Alignment options and robotics in total knee arthroplasty

  • Department of Orthopedic Surgery and Sports Medicine, Houston Methodist Hospital, Houston, TX, United States

Total knee arthroplasty is one of the most widely performed surgical procedures today. Its widespread popularity has helped drive innovation and improvement in the field. Different schools of thought have developed regarding the best way to perform this operation. Specifically, there are controversaries regarding the best alignment philosophy for the femoral and tibial components to optimize implant stability and longevity. Traditionally, neutral mechanical alignment has been the preferred alignment target. More recently, some surgeons advocate for alignment matching the patient's pre-arthritic anatomic alignment (“physiologic” varus or valgus), which has been described as kinematic alignment. Functional alignment is a hybrid technique that focuses on the coronal plane minimizing soft tissue releases. To date, there is no evidence demonstrating superiority of one method over another. There is growing popularity of robotic surgical techniques to improve accuracy of implant position and alignment. The choice of alignment philosophy is an important aspect of robotic assisted TKA surgery and has the potential to clarify the optimal alignment technique.

Introduction

Total knee arthroplasty (TKA) is a commonly performed orthopedic procedure with generally good outcomes. However, since its inception, differing schools of thought have developed regarding the best approach for reconstructing, aligning, and producing a balanced knee. In this paper, we will review the current concepts in TKA alignment as well as robotic-assisted surgical methods.

Mechanical alignment

Mechanical alignment is the approach whereby a knee is reconstructed with an intended goal of a hip-knee-ankle (HKA) angle of 180° by positioning both the tibial and femoral components perpendicular to the mechanical axis of each bone. In this method, coronal alignment is prioritized, and the same bone cuts are made in every knee, regardless of unique patient characteristics or deformity. This philosophy is backed by engineering principles that neutral limb mechanical alignment results in equal mechanical load on implants to avoid early prosthesis/polyethylene wear and tibial component loosening (1, 2).

Mechanical alignment is classically performed with a 5–6° valgus femur cut as measured with conventional instrumentation, which is most often intramedullary, that represents the anatomic axis. Newer methods to determine the mechanical axis include navigation, patient specific instrumentation, and robotic-assisted techniques. For the tibia, an extramedullary guide is commonly used to perform a perpendicular cut (90°) to the long axis (representing both the anatomic and mechanical axis). With the knee in flexion, rotation of the femoral component is established to make a posterior femoral cut that is parallel to the tibial cut and to create a rectangular flexion space. Femoral rotation targets can be achieved with measured resection or gap balancing techniques, which differ in the degree and timing of soft tissue releases (3).

Measured resection

Measured resection makes all bony cuts based on anatomic landmarks first, followed by soft tissue releases as needed to create symmetric flexion and extension gaps. This technique utilizes the posterior femoral condylar axis (PCA), anteroposterior (AP) axis, and/or trans-epicondylar axis (TEA) to gauge femoral rotation, and bony resections are typically made irrespective of soft tissue tension. Rotational positioning of the femoral component parallel to the TEA is associated with decreased femoral lift-off (4). Furthermore, the TEA does not significantly change with degenerative deformity or revision cases, and thus does not lose reliability as a landmark in these situations (3). However, some authors have noted that intraoperative identification of the TEA can be inconsistent even among experienced surgeons (57). The AP axis (Whiteside's line) line runs through the crest of the trochlea and intersects the middle of the posterior condylar axis. Although the AP axis can help determine femoral rotation, this line depends on normal trochlear anatomy. Thus, this method can prove inaccurate in the presence of femoral dysplasia or significant degenerative changes, which may result in internal rotation of the femoral component (8). The PCA can also be used to determine femoral rotation (arbitrarily sets external rotation at 3–5° relative to PCA), though this is reliant on normal anatomy and thus may not be appropriate for all cases (9, 10).

Gap balancing

Gap balancing utilizes the patient's soft tissue (medial/lateral collateral ligament) tension to determine bony resection. Proponents of a gap balancing technique argue this technique may lead to better coronal plane stability from extension to deep flexion.

Historically, gap balancing was achieved by first focusing on the flexion gap (11). In the flexion gap first method, a perpendicular proximal tibial resection, relative to the longitudinal axis of the tibia, is the first bony resection. A flexion gap tensioning jig is used to assure appropriate flexion gap symmetry (proximal tibia is parallel to the TEA) prior to making the posterior femoral condylar resection. Attention is then turned to the extension gap and tensioning devices that are set to the same tension level to the flexion gap are used to guide the distal femoral resection.

Alternatively, to minimize the risk of joint line alteration, a gap balancing technique that focuses on the extension gap first has been described and is now more commonly utilized (12). The distal femur resection (5–7° valgus) is performed with an intramedullary guide, followed by the proximal tibial resection perpendicular to the longitudinal tibial axis. A spacer block or tensioner can then be used to evaluate ligamentous structures that need to be released to achieve neutral limb alignment and a symmetric extension gap. Focus is then turned to the flexion gap. If the knee is well balanced in extension and the proximal tibial resection is accurate, the TEA should be parallel to the proximal tibia at 90 degrees of flexion. An implant specific tensioner, laminar spreader, or spacer block is used to equally tension the collateral ligaments with the knee at 90° of flexion to determine the rotation and magnitude of the posterior femoral condylar resection. This technique prioritizes an equal and rectangular flexion and extension gap.

Pitfalls

There is a paucity of evidence to suggest any significant clinical or functional difference between measured resection and gap balancing (13, 14). Gap balancing has been hypothesized to achieve better soft tissue balance due to this technique not relying on bone landmark identification with the knee in flexion (15). Furthermore, some studies attest that there is less femoral condylar lift-off (rectangular flexion gap) with the gap balancing approach than with the measured resection method throughout knee range of motion (16). Still, other studies indicate that there may be more soft tissue release, bone resection and joint line elevation with gap balancing, while the measured resection technique may have unequal medial/lateral (i.e., not rectangular) extension gaps (17).

Although restoration of neutral mechanical alignment is a reproducible goal during total knee arthroplasty, the philosophy of mechanical alignment is not without shortcomings. Making the distal femoral cut with 6° of valgus is based on this angle most often being in the range of 5–7° but does not account for patient specific anatomy. For example, shorter patients may need a distal femur resection of more than 6°, while 5° or less may be optimal for taller patients (18). Moreover, the intramedullary canal method to determine femoral anatomic alignment can result in substantial malalignment errors if minor malposition of the intramedullary rod occurs due variables such as canal diameter, starting hole orientation, rod length and width dimensions, or unique patient anatomy (19). Although asymmetric bearing wear is theoretically minimized with neutral mechanical alignment, component positioning to achieve this target may affect native femoral flexion axis, ligament tension, quadriceps function, and patella tracking (20, 21). Finally, if the patient has pre-arthritis varus or valgus, mechanical alignment will always require ligament balancing/releasing to achieve medial-lateral stability.

Kinematic (anatomic) alignment

Anatomic alignment is considered the precursor of kinematic alignment and intends to recreate the anatomic (6–7°) femoral valgus and anatomic (3°) tibial varus appreciated in normal knee anatomy (22). The rotation of the femoral component is set to the posterior condylar axis. Prior to navigated or robotic-assisted surgery, the reproducibility of performing precise valgus and varus cuts was called into question.

Kinematic alignment assumes the non-arthritic knee is perfectly balanced and attempts to restore the native anatomy of the knee to restore the native balance of the soft tissue envelope (23). The goal of kinematic alignment is to recreate or resurface the pre-arthritis anatomy to achieve symmetric balanced extension, medial pivot throughout arc, lateral laxity in flexion, and central patellar tracking. Typically, calipered bone resection is performed to take away only as much bone and cartilage as is replaced with metal and polyethylene (adjusted for bone/cartilage wear, saw blade thickness, and osteophytes). The goal is to perform no soft tissue releases and to achieve normal knee kinematics.

Kinematic alignment is supported with more recent data demonstrating that there is a large percentage of the normal population (32% of men, 17% of women) with constitutional varus and a HKA alignment of 3° or more at skeletal maturity (24). A particular benefit of kinematic alignment includes the avoidance of ligamentous releases to achieve medial-lateral balance in extension. In patients with constitutional varus and a varus joint line orientation, attempting to achieve a neutral mechanical axis may alter the natural joint line (25). In the sagittal plane, the tibial component is positioned such that the posterior tibial slope matches the flexion-extension plane of the native joint.

