Skip to main content

ORIGINAL RESEARCH article

Front. Chem., 11 November 2022
Sec. Electrochemistry

Electrochemical sensing of fentanyl as an anesthesia drug on NiO nanodisks combined with the carbon nanotube-modified electrode

Xi Li&#x;Xi Li1Bo Luo&#x;Bo Luo1Min Liao
Min Liao2*Abdullah MohamedAbdullah Mohamed3
  • 1Hospital of Chengdu University of Traditional Chinese Medicine, Chengdu, Sichuan, China
  • 2Department of Traditional Chinese Medicine, Sichuan Provincial People’s Hospital, University of Electronic Science and Technology of China, Chengdu, Sichuan, China
  • 3Research Centre, Future University in Egypt, New Cairo, Egypt

Fentanyl was successfully determined in the current effort based on hexagonal NiO nanodisks (HG-NiO-NDs) fabricated by the hydrothermal protocol. The synergism of HG-NiO-NDs with multiwall carbon nanotubes (MWCNTs), large specific surface area, and active material enabled the electrochemical sensor to show potent electrochemical behavior. Admirable performance was found for the fentanyl measurement by the MWCNT and HG-NiO-ND-modified pencil graphite electrode (MWCNT/HG-NiO-ND/PGE). The correlation of oxidation currents with the pH value, concentration, and sweep rate of supporting electrolytes was determined for the optimization of conditions to detect fentanyl. The surfaces of modified and unmodified electrodes were characterized as well. The diffusion-control processes were confirmed on the basis of anodic peak findings. The results also revealed a two-electron transfer process. The linear range was obtained to be 0.01–800.0 μM for the fentanyl concentrations on the developed electrode, with the sensitivity of 0.1044 μA/mM/cm2. The limit of detection (S/N = 3) was 6.7 nM. The results indicated the ability of the modified electrode to fabricate non-enzymatic fentanyl sensor applications.

1 Introduction

N-(1-Phenyl-4-piperidyl) propionanilide citrate, also called as fentanyl citrate, which is a strong man-made narcotic analgesic, is widely prescribed for analgesia and anesthesia in the intensive care unit and operating room. The effect of 100 μg of fentanyl is estimated to be equivalent to about 10 mg of morphine. Chronic pain can be relieved to some extent with transdermal fentanyl patches (Ebrahimzadeh et al., 2008). Fentanyl is metabolized in the liver after easy passage through plasma and the central nervous system (CNS) (Saffer et al., 2015). The serious complications have been reported for fentanyl, such as serotonin syndrome, respiratory distress, coma, addiction, gastrointestinal conditions, and hypotension. Therefore, it is critical to diagnose this substance as more medically dangerous than heroin (Sohouli et al., 2020). The conventional methods such as radioimmunoassay, surface-enhanced Raman spectroscopy (SERS), and other chromatographic techniques have already been employed to determine fentanyl in biological matrices such as blood and urine (Stiller et al., 1990; Saraji and Boroujeni, 2011; Haddad et al., 2018). Among these, special attention has been paid toward the electrochemical method for the fentanyl determination because of sensitivity, selectivity, low cost, rapidity, and safety (Foroughi et al., 2015; Foroughi and Ranjbar, 2017; Jahani, 2018; Maaref et al., 2018; Farvardin et al., 2020; Foroughi et al., 2021a; Moarefdoust et al., 2021; Santana et al., 2021; Dalkiran and Brett, 2022; Foroughi and Jahani, 2022; Jahani et al., 2022). It is very important to choose the right platform in the electrochemical method, for example, the pencil graphite electrode (PGE) is known as a suitable substrate in electrochemical applications with unique properties such as broad potential range and low background current (Antherjanam and Saraswathyamma, 2022). The electron transfer is slow on the surface of this green platform. This bottleneck can be bypassed by modifying the electrode surface with different strategies. Chemically modified electrodes (CMEs) have been recently introduced for different electroanalysis and (bio)sensors (Salajegheh et al., 2019; Al-Enizi et al., 2020a; Ubaidullah et al., 2020a; Vakili Fathabadi et al., 2020; Foroughi et al., 2021b; Vignesh et al., 2022).

Due to their high electrical conductivity, chemically modifiable surface area, large surface area, chemical stability, high mechanical strength, and high surface-to-volume ratio, carbon nanotubes (CNTs) are also attractive for electroanalysis. They are able to enable the oxidation of the analyte (You et al., 2021; Zhou et al., 2021). Carbon nanotubes containing metal oxides are expected to form a hybrid nanostructure for electrochemical sensors.

Metal oxide nanocrystals have distinct geometric shapes and sizes due to attractive size/shape/surface structure-dependent attributes and high potential as basic building blocks for nanoscale electronic and photonic equipment. Nickel oxide (NiO) is a p-type semiconductor and has a bandgap as broad as 3.6–4.0 eV (Yang et al., 2008; Ahmad et al., 2019; Fathi et al., 2020; Khand et al., 2021; Haunsbhavi et al., 2022). It has various applications such as catalysts, electrode materials for lithium ion batteries, electrochromic films, electrochemical supercapacitors, sensors, photovoltaic tools, and magnetic materials. Promising NiO applications and physicochemical properties of nanoscale materials have led to extensive studies to fabricate NiO nanomaterials with different morphologies (Ichiyanagia et al., 2003; Li et al., 2007; Al-Enizi et al., 2020b; Ubaidullah et al., 2020b).

The current work introduces a sensitive electrochemical sensor (MWCNT/HG-NiO-ND/PGE) with high reproducibility for the determination of fentanyl, based on the fabrication of a new nanostructure of hexagonal NiO nanodisks (HG-NiO-NDs), and then the surface modification of a pencil graphite electrode with the as-fabricated nanostructure. The characteristics of fentanyl were determined using cyclic voltammetry (CV), differential pulse voltammetry (DPV), and chronoamperometry (CHA) techniques.

2 Experimental

2.1 Chemicals and reagents

The multiwalled carbon nanotubes (MWCNTs) with a diameter of nanotubes (NTs) of OD = 6–13 nm, purity of >98%, and length of 2.5–20 μm were bought from Aldrich. Some local pharmacies were selected to purchase fentanyl tablets. Other required chemicals (belonging to Merck) with a practical grade were included. All solutions were freshly prepared with double distilled water (DDW). To prepare the fentanyl standard solution, an aliquot amount of fentanyl was dissolved in methanol (10 ml), followed by diluting to 50 ml at pH 7 (the final concentration of 1 mM). The lower concentrations were daily obtained by stepwise dilution.

2.2. Equipment

A potentiostat (Metrohm 757 VA Computrace, Herisau, Switzerland) with a conical vessel was utilized to perform all voltammetric determinations, and a three-electrode system consisted of one working electrode (pencil graphite electrode (PGE)), one axillary electrode, (Pt) and one reference electrode (SCE). All pH measurements were carried out using a digital Metrohm 710 pH/mV meter. The Rigaku D/MAX-3B powder diffractometer with Cu/Kα radiation at λ = 1.54056 Å was employed to record X-ray diffraction (XRD) with intensities as a function of 2θ. The specimen was scanned at 10–80° (2θ) in 0.02 steps. Surface analysis was performed using images from su35000 field emission scanning electron microscopy (FESEM; (Hitachi; Japan)) and Ametek energy-dispersive X-ray spectroscopy (EDS; Octane Prime; United States).

2.3 Construction of hexagonal NiO nanodisks

Porous hexagonal NiO nanodisks were prepared by dissolving nickel nitrate hexahydrate (Ni(NO3)2.6H2O, 1.45 g) and hexamethylenetetramine (1.0 g) in DDW (50 ml) while stirring rigorously for 15 min. The solution pH was adjusted to 13 using some drops of NaOH, followed by stirring vigorously for 60 min. The solution was then sealed and heated up to 150°C for 10 h in a Teflon-lined stainless steel autoclave, followed by cooling down to room temperature. The resultant precipitate was washed with DDW/ethanol and dried at ambient temperature overnight, followed by calcination at 650°C for 3 h. Diverse techniques were applied to characterize the final product.

2.4 Construction of the modified electrode

Due to the importance of electrode construction in the electrochemical analysis, the PGE surface was first washed thoroughly with water to remove possible impurities. Then, the MWCNT (1 mg) was dispersed in DDW (1 ml) to prepare the modifier under 15-min sonication. The PGE surface was coated with suspension (6 µL) via drop-casting, followed by drying at 50°C in an air oven. Identical methods were adopted to fabricate MWCNT/HG-NiO-ND/PGE with the addition of 6 µL of HG-NiO-ND suspension (1.0 mg HG-NiO-ND + 1.0 ml DDW) on the MWCNT/PGE surface.

2.5 Analysis of tablet and serum specimens

Some of the fentanyl powders required to prepare the stock solution (1.0 × 10−3 M) were dissolved in DDW under sonication. Aliquots of the clear supernatant of the tablet solution were diluted with phosphate-buffered solution (pH 7) for analysis. Certain amounts of pure drugs were added to the tablet solution under analysis to evaluate the effect of the tablet excipients, the accuracy of the technique.

The serum samples were obtained from a reputable medical laboratory (Pasteur Bam Hospital). The serum sample concentration was adjusted in phosphate-buffered solution before fentanyl analysis.

3 Results and discussion

3.1 Determination of hexagonal NiO nanodisk characteristics

As-fabricated HG-NiO-NDs were examined for crystallinity, phase, and morphology as follows: the XRD patterns (Figure 1) regarding the crystal structure verified an acceptable crystallinity. The peaks at 35.76°, 43.46°, 63.89°, 74.95°, and 78.68° were related to the planes of (111), (200), (220), (311), and (222), respectively. The XRD data are evidence of a cubic form of NiO (JCPDS No: 78-0643) (Zhou et al., 2018). According to XRD findings, the diffraction peaks corresponded to NiO only. Based on the XRD results, the Scherer equation of D = Kλ/ßcosθ was used to calculate the HG-NiO-ND crystallite size, where λ stands for the used X-ray wavelength (1.541 Å), β for the peak width at half maximum (FWHM), and θ for the Bragg diffraction angle, which was 106.43 nm.

FIGURE 1
www.frontiersin.org

FIGURE 1. XRD pattern of HG-NiO-NDs.

The morphological analysis of HG-NiO-NDs was performed by FE-SEM, the results of which are presented in Figure 2. Interestingly, the prepared HG-NiO-ND possesses hexagonal shape morphologies with high-porous surfaces and irregular pore sizes as shown in Figure 2B. Due to high-porous morphologies, some broken nanodisks are also observed in Figure 2A. It is fascinating to see that most of the nanodisks possess perfect hexagonal shapes with an internal angle of ∼120°; however, there are also some deformed randomly distributed nanodisks. The mean diagonal of nanodisks ranged from 0.4 to 1.5 μm although the micrograph shows some larger nanodisks. The mean typical nanodisk thickness was 10–20 nm, as shown in Figure 2C.

