Establishment and Application of Multiple Cross Displacement Amplification Coupled With Lateral Flow Biosensor (MCDA-LFB) for Visual and Rapid Detection of Candida albicans in Clinical Samples

\r\nFan Zhao&#x;Fan Zhao1Lina Niu,,&#x;Lina Niu2,3,4Linlin Yan&#x;Linlin Yan1Jinqing NongJinqing Nong1Chunmei WangChunmei Wang1Jing WangJing Wang1Naishu GaoNaishu Gao1Xiaoxue ZhuXiaoxue Zhu1Lei WuLei Wu1Fengzhi ZhengFengzhi Zheng1Shoukui Hu*Shoukui Hu1*
  • 1Department of Clinical Laboratory, Peking University Shougang Hospital, Beijing, China
  • 2Department of Pathogen Biology, School of Basic Medicine and Lifescience, Hainan Medical University, Haikou, China
  • 3Key Laboratory of Translation Medicine Tropical Diseases, Haikou, China
  • 4Hainan Medical University-University of Hong Kong Joint Laboratory of Tropical Infectious Diseases, Haikou, China

Candida albicans is an opportunistic pathogenic yeast that predominantly causes invasive candidiasis. The conventional diagnosis of C. albicans infection depends on time-consuming, culture-based gold-standard methods. Here, a multiple cross displacement amplification (MCDA) assay, combined with a gold nanoparticle-based lateral flow biosensor (LFB) visualization method, was developed for the rapid detection of C. albicans. The internal transcribed spacer II, a region between 5.8 and 28 S fungal ribosomal DNA, is a C. albicans species-specific sequence that was used as the MCDA assay target. As an isothermal amplification method, the MCDA reaction with optimized conditions could be completed within only 40 min at a constant temperature (64°C). Then, the amplification reaction products could be visibly detected by a LFB without special equipment. The developed MCDA-LFB assay for C. albicans detection was a specific and accurate method, and could distinguish C. albicans from other pathogens. Just 200 fg of genomic DNA template from pure cultures of C. albicans could be detected using the MCDA-LFB method. The limit of detection (LOD) of the new method was more sensitive than that of both qPCR and loop-mediated isothermal amplification (LAMP). Of 240 clinical sputum samples, all of the C. albicans-positive (87/240) samples identified by the gold-standard method were successfully detected by the MCDA-LFB assay. Moreover, the true positive rate of the newly developed assay was not only higher than that of qPCR (100 vs. 86.2%), but also higher than that of LAMP (100 vs. 94.3%). Thus, the MCDA-LFB assay might be a simple, specific, and sensitive method for the rapid diagnosis of C. albicans in clinical samples.

Introduction

Candida albicans is a member of the human gut flora and it is also an opportunistic pathogenic yeast (Gow and Yadav, 2017). C. albicans, together with other species of Candida such as C. tropicalis, C. parapsilosis, and C. glabrata, are responsible for 50–90% of all cases of candidiasis in humans (Pfaller and Diekema, 2007; Martins et al., 2014; Schlecht et al., 2015). As one of the most common agents responsible for invasive candidiasis, C. albicans infection has been reported to result in a mortality rate of 40% for patients with systemic candidiasis. From an estimated report, invasive candidiasis contracted in a hospital setting could cause up to 11,200 deaths annually in the United States (Pfaller and Diekema, 2007).

However, the current gold-standard method for C. albicans detection is based on phenotyping, including culture, microscopic examination, and biochemical identification. It typically requires more than 2 days for growth and identification, which is limited by slow turn-around times; therefore, positive identification may occur late in the course of infection. Such delayed diagnosis may cause poor outcomes in patients with systemic fungal disease. Thus, it is imperative to develop and validate a rapid and accurate method for C. albicans identification. Several non-culture methods have been developed for C. albicans detection, including mass spectrometry (Zehm et al., 2012), immunoassay (Gunasekera et al., 2015), polymerase chain reaction (PCR) (Vahidnia et al., 2015), real-time PCR (Maaroufi et al., 2003; Kasai et al., 2006), polymerase spiral reaction (PSR) (Jiang et al., 2016), and loop-mediated isothermal amplification (LAMP) (Noguchi et al., 2017). Compared with the gold-standard culture method, these tests for C. albicans identification can save considerable time. However, such assays conversely depend on technical expertise or complex equipment, which may not be available in some settings.

