Plan Quality Assessment of Fractionated Stereotactic Radiotherapy Treatment Plans in Patients With Brain Metastases

  • 1Department of Oncoradiology, Faculty of Medicine, University of Debrecen, Debrecen, Hungary
  • 2Doctoral School of Health Sciences, University of Pécs Faculty of Health Sciences, Pécs, Hungary

Background and Purpose: The treatment options available in the management of brain metastases includes fractionated stereotactic radiotherapy (FSRT) and stereotactic radiosurgery (SRS) treatments. FSRT treatments have proved to be useful mainly in the treatment of larger volumes. This study aims to evaluate the FSRT treatment technique used in our department based on various plan quality indices.

Methods and Materials: 24 treatment plans of 23 patients were analyzed. Volumetric modulated arc therapy (VMAT) plans were generated in line with the department protocol. The following parameters were extracted: Radiation Therapy Oncology Group conformity index (RTOG CI), Paddick conformity index (Paddick CI), gradient index (GI), quality index (Q), homogeneity index (HI), and V24.4 volume as a parallel index of V12 used at SRS plan evaluation.

Results: Plan conformity was acceptable, RTOG CI mean was 0.942; Paddick CI mean was 0.824. The mean GI value was 6.146. The mean of HI and Q indices were 1.263 and 0.94, respectively. V24.4 mean was 33.434 cm3. All plans achieved clinically acceptable organs-at-risk (OAR) constraints. PTV volumes were clustered into either 10 cm3 or 15 cm3 bins depending on the plan quality metric we used. The mean values show a balanced distribution of plan indices along the various PTV bins.

Discussion: Our results based on the derived indices show that our FSRT approach can achieve clinically acceptable treatment plans. Furthermore, the clustering of PTV volumes show that these plan quality metrics remain acceptable for a wide spectrum of PTV volumes.

Introduction

Brain metastases (BM) are considered to be a serious problem in the course of oncological diseases as 20-40% of patients affected by cancer will develop BMs (1). Although any primary site can metastasize to the brain, the most common sites are the lung (especially small cell lung cancer), breast, colon (colorectal cancer), skin, and kidney (2). Until recently, the major treatment option for these patients was whole brain radiotherapy (WBRT), but today the treatment for BMs depends on the number and size of lesions, on the patients performance state and the oncological status of the primary disease (3, 4). In cases with a limited number of metastases, stereotactic radiosurgery (SRS) was available, but due to the ablative nature of the treatment, the size of individual metastases was usually limited to 3.5 cm, and for multiple mets it was limited to 2.5 cm (5). SRS is a very high precision treatment given in one fraction in multiple static or moving beams to create a steep dose fall-off to spare normal brain tissue or adjacent organs at risk (OAR). High single doses near critical structures or large tumours are associated with significant risk of toxicity, and cognitive impairment has been associated with the brain volume receiving at least 12 Gy (6). There has been an attempt to utilize hypo fractionated stereotactic radiotherapy (FSRT) to achieve similar local control with acceptable toxicity rates. FSRT also uses high precision radiation therapy to treat BMs with non-invasive head fixation devices. The aim of FSRT is to minimize the normal tissue toxicity while preserving local control. It lacks the precision of SRS with the head frame, but can decrease the margins; therefore, in addition to decreasing normal tissue dose, it also makes use of the advantage of fractionation. Recent data suggest FSRT could be safer near critical structures and more effective than single fraction SRS (7, 8). In the present study we propose an FSRT treatment method implemented at our clinic. Plan quality indices are compared to values found in literature.

Materials and Methods

Patients

Between May 2019 and May 2020, 24 patients were treated using FSRT treatment at our department. Patients who had single or multiple intact BMs were selected for this study. Patients who underwent surgical resection, but also had intact lesions treated with FSRT were also included. All patients had a multidisciplinary board meeting indicating FSRT prior treatment. The number of lesions was between 1 and 7, and all of them were treated with a mono isocentre technique within the same treatment plan.

