Two decades of experience in explantation and graft preserving strategies following primary endovascular aneurysm repair and lessons learned

\r\nSherif Sultan,,
&#x;Sherif Sultan1,2,3*†Yogesh Acharya,&#x;Yogesh Acharya1,2†Mohieldin HezimaMohieldin Hezima1Keegan Chua Vi LongKeegan Chua Vi Long1Osama SolimanOsama Soliman3Juan Parodi,Juan Parodi4,5Niamh Hynes&#x;\r\nNiamh Hynes3†
  • 1Western Vascular Institute, Department of Vascular and Endovascular Surgery, University Hospital Galway, National University of Ireland, Galway, Ireland
  • 2Galway: Department of Vascular Surgery and Endovascular Surgery, Galway Clinic, Doughiska, Royal College of Surgeons in Ireland and National University of Ireland, Galway affiliated Hospital, Galway, Ireland
  • 3CORRIB-CURAM-Vascular Group, National University of Ireland, Galway, Ireland
  • 4Department of Vascular Surgery and Biomedical Engineering Department, Alma mater, University of Buenos Aires, and Trinidad Hospital, Buenos Aires, Argentina
  • 5Winston-Salem and St. Louis: Wake Forest University, Winston-Salem, North Carolina and Washington University in St. Louis, St. Louis, Missouri, United States of America

Objectives: We aim to scrutinize our evolving re-intervention strategies following primary endovascular aortic aneurysm repair (EVAR) - EVAR GORE SalvAge Fabric Technique (ARAFAT), aortic sac double breasting with endograft preservation, and stent-graft explantation.

Methods: We performed 1,555 aortic interventions over the study period, including 910 EVARs. Factors associated with the need for reintervention and the likelihood of chronic fabric fatigue failure (CFFF) were investigated. Using conventional and innovative diagnostic modalities with Prone contrASt enHanced computed tomography Angiography (PASHA), 136 endoleaks (ELs) were identified (15 type I, 98 type II; 18 type III; 5 type IV).

Results: Forty-four (4.84%) patients underwent re-intervention post-primary EVAR; 18 ARAFATs, 12 double breastings, and 14 explantations. Choice of re-intervention was based on patient fitness and mode of failure. Mean EL detection duration following primary EVAR was 53.3 ± 6.82 months, while mean time to re-intervention was 70.20 ± 6.98 months. The mean sac size before the primary EVAR and re-intervention was 6.00 ± 1.75 cm and 7.51 ± 1.94 cm, respectively. Polyester (61.40%) was the most commonly employed stent-graft material. Use of more than three modular stent-graft components (3.42 ± 1.31, p = 0.846); with the proximal stent-graft diameter of 31.6 ± 3.80 cm (p = 0.651) and the use of iliac limbs more than 17 mm (p = 0.364), all added together are contributing factors. We had one peri-operative mortality following explantation due to sepsis-induced multiorgan failure.

Conclusions: Our re-intervention strategies matured from stent graft explantation to graft preservation with endovascular relining of the stent-graft. Graft preservation with aortic sacotomy and double breasting were used to manage concealed ELs due to aortic hygroma.

Introduction

Endovascular aneurysm repair (EVAR) has revolutionised therapeutic tactics in managing aortic pathologies over the past three decades. EVAR has shown reduced early peri-operative morbidity and mortality compared to open surgical repair (OSR) (1). However, these advantages are lost in long-term follow-up due to stent-graft complications, including fabric and material failure (25).

EVAR effectiveness is determined by the aortic sac segregation from systematic pressure and sac shrinkage. Complications post-EVAR need close surveillance as aortic sac dynamics influences EVAR durability, and continuous sac expansion could result in rupture (3). Endoleaks (ELs) may thrombose, but if they persist, the consequences can be detrimental, mainly if they are high-flow and associated with continued aneurysmal sac expansion (4, 5). These complications require aggressive management with re-intervention to abolish the risk of rupture. Re-intervention could be achieved through salvage of the primary endograft via a graft preserving strategy or explantation as necessary. In this study, we aim to analyse our three evolving strategies of re-intervention following the primary EVAR - EVAR GORE SalvAge Fabric Technique (ARAFAT), double breasting, and explantation.

Methods

This is a retrospective observational study performed in our tertiary vascular center from 2002 to 2020. The primary outcome is aortic related mortality. The secondary outcomes are technical success, perioperative morbidity and mortality, and overall survival probabilities.

Society for Vascular Surgery (SVS) reporting standards is used to define the outcomes (6). Technical success is defined as the periprocedural events from the initiation of the procedure to the first 24-hour. Primary technical success is the successful introduction and deployment without surgical conversion or death, type I or III ELs, or graft limb obstruction.

A re-intervention is classified as any procedure performed for subsequent aneurysm and/or primary procedure-related complications during follow-up of the primary EVAR. Factors associated with the need for re-intervention and the likelihood of chronic fabric fatigue failure (CFFF) were investigated amongst the re-intervention groups.

Primary EVAR and follow-up strategy

Out of 22,349 aortic referrals to our tertiary referral centre, we performed 1,555 aortic interventions over twenty years. Amongst them, 910 were EVAR ± iliac branch device (IBD), and 96 were thoracic endovascular aortic repair (TEVAR)/branched endovascular aortic repair (BEVAR).

All our patients were followed-up with duplex ultrasonography (DUS) and plain film abdomen every six months. Patient with continuous sac expansion without detection of EL through conventional supine computed tomography angiography (CTA) underwent a Prone contrASt enHanced computed tomography Angiography (PASHA), a multiphase time-resolved four-phase positional (prone) CTA protocol for detection and classification of concealed ELs (7). The PASHA protocol enhanced the degree of contrast infiltration into the aortic sac when microleaks were present, which helped us to plan the re-intervention strategy. PASHA diagnosed all cases of type IIIb EL that were previously classified as concealed Type V EL in the context of continuous sac enlargement (7).

We identified 136 ELs (14.95%), including 15 type I ELs, 98 type II ELs, 18 type III ELs, and 5 type IV ELs. Decisions for re-interventions were formulated at the discretion of the vascular surgery multidisciplinary team, which considered factors like the rate of sac expansion, EL types, and patient's general condition (Figure 1). All the type I ELs were managed with proximal aortic cuff and/or distal extension or chimney endovascular aneurysm repair (ChEVAR). Four (26.70%) of these type I ELs required tertiary re-intervention at 5, 7, 8 years of post-secondary intervention. Amongst the 98 type II ELs, 12 (12.24%) that were associated with aortic sac expansion had a trial of embolisation initially; however, they all had persistent sac expansion despite embolisation. We applied PASHA diagnostic technique to them, which eventually showed aortic sac hygromas or type III ELs. Out of these 12, seven underwent double breasting for aortic sac hygromas, and the remaining five had explantations due to chronic fabric fatigue with type IIIB EL within three years of re-intervention. Our isolated type II EL had a 41.84% (n = 41) spontaneous resolution rate, and none of them ruptured.

