Introduction: Pregestational diabetes (PGDM) is an increasingly common and complex condition that infers risk to both mother and infant. To prevent serious morbidity, strict glycaemic control is essential. The aim of this review is to review the glucose sensing and insulin delivering technologies currently available for women with PGDM.
Methods: We reviewed online databases for articles relating to technology use in pregnancy using a combination of keywords and MeSH headings. Relevant articles are included below.
Results: A number of technological advancements have improved care and outcomes for women with PGDM. Real time continuous glucose monitoring (rtCGM) offers clear advantages in terms of infants size and neonatal intensive care unit admissions; and further benefits are seen when combined with continuous subcutaneous insulin delivery (insulin pump) and algorithms which continuously adjust insulin levels to glucose targets (hybrid closed loop). Other advancements including flash or intermittent scanning CGM (isCGM) and stand-alone insulin pumps do not confer as many advantages for women and their infants, however they are increasingly used outside of pregnancy and many women enter pregnancy already using these devices.
Discussion: This article offers a discussion of the most commonly used technologies in pregnancy and evaluates their current and future roles.
1 Introduction
Pregestational diabetes mellitus (PGDM) is a combination of complex, chronic conditions defined by hyperglycaemia and associated with adverse fetal and maternal complications. PGDM includes any form of diabetes which exists before the conception. The most common forms of PGDM are undoubtedly type 1 and type 2 diabetes, however other forms of diabetes including latent autoimmunity diabetes of adulthood (LADA), maturity onset diabetes of the young (MODY), cystic fibrosis related diabetes (CFRD) and diabetes related to endocrinopathies and medications can occur in women of childbearing age and can cause complications during pregnancy.
Type 1 and type 2 diabetes complicate between 0.5-2.4% of all pregnancies worldwide (1). While there is variation in prevalence across different regions, there has been a universal increase in the number of pregnancies complicated by PGDM and its rate has more doubled since 1995 (1, 2). This substantial increase is likely multifactorial. Firstly more women are entering pregnancy with type 2 diabetes and women with type 2 diabetes now account for between 30-50% of cases of PGDM (3, 4). Rates of type 2 diabetes have increased by roughly 30% in recent years (5) – most likely due to the increase in obesity in adolescents and young adults (6). Other factors contributing to the rise of type 2 diabetes include urbanisation, environmental factors like pollution and increased testing and detection (7). The incidence of type 1 diabetes, which is similarly rising by roughly 1.9% per year is less well understood (8).
The prompt recognition and treatment of PGDM is important due to both the short and long term complications faced by both mother and infant. During pregnancy, complications such as pre-eclampsia and Caesarean delivery are three times more common than the non-diabetic population (9) and half of women with PGDM will have at least one hospitalisation during their pregnancy (3). Infants are at risk of preterm delivery (OR 3.48); macrosomia (OR 1.51); being born large for gestational age (LGA) (OR 3.9) and have a 2-3.5 fold increased risk of neonatal death and stillbirth (10). Much of this risk of perinatal mortality comes from the increased risk of congenital anomalies seen in the infants of diabetic mothers (9). In the long term, infants of women with diabetes are more likely to be overweight and obese in childhood and display evidence of significant insulin resistance (even when adjusted for confounders like family history) (11). Similarly rates of cardiovascular disease are 29% higher in infants exposed to PGDM and these infants have a higher rate of hypertensive disorders and venous thromboembolism (10). The risk of complications in the offspring also seems to be associated with the number of diabetes related complications in the mother (12). More recently the risk of autism, attention-deficit hyperactivity disorder (ADHD) and other neurocognitive disorders have been increasingly recognised. A recent meta-analysis identified hazard ratios of 1.36 (95% CI 1.19-1.55) and 1.98 (95% CI 1.46-2.88) for ADHD and autism respectively (13). This risk appears to correlate with the degree of fetal exposure to hyperglycaemia as neurocognitive disorders are more common in type 1diabetes, and it is thought to be directly related to the effect of hyperglycaemia on the developing brain and neural pathways (14). Similarly insulin use in GDM is associated with neonatal hypoglycaemia and worse neonatal neural adaptability (15).
The serious and diverse range of complications faced by this cohort make the long term follow up of infants exposed to diabetes all the more important (16).
Strict glycaemic control is key to the prevention of many of these complications and the cornerstone of management. International guidelines like the American Diabetes Association (ADA) and the National Institute for Health & Care Excellence (NICE) recommend that women with PGDM who are planning a pregnancy aim for a HbA1c level below 48 mmol/mol (6.5%), if attainable without causing significant hypoglycaemia, to reduce the risk of complications associated with elevated glucose (17, 18). In pregnancy, the ADA recommends a HbA1c target of < 42mmol/mol (<6%) but highlights that this target can be relaxed to < 53mmol/mol (<7%) to prevent hypoglycaemia. These targets were selected as HbA1c levels above this level have been shown to be associated with an increase in adverse outcomes. A peri-conception HbA1c above 49mmol/mol (6.6%) [adjusted odds ratio, aOR=1.02 (95% CI: 1.00 - 1.04)], pre-pregnancy retinopathy [aOR=2.05 (95% CI: 1.04 - 4.05)] and lack of pre-pregnancy folic acid consumption [aOR=2.52 (95% CI: 1.12 - 5.65)] were all independently associated with increased odds of fetal and infant death (19). Observational population based studies have also demonstrated the increased risk of congenital anomaly that correlates directly with poor glycaemic control (20). Interventions including pre-pregnancy care (PPC) have successfully improved maternal and fetal outcomes and are cost-effective (21). A systematic review and meta-analysis of observational studies evaluating the effectiveness of PPC in improving maternal and perinatal outcomes suggested that PPC is associated with a reduction in first trimester HbA1c of 1.27% (22)- it is however worth noting that this was in high income countries with homogenous populations. The meta-analysis results showed that attendance at PPC reduced congenital malformation risk by 71%, [RR=0.29 (95% CI: 0.21–0.40)]. It also resulted in a reduction in the risk of preterm delivery by 15%, [RR=0.85 (95% CI: 0.73–0.99)] and a risk reduction of perinatal mortality by 54%, [RR=0.46 (95% CI: 0.30–0.73)]. Results of these and other studies which have demonstrated the importance of good pre- and ante-natal glycaemic control highlighted the need for better treatment options and changes in care for women with PGDM.