Localization

Patients were immobilized in a supine position with individual open face thermoplastic head masks (QFix, Avondale, PA, USA),lying on a baseplate and a mouldable head cushion (QFix, Avondale, PA, USA). These masks are 2.4 mm thick and are strengthened with Kevlar to increase cranial support. Open face masks leave the eyes and nose free, and are proven to reduce anxiety, and increase patient comfort (9). The localization CT was performed on a Philips Brilliance Big Bore CT (Philips, The Netherlands) with a specific protocol with 2 mm slice thickness. Prior to CT, preferably within 2 weeks, all patients had an MRI scan of the brain with gadolinium contrast.

Treatment Planning

All patients were contoured and planned on Pinnacle (Philips, The Netherlands) treatment planning system version 9.8. Before planning, a gadolinium contrast agent enhanced MRI was performed, and T1 post-contrast enhanced, and T2-weighted series were co-registered to the localization CT series with a rigid transformation. The GTV was contoured on the planning CT series based on MRI information; the department protocol defines GTV as the visible, contrast enhanced tumour volume on the T1 series. GTV was assumed to be equal to the clinical target volume (CTV). The GTV to PTV margin was an isotropic 3 mm. The OARs considered during planning were brainstem, optic chiasm, inner ears, lenses, optic nerves, eyes, and brain. Dose tolerances for the OARs are listed in Table 1 (1014). Single isocentre volumetric modulated arc therapy (VMAT) treatment plans were generated in the treatment planning system using 5 half arcs. The beam arrangement involved 2 coplanar and 3 non-coplanar half arcs (180 degrees); these required angular table movements (0°, ± 45°, +90°). In cases where the PTV or PTVs were localized laterally, one coplanar arc could be omitted to decrease overall brain dose.The isocentre was placed at the centre of mass of all the PTVs. A PRV margin of 3 mm was added to all serial organs at risk. The treatment energy was 6MV, and planners were allowed to use Flattening Filter Free (FFF) mode. The use of FFF mode in brain SRS has been described by several authors (15, 16). During treatment planning a PRV was generated with a 3 mm margin around each OAR (17).

TABLE 1
www.frontiersin.org

Table 1 OAR tolerances for FSRT planning.

Prescription

The prescribed dose was 30 Gy, either in 5 or 6 fractions. Dose was prescribed to the isocentre (single lesion) or to PTV mean (multiple lesions). Initially the prescription was 80% of the isocentre dose or the PTV mean dose. Planners could shift the prescription between 70% and 95% to conform to the 95% isodose line to the PTV.

Image Guidance and Treatment

Patients were treated and image guided on an Elekta Versa HD linear accelerator with HexaPod robotic treatment couch. A high resolution3D CBCT was acquired and registered to the localization CT series with an automatic bone match of a region, involving the entire skull. Before treatment, translational and rotational errors were corrected; thus, the residual error was significantly reduced. After the registration procedure a second 3D CBCT was performed to make sure that the position of the patient has not changed during the registration, and the patient is in the correct position for the treatment. The delivery started with the two coplanar beams, and the following arcs were scheduled in a manner that the gantry stop position of a half arc was the stat position of the following beam. This reduced the treatment time, and decreased the risk of positional deviations (18).

HexaPOD

The Medical Intelligence HexaPODtm couch top (Schwabmünchen, Germany) is a specifically designed robotic treatment couch top rigidly mounted on the standard treatment table. It has 6 degrees of freedom, and can correct both translational and rotational errors. The range of linear movement of the table is +-30, +-30; +-40 mm in longitudinal, lateral, and vertical directions, respectively. Also, it has an angular range of +-3 degrees along all three rotational axes, which are denoted as pitch, roll and yaw. It has sub-millimetre and sub-degree-level accuracy, and is capable of efficiently moving the target (19, 20). Rotational errors of single isocentre plans treating multiple lesions were shown to increase the risk of compromised coverage above 0.5° (21).

Analysis

The following indices were extracted and analyzed per patient across all patient plans: RTOG Conformity index (RTOG CI), Paddick Conformity Index (Paddick CI), Gradient Index (GI), Homogeneity Index (HI), and Quality Index (Q). The RTOG CI was defined as the ratio of the reference isodose volume (PIV) to the target volume (TV) (Eq 1.).

CIRTOG=PIVTV(Eq. 1.)

The Paddick CI is calculated by the ratio of the square of the volume of the target enclosed by reference isodose volume (TVPIV) to the multiplication of the target volume (TV) with the reference isodose volume (VRI) (Eq 2.).

CIPaddick=TVPIV2(TVVRI)(Eq. 2.)

The Gradient Index defined as the ratio of the 50% isodose (V50) volume to the prescription isodose volume (PIV) (Eq 3.).

GI=V50PIV(Eq. 3.)

By definition, Homogeneity Index is the ratio of the maximum dose in the target volume (Imax) to the reference isodose volume (PIV) (Eq 4.)

HI=ImaxPIV(Eq. 4.)

The Q index is the ratio of the minimum dose given to the target volume (Imin) to the reference isodose volume (PIV) (Eq 5.).

Q=IminPIV(Eq. 5.)

Furthermore, the V24.4Gy parameter was extracted, V24.4 Gy is the volume of the brain which receives at least 24.4 Gy. V24.4 Gy was used to parallel the V12 index used in SRS techniques based on Eq 6.

BED=nd(1+dαβ)(Eq. 6.)

Here α/β = 2, n refers to the number of fractions, and d represents the fractional dose.

The PTV volumes were binned into 10cm3 bins with the corresponding indices.

Results

Table 2 summarizes patient characteristics. Age was in a range from 42 to 80, with a median age of 63.9 years. Patients were in a good overall condition:15 of them (63%) was ECOG 0, and only 2 patients (8%) were ECOG 2. The RPA classification of the disease ranged from 1 to 3. Altogether 4 patients (17%) received some form of intracranial irradiation before the SRT treatment, 1 patient had WBRT, 2 patients had SRS with a gamma knife, and 1 patient had received SRT previously at our institute for different lesions. These previous treatments were taken into account during the preparation and planning of the present treatment plans in terms of critical structure doses. 4 patients underwent surgical resection prior treatment; 7 patients had solitary lesions, and 17 patients had oligometastatic disease. The total number of lesions was 65. Most frequent primary sites were NSCLC (non-small cell lung cancer) (46%), breast (17%) and melanoma (13%) The mean number of treated lesions per patient was 2.7, with a standard deviation of 1.7.

TABLE 2
www.frontiersin.org

Table 2 Patient characteristics.

Table 3 contains the laterality and localization specific lesion statistics. Out of 65 lesions, 32 were left sided and 32 were right sided, and laterality was not applicable for one lesion. In terms of localization, most of the metastases was supratentorial (45), 25% (18) was infratentorial, and 3% (2) were in the base of skull.

TABLE 3
www.frontiersin.org

Table 3 Lesion charactheristics.

The conformity of the plans was calculated based on several indices, such as RTOG CI, Paddick CI, GI, HI, and Q. The RTOG CI, Paddick CI, and GI were extracted by patient (Table 4).

TABLE 4
www.frontiersin.org

Table 4 RTOG conformity, paddick conformity, and gradient indices.

Q and HI were calculated on each lesion (Table 5).

TABLE 5
www.frontiersin.org

Table 5 Quality and homogeneity indices.

The RTOG CI mean value was 0.942 with a standard deviation (SD) of ±0.153. The Paddick CI mean value was 0.824 with an SD of ±0.090. GI mean was 6.146±3.085 (SD) with a median of 5.483. The mean value of Q index was 0.940 with a 0.118 SD. HI mean was 1.263, the SD was 0.103.

Figure 1 divides the RTOG CI and Paddick CI values per overall PTV volume for each patient. Table 6 shows the conformity and gradient indices for each PTV cluster.