FIGURE 2
www.frontiersin.org

FIGURE 2. (A) and (B) FESEM image. (C) High-resolution FESEM image of HG-NiO-NDs.

Figure 3 illustrates HG-NiO-ND EDS analysis. According to the EDS analysis, the compositions contained only Ni and O, with no impurity. The elemental mapping also confirms the distribution of Ni and O.

FIGURE 3
www.frontiersin.org

FIGURE 3. EDX spectra and elemental mapping of HG-NiO-NDs.

XPS studies (Figure 4) were performed to establish the binding energy and oxidation states of the elements present at the surface of HG-NiO-NDs. The wide XPS survey spectra (Figure 4A) reveal the presence of Ni and O. High-resolution spectra, as shown in Figures 4B–D, typically exhibit 2P3/2 which consists of the main peak at ∼853 eV. Similarly, 2P1/2 displays two peaks at ∼871 eV and ∼878 eV, corresponding to the main peak and the satellite peak, respectively. The Ni 2P3/2 peak at ∼853 eV and O 1s peak at ∼529 eV are from Ni2+ and lattice oxygen, respectively, and are incorporated with the Ni–O octahedral bonding of cubic rock salt (Al-Enizi et al., 2021). The carbon peak is originated from surface contamination in the process of handling and storage of the sample (Figure 4D).

FIGURE 4
www.frontiersin.org

FIGURE 4. XPS spectra of HG-NiO-NDs: (A) survey; high-resolution spectra of (B) Ni 2p, (C) O 1s, and (D) C 1s.

3.2 Electrochemical behaviors of MWCNT/HG-NiO-ND/PGE

The CV curves were recorded for bare PGE, MWCNT/PGE, and MWCNT/HG-NiO-ND/PGE in the redox probe. Figure 5A shows the redox peaks for MWCNT/HG-NiO-ND/PGE, possessing the peak-to-peak differences (∆Ep = Eanodicpeak − Ecathodicpeak) of 0.19 V. Figure 5B shows the performances of the MWCNT/HG-NiO-ND/PGE according to the Randles–Sevcik equation expressed in Eq. 1 (Bard and Faulkner, 2001):

Ip=±(2.69×105)n3/2AD1/2Cv1/2.(1)

FIGURE 5
www.frontiersin.org

FIGURE 5. (A) CVs of MWCNT/HG-NiO-ND/PGE in the presence of 0.2 mM [Fe(CN)6]3- solution in aqueous 0.1 M KCl at various scan rates (from inner to outer curves): 10, 20, 30, 40, 50, 60, 70, 80, 90, 100, 200, 300, 400, and 500 mV/s. (B) Plot of peak currents vs. υ1/2.

In this equation, the A values are 0.14, 0.19, and 0.32 cm2 for the surfaces of bare PGE (BPGE), MWCNT/PGE, and MWCNT/HG-NiO-ND/PGE, respectively.

The EIS method was used to electrochemically determine the characteristics of HG-NiO-NDs in which the charge-transfer resistance (Rct) shows the electron-transfer kinetics of the redox probe at the electrode interface, confirming the substrate bond on the surface of the modified electrode. Nyquist plots were drawn for the BPGE, MWCNT/PGE, and MWCNT/HG-NiO-ND/PGE in the redox probe (Figure 6). As seen, a large semicircular structure exists for the BPGE at high frequencies, as a high charge-transfer resistance (Rct = 1,194 Ω) related to a low charge and mass transfer rate. MWCNT/PGE and MWCNT/HG-NiO-ND/PGE showed a significant decrease in the Rct values (794 Ω and 483 Ω, respectively), which can be because of capacity of NiO ND and HG-NiO-ND to boost the electron transfer as well as the electrode surface area. The results showed that the resistance is lower after modification with MWCNTs and HG-NiO-NDs, demonstrating that modifiers increase the oxidation peak current of the fentanyl; consequently, it increases the sensitivity.

FIGURE 6
www.frontiersin.org

FIGURE 6. EIS diagrams and the equivalent circuit for 0.1 mM [Fe(CN)6]3- solution at (A) BPGE, (B) MWCNT/PGE, and (C) MWCNT/HG-NiO-ND/PGE in aqueous 0.1 M KCl. Frequency ranges from 100 KHz to 0.1 Hz.

Based on the EIS, the standard heterogeneous rate constant for electron standard transfer (k0, cm/s) was calculated to explore the modification of the electrode surface according to Eq. 2 (Bard and Faulkner, 2001):

k0=RTF2RctAC.(2)

In this equation, R stands for the global gas constant (squared with 8.314 J/K/mol), T for thermodynamic temperature (298.15 K), F for Faraday constant values (96.485 C/mol), A for the electrode surface area (cm2), and C for the concentration of 0.1 mM [Fe(CN)6]3-/4-.