FIGURE 1
www.frontiersin.org

Figure 1. Flowchart depicting the adoption of the re-intervention strategies (EVAR GORE SalvAge FAbric Technique (ARAFAT), Double Breasting, and Explantation) following Prone ContrASt EnHanced Computed Tomography Angiography (PASHA).

Re-intervention strategies

Our practice had evolved over twenty years from diagnostics to management. Our decision-making has been influenced by several factors, including the mode of endograft failure, aneurysm sac size and patient fitness for surgery. Our ability to make informed decisions on the mode of failure evolved after we spearheaded the PASHA CTA protocol to accurately differentiate between different types of ELs and aortic sac hygromas (7). Aortic sac hygromas with the Hounsfield unit (HU) < 25 and an associated sac size greater than 7.5 cm indicate aortic aneurysm sac failure and a loss of the ability of the aortic wall to remodel (7). In cases with no defined type I, II or III ELs, and HU was less than 25, a diagnosis of expanding aortic hygroma was confirmed, and aneurysmorrhaphy was performed. In our experience, patients developing hygromas had factors, which likely contributed to the ability of fluid to exudate through the endograft material. We witnessed that the predisposing factors for aortic sac hygroma are direct oral anti-coagulant medication, administration of tissue plasminogen activator (tPA) and episodes of hypoalbuminemia (7). However, in the absence of these factors, it was considered that exudation of fluid through the endograft is likely to represent general fabric fatigue and loss of crystallinity; we considered that relining the endograft with the ARAFAT technique was more useful to exclude the failed endograft from the circulation and prevent ongoing transudation into the aortic sac. As our patients are living longer and now present 10–15 years after the index procedure, they can be late octogenarians or early nonagenarians by the time they require reintervention. This forced us to employ the ARAFAT protocol more frequently. In essence, if a patient is fit for a definitive endograft explant and open surgical repair, then this is our procedure of choice; if a patient has precedent factors that likely contributed to a hygroma and those factors are likely not to recur, then we perform an aneurysmorrhaphy; however, if the patient has an expanding sac and is unlikely to be able to tolerate such an open surgical approach, we opt for ARAFAT physically (Figure 1).

Explantation

In our early series with type IIIB fabric failure, we performed explantation by infra-renal clamping after transfixing the left renal vein followed by partial endograft excision (Figures 2A,B). Once the fabric of an endograft has started to fray, it heralds the start of a more substantial issue and represents a more extensive reduction in fabric integrity and a loss of crystallinity. Placing a stitch in the fabric would be a temporary, and likely unsuccessful attempt to solve a chronic problem. In fact, stitching a frayed fabric would likely propagate further holes and fabric disintegration, causing more damage than good. Once the device has failed, it needs to be explanted or relined. During explantation, we routinely leave behind the suprarenal uncovered stents and their barbs and anastomosed the open surgical graft to it infra-renally after partial graft excision. However, the AFX endologix (Endologix Inc., Irvine, CA, USA) graft was totally removed as it depends on the iliac bifurcation for fixation and does not have barbs (Figure 2C). We routinely performed re-enforcement with polytetrafluoroethylene (PTFE) pledgets when anastomosing the silver Dacron graft to the old proximal failed aortic graft or the aortic wall. Pledgets were necessary for reinforcement as the aortic tissue was friable and required support.

FIGURE 2
www.frontiersin.org

Figure 2. Explantation. (A) Body of partially explanted Cook (Bloomington, IN, USA) polyester-based endograft depicting micro-fabric pores due to chronic fatigue failure close to the allies of the stent. (B) Computed tomography angiography (CTA) image showing partially explanted Cook graft with the suprarenal spare springs and hooks left in-situ. (C) Here, patient underwent primary endovascular aneurysm repair (EVAR) with Endologix AFX (Endologix, Irvine, CA). However, this polytetrafluoroethylene (PTFE) based endograft failed due to the fracture of the endoskeleton, which acted as a hinge against the PTFE. Attempts to salvage the EVAR resulted in implantation of two proximal cuffs over six years follow-up. The patient presented with rapidly expanding 8 cm abdominal aortic aneurysm and abdominal pain, necessitating immediate total graft explantation and aorto-bi-iliac reconstruction by 16 × 8 mm silver Dacron graft (Maquet, Rastatt, Germany).

However, many of our patients with type IIIB EL were between 4 and 9 years post-implantation and therefore, most were either septuagenarian or octogenarian. Open surgery and explantation are risky in these frail sarcopenic patients. Consequently, we developed the ARAFAT technique in patients at high risk for surgery, i.e., elderly patients with concurrent co-morbidities.

Aortic sac hygroma and double-breasting

Aortic sac hygromas are sac dilatations attributed to transudation through the stent-graft fabric. They expand slowly without evidence of EL forming a phlegmon with jelly-like consistency around the endograft. They have radio-density less than 25 Hounsfield units (HU) and are most commonly associated with polyester endografts (711). Diagnosis of aortic sac hygromas was neither feasible nor accurate previously; however, newer imaging techniques will identify them.

We performed aortic sacotomy after the diagnosis, with the evacuation of related hygroma and/or aortic thrombus, and all bleeding lumber vessels were transfixed. Subsequently, we filled the opened aortic sac with XenoSure® biologic patches (LeMaitre Vascular, Inc., MA, USA) to induce fibrosis. We then performed aneurysmorrhaphy by double breasting and plication of the aneurysm sac over EVAR graft to prevent contact with the bowel, thereby reducing the risk of subsequent graft infection (Figures 3A, B). Wrapping and cerclage of the proximal aneurysm neck were used to prevent stent-graft migration and EVAR dislodgment. Aortic sacotomy and double breasting made it possible to confirm endotension when no visible leak was seen on the preoperative CT scan. In these cases, the aortic sac hygroma presented as dark grey organised seroma with no patent back-bleeding vessels. All hygromas were sent for culture and sensitivity, and all returned sterile.