FIGURE 1
www.frontiersin.org

Figure 1 RTOG conformity and Paddick Conformity indices clustered into 15 cm3 bins.

TABLE 6
www.frontiersin.org

Table 6 RTOG, paddick conformity, and gradient indices for PTV clusters.

PTV volumes ranged from 1.23 cm3 to 61.63 cm3; the mean volume was 14.17 cm3 with a standard deviation of 16.5 cm3, while the median of the volumes was 4.89. Figure 2 shows the distribution of PTV volumes across all metastases. Figure 3 shows the GI values clustered by the overall PTV volume per patient.

FIGURE 2
www.frontiersin.org

Figure 2 Distribution of PTV volumes across bins. Each PTV had been assigned to a bin with according to the volume.

FIGURE 3
www.frontiersin.org

Figure 3 Gradient index per patients clustered in 15cm3 bins. The * symbol is the most extreme value compared to the mean for that dataset.

Figure 4 displays the HI and Q clustered by lesion size. Table 7 presents the Q and HI values for the individual PTV volume bins.

FIGURE 4
www.frontiersin.org

Figure 4 Homogeneity and quality indices by lesion clustered into 10cm3 bins. The * symbol is the most extreme value compared to the mean for that dataset.

TABLE 7
www.frontiersin.org

Table 7 Quality and homogeneity indices for individual PTV volume bins.

Our results show that the applied treatment planning technique was able to generate clinically acceptable plans in terms of these metrics across various size of metastases.

V24.4 was calculated after the exclusion of the given PTV volume from the 24.4 Gy area; for each lesion a 24.4 Gy volume was defined by dividing the 24.4 Gy area of the patient by the number of treated metastases. Table 8 contains the values for V24.4Gy.

TABLE 8
www.frontiersin.org

Table 8 V24.4Gy descriptive statictics.

Discussion

Stereotactic radiosurgery (SRS) is the gold standard treatment of brain metastases (BM); however, there is a limitation in treatable tumour size (<4 cm) and the risk of radio necrosis increasing with doses over 12Gy (11). Some comparative analyses have been published on SRS and FSRT (2224), suggesting FSRT may offer an alternative solution due to its comparable precision during treatments, as well as its ability to treat larger lesions. Regarding tumour progression and risk of radio necrosis, Putz et al. published better results for FSRT that single fraction SRS (25). The overall findings of our study have demonstrated the plausibility of using our FSRT treatment technique in the treatment of multiple brain metastases. We proposed a linac-based FSRT treatment method with non-invasive patient positioning and treatment planning, which can achieve clinically acceptable conformality based on various indices. Linac-based techniques proved to have improved conformity over gamma knife (GK) treatments in large lesions, and even compete GK in terms of V12 volume (26). Our aim was to derive conformity indices used in stereotactic treatments such as RTOG conformity index (RTOG CI), Paddick conformity index (Paddick CI), Quality index (Q), Homogeneity index (HI) (27),and biologically equate the V12 volume used in SRS technique based on Equation 6. Milano et al. paralleled the V12 index used in SRS techniques with V24.4 for 5 fraction FSRT treatments (11). Hsu et al. compared different treatment modalities based on similar indices based on tumour size and tumor distance from brainstem. Our mean PTV volume was 14.17cm3, the corresponding values for FFF-VMAT plans were 0.73 and 0.72 for Paddick CI, and 1.19 and 1.17 for HI. Our mean Paddick CI and HI values were 0.824 and 1.263, respectively. Ruggieri et al. evaluated treatment plans generated with HyperArc™ (HA) (Varian Medical System, Palo Alto, CA) and MultipleBrainMets™ (MBM) (Brainlab AG, Munchen, Germany) in terms of plan conformity. The mean number of treated lesions was 5. The mean Paddick CI values for HA and MBM plan were 0.94 and 0.75, respectively (28, 29). For RapidArc (RA) plans Ruggieri reported 0.87 and 6.08 for Paddick CI and GI values, respectively. Our gradient index (GI) mean value was 6.146. Individual PTV volumes were clustered into 10cm3 bins, and the derived mean of Q and HI indices shows good consistency among the various sized bins. In terms of RTOG CI, Paddick CI, and GI the overall PTV volumes were binned into 15cm3 bins. The results show consistent values of these indicators as well, regardless of PTV volume. In terms of V12 Ruggieri et al. published 23.7cm3 and 37.3cm3 for HA and MBM plans, and 42,2cm3 for RA plans, respectively. In our cohort the mean value for V24.4Gy was 33.434 cm3.

