ORIGINAL RESEARCH article

Front. Cell. Infect. Microbiol., 08 August 2024
Sec. Clinical Microbiology
This article is part of the Research Topic Value of a multidisciplinary approach for modern diagnosis of infectious diseases View all 7 articles

Real-time fluorescent multiple cross displacement amplification for rapid and sensitive Mycoplasma pneumoniae detection

Fei XiaoFei Xiao1Yu ZhangYu Zhang1Wenjian XuWenjian Xu2Jin FuJin Fu1Xiaolan HuangXiaolan Huang1Nan JiaNan Jia1Chunrong SunChunrong Sun1Zheng XuZheng Xu1Baoying Zheng*Baoying Zheng3*Juan Zhou*Juan Zhou1*Yi Wang*Yi Wang1*Lihui Meng*Lihui Meng4*
  • 1Experiment Research Center, Capital Institute of Pediatrics, Beijing, China
  • 2Laboratory Center, Children’s Hospital Affiliated to the Capital Institute of Pediatrics, Beijing, China
  • 3Respiratory Medicine, Children’s Hospital Affiliated to the Capital Institute of Pediatrics, Beijing, China
  • 4Department of Infectious Diseases, Children’s Hospital Affiliated to Capital Institute of Pediatrics, Beijing, China

Mycoplasma pneumoniae is a significant pathogen responsible for community-acquired pneumonia, predominantly affecting children and adolescents. Here, we devised a rapid method for M. pneumoniae that combined multiple cross displacement amplification (MCDA) with real-time fluorescence technology. A set of ten primers, which were specifically designed for M. pneumoniae detection, were employed in a real-time fluorescence MCDA reaction. Of these, one primer incorporated a restriction endonuclease recognition sequence, a fluorophore, and a quencher, facilitating real-time fluorescence detection. The real-time (RT)-MCDA reactions were monitored in a simple real-time fluorescence instrument and conducted under optimised conditions (64°C for 40 min). The detection limit of the M. pneumoniae RT-MCDA assay for genomic DNA extracted from M. pneumoniae culture was down to 43 fg/µl. This assay accurately identified M. pneumoniae strains without cross-reacting with other bacteria. To validate its practical application, we tested the M. pneumoniae RT-MCDA assay using genomic DNA extracted from clinical samples. The assay’s detection capability proved comparable with real-time PCR, MCDA-based biosensor detection, and visual inspection under blue light. The entire process, including rapid DNA extraction and real-time MCDA detection, was completed within 1 h. Overall, the M. pneumoniae RT-MCDA assay reported here is a simple and effective diagnostic tool for rapid M. pneumoniae detection, which holds significant potential for point-of-care testing and in resource-limited regions.

Introduction

Mycoplasma pneumoniae is a prevalent cause of community-acquired pneumonia (Hammerschlag, 2001; Lee, 2008; Kumar and Kumar, 2023). This infectious agent can affect individuals of all ages but is most common among children and adolescents, with its incidence rising annually (Chang et al., 2014; Chen et al., 2024; Liu et al., 2024). Infections with M. pneumoniae typically manifest with respiratory symptoms but it can also lead to complications outside the respiratory system, such as cardiovascular and neurological complications (Kumar and Kumar, 2023). The clinical signs of M. pneumoniae infection are non-specific and coupled with diagnostic complexities, making early and rapid identification challenging (Hammerschlag, 2001; Kumar and Kumar, 2023). Therefore, improving the accuracy of M. pneumoniae detection is particularly important for clinical diagnosis and management.

There is no significant difference in clinical symptoms and imaging between M. pneumonia infection-associated pneumonia and pneumonia caused by other pathogens. Thus, the identification and diagnosis of M. pneumoniae infection mainly relies on laboratory diagnosis (Hammerschlag, 2001; Daxboeck et al., 2003; Kumar and Kumar, 2023). Currently, the main techniques for detecting M. pneumoniae are the culture-based method, serological tests, and polymerase chain reaction (PCR) (Kumar and Kumar, 2023). The culture-based method is regarded as the “gold standard” for confirming M. pneumoniae infections (Daxboeck et al., 2003; Chang et al., 2014; Kumar and Kumar, 2023). However, it is time-consuming (2–6 weeks for results), labour-intensive, and costly, limiting its practical use in clinical settings (Chang et al., 2014; Kumar and Kumar, 2023). Serological testing is commonly used due to its ease of use (Daxboeck et al., 2003; Chang et al., 2014). However, challenges in producing highly specific M. pneumoniae antibodies, due to numerous stop codons in the M. pneumoniae gene sequence and difficulties in recombinant antigen production via prokaryotic expression, affect the sensitivity and specificity of this method. Factors such as patient age, immune status, and the quality of test kits also impact results (Daxboeck et al., 2003; Chang et al., 2014). In addition, the requirement for paired serum samples taken 1–2 weeks apart makes serological testing less accessible and generally retrospective (Chang et al., 2014). A PCR-based detection assay is emerging as a rapid and reliable diagnostic method, often regarded as the new “gold standard” for detecting M. pneumoniae due to its high sensitivity and specificity (Chang et al., 2014; Park, 2022). However, its widespread adoption is hampered by the need for sophisticated equipment and skilled operators, particularly in primary healthcare settings (Chang et al., 2014; Park, 2022). To address these limitations presented by PCR-based methods, various isothermal amplification techniques have been developed since the early 1990s as alternatives to PCR (Zhao et al., 2015; Park, 2022; Srivastava and Prasad, 2023). These methods allow for the rapid and efficient amplification of nucleic acids at a constant temperature, offering advantages such as lower equipment costs, quicker results, simpler procedures, and greater specificity and sensitivity, making them suitable for immediate clinical use (Wang et al., 2015, 2016).

