Front. Endocrinol., 21 June 2023
Sec. Thyroid Endocrinology

The value of conventional ultrasound combined with superb microvascular imaging and color Doppler flow imaging in the diagnosis of thyroid malignant nodules: a systematic review and meta-analysis

Li Jiang,&#x;Li Jiang1,2†Dian Zhang&#x;Dian Zhang2†Yue-Nan ChenYue-Nan Chen2Xue-Juan YuXue-Juan Yu2Mei-Fang PanMei-Fang Pan2Lian Lian,*Lian Lian1,3*
  • 1Medical College, Yangzhou University, Yangzhou, Jiangsu, China
  • 2Department of Ultrasound, Suzhou Xiangcheng People’s Hospital, Jiangsu, China
  • 3Department of Oncology, Suzhou Xiangcheng People’s Hospital, Jiangsu, China

Purpose: To evaluate and compare the value of conventional ultrasound-based superb microvascular imaging (SMI) and color Doppler flow imaging (CDFI) in the diagnosis of malignant thyroid nodule by meta-analysis.

Methods: The literature included in the Cochrane Library, PubMed, and Embase were searched by using “ superb microvascular imaging (SMI), color Doppler flow imaging (CDFI), ultrasound, thyroid nodules” as the keywords from inception through February 1, 2023. According to the inclusion and exclusion criteria, the clinical studies using SMI and CDFI to diagnose thyroid nodules were selected, and histopathology of thyroid nodules was used as reference standard. The diagnostic accuracy research quality assessment tool (QUADAS-2) was used to evaluate the quality of included literature, and the Review Manager 5.4 was used to make the quality evaluation chart. The heterogeneity test was performed on the literature that met the requirements, the combined sensitivity, specificity, positive likelihood ratio, and negative likelihood ratio were pooled, and a comprehensive ROC curve analysis was performed. Meta-DiSc version 1.4, StataSE 12, and Review Manager 5.4 software were used.

Results: Finally, 13 studies were included in this meta-analysis. A total of 815 thyroid malignant nodules were assessed. All thyroid nodules were histologically confirmed after SMI or CDFI. The combined sensitivity, specificity, PLR, NLR, DOR, and area under the SROC curve of SMI for the diagnosis of malignant thyroid nodules were 0.80(95%CI: 0.77-0.83), 0.79(95%CI: 0.77-0.82), 4.37(95%CI: 3.0-6.36), 0.23(95%CI: 0.15-0.35), 22.29(95%CI: 12.18-40.78), and 0.8944, respectively; the corresponding values of CDFI were 0.62(95%CI: 0.57-0.67), 0.81(95%CI: 0.78-0.85), 3.33(95%CI: 2.18-5.07), 0.41(95%CI: 0.27-0.64), 8.93(95%CI: 3.96-20.16), and 0.8498. Deek funnel pattern showed no significant publication bias.

Conclusion: The diagnostic efficiency of SMI for malignant thyroid nodules is better than CDFI, and SMI technology can provide significantly more information on vascularity, make up for the deficiency of CDFI, and has better clinical application value.

Systematic review registration: https://www.crd.york.ac.uk/PROSPERO, identifier CRD42023402064.

1 Introduction

In recent years, the incidence rate of thyroid cancer is on the rise. With the wide application of ultrasound technology, ultrasound has become one of the most important means of thyroid detection (1, 2). Fine needle aspiration biopsy (FNA) is the best non-surgical diagnostic method for thyroid malignant tumor at present, but it is an invasive examination (36). Improving the accuracy of non-invasive ultrasound diagnosis is helpful to reduce the number of FNA, which is of clinical significance. Although ultrasound has made significant progress in the differential diagnosis of benign and malignant thyroid nodules, at present, conventional ultrasound (CUS) combined with color Doppler flow imaging (CDFI) still has a certain rate of missed diagnosis and misdiagnosis in determining the nature of thyroid nodules, and the diagnostic accuracy is not high, and there are also technical limitations in detecting small blood vessels and low blood flow (7, 8). As a new vascular imaging technology, Superb microvascular imaging (SMI) can visualize low-velocity blood flow (9, 10). Compared with CDFI, it can describe blood flow in more detail, and can obtain high-quality microvascular images without relying on contrast media, it can also describe the blood flow around the sinus and in the nodule in more detail (11, 12). In recent years, with the wide application of SMI technology in breast, liver, and other fields, more and more research teams have begun to pay attention to its diagnostic value for thyroid nodules (1318). Although there are many related studies comparing SMI with CDFI, there are still disputes on the comparison of diagnostic indicators such as sensitivity and specificity. Therefore, this study included relevant literature and applied Meta-analysis to comprehensively compare the diagnostic efficacy of the two methods in order to provide a more objective and comprehensive evidence-based basis for the diagnosis and clinical management of thyroid nodule disease.