Kinematic alignment has been reported to achieve a neutrally aligned knee with no need for balancing the flexion gap (26) and, if properly executed, there are no differences in clinical or radiographic outcomes between this method and mechanically aligned TKAs (27). The reproducibility of a kinematically aligned TKA has been called into question by many arthroplasty surgeons and how well tolerated an oblique joint line is for a prosthetic knee is unknown and has been described as a possible downside (28). Inadvertently (via bone cut margin of error) creating too much tibial varus and/or femoral valgus can cause an overly oblique joint line. If natural varus is restored without sufficient external rotation of the femoral component, the patient may experience patellofemoral maltracking through increased contact stresses at the patellofemoral joint. Many surgeons do not agree that the flexion gap does not require balancing and are concerned that excessive lateral laxity in flexion can lead to flexion instability, femoral condylar lift-off, eccentric polyethylene loading, and premature loosening of the implants (29). Lastly, excessive posterior slope of the tibial component has been a reported risk factor for posterior subsidence of the tibial baseplate or excessive posterior edge wear of the polyethylene (30).

Functional alignment

The authors strive to gain the advantages and avoid the shortcomings of both the kinematic and mechanical alignment schools of thought to achieve a functionally aligned TKA. Functional alignment emphasizes the coronal plane, but components are positioned in a manner that compromises the soft tissue envelope of the knee as little as possible. Thus, the soft tissues dictate the plane of the joint line and soft tissue releases are typically not needed. Distal femur and proximal tibia resection can be achieved with valgus and varus correction, respectively.

In functional alignment, the surgeon aims to recreate the patient's pre-arthritic condition by using gap balancing techniques to first create a rectangular extension space with the knee at 0° by first resecting distal femur in 3–5° of valgus and then the proximal tibia. Less valgus (3–5°) of the femoral cut eliminates the typical lateral laxity with 5–7° of valgus. This is because the native knee has lateral laxity to allow for the screw home mechanism in extension and for the convexity of the lateral tibial plateau as the knee flexes. The prosthetic knee has neither of these, so less femoral valgus will tighten and stabilize the lateral side, producing balance of the medial and lateral collateral ligaments without release.

Preparation of the extension space is followed by creating a balanced flexion space at 90°. By aiming for a rectangular flexion gap, even when accounting for the margin of cutting error, the surgeon can reduce the risk of inadvertent internal femoral component rotation and creation of a pathologic Q-angle. By ensuring a balanced rectangular flexion gap, the surgeon can avoid further ligamentous release. This has clinical relevance as the addition of soft tissue releases after bony cuts may lead to worse patient-reported outcomes (31). Anecdotally, the authors have found that functional alignment consistently provides the patient with a stable and functional knee without ligamentous release in most cases (Figures 1, 2). The authors strive for a HKA angle of 2° of varus for all cases unless there is pre-operative valgus >2°, in which case the target HKA angle is 2° of valgus.

FIGURE 1
www.frontiersin.org

Figure 1. (A) example of pre-operative standing long-leg radiograph showing constitutional varus of the right lower extremity and varus malalignment of the left lower extremity secondary to end-stage osteoarthritis. (B) Post-operative standing long-leg leg radiograph using functional alignment to achieve slight varus alignment without any ligamentous releases.

FIGURE 2
www.frontiersin.org

Figure 2. (A) anteroposterior and (B) lateral radiograph of the same patient demonstrates a balanced left TKA.

Opponents of natural alignment argue that aligning the components in varus (or valgus) places excess load on half of the knee joint. Whether this varus alignment alters patient outcomes, results in polyethylene catastrophic failure, or tibial loosening is not fully elucidated (23, 32).

