AUTHOR=Liu Hui , Du Bingqing , Liu Bo , Fu Xinxing , Wang Yao TITLE=Clinical comparison of two automated audiometry procedures JOURNAL=Frontiers in Neuroscience VOLUME=16 YEAR=2022 URL=https://www.frontiersin.org/journals/neuroscience/articles/10.3389/fnins.2022.1011016 DOI=10.3389/fnins.2022.1011016 ISSN=1662-453X ABSTRACT=Objective

Automated pure-tone audiometry has been shown to provide similar hearing threshold estimates to conventional audiometry, but lower correlations were reported at high and low frequencies in audiometric tests than those of manual tests, while the correlations were better in the middle frequencies. In this paper, we used the same equipment and different test procedures for automated testing, and compared the results with manual test results.

Design

One hundred subjects aged 18–36 years were randomly divided into two groups to perform air-conduction pure-tone audiometry (0.25, 0.5, 1, 2, 4, 8 kHz) using the ascending and shortened ascending protocols built-in to the automated audiometer, respectively. Recorded testing time, the total number of responses and the subject’s preference tests were compared with those of manual tests.

Results

Significant difference was found at 250 Hz regarding the distribution of the absolute difference between the two automated and the manual thresholds. The testing time spend in the ascending method (9.8 ± 1.4 min, mean ± SD) was significantly longer than in the shorted ascending method (5.8 ± 0.9 min). The total numbers of responses of the ascending method (90.5 ± 10.8 times) and shorted ascending method (62.0 ± 11.4 times) were significantly different. Finally, no significant difference was found in preferences between automated and manual procedures.

Conclusion

The shorted ascending method can save lots of testing time. The difference between the two automated thresholds at 250 Hz is caused by the different test procedures, and the difference at 8,000 Hz between the automated test and the manual test can be due to the transducer types and allowable differences in calibration.