In recent years, the use of multiple cross displacement amplification (MCDA) technology within the realm of isothermal amplification has proven effective for detecting numerous pathogens (Zhao et al., 2019; Li et al., 2020; Jiang et al., 2022; Jia et al., 2023; Zhou et al., 2023). MCDA is a novel amplification strategy based on isothermal chain displacement polymerisation reactions, in which a set of 10 specific primers is designed for each target, spanning 10 different regions of the target sequence, achieving sensitivity in the femtogram range (Wang et al., 2015). At the operational temperatures (58°C–70°C), the dynamic interaction between primer-template hybridisation facilitates the annealing of a high concentration of primers onto the DNA template without requiring a denaturation phase, thus initiating synthesis (Wang et al., 2015). This leads to exponential amplification of the DNA products through chain displacement reactions. The MCDA reactions can be monitored through three primary methods: colorimetric analysis, agarose gel electrophoresis, and real-time turbidimetry. Therefore, the MCDA technique not only offers rapid results, high specificity, and sensitivity, but also has the added benefits of low equipment demands and simplicity in execution.

In this study, we successfully incorporated MCDA technology into the real-time PCR platform, creating a novel real-time MCDA detection method specifically tailored for M. pneumoniae. This approach enables direct monitoring of fluorescence signals and allows results to be read directly through the instrument without having to open the reaction tube, significantly reducing the risk of contamination. This rapid, sensitive, specific, and contamination-free method proves highly effective in diagnosing M. pneumoniae infection. During the establishment of this technique, we also assessed its clinical accuracy in detecting M. pneumoniae infection. The results of our study indicate that M. pneumoniae RT-MCDA is particularly useful in regions lacking dedicated product testing laboratories.

Materials and methods

Reagents and instruments

A DNA Isothermal Amplification Kit, biotin-14-dCTP, visual detection reagent (VDR), and a nanoparticle-based lateral flow biosensor (LFB) were obtained from Huidexin Biotech Co., Ltd (Tianjin, China). Restriction endonuclease (Nb.BsrDI) was purchased from New England BioLabs (Beijing, China). A nucleic acid extraction kit was obtained from Beijing Transgen Biotech Co., Ltd (Beijing, China) and real-time PCR kits were procured from HuNan SX Bio-Tech Co., Ltd (HuNan, China). The BluSight Pro (GD50502) was obtained from Manod Biotech. Co, Ltd (Suzhou, China). The real-time turbidimeter (LA-320C) was purchased from Eiken Chemical Co., Ltd, Japan. A fluorescence quantitative PCR instrument ABI 7500 was purchased from Applied Biosystem Inc., USA.

Primer design

A set of ten primers targeting the community-acquired respiratory distress syndrome (CARDS) toxin gene (LR214945.1) of M. pneumoniae were designed using primer premier 5.0 (https://primerexplorer.jp). The primer set includes two displacement primers (F1 and F2), two cross primers (CP1 and CP2), and six amplification primers (C1, C2, D1, D2, R1, and R2). The amplification primer D1 used in the M. pneumonia MCDA-LFB assay (termed D1#) was modified by assigning a fluorophore (FAM) at the 5' end, whereas the amplification primer C1 employed in the M. pneumoniae RT-MCDA assay (termed C1*) was modified by additionally adding a short sequence (Ss, TGCAATG) at the 5′ end and assigning a fluorophore and black hole quencher 1 (BHQ1) at the 5′ end and the middle of new primer. Primers and labelled primers used in this study were synthesised by AoKe Biotech Co., Ltd (Beijing, China). The sequences, location, and the modification of the primers are shown in Table 1 and Figure 1.

Table 1
www.frontiersin.org

Table 1 The primers used in the study.

Figure 1
www.frontiersin.org

Figure 1 Primers for the CARDS toxin gene used in the M. pneumoniae RT-MCDA assay in this study. The location of the primer sequences used in this study on the targeting CARDS toxin gene of M. pneumoniae. Right and left arrows show sense and complementary sequences, respectively. The coloured text indicates the position of the primers, including two displacement primers (F1 and F2), two cross primers (CP1 and CP2), and six amplification primers (C1, C2, D1, D2, R1, and R2).

DNA preparation

Genomic DNA of all the strains was extracted and purified using an EasyPure® Genomic DNA kit (TransGen Biotech, Beijing, China) according to the manufacturer’s instructions. Purity and concentration were determined using a Nanodrop 2000 (Thermo Fisher, Waltham, MA, USA) at A260/280. Before usage, the extracted DNA was kept at −20°C.

Standard M. pneumoniae -MCDA-LFB reaction

The volume of the MCDA reaction system was 25 μl, containing 12.5 μl of 2 × isothermal reaction buffer, 0.1 μmol/L each of displacement primers (F1 and F2), 0.4 μmol/L each of cross primers (CP1 and CP2), 0.2 μmol/L each of amplification primers [C1, C2, D1#(FAM labelled), D2, R1, and R2], 1.0 μl of Bst 2.0 DNA polymerase (8 U), 0.5 μl of biotin-14-dCTP, and 1 μl of DNA template from M. pneumoniae pure culture or 5 μl from clinical samples. In addition, 1 µl of DNA template of Klebsiella pneumoniae was employed as a negative control, and 1 µl of double distilled water (DW) was used as a blank control. The reaction mixtures were incubated at 64°C for 1 h using a real-time turbidimeter or traditional PCR instrument. The results were monitored using a real-time turbidimeter, VDR, and LFB test. For real-time turbidimeter, reaction with turbidity >0.1 was considered as positive, otherwise as negative. For VDR, reaction with a light blue colour was deemed as positive, while the colourless one was negative. For result interpretation by LFB, 5 µl of MCDA reaction products were added to the sample pad of LFB, followed by 100 μl of running buffer [10 mM PBS (PH 7.4) with 1% Tween 20]. The results were indicated within 2 min, with two red lines at the control line (CL) and testing line (TL) representing a positive result and one red line at the CL representing a negative result. In parallel, results of the M. pneumoniae MCDA reaction were real-time monitored by real-time turbidimeter and visually inspected by naked eye with VDR.