As for non-pregnant adults, advances in glucose sensing and insulin delivery technology offer potential improvements in care. From the use of rudimentary and cumbersome insulin pumps in the 1960 (23–27), to sophisticated, advanced hybrid closed loop insulin delivery systems (28), the treatment of diabetes has dramatically changed since the discovery of insulin in 1921 (29). A summary and timeline of the evolution of diabetes treatment for non-pregnant adults can be found in Table 1.
1.1 Aims
The aim of this article is to summarise the advancements made in the area of diabetes in pregnancy and to highlight the corresponding improvements seen in pregnancy outcomes.
2 Technologies
2.1 Glucose sensing technology
2.1.1 Real time continuous glucose monitoring
Real time continuous glucose monitoring (rtCGM) offers hundreds of real-time interstitial glucose readings per day and allows the wearer to enable alerts for hypo- and hyperglycaemia. This allows the patient to observe daily patterns in glucose levels to improve long term glycaemic control and to make treatment decisions to avoid hypo- and hyperglycaemia. Through this mechanism rtCGM has been shown to be effective in reducing hypoglycaemia and improving glycaemic control and quality of life for patients with type 1 diabetes (46). In certain countries rtCGM is now commonly used before and throughout pregnancy and has been studied during labour and delivery (47).
A summary of the commonly used terminology in rtCGM (specific to pregnancy) is shown in Box 1.
Box 1. Commonly used terms in diabetes technology (48).
Time in range (TIR): number of minutes/hours per day spent between 3.5 and 7.8 mmol/L. Patients with diabetes in pregnancy should aim to spend >70% (>16 hours, 48 mins) per day in this range. For non-pregnant adults this range is 4-10 mmol/L.
Time below range (TBR): number of minutes per day spent <3.5 mmol/L. Patients with diabetes in pregnancy should aim to spend <4% (<1 hour) per day in this range; and should aim to spend <1% (<15 mins) per day <3mmol/L.
Time above range (TAR): number of minutes per day spent >7.8 mmol/L. Patients with diabetes in pregnancy should aim to spend <25% (<6 hours) per day in this range.
Although rtCGM has been studied in diabetes since the 1970s and studied in diabetes in pregnancy since at least 2008 (49), the seminal trial in this field was published in 2017. The Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT) study was a landmark randomized controlled trial in diabetes in pregnancy (50). In this study, 325 women with type 1 diabetes who were pregnant or planning pregnancy were randomized to either rtCGM or self-monitoring of blood glucose (SMBG). Although no improvement was seen in hypoglycemia, a reduction in hyperglycemia in the rtCGM group resulted in a greater number of hours spent in range between 3.5-7.8 mmol/L (TIR). This resulted in improved fetal outcomes including a reduction in LGA births (odds ratio 0·51, 95% CI 0·28 to 0·90; p=0·0210), a reduction in neonatal hypoglycemia (odds ratio 0·45; 0·22 to 0·89) and a reduction in time spent in neonatal intensive care units (NICU). Further post-hoc analyses of the CONCEPTT study demonstrated:
a) even slight increases in the number of minutes per day spent in the target range in trimesters two and three can decrease the risk of neonatal hypoglycemia (51)
b) hyperglycemia in the early morning and late afternoon in trimester two and three respectively can increase rate of LGA (52) – underlining the importance of analyzing patterns and adjusting insulin regimes and
c) Post-prandial rises are more pronounced in those with LGA infants, again demonstrating the importance of appropriate and timely insulin bolus adjustments (53).
Similar observational studies have demonstrated that each additional 8.5% TIR in pregnancy correlates to a 1% or an 11 mmol/mol decrease in HbA1c (54).
These improvements in fetal outcomes prompted assessments of potential cost benefits and cost-analyses which demonstrated favorable results. The degree of cost-effectiveness differs internationally as the cost associated with NICU varies (55). For example in one Canadian study rtCGM was cost effective when paid for by the individual patient, and cost neutral when paid for by the healthcare provider or government (in this instance, the patient pays for the initial cost outlay and the government does not have to incorporate the cost of purchasing the rtCGM into their budget). In UK based cost-analysis, rtCGM was calculated to be very cost efficient (56). It is important to note that this study included the use of real time CGM rather than retrospective CGM. Retrospective CGM only allows retrospective evaluation and review of glucose readings and does not allow the patient to make changes to their insulin dose at the time of hyperglycemia. Retrospective CGM is very helpful in evaluating daily trends in hyperglycemia, however it has shown conflicting results in terms of fetal macrosomia (57).
In women with type 2 diabetes, studies of rtCGM have been small and findings are inconsistent. One study which enrolled 46 women with type 1 and 25 women with type 2 diabetes found an improvement in birth weight and macrosomia in the rtCGM group, although an exact breakdown was not given (49). A larger Danish study which enrolled 123 women with type 1 and 31 women with type 2 diabetes found no difference between rtCGM and SMBG users (58).
From a patient perspective, rtCGM appears to be generally well tolerated. The main drawbacks of rtCGM include sleep disturbance, discomfort/adhesive issues at the insertion site and false readings of hypoglycemia related to positioning (compression hypoglycemia) (59). Despite this more than 80% would recommend rtCGM and a similar proportion would use it again (60). This increased use and popularity of rtCGM also carries implications for healthcare providers (61). The analysis of rtCGM is resource intense and requires dedicated time with each patient. Patients derive the most benefit from rtCGM when it is reviewed by a clinician and adjustments can be made (62). As such the effective use of rtCGM relies on appropriately trained staff and regular patient contact.
Lastly, although rtCGM unequivocally confers benefits to women with type 1 diabetes, due to its cost it is not universally available and the inequity of available healthcare remains a major issue for people living with diabetes worldwide (63).
2.1.2 Intermittent scanning CGM
Intermittently scanned CGM (isCGM) or flash CGM provides immediate information about a patient’s current and predicted interstitial glucose. It differs from rtCTG as patients have to scan a wearable device with a device reader or use an app on their mobile phone (whereas rtCGM provides constant information without scanning) (64). Unlike rtCGM, first generation isCGMs do not have alarms and do not need to be calibrated against SMBG.
The glucose targets for isCGM are the same as for rtCGM and isCGM devices have also been studied and approved for use in pregnancy.
One measurement tool which is being increasingly used in discussions about real time and intermittent scanning CGM is the mean absolute relative difference (MARD). MARD is a single number which represents the accuracy of the glucose monitor. It is calculated using the difference between the CGM readings and the values measured at the same time by the reference measurement system eg central laboratory level (65).
When compared to SMBG in pregnancy, the MARD between isCGM and SMBG is 11%, and this same study found high rates of patient satisfaction (66). It is worth noting that this was a single study of 74 pregnancy women, however the MARD range is similar to that quoted by the manufactures and real life data (67, 68).
isCGM also has a number of benefits over SMBG including less hypoglycaemia and greater TIR (69, 70). Despite these advantages, discrepancies exist between isCGM and SMBG. In one study isCGM under-estimated glucose levels and potentially lead to differences in treatment decisions in up to 30% of instances (71). Finally improved fetal outcomes have not been demonstrated when compared to SMBG (69). In larger observational studies of over 300 women with type 1 diabetes, isCGM resulted in better glycaemic control in trimester 2 compared to conventional SMBG - however this translated to higher rates of neonatal hypoglycaemia and no improvement in rates of LGA or prematurity (72). The authors have suggested that the tendency of isCGM to under-read glucose readings could make patients more likely to overtreat hypoglycaemia, and thus expose themselves to rebound hyperglycaemia. It may also lead both patients and clinicians to be more cautious about insulin adjustments.
When compared to rtCGM, isCGM has some disadvantages. In one study of 20 pregnant women with type 1 diabetes, isCGM users reported a great TBR overnight and similar results have been found in more recent studies (73, 74). Although there was no significant difference in maternal or fetal outcomes such studies raise the question of using isCGM for treatment decisions, especially overnight. Other studies which evaluated clinical outcomes using older rtCGM and isCGM found better glycaemic control in the first trimester in rtCGM users. Despite better control in this critical period of organogenesis, the clinical outcomes were comparable and patients in both groups improved their glycaemic control (75).
The culmination of this evidence has resulted in the recommendation to offer isCGM to all pregnant women with PGDM who are unable or unwilling to use rtCGM (76).
2.2 Insulin delivery technology
2.2.1 Continuous subcutaneous insulin infusion
Despite the clear benefits of CSII outside of pregnancy, in pregnancies complicated by type 1 diabetes the use of CSII has been associated with some disadvantages and conflicting evidence. While initial studies showed benefit in neonatal hypoglycaemia, caesarean delivery, preterm delivery and the Apgar score at five minutes (77), other results have been disappointing. Some studies have failed to demonstrate glycaemic benefits and have found that CSII users require more support and staff resources (78). Others have shown better glycaemic control and less hypoglycaemia, but this has not translated to improved fetal outcomes (79–82). One meta-analysis of four randomised and 43 observational studies found that although glycaemic control and insulin requirements were better in CSII versus multiple daily injections (MDI), rates of gestational weight gain, large and small for gestational age (LGA, SGA) and second and third trimester glycaemic control were inferior in CSII users (83). An older systematic review of RCTs found no differences between CSII and MDI (84). Other studies similarly found higher rates of gestational weight gain and LGA in CSII users despite better glycaemic control (85).
A pre-specified analysis of the CONCEPTT study found that MDI users had better glycaemic control at 24 and 34 weeks, had less hypertension and less neonatal hypoglycaemia and NICU admissions (86). A similar study which also evaluated CSII and MDI in rtCGM users found no significant difference in outcomes (87).
It is worth noting that much of this evidence comes from observational studies and the lack of RCTs in this area mean that results must be interpreted in this light.
CSII have also been studied at the time of delivery and women who consistently use CSII throughout pregnancy and delivery have better TIR than those who switch to an intravenous insulin infusion in labour. It should be noted however that the decision to use CSII during labour is multifactorial and depends on staff familiarity with different diabetes technology and the ability to frequently and reliably check for ketones if needed (88).
2.3 Combination of rtCGM/CSII
Some patients require and benefit from a combination of both CSII and rtCGM. When used in combination this is termed sensor augmented pump (SAP). In some SAP the rtCGM simply provides information to the wearer, allowing them to make manual adjustments to their CSII. More recent SAPs have low glucose suspend or predictive low glucose suspend (PLGS) features, which will pause insulin delivery when glucose levels begin to fall below a certain level. The most recent advance in SAP is the use of a hybrid closed loop system, which can automatically make adjustments to the wearer’s basal insulin in the presence of both hypo- and hyperglycaemia.
A definition of the different types of SAP can be found in Box 2.
Box 2. Definitions of SAP (73).
1st generation SAP - the user has to make basal rate adjustments manually.
2nd generation SAP – “the insulin dosing software and the rtCGM values are coupled which allows for automated suspension of basal insulin delivery in response to a predicted or detected low glucose level”.
Hybrid closed loop – maintain glucose levels within a target range through the use of a computerized algorithm to adjust the basal rate of insulin and administer corrective bolus doses (74).
Sensor augmented and closed loop technology have been studied in pregnancy since the early 2000’s and have shown some benefits in selected groups of patients.
Sensor augmented pump therapy with low glucose suspend has been shown to reduce HbA1c levels without increasing hypoglycaemia compared to standard treatment in observational studies (91). When compared to CSII without rtCGM, individuals using senor augmented pumps with low glucose suspend features had better third trimester glycaemic control, although no other differences, namely in macrosomia or pre-term birth were noted (92).
Reassuringly, these changes do not correlate with any increased risk in diabetic ketoacidosis (DKA) (93).
The initial studies of closed loop insulin delivery (where basal insulin levels are automatically adjusted to bring glucose levels within a target range) were studied in small numbers of women with type 1 diabetes over very short periods of time. These early studies demonstrated the safety of closed loop systems for short term use and paved the way for larger studies which demonstrated improved safety and less nocturnal hypoglycaemia (94, 95). In more advanced closed loop studies, TIR approached 69% in studies which evaluated pregnancy, labour and delivery. TIR approached 80% during labour itself and closed loop technology offered less hypoglycaemia than standard sensor augmented pump (96–98).
Due to the tight glycaemic control required in pregnancy, currently only one commercially available closed loop system is approved for use in pregnancy – CamsAPS FX (99). This closed loop system allows the user to personalise their glucose target, allowing them to achieve tighter control. Other commercially available closed loop systems do not allow the patient to reduce the target glucose to the levels required in pregnancy, but are often used either off-label or as SAP during pregnancy (100, 101). There are a number of ongoing randomised controlled trials aiming to evaluate commercial closed loop technology in pregnancy (102), NCT03774186, NCT04902378, NCT04520971, NCT04938557.