Multiple cross displacement amplification (MCDA) is an isothermal amplification technology similar to LAMP, but with more sensitivity in bacterial detection (Wang et al., 2015). MCDA assays have been applied to detect several species of bacteria including Vibrio parahaemolyticus (Wang et al., 2016a), Shigella spp.(Wang et al., 2016b), Listeria monocytogenes (Wang et al., 2015), and Pseudomonas aeruginosa (Zhao et al., 2018). On the other hand, detection of amplification products from MCDA can be performed by gold nanoparticle-based lateral flow biosensors (LFBs). The LFB is a novel method to detect specific DNA fragments, which is based on the binding of antibodies (embedded on the LFB) and haptens (labeled on the 5′ side of primers) (Ang et al., 2015) (Chen et al., 2016; Toubanaki et al., 2016). In comparison with other methods, such as gel electrophoresis, with colorimetric indicators, and real-time turbidity for indicating MCDA results, LFB is relatively quick, objective, and simple (Wang et al., 2016a,b; Zhao et al., 2018). Therefore, a combined MCDA and LFB (MCDA-LFB) method could be used as an easy, rapid, specific, and very sensitive test for microbial detection. Internal transcribed spacer II (ITS II), a region between 5.8 and 28 S fungal ribosomal DNA, is frequently used for identification of fungal species, as the ITS II region of fungi exhibits sequence variability (Turenne et al., 1999; Jiang et al., 2016). Here, we aimed to establish a MCDA-LFB assay for C. albicans detection via the species-specific sequence from ITS II region with DNA template from pure cultures of the reference strain, following it, we applied the assay to clinical samples. We also compared the developed MCDA-LFB method with the gold-standard method and the more-commonly used molecular methods, in order to explore the application of our method in clinical practice.

Materials and Methods

Fungi, Bacteria Strains, and Clinical Samples

The 37 fungi strains and 2 bacteria strains used in this study are listed in Table 1, including 13 C. albicans and 26 non-C. albicans strains. Two reference strains were from American Type Culture Collection (ATCC), including one C. albicans strain (ATCC14053) and one Klebsiella pneumoniae strain (ATCC2146). Other strains were isolated from clinical samples, which were identified by the gold-standard method in the Clinical Laboratory of Peking University Shougang Hospital. All strains were stored in 15% (w/v) glycerol broth at –70°C after pure culture isolation. DNA from C. albicans ATCC14053 was used as a positive control, mainly for analysis of assay performance, and for determination of the optimal reaction temperature, while DNA from K. pneumoniae ATCC2146 and Pseudomonas aeruginosa SGH-PA001 strains were used as negative controls. To assess the limit of detection (LOD), DNA of C. albicans ATCC14053 was serially diluted, as indicated in the results section. DNA from 13 C. albicans strains and 26 non-C. albicans strains were used to analyze the specificity of the developed method.

TABLE 1
www.frontiersin.org

Table 1. Fungi and bacteria strains used in this study.

To explore the applicability of the C. albicans MCDA-LFB assay to clinical samples, human sputum samples were serially collected from January 1st to August 31st 2018 in the Clinical Laboratory, Peking University Shougang Hospital. This study was approved by the Ethics Committee of Shougang Hospital, and conducted according to the regulations of the Ministry of Health, China. Two hundred and forty human sputum samples were collected and 87 from the 240 samples were identified as C. albicans-positive by the gold-standard methods. All collected samples were stored at –70°C until use. The protocol was approved by the Ethics Committee of Shougang Hospital. Patients who provided sputum samples gave written informed consent, in accordance with the Declaration of Helsinki.

Genomic DNA Extraction

C. albicans strains and non-C. albicans fungi strains were subcultured on Emmons' modification of Sabouraud's agar plates at 30°C for 48 h before DNA extraction, while bacteria strains were subcultured on nutrient agar plates at 37°C. Genomic DNA was extracted from each strain using a DNA extraction kit (QIAamp DNA Mini Kits, Qiagen, Beijing, China) in accordance with the manufacturer's instructions. DNA concentration was measured by spectrophotometry using a NanoDrop 1,000 (Thermo Fisher Scientific, Wilmington, DE, USA). For clinical samples, DNA was extracted using the same QIAamp DNA extraction kit after the stored samples were thawed, in accordance with the manufacturer's instructions. The extracted DNAs were stored at −20°C until use.