The k0 values were computed to be 1.59 × 10−2, 1.67 × 10−2, and 1.72 × 10−2 cm/s for the BPGE, MWCNT/PGE, and MWCNT/HG-NiO-ND/PGE, respectively. The values of K0 approach the kinetic potential of the redox couple. Thus, a system with a higher k0 value has a longer balance in less time than a system with a lower k0 value. Therefore, a greater k0 value is obtained for the MWCNT/HG-NiO-ND/PGE sensor than MWCNT/HG-NiO-ND/PGE > MWCNT/PGE > BPGE, indicating faster electron transfer than other electrodes.

3.3 The voltammetric activity of fentanyl

The cyclic voltammetry (CV) method was followed to explore the electrochemical activity of fentanyl (160.0 μM) in 0.1 M PBS at pH 7.0 on the surfaces of MWCNT/HG-NiO-ND/PGE, MWCNT/PGE, and BPGE. The results showed that MWCNT/HG-NiO-ND/PGE does not undergo any oxidation reaction in the absence of fentanyl in 0.1 M PBS at a pH of 7.0 (Figure 7 (curve a)). Figure 7 (curve b) shows the fentanyl oxidation peak of 924 mV with an ultra-low peak current on the BPGE. Figure 7 (curve c) shows a distinct peak for fentanyl on the MWCNT/PGE at 821 mV. Figure 7 (curve c) also illustrates six times higher peak current for fentanyl on MWCNT/PGE than on BPGE (curve b) because of the catalytic ability of MWCNTs. Figure 7 (curve d) shows increased peak current due to the augmentation of MWCNTs and HG-NiO-NDs. The electro-catalytic synergism of MWCNTs with HG-NiO-NDs caused the MWCNT/HG-NiO-ND/PGE to improve its catalytic behavior toward fentanyl and was useful for detecting this compound.

FIGURE 7
www.frontiersin.org

FIGURE 7. CVs of (A) BPGE, (B) MWCNT/PEG and (C) MWCNT/HG-NiO-ND/PEG in 0.1 M PBS (pH 7.0) containing 160.0 μM fentanyl.

3.4 The effect of solution pH on fentanyl oxidation

The solution pH is a pivotal parameter influencing the fentanyl electro-oxidation because of the presence of protons in the electrode reaction. The measurements (Figure 8) were carried out by the CV method regarding the signal of MWCNT/HG-NiO-ND/PGE in 0.1 M buffer solutions at variable pH values from 4 to 8. There was a minor increase in the fentanyl peak currents with increasing solution pH until 7.0 and then decreased. The highest peak current was found at pH 7.0 for fentanyl. Gradually, elevated pH of the solution switched the peak oxidation potential of fentanyl toward less positive values, which shows the presence of protons during the electrode reactions. Since the PBS with pH 7.0 created an optimized reaction for peak current and peak shape and negative shift, the value was selected to be the best (working pH) for next testing. The plot of Ep versus pH was drawn for fentanyl at pH 7 (Figure 8B). The Ep values of fentanyl had a linear relationship with buffer solution pH as follows:

Fentanyl:Ep(V)=0.0493pH+1.1656(R2=0.9999).(3)

FIGURE 8
www.frontiersin.org

FIGURE 8. (A) Effect of pH on the peak currents for the oxidation of fentanyl (125.0 µM), pH = 4–8. (B) Plots of peak potential vs. pH. Scan rate: 50 mV/s.

Concerning the slope of 0.0493 V/pH for fentanyl, they were close to the Nernstian value predicted for an equal number of proton and electron electrochemical process (Scheme 1) (Bard and Faulkner, 2001).

SCHEME 1
www.frontiersin.org

SCHEME 1. Probable oxidation mechanism for fentanyl.

3.5 The effect of the scan rate on electrochemical behaviors of fentanyl

The effect of the scan rate on the peak current of fentanyl oxidation was explored by the CV method on the MWCNT/HG-NiO-ND/PGE. The peak current intensity was elevated with the increasing scan rate, as shown in Figure 9A. Figure 9B highlights the current fit with the scan rate square root (10–500 mV/s), which means the redox reactions are controlled by fentanyl diffusion. Data suggested that the scan rate of 50 mV/s was the best for peak currents and peak separation.

FIGURE 9
www.frontiersin.org

FIGURE 9. (A) CVs of MWCNT/HG-NiO-ND/PGE at pH 7.0 in the presence of fentanyl (20.0 µM) at various scan rates (from inner to outer curves): 10, 20, 30, 40, 50, 60, 70, 80, 90, 100, 200, 300, 400, and 500 mV/s. (B) Plots of peak currents vs. υ1/2.

3.6. The chronoamperometric analysis

Figure 10A shows the chronoamperometric determinations of fentanyl on the MWCNT/HG-NiO-ND/PGE at a potential of 871 mV for variable fentanyl contents in PBS at pH 7. For the fentanyl with a determined diffusion coefficient (D), the Cottrell equation refers to the electrochemical reaction current with a mass transport-limited rate (Bard and Faulkner, 2001).

I=nFAD1/2Cbπ1/2t1/2.(4)

FIGURE 10
www.frontiersin.org

FIGURE 10. (A) Chronoamperograms of fentanyl on MWCNT/HG-NiO-ND/PGE in 0.1 M PBS (pH 7.0) for different concentrations of fentanyl. The numbers 1–4 correspond to 0.1, 0.15, 0.2, and 0.25 mM of fentanyl, respectively. (B) Plots of I vs. t−1/2 obtained from chronoamperograms 1–4. (C) Plot of the slope of the straight lines against fentanyl concentration.