From a patient-reported outcome perspective, both benefits and burdens of closed-loop systems were described. Women reported having a sense of peace of mind and trust in the ease of use, however others described frustration with technical issues and being attached to diabetes related devices on a constant basis (103). While these considerations are very important they are not unique to pregnancy and commonly affect non-pregnant adults with type 1 diabetes (104).
2.4 Other
2.4.1 Telemedicine
Telemedicine or telehealth is defined as “technology-based virtual platform to deliver various aspects of health information, prevention, monitoring, and medical care” (105). It was already increasingly used prior to the Sars-CoV-2 pandemic however its use soared during this time due to universal lockdown restrictions (106). Telemedicine offers increased convenience and flexibility and is generally acceptable to patients however to be beneficial in PGDM hard outcomes like glycaemic control and fetal outcomes need to be evaluated (107). A systematic review of the use of telemedicine in treating both PGDM and gestational diabetes (GDM) did not find any benefit in maternal or fetal outcomes, although very few studies evaluated women with PGDM (108). In studies that did include small numbers of patients with PGDM, telemedicine resulted in fewer GP and nurse visits (109), improved satisfaction and quality of life (110) and some slight improvements in glycaemic control (111).
Further evaluation is needed before telemedicine can become a routine part of care for pregnant women with PGDM.
2.4.2 Smart pens
Smart pens are reusable insulin delivery devices which can help patients track their timing and doses of insulin. This prevents inadvertent insulin delivery when patients forget about previously administered doses and improves glycaemic control by helping to reduce the frequency of missed doses. Such interventions have been shown to improve HbA1c levels and are cost-effective in non-pregnant population (112). They are increasingly used and are highly acceptable to patients and improve confidence in diabetes self-management in both type 1 and type 2 diabetes (113, 114). Although it is reasonable to assume that these benefits would translate to women with PGDM there is no evidence in this area.
2.4.3 Bolus calculating apps
Another innovation which has shown promise in the management of type 1 diabetes is the use of applications which facilitate carbohydrate counting and bolus calculation. A number of applications are available to facilitate carbohydrate counting, however more recent developments include the launch of food identification software that calculates the carbohydrate content of food using photographs (115, 116). This software can improve carbohydrate counting and HbA1c in a population of young adults, although its use is limited and it remains unvalidated in pregnancy (117).
3 Conclusion
In conclusion, a number of technological developments have improved the care for women with diabetes. Advancements in rtCGM and to a lesser extent isCGM offered greater convenience for patients, and have translated in improvements in clinical outcomes. CSII therapy has shown more conflicting results, however it has a number of benefits outside of pregnancy and the numbers of women entering pregnancy using CSII is likely to increase. Hybrid closed loop technology has shown significant promise in pregnancy. As this technology advances, becomes more widespread and cost-effective a greater number of women will be able to avail of its use - offering better glycaemic control and pregnancy outcomes in the future.
Author contributions
CN performed the data search and reviewed articles; AE drafted the manuscript; CN and FD contributed to the original idea and design of the study. All authors contributed to the article and approved the submitted version.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1. Chivese T, Hoegfeldt CA, Werfalli M, Yuen L, Sun H, Karuranga S, et al. IDF diabetes atlas: The prevalence of pre-existing diabetes in pregnancy - a systematic reviewand meta-analysis of studies published during 2010-2020. Diabetes Res Clin Pract (2022) 183:109049. doi: 10.1016/j.diabres.2021.109049
2. Coton SJ, Nazareth I, Petersen I. A cohort study of trends in the prevalence of pregestational diabetes in pregnancy recorded in UK general practice between 1995 and 2012. BMJ Open (2016) 6(1):e009494. doi: 10.1136/bmjopen-2015-009494
3. Newman C, Egan AM, Ahern T, Al-Kiyumi M, Galan G, Brassill MJ, et al. Diabetes care and pregnancy outcomes for women with pregestational diabetes in Ireland. Diabetes Res Clin Pract (2021) 173:108685. doi: 10.1016/j.diabres.2021.108685
4. Murphy HR, Howgate C, O’Keefe J, Myers J, Morgan M, Coleman MA, et al. Characteristics and outcomes of pregnant women with type 1 or type 2 diabetes: a 5-year national population-based cohort study. Lancet Diabetes Endocrinol (2021) 9(3):153–64. doi: 10.1016/S2213-8587(20)30406-X
5. Lascar N, Brown J, Pattison H, Barnett AH, Bailet CH, Bellary S, et al. Type 2 diabetes in adolescents and young adults. Lancet Diabetes Endocrinol (2018) 6(1):69–80. doi: 10.1016/S2213-8587(17)30186-9
6. Rafei A, Elliott MR, Jones RE, Riosmena F, Cunningham SA, Mehta NK, et al. Obesity incidence in U.S. children and young adults: A pooled analysis. Am J Prev Med (2022) 63(1):51–9. doi: 10.1016/j.amepre.2021.12.021
7. Thibault V, LeBlanc M, Babin E, Halpine L, Greene S, Mancuso B, et al. Factors that could explain the increasing prevalence of type 2 diabetes among adults in a Canadian province: a critical review and analysis. Diabetol Metab Syndrome (2016) 8(1):71. doi: 10.1186/s13098-016-0186-9
9. Alexopoulos AS, Blair R, Peters AL. Management of preexisting diabetes in pregnancy: A review. Jama (2019) 321(18):1811–9. doi: 10.1001/jama.2019.4981
10. Yu L, Zeng XL, Cheng ML, Yang GZ, Wang B, Xiao ZW, et al. Quantitative assessment of the effect of pre-gestational diabetes and risk of adverse maternal, perinatal and neonatal outcomes. Oncotarget (2017) 8(37):61048–56. doi: 10.18632/oncotarget.17824
