CASE REPORT article

Front. Public Health, 03 November 2023
Sec. Infectious Diseases: Epidemiology and Prevention

Continuous false positive results by SARS-CoV-2 rapid antigen testing: a case report

Yannick GalipeauYannick Galipeau1Abishek XavierAbishek Xavier1Aaron DyksAaron Dyks2Curtis CooperCurtis Cooper2Marc-Andr Langlois,
Marc-André Langlois1,3*
  • 1Department of Biochemistry, Microbiology & Immunology, Faculty of Medicine, University of Ottawa, Ottawa, ON, Canada
  • 2The Ottawa Hospital Research Institute, Ottawa, ON, Canada
  • 3University of Ottawa Centre for Infection, Immunity and Inflammation (CI3), Ottawa, ON, Canada

Efficient and rapid identification of active SARS-CoV-2 infections has been key to monitoring and mitigating the spread of the virus. The implementation of nucleic acid testing (e.g., RT-PCR) was broadly adopted by most public health organizations at the national and community levels across the globe, which was followed by more accessible means of home testing including lateral flow immunochromatographic assay (LFA), also known as a rapid antigen test. Here we report the case of an adult female who repeatedly and consecutively tested positive by RAT (BTNX inc). This sustained false positive was not linked with an active SARS-CoV-2 infection, which was ruled out by RT-PCR and serological analyses. SARS-CoV-2 serology revealed no detectable levels of antibodies against the nucleocapsid suggesting no recent prior infection by SARS-CoV-2. This continuous false positive was limited to BTNX testing devices. This case report aims to describe that such continuous false positives can occur and describes alternative testing approaches that can be performed to confirm RAT results. In addition, broader awareness of such occurrences is warranted in the healthcare and public health community to avoid unnecessary negative impacts on individual’s day to day life.

Introduction

The ability to rapidly identify SARS-CoV-2 infected individuals has been central to most public health measures across the globe and is a major component to contact tracing initiatives. Population-level testing requirements prompted the establishment of testing centers that conducted viral nucleic acid detection (i.e., RT-PCR) (1, 2). Despite being the gold standard detection method of SARS-CoV-2 in clinical samples, and its broad use early in the pandemic, RT-PCR testing has some important limitations (i.e., expensive laboratory equipment, reliance on highly qualified laboratory and medical workers, non-point-of-care testing, etc.). In light of these limitations, countries like Canada have endorsed the usage of at-home Rapid Antigen Tests (RAT) (35).

Several studies have investigated RAT performances in real life settings. In a cohort of 72,382 paired RT-PCR and RAT, Eyre et al. reported an overall sensitivity of 63.2% (6). This estimate is in line with other studies (7, 8). While sensitivity is increased in symptomatic individuals, RATs overall have a lower sensitivity compared to RT-PCR. In addition, despite concerns that SARS-CoV-2 RAT could cross-react with other pathogens such as seasonal coronaviruses (OC43, 229E, NL63, HKU-1), RATs are regarded as highly specific. In fact, several studies reported specificities that ranged from 99.71% (6), 99.95% (9) 100% (10).

Given that a vast array of RATs have been authorized by the federal regulatory bodies, their relative real-world performance, technology used, and their usage vary (10, 11). In this case report, we will focus primarily on the RAT from BTNX inc (COV-19C25). This lateral immunochromatographic at-home test kit is designed to detect SARS-CoV-2 nucleocapsid in nasopharyngeal secretions (12). Here we report the case of a 47 year old female who consistently and consecutively tested positive by RAT despite any laboratory or clinical evidence of SARS-CoV-2 infection.