FIGURE 3
www.frontiersin.org

Figure 3. Aortic sacotomy with obliterating endo-aneurysmorrhaphy and stent-graft preservation post-EVAR. (A): Wrapping the opened aortic sac with XenoSure biological patch (LeMaitre Vascular, Inc., MA, USA) during double breasting. This 82-year-old patient had endovascular aneurysm repair (EVAR) with polytetrafluoroethylene (PTFE) GORE graft (GORE®, Flagstaff, Arizona, USA) in 2008, after which the aortic sac shrunk to 4.2 cm. However, the patient had tissue plasminogen activator (TPA) twice for acute MI in 2018, resulting in aortic sac hygroma with sac expansion to 8.9 cm. The formation of a hygroma as a consequence of ultrafiltration of blood through the stent-graft fabric led to continued sac enlargement without a detectable problem within the endograft, i.e., no structural stent-graft problem and no demonstration of EL. (B). Double breasting of the aortic sac with interrupted mattress prolene stitches to achieve hemostasis. Due to aortic sac hygroma, the patient experienced continuous sac expansion, abdominal pain, and low back pain. Computed tomography angiography (CTA) demonstrated the radiodensity of <25 Hounsfield units. The patient required aortic sacotomy, followed by wrapping the aortic ePTFE device with XenoSure biological patches and filling the post wall of the aortic sac with haemostatic powder HaemocerTM plus (Biocer Entwicklungs-Gmbh, Bayreuth, Germany), after which double breasting of the aortic sac was performed. This sealed the aortic sac and abolished the abdominal and low back pain. The aortic sac shrank to 4.5 cm during the subsequent follow-up.

EVAR GORE SalvAge fabric technique (ARAFAT)

Over the past 5 years, we utilized ARAFAT as our protocol to seal the type IIIB EL, particularly in patients at high risk for open conversion (7). ARAFAT helped us realign stent graft to seal microleaks and improve spiral arterial flow. Here, we deployed an oversized EXCLUDER® aortic cuff (GORE®, Flagstaff, Arizona, USA) into the previously implanted stent graft, followed by the simultaneous deployment of EXCLUDER® iliac extensions as necessary in double-barrel configuration from the main cuff (Figures 4A–C) (7).

FIGURE 4
www.frontiersin.org

Figure 4. EVAR GORE SalvAge Fabric Technique (ARAFAT) is our protocol used to seal the type IIIb EL. (A) Microleaks at the endograft. (B) Oversized EXCLUDER® aortic cuff (GORE®, Flagstaff, Arizona, USA) deployed into the previously implanted stent-graft. (C) Simultaneous deployment of an EXCLUDER® iliac extension, as necessary, in double-barrel configuration from the main cuff.

Ethical consideration

Ethical approval was obtained from the local Institutional Ethics Review Board (C.A. 1210). Data were collected from patients' records and anonymised. Utmost priority was given to maintain patients' confidentiality.

Statistical analysis

Continuous outcomes were summarized with mean and standard deviation (normal distribution) or median and interquartile range (non-normal distribution). The categorical outcomes were summarized with percentages and proportions. For statistical significance, Chi-squared and Kruskal-Wallis tests were used. p < 0.05 was taken as statistically significant. Statistical analyses were conducted with Minitab (Minitab® Ltd., UK).

Results

We had 44 patients who underwent reinterventions following primary EVAR, including 18 ARAFAT and 14 explantations for type IIIB ELs and 12 double breastings for aortic sac hygromas. The baseline characteristics of these patients are given in Table 1.

TABLE 1
www.frontiersin.org

Table 1. Baseline characteristics of the patients.

The average size of the aortic aneurysm sac during the primary EVAR was 6.00 ± 1.75 cm (Table 2). Twenty-seven polyester and 17 PTFE based endo-grafts were employed in these primary procedures. The average number of stent pieces used in the primary EVAR was 3.42 ± 1.31. The proximal main body diameter was 31.6 ± 3.80 mm (right limb size: 17.6 ± 4.20 mm and left limb size 17.9 ± 4.28 mm).

TABLE 2
www.frontiersin.org

Table 2. Endograft characteristics employed in the primary endovascular aneurysm repair.

The mean duration of ELs identification following primary EVAR was 53.3 ± 6.82 months. The mean aortic sac expansion rate was 0.43 ± 0.25 cm per year, and the mean aortic sac size before the re-intervention was 7.51 ± 1.94 cm. Patients who underwent double breasting had the highest sac size (ARAFAT: 6.68 ± 2.13 vs. double breasting: 8.41 ± 1.72 vs. explantation: 7.52 ± 1.61 cm, p = 0.029). The mean duration of re-intervention following primary-EVAR was 70.2 ± 6.98 months (ARAFAT: 94.2 ± 12.5 vs. double breasting: 67.3 ± 6.78 vs. 41.8 ± 9.54 months, p = 0.026).

There was no difference between stent-graft materials (27 PE vs. 17 ePTFE, p = 0.06) on the re-intervention rate (Table 3).

TABLE 3
www.frontiersin.org

Table 3. Stent-graft employed during the primary endovascular aneurysm repair (EVAR) and those requiring re-intervention.

All the patients had primary technical success. However, we had one sepsis-induced peri-operative mortality following explantation in a patient who had initial re-intervention for rapidly expanding aortic sac with type I EL before being referred to us. This patient also had prior embolisation of a lumbar branch, where the coil migrated to the spine resulting in paraparesis. After being referred to our centre, we performed explantation and aorto-bi-renal-bi-iliac bypass. The patient developed multiorgan failure due to sepsis and succumbed to death on the 28th postoperative day.

The overall survival plots during an average follow-up duration of 35.6 ± 6.24 months (ARAFAT: 9.00 ± 1.36 vs. double breasting: 42.5 ± 9.51 vs. explantation: 63.8 ± 14.2) is depicted in Figure 5.

FIGURE 5
www.frontiersin.org

Figure 5. The overall survival plot of the patients undergoing different re-intervention strategies (EVAR GORE SalvAge FAbric Technique (ARAFAT), Double Breasting, and Explantation).

Discussion

This study aims to scrutinise our three techniques of post-EVAR re-intervention, including an EVAR graft preservation strategy and/or explantation. Significant sac expansion over a short period needs scrutiny. Stent graft explantation with subsequent replacement is the definitive management approach. Graft preservation strategies include surgical double breasting of the aortic sac or endovascular relining of the stent-graft. For patients with significant co-morbidities and an enlarged sac with a maximum diameter less than 7.5 cm, we employed ARAFAT. Our results mimic Doumenc et al. (12) finding's that explantation and endovascular management, hybrid endovascular repair and ARAFAT can be achieved in high deliberate practice volume centres with satisfactory results.

All our re-interventions had aortic sac size greater than 6 cm at the time of their primary EVAR with more than three modular stent graft components. Those that underwent double breasting had the largest sac size due to aortic hygromas (more than 7.5 cm on average) (7). Polyester was the most common stent material used in the primary EVAR amongst the re-intervention groups (p = 0.258).

EVAR carries a higher reintervention risk than OSR, and the risk increases with time (2). Reintervention rate could range from 20% in low-risk cases to 25% in high-risk patients (13, 14). There is a minimal divergence between first, second or third-generation aortic endograft devices (15). However, there has been an acknowledged failure rate for all major commercial endografts, necessitating late aneurysmorrhaphy and double breasting (16).

