Skip to main content

BRIEF RESEARCH REPORT article

Front. Nucl. Med. , 28 February 2025

Sec. Physics and Data Analysis

Volume 5 - 2025 | https://doi.org/10.3389/fnume.2025.1556848

This article is part of the Research Topic Total Body Positron Emission Tomography: Science and Clinical applications View all articles

Validating image-derived input functions of dynamic 18F-FDG long axial field-of-view PET/CT studies

  • 1Department of Radiology and Nuclear Medicine, Amsterdam UMC Location Vrije Universiteit Amsterdam, Amsterdam, Netherlands
  • 2Imaging and Biomarkers, Cancer Center Amsterdam, Amsterdam, Netherlands
  • 3Department of Hematology, Amsterdam UMC Location Vrije Universiteit Amsterdam, Amsterdam, Netherlands
  • 4Department of Medical Oncology, Amsterdam UMC Location Vrije Universiteit Amsterdam, Amsterdam, Netherlands
  • 5Department of Pulmonary Medicine, Amsterdam UMC Location Vrije Universiteit Amsterdam, Amsterdam, Netherlands

Aim/background: Dynamic PET imaging requires an input function typically obtained through blood sampling. Image-derived input functions (IDIFs) of the ascending aorta (AA), aortic arch, descending aorta (DA), or left ventricle (LV) offer non-invasive alternatives, especially with long-axial field-of-view (LAFOV) PET/CT systems enabling whole-body dynamic 1⁸F-FDG imaging. This study aimed to validate uncorrected IDIFs derived from the AA, DA, aortic arch, and LV by comparing them to (late) venous whole-blood in patients undergoing LAFOV PET/CT.

Methods: Eleven oncology patients who underwent 70-min dynamic 18F-FDG PET/CT scans on a LAFOV PET/CT system after receiving an intravenous bolus injection of 3.0 MBq/kg were included. Seven venous blood samples were collected manually at approximately 5, 10, 15, 25, 35, 45, and 60 min post-injection (pi) and compared to IDIFs derived from the AA, aortic arch, DA, and LV. Bias between IDIFs and venous blood samples was assessed at each time point.

Results: IDIF accuracy relative to venous blood samples improved over time, with a median percentage bias <10% after 25 min pi. At 60 min pi, the aortic arch showed the smallest bias (median −1.1%, IQR 5.9%), followed by the AA (2.5%, IQR 7.0%), DA (5.1%, IQR 8.6%), and LV (7.4%, IQR 7.6%).

Conclusion: The high precision of aorta-derived IDIFs suggests that IDIFs are a reliable alternative to manual blood sampling for dynamic 18F-FDG PET imaging on a LAFOV PET/CT system. Using IDIFs reduces variability, simplifies protocols, minimizes radiation exposure, and enhances patient safety with a non-invasive approach.

Introduction

18F-fluoro-deoxy-glucose (18F-FDG) positron emission tomography-computed tomography (PET/CT) imaging is widely used in oncology (1, 2). Clinical 18F-FDG PET imaging often relies on static PET scans, where tracer uptake is primarily assessed by semi-quantitative metrics, such as the standardized uptake value (SUV) (3). However, examining 18F-FDG uptake through semi-quantitative metrics is debated since it has several limitations, such as not taking the variation of 18F-FDG availability in the blood into account (47).

Dynamic PET imaging overcomes many limitations by providing spatiotemporal information on tracer kinetics (811). This approach requires an accurate input function, typically obtained through arterial blood sampling (12). Yet, arterial blood sampling is invasive, technically challenging, and associated with potential adverse events, making it preferable to avoid if possible (1315). Arterialized venous blood sampling or venous blood sampling may be a potential alternative for 18F-FDG as there is an arteriovenous equilibrium after approximately 15–30 min (16, 17). However, these approaches still complicate scanning protocols and frequent blood sampling increases the radiation burden for technicians (12). Prior studies on short axial field-of-view (SAFOV) PET/CT systems have demonstrated that, for 18F-FDG, an image-derived input function (IDIF) from the aorta or left ventricle (LV) can replace manual blood sampling when assessing whole-blood activity (12, 18, 19). Additionally, Palard-Novello et al. (8) validated IDIFs corrected for plasma-to-whole-blood ratios with venous blood samples when performing dynamic 18F-FDG PET scans on a long axial field-of-view (LAFOV) PET/CT system. They suggested that manual blood sample calibration might be unnecessary, given the high precision of the IDIF.