Standard M. pneumoniae RT-MCDA reaction

The RT-MCDA assay was conducted in a 25 μl reaction system consisting of 12.5 μl of buffer, 0.1 μmol/L each of displacement primers (F1 and F2), 0.4 μmol/L each of cross primers (CP1and CP2), 0.2 μmol/L each of amplification primers [C1* (restriction endonuclease recognition sequence, FAM and BHQ1 labelled), C2, D1, D2, R1, and R2], 1.0 μl of Bst DNA polymerase (8 U), 1.0 μl of Nb.BsrDI restriction endonuclease, and 1 μl of template from pure culture (5 μl from the clinical sample). The reaction was carried out using real-time PCR instruments for 40 cycles with a setting mode of 5 s at 64°C for reaction and 55 s at 64°C for fluorescence monitoring. Reaction with apparent fluorescence generation indicated a positive result, while that with trace or none fluorescent signal implied a negative one. The positive threshold value was obtained by testing plenty of positive, negative, and blank reactions.

Optimisation of the M. pneumoniae RT-MCDA assay

The optimal reaction temperature for the M. pneumoniae MCDA reaction was determined by performing an M. pneumoniae MCDA reaction at temperatures from 60 to 67°C with a 1°C interval for 40 min followed by incubation at 85°C to stop the reaction. The amplification process was real-time monitored using a real-time turbidimeter. DNA templates from M. pneumoniae were used as a positive control and DW as a blank control. The temperature generating the best reaction performance was considered as the optimal one. Meanwhile, assays with different reaction times from 10 min to 40 min (with 10 min intervals) were conducted to determine the optimal reaction time. The shortest time required for the MCDA reaction was regarded as the optimal reaction time. All the tests were repeated three times. The obtained optimum reaction temperature and time were employed for the following tests.

Sensitivity and specificity evaluation of the M. pneumoniae RT-MCDA assay

To determine the sensitivity of the M. pneumoniae RT-MCDA assay, DNA templates from cultured M. pneumoniae were serially diluted to 4.3 ng/μl, 430 pg/μl, 43 pg/μl, 4.3 pg/μl, 430 fg/μl, 43 fg/μl, and 4.3 fg/μl. Accordingly, 1 μl or 2 μl each of the serial templates was added into the M. pneumoniae RT-MCDA reaction systems to test the limit of detection (LoD) of the assay. Each reaction was repeatedly tested in triplicate. For the specificity assessment of the M. pneumoniae RT-MCDA assay, DNA templates from three M. pneumoniae strains and 29 non-M. pneumoniae strains (Table 2) were examined by this assay with three replicates. Similarly, all the samples were tested by the MCDA-LFB assay for comparison.

Table 2
www.frontiersin.org

Table 2 Strains for specificity confirmation of M. pneumoniae RT-MCDA assay.

Clinical application assessment of the M. pneumoniae RT-MCDA assay

For evaluation of the clinical feasibility, DNA templates extracted from 48 sputum samples were retrospectively tested by the M. pneumoniae RT-MCDA assay and MCDA-LFB test. Particularly, 1 µl, 3 µl, 5 µl, and 7 µl of clinical sample template were examined separately to identify the appropriate volume for clinical evaluation. The 48 samples were collected from children hospitalised in the Children’s Hospital Affiliated to Capital Institute of Pediatrics from 12 October 2023 to 20 December 2023 and had been examined using the real-time PCR method for clinical diagnosis. All the samples were obtained with informed consent signed by the guardians of the participants.

Results

Principle of the M. pneumoniae RT-MCDA assay

The M. pneumoniae RT-MCDA assay was devised by combining an MCDA reaction with restriction endonuclease cleavage, which will generate fluorescent readouts for result interpretation. As illustrated in Figure 2, using primers (C1* in this study) adding a short sequence 5′-TGCAATG-3′, which could be specifically recognised and cleaved by restriction endonuclease Nb.BsrDI, the RT-MCDA reaction will produce plenty of products containing the Nb.BsrDI recognition sequence. Once the generated products were cleaved by Nb.BsrDI, fluorescent signals were yielded due to the separation of FAM and BHQ1, which flanked the Nb.BsrDI recognition sequence. It is worth mentioning that owing to the addition of the short sequence (Ss, TGCAATG) at the 5′ end of the amplification primer C1 (C1*), Nb.BsrDI accurately identified the target double-stranded DNA sequences and cut them to generate fluorescence signals, which will not be interfered with by the excess (single-stranded) primer cleavage. The fluorescent signals could be visually inspected with the naked eye under blue light in addition to a real-time PCR instrument. Thus, the presence or absence of target pathogen will be reported only with the naked eye rather than complicated instruments.

Figure 2
www.frontiersin.org

Figure 2 Schematic illustration of the M. pneumoniae RT-MCDA assay. In the M. pneumoniae RT- MCDA system, the primer C1* was modified by adding a short sequence (TGCAATG) at 5′ end that could be recognised by restriction endonuclease Nb.BsrDI. When reacted with C1*, the M. pneumoniae real-time MCDA system generated plenty of Ss-containing target amplicons that were then cleaved by restriction endonuclease Nb.BsrDI, which could separate the fluorophore FAM from BHQ1, resulting in the emission of fluorescence signals that could be visually inspected with the naked eye under blue light, in addition to a real-time PCR instrument.

