Skip to main content

CASE REPORT article

Front. Surg., 20 September 2022
Sec. Orthopedic Surgery

Entrapment of a metal foreign body in the heart during surgical procedure: A case report and literature review

\r\nFei Wang,&#x;Fei Wang1,†Ting Li,,&#x;Ting Li1,2,†Xinwei YuanXinwei Yuan1Jiang Hu
\r\nJiang Hu1*
  • 1The Orthopaedic Surgery Department of Sichuan Academy of Medical Sciences / Sichuan Provincial People’s Hospital, The Affiliated Hospital of University of Electronic Science and Technology of China, Chengdu, China
  • 2Department of Postgraduate, Chengdu Medical College, Chengdu, China

A rongeur had been used to remove thin bones in both orthopedic surgery and neurosurgery, featured with a tip holding and cutting bone effectively while protecting the underlying instruments. The authors describe a case of a 40-year-old man who proceeded with the second lumbar vertebrae osteotomy and presented to be ankylosing spondylitis with kyphosis and limited mobility for 10 years. During the surgery, we found that the rongeur tip was missing. C-arm fluoroscopy showed the high-density body just in front of the vertebral body intraoperatively. However, the CT scan showed the foreign body migrated to the right auricle of the heart postoperatively. This case is unique in that there was no exact vessel injury detected intraoperatively. There were few reports about the surgical instrument migrating to the heart. Our case showed the rare experience of the function of multidisciplinary collaboration in the migration of foreign bodies in the cervical spinal canal.

Introduction

A rongeur had been used to remove thin bones in both orthopedic surgery and neurosurgery, featured with a tip holding and cutting bone effectively while protecting the underlying instruments. Our team first reported a rare case where the tip was missing during the surgical procedure and migrated to the heart.

In this case, our team used posterior approach to correct the spine kyphosis deformity. When the instrument is broken and entrapped in the spinal canal during the surgical procedure, immediate removal is not possible. The tip finally migrated to the heart. It is lucky to remove the foreign body successfully without severe outcomes by a 6-month follow-up.

We report this case to warn that retaining foreign objects in the spinal canal is very dangerous. Checking and replacing orthopedic surgery instruments frequently is necessary. In addition, the design for the new rongeur should also consider durability.

Case report

Patient presentation

A male patient, aged 40 years, presented with intractable back pain for 10 years. He was not able to lie on the floor for 3 years. As such, he was referred to us for surgery. Whole spine lateral radiographs revealed spine sagittal imbalance (Figures 1A,D), preoperative computed tomography (CT) scan revealed the disappearance of bilateral sacroiliac joint space and the calcification of the vertebral ligaments, while the BLA-27 test was positive. The patient was diagnosed as having ankylosing spondylitis (AS) with kyphosis.

FIGURE 1
www.frontiersin.org

Figure 1. Male, aged 40, ankylosing spondylitis with kyphosis, preoperative whole spine lateral radiographs showed thoracolumbar kyphosis, spine sagittal imbalance (A), cobb angle 40° (D). The spine sagittal rebalanced (B,C) and the Cobb angle corrected to 20° (E,F), tested 1 month and 6 months after operation, respectively.

Clinical procedure

The patient underwent second lumbar vertebrae osteotomy, decompression, and stabilization in a prone position. During the decompression, the surgeon noted that the Kerrison rongeur was withdrawn without cutting edge and the tip was not visible within the incision (Figure 2). Meanwhile, little blood was encountered and the patient didn't experience hemodynamic compromise. Intraoperative immediate C-arm fluoroscopy manifested that there was a high-density shadow just in the right anterior between the fourth and fifth lumbar vertebrae (Figure 3A). By consultation with cardiac surgeons, vascular surgeons, and gastrointestinal surgeons, immediate anterior exploration was not recommended giving the hemodynamic stability. As per their recommendation, the referring surgeon completed the spine surgery without cerebrospinal fluid leakage while there was 1,000 ml bleeding. At the end of the surgery, we tried to confirm the location of the tip by C-arm fluoroscopy. The tip was found in the left thoracic space between the seventh and eighth rib (Figure 3C). We considered that the tip migrated to the heart through the blood circulation system.

