- 1Clinical Microbiology and Infectious Diseases Department, Hospital General Universitario Gregorio Marañón, Madrid, Spain
- 2Instituto de Investigación Sanitaria Gregorio Marañón, Madrid, Spain
- 3CIBER de Enfermedades Respiratorias (CIBERES CB06/06/0058), Madrid, Spain
- 4Medicine Department, Faculty of Medicine, Universidad Complutense de Madrid, Madrid, Spain
- 5Francisco Soria-Melguizo SA, Madrid, Spain
Despite the benefits of MALDI-TOF MS technology (Matrix-Assisted Laser Desorption-Ionization Time-Of-Flight Mass Spectrometry) reported worldwide and the continuous improving of the available databases, discrimination between Streptococcus pneumoniae and closely related species within the Streptococcus mitis group (SMG) using this methodology has been suboptimal. However, the accurate identification at the species level of this group of microorganisms is important for the appropriate management of infected patients. In this study, 216 SMG isolates -101 S. pneumoniae and 115 corresponding to 7 non-pneumococcal species within this group- were analyzed. All the isolates had been previously identified by conventional methods (optochin and bile solubility tests) and non-pneumococcal isolates were confirmed by sequence analysis (sodA and plys genes) when required. The isolates were also identified with the MALDI Biotyper 3.1 (Bruker Daltonics, Bremen, Germany) using an updated library containing 6,903 Main Spectra Profiles (MSPs). All the analyzed S. pneumoniae were correctly identified with MALDI-TOF MS at species level using the most updated database and all the non-pneumococcal SMG isolates were also identified at the group level. Several peaks (4,964.32, 6,888.90, and 9,516.46 m/z) have been found to be specific of S. pneumoniae, whilst a different set of peaks have proved to be present only in S. mitis (6,839.07 m/z) and S. oralis (5,297.61, 5822.53, and 6,839.07 m/z). Peak analysis allowed correct species assignment of 101/101 S. pneumoniae isolates (100%) and 102/105 S. mitis/oralis isolates (97.1%). Thus, the implementation of MALDI-TOF MS plus peak analysis for the identification of this group of microorganisms may provide precise species-level information that will allow the early implementation of directed antibiotic therapy.
Introduction
The wide implementation of MALDI-TOF MS (Matrix Assisted Laser Desorption Ionization Time-of-Flight Mass Spectrometry) has led to its use for species assignment within the viridans group streptococci (VGS) and its differentiation from Streptococcus pneumoniae. However, that task currently represents a challenge for the mass spectrometry techniques as well as for the molecular methods and new technologies (Doern and Burnham, 2010). Efforts to get an accurate identification of VGS species are directed to determine the clinical significance of the organism and to promptly initiate directed antibiotic treatment.
Conventional tests such as optochin sensitivity and bile solubility are still applied in laboratory routine and represent the reference method for the identification of S. pneumoniae isolates (Yahiaoui et al., 2016). However, S. pneumoniae isolates resistant to optochin have been reported in different geographical regions (Nagata et al., 2012) making this tests less than optimal also due to its low specificity (Yahiaoui et al., 2016).
The development and implementation of genotypic methods allowed a more complete view of the complexity within alpha-haemolytic Streptococci and defined six main groups of VGS: S. anginosus, S. bovis, S. mitis, S. mutans, S. salivarius, and S. sanguinis groups (Facklam, 2002; Doern and Burnham, 2010). DNA analysis demonstrated that several species within this group share >99% sequence homology. This is the case of Streptococcus mitis, S. oralis and S. pneumoniae (Richter et al., 2008). Thus, sequencing of housekeeping genes such as 16S rRNA does not allow enough discrimination of these species (Hoshino et al., 2005) and different targets are required. Genes such as tuf, sodA, rpoB, or gyrB (Simmon et al., 2008; Póntigo et al., 2015) have been analyzed in order to determine species-specific sequences. Efforts have also been directed to the amino acid sequencing of the protein encoded by the gyrB gene in order to reliably identify VGS species (Galloway-Peña et al., 2014).