In this study, an M. pneumoniae-MCDA-LFB assay was conducted in parallel for comparison. The principle of the MCDA-LFB assay is shown in Figure 3. Unlike the RT-MCDA assay, the MCDA-LFB assay was devised on the basis of the MCDA reaction and LFB detection. Using a FAM-labelled primer and biotin-14-dCTP, the MCDA reaction could produce abundant FAM/biotin-labelled amplicons, which could be transformed into a colorimetric readout with LFB. Particularly, the immobilised anti-FAM of LFB captured FAM-containing target amplification products and SA-GNPs (streptavidi-coated dyed polymer nanoparticles) enabled products containing biotin to be visualised, resulting in a red line occurring in the TL region of LFB. The remaining SA-GNPs were captured by the biotinylated bovine serum albumin (biotin-BSA) in the CL region, resulting in a red line in the CL region, indicating the effectiveness of the LFB (Figure 3).

Figure 3
www.frontiersin.org

Figure 3 Schematic illustration of the M. pneumoniae-MCDA-LFB assay. In the M. pneumoniae -MCDA-LFB system, anti-FAM immobilised on LFB can capture the double-labelled amplicon (produced by FAM-labelled amplification primer D1# and biotin-14-dCTP) in the reaction and is visualised by the reaction of biotin and SA-GNPs (dark red), resulting in a red line occurring in the TL region of the LFB. The remaining SA‐GNPs were captured by the immobilised biotin‐BSA at the CL region, leading to a red line occurring in the CL region, which indicated the effectiveness of the LFB.

Effectiveness of the primer set for the M. pneumoniae RT-MCDA assay

The effectiveness of the primer set (Figure 1, Table 1) for the M. pneumoniae MCDA assay was confirmed by performing an MCDA reaction with the selected primer set and monitoring results using three methods (real-time turbidity, VDR, and LFB). Using a real-time turbidimeter, the positive reaction (the reactions of M. pneumoniae) displayed a sharp increase in turbidity, whereas the negative (the reaction of K. pneumoniae) and blank control (the reaction of DW) reactions did not generate any turbidity (Figure 4A). Using VDR, the colour of the positive control changed into light green, whereas the others were colourless (Figure 4B). With LFB, two visible red lines in the CL and TL regions were observed with the positive control products, whereas only one line in CL was observed when the negative and blank control products were examined (Figure 4C). Thus, the primer set was determined to establish the M. pneumoniae RT-MCDA assay.

Figure 4
www.frontiersin.org

Figure 4 Effectiveness of the primer set for the M. pneumoniae RT-MCDA reaction. The DNA templates extracted from M. pneumoniae strains were effectively amplified with MCDA reactions at 64°C for 1 h, and there was no obvious amplification in the negative and blank controls. The products were indicated by real-time turbidity (A), the VDR (B), and the nanoparticle-based LFB test (C), and observed with the naked eye under blue light (D). Tubes/strips 1–2 represent DNA templates extracted from M. pneumoniae strains. Tube/strip 3 represents DNA template extracted from a Klebsiella pneumoniae strain, Tube/strip 4 represents the blank control. CL, control line; TL, test line.

Subsequently, the primer set was used for M. pneumoniae RT-MCDA assay confirmation. As shown in Figure 4D, positive reactions released fluorescent signals that showed an obvious fluorescence value in the real-time PCR platform and a light yellow colour under blue light, whereas the negative and blank controls did not present a fluorescent signal either in the real-time PCR platform or under blue light. These data indicated that the selected primer set was suitable for the M. pneumoniae RT-MCDA assay and M. pneumoniae detection.

Optimal reaction condition for the M. pneumoniae RT-MCDA assay

Temperatures ranging from 60°C to 67°C with a 1°C interval were employed to conduct the MCDA reactions for the optimisation of the reaction temperature. As shown in Figure 5, 64°C was the optimal reaction temperature for the primer set, as the faster peak was obtained under this condition. Therefore, 64°C was subsequently used as the optimal temperature for the M. pneumoniae MCDA assay. In addition, the M. pneumoniae MCDA assay was conducted at 64°C for 10 min, 20 min, 30 min, and 40 min to reveal the optimal reaction time. As shown in Figure 6, it was concluded that the LoD level (determined by sensitivity analysis as 43 fg/µl) of the genomic DNA of M. pneumoniae could be detected only after amplification at 64°C for 40 min. Thus, a reaction temperature of 64°C and a reaction time of 40 min were employed in the following tests, and the whole diagnostic procedure, from template extraction to result reporting, was finished within 1 h (Figure 2).

Figure 5
www.frontiersin.org

Figure 5 Temperature optimisation for the M. pneumoniae RT-MCDA assay. MCDA reactions detecting M. pneumoniae were conducted using a real-time turbidimeter. Turbidity >0.1 was considered a positive result. The kinetic curves at different temperatures ranging from 60 to 67°C (A–H) were acquired, showing that 64°C was the optimal temperature.

Figure 6
www.frontiersin.org

Figure 6 Time optimisation for the M. pneumoniae RT-MCDA assay. The standard MCDA reaction with four distinct times, 10 min, 20 min, 30 min, and 40 min, was tested and analysed by VDR (A), LFB (B), and the naked eye under blue light (C) at the optimal temperature of 64°C. The LoD level of M. pneumoniae genomic DNA could be detected only after amplification for 40 min. Strips/tubes 1–7, DNA templates extracted from serial dilutions (4.3 ng–4.3 fg/µl); strip/tube 8, blank control. CL, control line; TL, test line.

Sensitivity and specificity of the M. pneumoniae RT- MCDA assay

To assess the sensitivity of the M. pneumoniae RT-MCDA assay, an RT-MCDA assay was performed with serial dilutions of genomic DNA templates from M. pneumoniae strains. The detection limits of the M. pneumoniae RT-MCDA assay with 1 µl or 2 µl of pure culture template were both 43 fg/µl (Figures 7A, B), which was in complete accordance with those observed with the naked eye under blue light (Figure 7C) and that when using an MCDA-LFB assay (Figure 7D). Considering the risk of carryover contamination, a reaction volume of 1 µl was recommended.

Figure 7
www.frontiersin.org

Figure 7 Sensitivity confirmation of the M. pneumoniae RT-MCDA assay. Sensitivity of the M. pneumoniae RT-MCDA assay was analysed using DNA templates extracted from serial dilutions (4.3 ng, 430 pg, 43 pg, 4.3 pg, 430 fg, 43 fg, and 4.3 fg/µl) of cultured M. pneumoniae. Three detection formats were used for amplification products analysis: (A) real-time fluorescence with 1 µl of template; (B) real-time fluorescence with 2 µl of template; (C) visual detection with the naked eye under blue light; and (D) an LFB test. The results of the three tests showed that the lowest detection limit of the M. pneumoniae RT-MCDA was 43 fg/µl. The sensitivity was also confirmed through visual detection with the naked eye under blue light (C) and an LFB test (D). TL, test line; CL, control line.

To estimate the analytical specificity, the M. pneumoniae RT-MCDA assay was conducted using genomic DNA templates extracted from M. pneumoniae strains and non-M. pneumoniae strains (Table 2). When analysed using a real-time fluorometer, the positive results were only acquired from the reactions with DNA templates of M. pneumoniae strains; the non-M. pneumoniae strains were negative (Figure 8A). When observed under blue light, a light yellow colour was only observed with the products of three M. pneumoniae strains and not with those of the 29 non-M. pneumoniae pathogens (Figure 8B). When monitored by LFB, the three M. pneumoniae strains displayed two red lines at both TL and CL of LFB, whereas all the 29 non-M. pneumoniae pathogens displayed only one red line in CL of LFB (Figure 8C). All the above results proved that the specificity of the M. pneumoniae RT-MCDA assay was 100%.

Figure 8
www.frontiersin.org

Figure 8 Analytical specificity of the M. pneumoniae RT-MCDA assay. DNA templates from three M. pneumoniae strains and 29 non-M. pneumoniae strains were tested using the M. pneumoniae RT-MCDA assay. No reactions were recorded with the non-M. pneumoniae strains. The generated fluorescence signals were recorded by the real-time PCR platform (A) and visually interpreted with the naked eye under blue light (B). The results were also confirmed by an LFB test (C). Signals/tubes/strips 1–3 represent the three M. pneumoniae strains, and the others represent the non-M. pneumoniae strains (Table 2). TL, test line; CL, control line.

Application of the M. pneumoniae RT-MCDA assay in clinical specimens

To further validate the feasibility of the M. pneumoniae RT-MCDA assay in clinical application, the optimised process was used to detect the retrospectively collected DNA templates extracted from 48 sputum samples, which were previously detected using the real-time PCR method. Among the 48 samples, 26 tested positive for the M. pneumoniae RT-MCDA assay and 22 samples tested negative, regardless of whether the reaction volume was 1 µl, 3 µl, 5 µl, or 7 µl (Figures 9A-D). However, reactions with 1 µl and 3 µl of clinical sample template displayed more threshold results, which may be difficult to read with the naked eye, thus 5 µl of clinical sample template was selected for clinical evaluation analysis. Of note, the results obtained by the M. pneumoniae RT-MCDA assay (with 5 µl of template) (Figures 9C, E) were identical to that obtained by the M. pneumoniae MCDA-LFB assay (Figure 9F). In terms of the results obtained using the real-time PCR method, 26 positive specimens were previously determined positive and 22 specimens were negative. The complete accordance between the M. pneumoniae RT-MCDA assay and real-time PCR method implied the feasibility of the M. pneumoniae RT-MCDA assay in clinical practice.

Figure 9
www.frontiersin.org

Figure 9 Application of the M. pneumoniae RT- MCDA assay in clinical specimens. DNA templates from 48 sputum samples were detected by the M. pneumoniae RT-MCDA assay. The produced fluorescent signals were documented by the real-time PCR system with 1 µl of template (A), 3 µl of template (B), 5 µl of template (C), and 7 µl of template (D), and visually interpreted with the naked eye under blue light (E). The results were also confirmed by M. pneumoniae-MCDA-LFB (F). Signals/tubes/strips 1–26 represent the results of the templates from the positive samples, and tubes/strips 27–48 represent the results of the M. pneumoniae negative samples. TL, test line; CL, control line.

Discussion

In recent years, the global surge in macrolide resistance among M. pneumoniae strains has been particularly pronounced in China, where resistance levels have approached nearly 90%. This alarming trend has led to an increase in cases of severe pneumonia among patients who do not receive timely and effective treatment, posing a serious threat to public health (Kumar and Kumar, 2023; Chen et al., 2024). Furthermore, M. pneumoniae infections are characterised by non-specific clinical and imaging features, along with limited microbial diagnostic options, complicating the clinical diagnosis and increasing the likelihood of missing optimal treatment windows. This often exacerbates the patient’s disease burden (Kumar and Kumar, 2023). Consequently, there is a critical need for rapid and precise detection methods to identify and diagnose M. pneumoniae infections early and effectively.