Robotics in total knee arthroplasty

Computer guidance has been adopted by arthroplasty surgeons to control alignment issues, to plan for complex deformity or prior hardware, to assist when patient body habitus alters landmarks, to determine flexion/extension gaps, and to increase accuracy of bony resection. Robotic systems in arthroplasty surgery include robotic arm-assisted procedures, robot-guided cutting jigs, and robotic milling using a range of passive to active systems (33). Passive robotic systems provide positional guidance for bony resection and implant positioning but lack haptic feedback. Semi-active robotic systems enable the surgeon to guide a robotic arm to perform bony resections within the confines of haptic restraint but require constant input from the surgeon. Active robotic systems operate independently to perform bony resections without real-time guidance after the surgical approach is performed and retractors are placed. There are image-dependent systems that rely on preoperative computed tomography (CT) or magnetic resonance imaging (MRI) scans. There are also imageless systems that require a more detailed intraoperative registration of bony anatomy (handheld probes create the topography of the knee) and joint kinematics. A benefit of these techniques is not using an intramedullary femoral guide to establish the anatomic axis for distal femoral resection.

Robotic-assisted techniques were developed to increase the precision of prosthetic alignment and have evolved to include soft tissue balancing algorithms (34). Tracking arrays are used to determine pre-resection flexion/extension and varus/valgus knee kinematics. In combination with a robotic ligament tensioning tool, this technology can accurately predict postoperative flexion and extension gaps prior to femoral resection (35). Ultimately, this may lead to a reduced need to perform soft tissue releases. With robotic-assisted techniques, the surgeon can achieve component position and non-neutral limb alignment targets more reproducibly with reduced alignment outliers (3638). These techniques can be utilized for varying alignment goals (39). This enables the surgeon to better plan both component positioning and final polyethylene insert thickness than when compared to using a manual technique (40).

Arguments against robotics include the potential need for advanced preoperative imaging, added surgical time, integration into surgical workflow, and unknown cost-effectiveness. The risk for periprosthetic pin site fracture has been greatly reduced by using unicortical (instead of bicortical) pins and smaller diameter pins for tracking arrays. The clinical benefit of robotic-assisted techniques is yet to be determined and they have not been universally adopted by surgeons (41, 42). Augmented reality is an emerging technology in arthroplasty surgery that involves a computer-generated image that is superimposed on the surgeon's field of vision. This enabling technology can be achieved through a helmet attachment that can improve planning and execution of a procedure while allowing the surgeon to keep their eyes on the patient (43).

Discussion

Several methods are described for surgically achieving a well-balanced knee, however, to date, no one school of thought has proven superior regarding patient outcomes. This is exemplified by a randomized controlled trial of simultaneous gap-balancing TKA and measured-resection TKA in 50 patients with bilateral varus alignment, in which there was no statistical difference in patient-reported or functional outcomes between the two methods of knee balancing (14). As previously stated, benefits and drawbacks exist for each described method of knee balancing. Not surprisingly, patient satisfaction may be able to be achieved with accuracy of measurable TKA variables within a precise range of measurements or techniques instead of a single, ultimate method or measurement. Preserving soft tissue stability, maintaining joint line orientation/height, maintaining HKA axis alignment within certain parameters, and installing implants within acceptable ranges are likely the most important factors for patient satisfaction and implant longevity.

Conclusion

Alignment, or malalignment, in TKA can be a potential cause of patient dissatisfaction. When considering the optimal alignment target for implanting a TKA prosthesis, the relationship between the prosthesis and the soft tissue envelope may be more important than neutral mechanical alignment. Functional alignment represents a compromise between mechanical and kinematic alignment, and we believe represent the best of both techniques. It is our preferred method that leaves constitutional varus/valgus and sets rotation at 90° of flexion and avoids most ligament releases. However, regardless of the technique used, the goal remains to achieve a stable, functioning knee arthroplasty.

Author contributions

JOA helped in the research process as well as the writing and editing of the final paper. AEW helped in the research process as well as the writing and editing of the final paper. KJP and SJI helped guide the organization of, and participated in, the writing and editing of the final paper. All authors contributed to the article and approved the submitted version.