The availability of MALDI-TOF MS (Matrix-Assisted Laser Desorption Ionization Time of Flight Mass Spectrometry) has represented a new approach for the reliable identification of VGS at the species level. Several authors have evaluated the available MALDI-TOF instruments (Kärpänoja et al., 2014; Angeletti et al., 2015), improved samples preparation methods (Schulthess et al., 2013) and the use of peak analysis (Werno et al., 2012; Ikryannikova et al., 2013) in order to achieve accurate identification of VGS. In this study we applied an updated Biotyper database (Bruker Daltonics, Bremen, Germany) to the identification of SMG species -the most challenging species for MALDI-TOF MS to identify- from colonies grown on agar plates, without protein extraction procedures. Besides, peak analysis was performed on well characterized isolates by DNA sequencing (sodA and plys gens) in order to find putative species-specific peaks that may facilitate the identification of SMG isolates at the species level.
Materials and Methods
Isolates and Culture Conditions
We analyzed a total of 216 clinical isolates of SMG from the collection of isolates identified in our microbiology laboratory between 2014 and 2016. The isolates consisted of 101 S. pneumoniae from blood (68), bronchial aspirates (14), conjunctival and otic exudates (5), normally sterile fluids (5), bronchoalveolar lavage (3), sputum (3), and others (3) and 115 non-pneumococcal SMG isolates from blood (52), normally sterile fluids (16), abscesses (14), urine (13), different types of biopsies (11), surgical wounds (5), and bronchial or tracheal aspirates (4). Besides, 8 reference strains were also analyzed: S. anginosus (ATCC 12395), S. equinus (ATCC 33317), S. gordonii (ATCC 33399), S. intermedius (ATCC 27335), S. mutans (ATCC 25175), S. oralis (ATCC 35037), S. pneumoniae (ATCC 27336), and S. salivarius (ATCC 7073). All the isolates were subcultured on Columbia agar + 5% sheep blood (Biomérieux, Marcy L'Étoile, France) and incubated at 37°C in 5% CO2 for 48 h.
Conventional and Genotypic Identification
All the strains were identified as VGS by Gram staining, catalase test and alpha-haemolysis. Besides, the optochin susceptibility and bile solubility tests confirmed the identification of S. pneumoniae. On the other hand, the non-pneumococcal isolates were all confirmed by their resistance to optochin (Kellogg et al., 2001; Facklam, 2003).
DNA sequence analysis of the superoxide dismutase (sodA) and the pneumolysin (plys) genes was performed in order to provide species-level identification to non-pneumococcal isolates. A 480-bp fragment within the sodA gene was amplified using the degenerated primers d1 (5′'-CCITAYICITAYGAYGCIYTIGARCC-3′') and d2 (5′'-ARRTARTAIGCRTGYTCCCAIACRTC-3; Poyart et al., 1998). For the amplification of the plys gene, the forward primer (5′'-TGCAGAGCGTCCTTTGGTCTAT-3′') and reverse primers (5′'-TGCAGAGCGTCCTTTGGTCTAT-3′') were used in combination with a labeled probe (VIC–5′'-TGGCGCCCATAAGCAACACTCGAA-3′') as described by Corless et al. (2001). The amplified fragments were purified using the GFX PCR DNA and Gel Band Purification kit (GE Healthcare, UK) and sequenced with an AbiPrism 3,130 × l Genetic Analyzer equipment (Applied Biosystems, CA, USA). Sequences were compared with the GeneBank library using BLAST alignment software (http://www.ncbi.nlm.nih.gov/blast) and BIBI (bioInformatic bacterial identification tool: https://umr5558-bibiserv.univ-lyon1.fr/lebibi/lebibi.cgi). And the identification of the strains was carried out following the criteria established by the CLSI guidelines –(CLSI, 2008)-, with species level identification only when the homology was >99%.
Conventional methods and DNA sequencing analysis were considered as the reference methods for SMG identification.
MALDI-TOF Identification
All SMG isolates were analyzed by MALDI-TOF MS, using a Microflex LT bench top mass spectrometer (Bruker Daltonics, Bremen, Germany) as described previously –(Rodríguez-Sánchez et al., 2016)- Briefly, FlexControl 3.3 and MALDI Biotyper 3.1 software (Bruker Daltonics, Germany) was used for the acquisition of the spectra and for spectra analysis and subsequent comparison with the database (updated with 6,903 MSP entries), respectively. The available database contains 31 S. pneumoniae, 39 S. mitis and 38 S. oralis references. A commercial standard Bacterial Test Standard (Bruker Daltonics, Bremen, Germany) was used for calibration of the equipment and acquisition of mass spectra was performed using default settings.