FIGURE 2
www.frontiersin.org

Figure 2. The broken rongeur (arrow) was found in the operation (A). The tip was not invisible in the incision (B).

FIGURE 3
www.frontiersin.org

Figure 3. Intraoperative x-ray revealed that the broken metal tip (arrow) was located in the right anterior between the fourth and fifth lumbar vertebrae (A). After L2 vertebral body osteotomy, the metal fragment was found to migrate to the heart (B,C). MR imaging showed abnormal internal vertebral venous plexus in the dorsal part of the 5th lumbar vertebral body, on T1W (D), T2W (E), and axial (F) images. MR, magnetic resonance.

After surgery, the patient was taken for a dual-source computed tomography angiography (CTA), which revealed that the rongeur tip was located in the auricula dextra (Figure 4A). The patient was transferred to the cardiac surgery department by consulting the cardiologists. Considering the irregular size of the surgical instruments and the difficult technique for retrieval, the cardiac surgery team chose open-heart surgery. The patient underwent open-heart surgery to remove the foreign body (Figure 4). The rongeur tip, about 1 cm in length (Figure 4), was found in the auricula dextra, close to the right atrioventricular groove and right coronary artery.

FIGURE 4
www.frontiersin.org

Figure 4. Radiation and dual-source CTA showed that the rongeur tip was in the auricula dextra (A–C). The tip was found in the auricula dextra close to the right atrioventricular groove (D). Image of the extracted rongeur tip (E). CT was used to confirm (F).

The low back pain and kyphosis were relieved immediately without any signal of root damage and the patient could lie flat after surgery. Table 1 shows the sagittal score changes of the spine sagittal compared between the preoperative and postoperative stages according to the radiographs (Figure 1). The patient did not show dyspnea, stethalgia, or edema, and Table 2 shows the related data of laboratory inspection after orthopedic surgery and cardiac surgery. The reduction of myoglobin and fibrin and fibrinogen degradation products means that the risk factor is removed successfully, which may induce the thrombus. Echocardiography and coagulation function tests show a good prognosis in the patient's cardiovascular system. The patient returned to normal life 3 months after the surgery. This case was monitored for 6 months. Written informed consent was obtained from the patient for publication of this case report and accompanying images.

TABLE 1
www.frontiersin.org

Table 1. Male, aged 40, ankylosing spondylitis with kyphosis, the spine sagittal score showed the precise value of the spine sagittal according to the radiographs preoperatively, at 1 month, and at 6 months.

TABLE 2
www.frontiersin.org

Table 2. The index of the blood coagulation and cardiomyocyte injury markers were shown preoperatively, at 1 month, and at 3 months after open-heart surgery to guide the treatment and the prognosis.

Discussion

Physiologists often design instruments with a purpose to make the operation more effective, but the durability of surgical instruments is always neglected. Breakage of instruments during orthopedic procedures reportedly occurs in 0.35% of cases (1). Taking the rongeur as an example, the cutting edge could be designed flatter and smaller in order to make it easy to insert into the layer between the bone and spinal dural. In the operation, the broken tip may be rushed into the spinal canal or even into the abnormal vertebral vessel under pressure. In the spinal canal, the dural sac pulses with the heart beats and the dural pressure change in different postures (2). As such, the spring phenomenon likely pushes the foreign body to migrate or penetrate the intradural space, resulting in encroachment upon the spinal cord or nerve roots (3). Luckily, the patient did not show symptoms of nerve damage after spine surgery. However, while the foreign body is lost in the invisible spine canal and moved in the vertebral canal, it is difficult to withdraw the instrument in the limited cut. Lv et al. had reported the case that keeps the foreign body in the canal without neurological compression (4). However, in our case, the tip is big as 1 cm and retained in auricula dextra, making the situation conditionally to remove it from the body. In spite of the related operative complication, the patient reduced the risk of durative myocardial injury and safety to undertake the MRI test to reduce the radiation exposure caused by the CT scan. Inflammation of tendon and ligament attachment point and ectopic new bone formation were two characteristic pathological changes of AS. On the one hand, by the development of trabecular and cortical osteoporosis and, on the other hand, by the overproduction of cortical bone and entheseal ossification. Therefore, AS patients had aberrant calcification of entheses, which might augment the resistance of bone (5). Because using rongeur orthopedic instruments for a long time will cause certain damage, it is noticed that checking and replacing the orthopedic surgery instruments frequently is necessary, especially for patients with osteosclerosis. In addition, the design for a new rongeur or other orthopedic instruments should also consider durability (6).

