ORIGINAL RESEARCH article
Front. Oncol.
Sec. Radiation Oncology
Volume 15 - 2025 | doi: 10.3389/fonc.2025.1509449
The final, formatted version of the article will be published soon.
You have multiple emails registered with Frontiers:
Please enter your email address:
If you already have an account, please login
You don't have a Frontiers account ? You can register here
Purpose: This study introduces a deep learning (DL) model that leverages doses calculated from both a treatment planning system (TPS) and independent dose verification software using Monte Carlo (MC) simulations, aiming to predict the gamma passing rate (GPR) in VMAT patient-specific QA more accurately.: We utilized data from 710 clinical VMAT plans measured with an ArcCHECK phantom. These plans were recalculated on an ArcCHECK phantom image using Pinnacle TPS and MC algorithms, and the planar dose distributions corresponding to the detector element surfaces were utilized as input for the DL model. A convolutional neural network (CNN) comprising four layers was employed for model training. The model's performance was evaluated through multiple predictive error metrics and receiver operator characteristic (ROC) curves for various gamma criteria.The mean absolute errors (MAE) between measured GPR and predicted GPR are 1.1%, 1.9%, 1.7%, and 2.6% for the 3%/3mm, 3%/2mm, 2%/3mm, and 2%/2mm gamma criteria, respectively. The correlation coefficients between predicted GPR and measured GPR are 0.69, 0.72, 0.68, and 0.71 for each gamma criterion. The AUC (Area Under the Curve) values based on ROC curve for the four gamma criteria are 0.90, 0.92, 0.93, and 0.89, indicating high classification performance.This DL-based approach showcases significant potential in enhancing the efficiency and accuracy of VMAT patient-specific QA. This approach promises to be a useful tool for reducing the workload of patientspecific quality assurance.
Keywords: Gamma passing rate, deep learning, monte carlo, Quality Assurance, ArcCHECK
Received: 11 Oct 2024; Accepted: 04 Mar 2025.
Copyright: © 2025 Miao, XU, Men and Dai. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
* Correspondence:
Jianrong Dai, Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China
Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.