Sample preparation prior to MALDI-TOF identification has been described before –(Rodríguez-Sánchez et al., 2014)-. A small amount of bacteria was transferred to a polished steel MALDI target plate using a wooden toothpick. The spots were covered with 1 μl of 100% formic acid and allowed to dry at room temperature prior to the addition of 1 μl of matrix (α-cyano-4-hydroxy-cinnamic acid solution in 50% acetonitrile and 2.5% trifluoroacetic acid) -prepared following the manufacturer's instructions-. The spots were allowed to air dry before spectra acquisition and comparison with the database. All the isolates were analyzed by MALDI-TOF MS in duplicates and the higher score value was recorded as well as the identification provided by MALDI-TOF MS. Score values for species level (≥2.0) and genus-level identification (≥1.7) used in this study are those proposed by the manufacturer. Score values below 1.7 were, accordingly, considered as not reliable. Identification agreement at the species level was considered when MALDI-TOF provided the same result as the sodA gene sequencing and the conventional methods. However, the identification was considered a major error when MALDI-TOF MS identified an isolate as S. pneumoniae whilst the conventional methods and sodA gene sequencing indicated that the isolate was a non-pneumococcal isolate and viceversa. When the discrepancy was between two non-pneumococcal SMG species it was considered as a minor error.
Peak Analysis
For peak analysis, 20 S. pneumoniae, 20 S. mitis and 20 S. oralis isolates were submitted to a standard protein extraction procedure: a 1 μl-loopful of bacteria was resuspended in 300 μl of distilled water and 900 μl ethanol, vortexed and centrifuged at maximal speed for 1 min; the pellet was then resuspended in 20 μl of 70% formic acid and the same amount of acetonitrile, vortexed again and centrifuged again for another minute at maximal speed. 1 μl of supernatant was pipetted onto the MALDI plate, allowed to dry and covered with 1 μl of matrix. The supernatant from every strain was spotted onto 6 positions. In a different spot, 1 μl of BTS was spotted on top of the dried supernatant, allowed to dry and covered with 1 μl of matrix. This position was used for calibration purposes.
FlexAnalysis 3.3 (Bruker Daltonics, Bremen, Germany) was applied for the comparison of mass spectra from the different SMG species. Spectra were smoothed, the baseline was subtracted and BTS was used as references for calibration purposes. Peaks from individual spectra were compared for strain-specific variations and spectra from all 20 isolates from the same species were considered together as a single MSP entry in order to find species-specific peaks.
The 60 isolates submitted to protein extraction were used for the characterization of species-specific peaks present in each Streptococcus species. Afterwards, the presence of species-specific peaks was analyzed in all S. pneumoniae-101-, S. mitis-50-, and S. oralis-55- isolates included in this study.
Ethics Statement
The hospital Ethics Committee approved this study and gave consent for its development. Since this study has been performed on microbiological samples, not human products, all the conditions to waive the informed consent have been met.
Results
Routine Identification of SMG Isolates with Maldi-TOF MS
During the study period, 216 SMG isolates were identified with MALDI-TOF MS as part of the clinical microbiology laboratory routine. MALDI Biotyper with 6,903 MSPs correctly discriminated S. pneumoniae from non-pneumococcal SMG isolates. Therefore, no major errors were detected when this database was implemented. Concordantly, all non-pneumococcal isolates were correctly identified at group level, although at the species level 14 S. mitis/oralis isolates were misidentified according to sodA sequencing analysis results (Table 1).
Table 1. Isolates included in this study and their identification using MALDI-TOF MS with the Biotyper database 6903 MSPs.
Interestingly, one of the isolates selected for this study and wrongly listed as a non-pneumococcal SMG isolate was consistently identified by MALDI-TOF MS as S. pneumoniae (score value = 2.524). When its identification by the optochin test, the bile solubility test and sodA gene sequencing was double-checked it turned out that the isolate was actually a S. pneumoniae one, thus, confirming MALDI-TOF MS identification.
