard DSC MRI analysis (Vu et al., 2021).

The simulations reveal that bolus duration influences MTT and rCBF calculation (Figure 9). We also found that the thresholding of singular values causes a duration dependency—longer boluses resulted in the removal of more singular values from the AIF. It is known that as more singular values are removed, there is a greater underestimation of rCBF and, therefore, greater overestimation of MTT (Bjørnerud and Emblem, 2010). Thus, a Gd bolus with shorter duration and less removed singular values results in the calculation of a lower and more accurate MTT relative to the MTT calculated from a dOHb bolus. In our simulation work, when the dOHb bolus is shortened to the duration of Gd, the calculated MTT is more like that obtained with Gd (Figure 9). In previous work where Gd was doubled in concentration, an increase in the calculated MTT was found relative to a standard dose (Wirestam et al., 2010). Another study found that reducing the injection rate of Gd results in an increase to the calculated MTT (van Osch et al., 2003). As the bolus had increased in duration for these two studies, our findings support and help explain the observations.

Although noise was not simulated in our work, it is known that SNR is another major parameter influencing MTT calculation, wherein more singular values from the AIF matrix are removed when there is a lower SNR, resulting in a greater underestimation of rCBF and overestimation of MTT (Willats and Calamante, 2013). This may explain why the different dOHb paradigms in our work, although equal in duration, show an increasing estimation of MTT as the SNR decreases (Supplementary Table 1). Thus, a combination of low SNR and long bolus duration leads to higher estimations of MTT, supporting the finding that a dOHb bolus yields a higher MTT estimation than a Gd bolus.

Considerations for clinical implementation of dOHb and Gd

For dOHb contrast to attain an SNR like that of Gd in a single bolus, the oxygen saturation must drop below physiological tolerance (Supplementary Figure 1 and Supplementary Table 1). However, DSC using dOHb has several advantages (Poublanc et al., 2021; Vu et al., 2021). Firstly, as dOHb is an endogenous contrast agent, it does not share the safety limitations associated with Gd (Schlaudecker and Bernheisel, 2009; Voth et al., 2011; Kanda et al., 2015; Lord et al., 2018; Rogowska et al., 2018). As well, [dOHb] does not saturate, recirculate, or accumulate, as the blood oxygenation is reset to a pre-determined value in the lungs using the RespirAct®. Unlike Gd, the concentration of dOHb in the arterial blood can be determined from end-tidal gas analysis which may be useful as a precise AIF and aid in quantitative modeling (Ito et al., 2008; Fierstra et al., 2011). Finally, dOHb acquisitions can be repeated and averaged to improve the SNR (“Average dOHb” in Supplementary Table 1).

While the smaller hypoxic challenge (ΔSaO2 of 8%) administered in our study is equivalent to experiencing reduced SaO2 in moderately high altitudes (e.g., Aspen, Colorado), the larger challenge (ΔSaO2 of 24%) is closer to the SaO2 in the Everest Base Camp (Rojas-Camayo et al., 2018), albeit only for a short duration. Individuals may spend many months in these environments without additional oxygen or altitude sickness, minimizing the safety concern in our study where hypoxia only lasts for roughly 30 s.

Limitations and directions

We have not included T1 effects in our simulations. According to Calamante et al., a TR of 1.5 s and flip angle of 73° would result in, roughly, a 10% underestimation of CBF relative to that calculated in the absence of T1 effects, although this depends on the inflow rate (Calamante et al., 2007). This is to be considered in future studies but is not of major concern in our work as we are not investigating absolute values, but rather how relative quantification depends on various paradigms. It is not expected that the general patterns of the findings will change by including T1 effects, especially in the absence of contrast agent extraversion. As well, it has been shown in previous work that there is no significant T1 change in response to changes in dOHb, limiting the T1 effect concern to Gd (Thulborn et al., 1982). A further limitation of our study and DSC MRI theory in general is the correct simulation of [Gd]max. Studies have reported varying quantities for [Gd]max (Patil et al., 2013; Kellner et al., 2018; Lind et al., 2020). Simulating the exact value of [Gd]max is not a major concern in this work, but it should be noted that in vivo, a standard Gd dose likely results in a [Gd]max that is larger than the simulated 2 mM used in the majority of our simulations.

In the model, we do not include contrast agent extravasation, which is known from previous modeling and experiments to elicit additional T2* and T1 relaxation effects, limiting perfusion quantification accuracy (Quarles et al., 2005, 2009). Additional relaxation effects that result from leakage, dependent on the molecular size of contrast agent, can be implemented in future work that simulates pathological cerebral tissue. The limitation of neglecting this effect in our modeling is likely minimal as healthy subject tissue is assumed to have an intact blood-brain-barrier, with little to no leakage (Quarles et al., 2009).

Vasodilation from the hypoxic stimulus is another potential experimental limitation. Previous work showed that dropping SaO2 from 100 to 90% for a 20-min duration is the threshold for statistically significant cerebral vasodilation in humans (Gupta et al., 1997). According to more recent work, an SaO2 of ~70% results in increases to the CBF (Mardimae et al., 2012); however, the vascular response is delayed by about 3 minutes following hypoxia (Harris et al., 2013). We expect that there is no significant vasodilation in this work as the lowest SaO2 of 75% only lasts for approximately 30 s. Nevertheless, vasodilation would have to be accounted for in more drastic hypoxic paradigms, or in a diseased state, to avoid errors in quantification—this can easily be done with our model. Furthermore, for isocapnic hypoxia, as utilized in this study, no changes in the tissue oxygen metabolism or oxygen extraction fraction are expected (Ainslie et al., 2014; Gu and Jun, 2018).