Funding

The authors received no external funding of any kind during the creation of this manuscript.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Abdel MP, Oussedik S, Parratte S, Lustig S, Haddad FS. Coronal alignment in total knee replacement: historical review, contemporary analysis, and future direction. Bone Joint J. (2014) 96-B:857–62. doi: 10.1302/0301-620X.96B7.33946

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Jeffery RS, Morris RW, Denham RA. Coronal alignment after total knee replacement. J Bone Joint Surg Br. (1991) 73:709–14. doi: 10.1302/0301-620X.73B5.1894655

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Daines BK, Dennis DA. Gap balancing vs. measured resection technique in total knee arthroplasty. Clin Orthop Surg. (2014) 6:1–8. doi: 10.4055/cios.2014.6.1.1

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Insall JN, Scuderi GR, Komistek RD, Math K, Dennis DA, Anderson DT. Correlation between condylar lift-off and femoral component alignment. Clin Orthop Relat Res. (2002) 403:143–52. doi: 10.1097/00003086-200210000-00022

CrossRef Full Text | Google Scholar

5. Jerosch J, Peuker E, Philipps B, Filler T. Interindividual reproducibility in perioperative rotational alignment of femoral components in knee prosthetic surgery using the transepicondylar axis. Knee Surg Sports Traumatol Arthrosc. (2002) 10:194–7. doi: 10.1007/s00167-001-0271-x

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Kinzel V, Ledger M, Shakespeare D. Can the epicondylar axis be defined accurately in total knee arthroplasty? Knee. (2005) 12:293–6. doi: 10.1016/j.knee.2004.09.003

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Siston RA, Patel JJ, Goodman SB, Delp SL, Giori NJ. The variability of femoral rotational alignment in total knee arthroplasty. J Bone Joint Surg Am. (2005) 87:2276–80. doi: 10.2106/JBJS.D.02945

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Poilvache PL, Insall JN, Scuderi GR, Font-Rodriguez DE. Rotational landmarks and sizing of the distal femur in total knee arthroplasty. Clin Orthop Relat Res. (1996) 331:35–46. doi: 10.1097/00003086-199610000-00006

CrossRef Full Text | Google Scholar

9. Laskin RS, Rieger MA. The surgical technique for performing a total knee replacement arthroplasty. Orthop Clin North Am. (1989) 20:31–48.2919077

PubMed Abstract | Google Scholar

10. Pagnano MW, Hanssen AD. Varus tibial joint line obliquity: a potential cause of femoral component malrotation. Clin Orthop Relat Res. (2001) 392:68–74. doi: 10.1097/00003086-200111000-00009

CrossRef Full Text | Google Scholar

11. Freeman MA, Samuelson KM, Bertin KC. Freeman-Samuelson total arthroplasty of the knee. Clin Orthop Relat Res. (1985) 192:46–58. doi: 10.1097/00003086-198501000-00007

CrossRef Full Text | Google Scholar

12. Sugama R, Kadoya Y, Kobayashi A, Takaoka K. Preparation of the flexion gap affects the extension gap in total knee arthroplasty. J Arthroplasty. (2005) 20:602–7. doi: 10.1016/j.arth.2003.12.085

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Churchill JL, Khlopas A, Sultan AA, Harwin SF, Mont MA. Gap-Balancing versus measured resection technique in total knee arthroplasty: a comparison study. J Knee Surg. (2018) 31:13–6. doi: 10.1055/s-0037-1608820

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Tapasvi SR, Shekhar A, Patil SS, Dipane MV, Chowdhry M, McPherson EJ. Comparison of gap balancing vs measured resection technique in patients undergoing simultaneous bilateral total knee arthroplasty: one technique per knee. J Arthroplasty. (2020) 35:732–40. doi: 10.1016/j.arth.2019.10.002

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Katz MA, Beck TD, Silber JS, Seldes RM, Lotke PA. Determining femoral rotational alignment in total knee arthroplasty: reliability of techniques. J Arthroplasty. (2001) 16:301–5. doi: 10.1054/arth.2001.21456

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Dennis DA, Komistek RD, Kim RH, Sharma A. Gap balancing versus measured resection technique for total knee arthroplasty. Clin Orthop Relat Res. (2010) 468:102–7. doi: 10.1007/s11999-009-1112-3