Case description

A 47-year-old Caucasian female first tested positive by rapid antigen test (RAT) on December 28th 2021. At that time, she remained positive for over 2 weeks until she stopped self-testing. Following Canadian public health recommendations, she tested herself regularly by RAT after a high-risk contact, or when experiencing COVID-19 associated symptoms. Repeated self-testing with RAT yielded a constant positive result. After each positive RAT result, she followed recommendations from public health to self-isolate. She was careful about masking and wore N95 in public spaces. Given that RATs were repeatedly positive, she contacted public health authorities for guidance and they recommended she stop testing. On the other hand, her family physician told her to keep testing, thereby suggesting that she was still contagious. These conflicting recommendations have led to confusion, distress and uncertainty for the subject. The numerous isolation periods have impacted her day-to-day life, impacted her family, and prevented her from visiting friends and relatives.

We were contacted in early fall 2022 to offer guidance on the continuous SARS-CoV-2 positive results on RAT given the mixed recommendations from her physician and local public health. Her medical history included an allergy to dust mites and mild asthma that did not require medication. Her SARS-CoV-2 vaccination history included a first dose of (Pfizer) on May 1st 2021, a second dose of (Moderna) on June 19th 2021, a third dose of (Moderna) on November 19th 2021, and a fourth dose of (Pfizer) on August 9th 2022.

We collected a blood sample to investigate SARS-CoV-2 associated antibodies in serum. We also requested that she provide a saliva sample every week for a month for molecular testing. At the time of providing the saliva sample, we also asked that she perform a set of RAT tests, which included 3 BTNX tests. One test was performed following the manufacturer’s instructions (swabbing of the nasal cavity), the other was a swab of the oral cavity, and the third comprised of combination swabbing techniques of both sites. For the first week, a RAT from another supplier was also performed simultaneously (Figure 1).

FIGURE 1
www.frontiersin.org

Figure 1. Timeline overview of positive RATs and specimen collection. A visual representation of both the subject testimony in regards to the history of positive RATs and sample collection (Blood, RATs) as part of this case report. Red vertical bars represent the four weeks where we documented positive RAT for this case report, and where the subject provided saliva samples.

All four weekly RATs that were performed following manufacturer recommended swabbing techniques or the combination of oral and nasal swabbing techniques yielded unambiguous positive results. All four weekly RATs performed by oral swabbing were negative (Figure 2). Interestingly, a RAT from another supplier (FlowFlex) was performed in the first week and was negative, suggesting that the positivity might be limited to BTNX kits only. Nucleic acid testing from the saliva samples provided for each week was negative for SARS-CoV-2 E and N gene, thus suggesting the absence of an active or persistent infection. SARS-CoV-2 IgG antibody measurements against the N and S proteins were performed. Only antibodies against the Spike protein were detected beyond our positivity threshold (Figure 3). Given the lack of antibodies against the N protein, a recent SARS-CoV-2 infection was deemed unlikely.

FIGURE 2
www.frontiersin.org

Figure 2. Weekly results of RAT by swabbing technique and location. (A) An overview of the RATs by swabbing technique for week 4 from photographed RAT performed at home. As described, the nasal cavity was swabbed (deep and anterior nasal swab), the oral cavity (throat and cheeks) as well as the combination of both swabbing techniques. The RATs were performed sequentially in the same session and photographed. (B) An overview of the RAT’s band intensity in relation to the control band for every week and for the three swabbing techniques.

FIGURE 3
www.frontiersin.org

Figure 3. Serology results of serum sample provided prior to week 1. A blood sample was provided prior to performing the RATs, and serum was isolated. Antibodies against the SARS-CoV-2 (A) nucleocapsid and (B) Spike protein was performed using a previously established chemiluminescent ELISA serological assay. IgG antibodies were measured by titrating the subject’s serum sample (red). In addition, a titration of a negative (blue) and positive (green) pooled serum from residual blood was included as positive and negative controls. Here the blank adjusted relative luminescence units (RLU) intensity is shown. IgG antibody titer against the nucleocapsid was below conversion threshold, IgG titer against the spike protein was estimated to be 130.41 BAU/mL.