In our experience of 44 cases, we were surgically successful for a range of different endovascular scenarios and accomplished reasonable outcomes. There was no aneurysm-related mortality. However, amongst the nine mortalities over the mean follow-up duration, 60% were cancer-related, and 40% were cardiac-related.

Endograft twisting, dislodgment, kinking, outflow obstruction, and ELs are some of the complications encountered post-EVAR. However, the main reason for re-intervention is EL with endotension and sac expansion regardless of the EL types (17). Gambardella et al. (18) reported that EL is also the utmost reason for aortic sacotomy, aneurysmorrhaphy and double breasting. Also, our strategy mirrors their advice in favouring infrarenal clamping for partial explanation of failed supra renal fixated EVAR grafts.

Controversial views about EL and endotension are plentiful within the vascular literature. Studies have shown that persistent EL increases the risk of rupture (19). However, some advocates that type II ELs are protective against aortic rupture (20). Current guidelines recommend intervention in patients with continuous sac expansion (21). Variable results with trans-arterial, trans-caval, direct sac access and trans-iliac-lumbar embolization for type II EL had been reported (21). Alternative approaches such as the laparoscopic approach for type II EL has been described as tricky, even by the most proficient surgeons, due to dense periaortic inflammation (22). Minimal invasive therapies for type II ELs risk repeated interventions, with poor outcomes at five years (21, 23). In our opinion, type II EL should not be accepted as the cause of aortic sac expansion until multimodal investigations with temporal information regarding perigraft blood flow have been completed.

Our type IA ELs were challenging to solve as there was no consensus. Our cases were managed by Design Reconfigure Elongate Straighten Stiffen (DRESS) technique with proximal cuff and adjuvant ChEVAR.

We agree with the recommendation of Hinchliffe et al. (24) that in frail patients who were turned down for open repair a few years earlier, less invasive approaches with a lower threshold for endograft preservation should be considered. In this context, an aortic sacotomy with ligation of patent back-bleeding vessels and preservation of the EVAR graft can be a wise alternative to a more invasive explantation to prevent rupture in an expanding aortic sac.

Almost half of the ELs seal spontaneously during the first year; however, the likelihood of spontaneous closure decreases with time (4, 25). There is no consensus regarding how long to follow up and when to re-intervene (1315). Only 1.97% of our patients had an endovascular reintervention before aneurysmorrhaphy, as we are highly selective in offering EVAR. The mean age of our patients who underwent re-intervention was 79.7 ± 7.03 years. In frail patients, we considered open repair only if endovascular options were impractical or exhausted. However, in cases in which the aortic sac diameter was more than 75 mm with abdominal or low back pain, we performed aneurysmorrhaphy. Any prolongation of such a critical situation could have ended in aortic sac rupture. This contradicts some authors who had one-third of all their patients subjected to endovascular manipulation before sacotomy, aneurysmorrhaphy, graft preservation and double breasting (26, 27).

Complications post-EVAR could be attributed to various factors. A head to head comparison with Endurant and Excluder grafts implanted in two groups of patients having similar anatomical characteristics demonstrated that two different types of EVAR endografts implanted in similar AAAs could provoke diverse flow properties (28). The study concluded that delineation of the hemodynamic features associated with the various commercially available EVAR grafts could further promote the personalization of treatment offered to aneurysmal patients and instigate concepts for design perfection in the future.

Published data has reported several factors associated with major adverse events post-EVAR. These factors include an aortic bifurcation with 50% calcification and diameter less than 20 mm; endograft iliac limb diameters greater than 27 mm; nitinol endograft stents; and the ratio of endograft iliac limb diameters to aortic bifurcation diameter greater than 1.4 (29). Conversely, the displacement forces in tapered iliac stent-grafts with asymmetric curvatures will impact stent-graft performance. When arterial blood pressure is on a curved stent-graft, it will generate an axially oriented force. The larger the cross-section, the more significant the force. In addition, the force generated by the flow velocity in the curvature of a vessel or a stent-graft acts in the axial direction. These two drag forces are due to the flow reaction to a change in direction due to the kinetic energy of the moving blood volume. The higher this velocity, the higher the energy. When the flow rallies into the curved wall, the kinetic energy is converted into a strong force. In smaller vessels, less than 11 mm, the force-velocity is larger than the force pressure. This breeds trepidations, particularly for stent-graft designs with 11 mm contralateral gates, which necessitates 13–14 mm contralateral docking limbs as they induce mega forces that result in major adverse events. Tapered grafts increase the axial forces applied at both ends by 50% as the flow velocity increases with a smaller diameter, increasing force-velocity. These forces increase with angulation, and implantations in angulated iliacs must be avoided to minimise migration risks. These haemodynamic forces have implications for stent-graft design for both tapered and bell-bottom geometries. A tapered graft should be outsized at both ends to augment radial and frictional force with the vessel and counterbalance velocity increase (30).

Morris et al. (31) had mirrored the above findings in the abdominal aortic endografts by analysing the Zenith (Cook, Bloomington, IN) and Endurant II (Medtronic Santa Rosa, CA) devices and documented the highest radial resistive force up to 3 N/cm. The supra-renal and infrarenal compliances were 6.9–5.1 × 104/mmHg and 4.8–5.4 × 104/mmHg. In contrast, the Fortron device (Cordis Endovascular, Santa Clara, CA) had the lowest at 0.11 N/cm. The Endurant II and Excluder devices had significantly decreased infrarenal compliance by 13%–26%. All four devices increased the pulsatile arterial energy loss (PAEL) by 44%, significantly lowering aortic wall compliance after EVAR. Choosing the most compliant devices for treating AAA minimises micro and macro-ELs and graft material fatigue and failure with later explantation with avoiding long-term cardiovascular events.

Further, long term renal outcomes with proximal aortic fixation are questionable, problematic and not yet established (32). Morris et al. (31) showed a frank dissimilarity between nitinol-based endografts with Dacron and suprarenal fixation compared to nitinol-based endografts with PTFE. Zenith (Cook Medical) and Endurant II (Medtronic) had the highest aortic stiffness (radial resistive force). Moreover, significant lower infra-renal compliance was observed in Endurant II and Excluder. Similarly, the selection of the most compliant devices will enhance aortic elastic recoil with lower post-procedural complications. The metallic endograft skeleton, whether nitinol, stainless steel or cobalt alloy, reduces aortic compliance and stiffens the aorta. This stiffness induces a mismatch in the physio-mechanical properties between the native and the stented aorta, which results in PWV intensification.

Proximal fixation devices by suprarenal barbs for wall anchoring are contemplated to lessen the risk of distal migration, perfect proximal seal, and minimise type I EL. These attributes present us with a more challenging task as the presence of suprarenal stents often leads to suprarenal or supra-visceral proximal aortic clamping. Furthermore, detachment of wall-anchoring barbs risks injury to the aortic wall and renal ostia. These challenges increase operative and aortic clamp times, explaining the high reported mortality in these patients (33, 34).