Eliminating the need for manual blood sampling in IDIF calibration would simplify scanning protocols, reduce patient invasiveness, and decrease radiation exposure for personnel. However, the accuracy of whole-blood 1⁸F-FDG activity derived from manual blood samples compared to time-activity curves (TACs) from uncorrected IDIFs in LAFOV PET/CT systems remains unclear. Therefore, this study aimed to validate uncorrected IDIFs derived from the ascending aorta (AA), descending aorta (DA), aortic arch, and LV in patients undergoing LAFOV PET/CT. The accuracy of the IDIF was assessed by comparing it with (late) venous blood samples.

Method

Study population

Eleven consecutive patients were included from March 2023 to December 2023 at Amsterdam UMC, location VUmc. Each patient underwent a 70-min dynamic 18F-FDG PET scan on the Siemens Biograph Vision Quadra PET/CT system (Siemens Healthineers, Knoxville, TN, USA) (LAFOV PET/CT system). Before any study procedures written informed consent was obtained from all patients, and the study was approved by the Medical Ethics Review Committee of the Amsterdam UMC (NL80924.029.2).

18F-FDG PET/CT acquisition

PET/CT scans were performed and reconstructed according to the European Association of Nuclear Medicine Research Ltd (EARL2) guidelines (20). Each patient received 3.0 MBq/kg 18F-FDG at the start of the scan through an intravenous bolus injection. Patients fasted for 4-6 h before tracer administration and plasma glucose levels were <7.0 mmol/L (20). Patients were scanned in one bed position (106 cm), covering the skull vertex to mid-thigh. List-mode PET data started immediately after tracer administration and continued for 70 min. The dynamic PET images were binned in 20 data frames (1 × 15 s, 3 × 5 s, 3 × 10 s, 4 × 60 s, 2 × 150 s, 2 × 300 s, 5 × 600 s). Images were reconstructed using the PSF + TOF OSEM algorithm with 5 subsets, 4 iterations, a matrix size of 220 × 220, slice thickness of 3, voxel size of 3.3 × 3.3 × 2 mm3 and a Gaussian filter of 4 mm.

Seven venous blood samples were manually collected via the median cephalic vein or port-a-cath at approximately 5, 10, 15, 25, 35, 45, and 60 min post-injection (pi) and compared with the IDIFs. IDIFs were obtained by manually placing multiple circular-shaped regions of interest (ROI) within the AA, DA, aortic arch, and LV. Ten ROIs were sequentially arranged in the center of each structure, creating a total volume of 1.9602 ml. ROI placement was guided using the early PET frames and the low-dose CT scan, and the same ROIs were then copied to all PET frames. ROI positioning was checked and adjusted if needed to ensure accurate placement in every frame. In cases of PET-CT mismatch, ROIs were realigned based on the PET images.

TACs were generated by projecting the ROIs onto the dynamic PET frames and compared with venous blood samples. Bias between whole-blood activity from the venous blood samples compared to the IDIFs was determined for each time point that the venous blood samples were collected by interpolating the IDIF to the corresponding time. Relative bias for each IDIF at each time point was calculated using the following equation:

%bias=((interpolatedIDIFvenousbloodsample)/venousbloodsample)*100

A relative bias below 10% was considered an acceptable agreement between IDIF and venous blood samples and an acceptable agreement was expected after approximately 15–30 min when the arteriovenous equilibrium is established (16, 17). The relative bias is summarized as medians and interquartile ranges (IQR) and represented with Tukey's boxplots.

Results

Patient demographics

Eleven oncologic patients (36.4% female) were included in the study. The cohort included four patients with multiple myeloma, one with breast cancer, one with melanoma, one with a Kaposi sarcoma, two with large B-cell lymphoma (DLBCL), one with a T-cell lymphoma and one with a salivary gland tumor. Nobody had a history of type I or II diabetes mellitus. Patient demographics are summarized in Table 1.

Table 1
www.frontiersin.org

Table 1. Patient demographics.