Discussion

While false positives occur with any diagnostic tests, including RATs, here we report the case of a continuous false positive. To document the events related to the positive RAT and to investigate potential sources of cross-reactivity, a series of RATs were performed. One weekly RAT was performed for four weeks by anterior and deep nasal swabbing as per manufacturer’s instructions. Each of these RATs were positive, confirming the subject’s testimony (Figure 2B). Given that Canadian public health authorities were also recommending individuals to swab the oral cavity (throat and cheeks) in combination with anterior and deep nasal swabbing to increase sensitivity (13, 14), we also asked that a RAT with both swabbing techniques (oral and nasal swabbing) and oral swabbing alone be performed sequentially in one session. Interestingly, all of the oral cavity RATs were negative (Figure 2). The RATs performed with the combinatorial swabbing technique echoed the results seen with the nasal cavity swabbing technique and were all positive. Notably, the signal intensity of the RATs on week four was higher than previous three weeks. Given that RATs are non-quantitative tests it is difficult to comment if this is caused by a better swabbing technique resulting in a greater amount of biological material applied to the RAT or a true increase in signal. In addition, a RAT from another supplier was performed in parallel on week one and resulted in a negative result. This echoes previous documented RATs from the participant outside of this study’s timeframe where RAT from FlowFlex and Abbott remained negative despite a concurrent positive on BTNX RAT.

While the subject did not display any COVID-19 related symptoms at the time of taking the RATs, we still collected saliva samples to perform nucleic acid testing given that subclinical asymptomatic SARS-CoV-2 infections have been documented (1518). The multiplex qPCR showed no detectable levels of SARS-CoV-2 N and E genes. An internal control (Human RPP30 gene) was also included to validate RNA extraction efficiency from the saliva. These data also align with the fact that no close contacts to the subject tested positive or developed COVID-19 related symptoms. In light of this, the possibility of an active SARS-CoV-2 infection was ruled out.

SARS-CoV-2 antigen deposition in various tissue following infection have been documented previously (19, 20). As such, we next investigated whether the positive RAT could be from deposited nucleocapsid protein from a prior infection. We measured IgG antibodies against the spike protein and the nucleocapsid of SARS-CoV-2 (Wuhan). We calculated an overall titer of 130.41 BAU/mL of anti-spike IgG antibodies, while antibodies against the N protein were below the seroprevalence threshold. Given the lack of antibodies recognizing the nucleocapsid, we can conclude that the spike antibodies are vaccine-elicited and that prior to performing the RAT, this subject had no serological markers suggesting a prior infection. While it is true that antibodies decay with time, it is unlikely that a RAT would detect true nucleocapsid protein without an associated detectable antibody response and/or a positive RT-PCR result.

The source driving the false positive signal on BTNX devices remains unclear. In principle, endogenous peptides could cross-react with BTNX RAT antibodies within the devices. It is also possible that cross-reactivity be driven by exogenous peptides from the microbiome. Given the proprietary nature of the antibodies used in each RATs by the various manufacturers, further investigation on how the antibody selection contributes to false positives is difficult. Anecdotally, the subject’s children and family direct members did not generate a false positive result on RATs.

Conclusion

Continuous false positives can occur with RATs despite their overall high positive predictive value. A confirmatory RAT from another supplier could be used as confirmation. Molecular testing remains the gold standard to confirm active infections. Serological testing can rule out chronic or previous SARS-CoV-2 infections in the vast majority of cases. Increased awareness by the medical and public health community of the existence of individuals that continuously test positive is warranted to avoid unnecessary isolation periods and subsequent impact on an individual’s mental health, family, professional activities, social and day-to-day life.

Detailed descriptions of the methods are available as Supplementary material.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

Ethical approval was obtained from the University of Ottawa REB (H-11-20-6172) for the studies involving humans. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

YG and M-AL wrote the manuscript. YG performed the serology experiments and created the figures. AX performed the qPCR experiments. AD provided participant coordination and sample collection. CC offered clinical perspective and insights. All authors contributed to the article and approved the submitted version.

