AUTHOR=Reilly Leanne , Mac Giolla Eain Marc , Murphy Sarah , O’Sullivan Andrew , Joyce Mary , MacLoughlin Ronan TITLE=An in vitro study of the effects of respiratory circuit setup and parameters on aerosol delivery during mechanical ventilation JOURNAL=Frontiers in Medicine VOLUME=10 YEAR=2024 URL=https://www.frontiersin.org/journals/medicine/articles/10.3389/fmed.2023.1307301 DOI=10.3389/fmed.2023.1307301 ISSN=2296-858X ABSTRACT=Introduction

Aerosol therapy is often prescribed concurrently during invasive mechanical ventilation (IMV). This study determines the effects of nebuliser position, circuit humidification source, and most importantly, lung health on the delivery of aerosol in simulated adult and paediatric IMV patients. Furthermore, the influence of closed suction catheters on aerosol delivery is also addressed.

Methods

A vibrating mesh nebuliser was used to deliver Albuterol to simulated adult and paediatric IMV patients with differing states of lung health. Four different nebuliser positions and two types of humidification were analysed. Closed suction catheter mounts, a mainstay in IMV therapy, were incorporated into the circuits. The mean ± SD dose of aerosol (%) was assayed from a filter at the distal end of the endotracheal tube.

Results

Nebuliser placement and circuit humidification source had no effect on the delivered dose (%) in adults, yet both significantly did in the simulated paediatric patients. The use of closed suction catheter mounts significantly reduced the delivered dose (%) in adults but not in paediatric patients. A simulated healthy lung state generated the largest delivered dose (%), irrespective of nebuliser position in the adult. However, different lung health and nebuliser positions yielded higher delivered doses (%) in paediatrics.

Conclusion

Lung health and respiratory circuit composition significantly affect aerosol delivery in both adult and paediatric IMV patients. Nebuliser placement and respiratory circuit humidification source do not affect the delivered dose in adult but do in paediatric IMV patients.