Vascular injury is one of the complications, although rare, and the rate ranges from 0.04% to 1.11% in the literature (7, 8). In many studies, the posterior approach was reported safer in reducing the vascular injury to proceed with the interbody fusions. Uribe and Deukmedjian (9) reported 0.1% vascular injury following over 13,000 lateral interbody fusions, while in the paper by Liu (10), it was 0.29% for post-exposure in retrospectively reviewed lumbar spine surgery 1,159 patients. He demonstrated that the incidence was higher in anterior exposure (9.1%) than in posterior exposure. Zahradnik et al. (11) showed a vascular injury in 37 cases of 260 patients during anterior exposure of the thoracolumbar spine (13.8%). Klezl et al. (12) indicated that injury to a major vessel was encountered in 14 (1.11%) cases of 1,262 patients, of which 9 involved an injury to the common iliac vein. As shown in Figure 3, we found that there is a signal of abnormal internal vertebral venous plexus in the dorsal part of the fifth lumbar vertebral body. It explained how the foreign body migrated to the heart while the patient did not experience hemodynamic compromise and no damaged port of the inferior vena cava was found. During the surgery, the blood loss was 1000 ml and 900 ml post operation, which we considered as blood from vertebral bone. “Batson's plexus” was also known as the intraspinal venous plexus. In the spinal canal, the dural pressure changed with heartbeats and respiratory movements, which might allow the tip to enter the “Batson's plexus” with the changes in dural sac pressure. Moreover, we found that there is a signal of abnormal internal vertebral venous plexus in the dorsal part of the fifth lumbar vertebral body, which also might be a significant reason to allow the tip to enter “Batson's plexus.” The tip passes through the Batson plexus to the inferior vena cava and finally to the right atrium. Therefore, we consider that the abnormal internal vertebral venous plexus is responsible for the rare case in the patient with both hemorrhage and foreign body movement.

The guideline of treatment for foreign body migration to the heart was not established due to the complicated situation in these cases. In our case, we found that a foreign body entered the cardiovascular system and reached the right atrium. This is unique. A review of the literature reveals that most of the foreign bodies that migrate to the heart are endovascular devices, such as catheters, guidewires, or pacemakers (13, 14). Naito et al. reported a case of a piece of rubber catheter that migrated to the inferior vena cava during surgery (15). Leonardi and Rivera adopted conservative treatment for their broken cerclage wire in the left heart case considering the high risk of surgical left ventriculotomy associated with searching for the wire that had migrated into the myocardial wall (13). However, the instrument migrating into the heart during spine surgery was less reported so far, and a less-invasive endovascular retrieval would have been preferred in most cases. Bydon et al. recommend percutaneous endovascular or surgical intervention when the diameter of the foreign object exceeds 5 mm, its shape is irregular, or when the patient exhibits clinical symptoms (16). However, the rongeur tip has about 1 cm in length, and its shape is irregular. Minimally invasive surgery might carry greater risks, so we opted for open surgery. The evidence showed that the body migrated to the inferior vena cava from the abnormal internal vertebral venous plexus. In a few reported cases, the retained foreign bodies typically lead to delayed complications within several months or years, including pain, infections, and organ dysfunction. Additionally, migration to the heart of foreign bodies can lead to arrhythmia, perforation, endocarditis, valvular malfunctions, and pericarditis. Some of the complications may happen years later after the initial event (17). On the other hand, some of the complications can be life threatening and be emergency events (18). The foreign body needs to be CT scanned every month while MRI would make the body migrate and cause damage while it migrated into the apparatus; as such, it results in a tumorous tendency. The presentation of foreign bodies varies greatly, and the object that is ingested, the location of the object, whether the event was witnessed, the age of the patient, as well as the timeframe in which ingestion occurred are very important (19). For the same considering factors in our case, the patient was young and in good health, and we have identified what was ingested (a broken rongeur tip) by C-arm. In addition, the location of the object migrated from the Batson's plexus” to the right atrium. Finally, the timeframe in which ingestion occurred is well defined. Therefore, open surgery was chosen as a definitive method for broken rongeur tip retrieval. The patient is lucky to have the foreign body removed without a severe outcome. The metal piece was successfully removed from the auricula dextra by the cardiologist without cardiovascular complications in a 6-month follow-up.