Figure 10B illustrates a linear plot of I versus t−1/2 under the diffusion-controlled process; the D value can be obtained for fentanyl based on the linear part of the slope of the Cottrell plot, as shown in Figure 10C, which was 4.1 × 10–6 cm2/s.

3.7 The determination of fentanyl

The DPV method was utilized to clarify the relationship between the peak current and fentanyl content (Figure 11). The DPV curves show the oxidation peaks. The peak currents of fentanyl oxidation on the surface of MWCNT/HG-NiO-ND/PGE were linearly related to variable fentanyl contents (0.01–800.0 μM), and the LOD value was as low as 6.7 nM.

FIGURE 11
www.frontiersin.org

FIGURE 11. (A) DPVs of fentanyl on MWCNT/HG-NiO-ND/PGE in PBS (pH 7.0) at the scan rate of 50 mV s−1. Concentrations of fentanyl from inner to outer curves: 0.01, 1.0, 10.0, 20.0, 30.0, 40.0, 50.0, 60.0, 70.0, 80.0, 90.0, 100.0, 200.0, 300.0, 400.0, 500.0, 600.0, 700.0, and 800.0 (μM). (B) Plots of I vs. concentrations.

The fentanyl content was determined by seven consecutive measurements (150.0 μM), and the relative standard deviation (RSD%) was obtained to be 1.8, which means high efficiency of the proposed protocol for the fentanyl determination.

3.8 The interference analysis

The possible interferences of other analytes were explored by DPV in the fentanyl determination. According to the results, there was no significant interference from both diverse anions and cations (Na+, Ca2+, K+, Zn2+, Ni2+, NO2-, Cl, and SO42-) and l-phenylalanine, leucine, alanine, uric acid, l-glutamic acid, l-tryptophan, ascorbic acid, l-tyrosine, and epinephrine at 150.0 M concentration with the signal deviations of less than 5%.

3.9 The analysis of real specimens

The practical potential of the proposed fentanyl sensor was explored in the drug tablets and blood serum specimens. The reliability of the method was tested by sensing the fentanyl in the specimens spiked with certain content of the analyte (Table 1). The as-fabricated fentanyl sensor applicability was confirmed on the basis of impressive recovery rates obtained between 98.62 and 101.5%.

TABLE 1
www.frontiersin.org

TABLE 1. Determination of fentanyl in fentanyl tablets and human blood serum samples. All concentrations are given in μM (n = 5).

3.10 Reproducibility and stability

Repeatability and stability of the electrochemically as-fabricated fentanyl sensor were tested by the DPV method under the optimized circumstances. Six successive applications of MWCNT/HG-NiO-ND/PGE to quantify 150.0 μM of fentanyl showed no clear alteration in the DPV response. The RSD value of 0.94% validated the commendable repeatability of our sensor. To determine the sensor stability, it was incubated at room temperature for three consecutive weeks, the results of which revealed no distinct change in the peak current (2.82%), which means long-lasting stability of our electrode under the optimized circumstances.

3.11 Comparison of our method with others in the literature

The comparison of analytical efficacy between the as-fabricated electrode and other electrochemical methods was performed for fentanyl (Tables). Based on Table 2, the performance of our proposed electrochemical electrode for sensing fentanyl displayed a comparable linear range and better detection limit and sensitivity when compared with the other methods (Najafi and Sohuli, 2018; Naghian et al., 2020; Sohouli et al., 2020; Wester et al., 2020). Accordingly, the as-fabricated sensor is potentially able to determine the trace amounts of studied drugs in various media. Moreover, the electrode used for the sensor fabrication is a PGE that has various advantages, like cost-effectiveness, facile modification, admirable accessibility, and lower background current, when compared with other electrodes.

TABLE 2
www.frontiersin.org

TABLE 2. Performance comparison of MWCNT/HG-NiO-ND/PGE for the determination of fentanyl with other electroanalytical methods.

4 Conclusion

This study introduces a new fentanyl electrochemical sensor based on carbon nanotubes combined with HG-NiO-NDs. The production of HG-NiO-NDs added merits and was considered a green approach in the construction of non-toxic sensing systems. The electrochemical behavior of the modified electrode was evaluated for the electrocatalytic oxidation of fentanyl under optimal conditions of electrolytes (0.1 M PBS, pH = 7; a scan rate of 50 mV/s). It was found that the modified electrode could sense fentanyl in the linear range of 0.01 × 10–6 to 800.0 × 10–6 M. These research findings demonstrated the synergetic impact of MWCNTs and HG-NiO-NDs on the oxidation of fentanyl via declining oxidation over-potential and enhancing the oxidation peak current. Reproducible responses together with a lower limit of detection (6.7 nM), in comparison with other literature works, can be attained through the utility of this electrode. The modified electrode was able to provide a very sensitive and stable behavior to fentanyl determination without interference under optimal conditions. In addition, sensitiveness and functionality of the suggested electrochemical sensor toward fentanyl were explored by analyzing pharmaceutical samples, which was accompanied by acceptable outcomes.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material; further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving human participants were reviewed and approved by the Ethics Committee of the University of Electronic Science and Technology of China. The patients/participants provided their written informed consent to participate in this study.