11. Pitchika A, Winklet M, Hummel C, Hummel S, Krumsiek N, Kastenmuller J, et al. Associations of maternal type 1 diabetes with childhood adiposity and metabolic health in the offspring: a prospective cohort study. Diabetologia (2018) 61(11):2319–32. doi: 10.1007/s00125-018-4688-x
12. Yu Y, Onyebuchi A, Liew ZC, Olsen S, Sorensen J, Toft Qin H, et al. Maternal diabetes during pregnancy and early onset of cardiovascular disease in offspring: population based cohort study with 40 years of follow-up. BMJ (2019) 367:l6398. doi: 10.1136/bmj.l6398
13. Yamamoto JM, Benham JL, Dewey D, Sanchez JJ, Murphy HR, Feig DS, et al. Neurocognitive and behavioural outcomes in offspring exposed to maternal pre-existing diabetes: a systematic review and meta-analysis. Diabetologia (2019) 62(9):1561–74. doi: 10.1007/s00125-019-4923-0
14. Xiang AH, Wang X, Martinez MP, Page K, Buchanan TA, Feldman RK, et al. Maternal type 1 diabetes and risk of autism in offspring. Jama (2018) 320(1):89–91. doi: 10.1001/jama.2018.7614
15. Qiao L-X, Wang J, Yan JH, Xu SZ, Wang H, Zhu WY, et al. Follow-up study of neurodevelopment in 2-year-old infants who had suffered from neonatal hypoglycemia. BMC Pediatr (2019) 19(1):133. doi: 10.1186/s12887-019-1509-4
16. Feig D, Sanchez JJ, Asztalos K, Zinman E, Simmons B, Haqq B, et al. 248-OR: MiTy kids: Follow-up of offspring exposed to metformin in-utero in mothers with type 2 diabetes in the MiTy trial. Diabetes (2022) 71(Supplement_1):248–OR. doi: 10.2337/db22-248-OR
17. American Diabetes Association Professional Practice, C 15. management of diabetes in pregnancy: Standards of medical care in diabetes–2022. Diabetes Care (2021) 45(Supplement_1):S232–43. doi: 10.2337/dc22-SINT
18. National Institute for Health and Care Excellence. Guidelines, in diabetes in pregnancy: management from preconception to the postnatal period. 2020. London: National Institute for Health and Care Excellence (NICE (2020).
19. Tennant PW, Glinianaia SV, Bilous RW, Rankin J, Bell R. Pre-existing diabetes, maternal glycated haemoglobin, and the risks of fetal and infant death: a population-based study. Diabetologia (2014) 57(2):285–94. doi: 10.1007/s00125-013-3108-5
20. Bell R, Glinianaia SV, Tannant PW, Bilous RW, Rankin J. Peri-conception hyperglycaemia and nephropathy are associated with risk of congenital anomaly in women with pre-existing diabetes: a population-based cohort study. Diabetologia (2012) 55:936–47. doi: 10.1007/s00125-012-2455-y
21. Egan AM, et al. A prepregnancy care program for women with diabetes: Effective and cost saving. J Clin Endocrinol Metab (2016) 101(4):1807–15. doi: 10.1210/jc.2015-4046
22. Wahabi HA, et al. Systematic review and meta-analysis of the effectiveness of pre-pregnancy care for women with diabetes for improving maternal and perinatal outcomes. PloS One (2020) 15(8):e0237571. doi: 10.1371/journal.pone.0237571
23. Pfeiffer EF, Thum C, Clemens AH. The artificial beta cell–a continuous control of blood sugar by external regulation of insulin infusion (glucose controlled insulin infusion system). Horm Metab Res (1974) 6(5):339–42. doi: 10.1055/s-0028-1093841
24. Clemens AH, Chang PH, Myers RW. The development of biostator, a glucose controlled insulin infusion system (GCIIS). Horm Metab Res (1977) Suppl 7:23–33.
25. Fogt EJ, et al. Development and evaluation of a glucose analyzer for a glucose controlled insulin infusion system ((Biostator). Clin Chem (1978) 24(8):1366–72. doi: 10.1093/clinchem/24.8.1366
26. Pickup JC, et al. Continuous subcutaneous insulin infusion:good blood glucose control for up to 4 days. Diabetologia (1979) 16(6):385–9. doi: 10.1007/BF01223159
27. Santiago JV, et al. Closed-loop and open-loop devices for blood glucose control in normal and diabetic subjects. Diabetes (1978) 28(1):71–84. doi: 10.2337/diab.28.1.71
28. Choudhary P, et al. Advanced hybrid closed loop therapy versus conventional treatment in adults with type 1 diabetes (ADAPT): a randomised controlled study. Lancet Diabetes Endocrinol (2022) 10(10):720–31. doi: 10.1016/S2213-8587(22)00212-1
29. Karamitsos DT. The story of insulin discovery. Diabetes Res Clin Pract (2011) 93 Suppl 1:S2–8. doi: 10.1016/S0168-8227(11)70007-9
30. Quianzon CC, Cheikh I. History of insulin. J Community Hosp Intern Med Perspect (2012) 2(2). doi: 10.3402/jchimp.v2i2.18701
31. Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol (2009) 3(4):971–80. doi: 10.1177/193229680900300446
32. Alsaleh FM, et al. Insulin pumps: from inception to the present and toward the future. J Clin Pharm Ther (2010) 35(2):127–38. doi: 10.1111/j.1365-2710.2009.01048.x
33. Selam JL. Evolution of diabetes insulin delivery devices. J Diabetes Sci Technol (2010) 4(3):505–13. doi: 10.1177/193229681000400302
34. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. New Engl J Med (1993) 329(14):977–86. doi: 10.1016/s0022-3476(94)70190-3.
35. Olczuk D, Priefer R. A history of continuous glucose monitors (CGMs) in self-monitoring of diabetes mellitus. Diabetes Metab Syndrome: Clin Res Rev (2018) 12(2):181–7. doi: 10.1016/j.dsx.2017.09.005
36. Bergenstal RM, et al. Threshold-based insulin-pump interruption for reduction of hypoglycemia. New Engl J Med (2013) 369(3):224–32. doi: 10.1056/NEJMoa1303576
37. Bergenstal RM, et al. Sensor-augmented pump therapy for A1C reduction (STAR 3) study: results from the 6-month continuation phase. Diabetes Care (2011) 34(11):2403–5. doi: 10.2337/dc11-1248
38. Battelino T, et al. The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial. Diabetologia (2012) 55(12):3155–62. doi: 10.1007/s00125-012-2708-9
39. Bolinder J, et al. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet (2016) 388(10057):2254–63. doi: 10.1016/s0140-6736(16)31535-5
40. Kesavadev J, et al. The do-It-Yourself artificial pancreas: A comprehensive review. Diabetes Ther (2020) 11(6):1217–35. doi: 10.1007/s13300-020-00823-z
41. Stone MP, et al. Retrospective analysis of 3-month real-world glucose data after the MiniMed 670G system commercial launch. Diabetes Technol Ther (2018) 20(10):689–92. doi: 10.1089/dia.2018.0202
42. Brown SA, et al. Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. New Engl J Med (2019) 381(18):1707–17. doi: 10.1056/NEJMoa1907863
43. Multicenter, randomized trial of a bionic pancreas in type 1 diabetes. New Engl J Med (2022) 387(13):1161–72. doi: 10.1056/NEJMoa2205225.