The main limitation of this study is the limited number of patients and the lack of comparability to other treatment techniques such as GK.

In summary, our approach to treat multiple brain metastases with a single isocenter VMAT technique using non-invasive mask fixation proved to be clinically plausible and based on the metrics the results are comparable to the results published in literature. Furthermore, our approach can reach clinically acceptable values for plan quality indices in a wide spectrum of PTV volumes.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Author Contributions

MS: data collection, statistical analyses, manuscript preparation. JP and EC: tables, data collection, manuscript preparation. ÁK: manuscript preparation. All authors contributed to the article and approved the submitted version.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Patchell RA. The Management of Brain Metastases. Cancer Treat Rev (2003) 29(6):533–40. doi: 10.1016/S0305-7372(03)00105-1

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Mehta MP, Tsao MN, Whelan TJ, Morris DE, Hayman JA, Flickinger JC, et al. ASTRO. Radiosurgery in BM. Week 25), and the Cochrane Library (2004 Issue 2) Databases Were Searched Using OVID. In Addition, the Physician Data Query Clinical Trials Database, the Proceedings of the. Eur Soc Ther Radiol Oncol (ESTRO (2005) 63(1):37–46. doi: 10.1016/j.ijrobp.2005.05.023

CrossRef Full Text | Google Scholar

3. Shinde A, Akhavan D, Sedrak M, Glaser S, Amini A. Shifting Paradigms: Whole Brain Radiation Therapy Versus Stereotactic Radiosurgery for Brain Metastases. CNS Oncol (2019) 8(1):CNS27. doi: 10.2217/cns-2018-0016

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Balducci M, Autorino R, Chiesa S, Mattiucci G, Pompucci A, Azario L, et al. Radiosurgery or Fractionated Stereotactic Radiotherapy Plus Whole-Brain Radioherapy in Brain Oligometastases a Long-Term Analysis. Anticancer Res (2015) 35(5):3055–60.

PubMed Abstract | Google Scholar

5. Kocher M, Wittig A, Piroth MD, Treuer H, Seegenschmiedt H, Ruge M, et al. Stereotactic Radiosurgery for Treatment of Brain Metastases: A Report of the DEGRO Working Group on Stereotactic Radiotherapy. Strahlenther und Onkologie (2014) 190(6):521–32. doi: 10.1007/s00066-014-0648-7

CrossRef Full Text | Google Scholar

6. Jacob J, Durand T, Feuvret L, Mazeron JJ, Delattre JY, Hoang-Xuan K, et al. Cognitive Impairment and Morphological Changes After Radiation Therapy in Brain Tumors: A Review. Radiother Oncol (2018) 128(2):221–8. doi: 10.1016/j.radonc.2018.05.027

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Minniti G, Scaringi C, Paolini S, Lanzetta G, Romano A, Cicone F, et al. Single-Fraction Versus Multifraction (3 × 9 Gy) Stereotactic Radiosurgery for Large (>2 Cm) Brain Metastases: A Comparative Analysis of Local Control and Risk of Radiation-Induced Brain Necrosis. Int J Radiat Oncol Biol Phys (2016) 95(4):1142–8. doi: 10.1016/j.ijrobp.2016.03.013