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Moon YW, Kim HJ, Ahn HS, Park CD, Lee DH. Comparison of soft tissue balancing, femoral component rotation, and joint line change between the gap balancing and measured resection techniques in primary total knee arthroplasty: a meta-analysis. Medicine (Baltimore. (2016) 95(39):e5006. doi: 10.1097/MD.0000000000005006

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Drexler M, Abolghasemian M, Barbuto R, Naini MS, Voshmeh N, Rutenberg TF, et al. Patient's height and hip medial offset are the main determinants of the Valgus cut angle during total knee arthroplasty. J Arthroplasty. (2017) 32:1496–501. doi: 10.1016/j.arth.2016.12.021

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Reed SC, Gollish J. The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty. (1997) 12:677–82. doi: 10.1016/s0883-5403(97)90141-8

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Almaawi AM, Hutt JRB, Masse V, Lavigne M, Vendittoli PA. The impact of mechanical and restricted kinematic alignment on knee anatomy in total knee arthroplasty. J Arthroplasty. (2017) 32:2133–40. doi: 10.1016/j.arth.2017.02.028

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Eckhoff D, Hogan C, DiMatteo L, Robinson M, Bach J. Difference between the epicondylar and cylindrical axis of the knee. Clin Orthop Relat Res. (2007) 461:238–44. doi: 10.1097/BLO.0b013e318112416b

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Hungerford DS, Krackow KA. Total joint arthroplasty of the knee. Clin Orthop Relat Res. (1985) 192:23–33. doi: 10.1097/00003086-198501000-00004

CrossRef Full Text | Google Scholar

23. Howell SM, Howell SJ, Kuznik KT, Cohen J, Hull ML. Does a kinematically aligned total knee arthroplasty restore function without failure regardless of alignment category? Clin Orthop Relat Res. (2013) 471:1000–7. doi: 10.1007/s11999-012-2613-z

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Bellemans J, Colyn W, Vandenneucker H, Victor J. The chitranjan ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res. (2012) 470:45–53. doi: 10.1007/s11999-011-1936-5

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Victor JM, Bassens D, Bellemans J, Gürsu S, Dhollander AA, Verdonk PC. Constitutional varus does not affect joint line orientation in the coronal plane. Clin Orthop Relat Res. (2014) 472:98–104. doi: 10.1007/s11999-013-2898-6

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Oussedik S, Scholes C, Ferguson D, Roe J, Parker D. Is femoral component rotation in a TKA reliably guided by the functional flexion axis? Clin Orthop Relat Res. (2012) 470:3227–32. doi: 10.1007/s11999-012-2515-0

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Young SW, Sullivan NPT, Walker ML, Holland S, Bayan A, Farrington B. No difference in 5-year clinical or radiographic outcomes between kinematic and mechanical alignment in TKA: a randomized controlled trial. Clin Orthop Relat Res. (2020) 478:1271–9. doi: 10.1097/CORR.0000000000001150

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Shelton TJ, Nedopil AJ, Howell SM, Hull ML. Do varus or valgus outliers have higher forces in the medial or lateral compartments than those which are in-range after a kinematically aligned total knee arthroplasty? Limb and joint line alignment after kinematically aligned total knee arthroplasty. Bone Joint J. (2017) 99-B:1319–28. doi: 10.1302/0301-620X.99B10.BJJ-2017-0066.R1

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Dennis DA, Komistek RD, Walker SA, Cheal EJ, Stiehl JB. Femoral condylar lift-off in vivo in total knee arthroplasty. J Bone Joint Surg Br. (2001) 83:33–9. doi: 10.1302/0301-620x.83b1.10632

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Nedopil AJ, Howell SM, Hull ML. What mechanisms are associated with tibial component failure after kinematically-aligned total knee arthroplasty? Int Orthop. (2017) 41:1561–9. doi: 10.1007/s00264-017-3490-6

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Vigdorchik JM, Wakelin EA, Koenig JA, Ponder CE, Plaskos C, DeClaire JH, et al. Impact of component alignment and soft tissue release on 2-year outcomes in total knee arthroplasty. J Arthroplasty. (2022) 37:2035–40.e5. doi: 10.1016/j.arth.2022.04.042

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Parratte S, Pagnano MW, Trousdale RT, Berry DJ. Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am. (2010) 92:2143–9. doi: 10.2106/JBJS.I.01398