Numerous isothermal amplification techniques have been developed for diagnosing pathogen infections, including nucleic acid sequence-based amplification, self-sustained sequence replication, strand displacement amplification, exponential amplification reaction, helicase-dependent amplification, recombinase polymerase amplification, single primer isothermal amplification, rolling circle amplification, loop-mediated isothermal amplification (LAMP), and cross priming amplification (Wang et al., 2016). However, these methods face technical challenges and do not entirely fulfill the point-of-care detection standards set by the World Health Organization (Srivastava and Prasad, 2023). MCDA is a relatively recent advancement in this area and addresses many limitations of earlier techniques (Wang et al., 2015). MCDA has gained broad application in fields such as basic research, medical diagnostics, and food safety (Qi et al., 2018; Boonbanjong et al., 2022; Li et al., 2022). Here, we introduce a novel real-time fluorescence MCDA method for M. pneumoniae that integrates real-time PCR technology with MCDA for efficient clinical detection. This approach uses a set of 10 sequence-specific primers that target 10 distinct regions (F1, F2, C1, C2, P1, P2, R1, R2, D1, and D2), ensuring high specificity (Wang et al., 2015, 2016). Critically, our technique employs the 5′ endonuclease recognition sites and fluorophore labelling to achieve precise cleavage of double-stranded DNA by the restriction enzyme Nb.BsrDI, releasing a fluorescence signal. Thus, the new real-time fluorescence MCDA method enables rapid multiplex detection and identification in a single isothermal step without the need for thermal cycling or subsequent analysis (Wang et al., 2015, 2016; Srivastava and Prasad, 2023). Additionally, the early high expression of the CARDS toxin gene during infection and the critical role of CARDS toxin in the pathogenesis of M. pneumoniae facilitate the CARDS toxin gene as a preferable biomarker for early and accurate M. pneumoniae detection (Chaudhry et al., 2016; Su et al., 2021). It was reported that detection assays targeting the CARDS toxin gene exhibited superior sensitivity compared with those targeting other genes such as the ATPase (Mp3 and Mp7) gene (Winchell et al., 2008). Thus, the real-time MCDA targeting the CARDS toxin gene in this study offers a more timely and efficient diagnosis, increasing early detection capabilities. In addition, a real-time PCR method targeting the P1 gene was used for clinical feasibility comparison.

In this study, we employed the restriction enzyme Nb.BsrDI and a modified primer, C1*, to detect MCDA reaction products through fluorescence. Additionally, another modified primer, D1#, and biotin-14-dCTP were used in a local feedback mechanism to monitor these products. We evaluated the sensitivity and specificity of these two methods through detailed analytical comparisons, finding both to be equally effective in detecting M. pneumoniae. Furthermore, the optimised real-time fluorescence MCDA detection process we developed can complete the identification of M. pneumoniae within 1 h, allowing for prompt treatment and control of the infection. Given its efficiency and reliability, the real-time MCDA detection method introduced in this study is highly recommended for the early diagnosis and management of M. pneumoniae infections.

The newly developed RT-MCDA assay demonstrated high sensitivity in detecting M. pneumoniae, with a LoD of 43 fg (approximately 47 copies) genomic DNA per reaction. This sensitivity is comparable with that achieved by real-time PCR (approximately 10 copies) (Morozumi et al., 2006) and slightly higher than that achieved by LAMP assays (50–600 fg) (Saito et al., 2005; Zhao et al., 2013; Wang et al., 2019; Xiao et al., 2022). A comparison of different methods for M. pneumoniae infection diagnosis is summarised in Table 3 (Daxboeck et al., 2003; Templeton et al., 2003; Saito et al., 2005; Morozumi et al., 2006; Loens et al., 2010; Zhao et al., 2013; Wang et al., 2019; Xiao et al., 2022). Table 3 conclusively shows that the RT-MCDA assay had superior sensitivity and timeliness but a relatively higher cost in M. pneumoniae detection; the LAMP-based assay displayed moderate sensitivity and timeliness as well as a relatively high cost; the real-time PCR method exhibited superior sensitivity and moderate timeliness and cost; the serological test displayed poor sensitivity but was superior in timeliness and cost; and the culture method showed poor sensitivity and timeliness but was the least costly. As a conclusion, the RT-MCDA assay was a superior option for M. pneumoniae detection with a further optimisation of sensitivity and cost. Moreover, the RT-MCDA assay exhibited no cross-reactivity with non-M. pneumoniae pathogens, achieving a specificity of 100%. To assess the clinical applicability of this method, we analysed 48 suspect samples, confirming M. pneumoniae in 26 cases (54.17%)—a rate consistent with that of real-time PCR. Although there were no significant statistical differences between the RT-MCDA method and conventional real-time PCR, further validation with a larger sample set is warranted. Additionally, the outcomes of the M. pneumoniae RT-MCDA assay are not only trackable through a real-time fluorescence monitoring system but can also be visually confirmed under blue light, significantly reducing the contamination risks typically associated with opening the reaction container. Given its robust performance and operational flexibility, this novel method holds promise for its widespread use in clinical diagnostics, field testing, and resource-constrained environments.

Table 3
www.frontiersin.org

Table 3 Sensitivity, time, and cost comparison of different methods for M. pneumoniae detection.

In summary, this study successfully developed a real-time MCDA detection method for M. pneumoniae that offers timely, simple, and reliable testing. This novel assay represents a promising tool for the rapid, sensitive, and specific identification of M. pneumoniae infections and stands as a viable option for increasing diagnostic capabilities.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding authors.

Ethics statement

The studies involving human participants were reviewed and approved by the ethic committee of Capital Institute of Pediatrics (Ethical approval number: SHERLLM2021010). The patients/participants [legal guardian/next of kin] provided written informed consent to participate in the study.

Author contributions

FX: Writing – original draft, Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Validation, Writing – review & editing. YZ: Data curation, Formal Analysis, Writing – original draft. WX: Conceptualization, Writing – original draft. JF: Project administration, Writing – review & editing. XH: Project administration, Writing – review & editing. NJ: Formal Analysis, Writing – review & editing. CS: Software, Writing – review & editing. ZX: Data curation, Writing – original draft. BZ: Conceptualization, Resources, Writing – review & editing. JZ: Formal Analysis, Supervision, Writing – review & editing. YW: Data curation, Formal Analysis, Funding acquisition, Methodology, Project administration, Resources, Software, Writing – review & editing. LM: Project administration, Resources, Supervision, Validation, Writing – review & editing.

