Symmetric biphasic pulses have been shown to increase the therapeutic window compared to standard cathodic pulses in ET Vim-DBS patients. Furthermore, three hours of stimulation with biphasic pulses caused less stimulation-induced ataxia compared to cathodic pulses. Therefore, an investigation of the longer-term safety of biphasic pulses is warranted.
Seven ET patients were included in a randomized double-blind, cross-over design of one week home-use of symmetric biphasic stimulation (anodic phase first) versus cathodic stimulation. Amplitude was set in a double-blinded way, at the tremor arrest threshold. The primary outcome was safety assessed by documenting the adverse events. Secondary outcome parameters were stimulation amplitude, tremor (Fahn-Tolosa-Marin Tremor Rating Scale) and ataxia (International Cooperative Ataxia Rating Scale) severity, quality of life (Quality of Life in Essential Tremor Questionnaire) and cognition (Montreal Cognitive Assessment). Three patients continued in the open-label extension phase for 3 months, during which biphasic stimulation-only was further assessed by the same outcome parameters.
During the 1 week testing, no adverse effects were reported. To obtain equivalent tremor control, the amplitude of the biphasic pulse was significantly higher compared to that of the cathodic pulse (
In a small cohort, when tested for one week, symmetric biphasic pulses suggest to be safe, but require higher stimulation amplitudes. Further follow-up studies are needed to investigate long-term effects and safety.