AUTHOR=Lei Rongwei , Vu Binh , Kourentzi Katerina , Soomro Sanam , Danthanarayana Adheesha N. , Brgoch Jakoah , Nadimpalli Suma , Petri Michelle , Mohan Chandra , Willson Richard C. TITLE=A novel technology for home monitoring of lupus nephritis that tracks the pathogenic urine biomarker ALCAM JOURNAL=Frontiers in Immunology VOLUME=13 YEAR=2022 URL=https://www.frontiersin.org/journals/immunology/articles/10.3389/fimmu.2022.1044743 DOI=10.3389/fimmu.2022.1044743 ISSN=1664-3224 ABSTRACT=Introduction

The gold standard for diagnosis of active lupus nephritis (ALN), a kidney biopsy, is invasive with attendant morbidity and cannot be serially repeated. Urinary ALCAM (uALCAM) has shown high diagnostic accuracy for renal pathology activity in ALN patients.

Methods

Lateral flow assays (LFA) for assaying uALCAM were engineered using persistent luminescent nanoparticles, read by a smartphone. The stability and reproducibility of the assembled LFA strips and freeze-dried conjugated nanoparticles were verified, as was analyte specificity.

Results

The LFA tests for both un-normalized uALCAM (AUC=0.93) and urine normalizer (HVEM)-normalized uALCAM (AUC=0.91) exhibited excellent accuracies in distinguishing ALN from healthy controls. The accuracies for distinguishing ALN from all other lupus patients were 0.86 and 0.74, respectively.

Conclusion

Periodic monitoring of uALCAM using this easy-to-use LFA test by the patient at home could potentially accelerate early detection of renal involvement or disease flares in lupus patients, and hence reduce morbidity and mortality.