As a future direction, dOHb contrast can be studied with a higher field strength scanner to further bolster SNR, which might be the best way to maximize the benefits of dOHb as long as the arterial signal does not become saturated. It is also known that at higher field strengths, the intravascular contribution in T2*-weighted acquisitions is heavily reduced relative to extravascular signal contribution (Uludag et al., 2009), which would lead to lower errors in perfusion quantification. It might be worthwhile, for a more detailed validation of the Gd to dOHb relationship determined in our modeling, to compare dOHb with low-dose Gd as contrast, where the low-dose of Gd better approximates the susceptibility induced by dOHb. We are currently in the planning stages of such a study. Another potential direction would be to investigate the use of dOHb contrast in clinical conditions, such as brain tumors, steno-occlusive vascular disease, and dementia. This could first be simulated with our framework and then validated with clinical scans. Finally, it would be useful to explore either a model-based DSC MRI analysis method or calibration approach using our DSC model, which should provide more consistent and accurate quantifications of perfusion.

Conclusion

We developed a DSC MRI signal framework to simulate cerebral tissue signal changes in response to dOHb and Gd contrast at 3T. We used this framework to investigate how various contrast agent parameters and tissue properties influence perfusion estimation. In doing so, perfusion quantification dependencies were discovered that are in agreement between the simulations and validations. We hope our model and findings will help inform future research and clinical protocols that employ dOHb and Gd as contrast agents.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by University Health Network Research Ethics Board. The patients/participants provided their written informed consent to participate in this study.

Author contributions

JS: involved in all areas of project, including project and methods design, simulation development, data collection, data analysis/interpretation, and writing. ES and JP involved in methods design, data collection, and writing. AM: involved in methods design, simulation development, and writing. OS: involved in data collection and writing. JD: involved in writing. JF and DM: involved in data interpretation and writing. KU: involved in all areas of project, including project and methods design, simulation development, data collection, data analysis/interpretation, and writing. All authors contributed to the article and approved the submitted version.

Funding

The research conducted in this paper was supported by funding from the Canadian Institutes of Health Research (CIHR). The study was also supported by the Institute for Basic Science, Suwon, Republic of Korea (IBS-R015-D1) to KU.

Acknowledgments

We would like to thank Dr. Jean J. Chen for her valuable feedback on a previous version of the manuscript.

Conflict of interest

The RespirAct device used in this study was developed by Thornhill Medical Inc., (TMI) a for profit spinoff from the University Health Network, University of Toronto, to enable cerebrovascular reactivity studies. It is not a commercial product and is made available to academic centers for certified research under ethics board approval. JF, OS, JD, and DM are appointees at the University of Toronto and employees of, and/or own shares in, TMI.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fnimg.2023.1048652/full#supplementary-material

Supplementary Figure 1. Susceptibility Relationship of Gd and dOHb in Simulations. Peak ΔSaO2 (%) and Gd (mM) as a function of peak susceptibility induced by contrast agent. The region shaded in blue is the typical peak susceptibility elicited in a simulated hypoxic bolus (Peak ΔSaO2 ≈ 10-25%). The region shaded in red is the typical peak susceptibility elicited in a simulated Gd bolus (Peak [Gd] ≈ 2-5 mM).

Supplementary Figure 2. Summary of Simulation Method. (A) Forward Model (top). Bolus time courses for Gd [mM] and dOHb [SaO2 (%)] are simulated; tissue, arterial, and venous properties (CBV, MTT, CBF) are then applied; the ΔR2EES* and ΔR2IVS* responses for water protons are modeled; finally, the resultant ΔR2*(t) time courses are fed into a signal model. (B) Inverse Solution (bottom). Tissue, arterial, and venous S(t) time courses are generated from Eq. 1—these are then converted into ΔR2*(t) (Eq. 9). The arterial input and tissue response curves are then used to calculate rCBV, rCBF, and the MTT.

Supplementary Figure 3. Visual Illustration of Workflow for Signal Model. Parameters, in red, are based on the values in Table 2. This process was used to simulate the signal and relaxation time courses for arterial, tissue, and venous voxels.

Supplementary Figure 4. GM, WM, and Large Vessel Masks Created in Functional Space. Red. Large vessels. Blue. GM. White. WM. For WM, voxels containing values within the lowest 15–20% of the ΔS data range were kept; for GM, the highest 75% were kept. The maps were binarized and linearly registered to the functional space of each dOHb paradigm. GM and WM were further masked with a thresholded εt map (voxels with values in the lowest 10% of data range were kept in mask) to remove noisy voxels in peripheral brain tissue, skull base, vein, and artery. The GM and WM regions were finally masked with a thresholded Gd MTT map (lowest 70% of robust range voxels were kept in mask) to remove voxels in the CSF. Large vessel mask (only used for results in Figure 4) generated by taking the highest 50% of ΔSGd or ΔSdOHb robust range.