We performed a partial explantation technique by leaving the suprarenal components of stent-grafts in-situ in the absence of sepsis. This is our preferred revascularisation option to reduce the risk of intraoperative injury to the aortic wall and branch vessels. It also helps to minimise para-aortic dissection and supra visceral aortic clamp level by retaining the proximal aortic endoskeleton. We anastomose the Dacron graft enforced with pledges to the endograft components to avoid suprarenal clamping and mortality.

The incidence of late aneurysmorrhaphy and double breasting after EVAR has been reported in up to 50% of reinterventions, which is multifactorial and depends on patient selection, follow-up protocols, endograft generation and type, and expertise in both endovascular and open aortic management (3537).

In aortic hygroma patients with aortic sac more than 8 cm with abdominal and low back pain, we perform aortic sacotomy with obliterating endo-aneurysmorrhaphy and stent-graft preservation post-EVAR.

Aortic sacotomy with obliterating endo-aneurysmorrhaphy and stent-graft preservation post-EVAR is appealing as it averts the physiologic stresses of aortic cross clamping. Our results mirror Mohapatra et al. (38) and contradict Kansal et al. (39) for their striking 43% 30-day mortality.

In our experience, we noticed no difference in patients undergoing graft preservation vs. graft explantation. However, trends indicate that the graft preservation patients' were older and a higher risk cohort.

In proceeding with graft preservation, external banding of the neck combined with ligation of all branch vessels, including inferior mesenteric artery and median sacral artery, is performed. Subsequently, the aortic sac is filled with XenoSure biological patches, and the preserved graft is wrapped with silver Dacron patches to induce fibrosis and prevent any future chance of aortic sac expansion.

Some authors advocate routine CTA follow-up post-EVAR (40, 41). However, less than 50% adhered to imaging in the EVAR-1 trial and Medicare beneficiaries after five years (42, 43). We adopted the PASHA technique over contrast-enhanced ultrasound imaging, which is operator-dependant, as PASHA is standardised, easily reproducible and can be read by everyone (7, 44, 45). We recommend lifelong follow-up every six months by DUS for both iliac arteries post explantation of aorto-bi-iliac endograft for fear of iliac degeneration and rupture, as reported by Arnaoutakis et al. (46) Postoperative CTA provides better diagnostic utility for proximal and distal neck dilatation or disconnection of the stent-graft components, which DUS could miss on regular follow-ups.

ELs are triggered by the instability of the longitudinal growth of the aorta due to the cone-shaped necks or steep angulations. The shape of the proximal portion of the stent's main body gets flattened or crushed due to the cardiovascular aortic oscillation, increasing 2–4 mm in the aortic diameter of the proximal landing zone every year. The main body migrates continuously and slowly down towards the aortic bifurcation. This creates “autologous” strut perforations with type III fabric failure and ultimately type IA that can only be salvaged by explantation or ARAFAT technique (7, 47, 48).

The ability to salvage such cases by adding FEVAR or BEVAR is deemed to fail because bridging stents are needed to provide adequate stability over time. Furthermore, new challenges have arisen concerning patency as vascular territories that are primarily unaffected are incorporated into the disease process. Moreover, devices with suprarenal fixation components preclude suitable entry to visceral and renal vessels (49). Cognisant of such findings, we strongly advocate precise and meticulous strategic primary EVAR planning.

Post-implantation syndrome (PIS) following primary procedure is also a long-term determinate of EL and micro-fabric fatigue failure. Ito et al. (50) Voûte et al. (51) and Sartipy (52) associated polyester grafts with higher postoperative pyrexia, PIS and longer in-hospital stay compared to ePTFE grafts following EVAR. Post-implantation syndrome (PIS) has been reported in up to 60% following EVAR (50, 53). Polyester triggers a higher release of inflammatory biomarkers (tumour necrosis factor-α, IL-6, IL-10, and CRP) than ePTFE in vitro (51, 54). However, implantation of stent-grafts made with woven polyester is not just independently associated with a stronger inflammatory response; it also results in endothelial damage. Furthermore, active fixation using penetrating hooks or barbs at the proximal aortic implantation site leads to endothelial aggression with the penetration of the foreign material. The precise balance between nickel and titanium, or even cutting and polishing the metal, will affect the antigenic properties of the nitinol (51).

It is not surprising that most of the EVAR device technology introduced over the past decade has been withdrawn from the market due to failure in sealing technology, material durability, unsupported body, stent fracture, and avulsion. Failures have also arisen when aortic device companies have iteratively lowered the profile of their devices to make them more attracted to non-surgical interventionalists (15). It is surprising that better stents have not yet been crafted after more than three decades of EVAR.

Choosing the most compliant devices for treating AAA minimises micro and macro-ELs and graft material fatigue, thereby avoiding late failure, explantation and long-term cardiovascular events. The physician involved in the decision-making should select the most appropriate EVAR graft. In general, approximately one-third of all of our EVARs were done out of instructions-for-use (IFU) with neck less than 1.5 cm and neck angulation more than 75%, and thrombus more than 3 mm at the proximal implantation site. However, the objective going forward is to recommend open surgery if the patient is physically fit. Otherwise, EVAR can be offered but cannot be undertaken outside IFU, with rare exceptions depending on patient fitness and urgency of presentation (47).

A strong policy of obeying the indications for use or abstaining from exploiting EVAR in challenging patients will alleviate the need for late aneurysmorrhaphy and reduce the need for reintervention to exceptional cases. LOCOS-1 (55) investigators documented that the broad applicability of EVAR increased late open conversion, independent of endovascular techniques innovations or advancements in-stent and fabrics materials. However, there are limited studies on the direct comparison between endografts and long-term outcomes, which need to be interpreted with caution. The majority of the available evidence-based studies represent a retrospective single centre experience with a limited subset of patients. Also, comparison among stent-grafts is restricted to devices based on personal and/or institutional preferences.

We believe that patient choice to a less invasive option is essential in the decision-making process; however, patients must be told of their alternatives during the informed consent process. Patients need to be well informed on the advantages and disadvantages of EVAR and understand that post-EVAR complication rates are still substantial.

Study limitations

The current study is limited by its observational nature, with potential selection bias. Although the number of patients included in this study is limited, this is one of the most extensive series to date evaluating aortic sacotomy with obliterating endo-aneurysmorrhaphy and stent-graft preservation post-EVAR. We have discussed various re-intervention strategies employed in our vascular setting; however, we understand that the indications for the re-intervention are different. Furthermore, it is not feasible to make a head-on-head comparison between them. Also, most of these re-interventions were performed recently, and we lack a long-term follow-up.

