AUTHOR=Ma Fang-Zhou , Liu De-Feng , Yang An-Chao , Zhang Kai , Meng Fan-Gang , Zhang Jian-Guo , Liu Huan-Guang TITLE=Application of the robot-assisted implantation in deep brain stimulation JOURNAL=Frontiers in Neurorobotics VOLUME=16 YEAR=2022 URL=https://www.frontiersin.org/journals/neurorobotics/articles/10.3389/fnbot.2022.996685 DOI=10.3389/fnbot.2022.996685 ISSN=1662-5218 ABSTRACT=Introduction

This work aims to assess the accuracy of robotic assistance guided by a videometric tracker in deep brain stimulation (DBS).

Methods

We retrospectively reviewed a total of 30 DBS electrode implantations, assisted by the Remebot robotic system, with a novel frameless videometric registration workflow. Then we selected 30 PD patients who used stereotactic frame surgery to implant electrodes during the same period. For each electrode, accuracy was assessed using radial and axial error.

Results

The average radial error of the robot-assisted electrode implantation was 1.28 ± 0.36 mm, and the average axial error was 1.20 ± 0.40 mm. No deaths or associated hemorrhages, infections or poor incision healing occurred.

Conclusion

Robot-assisted implantation guided by a videometric tracker is accurate and safe.