AUTHOR=Garcia-Tabar Ibai , Eclache Jean P. , Aramendi José F. , Gorostiaga Esteban M. TITLE=Gas analyzer's drift leads to systematic error in maximal oxygen uptake and maximal respiratory exchange ratio determination JOURNAL=Frontiers in Physiology VOLUME=6 YEAR=2015 URL=https://www.frontiersin.org/journals/physiology/articles/10.3389/fphys.2015.00308 DOI=10.3389/fphys.2015.00308 ISSN=1664-042X ABSTRACT=
The aim was to examine the drift in the measurements of fractional concentration of oxygen (FO2) and carbon dioxide (FCO2) of a Nafion-using metabolic cart during incremental maximal exercise in 18 young and 12 elderly males, and to propose a way in which the drift can be corrected. The drift was verified by comparing the pre-test calibration values with the immediate post-test verification values of the calibration gases. The system demonstrated an average downscale drift (