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Jacofsky DJ, Allen M. Robotics in arthroplasty: a comprehensive review. J Arthroplasty. (2016) 31:2353–63. doi: 10.1016/j.arth.2016.05.026

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Lang JE, Mannava S, Floyd AJ, Goddard MS, Smith BP, Mofidi A, et al. Robotic systems in orthopaedic surgery. J Bone Joint Surg Br. (2011) 93:1296–9. doi: 10.1302/0301-620X.93B10.27418

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Shalhoub S, Lawrence JM, Keggi JM, Randall AL, DeClaire JH, Plaskos C. Imageless, robotic-assisted total knee arthroplasty combined with a robotic tensioning system can help predict and achieve accurate postoperative ligament balance. Arthroplast Today. (2019) 5:334–40. doi: 10.1016/j.artd.2019.07.003

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Joo PY, Chen AF, Richards J, Law TY, Taylor K, Marchand K, et al. Clinical results and patient-reported outcomes following robotic-assisted primary total knee arthroplasty: a multicentre study. Bone Jt Open. (2022) 3:589–95. doi: 10.1302/2633-1462.37.BJO-2022-0076.R1

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty. (2007) 22:1097–106. doi: 10.1016/j.arth.2007.08.001

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Song EK, Seon JK, Yim JH, Netravali NA, Bargar WL. Robotic-assisted TKA reduces postoperative alignment outliers and improves gap balance compared to conventional TKA. Clin Orthop Relat Res. (2013) 471:118–26. doi: 10.1007/s11999-012-2407-3

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Yim JH, Song EK, Khan MS, Sun ZH, Seon JK. A comparison of classical and anatomical total knee alignment methods in robotic total knee arthroplasty: classical and anatomical knee alignment methods in TKA. J Arthroplasty. (2013) 28:932–7. doi: 10.1016/j.arth.2013.01.013

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Deckey DG, Rosenow CS, Verhey JT, Brinkman JC, Mayfield CK, Clarke HD, et al. Robotic-assisted total knee arthroplasty improves accuracy and precision compared to conventional techniques. Bone Joint J. (2021) 103-B:74–80. doi: 10.1302/0301-620X.103B6.BJJ-2020-2003.R1

PubMed Abstract | CrossRef Full Text | Google Scholar

41. Kim YH, Yoon SH, Park JW. Does robotic-assisted TKA result in better outcome scores or long-term survivorship than conventional TKA? A randomized, controlled trial. Clin Orthop Relat Res. (2020) 478:266–75. doi: 10.1097/CORR.0000000000000916

PubMed Abstract | CrossRef Full Text | Google Scholar

42. Kort N, Stirling P, Pilot P, Müller JH. Robot-assisted knee arthroplasty improves component positioning and alignment, but results are inconclusive on whether it improves clinical scores or reduces complications and revisions: a systematic overview of meta-analyses. Knee Surg Sports Traumatol Arthrosc. (2022) 30:2639–53. doi: 10.1007/s00167-021-06472-4

PubMed Abstract | CrossRef Full Text | Google Scholar

43. Tanzer M, Laverdière C, Barimani B, Hart A. Augmented reality in arthroplasty: an overview of clinical applications, benefits, and limitations. J Am Acad Orthop Surg. (2022) 30:e760–8. doi: 10.5435/JAAOS-D-21-00964

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: total knee arthoplasty, alignment, mechanical, kinematic, robotics, functional

Citation: Aflatooni JO, Wininger AE, Park KJ and Incavo SJ (2023) Alignment options and robotics in total knee arthroplasty. Front. Surg. 10:1106608. doi: 10.3389/fsurg.2023.1106608

Received: 23 November 2022; Accepted: 23 January 2023;
Published: 9 February 2023.

Edited by:

Yan Yu, Tongji University School of Medicine, China

Reviewed by:

Wang Deng, Beijing Jishuitan Hospital, China

© 2023 Aflatooni, Wininger, Park and Incavo. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Stephen J. Incavo sjincavo@houstonmethodist.org

Specialty Section: This article was submitted to Orthopedic Surgery, a section of the journal Frontiers in Surgery

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.