Supplementary Figure 5. Simulated ΔR2*(t) for dOHb and Gd Contrast. (A) ΔR2*(t) for arterial voxel with varying intravascular space in response to dOHb. (B) Same for Gd (bottom). (C) ΔR2*(t) for tissue voxel with varying intravascular space in response to dOHb. (D) Same for Gd (bottom). (E) ΔR2*(t) for venous voxel with varying intravascular space in response to dOHb. (F) Same for Gd (bottom). Note that tissue curves (MTT = 3 s) are convolved with a more quickly decaying exponential than venous curves (MTT = 4 s).

Supplementary Figure 6. Perfusion Ratios as a Function of Peak Susceptibility Induced by Contrast Agent. The region shaded in blue is the typical peak susceptibility elicited in a hypoxic dOHb bolus (Peak ΔSaO2 ≈ 10-25%). The region shaded in red is the typical peak susceptibility elicited in a Gd bolus (Peak [Gd] ≈ 2-5 mM). (A) rCBV (%) as a function of peak susceptibility induced by contrast agent. Simulated artery CBV = 100%; simulated tissue CBV = 4%. (B) Ratio of arterial to venous relaxation rate time integral as a function of peak susceptibility induced by contrast agent. Simulated arterial and venous CBV = 100%.

Supplementary Figure 7. dOHb rCBV as a Function of Simulated Arterial or Venous CBV (%). rCBV accuracy (y-axis) for tissue (CBV = 4%) when normalized to artery (or vein) of varying simulated CBV (x-axis). rCBV accuracy calculated as rCBVcalculated divided by rCBVsimulated. All values above 1 (dotted line) overestimate the rCBVsimulated. (A) Three scenarios are described in this figure: only extravascular component of signal model used in simulated artery (black plus); only intravascular component of signal model used in simulated artery (dark gray square); full signal model used in simulated artery (light gray diamond). (B). Tissue data normalized to AIF (black) or VOF (gray). Four scenarios are described in this figure: large ΔSaO2 with high base-SaO2 (diamond), medium ΔSaO2 with high base-SaO2 (square), small ΔSaO2 with high base-SaO2 (plus), small ΔSaO2 with low base-SaO2 (triangle).

Supplementary Figure 8. Gd and dOHb Experimental Relaxation Rate Time Courses. (Top) Gd time course for a single bolus. (Bottom) dOHb time course for three boluses. Red represents the AIF; blue represents the VOF; dark gray represents gray matter; light gray represents white matter.

Supplementary Table 1. Summary of Perfusion Data Collected with dOHb and Gd Contrast Agents. Average and standard deviations are pooled from a total of six healthy participants (age 22–60; one female). Low baseline dOHb is only pooled from five participants due to motion.

Supplementary Table 2. Table of Parameters.

Footnotes

1. ^As the susceptibility values change with time (i.e., due to changes in blood oxygenation or in contrast agent concentration), the relaxation rates, MRI signal contributions, and the total signal are also a function of time. However, for simplification, we omit “time” in this and the following equations in the forward model section.

2. ^In Uludag et al. (2009), SEES and SIVS were termed Sex and Sin, respectively.

References

Ainslie, P. N., Shaw, A. D., Smith, K. J., Willie, C. K., Ikeda, K., Graham, J., et al. (2014). Stability of cerebral metabolism and substrate availability in humans during hypoxia and hyperoxia. Clin. Sci. 126, 661–670. doi: 10.1042/CS20130343

PubMed Abstract | CrossRef Full Text | Google Scholar

Alger, J. R., Schaewe, T. J., Lai, T. C., Frew, A. J., Vespa, P. M., Etchepare, M., et al. (2009). Contrast agent dose effects in cerebral dynamic susceptibility contrast magnetic resonance perfusion imaging. J. Magn. Reson. Imaging 29, 52–64. doi: 10.1002/jmri.21613

PubMed Abstract | CrossRef Full Text | Google Scholar

Bandettini, P. A., Wong, E. C., Hinks, R. S., Tikofsky, R. S., Hyde, J. S. (1992). Time course EPI of human brain function during task activation. Magnet. Reson. Med. 25, 390–397. doi: 10.1002/mrm.1910250220

PubMed Abstract | CrossRef Full Text | Google Scholar

Bjørnerud, A., Emblem, K. E. (2010). A fully automated method for quantitative cerebral hemodynamic analysis using DSC–MRI. J. Cereb. Blood. Flow. Metab 30, 1066–1078. doi: 10.1038/jcbfm.2010.4

PubMed Abstract | CrossRef Full Text | Google Scholar

Bleeker, E. J., van Buchem, M. A., van Osch, M. J. (2009). Optimal location for arterial input function measurements near the middle cerebral artery in first-pass perfusion MRI. J. Cereb. Blood. Flow. Metab 29, 840–852. doi: 10.1038/jcbfm.2008.155

PubMed Abstract | CrossRef Full Text | Google Scholar

Blockley, N. P., Jiang, L., Gardener, A. G., Ludman, C. N., Francis, S. T., Gowland, P. A., et al. (2008). Field strength dependence of R1 and R relaxivities of human whole blood to prohance, vasovist, and deoxyhemoglobin. Magnet. Reson. Med. 60, 1313–1320. doi: 10.1002/mrm.21792