2 Materials and methods

2.1 Search strategy

We searched PubMed, Embase, and Cochrane Library from inception through February 1, 2023. The following keywords and MeSH terms were used: [“thyroid cancer” or “thyroid tumor” or “thyroid nodule”] and [“superb microvascular imaging” or “SMI”] and [“color Doppler flow imaging” or “CDFI”] and [“ultrasound”]. We also performed a manual search to find other potential articles. Two investigators (LJ and DZ) searched online to obtain the original data, and the reference lists of all relevant articles were also scanned. All retrieved citations were exported to Zotero and checked for duplicates.

2.2 Inclusion and exclusion criteria

The inclusion criteria of this study were as follows:(1) Clinical cohort study or diagnostic test; (2) All thyroid nodules were histologically confirmed after ultrasonography; (3) All studies can directly or indirectly obtain original data such as true positive (TP), false positive (FP), false negative (FN), true negative values (TN). Exclusion criteria: (1) Abstracts, reviews, or case reports; (2) Repeated publication of data; (3) Incomplete original data.

2.3 Data extraction

Two evaluators (LJ and DZ) independently screened the literature according to the inclusion and exclusion criteria of the literature, and finally obtained two copies of data, and then cross-checked the data. If there were different opinions, discuss and negotiate together or ask a third party to help decide. Extract literature data from the data, including the first author, publication year, country, number of malignant lesions, the average age of patients, instrument system parameters, diagnostic gold standard, and research type. This study was conducted by the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines (19), the protocol was registered in the PROSPERO:(CRD42023402064).

2.4 Literature quality assessment

The diagnostic accuracy research quality evaluation tool (QUADAS-2) recommended by Cochrane was used to evaluate the quality of the included literature, and 14 items such as the population to be evaluated in the diagnostic test and the gold standard were scored (20). According to the specific content of each item, the answer options are divided into “Yes”, “No” and “Unclear”, and then the risk of bias and clinical applicability is comprehensively evaluated. Revman 5.4 software is used to make quality evaluation charts.

2.5 Statistical analysis

Meta-DiSc version 1.4 (Universidad Complutense, Madrid, Spain) software was used for meta-analysis. The sensitivity (Sen), specificity (Spe), positive likelihood ratio (LR+), and negative likelihood ratio (LR-) are calculated, and the threshold effect is evaluated using the 95% confidence interval (CI). The summary receiver operating characteristic (SROC) curve and the corresponding area under the curve was determined. Quality evaluation chart was created using Review Manager (RevMan, version 5.4, Cochrane IMS). The heterogeneity of the diagnostic odds ratio (DOR) of each study was analyzed by Meta-Disc 1.4 software. If I2>50% or P<0.05 among the included literature, there is a high heterogeneity between the results; I2<25% indicates that the heterogeneity between results is small, and 25% ≤ I2 ≤ 50% indicates that the heterogeneity of results is medium; P ≥ 0.05 indicates that there is no heterogeneity in the results. If there is heterogeneity, try to explore the source of heterogeneity by using Meta-regression. The sensitivity analysis was carried out by StataSE 12 (Stata Corporation, College Station, TX). P<0.05 indicated a statistically significant difference.

3 Results

3.1 Overall characteristics of selected studies

After the primary election, the abstracts and titles of 659 published articles (42 from PubMed, 602 from Embase, and 15 from Cochrane) were preliminarily reviewed. After eliminating duplicate results and abstract screening, the number of complete publications that may meet the criteria was reviewed, and finally, 13 articles were included (2133). Other 646 articles were excluded because of duplication, unrelated research, inappropriate data, review, no or inappropriate control group, other test methods or non-full text. Figure 1 shows a flowchart that reflects the search process and the reasons of exclusion.

FIGURE 1
www.frontiersin.org

Figure 1 Flowchart of the study selection process.

3.2 Characteristics of the included studies

The detailed characteristics of the included study are shown in Table 1, including the name of the first author, year of publication, the age range of patients, gender distribution (male/female), design, sample size, and equipment. All these articles were published from 2017 to 2022. The study was conducted in European countries (n=1), China (n=9), and Korea (n=3). The sample size is between 52 and 254. A total of 1712 nodules were included in these studies. After excluding invalid nodules, there were 897 benign nodules and 815 malignant nodules, respectively.