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Kirkpatrick JP, Soltys SG, Lo SS, Beal K, Shrieve DC, Brown PD. The Radiosurgery Fractionation Quandary: Single Fraction or Hypofractionation? Neuro-Oncology (2017) 19(2):ii38–49. doi: 10.1093/neuonc/now301

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Wiant D, Squire S, Liu H, Maurer J, Lane Hayes T, Sintay B. A Prospective Evaluation of Open Face Masks for Head and Neck Radiation Therapy. Pract Radiat Oncol (2016) 6(6):e259–67. doi: 10.1016/j.prro.2016.02.003

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Milano MT, Usuki KY, Walter KA, Clark D, Schell MC. Stereotactic Radiosurgery and Hypofractionated Stereotactic Radiotherapy: Normal Tissue Dose Constraints of the Central Nervous System. Cancer Treat Rev (2011) 37(7):567–78. doi: 10.1016/j.ctrv.2011.04.004

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Milano MT, Grimm J, Niemierko A, Soltys SG, Moiseenko V, Redmond KJ, et al. Single- and Multifraction Stereotactic Radiosurgery Dose/Volume Tolerances of the Brain. Int J Radiat Oncol Biol Phys (2020) 110(1):68–86. doi: 10.1016/j.ijrobp.2020.08.013

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Hanna GG, Murray L, Patel R, Jain S, Aitken KL, Franks KN, et al. UK Consensus on Normal Tissue Dose Constraints for Stereotactic Radiotherapy. Clin Oncol (2018) 30(1):5–14. doi: 10.1016/j.clon.2017.09.007

CrossRef Full Text | Google Scholar

13. Marcrom SR, McDonald AM, Thompson JW, Popple RA, Riley KO, Markert JM, et al. Fractionated Stereotactic Radiation Therapy for Intact Brain Metastases. Adv Radiat Oncol (2017) 2(4):564–71. doi: 10.1016/j.adro.2017.07.006

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic Body Radiation Therapy: The Report of AAPM Task Group 101. Med Phys (2010) 37(8):4078–101. doi: 10.1118/1.3438081

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Ruggieri R, Naccarato S, Mazzola R, Ricchetti F, Corradini S, Fiorentino A, et al. Linac-Based VMAT Radiosurgery for Multiple Brain Lesions: Comparison Between a Conventional Multi-Isocenter Approach and a New Dedicated Mono-Isocenter Technique. Radiat Oncol (2018) 13(1):1–9. doi: 10.1186/s13014-018-0985-2

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Clark GM, Popple RA, Young PE, Fiveash JB. Feasibility of Single-Isocenter Volumetric Modulated Arc Radiosurgery for Treatment of Multiple Brain Metastases. Int J Radiat Oncol Biol Phys (2010) 76(1):296–302. doi: 10.1016/j.ijrobp.2009.05.029

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Stroom JC, Heijmen BJM. Limitations of the Planning Organ at Risk Volume (PRV) Concept. Int J Radiat Oncol Biol Phys (2006) 66(1):279–86. doi: 10.1016/j.ijrobp.2006.05.009

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Tarnavski N, Engelholm SA, Af Rosenschold PM. Fast Intra-Fractional Image-Guidance With 6D Positioning Correction Reduces Delivery Uncertainty for Stereotactic Radiosurgery and Radiotherapy. J Radiosurg SBRT (2016) 4(1):15–20.

PubMed Abstract | Google Scholar

19. Meyer J, Guckenberger M, Wilbert J, Baier K. On the Use of a Hexapod Table to Improve Tumour Targeting in Radiation Therapy. Bioinspiration Rob Walking Climbing Robots (September) (2007). doi: 10.5772/5519

CrossRef Full Text | Google Scholar

20. Chung H, Jin H, Suh TS, Palta J, Kim S. Characterization of a Commercial Add-on Couch, hexaPOD™ 6D Robotic Treatment couchTOP. IFMBE Proc (2007) 14(1):1945–7. doi: 10.1007/978-3-540-36841-0_485