Funding

M-AL holds a Faculty of Medicine Chair of Excellence in Pandemic Viruses and Preparedness Research. YG holds a Canadian Institute of Health Research (CIHR) Frederick Banting and Charles Best graduate scholarship (CGS-Doctoral, 476885). This study was supported in part by a COVID-19 Rapid Response grant by the Canadian Institute of Health Research (CIHR, VR2-172722) and by a grant supplement by the Canadian Immunity Task Force (CITF) to M-AL. The funding sources played no role in the study design, data collection, analysis, writing of the manuscript, and the decision to publish.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fpubh.2023.1240308/full#supplementary-material

References

1. Honein, MA, Christie, A, Rose, DA, Brooks, JT, Meaney-Delman, D, Cohn, A, et al. Summary of guidance for public health strategies to address high levels of community transmission of SARS-CoV-2 and related deaths, December 2020. MMWR Morb Mortal Wkly Rep. (2020) 69:1860–7. doi: 10.15585/mmwr.mm6949e2

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Chen, YT, Yen, YF, Yu, SH, and Su, EC. An examination on the transmission of COVID-19 and the effect of response strategies: a comparative analysis. Int J Environ Res Public Health. (2020) 17:5687. doi: 10.3390/ijerph17165687

CrossRef Full Text | Google Scholar

3. Blanchard, AC, Desforges, M, Labbé, A-C, Nguyen, CT, Petit, Y, Besner, D, et al. Evaluation of real-life use of point-of-care rapid antigen testing for SARS-CoV-2 in schools (EPOCRATES): a cohort study. CMAJ Open. (2022) 10:E1027–33. doi: 10.9778/cmajo.20210327

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Schwartz, KL, Bogoch, II, MacInTosh, D, Barrow, J, Sindrey, D, Jha, P, et al. SARS-CoV-2 rapid antigen screening of asymptomatic employees: a pilot project. Can J Public Health. (2022) 113:898–903. doi: 10.17269/s41997-022-00691-z

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Government of Canada. Interim guidance on the use of rapid antigen detection tests for the identification of SARS-CoV-2 infection (2022) [Available at: https://www.canada.ca/en/public-health/services/diseases/2019-novel-coronavirus-infection/guidance-documents/use-rapid-antigen-detection-tests.html.

Google Scholar

6. Eyre, DW, Futschik, M, Tunkel, S, Wei, J, Cole-Hamilton, J, Saquib, R, et al. Performance of antigen lateral flow devices in the UK during the alpha, delta, and omicron waves of the SARS-CoV-2 pandemic: a diagnostic and observational study. Lancet Infect Dis. (2023) 23:922–32. doi: 10.1016/S1473-3099(23)00129-9

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Chu, VT, Schwartz, NG, Donnelly, MAP, Chuey, MR, Soto, R, Yousaf, AR, et al. Comparison of home antigen testing with RT-PCR and viral culture during the course of SARS-CoV-2 infection. JAMA Intern Med. (2022) 182:701–9. doi: 10.1001/jamainternmed.2022.1827

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Wölfl-Duchek, M, Bergmann, F, Jorda, A, Weber, M, Müller, M, Seitz, T, et al. Sensitivity and specificity of SARS-CoV-2 rapid antigen detection tests using Oral, anterior nasal, and nasopharyngeal swabs: a diagnostic accuracy study. Microbiology Spectrum. (2022) 10:e02029–1. doi: 10.1128/spectrum.02029-21

CrossRef Full Text | Google Scholar

9. Gans, JS, Goldfarb, A, Agrawal, AK, Sennik, S, Stein, J, and Rosella, L. False-positive results in rapid antigen tests for SARS-CoV-2. JAMA. (2022) 327:485–6. doi: 10.1001/jama.2021.24355

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Navero-Castillejos, J, Casals-Pascual, C, Narváez, S, Cuesta, G, Hurtado, JC, Fernandez, M, et al. Diagnostic performance of six rapid antigen tests for SARS-CoV-2. Microbiology Spectrum. (2022) 10:e02351–21. doi: 10.1128/spectrum.02351-21

