AUTHOR=Chen Shengsen , Zhou Danping , Ruan Rongwei , Yu Jiangping , Li Yandong , Liu Yuanshun , Wang Shi TITLE=A Novel Bipolar Polypectomy Snare Can Be an Alternative Choice for Endoscopic Resection JOURNAL=Frontiers in Medicine VOLUME=7 YEAR=2021 URL=https://www.frontiersin.org/journals/medicine/articles/10.3389/fmed.2020.619844 DOI=10.3389/fmed.2020.619844 ISSN=2296-858X ABSTRACT=

Objective: Endoscopic resection (ER) is more difficult and has a higher rate of complications, such as perforation and bleeding. The aim of this study was to evaluate the safety and feasibility of a bipolar polypectomy snare for ER.

Methods: Initial ER procedures in live pigs were carried out. Then, a human feasibility study was performed in patients with colorectal polyps. Finally, the finite element method was used to evaluate the safety and effectiveness of the new bipolar snare.

Results: In the live animal model, there were no significant differences in wound size and cutting time between monopolar and bipolar groups. The histological results (histological scores) of the two groups in porcine experiments were almost the same except that the incision flatness of bipolar group was better than that of the monopolar group. Incidence of bleeding and perforation was similar between the two groups in pigs' and patients' study. At last, the finite element model showed that the vertical thermal damage depth produced by bipolar snare system was approximately 71–76% of that produced by monopolar snare system at the same power.