Funding

The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. This study was funded by Beijing Nova Program (Grant No. Z211100002121042), National Natural Science Foundation of China (82200115), National Key Research and Development Program of China (Grant No. 2021YFC2301101 and 2021YFC2301102), Special Innovation and Promotion Project of Beijing Medical Administration (Grant No. XTCX201820) and the Research Foundation of Capital Institute of Pediatrics (Grant No. PX2021050).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Boonbanjong, P., Treerattrakoon, K., Waiwinya, W., Pitikultham, P., Japrung, D. (2022). Isothermal amplification technology for disease diagnosis. Biosensors (Basel) 12(9):677. doi: 10.3390/bios12090677

PubMed Abstract | CrossRef Full Text | Google Scholar

Chang, H. Y., Chang, L. Y., Shao, P. L., Lee, P. I., Chen, J. M., Lee, C. Y., et al. (2014). Comparison of real-time polymerase chain reaction and serological tests for the confirmation of Mycoplasma pneumoniae infection in children with clinical diagnosis of atypical pneumonia. J. Microbiol. Immunol. Infect. 47, 137–144. doi: 10.1016/j.jmii.2013.03.015

PubMed Abstract | CrossRef Full Text | Google Scholar

Chaudhry, R., Ghosh, A., Chandolia, A. (2016). Pathogenesis of Mycoplasma pneumoniae: An update. Indian J. Med. Microbiol. 34, 7–16. doi: 10.4103/0255-0857.174112

PubMed Abstract | CrossRef Full Text | Google Scholar

Chen, Y., Zhang, Y., Tang, Q. N., Shi, H. B. (2024). Efficacy of doxycycline therapy for macrolide-resistant Mycoplasma pneumoniae pneumonia in children at different periods. Ital. J. Pediatr. 50, 38. doi: 10.1186/s13052-024-01615-y

PubMed Abstract | CrossRef Full Text | Google Scholar

Daxboeck, F., Krause, R., Wenisch, C. (2003). Laboratory diagnosis of Mycoplasma pneumoniae infection. Clin. Microbiol. Infect. 9, 263–273. doi: 10.1046/j.1469-0691.2003.00590.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Hammerschlag, M. R. (2001). Mycoplasma pneumoniae infections. Curr. Opin. Infect. Dis. 14, 181–186. doi: 10.1097/00001432-200104000-00012

PubMed Abstract | CrossRef Full Text | Google Scholar

Jia, N., Wang, C., Liu, X., Huang, X., Xiao, F., Fu, J., et al. (2023). A CRISPR-Cas12a-based platform for ultrasensitive rapid highly specific detection of Mycobacterium tuberculosis in clinical application. Front. Cell Infect. Microbiol. 13. doi: 10.3389/fcimb.2023.1192134

CrossRef Full Text | Google Scholar

Jiang, L., Gu, R., Li, X., Song, M., Huang, X., Mu, D. (2022). Multiple Cross Displacement Amplification Coupled with Lateral Flow Biosensor (MCDA-LFB) for rapid detection of Legionella pneumophila. BMC Microbiol. 22, 20. doi: 10.1186/s12866-021-02363-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Kumar, S., Kumar, S. (2023). Mycoplasma pneumoniae: Among the smallest bacterial pathogens with great clinical significance in children. Indian J. Med. Microbiol. 46, 100480. doi: 10.1016/j.ijmmb.2023.100480

PubMed Abstract | CrossRef Full Text | Google Scholar

Lee, K. Y. (2008). Pediatric respiratory infections by Mycoplasma pneumoniae. Expert Rev. Anti Infect. Ther. 6, 509–521. doi: 10.1586/14787210.6.4.509

PubMed Abstract | CrossRef Full Text | Google Scholar

Li, S., Liu, C., Liu, Y., Ma, Q., Wang, Y., Wang, Y. (2020). Establishment and application of a multiple cross displacement amplification combined with nanoparticles-based biosensor method for the detection of Bordetella pertussis. BMC Microbiol. 20, 263. doi: 10.1186/s12866-020-01945-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Li, X., Zhang, X., Shi, X., Shi, H., Wang, Z., Peng, C. (2022). Review in isothermal amplification technology in food microbiological detection. Food Sci. Biotechnol. 31, 1501–1511. doi: 10.1007/s10068-022-01160-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Liu, X., Zhang, Q., Chen, H., Hao, Y., Zhang, J., Zha, S., et al. (2024). Comparison of the clinical characteristics in parents and their children in a series of family clustered Mycoplasma pneumoniae infections. BMC Pulm. Med. 24, 107. doi: 10.1186/s12890-024-02922-0

PubMed Abstract | CrossRef Full Text | Google Scholar

Loens, K., Goossens, H., Ieven, M. (2010). Acute respiratory infection due to Mycoplasma pneumoniae: current status of diagnostic methods. Eur. J. Clin. Microbiol. Infect. Dis. 29, 1055–1069. doi: 10.1007/s10096-010-0975-2

PubMed Abstract | CrossRef Full Text | Google Scholar

Morozumi, M., Ito, A., Murayama, S. Y., Hasegawa, K., Kobayashi, R., Iwata, S., et al. (2006). Assessment of real-time PCR for diagnosis of Mycoplasma pneumoniae pneumonia in pediatric patients. Can. J. Microbiol. 52, 125–129. doi: 10.1139/w05-118

PubMed Abstract | CrossRef Full Text | Google Scholar

Park, J. W. (2022). Principles and applications of loop-mediated isothermal amplification to point-of-care tests. Biosensors (Basel) 12 (10), 857. doi: 10.3390/bios12100857

PubMed Abstract | CrossRef Full Text | Google Scholar

Qi, H., Yue, S., Bi, S., Ding, C., Song, W. (2018). Isothermal exponential amplification techniques: From basic principles to applications in electrochemical biosensors. Biosens. Bioelectron. 110, 207–217. doi: 10.1016/j.bios.2018.03.065