Peak Analysis
Despite MALDI-TOF MS high level of accuracy in the identification of SMG isolates, in this study we have observed that challenging strains may still be found, leading to misidentifications. Thus, the protein spectra from the 101 analyzed S. pneumoniae isolates were compared with the spectra from those belonging to the S. mitis group, in order to single out S. pneumoniae specific peaks. One peak was found in at least 95.0% of the S. pneumoniae isolates -4,964.32 m/z- and other two peaks in 90.5% of the isolates: 6,888.90 and 9,516.46 m/z –Table 2. Their presence was also confirmed in the reference strain ATCC 27336 (Figure 1).
Figure 1. Distinctive species-specific peaks present in close related Streptococcus species. The 6,839.07 m/z has been shown to be present in protein spectra from S. mitis isolates (above). This peak is also present in S. oralis spectra (middle) in combination with the 5,297.61 and the 5,822.53 m/z peaks. On the other hand, S. pneumoniae isolates (bottom) showed three unique peaks of 4,964.32, 6,888.90, and 9,516.46 m/z.
On the other hand, 92.0% of the S. oralis strains showed two specific peaks of 5,297.61 and 5,822.53 m/z also present in the reference strain ATCC 35037. These peaks were not detected in S. pneumoniae isolates neither in the other species of non-pneumococcal VGS analyzed. Besides, S. mitis and S. oralis isolates shared a peak of 6,839.07 m/z. This peak was the only species-specific feature of S. mitis isolates, whilst S. oralis displayed it together with the 5,297.61 and 5,822.53 m/z peaks (Figure 1).
Finally, although MALDI-TOF MS accurately identifies the remaining VGS groups (S. anginosus, S. bovis, S. mitis, S. mutans, S. salivarius, and S. sanguinis) and no misidentifications with S. pneumoniae have been reported as far as the authors know, peak comparison was performed between S. pneumoniae isolates and the rest of non-pneumococcal VGS, including the reference strains (data not shown). None of the S. pneumoniae specific peaks were found alone or in combination in the remaining non-pneumococcal VGS included in this study.
In our study, out of the 14 non-pneumococcal SMG isolates misidentified by the Biotyper system with 6,903 MSPs, all but two isolates could be correctly identified by peak analysis. The discrepant identifications corresponded to two S. mitis isolates identified by sodA sequencing analysis that were identified as S. oralis by the updated database and by peak analysis –Table 3. Further genomic targets should be analyzed in order to resolve these discrepancies.
Table 3. Reidentification by peak-analysis of SMG isolates that showed discrepancies in the species assignment by MALDI-TOF and the conventional or the molecular methods.
In summary, the updated database allowed the correct species-level identification of 100% of the S. pneumoniae isolates and 100% of the non-pneumococcal isolates at the group-level. Besides, the performance of peak analysis confirmed the identification of all S. pneumoniae isolates and provided a correct species-level discrimination of 103/105 S. mitis and S. oralis isolates (98.1%).
Discussion
The identification of SMG isolates with MALDI-TOF MS has been improved in the last years thanks to the availability of updated databases. In this study, 100% of the S. pneumoniae and 87.8% of the non-pneumococcal SGM isolates were correctly identified at the species level using an updated database. Similar studies have reported 100% accuracy of the Biotyper system with 5,267 entries in the identification of S. pneumoniae isolates, but only 61.0–66.7% species-level correct identification of non-pneumococcal VGS isolates and up to 23.4% of misidentification of these species as S. pneumoniae (Angeletti et al., 2015; Chen et al., 2015). Another study that applied the Biotyper database updated with 5,989 entries for the identification of VGS species reported 21 isolates of S. mitis/oralis/pseudopneumoniae out of 107 misidentified (19.6%) –(Zhou et al., 2016)-. These results are clearly far from the 14 misidentified isolates (6.5%) observed in our study.
A recent paper by Harju et al. (2017) has reported 100% correct species assignment to S. pneumoniae isolates and 99.0% correct group level identification of non-pneumoccocal isolates using the Biotyper library with 5,627 MSPs. By applying an algorithm that assigns a numerical value to each of the 10 identifications in the ranking list provided by the Biotyper software, the authors achieved 100% correct group-level identification of non-pneumococcal isolates.