PubMed Abstract | CrossRef Full Text | Google Scholar

Boxerman, J. L., Hamberg, L. M., Rosen, B. R., Weisskoff, R. M. (1995). MR contrast due to intravascular magnetic susceptibility perturbations. Magn. Reson. Med. 34, 555–566. doi: 10.1002/mrm.1910340412

PubMed Abstract | CrossRef Full Text | Google Scholar

Calamante, F. (2013). Arterial input function in perfusion MRI: a comprehensive review. Prog. Nucl. Magn. Reson. Spectrosc 74, 1–32. doi: 10.1016/j.pnmrs.2013.04.002

PubMed Abstract | CrossRef Full Text | Google Scholar

Calamante, F., Gadian, D. G., Connelly, A. (2000). Delay and Dispersion Effects in Dynamic Susceptibility Contrast MRI: Simulations Using Singular Value Decomposition. Magn. Reson. Med. 44, 466–473. doi: 10.1002/1522-2594(200009)44:3<466::AID-MRM18>3.0.CO;2-M

PubMed Abstract | CrossRef Full Text | Google Scholar

Calamante, F., Vonken, E. P. A., van Osch, M. J. P. (2007). Contrast agent concentration measurements affecting quantification of bolus-tracking perfusion MRI. Magnet. Reson. Med. 58, 544–553. doi: 10.1002/mrm.21362

PubMed Abstract | CrossRef Full Text | Google Scholar

Chappell, M. A., Mehndiratta, A., Calamante, F. (2015). Correcting for large vessel contamination in dynamic susceptibility contrast perfusion MRI by extension to a physiological model of the vasculature: Correcting Contamination in DSC Perfusion. Magn. Reson. Med. 74, 280–290. doi: 10.1002/mrm.25390

PubMed Abstract | CrossRef Full Text | Google Scholar

Copen, W. A., Schaefer, P. W., Wu, O. (2011). MR Perfusion Imaging in Acute Ischemic Stroke. Neuroimaging. Clin. N. Am 21, 259–283. doi: 10.1016/j.nic.2011.02.007

PubMed Abstract | CrossRef Full Text | Google Scholar

Davenport, R. (1983). The Derivation of the Gamma-Variate Relationship for Tracer Dilution Curves. J. Nucl. Med. 24, 945–948.

PubMed Abstract | Google Scholar

Digernes, I., Bjørnerud, A., Vatnehol, S. A. S., Løvland, G., Courivaud, F., Vik-Mo, E., et al. (2017). A theoretical framework for determining cerebral vascular function and heterogeneity from dynamic susceptibility contrast MRI. J. Cereb. Blood. Flow. Metab 37, 2237–2248. doi: 10.1177/0271678X17694187

PubMed Abstract | CrossRef Full Text | Google Scholar

Ellinger, R., Kremser, C., Schocke, M. F. H., Kolbitsch, C., Griebel, J., Felber, S. R., et al. (2000). The impact of peak saturation of the arterial input function on quantitative evaluation of dynamic susceptibility contrast-enhanced MR studies. J. Comput. Assist. Tomogr. 24, 942–948. doi: 10.1097/00004728-200011000-00022

PubMed Abstract | CrossRef Full Text | Google Scholar

Fierstra, J., Machina, M., Battisti-Charbonney, A., Duffin, J., Fisher, J. A., Minkovich, L., et al. (2011). End-inspiratory rebreathing reduces the end-tidal to arterial PCO2 gradient in mechanically ventilated pigs. Intensive. Care. Med. 37, 1543–1550. doi: 10.1007/s00134-011-2260-y

PubMed Abstract | CrossRef Full Text | Google Scholar

Gu, C., Jun, J. C. (2018). Does Hypoxia Decrease the Metabolic Rate? Front. Endocrinol. 9,-668. doi: 10.3389/fendo.2018.00668

PubMed Abstract | CrossRef Full Text | Google Scholar

Gupta, A. K., Menon, D. K., Czosnyka, M., Smielewski, P., Jones, J. G. (1997). Thresholds for Hypoxic Cerebral Vasodilation in Volunteers. Anesth. Analg. 85, 817. doi: 10.1213/00000539-199710000-00018

PubMed Abstract | CrossRef Full Text | Google Scholar

Harris, A. D., Murphy, K., Diaz, C. M., Saxena, N., Hall, J. E., Liu, T. T., et al. (2013). Cerebral blood flow response to acute hypoxic hypoxia. NMR Biomed. 26, 1844–1852. doi: 10.1002/nbm.3026

PubMed Abstract | CrossRef Full Text | Google Scholar

Harris, G. J., Lewis, R. F., Satlin, A., English, C. D., Scott, T. M., Yurgelun-Todd, D. A., et al. (1998). Dynamic Susceptibility Contrast MR Imaging of Regional Cerebral Blood Volume in Alzheimer Disease: A Promising Alternative to Nuclear Medicine. AJNR Am. J. Neuroradiol. 19, 1727–1732.