44. Burnside MJ, et al. Open-source automated insulin delivery in type 1 diabetes. New Engl J Med (2022) 387(10):869–81. doi: 10.1056/NEJMoa2203913
45. Leelarathna L, et al. Intermittently scanned continuous glucose monitoring for type 1 diabetes. New Engl J Med (2022) 387:1477–87. doi: 10.1056/NEJMoa2205650
46. Beck RW, et al. Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: The DIAMOND randomized clinical trial. Jama (2017) 317(4):371–8. doi: 10.1001/jama.2016.19975
47. Cordua S, et al. Real-time continuous glucose monitoring during labour and delivery in women with type 1 diabetes - observations from a randomized controlled trial. Diabetes Med (2013) 30(11):1374–81. doi: 10.1111/dme.12246
48. Gabbay MAL, et al. Time in range: a new parameter to evaluate blood glucose control in patients with diabetes. Diabetol Metab Syndrome (2020) 12(1):22. doi: 10.1186/s13098-020-00529-z
49. Murphy HR, et al. Effectiveness of continuous glucose monitoring in pregnant women with diabetes: randomised clinical trial. BMJ (2008) 337:a1680. doi: 10.1136/bmj.a1680
50. Feig DS, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT): a multicentre international randomised controlled trial. Lancet (2017) 390(10110):2347–59. doi: 10.1016/S0140-6736(17)32400-5
51. Yamamoto JM, et al. Maternal glycaemic control and risk of neonatal hypoglycaemia in type 1 diabetes pregnancy: a secondary analysis of the CONCEPTT trial. Diabetes Med (2019) 36(8):1046–53. doi: 10.1111/dme.13988
52. Law GR, et al. Analysis of continuous glucose monitoring in pregnant women with diabetes: Distinct temporal patterns of glucose associated with Large-for-Gestational-Age infants. Diabetes Care (2015) 38(7):1319–25. doi: 10.2337/dc15-0070
53. Scott EM, et al. Continuous glucose monitoring metrics and birth weight: Informing management of type 1 diabetes throughout pregnancy. Diabetes Care (2022) 45(8):1724–34. doi: 10.2337/dc22-0078
54. Ling P, et al. Achieving the HbA1c target requires longer time in range in pregnant women with type 1 diabetes. J Clin Endocrinol Metab (2021) 106(11):e4309–17. doi: 10.1210/clinem/dgab502
55. Ahmed RJ, et al. The cost implications of continuous glucose monitoring in pregnant women with type 1 diabetes in 3 Canadian provinces: a posthoc cost analysis of the CONCEPTT trial. CMAJ Open (2021) 9(2):E627–e634. doi: 10.9778/cmajo.20200128
56. Murphy HR, et al. Modelling potential cost savings from use of real-time continuous glucose monitoring in pregnant women with type 1 diabetes. Diabetes Med (2019) 36(12):1652–8. doi: 10.1111/dme.14046
57. Voormolen DN, et al. Continuous glucose monitoring during diabetic pregnancy (GlucoMOMS): A multicentre randomized controlled trial. Diabetes Obes Metab (2018) 20(8):1894–902. doi: 10.1111/dom.13310
58. Secher AL, et al. The effect of real-time continuous glucose monitoring in pregnant women with diabetes: a randomized controlled trial. Diabetes Care (2013) 36(7):1877–83. doi: 10.2337/dc12-2360
59. Mensh BD, et al. Susceptibility of interstitial continuous glucose monitor performance to sleeping position. J Diabetes Sci Technol (2013) 7(4):863–70. doi: 10.1177/193229681300700408
60. Secher AL, et al. Patient satisfaction and barriers to initiating real-time continuous glucose monitoring in early pregnancy in women with diabetes. Diabetic Med J Br Diabetic Assoc (2012) 29(2):272–7. doi: 10.1111/j.1464-5491.2011.03426.x
61. Levy CJ, et al. 1424-p: Changes in device uptake and glycemic control among pregnant women with type 1 diabetes: Data from the T1D exchange. Diabetes (2019) 68(Supplement_1):1424–P. doi: 10.2337/db19-1424-P
62. Polsky S, et al. Continuous glucose monitor use with and without remote monitoring in pregnant women with type 1 diabetes: A pilot study. PloS One (2020) 15(4):e0230476. doi: 10.1371/journal.pone.0230476
63. Onisie O, Crocket H, de Bock M. The CGM grey market: a reflection of global access inequity. Lancet Diabetes Endocrinol (2019) 7(11):823–5. doi: 10.1016/S2213-8587(19)30263-3
64. Bailey CJ, Gavin JR. 3rd, flash continuous glucose monitoring: A summary review of recent real-world evidence. Clin Diabetes (2021) 39(1):64–71. doi: 10.2337/cd20-0076
65. Reiterer F, et al. Significance and reliability of MARD for the accuracy of CGM systems. J Diabetes Sci Technol (2017) 11(1):59–67. doi: 10.1177/1932296816662047
66. Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, user acceptability, and safety evaluation for the FreeStyle libre flash glucose monitoring system when used by pregnant women with diabetes. Diabetes Technol Ther (2018) 20(3):180–8. doi: 10.1089/dia.2017.0386
68. Ólafsdóttir AF, et al. A clinical trial of the accuracy and treatment experience of the flash glucose monitor FreeStyle libre in adults with type 1 diabetes. Diabetes Technol Ther (2017) 19(3):164–72. doi: 10.1089/dia.2016.0392
69. Tumminia A, et al. Efficacy of flash glucose monitoring in pregnant women with poorly controlled pregestational diabetes (FlashMom): A randomized pilot study. Nutr Metab Cardiovasc Dis (2021) 31(6):1851–9. doi: 10.1016/j.numecd.2021.03.013
70. Li SY, et al. Effects of intermittently scanned continuous glucose monitoring on blood glucose control and the production of urinary ketone bodies in pregestational diabetes mellitus. Diabetol Metab Syndr (2021) 13(1):39. doi: 10.1186/s13098-021-00657-0