TABLE 1
www.frontiersin.org

Table 1 Summary of included study characteristics.

3.3 Methodologic quality of the included studies

Figure 2 illustrates the methodological quality of the included studies based on QUADAS-2. Figure 3 summarizes the risk of bias and adherence of individual studies to these items. Table 2 provides data on the other items of methodological quality. All included studies had a low risk of bias and were of high quality.

FIGURE 2
www.frontiersin.org

Figure 2 Percentage of included studies with the risk of bias.

FIGURE 3
www.frontiersin.org

Figure 3 The assessment of the risk of bias for included study. Quality is represented by colors using green (+) as yes (high quality), yellow (?) as unclear, and red (–) as no (low quality).

TABLE 2
www.frontiersin.org

Table 2 Summary of risk of bias for nine additional items assessing the methodologic quality of the included studies.

3.4 Threshold effects and heterogeneity

The Spearman correlation coefficients were 0.429 and 0.512 by heterogeneity analysis (P>0.05), indicating that there was no threshold effect. At the same time, the results showed that heterogeneity for sensitivity (I= 86.4%), specificity (I= 88.0%) in SMI group and sensitivity (I= 91.0%), specificity (I= 66.5%) in CDFI group. The included literature has high heterogeneity, so it is necessary to use random effect model to summarize and evaluate and draw SROC curve.

3.5 Sensitivity analysis

To observe the stability of the synthetic results, the data included in the literature were excluded one by one and the sensitivity and specificity were summarized again. It showed that the combined effect of various indicators changed little, indicating that the stability of the included literature was good and the reliability of the results was high (Figure 4).

FIGURE 4
www.frontiersin.org

Figure 4 Sensitivity analysis of studies. [(A) CDFI; (B) SMI].

3.6 Meta-regression analysis

As a result of the significant heterogeneity, meta-regression analysis was used to explore the source of heterogeneity. The covariates of the regression model are set as follows: (1) The sample size ≥ 100 is set as 1, and the sample size<100 is set as 0; (2) The age≥50 years old is set as 1, and the age<50 years old is set as 0; (3) The prospective study was set as 1, the retrospective study was set at 0; (4) The research object from China is set as 1, and that from other countries is set as 0; (5) The TI-RADS classification of nodules in the literature is not set as 1, and the TI-RADS classification of only 4 types of nodules in the literature is 0. Meta-regression analysis of the CDFI group and SMI group showed that there was no significant difference between the sources of heterogeneity and the covariates (P>0.05).

3.7 Diagnostic accuracy

A random effect model was used to analyze the combined effect quantity of the diagnostic four-grid data of CDFI and SMI included in the literature. The combined sensitivity of CDFI and SMI in the diagnosis of thyroid nodules was 0.62(95% CI: 0.57-0.67) and 0.80(95% CI: 0.77-0.83), respectively; The combined specificity was 0.81(95% CI: 0.78-0.85) and 0.79(95% CI: 0.77-0.82) respectively; The positive likelihood ratios were 3.33(95% CI: 2.18~5.07) and 4.37(95% CI: 3.0~6.36) respectively; The negative likelihood ratios were 0.41(95% CI: 0.27~0.64) and 0.23(95% CI: 0.15~0.35) respectively; The DOR was 8.93(95% CI: 3.96~20.16) and 22.29(95% CI: 12.18~40.78) respectively. The area under the SROC curve is 0.8498 and 0.8944 respectively. The diagnostic sensitivity, positive likelihood ratio, and DOR of SMI were higher than those of CDFI, while the negative likelihood ratio was lower than that of CDFI. The SROC curve of SMI diagnosis is closer to the upper left corner of the image, and its area under the SROC curve is larger, indicating that the diagnosis efficiency of SMI is better than that of CDFI (Figures 5, 6).

FIGURE 5
www.frontiersin.org

Figure 5 Estimates of CDFI assessment for the diagnosis of malignancy thyroid nodules. (A–E), Forest plots illustrate pooled estimates (diamonds) for sensitivity (A), specificity (B), positive likelihood ratio (LR) (C), negative LR (D), and diagnostic odds ratio (E) and corresponding 95% CIs for pooled estimates. (F), Summary receiver operating characteristic (SROC) plot for assessing accuracy with corresponding curves indicative of upper and lower bounds of 95% CI. AUC, area under curve; SE, standard error; Q* = summary measure of accuracy derived from the SROC curve.