PubMed Abstract | Google Scholar

Ibaraki, M., Ito, H., Shimosegawa, E., Toyoshima, H., Ishigame, K., Takahashi, K., et al. (2007). Cerebral vascular mean transit time in healthy humans: a comparative study with PET and dynamic susceptibility contrast-enhanced MRI. J. Cereb. Blood Flow Metab. 27, 404–413. doi: 10.1038/sj.jcbfm.9600337

PubMed Abstract | CrossRef Full Text | Google Scholar

Ito, S., Mardimae, A., Han, J., Duffin, J., Wells, G., Fedorko, L., et al. (2008). Non-invasive prospective targeting of arterial P(CO2) in subjects at rest. J. Physiol. 586, 3675–3682. doi: 10.1113/jphysiol.2008.154716

PubMed Abstract | CrossRef Full Text | Google Scholar

Jahng, G.-H., Li, K-L., Ostergaard, L., Calamante, F. (2014). Perfusion magnetic resonance imaging: a comprehensive update on principles and techniques. Korean J. Radiol. 15, 554–577. doi: 10.3348/kjr.2014.15.5.554

PubMed Abstract | CrossRef Full Text | Google Scholar

Jensen, J. H., Chandra, R. (2000). NMR relaxation in tissues with weak magnetic inhomogeneities. Magnet. Reson. Med. 44, 144–156. doi: 10.1002/1522-2594(200007)44:1&lt;144::AID-MRM21&gt;3.0.CO;2-O

PubMed Abstract | CrossRef Full Text | Google Scholar

Kanda, T., Fukusato, T., Matsuda, M., Toyoda, K., Oba, H., Kotoku, J., et al. (2015). Gadolinium-based contrast agent accumulates in the brain even in subjects without severe renal dysfunction: evaluation of autopsy brain specimens with inductively coupled plasma mass spectroscopy. Radiology 276, 228–232. doi: 10.1148/radiol.2015142690

PubMed Abstract | CrossRef Full Text | Google Scholar

Kellner, E., Mader, I., Reisert, M., Urbach, H., Kiselev, V. G. (2018). Arterial input function in a dedicated slice for cerebral perfusion measurements in humans. MAGMA 31, 439–448. doi: 10.1007/s10334-017-0663-7

PubMed Abstract | CrossRef Full Text | Google Scholar

Kiselev, V. G. (2001). On the theoretical basis of perfusion measurements by dynamic susceptibility contrast MRI. Magnet. Reson. Med. 46, 1113–1122. doi: 10.1002/mrm.1307

PubMed Abstract | CrossRef Full Text | Google Scholar

Kjølby, B. F., Mikkelsen, I. K., Pedersen, M., Østergaard, L., Kiselev, V. G. (2009). Analysis of partial volume effects on arterial input functions using gradient echo: A simulation study. Magnet. Reson. Med. 61, 1300–1309. doi: 10.1002/mrm.21849

PubMed Abstract | CrossRef Full Text | Google Scholar

Kjølby, B. F., Østergaard, L., Kiselev, V. G. (2006). Theoretical model of intravascular paramagnetic tracers effect on tissue relaxation. Magnet. Reson. Med. 56, 187–197. doi: 10.1002/mrm.20920

PubMed Abstract | CrossRef Full Text | Google Scholar

Knutsson, L., Börjesson, S., Larsson, E-.M, Risberg, J., Gustafson, L., et al. (2007). Absolute quantification of cerebral blood flow in normal volunteers: Correlation between Xe-133 SPECT and dynamic susceptibility contrast MRI. J. Magnet. Reson. Imag. 26, 913–920. doi: 10.1002/jmri.21093

PubMed Abstract | CrossRef Full Text | Google Scholar

Knutsson, L., Ståhlberg, F., Wirestam, R. (2010). Absolute quantification of perfusion using dynamic susceptibility contrast MRI: pitfalls and possibilities. Magn. Reson. Mater. Phy 23, 1–21. doi: 10.1007/s10334-009-0190-2

PubMed Abstract | CrossRef Full Text | Google Scholar

Kwong, K. K., Belliveau, J. W., Chesler, D. A., Goldberg, I. E., Weisskoff, R. M., Poncelet, B. P., et al. (1992). Dynamic magnetic resonance imaging of human brain activity during primary sensory stimulation. Proc. Natl. Acad. Sci. U. S. A 89, 5675–5679. doi: 10.1073/pnas.89.12.5675

PubMed Abstract | CrossRef Full Text | Google Scholar

Law, M., Young, R. J., Babb, J. S., Peccerelli, N., Chheang, S., Gruber, M. L., et al. (2008). Gliomas: predicting time to progression or survival with cerebral blood volume measurements at dynamic susceptibility-weighted contrast-enhanced perfusion MR imaging. Radiology 247, 490–498. doi: 10.1148/radiol.2472070898

PubMed Abstract | CrossRef Full Text | Google Scholar

Lee, D., Le, T. T., Im, G. H., Kim, S. G. (2022). Whole-brain perfusion mapping in mice by dynamic BOLD MRI with transient hypoxia. J. Cereb. Blood. Flow. Metab. 42, 2270–2286. doi: 10.1177/0271678X221117008