IDIF validation

The accuracy of the IDIFs compared to the venous blood samples improved over uptake time. Substantial differences between the IDIFs and venous blood samples were found in 5, 10, and 15 min pi. The median percentage bias between the venous blood samples and the IDIFs was <10% from 25 min pi and onwards, with the best alignment at 60 min pi. At 25 min pi, the IDIF derived from the AA showed the smallest relative bias compared to the venous blood samples (median 5.8%, IQR 13.3%), followed by the aortic arch (median 7.9%, IQR 10.2%) and DA (median 7.9%, IQR 12.4%). The IDIF derived from the LV showed the largest bias (median 9.2%, IQR 11.7%). At 60 min pi, the IDIF from the aortic arch showed the smallest relative bias compared with venous blood samples (median −1.1%, IQR 5.9%), followed by the AA (median 2.5%, IQR 7.0%), DA (median 5.1%, IQR 8.6%), and LV (median 7.4%, IQR 7.6%). The relative median and IQR for each IDIF compared to the venous blood samples for each time point are summarized in Table 2 and illustrated in Figure 1.

Table 2
www.frontiersin.org

Table 2. Relative bias in whole-blood 18F-FDG activity between IDIFs and venous blood samples at seven time points post-injection.

Figure 1
www.frontiersin.org

Figure 1. Relative bias in whole-blood 18F-FDG activity between venous blood samples and image-derived input functions (IDIFs) of the ascending aorta (AA), descending aorta (DA), aortic arch, and left ventricle (LV) for seven time points post-injection (pi). The horizontal dotted line indicates 0% bias and the two dashed lines indicate −10% and 10% bias. The central line of the box represents the median and the edges are the 25th and 75th percentiles. The extreme data points which are not considered outliers are illustrated with the whiskers.

Discussion

With this study, we aimed to validate the performance of IDIFs obtained from the AA, DA, aortic arch, and LV by comparing them to (late, >30 min p.i.) whole-blood venous blood samples in patients undergoing LAFOV PET/CT imaging.

We observed an acceptable agreement (<10% relative bias) between venous blood and IDIFs from 25 min pi onward, with the best alignment at 60 min pi. Substantial relative bias in whole-blood 18F-FDG activity was found between IDIFs and venous blood at 5, 10, and 15 min pi, suggesting that approximately 15 to 25 min were needed to establish an equilibrium in whole-blood 18F-FDG activity between venous and arterial blood. The best alignment at 60 min pi is likely due to the distribution of 18F-FDG into tissues as a steady-state appears after approximately 60 min (20). Among the evaluated IDIFs, the aortic arch showed the smallest bias relative to venous blood, closely followed by the AA and DA. The superior performance of aorta-derived IDIFs found in this study aligns well with prior research on SAFOV (12, 18, 19) and LAFOV PET/CT systems (8). In contrast, the IDIF derived from the LV exhibited an increased bias, likely due to myocardial spillover, as previously reported by Weerdt et al. (12) and Sari et al. (9) using SAFOV PET/CT systems.

Given the high precision of IDIFs at later time points, our findings suggest that manual blood sampling for IDIF calibration may not be necessary. This would allow IDIFs to serve as the sole input function for dynamic 18F-FDG PET imaging. This would make dynamic 18F-FDG PET imaging more clinically feasible by simplifying study protocols, reducing invasiveness for patients, and minimizing radiation exposure for personnel (1315). If IDIF calibration is required, it may be feasible to use venous blood samples collected at later time points instead of arterial blood sampling. While this approach is not entirely non-invasive, it is significantly less invasive and more practical for clinical implementation than arterial blood sampling (1315). Our results align with the findings of Palard-Novello et al. (8) and support prior SAFOV PET/CT research (12, 18, 19), which indicated that manual blood sampling can be replaced with an aorta-derived IDIF. Additionally, using IDIFs may reduce variability introduced by delays in venous blood sampling, which would otherwise require correction.

A limitation of this study is the substantial relative bias observed during the first 15 min pi. However, this aligns with previous findings (16, 17), which indicate that it takes approximately 15–30 min for arteriovenous equilibrium to be established. Therefore, the early differences observed are likely attributable to physiological disparities between arterial and venous blood rather than inaccuracies in the IDIFs. Had arterial whole-blood samples been available for comparison, these early discrepancies might not have been present. Nevertheless, venous blood sampling was preferred over arterial sampling due to its lower invasiveness and reduced risks such as arterial occlusions and peripheral tissue ischemia (1315).