FIGURE 6
www.frontiersin.org

Figure 6 Estimates of SMI assessment for the diagnosis of malignancy thyroid nodules. (A–E), Forest plots illustrate pooled estimates (diamonds) for sensitivity (A), specificity (B), positive likelihood ratio (LR) (C), negative LR (D), and diagnostic odds ratio (E) and corresponding 95% CIs for pooled estimates. (F), Summary receiver operating characteristic (SROC) plot for assessing accuracy with corresponding curves indicative of upper and lower bounds of 95% CI. AUC, area under curve; SE, standard error; Q* = summary measure of accuracy derived from the SROC curve.

3.8 Evaluation of publication bias and clinical applicability

The meta-analysis of the value of SMI and CDFI showed no significant asymmetry (P>0.05), that is, there was no significant publication bias (Figure 7). At the same time, it can be seen from the Fagan diagram that the post-test probability of SMI and CDFI is 83% and 78% respectively, which is 50% higher than the pre-test probability. The combined negative likelihood ratio of SMI and CDFI in the diagnosis of malignant thyroid nodules is more than 0.1, and the combined positive likelihood ratio is less than 10, indicating that both methods are effective in the diagnosis of malignant thyroid nodules (Figure 8).

FIGURE 7
www.frontiersin.org

Figure 7 Funnel diagram of SMI and CDFI. Panel (A) is the funnel diagram of CDFI; panel (B) is the funnel diagram of SMI.

FIGURE 8
www.frontiersin.org

Figure 8 Fagan diagram of SMI and CDFI. Panel (A) is Fagan diagram of CDFI; Panel (B) is the Fagan diagram of SMI.

4 Discussion

At present, the effective non-invasive diagnosis and differential diagnosis of benign and malignant thyroid nodules is an urgent problem to be solved. Although CUS can display the size, echo, boundary, calcification and blood flow of nodules, it has certain limitations in displaying the blood flow of lesions. SMI has unique advantages in displaying blood vessels. It can better display low-velocity blood flow without the use of contrast media. Diseases closely related to angiogenesis can be diagnosed at a relatively early stage through SMI. It is of great significance to evaluate the grading and treatment effect. At the same time, SMI can avoid adverse reactions related to CEUS contrast agents. However, SMI is at the stage of clinical exploration and has not been widely used. Specific clinical standards have not been established for the diagnosis of some diseases. There are still many disputes about the differential diagnosis of benign and malignant tumors. Therefore, this study uses the meta-analysis method to systematically evaluate and compare the value of CUS combined with SMI and CDFI in predicting thyroid malignant nodules, to provide an evidence-based basis for the differential diagnosis of thyroid malignant nodules.

Neovascularization and the growth of irregular vascular structures are the prominent characteristics of malignant tumors (34). In contrast-enhanced ultrasound (CEUS) examination, thyroid cancer below 1 cm usually shows low enhancement (35, 36). However, some researchers found that with the enlargement of the nodule, the blood flow inside the nodule is gradually enriched, 1-2 cm thyroid cancer often shows moderate to low enhancement, and>2 cm thyroid cancer often shows high enhancement, at the same time, benign nodules mainly have peripheral and mixed blood flow distribution patterns, and the central blood flow distribution pattern is a significant sign of malignant nodules (3740). Frates described the CDFI features that may help to distinguish benign and malignant nodules (8). Cappelli evaluated the number of blood vessels in thyroid benign and malignant nodules and showed that thyroid malignant nodules had rich blood flow signals, which confirmed that the number of blood vessels was an important parameter for distinguishing thyroid benign and malignant nodules (33). Kong showed that the specificity of SMI in predicting intra-nodular vessels was 91.2%, and intra-nodular vessels were an effective feature in predicting malignant tumors (28). Zhao performed SMI and CEUS examination on 102 thyroid nodules, the diagnostic value of SMI-BF of the periphery is comparable to CEUS hypoen-enhancement to differentiate thyroid cancer, analysis of SMI-BF as diagnostic of thyroid carcinoma specifically at the nodule periphery indicated rates of 82.2%, 79.3%, and 81.3% for sensitivity, specificity, and accuracy, the corresponding rates for CEUS hypoen-enhancement were 80.8%, 86.2%, and 82.3%, respectively (29). Compared with SMI, CEUS can display more blood flow information inside the nodule, but the price is relatively expensive. The detection rate of CDFI for low-speed blood flow is low, especially for low-speed blood flow<1cm/s, so the evaluation of small and low-speed blood flow has limitations. When the penetrating vessels were used as the characteristics of malignant nodules, the results showed that the sensitivity, specificity, accuracy, and other results of SMI diagnosis of thyroid nodules with a diameter of less than 1 cm were not statistically significant with those of CEUS, which showed that SMI and CEUS had similar diagnostic capabilities in the differential diagnosis of thyroid nodules.