PubMed Abstract | CrossRef Full Text | Google Scholar

Saito, R., Misawa, Y., Moriya, K., Koike, K., Ubukata, K., Okamura, N. (2005). Development and evaluation of a loop-mediated isothermal amplification assay for rapid detection of Mycoplasma pneumoniae. J. Med. Microbiol. 54, 1037–1041. doi: 10.1099/jmm.0.46071-0

PubMed Abstract | CrossRef Full Text | Google Scholar

Srivastava, P., Prasad, D. (2023). Isothermal nucleic acid amplification and its uses in modern diagnostic technologies. 3 Biotech. 13, 200. doi: 10.1007/s13205-023-03628-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Su, X., You, X., Luo, H., Liang, K., Chen, L., Tian, W., et al. (2021). Community-acquired respiratory distress syndrome toxin: unique exotoxin for M. pneumoniae. Front. Microbiol. 12. doi: 10.3389/fmicb.2021.766591

CrossRef Full Text | Google Scholar

Templeton, K. E., Scheltinga, S. A., Graffelman, A. W., Van Schie, J. M., Crielaard, J. W., Sillekens, P., et al. (2003). Comparison and evaluation of real-time PCR, real-time nucleic acid sequence-based amplification, conventional PCR, and serology for diagnosis of Mycoplasma pneumoniae. J. Clin. Microbiol. 41, 4366–4371. doi: 10.1128/JCM.41.9.4366-4371.2003

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Wang, Y., Jiao, W., Li, J., Quan, S., Sun, L., et al. (2019). Development of loop-mediated isothermal amplification coupled with nanoparticle-based lateral flow biosensor assay for Mycoplasma pneumoniae detection. AMB Express 9, 196. doi: 10.1186/s13568-019-0921-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Wang, Y., Ma, A. J., Li, D. X., Luo, L. J., Liu, D. X., et al. (2015). Rapid and sensitive isothermal detection of nucleic-acid sequence by multiple cross displacement amplification. Sci. Rep. 5, 11902. doi: 10.1038/srep11902

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Wang, Y., Zhang, L., Liu, D., Luo, L., Li, H., et al. (2016). Multiplex, rapid, and sensitive isothermal detection of nucleic-acid sequence by endonuclease restriction-mediated real-time multiple cross displacement amplification. Front. Microbiol. 7. doi: 10.3389/fmicb.2016.00753

CrossRef Full Text | Google Scholar

Winchell, J. M., Thurman, K. A., Mitchell, S. L., Thacker, W. L., Fields, B. S. (2008). Evaluation of three real-time PCR assays for detection of Mycoplasma pneumoniae in an outbreak investigation. J. Clin. Microbiol. 46, 3116–3118. doi: 10.1128/JCM.00440-08

PubMed Abstract | CrossRef Full Text | Google Scholar

Xiao, F., Zhou, J., Sun, C., Huang, X., Zheng, B., Fu, J., et al. (2022). Loop-mediated isothermal amplification coupled with nanoparticle-based biosensor: A rapid and sensitive method to detect mycoplasma pneumoniae. Front. Cell Infect. Microbiol. 12. doi: 10.3389/fcimb.2022.882855

CrossRef Full Text | Google Scholar

Zhao, F., Liu, Z., Gu, Y., Yang, Y., Xiao, D., Tao, X., et al. (2013). Detection of Mycoplasma pneumoniae by colorimetric loop-mediated isothermal amplification. Acta Microbiol. Immunol. Hung 60, 1–9. doi: 10.1556/AMicr.60.2013.1.1

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhao, F., Niu, L., Yan, L., Nong, J., Wang, C., Wang, J., et al. (2019). Establishment and application of multiple cross displacement amplification coupled with lateral flow biosensor (MCDA-LFB) for visual and rapid detection of candida albicans in clinical samples. Front. Cell Infect. Microbiol. 9. doi: 10.3389/fcimb.2019.00102

CrossRef Full Text | Google Scholar

Zhao, Y., Chen, F., Li, Q., Wang, L., Fan, C. (2015). Isothermal amplification of nucleic acids. Chem. Rev. 115, 12491–12545. doi: 10.1021/acs.chemrev.5b00428

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhou, J., Xiao, F., Fu, J., Jia, N., Huang, X., Sun, C., et al. (2023). Rapid detection of monkeypox virus by multiple cross displacement amplification combined with nanoparticle-based biosensor platform. J. Med. Virol. 95, e28479. doi: 10.1002/jmv.28479

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: Mycoplasma pneumoniae, real-time detection, multiple cross displacement amplification, restriction endonuclease, rapid diagnosis

Citation: Xiao F, Zhang Y, Xu W, Fu J, Huang X, Jia N, Sun C, Xu Z, Zheng B, Zhou J, Wang Y and Meng L (2024) Real-time fluorescent multiple cross displacement amplification for rapid and sensitive Mycoplasma pneumoniae detection. Front. Cell. Infect. Microbiol. 14:1423155. doi: 10.3389/fcimb.2024.1423155

Received: 25 April 2024; Accepted: 19 July 2024;
Published: 08 August 2024.

Edited by:

Stefano Stracquadanio, University of Catania, Italy

Reviewed by:

Dimitra K. Toubanaki, Pasteur Hellenic Institute, Greece
Jun He, University of South China, China

Copyright © 2024 Xiao, Zhang, Xu, Fu, Huang, Jia, Sun, Xu, Zheng, Zhou, Wang and Meng. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Yi Wang, wildwolf0101@163.com; Juan Zhou, zhoujuan2015@126.com; Baoying Zheng, zbaoy1981@163.com; Lihui Meng, mengmlh@163.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.