In conclusion, venous whole-blood activity showed accurate alignment with IDIFs from 25 min pi onward, with the best alignment at 60 min pi. IDIFs derived from the AA, aortic arch, and DA exhibited the highest precision, whereas the LV showed increased bias, making aorta-derived IDIFs preferable. The high precision of aorta-derived IDIFs suggests that uncorrected IDIFs may serve as a reliable alternative to manual blood sampling in dynamic 18F-FDG PET scans conducted on a LAFOV PET/CT system. The use of IDIFs reduces variability, simplifies study protocols, and enhances safety for both patients and personnel, as IDIFs are non-invasive and minimize radiation exposure.

Data availability statement

The datasets generated during and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Ethics statement

The studies involving humans were approved by Amsterdam UMC location Vrije Universiteit Amsterdam, Boelelaan 1117, Amsterdam, The Netherlands. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

CS: Conceptualization, Formal analysis, Investigation, Methodology, Project administration, Visualization, Writing – original draft. GZ: Conceptualization, Methodology, Supervision, Writing – review & editing. MH: Data curation, Project administration, Writing – review & editing. HG: Methodology, Writing – review & editing. NG: Investigation, Methodology, Writing – review & editing. JZ: Investigation, Writing – review & editing. CM: Investigation, Writing – review & editing. IB: Investigation, Writing – review & editing. MY: Conceptualization, Methodology, Supervision, Writing – review & editing. RB: Conceptualization, Funding acquisition, Methodology, Supervision, Writing – review & editing.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declare that no Generative AI was used in the creation of this manuscript.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Boellaard R, Oyen WJ, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, et al. The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging. (2008) 35:2320–33. doi: 10.1007/s00259-008-0874-2

PubMed Abstract | Crossref Full Text | Google Scholar

2. Wechalekar K, Sharma B, Cook G. PET/CT in oncology–a major advance. Clin Radiol. (2005) 60:1143–55. doi: 10.1016/j.crad.2005.05.018

PubMed Abstract | Crossref Full Text | Google Scholar

3. Huang SC. Anatomy of SUV. Standardized uptake value. Nucl Med Biol. (2000) 27:643–6. doi: 10.1016/S0969-8051(00)00155-4

PubMed Abstract | Crossref Full Text | Google Scholar

4. Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med. (2009) 50(Suppl 1):11S–20. doi: 10.2967/jnumed.108.057182

PubMed Abstract | Crossref Full Text | Google Scholar

5. Keyes JW Jr. SUV: standard uptake or silly useless value? J Nucl Med. (1995) 36:1836–9.7562051

PubMed Abstract | Google Scholar

6. Boellaard R. Need for standardization of 18F-FDG PET/CT for treatment response assessments. J Nucl Med. (2011) 52(Suppl 2):93S–100. doi: 10.2967/jnumed.110.085662

PubMed Abstract | Crossref Full Text | Google Scholar

7. Zaidi H, Karakatsanis N. Towards enhanced PET quantification in clinical oncology. Br J Radiol. (2018) 91:20170508. doi: 10.1259/bjr.20170508

PubMed Abstract | Crossref Full Text | Google Scholar

8. Palard-Novello X, Visser D, Tolboom N, Smith CLC, Zwezerijnen G, van de Giessen E, et al. Validation of image-derived input function using a long axial field of view PET/CT scanner for two different tracers. EJNMMI Phys. (2024) 11:25. doi: 10.1186/s40658-024-00628-0

PubMed Abstract | Crossref Full Text | Google Scholar

9. Sari H, Mingels C, Alberts I, Hu J, Buesser D, Shah V, et al. First results on kinetic modelling and parametric imaging of dynamic (18)F-FDG datasets from a long axial FOV PET scanner in oncological patients. Eur J Nucl Med Mol Imaging. (2022) 49:1997–2009. doi: 10.1007/s00259-021-05623-6

PubMed Abstract | Crossref Full Text | Google Scholar

10. Freedman NM, Sundaram SK, Kurdziel K, Carrasquillo JA, Whatley M, Carson JM, et al. Comparison of SUV and patlak slope for monitoring of cancer therapy using serial PET scans. Eur J Nucl Med Mol Imaging. (2003) 30:46–53. doi: 10.1007/s00259-002-0981-4

PubMed Abstract | Crossref Full Text | Google Scholar

11. Samimi R, Kamali-Asl A, Geramifar P, van den Hoff J, Rahmim A. Short-duration dynamic FDG PET imaging: optimization and clinical application. Phys Med. (2020) 80:193–200. doi: 10.1016/j.ejmp.2020.11.004