Author contributions

All authors listed have made a substantial, direct, and intellectual contribution to the work and approved it for publication.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors, and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Ahmad, N., Alam, M., Wahab, R., Ahmad, J., Ubaidullah, M., Ansari, A. A., et al. (2019). Synthesis of NiO–CeO2 nanocomposite for electrochemical sensing of perilous 4-nitrophenol. J. Mat. Sci. Mat. Electron. 30, 17643–17653. doi:10.1007/s10854-019-02113-2

CrossRef Full Text | Google Scholar

Al-Enizi, A. M., Ahmed, J., Ubaidullah, M., Shaikh, S. F., Ahamad, T., Naushad, M., et al. (2020). Utilization of waste polyethylene terephthalate bottles to develop metal-organic frameworks for energy applications: A clean and feasible approach. J. Clean. Prod. 248, 119251. doi:10.1016/j.jclepro.2019.119251

CrossRef Full Text | Google Scholar

Al-Enizi, A. M., Ubaidullah, M., Ahmed, J., Ahamad, T., Ahmad, T., Shaikh, S. F., et al. (2020). Synthesis of NiOx@NPC composite for high-performance supercapacitor via waste PET plastic-derived Ni-MOF. Compos. Part B Eng. 183, 107655. doi:10.1016/j.compositesb.2019.107655

CrossRef Full Text | Google Scholar

Al-Enizi, A. M., Ubaidullah, M., and Kumar, D. (2021). Carbon quantum dots (CQDs)/Ce doped NiO nanocomposite for high performance supercapacitor. Mat. Today Commun. 27, 102340. doi:10.1016/j.mtcomm.2021.102340

CrossRef Full Text | Google Scholar

Antherjanam, S., and Saraswathyamma, B. (2022). Simultaneous electrochemical determination of hydrazine and hydroxylamine on a thiadiazole derivative modified pencil graphite electrode. Mat. Chem. Phys. 275, 125223. doi:10.1016/j.matchemphys.2021.125223

CrossRef Full Text | Google Scholar

Bard, A. J., and Faulkner, L. R. (2001). Electrochemical methods: Fundamentals and applications. second ed. New York: Wiley.

Google Scholar

Dalkiran, B., and Brett, C. M. A. (2022). Poly(safranine T)-deep eutectic solvent/copper oxide nanoparticle-carbon nanotube nanocomposite modified electrode and its application to the simultaneous determination of hydroquinone and catechol. Microchim. J. 179, 107531. doi:10.1016/j.microc.2022.107531

CrossRef Full Text | Google Scholar

Ebrahimzadeh, H., Yamini, Y., Gholizade, A., Sedighi, A., and Kasraee, S. (2008). Determination of fentanyl in biological and water samples using single-drop liquid–liquid–liquid microextraction coupled with high-performance liquid chromatography. Anal. Chim. Acta X. 626, 193–199. doi:10.1016/j.aca.2008.07.047

PubMed Abstract | CrossRef Full Text | Google Scholar

Farvardin, N., Jahani, Sh., Kazemipour, M., and Foroughi, M. M. (2020). The synthesis and characterization of 3D mesoporous CeO2 hollow spheres as a modifier for the simultaneous determination of amlodipine, hydrochlorothiazide and valsartan. Anal. Methods 12, 1767–1778. doi:10.1039/d0ay00022a

CrossRef Full Text | Google Scholar

Fathi, Z., Jahani, Sh., Shahidi Zandi, M., and Foroughi, M. M. (2020). Synthesis of bifunctional cabbage flower–like Ho3+/NiO nanostructures as a modifier for simultaneous determination of methotrexate and carbamazepine. Anal. Bioanal. Chem. 412, 1011–1024. doi:10.1007/s00216-019-02326-8

PubMed Abstract | CrossRef Full Text | Google Scholar

Foroughi, M. M., Jahani, Sh., Aramesh-Boroujeni, Z., Vakili Fathabadi, M., Hashemipour Rafsanjani, H., and Rostaminasab Dolatabad, M. (2021). Template-free synthesis of ZnO/Fe3O4/Carbon magnetic nanocomposite: Nanotubes with hexagonal cross sections and their electrocatalytic property for simultaneous determination of oxymorphone and heroin. Microchem. J. 170, 106679. doi:10.1016/j.microc.2021.106679

CrossRef Full Text | Google Scholar

Foroughi, M. M., Jahani, Sh., Aramesh-Boroujeni, Z., Rostaminasab Dolatabad, M., and Shahbazkhani, K. (2021). Synthesis of 3D cubic of Eu3+/Cu2O with clover-like faces nanostructures and their application as an electrochemical sensor for determination of antiretroviral drug nevirapine. Ceram. Int. 47, 19727–19736. doi:10.1016/j.ceramint.2021.03.311

CrossRef Full Text | Google Scholar

Foroughi, M. M., and Jahani, Sh. (2022). Investigation of a high-sensitive electrochemical DNA biosensor for determination of Idarubicin and studies of DNA-binding properties. Microchem. J. 179, 107546. doi:10.1016/j.microc.2022.107546

CrossRef Full Text | Google Scholar

Foroughi, M. M., Noroozifar, M., and Khorasani-Motlagh, M. (2015). Simultaneous determination of hydroquinone and catechol using a modified glassy carbon electrode by ruthenium red/carbon nanotube. J. Iran. Chem. Soc. 12, 1139–1147. doi:10.1007/s13738-014-0575-7

CrossRef Full Text | Google Scholar

Foroughi, M. M., and Ranjbar, M. (2017). Microwave-assisted synthesis and characterization photoluminescence properties: A fast, efficient route to produce ZnO/GrO nanocrystalline. J. Mat. Sci. Mat. Electron. 28, 1359–1363. doi:10.1007/s10854-016-5668-x