PubMed Abstract | CrossRef Full Text | Google Scholar

Lind, E., Knutsson, L., Ståhlberg, F., Wirestam, R. (2020). Dynamic contrast-enhanced QSM for perfusion imaging: a systematic comparison of ΔR2*- and QSM-based contrast agent concentration time curves in blood and tissue. MAGMA 33, 663–676. doi: 10.1007/s10334-020-00831-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Lord, M. L., Chettle, D. R., Gräfe, J. L., Noseworthy, M. D., McNeill, F. E. (2018). Observed deposition of gadolinium in bone using a new noninvasive in vivo biomedical device: results of a small pilot feasibility study. Radiology 287, 96–103. doi: 10.1148/radiol.2017171161

PubMed Abstract | CrossRef Full Text | Google Scholar

Lu, H., Ge, Y. (2008). Quantitative evaluation of oxygenation in venous vessels using T2-Relaxation-Under-Spin-Tagging MRI. Magn. Reson. Med. 60, 357–363. doi: 10.1002/mrm.21627

PubMed Abstract | CrossRef Full Text | Google Scholar

Mardimae, A., Balaban, D. Y., Machina, M. A., Han, J. S., Katznelson, R., Minkovich, L. L., et al. (2012). The interaction of carbon dioxide and hypoxia in the control of cerebral blood flow. Pflugers. Arch. Eur. J. Physiol. 464, 345–351. doi: 10.1007/s00424-012-1148-1

PubMed Abstract | CrossRef Full Text | Google Scholar

Mehndiratta, A., Calamante, F., MacIntosh, B. J., Crane, D. E., Payne, S. J., Chappell, M. A., et al. (2014). Modeling the residue function in DSC-MRI simulations: Analytical approximation to in vivo data. Magnet. Reson. Med. 72, 1486–1491. doi: 10.1002/mrm.25056

PubMed Abstract | CrossRef Full Text | Google Scholar

Meier, P., Zierler, K. L. (1954). On the theory of the indicator-dilution method for measurement of blood flow and volume. J. Appl. Physiol 6, 731–744. doi: 10.1152/jappl.1954.6.12.731

PubMed Abstract | CrossRef Full Text | Google Scholar

Ogawa, S., Lee, T. M., Kay, A. R., Tank, D. W. (1990). Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc. Natl. Acad. Sci. U. S. A. 87, 9868–9872. doi: 10.1073/pnas.87.24.9868

PubMed Abstract | CrossRef Full Text | Google Scholar

Ogawa, S., Tank, D. W., Menon, R., Ellermann, J. M., Kim, S. G., Merkle, H., et al. (1992). Intrinsic signal changes accompanying sensory stimulation: functional brain mapping with magnetic resonance imaging. Proc. Natl. Acad. Sci. U. S. A. 89, 5951–5955. doi: 10.1073/pnas.89.13.5951

PubMed Abstract | CrossRef Full Text | Google Scholar

Østergaard, L., Weisskoff, R. M., Chesler, D. A., Gyldensted, C., Rosen, B. R. (1996). High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: Mathematical approach and statistical analysis. Magn. Reson. Med. 36, 715–725. doi: 10.1002/mrm.1910360510

PubMed Abstract | CrossRef Full Text | Google Scholar

Patil, V., Jensen, J. H., Johnson, G. (2013). Intravascular contrast agent T2* relaxivity in brain tissue. NMR. Biomed 26, 392–399. doi: 10.1002/nbm.2876

PubMed Abstract | CrossRef Full Text | Google Scholar

Patil, V., Johnson, G. (2011). An improved model for describing the contrast bolus in perfusion MRI. Med. Phys 38, 6380–6383. doi: 10.1118/1.3658570

PubMed Abstract | CrossRef Full Text | Google Scholar

Pauling, L., Coryell, C. (1936). The Magnetic Properties and Structure of Hemoglobin, Oxyhemoglobin and Carbonmonoxyhemoglobin. PNAS 22, 210–216. doi: 10.1073/pnas.22.4.210

PubMed Abstract | CrossRef Full Text | Google Scholar

Poublanc, J., Sobczyk, O., Shafi, R., Sayin, E. S., Schulman, J., Duffin, J., et al. (2021). Perfusion MRI using endogenous deoxyhemoglobin as a contrast agent: Preliminary data. Magnet. Reson. Med. 86, 3012–3021. doi: 10.1002/mrm.28974

PubMed Abstract | CrossRef Full Text | Google Scholar

Qiao, P-.G, Han, C., Zuo, Z-W., Wang, Y-., Pfeuffer, J., et al. (2017). Clinical assessment of cerebral hemodynamics in Moyamoya disease via multiple inversion time arterial spin labeling and dynamic susceptibility contrast-magnetic resonance imaging: A comparative study. J. Neuroradiol. 44, 273–280. doi: 10.1016/j.neurad.2016.12.006

PubMed Abstract | CrossRef Full Text | Google Scholar

Quarles, C. C., Gochberg, D. F., Gore, J. C., Yankeelov, T. E. (2009). A Theoretical Framework to Model DSC-MRI Data Acquired in the Presence of Contrast Agent Extravasation. Phys. Med. Biol. 54, 5749. doi: 10.1088/0031-9155/54/19/006

PubMed Abstract | CrossRef Full Text | Google Scholar

Quarles, C. C., Ward, B. D., Schmainda, K. M. (2005). Improving the reliability of obtaining tumor hemodynamic parameters in the presence of contrast agent extravasation. Magnet. Reson. Med. 53, 1307–1316. doi: 10.1002/mrm.20497