PubMed Abstract | Crossref Full Text | Google Scholar

12. van der Weerdt AP, Klein LJ, Boellaard R, Visser CA, Visser FC, Lammertsma AA. Image-derived input functions for determination of MRGlu in cardiac (18)F-FDG PET scans. J Nucl Med. (2001) 42:1622–9.11696630

PubMed Abstract | Google Scholar

13. Hall R. Vascular injuries resulting from arterial puncture of catheterization. Br J Surg. (1971) 58:513–6. doi: 10.1002/bjs.1800580711

PubMed Abstract | Crossref Full Text | Google Scholar

14. Chim H, Bakri K, Moran SL. Complications related to radial artery occlusion, radial artery harvest, and arterial lines. Hand Clin. (2015) 31:93–100. doi: 10.1016/j.hcl.2014.09.010

PubMed Abstract | Crossref Full Text | Google Scholar

15. Everett BA, Oquendo MA, Abi-Dargham A, Nobler MS, Devanand DP, Lisanby SH, et al. Safety of radial arterial catheterization in PET research subjects. J Nucl Med. (2009) 50:1742. doi: 10.2967/jnumed.109.063206

PubMed Abstract | Crossref Full Text | Google Scholar

16. Yokoyama I, Inoue Y, Moritan T, Ohtomo K, Nagai R. Measurement of skeletal muscle glucose utilization by dynamic 18F-FDG PET without arterial blood sampling. Nucl Med Commun. (2005) 26:31–7. doi: 10.1097/00006231-200501000-00006

PubMed Abstract | Crossref Full Text | Google Scholar

17. Reed MB, Handschuh PA, Schmidt C, Murgas M, Gomola D, Milz C, et al. Validation of cardiac image-derived input functions for functional PET quantification. Eur J Nucl Med Mol Imaging. (2024) 51:2625–37. doi: 10.1007/s00259-024-06716-8

PubMed Abstract | Crossref Full Text | Google Scholar

18. Ohtake T, Kosaka N, Watanabe T, Yokoyama I, Moritan T, Masuo M, et al. Noninvasive method to obtain input function for measuring tissue glucose utilization of thoracic and abdominal organs. J Nucl Med. (1991) 32:1432–8.2066802

PubMed Abstract | Google Scholar

19. Huisman MC, van Golen LW, Hoetjes NJ, Greuter HN, Schober P, Ijzerman RG, et al. Cerebral blood flow and glucose metabolism in healthy volunteers measured using a high-resolution PET scanner. EJNMMI Res. (2012) 2:63. doi: 10.1186/2191-219X-2-63

PubMed Abstract | Crossref Full Text | Google Scholar

20. Boellaard R, Delgado-Bolton R, Oyen WJ, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. (2015) 42:328–54. doi: 10.1007/s00259-014-2961-x

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: IDIF, LAFOV PET, 18F-FDG, dynamic PET imaging, non-invasive quantification

Citation: Smith CLC, Zwezerijnen GJC, den Hollander ME, Greuter HNJM, Gerards NR, Zijlstra J, Menke-van der Houven van Oordt CW, Bahce I, Yaqub M and Boellaard R (2025) Validating image-derived input functions of dynamic 18F-FDG long axial field-of-view PET/CT studies. Front. Nucl. Med. 5:1556848. doi: 10.3389/fnume.2025.1556848

Received: 7 January 2025; Accepted: 11 February 2025;
Published: 28 February 2025.

Edited by:

Charalampos Tsoumpas, University Medical Center Groningen, Netherlands

Reviewed by:

Florent L. Besson, Assistance Publique Hopitaux De Paris, France
Ole Munk, Aarhus University Hospital, Denmark

Copyright: © 2025 Smith, Zwezerijnen, den Hollander, Greuter, Gerards, Zijlstra, Menke-van der Houven van Oordt, Bahce, Yaqub and Boellaard. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Charlotte L. C. Smith, Yy5sLmMuc21pdGhAYW1zdGVyZGFtdW1jLm5s

ORCID:
Charlotte L. C. Smith
orcid.org/0009-0008-8558-4872

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Research integrity at Frontiers

Man ultramarathon runner in the mountains he trains at sunset

94% of researchers rate our articles as excellent or good

Learn more about the work of our research integrity team to safeguard the quality of each article we publish.


Find out more