CrossRef Full Text | Google Scholar

Haddad, A., Comanescu, M. A., Green, O., Kubic, T. A., and Lombardi, J. R. (2018). Detection and quantitation of trace fentanyl in heroin by surface-enhanced Raman spectroscopy. Anal. Chem. 90, 12678–12685. doi:10.1021/acs.analchem.8b02909

PubMed Abstract | CrossRef Full Text | Google Scholar

Haunsbhavi, K., Arun Kumar, K. D., Ubaidullah, M., Shaikh, S. F., Venkatesh, R., Alagarasan, D., et al. (2022). The effect of rare-earth element (Gd, Nd, La) doping of NiO films on UV photodetector. Phys. Scr. 97, 055815. doi:10.1088/1402-4896/ac64d4

CrossRef Full Text | Google Scholar

Ichiyanagia, Y., Wakabayashia, N., and Yamazaki, J. (2003). Magnetic properties of NiO nanoparticles. Phys. B Condens. Matter 329, 862–863. doi:10.1016/s0921-4526(02)02578-4

CrossRef Full Text | Google Scholar

Jahani, Sh. (2018). Evaluation of the usefulness of an electrochemical sensor in detecting ascorbic acid using a graphite screen-printed electrode modified with NiFe2O4 nanoparticles. Anal. Bioanal. Electrochem. 10, 739–750.

Google Scholar

Jahani, Sh., Sedighi, A., Toolabi, A., and Foroughi, M. M. (2022). Development and characterization of La2O3 nanoparticles@snowflake-like Cu2S nanostructure composite modified electrode and application for simultaneous detection of catechol, hydroquinone and resorcinol as an electrochemical sensor. Electrochim. Acta 416, 140261. doi:10.1016/j.electacta.2022.140261

CrossRef Full Text | Google Scholar

Khand, A. A., Lakho, S. A., Tahira, A., Ubaidullah, M., Alothman, A. A., Aljadoa, K., et al. (2021). Facile electrochemical determination of methotrexate (MTX) using glassy carbon electrode-modified with electronically disordered NiO nanostructures. Nanomaterials 11, 1266. doi:10.3390/nano11051266

PubMed Abstract | CrossRef Full Text | Google Scholar

Li, Q., Wang, L. S., Hu, B. Y., Yang, C., Zhou, L., and Zhang, L. (2007). Preparation and characterization of NiO nanoparticles through calcination of malate gel. Mat. Lett. 61, 1615–1618. doi:10.1016/j.matlet.2006.07.113

CrossRef Full Text | Google Scholar

Maaref, H., Foroughi, M. M., Sheikhhosseini, E., and Akhgar, M. R. (2018). Electrocatalytic oxidation of sulfite and its highly sensitive determination on graphite screen printed electrode modified with new schiff base compound. Anal. Bioanal. Electrochem. 10, 1080–1092.

Google Scholar

Moarefdoust, M. M., Jahani, Sh., Moradalizadeh, M., Motaghi, M. M., and Foroughi, M. M. (2021). An electrochemical sensor based on hierarchical nickel oxide nanostructures doped with indium ions for voltammetric simultaneous determination of sunset yellow and tartrazine colorants in soft drink powders. Anal. Methods 13, 2396–2404. doi:10.1039/d1ay00306b

PubMed Abstract | CrossRef Full Text | Google Scholar

Naghian, E., Marzi Khosrowshahi, E., Sohouli, E., Ahmadi, F., Rahimi-Nasrabadi, M., and Safarifard, V. (2020). A new electrochemical sensor for the detection of fentanyl lethal drug by a screen-printed carbon electrode modified with the open-ended channels of Zn(II)-MOF. New J. Chem. 44, 9271–9277. doi:10.1039/d0nj01322f

CrossRef Full Text | Google Scholar

Najafi, M., and Sohuli, S. (2018). Electrochemical sensor for fentanyl determination by modified electrode with carbon nanotube and iron (III) oxide nanoparticles. J. Appl. Res. Chem. 12, 103.

Google Scholar

Saffer, C. S., Minkowitz, H. S., Ding, L., Danesi, H., and Jones, J. B. (2015). Fentanyl iontophoretic transdermal system versus morphine intravenous patient-controlled analgesia for pain management following gynecological surgery: A meta-analysis of randomized, controlled trials. Pain Manag. 5, 339–348. doi:10.2217/pmt.15.29

PubMed Abstract | CrossRef Full Text | Google Scholar

Salajegheh, M., Ansari, M., Foroughi, M. M., and Kazemipour, M. (2019). Computational design as a green approach for facile preparation of molecularly imprinted polyarginine-sodium alginate-multiwalled carbon nanotubes composite film on glassy carbon electrode for theophylline sensing. J. Pharm. Biomed. Anal. 162, 215–224. doi:10.1016/j.jpba.2018.09.032

PubMed Abstract | CrossRef Full Text | Google Scholar

Santana, E. R., Martins, E. C., and Spinelli, A. (2021). Electrode modified with nitrogen-doped graphene quantum dots supported in chitosan for triclocarban monitoring. Microchem. J. 167, 106297. doi:10.1016/j.microc.2021.106297