71. Sola-Gazagnes A, et al. Disagreement between capillary blood glucose and flash glucose monitoring sensor can lead to inadequate treatment adjustments during pregnancy. Diabetes Metab (2020) 46(2):158–63. doi: 10.1016/j.diabet.2019.08.001
72. Perea V, et al. Addition of intermittently scanned continuous glucose monitoring to standard care in a cohort of pregnant women with type 1 diabetes: effect on glycaemic control and pregnancy outcomes. Diabetologia (2022) 65(8):1302–14. doi: 10.1007/s00125-022-05717-2
73. Nørgaard SK, et al. Comparison of glycemic metrics measured simultaneously by intermittently scanned continuous glucose monitoring and real-time continuous glucose monitoring in pregnant women with type 1 diabetes. Diabetes Technol Ther (2021) 23(10):665–72. doi: 10.1089/dia.2021.0109
74. Kristensen K, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes: an observational cohort study of 186 pregnancies. Diabetologia (2019) 62(7):1143–53. doi: 10.1007/s00125-019-4850-0
75. Petrovski G, et al. Is there a difference in pregnancy and glycemic outcome in patients with type 1 diabetes on insulin pump with constant or intermittent glucose monitoring? a pilot study. Diabetes Technol Ther (2011) 13(11):1109–13. doi: 10.1089/dia.2011.0081
76. Yamamoto JM, Murphy HR. Benefits of real-time continuous glucose monitoring in pregnancy. Diabetes Technol Ther (2021) 23(S1):S–8-S-14. doi: 10.1089/dia.2020.0667
77. Talaviya PA, et al. Pregnancy outcome and glycemic control in women with type 1 diabetes: a retrospective comparison between CSII and MDI treatment. Diabetes Metab Syndr (2013) 7(2):68–71. doi: 10.1016/j.dsx.2013.02.032
78. Abell SK, et al. Pregnancy outcomes and insulin requirements in women with type 1 diabetes treated with continuous subcutaneous insulin infusion and multiple daily injections: Cohort study. Diabetes Technol Ther (2017) 19(5):280–7. doi: 10.1089/dia.2016.0412
79. Kallas-Koeman MM, et al. Insulin pump use in pregnancy is associated with lower HbA1c without increasing the rate of severe hypoglycaemia or diabetic ketoacidosis in women with type 1 diabetes. Diabetologia (2014) 57(4):681–9. doi: 10.1007/s00125-014-3163-6
80. Bruttomesso D, et al. Type 1 diabetes control and pregnancy outcomes in women treated with continuous subcutaneous insulin infusion (CSII) or with insulin glargine and multiple daily injections of rapid-acting insulin analogues (glargine-MDI). Diabetes Metab (2011) 37(5):426–31. doi: 10.1016/j.diabet.2011.02.002
81. Kekäläinen P, et al. Continuous subcutaneous insulin infusion during pregnancy in women with complicated type 1 diabetes is associated with better glycemic control but not with improvement in pregnancy outcomes. Diabetes Technol Ther (2016) 18(3):144–50. doi: 10.1089/dia.2015.0165
82. Wender-Ozegowska E, et al. Multiple daily injections of insulin versus continuous subcutaneous insulin infusion for pregnant women with type 1 diabetes. Aust N Z J Obstet Gynaecol (2013) 53(2):130–5. doi: 10.1111/ajo.12027
83. Rys PM, et al. Continuous subcutaneous insulin infusion vs multiple daily injections in pregnant women with type 1 diabetes mellitus: a systematic review and meta-analysis of randomised controlled trials and observational studies. Eur J Endocrinol (2018) 178(5):545–63. doi: 10.1530/EJE-17-0804
84. Farrar D, et al. Continuous subcutaneous insulin infusion versus multiple daily injections of insulin for pregnant women with diabetes. Cochrane Database Systematic Rev (2016) 6). doi: 10.1002/14651858.CD005542.pub3
85. Hauffe F, et al. Higher rates of large-for-gestational-age newborns mediated by excess maternal weight gain in pregnancies with type 1 diabetes and use of continuous subcutaneous insulin infusion vs multiple dose insulin injection. Diabetic Med (2019) 36(2):158–66. doi: 10.1111/dme.13861
86. Feig DS, et al. Pumps or multiple daily injections in pregnancy involving type 1 diabetes: A prespecified analysis of the CONCEPTT randomized trial. Diabetes Care (2018) 41(12):2471–9. doi: 10.2337/dc18-1437
87. Kjölhede K, et al. Glycemic, maternal and neonatal outcomes in women with type 1 diabetes using continuous glucose monitoring during pregnancy - pump vs multiple daily injections, a secondary analysis of an observational cohort study. Acta Obstet Gynecol Scand (2021) 100(5):927–33. doi: 10.1111/aogs.14039
88. Drever E, et al. Insulin pump use compared with intravenous insulin during labour and delivery: the INSPIRED observational cohort study. Diabetes Med (2016) 33(9):1253–9. doi: 10.1111/dme.13106
89. Steineck I, et al. Sensor-augmented insulin pumps and hypoglycemia prevention in type 1 diabetes. J Diabetes Sci Technol (2017) 11(1):50–8. doi: 10.1177/1932296816672689
90. Templer S. Closed-loop insulin delivery systems: Past, present, and future directions. Front Endocrinol (Lausanne) (2022) 13:919942. doi: 10.3389/fendo.2022.919942
91. Gómez AM, et al. Maternal-fetal outcomes in 34 pregnant women with type 1 diabetes in sensor-augmented insulin pump therapy. Diabetes Technol Ther (2017) 19(7):417–22. doi: 10.1089/dia.2017.0030
92. Lason I, et al. Continuous glucose monitoring and insulin pump therapy in pregnant women with type 1 diabetes mellitus. Ginekol Pol (2021) 92(10):675–81. doi: 10.5603/GP.a2021.0029
93. Benhalima K, et al. Risk for ketonaemia in type 1 diabetes pregnancies with sensor-augmented pump therapy with predictive low glucose suspend compared with low glucose suspend: a crossover RCT. Diabetologia (2021) 64(12):2725–30. doi: 10.1007/s00125-021-05589-y
