AUTHOR=Kim Jihun , Han Min Cheol , Chang Jee Suk , Hong Chae-Seon , Kim Kyung Hwan , Byun Hwa Kyung , Park Ryeong Hwang , Koom Woong Sub , Park Seong Yong , Kim Jin Sung TITLE=Development of a Margin Determination Framework for Tumor-Tracking Radiation Therapy With Intraoperatively Implanted Fiducial Markers JOURNAL=Frontiers in Oncology VOLUME=11 YEAR=2021 URL=https://www.frontiersin.org/journals/oncology/articles/10.3389/fonc.2021.753246 DOI=10.3389/fonc.2021.753246 ISSN=2234-943X ABSTRACT=Purpose

To develop an internal target volume (ITV) margin determination framework (or decision-supporting framework) for treating multiple lung metastases using CyberKnife Synchrony with intraoperatively implanted fiducial markers (IIFMs). The feasibility of using non-ideally implanted fiducial markers (a limited number and/or far from a target) for tracking-based lung stereotactic ablative radiotherapy (SABR) was investigated.

Methods

In the developed margin determination framework, an optimal set of IIFMs was determined to minimize a tracking uncertainty-specific ITV (ITVtracking) margin (margin required to cover target-to-marker motion discrepancy), i.e., minimize the motion discrepancies between gross tumor volume (GTV) and the selected set of fiducial markers (FMs). The developed margin determination framework was evaluated in 17 patients with lung metastases. To automatically calculate the respiratory motions of the FMs, a template matching-based FM tracking algorithm was developed, and GTV motion was manually measured. Furthermore, during-treatment motions of the selected FMs were analyzed using log files and compared with those calculated using 4D CTs.

Results

For 41 of 42 lesions in 17 patients (97.6%), an optimal set of the IIFMs was successfully determined, requiring an ITVtracking margin less than 5 mm. The template matching-based FM tracking algorithm calculated the FM motions with a sub-millimeter accuracy compared with the manual measurements. The patient respiratory motions during treatment were, on average, significantly smaller than those measured at simulation for the patient cohort considered.

Conclusion

Use of the developed margin determination framework employing CyberKnife Synchrony with a limited number of IIFMs is feasible for lung SABR.