This study aimed to introduce a sternoclavicular joint (SCJ)-specific plate for the treatment of medial-end clavicle fracture and evaluate the clinical and radiological results of this method.
From January 2006 to December 2020, 31 patients with displaced medial-end clavicle fractures were included in this study, with 8 patients with accompanying SCJ dislocation. Abduction and forward elevation of the shoulder, the Visual Analogue Scale (VAS), and the American Shoulder and Elbow Surgeons Score (ASES) were used for evaluation before index surgery and at the latest follow-up.
After an average of 98.5 (range, 13 to 171) months, the mean VAS significantly decreased from 6.8 ± 1.0 preoperatively to 0.9 ± 0.8 at the latest follow-up (
This SCJ-specific plate provided excellent long-term results for the treatment of displaced medial-end clavicle fractures and was an alternative implant for medial-end clavicle fractures with or without small or comminuted medial fragments, especially those associated with SCJ dislocation.