Conclusion

With the increasing popularity of EVAR, we forecast high numbers of post-procedural complications in terms of graft failure, short- and long-term hemodynamic alterations, and related morbidity and mortality. These complications necessitate the development and study of re-intervention strategies to salvage existing endograft and/or address graft-related complications.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by Galway Clinical Research Ethics Committee. Written informed consent for participation was not required for this study in accordance with the national legislation and the institutional requirements.

Author contributions

All the authors are involved in the conception and design with acquisition, analysis and interpretation of data, and drafting the work or revising it critically for important intellectual content. All authors contributed to the article and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Yin K, Locham SS, Schermerhorn ML, Malas MB. Trends of 30-day mortality and morbidities in endovascular repair of intact abdominal aortic aneurysm during the last decade. J Vasc Surg. (2019) 69:64–73. doi: 10.1016/j.jvs.2018.04.032

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Li B, Khan S, Salata K, Hussain MA, de Mestral C, Greco E, et al. A systematic review and meta-analysis of the long-term outcomes of endovascular versus open repair of abdominal aortic aneurysm. J Vasc Surg. (2019) 70:954–969.e30. doi: 10.1016/j.jvs.2019.01.076

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Maitrias P, Kaladji A, Plissonnier D, Amiot S, Sabatier J, Coggia M, et al., Association Universitaire de Recherche en Chirurgie Vasculaire (AURC), Treatment of sac expansion after endovascular aneurysm repair with obliterating endoaneurysmorrhaphy and stent graft preservation. J Vasc Surg. (2016) 63:902–8. doi: 10.1016/j.jvs.2015.10.059

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Karthikesalingam A, Thrumurthy SG, Jackson D, Choke E, Sayers RD, Loftus IM, et al. Current evidence is insufficient to define an optimal threshold for intervention in isolated type II endoleak after endovas- cular aneurysm repair. J Endovasc Ther. (2012) 19:200–8. doi: 10.1583/11-3762R.1

PubMed Abstract | CrossRef Full Text | Google Scholar

5. ElBatti S, Cochennec F, Roudot-Thoraval F, Becquemin JP. Type II endoleaks after endovascular repair of abdominal aortic aneurysm are not always a benign condition. J Vasc Surg. (2013) 57:1291–7. doi: 10.1016/j.jvs.2012.10.118

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Chaikof EL, Blankensteijn JD, Harris PL, White GH, Zarins CK, Bernhard VM, et al. Ad hoc committee for standardized reporting practices in vascular surgery of the society for vascular Surgery/American Association for Vascular Surgery. Reporting standards for endovascular aortic aneurysm repair. J Vasc Surg. (2002) 35(5):1048–60. doi: 10.1067/mva.2002.123763

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Sultan S, Acharya Y, Atteia E, Hynes N. Management of concealed Type IV endoleak and aortic sac hygroma by prone ContrASt EnHancement computed tomography angiography. Ann Vasc Surg. (2021) 72:647–61. doi: 10.1016/j.avsg.2020.11.039

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Cuff RF, Thomas JH. Recurrent symptomatic aortic sac seroma after open abdominal aortic aneurysm repair. J Vasc Surg. (2005) 41:1058–60. doi: 10.1016/j.jvs.2005.02.031

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Kadakol AK, Nypaver TJ, Lin JC, Weaver MR, Karam JL, Reddy DJ, et al. Frequency, risk factors, and management of perigraft seroma after open abdominal aortic aneurysm repair. J Vasc Surg. (2011) 54:637–43. doi: 10.1016/j.jvs.2011.03.258

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Jacobs TS, Won J, Gravereaux EC, Faries PL, Morrissey N, Teodorescu VJ, et al. Mechanical failure of prosthetic human implants: A 10-year experience with aortic stent graft devices. J Vasc Surg. (2003) 37:16–26. doi: 10.1067/mva.2003.58

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Riepe G, Loos J, Imig H, Schröder A, Schneider E, Petermann J, et al. Long-term in vivo alterations of polyester vascular grafts in humans. Eur J Vasc Endovasc Surg. (1997) 13:540–8. doi: 10.1016/S1078-5884(97)80062-7

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Doumenc B, Mesnard T, Patterson BO, Azzaoui R, De Préville A, Haulon S, et al. Management of type IA endoleak after EVAR by explantation or custom made fenestrated endovascular aortic aneurysm repair. Eur J Vasc Endovasc Surg. (2021) 61:571–8. doi: 10.1016/j.ejvs.2020.10.033

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Hiramoto JS, Reilly LM, Schneider DB, Sivamurthy N, Rapp JH, Chuter TAM. Long-term outcome and reintervention after endovascular abdominal aortic aneu- rysm repair using the Zenith stent graft. J Vasc Surg. (2007) 45:461–6. doi: 10.1016/j.jvs.2006.11.034

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Greenberg RK, Chuter TAM, Cambria RP, Sternbergh WC, Fearnot NE. Zenith abdominal aortic aneurysm endovascular graft. J Vasc Surg. (2008) 48:1–9. doi: 10.1016/j.jvs.2008.02.051

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Sultan S, Barrett N, Tawfick W, Parodi JC, Hynes N. Contemporary abdominal aortic aneurysm devices, three decades of research and development with big data. Why has the best graft not been produced yet? A missed opportunity. Ital J Vasc Endovasc. (2019) 26:121–34. doi: 10.23736/S1824-4777.19.01417-7

CrossRef Full Text | Google Scholar

16. Hobo R, Buth J, EUROSTAR collaborators. Secondary interventions following endovascular abdominal aortic aneurysm repair using current endografts. A EUROSTAR report. J Vasc Surg. (2006) 43:896–902. doi: 10.1016/j.jvs.2006.01.010

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Geraedts ACM, Mulay S, Vahl A, Wisselink W, Koelemay MJW, Balm R, et al. Secondary interventions and long-term follow-up after endovascular abdominal aortic aneurysm repair. Ann Vasc Surg. (2021) 71:381–91. doi: 10.1016/j.avsg.2020.07.042

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Gambardella I, Antoniou GA, Gaudino M, D'Ayala M, Girardi LN, Torella F. State of the art and meta-analysis of secondary open aortic procedure after abdominal endovascular aortic repair. J Vasc Surg. (2019) 70:1341–1350.e4. doi: 10.1016/j.jvs.2019.01.092

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Veith FJ, Baum RA, Ohki T, Amor M, Adiseshiah M, Blankensteijn JD, et al. Nature and significance of endoleaks and endotension: Summary of opinions expressed at an international conference. J Vasc Surg. (2002) 35:1029–35. doi: 10.1067/mva.2002.123095