In our study, 100% of the S. pneumoniae isolates were correctly identified at the species level using a more updated library and their identification was confirmed by peak analysis. Besides, group-level identification of non-pneumococcal isolates was obtained in 100% of the cases. Peak analysis of these isolates allowed species-level identification of 103/105 S. mitis and S. oralis isolates (98.1%) according to DNA sequencing analysis. Although these results may have little clinical value, the detection of the species-specific peaks reported in this study may provide important epidemiological information.
Peak analysis has been applied before for confirmation of MALDI-TOF MS identification (Werno et al., 2012; Ikryannikova et al., 2013; Chen et al., 2015). The peaks found by Werno et al. for the specific detection of S. pneumoniae (2,937.5 and 5,877 m/z) and S. mitis/oralis (2,911, 5,824, and 6,955 m/z) have been applied in our laboratory for routine discrimination of S. pneumoniae from non-pneumococcal VGS strains. However, none of the peaks propose by Ikryannikova et al. (2013) and Chen et al. (2015) for species discrimination was found in our isolates. The peak of 5,824 m/z described by Werno et al. as specific for S. mitis/oralis is in agreement with the 5,822.53 m/z found in our study as S. oralis specific. The species specific peaks described in this study allowed the confirmation of MALDI-TOF MS identification in 214/216 cases. Peak analysis enabled the correct identification of 12/14 S. mitis/oralis isolates misidentified by MALDI-TOF MS.
Thus, we can conclude that the improvement of the Biotyper database allows an accurate identification of SMG isolates from single colonies. However, this group of microorganisms contains highly similar species which pose a challenge for DNA- and protein-based diagnostic methods. For instance, Biotyper database with 6,903 entries contains only 5 representative spectra of S. pseudopneumoniae. Therefore, the chances to correctly identify this species using MALDI-TOF MS are scarce, and so were the possibilities to identify S. pseudopneumoniae by sequencing the sodA and the plys genes.
Despite the fact that the updated database misidentified a high number of S. mitis/oralis isolates (13.3%), peak analysis may allow a correct species assignment. Besides, this kind of misidentification is considered a minor error that does not hamper the early implementation of directed antibiotic therapy. Thus, our recommendation is to apply MALDI-TOF MS for the identification of S. pneumoniae and SMG isolates from single colonies and perform peak analysis only for S. mitis/oralis isolates for which both identifications alternate in the ranking list provided by MALDI-TOF MS and those that obtain low score values or to confirm the identity of S. pneumoniae isolates. The implementation of this approach may also save extra costs derived from the need to sequence the SGM isolates in order to get a final identification.
Author Contributions
MM carried out the sequencing of the isolates; EC performed the conventional methods for bacterial identification, discussed the results and reviewed the manuscript; CS collaborated in the design of the study and reviewed the manuscript; AR performed the analysis of the isolates by MALDI-TOF and analyzed the protein spectra; AG assisted in the spectra analysis; EB reviewed and discussed the manuscript and BR collaborated in the design of the study, the spectra analysis, the discussion of the results and the manuscript writting.
Conflict of Interest Statement
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Acknowledgments
This study has been supported by the Miguel Servet Program (ISCIII-MICINN CP14/00220) from the Health Research Fund (FIS) of the Carlos III Health Institute (ISCIII), Madrid, Spain, partially financed by the by the European Regional Development Fund (FEDER) “A way of making Europe.” The funders had no role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. ÁG works for F. Soria Melguizo. He assisted in the spectra analysis of the analyzed isolates as an expert in the use of FlexAnalysis software (Bruker Daltonics) for peak analysis. No commercial relationship was established with him or his company.
References
Angeletti, S., Dicuonzo, G., Avola, A., Crea, F., Dedej, E., Vailati, F., et al. (2015). Viridans Group Streptococci clinical isolates: MALDI-TOF mass spectrometry versus gene sequence-based identification. PLoS ONE 10:e0120502. doi: 10.1371/journal.pone.0120502
Chen, J. H., She, K. K., Wong, O. Y., Teng, J. L., Yam, W. C., Lau, S. K., et al. (2015). Use of MALDI Biotyper plus ClinProTools mass spectra analysis for correct identification of Streptococcus pneumoniae and Streptococcus mitis/oralis. J. Clin. Pathol. 68, 652–656. doi: 10.1136/jclinpath-2014-202818
CLSI (2008). Interpretive Criteria for Identification of Bacteria and Fungi by DNA Target Sequencing; Approved Guideline. CLSI Document MM18-a, Clinical and Laboratory Standards Institute. Wayne, PA: CLSI.