This study used the diagnostic experimental evaluation tool QUADAS-2 to evaluate the quality of the included literature. The results showed that the overall quality of the included research was high and the risk of bias was low. However, the risk of bias is high in the selection of study subjects, which may be since some studies only include patients suspected of thyroid cancer, which leads to bias, and some studies do not report the mass spectrum composition of the study population, which leads to the fact that the included population cannot represent the clinically accepted population, which has an impact on the accuracy of the test. Therefore, the criteria for the inclusion of cases should be clarified and the process of inclusion of cases should be described in detail to improve the possibility of clinical application of diagnostic tests. In addition, more than half of the documents included in this study did not report the time interval of imaging examination and diagnostic criteria, which may also lead to the bias of results. The heterogeneity included in the study is large, but the heterogeneity test indicates that there is no threshold effect, and the heterogeneity may be mainly caused by the non-threshold effect. This may be caused by the characteristics of the patients included in each study (nodule classification grade, nodule diameter size), the technical level of the operator, the type of research design, and the use of different diagnostic criteria. Meta-regression analysis shows that the relationship between the heterogeneity sources of CDFI and SMI and the covariates is not statistically significant, and it is speculated that the operator dependence of the relevant measurement will also bring some deviation. The heterogeneity of this study is not caused by the threshold effect, so a random model is used for summary analysis. In addition, this study is a diagnostic meta-analysis, so there is inevitable clinical heterogeneity.

To evaluate the ability of SMI and CDFI in the differential diagnosis of thyroid nodules, a total of 13 articles were included in this meta-analysis, including 1712 nodules. The combined sensitivity of SMI in the diagnosis of thyroid malignant nodules was 80%, the specificity was 79%, and the misdiagnosis rate was 20%, which was relatively high. The positive likelihood ratio was 4.37, the negative likelihood ratio was 0.23, the diagnostic odds ratio was 22.29, and the area under the SROC curve (AUC) was 0.8944. The combined sensitivity and specificity of CDFI in the diagnosis of thyroid malignant nodules are 62% and 81%, the positive likelihood ratio was 3.33, the negative likelihood ratio was 0.41, the diagnostic odds ratio was 8.93, and the area under the SROC curve (AUC) was 0.8498. SROC curve is a comprehensive index that directly observes the accuracy of diagnostic tests and reflects the sensitivity and specificity of diagnostic tests and target disease. AUC value is an important test accuracy index. The closer it is to 1, the better the diagnostic efficiency of this diagnostic method is. The results of this meta-analysis showed that the combined sensitivity, and ratio of SMI were higher than those of CDFI, and the area under the SROC curve was also larger than that of CDFI, indicating that SMI was more effective than CDFI in diagnosis.

This study has limitations. First, most of the included documents are from China, which may lead to potential regional bias. Second, not all the literature is prospective, and retrospective literature may improve the accuracy of diagnosis. Thirdly, due to the insufficient number of documents with nodule diameter<1cm included in the study, the ability of SMI to differentiate and diagnose thyroid micro-nodules could not be evaluated separately, and more research reports on SMI to differentiate and diagnose thyroid micro-nodules are needed. Fourth, the two methods of examination and diagnosis are not quantified, and there is no unified diagnostic standard. Fifth, although diagnostic meta-analysis has inevitable clinical heterogeneity, reducing heterogeneity is conducive to improving the accuracy of meta-analysis. Because there are relatively few reports on the diagnostic performance of SMI in differentiating benign and malignant thyroid nodules, a large-scale multi-center study is still needed.

In conclusion, SMI is superior to CDFI in the diagnosis of thyroid malignant nodules, and SMI technology can reflect the distribution of blood flow, avoid the overflow of blood flow signals, and can sensitively detect low-velocity blood flow and micro vessels to make up for the deficiency of CDFI, with good clinical application value.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author.