94. Murphy HR, et al. Closed-loop insulin delivery during pregnancy complicated by type 1 diabetes. Diabetes Care (2011) 34(2):406–11. doi: 10.2337/dc10-1796
95. Murphy HR KK, Elleri D, Allen JM, Caldwell K, Biagioni M, Simmons D, et al. Safety and efficacy of 24-h closed-loop insulin delivery in well-controlled pregnant women with type 1 diabetes: a randomized crossover case series. Diabetes Care (2012) 35(191):2527–9. doi: 10.2337/dc11-1430
96. Stewart ZA, et al. Closed-loop insulin delivery during pregnancy in women with type 1 diabetes. N Engl J Med (2016) 375(7):644–54. doi: 10.1056/NEJMoa1602494
97. Stewart ZA, et al. Day-and-Night closed-loop insulin delivery in a broad population of pregnant women with type 1 diabetes: A randomized controlled crossover trial. Diabetes Care (2018) 41(7):1391–9. doi: 10.2337/dc17-2534
98. Stewart ZA, et al. Adaptability of closed loop during labor, delivery, and postpartum: A secondary analysis of data from two randomized crossover trials in type 1 diabetes pregnancy. Diabetes Technol Ther (2018) 20(7):501–5. doi: 10.1089/dia.2018.0060
99. Chen NS, et al. User engagement with the CamAPS FX hybrid closed-loop app according to age and user characteristics. Diabetes Care (2021) 44(7):e148–50. doi: 10.2337/dc20-2762
100. Guzmán Gómez GE, et al. The closed-loop system improved the control of a pregnant patient with type 1 diabetes mellitus. Case Rep Endocrinol (2021) 2021:7310176. doi: 10.1155/2021/7310176
101. Ozaslan B, et al. Feasibility of closed-loop insulin delivery with a pregnancy-specific zone model predictive control algorithm. Diabetes Technol Ther (2022) 24(7):471–80. doi: 10.1089/dia.2021.0521
102. Lee TTM, et al. AiDAPT: automated insulin delivery amongst pregnant women with type 1 diabetes: a multicentre randomized controlled trial – study protocol. BMC Pregnancy Childbirth (2022) 22(1):282. doi: 10.1186/s12884-022-04543-z
103. Farrington C, et al. Women's experiences of day-and-Night closed-loop insulin delivery during type 1 diabetes pregnancy. J Diabetes Sci Technol (2018) 12(6):1125–31. doi: 10.1177/1932296818800065
104. Shivers JP, et al. "Turn it off!": diabetes device alarm fatigue considerations for the present and the future. J Diabetes Sci Technol (2013) 7(3):789–94. doi: 10.1177/193229681300700324
105. Mechanic OJ, Persaud Y, Kimball AB. Telehealth systems. In: StatPearls. Treasure Island (FL: StatPearls Publishing (2022).
106. Ahmed S, Sanghvi K, Yeo D. Telemedicine takes centre stage during COVID-19 pandemic. BMJ Innov (2020) 6(4):252. doi: 10.1136/bmjinnov-2020-000440
107. Knapp A, et al. Use of patient-reported outcome measures and patient-reported experience measures within evaluation studies of telemedicine applications: Systematic review. J Med Internet Res (2021) 23(11):e30042. doi: 10.2196/30042
108. Laursen SH, et al. Effectiveness of telemedicine in managing diabetes in pregnancy: A systematic review and meta-analysis. J Diabetes Sci Technol (2022) 17:19322968221094626. doi: 10.1177/19322968221094626
109. Carral F, et al. Web-based telemedicine system is useful for monitoring glucose control in pregnant women with diabetes. Diabetes Technol Ther (2015) 17(5):349–54. doi: 10.1089/dia.2014.0223
110. Dalfrà MG, Nicolucci A, Lapolla A. The effect of telemedicine on outcome and quality of life in pregnant women with diabetes. J Telemed Telecare (2009) 15(5):238–42. doi: 10.1258/jtt.2009.081213
111. Wojcicki JM, et al. What we can really expect from telemedicine in intensive diabetes treatment: results from 3-year study on type 1 pregnant diabetic women. Diabetes Technol Ther (2001) 3(4):581–9. doi: 10.1089/15209150152811207
112. Jendle J, et al. Smart insulin pens are associated with improved clinical outcomes at lower cost versus standard-of-Care treatment of type 1 diabetes in Sweden: A cost-effectiveness analysis. Diabetes Ther (2021) 12(1):373–88. doi: 10.1007/s13300-020-00980-1
113. Heinemann L, et al. Digital diabetes management: A literature review of smart insulin pens. J Diabetes Sci Technol (2022) 16(3):587–95. doi: 10.1177/1932296820983863
114. Galindo RJ, et al. Efficacy of a smart insulin pen cap for the management of patients with uncontrolled type 2 diabetes: A randomized cross-over trial. J Diabetes Sci Technol (2021) p:19322968211033837. doi: 10.1177/19322968211033837
115. Alfonsi JE, et al. Carbohydrate counting app using image recognition for youth with type 1 diabetes: Pilot randomized control trial. JMIR Mhealth Uhealth (2020) 8(10):e22074. doi: 10.2196/22074
116. Joubert M, et al. Prospective independent evaluation of the carbohydrate counting accuracy of two smartphone applications. Diabetes Ther (2021) 12(7):1809–20. doi: 10.1007/s13300-021-01082-2
Keywords: pregnancy, technology, continuous glucose monitoring (CGM), continuous subcutaneous insulin infusion (CSII), pumps
Citation: Newman C, Ero A and Dunne FP (2023) Glycaemic control and novel technology management strategies in pregestational diabetes mellitus. Front. Endocrinol. 13:1109825. doi: 10.3389/fendo.2022.1109825
Received: 28 November 2022; Accepted: 21 December 2022;
Published: 12 January 2023.
Edited by:
Lene Ringholm, Rigshospitalet, DenmarkReviewed by:
Robert Lindsay, University of Glasgow, United KingdomCopyright © 2023 Newman, Ero and Dunne. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Christine Newman, TmV3bWFuLmNocmlzdGluZTE3QGdtYWlsLmNvbQ==