CrossRef Full Text | Google Scholar

21. Roper J, Chanyavanich V, Betzel G, Switchenko J, Dhabaan A. TH-A-9a-11: Single-Isocenter Multiple-Target SRS: Risk of Compromised Coverage. Med Phys (2014) 41(6Part32):545–5. doi: 10.1118/1.4889581

CrossRef Full Text | Google Scholar

22. Soffietti R, Abacioglu U, Baumert B, Combs SE, Kinhult S, Kros JM, et al. Diagnosis and Treatment of Brain Metastases From Solid Tumors: Guidelines From the European Association of Neuro-Oncology (EANO). Neuro-Oncology (2017) 19(2):162–74. doi: 10.1093/neuonc/now241

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Kim YJ, Cho KH, Kim JY, Lim YK, Min HS, Lee SH, et al. Single-Dose Versus Fractionated Stereotactic Radiotherapy for Brain Metastases. Int J Radiat Oncol Biol Phys (2011) 81(2):483–9. doi: 10.1016/j.ijrobp.2010.05.033

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Fokas E, Henzel M, Surber G, Kleinert G, Hamm K, Engenhart-Cabillic R. Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy: Comparison of Efficacy and Toxicity in 260 Patients With Brain Metastases. J Neuro-Oncol (2012) 109(1):91–8. doi: 10.1007/s11060-012-0868-6

CrossRef Full Text | Google Scholar

25. Putz F, Weissmann T, Oft D, Schmidt MA, Roesch J, Siavooshhaghighi H, et al. FSRT vs. SRS in Brain Metastases—Differences in Local Control and Radiation Necrosis—A Volumetric Study. Front Oncol (2020) 10:559193. doi: 10.3389/fonc.2020.559193

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Vergalasova I, Liu H, Alonso-Basanta M, Dong L, Li J, Nie K, et al. Multi-Institutional Dosimetric Evaluation of Modern Day Stereotactic Radiosurgery (SRS) Treatment Options for Multiple Brain Metastases. Front Oncol (2019) 9:483. doi: 10.3389/fonc.2019.00483

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Stanley J, Breitman K, Dunscombe P, Spencer DP, Lau H. Evaluation of Stereotactic Radiosurgery Conformity Indices for 170 Target Volumes in Patients With Brain Metastases. J Appl Clin Med Phys (2011) 12(2):245–53. doi: 10.1120/jacmp.v12i2.3449

CrossRef Full Text | Google Scholar

28. Hsu SM, Lai YC, Jeng CC, Tseng CY. Dosimetric Comparison of Different Treatment Modalities for Stereotactic Radiotherapy. Radiat Oncol (2017) 12(1):1–11. doi: 10.1186/s13014-017-0890-0

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Ruggieri R, Naccarato S, Mazzola R, Ricchetti F, Corradini S, Fiorentino A, et al. Linac-Based Radiosurgery for Multiple Brain Metastases: Comparison Between Two Mono-Isocenter Techniques With Multiple Non-Coplanar Arcs. Radiother Oncol (2019) 132:70–8. doi: 10.1016/j.radonc.2018.11.014

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: FSRT = fractionated stereotactic radiotherapy, brain metastases, conformity index (CI), linac based, VMAT = volumetric modulated arc therapy

Citation: Simon M, Papp J, Csiki E and Kovács Á (2022) Plan Quality Assessment of Fractionated Stereotactic Radiotherapy Treatment Plans in Patients With Brain Metastases. Front. Oncol. 12:846609. doi: 10.3389/fonc.2022.846609

Received: 31 December 2021; Accepted: 11 February 2022;
Published: 08 March 2022.

Edited by:

James Chow, University of Toronto, Canada

Reviewed by:

John E. Mignano, Tufts University School of Medicine, United States
Enrico Clarke, University Hospital Southampton NHS Foundation Trust, United Kingdom

Copyright © 2022 Simon, Papp, Csiki and Kovács. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Mihály Simon, simonm@med.unideb.hu

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.