AUTHOR=Roy Aritra , Singh Pushpak , Saha Simran , Sen Arvil , Ray Mayukh , Deb Majumder Samarpan , Dutta Sagnik , Chowdhuri Shubho , Mundhra Raghav TITLE=Design of a Multi-Control Objective Rescue Mechanical Ventilation System (Linshomator) JOURNAL=Frontiers in Medical Technology VOLUME=2 YEAR=2020 URL=https://www.frontiersin.org/journals/medical-technology/articles/10.3389/fmedt.2020.575964 DOI=10.3389/fmedt.2020.575964 ISSN=2673-3129 ABSTRACT=
In recent years, the need for a low-cost emergency ventilation system has increased unprecedentedly. Mechanical ventilation systems are widely used to cater to sudden oxygen demands, low breathing rates, and critical conditions during bradycardia and tachycardia. In this research, a new design of mechanical ventilation system synced with the tidal volume requirements of the patient using a piezoelectric belt has been proposed. The device proposed has various modes of ventilation—ACV (assisted controlled ventilation), SIMV (synchronized intermittent mandatory ventilation), and NIV (non-invasive ventilation), depending on the patient's requirements. A digital interface or user-friendly software has also been developed to help medical professionals easily monitor a patient's medical conditions. Finally, the automation that controls the ventilation mechanism of the device has been tested and validated with a conventional ventilator, and it has been found that the accuracy of the device in terms of delivering the exact quantity of air into the patient according to his requirements has been improved significantly. Further, the comparative study of the experimental data indicated that 5–10% error in detecting inhale and exhale attempt of a patient was detected with the conventional ventilator.