PubMed Abstract | CrossRef Full Text | Google Scholar

20. van Marrewijk CJ, Fransen G, Laheij RJ, Harris PL, Buth J, EURO- STAR Collaborators. Is a type II endoleak after EVAR a harbinger of risk? Causes and outcome of open conversion and aneurysm rupture during follow-up. Eur J Vasc Endovasc Surg. (2004) 27:128–37. doi: 10.1016/j.ejvs.2003.10.016

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Sidloff DA, Stather PW, Choke E, Bown MJ, Sayers RD. Type II endoleak after endovascular aneurysm repair. Br J Surg. (2013) 100:1262–70. doi: 10.1002/bjs.9181

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Touma J, Coscas R, Javerliat I, Colacchio G, Goëau-Brissonnière O, Coggia M. A technical tip for total laparoscopic type II endoleak repair. J Vasc Surg. (2015) 61:817–20. doi: 10.1016/j.jvs.2014.11.002

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Uthoff H, Katzen BT, Gandhi R, Peña CS, Benenati JF, Geisbüsch P. Direct percutaneous sac injection for postoperative endoleak treatment after endovascular aortic aneurysm repair. J Vasc Surg. (2012) 56:965–72. doi: 10.1016/j.jvs.2012.03.269

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Hinchliffe RJ, Singh-Ranger R, Whitaker SC, Hopkinson BR. Type II endoleak: Transperitoneal sacotomy and ligation of side branch endo- leaks responsible for aneurysm sac expansion. J Endovasc Ther. (2002) 9:539–42. doi: 10.1177/152660280200900425

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Jones JE, Atkins MD, Brewster DC, Chung TK, Kwolek CJ, LaMuraglia GM, et al. Persistent type 2 endoleak after endovascular repair of abdominal aortic aneurysm is associated with adverse late outcomes. J Vasc Surg. (2007) 46:1–8. doi: 10.1016/j.jvs.2007.02.073

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Ben Abdallah I, El Batti S, Abou-Rjeili M, Fabiani JN, Julia P, Alsac JM. Open conversion after endovascular abdominal aneurysm repair: An 8 year single centre experience. Eur J Vasc Endovasc Surg. (2017) 53:831–6. doi: 10.1016/j.ejvs.2017.03.002

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Klonaris C, Lioudaki S, Katsargyris A, Psathas E, Kouvelos G, Doulaptsis M, et al. Late open conversion after failed endovascular aortic aneurysm re- pair. J Vasc Surg. (2014) 59:291–7. doi: 10.1016/j.jvs.2013.07.106

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Raptis A, Xenos M, Georgakarakos E, Kouvelos G, Giannoukas A, Matsagkas M. Hemodynamic profile of two aortic endografts accounting for their postimplantation position. ASME J Med Devices. (2017) 11:021003–021003-8. doi: 10.1115/1.4035687

CrossRef Full Text | Google Scholar

29. Massoni CB, Gargiulo M, Freyrie A, Gallitto E, Matteis MDE, Mascoli C, et al. Abdominal aortic bifurcation anatomy and endograft limb size affect the use of adjunctive iliac stenting after bifurcated endograft deployment for abdominal aortic aneurysm. J Cardiovasc Surg. (2018) 59:237–42. doi: 10.23736/S0021-9509.16.08871-6

CrossRef Full Text | Google Scholar

30. Sultan S, Acharya Y, Parodi JC, Hynes N. Limits to endovascular approaches in aortic disease. In: Settembrini P, Settembrini A, editors. Vascular surgery: A clinical guide to decision-making. 1st ed. Cambridge, MA: Academic Press (2021). p. 360.

31. Morris L, Stefanov F, Hynes N, Diethrich EB, Sultan S. An experimental evaluation of device/arterial wall compliance mismatch for four stent-graft devices and a multi-layer flow modulator device for the treatment of abdominal aortic aneurysms. Eur J Vasc Endovasc Surg. (2016) 51:44–55. doi: 10.1016/j.ejvs.2015.07.041

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Stather PW, Ferguson J, Awopetu A, Boyle JR. Meta-analysis of renal function following infrarenal EVAR using suprarenal or infrarenal fixation devices. Eur J Vasc Endovasc Surg. (2018) 56:486e96. doi: 10.1016/j.ejvs.2018.01.021

CrossRef Full Text | Google Scholar

33. Kelso RL, Lyden SP, Butler B, Greenberg RK, Eagleton MJ, Clair DG. Late conversion of aortic stent grafts. J Vasc Surg. (2009) 49:589–95. doi: 10.1016/j.jvs.2008.10.020

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Jimenez JC, Moore WS, Quinones-Baldrich WJ. Acute and chronic open conversion after endovascular aortic aneurysm repair: A 14-year review. J Vasc Surg. (2007) 46:642–7. doi: 10.1016/j.jvs.2007.05.030

PubMed Abstract | CrossRef Full Text | Google Scholar

35. EVAR Trial Participants. Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): Randomised controlled trial. Lancet. (2005) 365:2179–86. doi: 10.1016/S0140-6736(05)66627-5

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Terramani TT, Chaikof EL, Rayan SS, Lin PH, Najibi S, Bush RL, et al. Secondary conversion due to failed endovascular abdominal aortic aneurysm repair. J Vasc Surg. (2003) 38:473–8. doi: 10.1016/S0741-5214(03)00417-8

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Jacobowitz GR, Lee AM, Riles TS. Immediate and late explantation of endovascular aortic grafts: The endovascular technologies experience. J Vasc Surg. (1999) 29:309–16. doi: 10.1016/S0741-5214(99)70383-6

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Mohapatra A, Robinson D, Malak O, Madigan MC, Avgerinos ED, Chaer RA, et al. Increasing use of open conversion for late complications after endovascular aortic aneurysm repair. J Vasc Surg. (2019) 69(6):1766–75. doi: 10.1016/j.jvs.2018.09.049

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Kansal V, Nagpal S, Jetty P. Editor's choicedlate open surgical conversion after endovascular abdominal aortic aneurysm repair. Eur J Vasc Endovasc Surg. (2018) 55:163–9. doi: 10.1016/j.ejvs.2017.10.011

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Black SA, Carrell TW, Bell RE, Waltham M, Reidy J, Taylor PR. Long-term surveillance with computed tomography after endovascular aneurysm repair may not be justified. Br J Surg. (2009) 96(11):1280–3. doi: 10.1002/bjs.6732

PubMed Abstract | CrossRef Full Text | Google Scholar

41. Dias NV, Riva L, Ivancev K, Resch T, Sonesson B, Malina M. Is there a benefit of frequent CT follow-up after EVAR? Eur J Vasc Endovasc Surg. (2009) 37(4):425–30. doi: 10.1016/j.ejvs.2008.12.019