Author contributions

LJ and DZ: Software, Writing - original draft. Y-NC: Writing – review and editing. X-JY: Visualization, Investigation. M-FP: Supervision. LL: Conceptualization, Methodology, Supervision. All authors contributed to the article and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Durante C, Grani G, Lamartina L, Filetti S, Mandel SJ, Cooper DS. The diagnosis and management of thyroid nodules: a review. JAMA (2018) 319(9):914–24. doi: 10.1001/jama.2018.0898

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Gharib H, Papini E, Paschke R, Duick DS, Valcavi R, Hegedüs L, et al. American Association of clinical endocrinologists, associazione Medici endocrinologi, and European thyroid association medical guidelines for clinical practice for the diagnosis and management of thyroid nodules: executive summary of recommendations. J Endocrinol Invest (2010) 33(5):287–91. doi: 10.1007/BF03346587

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Sorensen JR, Johansen J, Gano L, Sørensen JA, Larsen SR, Andersen PB, et al. A "package solution" fast track program can reduce the diagnostic waiting time in head and neck cancer. Eur Arch Otorhinolaryngol (2014) 271(5):1163–70. doi: 10.1007/s00405-013-2584-z

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Lyu YJ, Shen F, Yan Y, Situ MZ, Wu WZ, Jiang GQ, et al. Ultrasound-guided fine-needle aspiration biopsy of thyroid nodules <10 mm in the maximum diameter: does size matter? Cancer Manag Res (2019) 11:1231–6. doi: 10.2147/CMAR.S189358

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Egset AV, Holm C, Larsen SR, Nielsen SH, Bach J, Helweg-Larsen JP, et al. Risk of malignancy in fine-needle aspiration biopsy in patients with thyroid nodules. Dan Med J (2017) 64(2):A5320.

PubMed Abstract | Google Scholar

6. Todsen T, Bennedbaek FN, Kiss K, Hegedüs L. Ultrasound-guided fine-needle aspiration biopsy of thyroid nodules. Head Neck (2021) 43(3):1009–13. doi: 10.1002/hed.26598

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Foster FS, Burns PN, Simpson DH, Wilson SR, Christopher DA, Goertz DE. Ultrasound for the visualization and quantification of tumor microcirculation. Cancer Metastasis Rev (2000) 19(1-2):131–8. doi: 10.1023/a:1026541510549

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Frates MC, Benson CB, Doubilet PM, Cibas ES, Marqusee E. Can color Doppler sonography aid in the prediction of malignancy of thyroid nodules? J Ultrasound Med (2003) 22(2):127–31;quiz 132-4. doi: 10.7863/jum.2003.22.2.127

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Ma Y, Li G, Li J, Ren WD. The diagnostic value of superb microvascular imaging (SMI) in detecting blood flow signals of breast lesions: a preliminary study comparing SMI to color Doppler flow imaging. Med (Baltimore) (2015) 94(36):e1502. doi: 10.1097/MD.0000000000001502

CrossRef Full Text | Google Scholar

10. Machado P, Segal S, Lyshchik A, Forsberg F. A novel microvascular flow technique: initial results in thyroids. Ultrasound Q (2016) 32(1):67–74. doi: 10.1097/RUQ.0000000000000156

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Mao Y, Mu J, Zhao J, Zhao L, Xin X. The value of superb microvascular imaging in differentiating benign renal mass from malignant renal tumor: a retrospective study. Br J Radiol (2018) 91(1082):20170601. doi: 10.1259/bjr.20170601

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Lu R, Meng Y, Zhang Y, Zhao W, Wang X, Jin M, et al. Superb microvascular imaging (SMI) compared with conventional ultrasound for evaluating thyroid nodules. BMC Med Imaging (2017) 17(1):65. doi: 10.1186/s12880-017-0241-5

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Xiao XY, Chen X, Guan XF, Wu H, Qin W, Luo BM. Superb microvascular imaging in diagnosis of breast lesions: a comparative study with contrast-enhanced ultrasonographic microvascular imaging. Br J Radiol (2016) 89(1066):20160546. doi: 10.1259/bjr.20160546

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Dubinsky TJ, Revels J, Wang S, Toia G, Sonneborn R, Hippe DS, et al. Comparison of superb microvascular imaging with color flow and power Doppler imaging of small hepatocellular carcinomas. J Ultrasound Med (2018) 37(12):2915–24. doi: 10.1002/jum.14654

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Oura K, Kato T, Ohba H, Terayama Y. Evaluation of intraplaque neovascularization using superb microvascular imaging and contrast-enhanced ultrasonography. J Stroke Cerebrovasc Dis (2018) 27(9):2348–53. doi: 10.1016/j.jstrokecerebrovasdis.2018.04.023