CrossRef Full Text | Google Scholar

Saraji, M., and Boroujeni, M. K. (2011). Analysis of narcotic drugs in biological samples using hollow fiber liquid–phase microextraction and gas chromatography with nitrogen phosphorus detection. Microchim. Acta 174, 159–166. doi:10.1007/s00604-011-0612-5

CrossRef Full Text | Google Scholar

Sohouli, E., Keihan, A. H., Shahdost-fard, F., Naghian, E., Plonska-Brzezinska, M. E., Rahimi-Nasrabadig, M., et al. (2020). A glassy carbon electrode modified with carbon nanoonions for electrochemical determination of fentanyl. Mater. Sci. Eng. C 110, 110684. doi:10.1016/j.msec.2020.110684

PubMed Abstract | CrossRef Full Text | Google Scholar

Stiller, R. L., Scierka, A. M., Davis, P. J., and Cook, D. R. (1990). A brief technical communication: Detection of fentanyl in urine. Forensic Sci. Int. 44, 1–6. doi:10.1016/0379-0738(90)90160-z

PubMed Abstract | CrossRef Full Text | Google Scholar

Ubaidullah, M., Ahmed, J., Ahamad, T., Shaikh, S. F., Alshehri, S. M., and Al-Enizi, A. M. (2020). Hydrothermal synthesis of novel nickel oxide@nitrogenous mesoporous carbon nanocomposite using costless smoked cigarette filter for high performance supercapacitor. Mat. Lett. 266, 127492. doi:10.1016/j.matlet.2020.127492

CrossRef Full Text | Google Scholar

Ubaidullah, M., Al-Enizi, A. M., Shaikh, S., Ghanem, M. A., and Mane, R. S. (2020). Waste PET plastic derived ZnO@NMC nanocomposite via MOF-5 construction for hydrogen and oxygen evolution reactions. J. King Saud Univ. - Sci. 32, 2397–2405. doi:10.1016/j.jksus.2020.03.025

CrossRef Full Text | Google Scholar

Vakili Fathabadi, M., Hashemipour Rafsanjani, H., Foroughi, M. M., Jahani, Sh., and Arefi Nia, N. (2020). Synthesis of magnetic ordered mesoporous carbons (OMC) as an electrochemical platform for ultrasensitive and simultaneous detection of thebaine and papaverine. J. Electrochem. Soc. 167, 027509. doi:10.1149/1945-7111/ab6446

CrossRef Full Text | Google Scholar

Vignesh, S., Suganthi, S., Srinivasan, M., Tamilmani, A., Sundar, J. K., Gedi, S., et al. (2022). Investigation of heterojunction between α-Fe2O3/V2O5 and g-C3N4 ternary nanocomposites for upgraded photo-degradation performance of mixed pollutants: Efficient dual Z-scheme mechanism. J. Alloys Compd. 902, 163705. doi:10.1016/j.jallcom.2022.163705

CrossRef Full Text | Google Scholar

Wester, N., Mynttinen, E., Etula, J., Lilius, T., Kalso, E., Mikladal, B. F., et al. (2020). Single-walled carbon nanotube network electrodes for the detection of fentanyl citrate. ACS Appl. Nano Mat. 3, 1203–1212. doi:10.1021/acsanm.9b01951

CrossRef Full Text | Google Scholar

Yang, H. M., Tao, Q. F., Zhang, X. C., Tang, A., and Ouyang, J. (2008). Solid-state synthesis and electrochemical property of SnO2/NiO nanomaterials. J. Alloys Compd. 459, 98–102. doi:10.1016/j.jallcom.2007.04.258

CrossRef Full Text | Google Scholar

You, H., Chen, Z., Yu, Q., Zhu, W., Chen, B., Lv, Z., et al. (2021). Preparation of a three-dimensional porous PbO2-CNTs composite electrode and study of the degradation behavior of p-nitrophenol. Sep. Purif. Technol. 276, 119406. doi:10.1016/j.seppur.2021.119406

CrossRef Full Text | Google Scholar

Zhou, Q., Umar, A., Sodki, E. M., Amine, A., Xu, L., Gui, Y., et al. (2018). Fabrication and characterization of highly sensitive and selective sensors based on porous NiO nanodisks. Sensors Actuators B Chem. 259, 604–615. doi:10.1016/j.snb.2017.12.050

CrossRef Full Text | Google Scholar

Zhou, Y., Cheng, X., Tynan, B., Shao, Z., Huang, F., Islam, M. S., et al. (2021). High-performance hierarchical MnO2/CNT electrode for multifunctional supercapacitors. Carbon 184, 504–513. doi:10.1016/j.carbon.2021.08.051

CrossRef Full Text | Google Scholar

Keywords: fentanyl, anesthesia drug, hexagonal NiO nanodisks, pencil graphite electrode, voltammetry

Citation: Li X, Luo B, Liao M and Mohamed A (2022) Electrochemical sensing of fentanyl as an anesthesia drug on NiO nanodisks combined with the carbon nanotube-modified electrode. Front. Chem. 10:997662. doi: 10.3389/fchem.2022.997662

Received: 19 July 2022; Accepted: 24 October 2022;
Published: 11 November 2022.

Edited by:

Pramod K. Kalambate, Chulalongkorn University, Thailand

Reviewed by:

Swamidoss Abraham John, The Gandhigram Rural Institute, India
Riyaz Ahmad Dar, University of Mumbai, India

Copyright © 2022 Li, Luo, Liao and Mohamed. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Min Liao, liaomin79@tom.com

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.