AUTHOR=Krout Adam J. , Weissinger Mathew J. , Mertens Joseph C. , Allyn Katheryn J. , Larsen Brian G. , McCarthy Nicholas K. , Garbini Joseph L. , Sanders Joan E. TITLE=Distal weight bearing in transtibial prosthesis users wearing pin suspension JOURNAL=Frontiers in Rehabilitation Sciences VOLUME=4 YEAR=2023 URL=https://www.frontiersin.org/journals/rehabilitation-sciences/articles/10.3389/fresc.2023.1322202 DOI=10.3389/fresc.2023.1322202 ISSN=2673-6861 ABSTRACT=Introduction

Low-level distal weight bearing in transtibial prosthesis users may help maintain perfusion and improve both proprioception and residual limb tissue health.

Methods

The primary objectives of this research were to develop a sensor to continuously measure distal weight bearing, evaluate how prosthesis design variables affected weight bearing levels, and assess fluctuations in distal weight bearing during at-home and community use.

Results

In-lab testing on a small group of participants wearing adjustable sockets demonstrated that if distal contact was present, when socket size was increased distal weight bearing increased and when socket size was reduced distal weight bearing decreased. During take-home use, participants accepted the distal weight bearing level set by the research team. It ranged between 1.1% and 6.4% BW for all days tested. The coefficient of variation (standard deviation/mean) ranged from 25% to 43% and was expected due in part to differences in walking style, speed, terrain, direction of ambulation, and bout duration. Two participants commented that they preferred presence of distal weight bearing to non-presence.

Discussion

Next steps in this research are to develop clinical practices to determine target distal weight bearing levels and ranges, and to simplify the design of the sensor and weight bearing adjustment mechanism for clinical use.