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Yokota K, Tsuzuki Wada T, Akiyama Y, Mimura T. Detection of synovial inflammation in rheumatic diseases using superb microvascular imaging: comparison with conventional power Doppler imaging. Mod Rheumatol (2018) 28(2):327–33. doi: 10.1080/14397595.2017.1337288

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Chen J, Chen L, Wu L, Wang R, Liu JB, Hu B, et al. Value of superb microvascular imaging ultrasonography in the diagnosis of carpal tunnel syndrome: compared with color Doppler and power Doppler. Med (Baltimore) (2017) 96(21):e6862. doi: 10.1097/MD.0000000000006862

CrossRef Full Text | Google Scholar

18. Fischerova D, Cibula D, Stenhova H, Vondrichova H, Calda P, Zikan M, et al. Transrectal ultrasound and magnetic resonance imaging in staging of early cervical cancer. Int J Gynecol Canc (2008) 18(4):766–72. doi: 10.1111/j.1525-1438.2007.01072.x

CrossRef Full Text | Google Scholar

19. Knobloch K, Yoon U, Vogt PM. Preferred reporting items for systematic reviews and meta-analyses (PRISMA) statement and publication bias. J Craniomaxillofac Surg (2011) 39(2):91–2. doi: 10.1016/j.jcms.2010.11.001

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med (2011) 155(8):529–36. doi: 10.7326/0003-4819-155-8-201110180-00009

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Zhu YC, Zhang Y, Deng SH, Jiang Q. A prospective study to compare superb microvascular imaging with grayscale ultrasound and color Doppler flow imaging of vascular distribution and morphology in thyroid nodules. Med Sci Monit (2018) 24:9223–31. doi: 10.12659/MSM.911695

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Zhu YC, Zhang Y, Shan J, Deng SH, Shi XR, Jiang Q. Added value of superb microvascular imaging and virtual touch imaging quantification in assisting thyroid cancer classification. Ultrasound Med Biol (2021) 47(12):3364–71. doi: 10.1016/j.ultrasmedbio.2021.07.017

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Yoon JH, Kim EK, Kwak JY, Park VY, Moon HJ. Application of various additional imaging techniques for thyroid ultrasound: direct comparison of combined various elastography and Doppler parameters to Gray-scale ultrasound in differential diagnosis of thyroid nodules. Ultrasound Med Biol (2018) 44(8):1679–86. doi: 10.1016/j.ultrasmedbio.2018.04.006

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Xue SW, Luo YK, Jiao ZY, Xu L. Clinical value of SMI combined with low-dose CT scanning in differential diagnosis of thyroid lesions and tumor staging. Pak J Med Sci (2021) 37(5):1347–52. doi: 10.12669/pjms.37.5.4144

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Pei S, Cong S, Zhang B, Liang C, Zhang L, Liu J, et al. Diagnostic value of multimodal ultrasound imaging in differentiating benign and malignant TI-RADS category 4 nodules. Int J Clin Oncol (2019) 24(6):632–9. doi: 10.1007/s10147-019-01397-y

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Ahn HS, Lee JB, Seo M, Park SH, Choi BI. Distinguishing benign from malignant thyroid nodules using thyroid ultrasonography: utility of adding superb microvascular imaging and elastography. Radiol Med (2018) 123(4):260–70. doi: 10.1007/s11547-017-0839-2

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Hong MJ, Ahn HS, Ha SM, Park HJ, Oh J. Quantitative analysis of vascularity for thyroid nodules on ultrasound using superb microvascular imaging: can nodular vascularity differentiate between malignant and benign thyroid nodules? Med (Baltimore) (2022) 101(5):e28725. doi: 10.1097/MD.0000000000028725

CrossRef Full Text | Google Scholar

28. Kong J, Li JC, Wang HY, Wang YH, Zhao RN, Zhang Y, et al. Role of superb micro-vascular imaging in the preoperative evaluation of thyroid nodules: comparison with power Doppler flow imaging. J Ultrasound Med (2017) 36(7):1329–37. doi: 10.7863/ultra.16.07004

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Yongfeng Z, Ping Z, Hong P, Wengang L, Yan Z. Superb microvascular imaging compared with contrast-enhanced ultrasound to assess microvessels in thyroid nodules. J Med Ultrason (2001) (2020) 47(2):287–97. doi: 10.1007/s10396-020-01011-z

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Zhang L, Gu J, Zhao Y, Zhu M, Wei J, Zhang B. The role of multimodal ultrasonic flow imaging in thyroid imaging reporting and data system (TI-RADS) 4 nodules. Gland Surg (2020) 9(5):1469–77. doi: 10.21037/gs-20-641

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Zhang L, Zhang B, Cao J, Pu H. Value of superb micro-vascular imaging in TI-RADS 4 thyroid nodules. Chin J Ultrasonogr (2017) 26(12):1029–33. doi: 10.3760/cma.j.issn.1004-4477.2017.12.004

CrossRef Full Text | Google Scholar

32. Li Y, Wen D, Li C, Li X, Xue G. The role of ATA (2015) guidelines, superb microvascular imaging, and spectral Doppler in differentiation between malignant and benign thyroid nodules. J Clin Otorhinolaryngol Head Neck Surg (2017) 31(15):1152–6. doi: 10.13201/j.issn.1001-1781.2017.15.003

CrossRef Full Text | Google Scholar

33. Cappelli C, Pirola I, Gandossi E, Marini F, Cristiano A, Casella C, et al. Ultrasound microvascular blood flow evaluation: a new tool for the management of thyroid nodule? Int J Endocrinol (2019) 2019:7874890. doi: 10.1155/2019/7874890

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Jászai J, Schmidt MHH. Trends and challenges in tumor anti-angiogenic therapies. Cells. (2019) 8(9):1102. doi: 10.3390/cells8091102

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Jiang J, Shang X, Zhang H, Ma W, Xu Y, Zhou Q, et al. Correlation between maximum intensity and microvessel density for differentiation of malignant from benign thyroid nodules on contrast-enhanced sonography. J Ultrasound Med (2014) 33(7):1257–63. doi: 10.7863/ultra.33.7.1257

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Zhou Q, Jiang J, Shang X, Zhang HL, Ma WQ, Xu YB, et al. Correlation of contrast-enhanced ultrasonographic features with microvessel density in papillary thyroid carcinomas. Asian Pac J Cancer Prev (2014) 15(17):7449–52. doi: 10.7314/apjcp.2014.15.17.7449

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Phuttharak W, Somboonporn C, Hongdomnern G. Diagnostic performance of gray-scale versus combined gray-scale with colour doppler ultrasonography in the diagnosis of malignancy in thyroid nodules. Asian Pac J Cancer Prev (2009) 10(5):759–64.

PubMed Abstract | Google Scholar

38. Hornung M, Jung EM, Georgieva M, Schlitt HJ, Stroszczynski C, Agha A. Detection of microvascularization of thyroid carcinomas using linear high-resolution contrast-enhanced ultrasonography (CEUS). Clin Hemorheol Microcirc (2012) 52(2-4):197–203. doi: 10.3233/CH-2012-1597

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Kim DW, Jung SJ, Eom JW, Kang T. Color Doppler features of solid, round, isoechoic thyroid nodules without malignant sonographic features: a prospective cytopathological study. Thyroid. (2013) 23(4):472–6. doi: 10.1089/thy.2012.0238

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Kim DW, In HS, Choo HJ, Jung SJ, Kim YW. Solid and isoechoic thyroid nodules without malignant sonographic features: comparison of malignancy rate according to nodule size, shape and color Doppler pattern. Ultrasound Med Biol (2013) 39(2):269–74. doi: 10.1016/j.ultrasmedbio.2012.09.018

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: malignant, thyroid nodules, ultrasonography, Superb microvascular imaging, color Doppler flow imaging, meta-analysis

Citation: Jiang L, Zhang D, Chen Y-N, Yu X-J, Pan M-F and Lian L (2023) The value of conventional ultrasound combined with superb microvascular imaging and color Doppler flow imaging in the diagnosis of thyroid malignant nodules: a systematic review and meta-analysis. Front. Endocrinol. 14:1182259. doi: 10.3389/fendo.2023.1182259

Received: 08 March 2023; Accepted: 30 May 2023;
Published: 21 June 2023.

Edited by:

Joseph V. Martin, Rutgers University Camden, United States

Reviewed by:

Natalia D. Gladkova, Privolzhsky Research Medical University (PIMU), Russia
Xiao-Yan Zhang, Chinese Academy of Medical Sciences and Peking Union Medical College Hospital, China

Copyright © 2023 Jiang, Zhang, Chen, Yu, Pan and Lian. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Lian Lian, dr_lianlian@163.com

These authors have contributed equally to this work and share first authorship

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.