To assess corneal epithelial-thickness (ET) mapping resulting from spectral-domain-optical-coherence-tomography (SD-OCT) by analysing its repeatability and reproducibility and its utility for screening corneal-refractive-surgery (CRS) candidates.
ET was measured in 25-sectors by two-operators. Intra-subject-standard-deviation, coefficient-of-repeatability (CoR) and coefficient-of-variability (CoV) were calculated to evaluate repeatability. Reproducibility was evaluated using a Bland–Altman analysis. Scheimpflug-tomography, refraction, visual acuity, and patient history were used to make a decision on eligibility for CRS. After this decision, the surgeon was shown the patient’s ET map and was asked to reconsider his analysis. The percentage of screenings that changed after evaluating the ET maps was determined.
Forty-three eyes with normal corneas (CRS-group) and 21 eyes not suitable for CRS (non-CRS-group) were studied. For the CRS-group, CoR ranged from 2.03 (central) to 19.73 μm (outer-inferonasal), with the central-sector showing the highest repeatability (CoV: 1.53–1.80%). For the non-CRS-group, CoR ranged from 3.82 (central-middle-superonasal) to 13.42 μm (middle-inferotemporal), with the inner-superonasal-sector showing the highest repeatability (CoV: 2.86–4.46%). There was no statistically significant difference between operators (
The SD-OCT provided repeatable and reproducible corneal ET measurements and may alter candidacy for CRS.
German Clinical Trials Register: