AUTHOR=Schaufelberger Matthias , Kühle Reinald Peter , Wachter Andreas , Weichel Frederic , Hagen Niclas , Ringwald Friedemann , Eisenmann Urs , Hoffmann Jürgen , Engel Michael , Freudlsperger Christian , Nahm Werner TITLE=Impact of data synthesis strategies for the classification of craniosynostosis JOURNAL=Frontiers in Medical Technology VOLUME=5 YEAR=2023 URL=https://www.frontiersin.org/journals/medical-technology/articles/10.3389/fmedt.2023.1254690 DOI=10.3389/fmedt.2023.1254690 ISSN=2673-3129 ABSTRACT=Introduction

Photogrammetric surface scans provide a radiation-free option to assess and classify craniosynostosis. Due to the low prevalence of craniosynostosis and high patient restrictions, clinical data are rare. Synthetic data could support or even replace clinical data for the classification of craniosynostosis, but this has never been studied systematically.

Methods

We tested the combinations of three different synthetic data sources: a statistical shape model (SSM), a generative adversarial network (GAN), and image-based principal component analysis for a convolutional neural network (CNN)–based classification of craniosynostosis. The CNN is trained only on synthetic data but is validated and tested on clinical data.

Results

The combination of an SSM and a GAN achieved an accuracy of 0.960 and an F1 score of 0.928 on the unseen test set. The difference to training on clinical data was smaller than 0.01. Including a second image modality improved classification performance for all data sources.

Conclusions

Without a single clinical training sample, a CNN was able to classify head deformities with similar accuracy as if it was trained on clinical data. Using multiple data sources was key for a good classification based on synthetic data alone. Synthetic data might play an important future role in the assessment of craniosynostosis.