PubMed Abstract | CrossRef Full Text | Google Scholar

42. Patel R, Sweeting MJ, Powell JT, Greenhalgh RM, EVAR trial investigators. Endovascular versus open repair of abdominal aortic aneurysm in 15-years’ follow-up of the UK endovascular aneurysm repair trial 1 (EVAR trial 1): A randomised controlled trial. Lancet. (2016) 388(10058):2366–74. doi: 10.1016/S0140-6736(16)31135-7

PubMed Abstract | CrossRef Full Text | Google Scholar

43. Schanzer A, Messina LM, Ghosh K, Simons JP, Robinson WP III, Aiello FA, et al. Follow-up compliance after endovascular abdominal aortic aneurysm repair in medicare beneficiaries. J Vasc Surg. (2015) 61(1):16–22.e1. doi: 10.1016/j.jvs.2014.06.006

PubMed Abstract | CrossRef Full Text | Google Scholar

44. Mirza TA, Karthikesalingam A, Jackson D, Walsh SR, Holt PJ, Hayes PD, et al. Duplex ultrasound and contrast-enhanced ultrasound versus computed tomography for the detection of endoleak after EVAR: Systematic review and bivariate meta-analysis. Eur J Vasc Endovasc Surg. (2010) 39:418–28. doi: 10.1016/j.ejvs.2010.01.001

PubMed Abstract | CrossRef Full Text | Google Scholar

45. Perini P, Sediri I, Midulla M, Delsart P, Mouton S, Gautier C, et al. Single-centre prospective comparison between contrast-enhanced ultrasound and computed tomography angiography after EVAR. Eur J Vasc Endovasc Surg. (2011) 42:797–802. doi: 10.1016/j.ejvs.2011.09.003

PubMed Abstract | CrossRef Full Text | Google Scholar

46. Arnaoutakis DJ, Sharma G, Blackwood S, Shah SK, Menard M, Ozaki CK, et al. Strategies and outcomes for aortic endograft explantation. J Vasc Surg. (2019) 69:80–5. doi: 10.1016/j.jvs.2018.03.426

PubMed Abstract | CrossRef Full Text | Google Scholar

47. Juraszek A, Rylski B, Kondov S, Scheumann J, Kreibich M, Morlock J, et al. Late surgical conversions after abdominal endovascular aortic repair: Underlying mechanisms, clinical results and strategies for prevention. Interact CardioVasc Thorac Surg. (2019) 29:944–9. doi: 10.1093/icvts/ivz207

PubMed Abstract | CrossRef Full Text | Google Scholar

48. Vukovic E, Czerny M, Beyersdorf F, Wolkewitz M, Berezowski M, Siepe M, et al. Abdominal aortic aneurysm neck remodeling after Anaconda stent graft implantation. J Vasc Surg. (2018) 68:1354–9.e2. doi: 10.1016/j.jvs.2018.02.041

PubMed Abstract | CrossRef Full Text | Google Scholar

49. Rylski B, Czerny M, Südkamp M, Russe M, Siep M, Beyersdorf F. Fenestrated and branched aortic grafts. Dtsch Arztebl Int. (2015) 112:816–22. doi: 10.3238/arztebl.2015.0816

PubMed Abstract | CrossRef Full Text | Google Scholar

50. Ito E, Toya N, Fukushima S, Nishie R, Akiba T, Ohki T. Polyester grafts are a risk factor for postimplantation syndrome after abdominal endovascular aneurysm repair: Retrospective analysis for polyester graft, excluder®, and endologix powerlink®/AFX®. Ann Vasc Dis. (2018) 11:520–4. doi: 10.3400/avd.oa.18-00058

PubMed Abstract | CrossRef Full Text | Google Scholar

51. Voûte MT, Bastos Gonçalves FM, van de Luijtgaarden KM, Klein Nulent CG, Hoeks SE, Stolker RJ, et al. Stent graft composition plays a material role in the postimplantation syndrome. J Vasc Surg. (2012) 56:1503–9. doi: 10.1016/j.jvs.2012.06.072

CrossRef Full Text | Google Scholar

52. Sartipy F, Lindström D, Gillgren P, Ternhag A. The impact of stent graft material on the inflammatory response after EVAR. Vasc Endovascular Surg. (2015) 49:79–83. doi: 10.1177/1538574415595209

PubMed Abstract | CrossRef Full Text | Google Scholar

53. Syk I, Brunkwall J, Ivancev K, Lindblad B, Montgomery A, Wellander E, et al. Postoperative fever, bowel ischaemia and cytokine response to abdominal aortic aneurysm repair–a comparison between endovascular and open surgery. Eur J Vasc Endovasc Surg. (1998) 15:398–405. doi: 10.1016/S1078-5884(98)80200-1

PubMed Abstract | CrossRef Full Text | Google Scholar

54. Swartbol P, Truedsson L, Pärsson H, Norgren L. Tumor necrosis factor-alpha and interleukin-6 release from white blood cells induced by different graft materials in vitro are affected by pentoxifylline and iloprost. J Biomed Mater Res. (1997) 36:400–6. doi: 10.1002/(SICI)1097-4636(19970905)36:3<400::AID-JBM15>3.0.CO;2-H

PubMed Abstract | CrossRef Full Text | Google Scholar

55. Perini P, Gargiulo M, Silingardi R, Bonardelli S, Bellosta R, Bonvini S, et al., on behalf of the LOCOS-1 investigators, Twenty-two year multicentre experience of late open conversions after endovascular abdominal aneurysm repair. Eur J Vasc Endovasc Surg. (2020) 59:757–65. doi: 10.1016/j.ejvs.2020.01.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: endovascular procedures, complications, reintervention, explantation, graft preservation

Citation: Sultan S, Acharya Y, Hezima M, Chua Vi Long K, Soliman O, Parodi J and Hynes N (2022) Two decades of experience in explantation and graft preserving strategies following primary endovascular aneurysm repair and lessons learned. Front. Surg. 9:963172. doi: 10.3389/fsurg.2022.963172

Received: 7 June 2022; Accepted: 26 July 2022;
Published: 9 August 2022.

Edited by:

Efstratios Georgakarakos, University Hospital of Alexandroupolis, Democritus University of Thrace, Greece

Reviewed by:

Konstantinos Moulakakis, University of Patras, Greece
Yuji Kanaoka, Kawasaki Medical School, Japan
Vincent Rowe, Keck School of Medicine, University of Southern California, United States

© 2022 Sultan, Acharya, Hezima, Chua Vi Long, Soliman, Parodi and Hynes. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Sherif Sultan sherif.sultan@hse.ie, sherif.sultan@nuigalway.ie

ORCID Sherif Sultan orcid.org/0000-0001-8767-4929 Yogesh Acharya orcid.org/0000-0003-1829-5911 Niamh Hynes orcid.org/0000-0002-6133-3322

Speciality Section: This article was submitted to Vascular Surgery, a section of the journal Frontiers in Surgery

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.