Conclusion

Broken rongeur tip migration to the heart during spine surgery is a rare complication. Migration can lead to arrhythmia, perforation, endocarditis, valvular malfunctions, and pericarditis. Although the complication is rare, serious consideration should be given to the removal of the broken rongeur tip in the heart space when they are discovered. Our case indicates the necessity of rapid diagnosis and the need for multidisciplinary collaboration and care in the treatment. For young and healthy patients, we suggest surgical intervention to be the treatment.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Ethics statement

Ethical review and approval were not required for the study on human participants in accordance with the local legislation and institutional requirements. Written informed consent from the patients was not required to participate in this study in accordance with the national legislation and institutional requirements. Written informed consent was obtained from the individual for the publication of any potentially identifiable images or data included in this article.

Author contributions

TL wrote the manuscript and produced the figures. JH and FW performed the surgery. XY, FW, and JH were responsible for English editing and reviewing the manuscript. All authors contributed to the article and approved the submitted version.

Funding

This study was supported by the Foundation Project of Sichuan Provincial People’s Hospital (2021LY17).

Acknowledgments

Thanks to the colleagues in the Radiology Department of Sichuan Academy of Medical Sciences / Sichuan Provincial People's Hospital for their support in providing operative x-ray and chest computed tomography. We also appreciate the collaborative work of colleagues in the Cardiac Surgery Department and the Intensive Care Unit of the Sichuan Academy of Medical Sciences / Sichuan Provincial People’s Hospital.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Dominguez ED, Rocos B. Patient safety incidents caused by poor quality surgical instruments. Cureus. (2019) 11(6):e4877. doi: 10.7759/cureus.4877

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Guan X, Wu X, Fan G, Zhao S, Gu G, Zhang H, et al. Endoscopic retrieval of a broken guidewire during spinal surgery. Pain Physician. (2016) 19(2):E339–42.26815261

PubMed Abstract | Google Scholar

3. Leiderman DBD, Zerati AE, Wolosker N, Hoffmann Melo HA, da Silva ES, De Luccia N. Endovascular treatment of penetrating injury to the vertebral artery by a stab wound: case report and literature review. Ann Vasc Surg. (2017) 45:267.e1–e5. doi: 10.1016/j.avsg.2017.06.145

CrossRef Full Text | Google Scholar

4. Lv X, Lu X, Wang Y. Entrapment of a metal foreign body in the cervical spinal canal during surgical procedure: a case report. Medicine. (2018) 97(17):e0548. doi: 10.1097/MD.0000000000010548

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Talotta R, Rucci F, Scaglione F. Calcium physiology, metabolism and supplementation: a glance at patients with ankylosing spondylitis. Reumatologia. (2020) 58(5):297–311. doi: 10.5114/reum.2020.100112

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Menger RP, Lazzari Z, Nanda A. Stepwise retrieval of a broken pituitary disc rongeur. Cureus. (2020) 12(6):e8561. doi: 10.7759/cureus.8561

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Postacchini R, Cinotti G, Postacchini F. Injury to major abdominal vessels during posterior lumbar interbody fusion. A case report and review of the literature. Spine J. (2013) 13(1):e7–e11. doi: 10.1016/j.spinee.2012.11.016