Corless, C. E., Guiver, M., Borrow, R., Edwards-Jones, V., Fox, A. J., and Kaczmarski, E. B. (2001). Simultaneous detection of Neisseria meningitidis, Haemophilus influenzae, and Streptococcus pneumoniae in suspected cases of meningitis and septicemia using real-time PCR. J. Clin. Microbiol. 39, 1553–1558. doi: 10.1128/JCM.39.4.1553-1558.2001
Doern, C. D., and Burnham, C. A. (2010). It's not easy being green: the viridans group streptococci, with a focus on pediatric clinical manifestations. J. Clin. Microbiol. 48, 3829–3835. doi: 10.1128/JCM.01563-10
Facklam, R. (2002). What happened to the streptococci: overview of taxonomic and nomenclature changes. Clin. Microbiol. Rev. 15, 613–630. doi: 10.1128/CMR.15.4.613-630.2002
Facklam, R. (2003). “Streptococcus pneumoniae, confirmatory identification and antimicrobial susceptibility testing,” in Manual for the Laboratory Identification and Antimicrobial Susceptibility Testing of Bacterial Pathogens of Public Health Importance in the Developing Word, ed J. Perilla Mindy (Atlanta, GA: WHO), 45–62.
Galloway-Peña, J., Sahasrabhojane, P., Tarrand, J., Han, X. Y., and Shelburne, S. A. (2014). GyrB polymorphisms accurately assign invasive viridans group streptococcal species. J. Clin. Microbiol. 52, 2905–2912. doi: 10.1128/JCM.01068-14
Harju, I., Lange, C., Kostrzewa, M., Maier, T., Rantakokko-Jalava, K., and Haanperä, M. (2017). Improved differentiation of Streptococcus pneumoniae and Other, S. mitis Group Streptococci by MALDI Biotyper Using an Improved MALDI Biotyper Database Content and a Novel Result Interpretation Algorithm. J. Clin. Microbiol. 55, 914–922. doi: 10.1128/JCM.01990-16
Hoshino, T., Fujiwara, T., and Kilian, M. (2005). Use of phylogenetic and phenotypic analyses to identify nonhemolytic streptococci isolated from bacteremic patients. J. Clin. Microbiol. 43, 6073–6085. doi: 10.1128/JCM.43.12.6073-6085.2005
Ikryannikova, L. N., Filimonova, A. V., Malakhova, M. V., Savinova, T., Filimonova, O., Ilina, E. N., et al. (2013). Discrimination between Streptococcus pneumoniae and Streptococcus mitis based on sorting of their MALDI mass spectra. Clin. Microbiol. Infect. 19, 1066–171. doi: 10.1111/1469-0691.12113
Kärpänoja, P., Harju, I., Rantakokko-Jalava, K., Haanperä, M., and Sarkkinen, H. (2014). Evaluation of two matrix-assisted laser desorption ionization-time of flight mass spectrometry systems for identification of viridans group streptococci. Eur. J. Clin. Microbiol. Infect. Dis. 33, 779–788. doi: 10.1007/s10096-013-2012-8
Kellogg, J. A., Bankert, D. A., Elder, C. J., Gibbs, J. L., and Smith, M. C. (2001). Identification of Streptococcus pneumoniae revisited. J. Clin. Microbiol. 39, 3373–3375. doi: 10.1128/JCM.39.9.3373-3375.2001
Nagata, M., Ueda, O., Shobuike, T., Muratani, T., Aoki, Y., and Miyamoto, H. (2012). Emergence of Optochin Resistance among Streptococcus pneumoniae in Japan. Open J. Med. Microbiol. 2, 8–15. doi: 10.4236/ojmm.2012.21002
Póntigo, F., Moraga, M., and Flores, S. V. (2015). Molecular phylogeny and a taxonomic proposal for the genus Streptococcus. Genet. Mol. Res. 14, 10905–10918. doi: 10.4238/2015.September.21.1
Poyart, C., Quesne, G., Coulon, S., Berche, P., and Trieu-Cuot, P. (1998). Identification of streptococci to species level by sequencing the gene encoding the manganese-dependent superoxide dismutase. J. Clin. Microbiol. 36, 41–47.