PubMed Abstract | CrossRef Full Text | Google Scholar

Rogowska, J., Olkowska, E., Ratajczyk, W., Wolska, L. (2018). Gadolinium as a new emerging contaminant of aquatic environments. Environ. Toxicol. Chem. 37, 1523–1534. doi: 10.1002/etc.4116

PubMed Abstract | CrossRef Full Text | Google Scholar

Rojas-Camayo, J., Mejia, C. R., Callacondo, D., Dawson, J. A., Posso, M., Galvan, C. A., et al. (2018). Reference values for oxygen saturation from sea level to the highest human habitation in the Andes in acclimatised persons. Thorax 73, 776–778. doi: 10.1136/thoraxjnl-2017-210598

PubMed Abstract | CrossRef Full Text | Google Scholar

Rosen, B. R., Belliveau, J. W., Vevea, J. M., Brady, T. J. (1990). Perfusion imaging with NMR contrast agents. Magn. Reson. Med. 14, 249–265. doi: 10.1002/mrm.1910140211

PubMed Abstract | CrossRef Full Text | Google Scholar

Schlaudecker, J. D., Bernheisel, C. R. (2009). Gadolinium-associated nephrogenic systemic fibrosis. Am. Fam. Physician 80, 711–714.

Google Scholar

Semmineh, N. B., Xu, J., Boxerman, J. L., Delaney, G. W., Cleary, P. W., Gore, J. C., et al. (2014). An Efficient Computational Approach to Characterize DSC-MRI Signals Arising from Three-Dimensional Heterogeneous Tissue Structures. PLoS ONE 9, e84764. doi: 10.1371/journal.pone.0084764

PubMed Abstract | CrossRef Full Text | Google Scholar

Silvennoinen, M. J., Clingman, C. S., Golay, X., Kauppinen, R. A., van Zijl, P. C. M. (2003). Comparison of the dependence of blood R2 and R on oxygen saturation at 1.5 and 4.7 Tesla. Magnet. Reson. Med. 49, 47–60. doi: 10.1002/mrm.10355

PubMed Abstract | CrossRef Full Text | Google Scholar

Sobczyk, O., Crawley, A. P., Poublanc, J., Sam, K., Mandell, D. M., Mikulis, D. J., et al. (2016). Identifying Significant Changes in Cerebrovascular Reactivity to Carbon Dioxide. AJNR. Am. J. Neuroradiol. 37, 818–824. doi: 10.3174/ajnr.A4679

PubMed Abstract | CrossRef Full Text | Google Scholar

Tanaka, Y., Nagaoka, T., Nair, G., Ohno, K., Duong, T. Q. (2011). Arterial spin labeling and dynamic susceptibility contrast CBF MRI in postischemic hyperperfusion, hypercapnia, and after mannitol injection. J. Cereb. Blood. Flow. Metab. 31, 1403–1411. doi: 10.1038/jcbfm.2010.228

PubMed Abstract | CrossRef Full Text | Google Scholar

Thompson, H. K., Starmer, C. F., Whalen, R. E., Mcintosh, H. D. (1964). Indicator transit time considered as a gamma variate. Circ. Res. 14, 502–515. doi: 10.1161/01.RES.14.6.502

PubMed Abstract | CrossRef Full Text | Google Scholar

Thulborn, K. R., Waterton, J. C., Matthews, P. M., Radda, G. K. (1982). Oxygenation dependence of the transverse relaxation time of water protons in whole blood at high field. Biochim. Biophys. Acta. 714, 265–270. doi: 10.1016/0304-4165(82)90333-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Tudorica, A., Li, H. F., Hospod, F., Delucia-Deranja, E., Huang, W., Patlak, C. S., et al. (2002). Cerebral blood volume measurements by rapid contrast infusion and T-weighted echo planar MRI. Magnet. Reson. Med. 47, 1145–1157. doi: 10.1002/mrm.10167

PubMed Abstract | CrossRef Full Text | Google Scholar

Uludag, K., Müller-Bierl, B., Ugurbil, K. (2009). An integrative model for neuronal activity-induced signal changes for gradient and spin echo functional imaging. NeuroImage 48, 150–165. doi: 10.1016/j.neuroimage.2009.05.051

PubMed Abstract | CrossRef Full Text | Google Scholar

van Osch, M. J. P., Vonken, E. P. A., Viergever, M. A., van der Grond, J., Bakker, C. J. G. (2003). Measuring the arterial input function with gradient echo sequences. Magn. Reson. Med. 49, 1067–1076. doi: 10.1002/mrm.10461

PubMed Abstract | CrossRef Full Text | Google Scholar

van Zijl, P. C., Eleff, S. M., Ulatowski, J. A., Oja, J. M., Ulu,g, A. M., Traystman, R. J., et al. (1998). Quantitative assessment of blood flow, blood volume and blood oxygenation effects in functional magnetic resonance imaging. Nat. Med. 4, 159–167. doi: 10.1038/nm0298-159

PubMed Abstract | CrossRef Full Text | Google Scholar

Voth, M., Rosenberg, M., Breuer, J. (2011). Safety of gadobutrol, a new generation of contrast agents: experience from clinical trials and postmarketing surveillance. Invest. Radiol. 46, 663–671. doi: 10.1097/RLI.0b013e3182218dc3