CrossRef Full Text | Google Scholar

11. Pickering, S, Batra, R, Merrick, B, Snell, LB, Nebbia, G, Douthwaite, S, et al. Comparative performance of SARS-CoV-2 lateral flow antigen tests and association with detection of infectious virus in clinical specimens: a single-Centre laboratory evaluation study. The Lancet Microbe. (2021) 2:e461–71. doi: 10.1016/S2666-5247(21)00143-9

PubMed Abstract | CrossRef Full Text | Google Scholar

12. BTNX INC. (n.d.)COVID-19 Antigen Rapid Test Device Available at: https://www.btnx.com/Product?id=2010.

Google Scholar

13. Jüni, P, Baert, S, Corbeil, A, Johnstone, J, Patel, SN, and Bobos, P. Use of Rapid Antigen Tests during the omicron wave Science Briefs of the Ontario COVID-19 Science Advisory Table. (2022).

Google Scholar

14. Ontario Government. (n.d.) Rapid testing for at-home use Available at: https://www.ontario.ca/page/rapid-testing-home-use#section-2.

Google Scholar

15. Jonczyk, R, Stanislawski, N, Seiler, LK, Blume, H, Heiden, S, Lucas, H, et al. Combined prospective seroconversion and PCR data of selected cohorts indicate a high rate of subclinical SARS-CoV-2 infections—an open observational study in Lower Saxony. Germany Microbiology Spectrum. (2022) 10:e01512–21. doi: 10.1128/spectrum.01512-21

CrossRef Full Text | Google Scholar

16. Chen, Y, Tong, X, Wang, J, Huang, W, Yin, S, Huang, R, et al. High SARS-CoV-2 antibody prevalence among healthcare workers exposed to COVID-19 patients. J Infect. (2020) 81:420–6. doi: 10.1016/j.jinf.2020.05.067

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Suzuki, T, Aizawa, K, Shibuya, K, Yamanaka, S, Anzai, Y, Kurokawa, K, et al. Prevalence of asymptomatic SARS-CoV-2 infection in Japan. JAMA Netw Open. (2022) 5:e2247704. doi: 10.1001/jamanetworkopen.2022.47704

CrossRef Full Text | Google Scholar

18. Avanzato, VA, Matson, MJ, Seifert, SN, Pryce, R, Williamson, BN, Anzick, SL, et al. Case study: prolonged infectious SARS-CoV-2 shedding from an asymptomatic immunocompromised individual with cancer. Cells. (2020) 183:1901–12.e9. doi: 10.1016/j.cell.2020.10.049

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Gaebler, C, Wang, Z, Lorenzi, JCC, Muecksch, F, Finkin, S, Tokuyama, M, et al. Evolution of antibody immunity to SARS-CoV-2. Nature. (2021) 591:639–44. doi: 10.1038/s41586-021-03207-w

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Cheung, CCL, Goh, D, Lim, X, Tien, TZ, Lim, JCT, Lee, JN, et al. Residual SARS-CoV-2 viral antigens detected in GI and hepatic tissues from five recovered patients with COVID-19. Gut. (2022) 71:226–9. doi: 10.1136/gutjnl-2021-324280

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: SARS-CoV-2, rapid antigen test (RAT), false positive, case report, COVID-19

Citation: Galipeau Y, Xavier A, Dyks A, Cooper C and Langlois M-A (2023) Continuous false positive results by SARS-CoV-2 rapid antigen testing: a case report. Front. Public Health. 11:1240308. doi: 10.3389/fpubh.2023.1240308

Received: 23 June 2023; Accepted: 29 September 2023;
Published: 03 November 2023.

Edited by:

Svetlana Khaiboullina, University of Nevada, Reno, United States

Reviewed by:

José Eduardo Levi, University of São Paulo, Brazil
Io Cheong, Shanghai Jiao Tong University, China

Copyright © 2023 Galipeau, Xavier, Dyks, Cooper and Langlois. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Marc-André Langlois, langlois@uottawa.ca

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.