Richter, S. S., Heilmann, K. P., Dohrn, C. L., Riahi, F., Beekmann, S. E., and Doern, G. V. (2008). Accuracy of phenotypic methods for identification of Streptococcus pneumoniae isolates included in surveillance programs. J. Clin. Microbiol. 46, 2184–2188. doi: 10.1128/JCM.00461-08
Rodríguez-Sánchez, B., Alcalá, L., Marín, M., Ruiz, A., Alonso, E., and Bouza, E. (2016). Evaluation of MALDI-TOF MS (Matrix-Assisted Laser Desorption-Ionization Time-of-Flight Mass Spectrometry) for routine identification of anaerobic bacteria. Anaerobe 42, 101–107. doi: 10.1016/j.anaerobe.2016.09.009
Rodríguez-Sánchez, B., Marín, M., Sánchez-Carrillo, C., Cercenado, C., Ruiz, A., Rodríguez-Créixems, M., et al. (2014). Improvement of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry identification of difficult-to-identify bacteria and its impact in the workflow of a clinical microbiology laboratory, Diagn. Microbiol. Infect. Dis. 79, 1–6. doi: 10.1016/j.diagmicrobio.2014.01.021
Schulthess, B., Brodner, K., Bloemberg, G. V., Zbinden, R., Böttger, E. C., and Hombach, M. (2013). Identification of Gram-positive cocci by use of matrix-assisted laser desorption ionization-time of flight mass spectrometry: comparison of different preparation methods and implementation of a practical algorithm for routine diagnostics. J. Clin. Microbiol. 51, 1834–140. doi: 10.1128/JCM.02654-12
Simmon, K. E., Hall, L., Woods, C. W., Marco, F., Miro, J. M., Cabell, C., et al. (2008). Phylogenetic analysis of viridans group streptococci causing endocarditis. J. Clin. Microbiol. 46, 3087–3090. doi: 10.1128/JCM.00920-08
Werno, A. M., Christner, M., Anderson, T. P., and Murdoch, D. R. (2012). Differentiation of Streptococcus pneumoniae from nonpneumococcal streptococci of the Streptococcus mitis group by matrix-assisted laser desorption ionization-time of flight mass spectrometry. J. Clin. Microbiol. 50, 2863–2867. doi: 10.1128/JCM.00508-12
Yahiaoui, R. Y., den Heijer, C. D., Wolfs, P., Bruggeman, C. A., and Stobberingh, E. E. (2016). Evaluation of phenotypic and molecular methods for identification of Streptococcus pneumoniae. Future Microbiol. 11, 43–50. doi: 10.2217/fmb.15.124
Zhou, M., Yang, Q., Kudinha, T., Zhang, L., Xiao, M., Kong, F., et al. (2016). Using Matrix-Assisted Laser Desorption Ionization-Time of Flight (MALDI-TOF) Complemented with Selected 16S rRNA and gyrB Genes Sequencing to Practically Identify Clinical Important Viridans Group Streptococci (VGS). Front. Microbiol. 7:1328. doi: 10.3389/fmicb.2016.01328
Keywords: Streptococcus pneumoniae, Streptococcus mitis group, routine identification, MALDI-TOF MS, peak analysis
Citation: Marín M, Cercenado E, Sánchez-Carrillo C, Ruiz A, Gómez González Á, Rodríguez-Sánchez B and Bouza E (2017) Accurate Differentiation of Streptococcus pneumoniae from other Species within the Streptococcus mitis Group by Peak Analysis Using MALDI-TOF MS. Front. Microbiol. 8:698. doi: 10.3389/fmicb.2017.00698
Received: 08 March 2017; Accepted: 05 April 2017;
Published: 25 April 2017.
Edited by:
Jorge Blanco, Universidade de Santiago de Compostela, SpainReviewed by:
Javier Fernández, Central University Hospital of Asturias, SpainMarina Oviaño García, Complexo Hospitalario Universitario A Coruña, Spain
Copyright © 2017 Marín, Cercenado, Sánchez-Carrillo, Ruiz, Gómez González, Rodríguez-Sánchez and Bouza. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Belén Rodríguez-Sánchez, bWJlbGVuLnJvZHJpZ3VlekBpaXNnbS5jb20=
†These authors have contributed equally to this work.