PubMed Abstract | CrossRef Full Text | Google Scholar

Vovenko, E. (1999). Distribution of oxygen tension on the surface of arterioles, capillaries and venules of brain cortex and in tissue in normoxia: an experimental study on rats. Pflugers. Arch 437, 617–623. doi: 10.1007/s004240050825

PubMed Abstract | CrossRef Full Text | Google Scholar

Vu, C., Chai, Y., Coloigner, J., Nederveen, A. J., Borzage, M., Bush, A., et al. (2021). Quantitative perfusion mapping with induced transient hypoxia using BOLD MRI. Magn. Reson. Med. 85, 168–181. doi: 10.1002/mrm.28422

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, D. J. J., Alger, J. R., Qiao, J. X., Gunther, M., Pope, W. B., Saver, J. L., et al. (2013). Multi-delay multi-parametric arterial spin-labeled perfusion MRI in acute ischemic stroke—Comparison with dynamic susceptibility contrast enhanced perfusion imaging. NeuroImage: Clin. 3, 1–7. doi: 10.1016/j.nicl.2013.06.017

PubMed Abstract | CrossRef Full Text | Google Scholar

Weisskoff, R., Zuo, C. S., Boxerman, J. L., Rosen, B. R. (1994). Microscopic susceptibility variation and transverse relaxation: Theory and experiment. Magnet. Reson. Med. 31, 601–610. doi: 10.1002/mrm.1910310605

PubMed Abstract | CrossRef Full Text | Google Scholar

Willats, L., Calamante, F. (2013). The 39 steps: evading error and deciphering the secrets for accurate dynamic susceptibility contrast MRI. NMR. Biomed 26, 913–931. doi: 10.1002/nbm.2833

PubMed Abstract | CrossRef Full Text | Google Scholar

Wirestam, R., Andersson, L., Ostergaard, L., Bolling, M., Aunola, J-., Lindgren, A., et al. (2000). Assessment of regional cerebral blood flow by dynamic susceptibility contrast MRI using different deconvolution techniques. Magnet. Reson. Med. 43, 691–700. doi: 10.1002/(SICI)1522-2594(200005)43:5<691::AID-MRM11>3.0.CO;2-B

PubMed Abstract | CrossRef Full Text | Google Scholar

Wirestam, R., Thilmann, O., Knutsson, L., Björkman-Burtscher, I. M., Larsson, E-.M, et al. (2010). Comparison of quantitative dynamic susceptibility-contrast MRI perfusion estimates obtained using different contrast-agent administration schemes at 3 T. Eur. J. Radiol. 75, e86–e91. doi: 10.1016/j.ejrad.2009.07.038

PubMed Abstract | CrossRef Full Text | Google Scholar

Woolrich, M. W., Jbabdi, S., Patenaude, B., Chappell, M., Makni, S., Behrens, T., et al. (2009). Bayesian analysis of neuroimaging data in FSL. Neuroimage 45, S173–186. doi: 10.1016/j.neuroimage.2008.10.055

PubMed Abstract | CrossRef Full Text | Google Scholar

Yablonskiy, D. A., Haacke, E. M. (1994). Theory of NMR signal behavior in magnetically inhomogeneous tissues: the static dephasing regime. Magn. Reson. Med. 32, 749–763. doi: 10.1002/mrm.1910320610

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhao, J. M., Clingman, C. S., Närväinen, M. J., Kauppinen, R. A., van Zijl, P. C. M. (2007). Oxygenation and hematocrit dependence of transverse relaxation rates of blood at 3T. Magn. Reson. Med. 58, 592–597. doi: 10.1002/mrm.21342

PubMed Abstract | CrossRef Full Text | Google Scholar

Zierler, K. L. (1962). Theoretical Basis of indicator-dilution methods for measuring flow and volume. Circ. Res. 10, 393–407. doi: 10.1161/01.RES.10.3.393

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: perfusion, DSC, gadolinium, simulations, deoxyhemoglobin

Citation: Schulman JB, Sayin ES, Manalac A, Poublanc J, Sobczyk O, Duffin J, Fisher JA, Mikulis D and Uludağ K (2023) DSC MRI in the human brain using deoxyhemoglobin and gadolinium—Simulations and validations at 3T. Front. Neuroimaging 2:1048652. doi: 10.3389/fnimg.2023.1048652

Received: 19 September 2022; Accepted: 01 February 2023;
Published: 20 February 2023.

Edited by:

Alberto L. Vazquez, University of Pittsburgh, United States

Reviewed by:

Valur T. Olafsson, Northeastern University, United States
Linda Knutsson, Kennedy Krieger Institute, United States

Copyright © 2023 Schulman, Sayin, Manalac, Poublanc, Sobczyk, Duffin, Fisher, Mikulis and Uludağ. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Jacob Benjamin Schulman, yes amFjb2Iuc2NodWxtYW4mI3gwMDA0MDttYWlsLnV0b3JvbnRvLmNh; Kâmil Uludağ, yes S2FtaWwuVWx1ZGFnJiN4MDAwNDA7cm1wLnVobi5jYQ==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.