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Charosky S, Guigui P, Blamoutier A, Roussouly P, Chopin D. Complications and risk factors of primary adult scoliosis surgery: a multicenter study of 306 patients. Spine. (2012) 37(8):693–700. doi: 10.1097/BRS.0b013e31822ff5c1

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Uribe JS, Deukmedjian AR. Visceral, vascular, and wound complications following over 13,000 lateral interbody fusions: a survey study and literature review. Eur Spine J. (2015) 24(Suppl 3):386–96. doi: 10.1007/s00586-015-3806-4

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Liu Y. Analysis of vascular injury in lumbar spine surgery. Pak J Med Sci. (2012) 28:791–4.

Google Scholar

11. Zahradnik V, Lubelski D, Abdullah KG, Kelso R, Mroz T, Kashyap VS. Vascular injuries during anterior exposure of the thoracolumbar spine. Ann Vasc Surg. (2013) 27(3):306–13. doi: 10.1016/j.avsg.2012.04.023

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Klezl Z, Swamy GN, Vyskocil T, Kryl J, Stulik J. Incidence of vascular complications arising from anterior spinal surgery in the thoraco-lumbar spine. Asian Spine J. (2014) 8(1):59–63. doi: 10.4184/asj.2014.8.1.59

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Leonardi F, Rivera F. Intravascular migration of a broken cerclage wire into the left heart. Orthopedics. (2014) 37(10):e932–5. doi: 10.3928/01477447-20140924-90

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Schechter MA, O’Brien PJ, Cox MW. Retrieval of iatrogenic intravascular foreign bodies. J Vasc Surg. (2013) 57(1):276–81. doi: 10.1016/j.jvs.2012.09.002

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Naito N, Abe M, Fukasawa M, Takeshi A. Pulmonary embolism by a foreign body that migrated in the inferior vena cava during lumbar spine surgery. BMJ Case Rep. (2014) 2014:bcr2014205423. doi: 10.1136/bcr-2014-205423

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Bydon A, Xu R, Conte JV, Gokaslan ZL, Brinker JA, Witham TF, et al. Surgical mystery: where is the missing pituitary rongeur tip?. Spine (Phila Pa 1976). (2010) 35(17):E867–E72. doi: 10.1097/BRS.0b013e3181d85bec

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Steiner J, Dhingra R, Devries JT. Needle in the haystack: purulent pericarditis from injection drug use. Catheter Cardiovasc Interv. (2012) 80(3):493–6. doi: 10.1002/ccd.23259

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Ishikawa K, Hayashi H, Mori H. Cardiac perforation caused by cement after percutaneous balloon kyphoplasty. Asian Cardiovasc Thorac Ann. (2017) 25(3):204–6. doi: 10.1177/0218492315613697

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Mulita F, Panagiotopoulos I, Verras GI, Liolis E, Tchabashvili L, Paraskevas T, et al. Accidental ingestion of a dental bur in an 84-year-old male. Clin Case Rep. (2022) 10(3):e05488. doi: 10.1002/ccr3.5488

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: foreign body migration, vascular complication, orthopedic surgery, complication, multidisciplinary collaboration

Citation: Wang F, Li T, Yuan X and Hu J (2022) Entrapment of a metal foreign body in the heart during surgical procedure: A case report and literature review. Front. Surg. 9:963021. doi: 10.3389/fsurg.2022.963021

Received: 7 June 2022; Accepted: 29 August 2022;
Published: 20 September 2022.

Edited by:

Marina Leitman, Yitzhak Shamir Medical Center, Israel

Reviewed by:

Leopoldo Mandic Furtado, Federal University of Minas Gerais, Brazil
Francesk Mulita, General University Hospital of Patras, Greece

© 2022 Wang, Li, Yuan and Hu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Jiang Hu hujiang8711@163.com

These authors have contributed equally to this work and share first authorship

Specialty Section: This article was submitted to Orthopedic Surgery, a section of the journal Frontiers in Surgery

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.