The MCDA-LFB Assay for C. albicans Detection

According to the standard MCDA method (Wang et al., 2015), a set of 10 primers encompasses 10 distinct regions from one selected fragment, including the species-specific sequence. In this study, we designed three sets of MCDA primers targeting three fragments of the selected species-specific sequence of C. albicans (Table S1). All primers were designed using PrimerExplorer V4 (Eiken Chemical) and primer software PRIMER PREMIER 5.0. The locations and sequences of the MCDA primers are shown in Figure 1 and Table 2.

FIGURE 1
www.frontiersin.org

Figure 1. Appropriate primer design for C. albicans MCDA-LFB assay and primer sequence positions in the selected fragment including ITS II.

TABLE 2
www.frontiersin.org

Table 2. Primers used for MCDA, LAMP and qPCR assays.

The standard MCDA reaction mixture was 25 μl with the compound as described previously (Wang et al., 2016a,b) with some modification (Zhao et al., 2018). In brief, the reaction mixture included: 2 × reaction mix (Isothermal Amplification Kit, Jieyite, Tianjin, China), 12.5 μl; displacement primers F1 and F2, 0.4 μM each; amplification primers C1 and C2, 0.8 μM each; amplification primers R1, R2, D1, and D2, 1.2 μM each; cross primer CP1,1.2 μM; cross primer CP2, 2.4 μM; Bst DNA polymerase, 1.25 μl (10 U); and DNA template, 1 μl. Negative control mixtures contained 10 ng genomic DNA of K. pneumoniae ATCC2146 or P. aeruginosa SGH-PA001, and blank controls used 1 μl double distilled water. For LFB detection, primer C1, labeled with fluorescein isothiocyanate (FITC), and biotin-labeled primer D1, were used in the MCDA reaction instead of unlabeled ones. All of the primers were produced and purified by Ruiboxingke Biological Technology Ltd (Beijing, China).

Amplicons were analyzed by one or more of four different methods: (1). 2% agarose gel electrophoresis; (2). a colorimetric indicator (malachite green, MG); (3). by using a turbidimeter (Loopamp Realtime Turbidimeter LA-320C, Tokyo, Japan); (4), or by LFB. We constructed a gold nanoparticle-based dipstick biosensor, according to the description in a previous paper (Wang et al., 2016b) (Wang et al., 2016a). For electrophoresis, reaction products were separated by applying a voltage of 5 volts/cm across the gel for at least 30 min. DNA amplification products were visualized by exposing the gel to UV light; a specific ladder of bands is observed for positive amplifications, but not for negative controls. When using MG, the amplified products cause a change in the solution color, from colorless to sky blue (negative/blank controls remain colorless). With LFB, two red lines, including the test line (the lower one), and the control line (the upper one), are observed for positive reactions, but the control line only is visualized for negative and blank control reactions.

The process for assessing and verifying the effectiveness of the primers we designed was as described previously (Zhao et al., 2018). Briefly, the reactions were initially set at 62°C for 40 min and then amplification was terminated by heating at 85°C for 5 min. The optimal reaction temperature for C. albicans-MCDA assay was then determined in reactions with the most effectiveness, using one of the three sets primers at fixed temperatures starting from 60 to 65°C with intervals.

The Specificity and Sensitivity of MCDA-LFB Assay for C. albicans Detection

The specificity of the developed assay for C. albicans detection was analyzed using DNA from 39 strains, including 13 C. albicans strains, 24 non-C. albicans fungi strains, and 2 bacteria strains (Table 1). Except C. albicans, other microbe including the fungi and bacteria species used in this study are also common in hospitals. The sensitivity of the MCDA-LFB assay indicates the detection limit. This was defined in terms of the amount of genomic DNA template. Genomic DNA template from ATCC14053 was serially diluted from 2 ng/μl to 2 fg/μl (2 ng/μl, 20 pg/μl, 2 pg/μl, 500 fg/μl, 200 fg/μl, 20 fg/μl, and 2 fg/μl). The detection of C. albicans-MCDA by LFB was compared to those using the other three methods (see above), with three replicates.

To compare the sensitivity of MCDA with that of LAMP and qPCR assays for pure cultures, template DNA from the reference strains was diluted from 2 ng/μl to 2 fg/μl. For LAMP assays, the Loopamp Specified Microorganism “Fungi Candida albicans” Detection Kit (Eiken Chemical Co., Tokyo, Japan) was used in accordance with the manufacturer's instructions (Noguchi et al., 2017). In brief, the primer sequence for LAMP assay is targeting ITS region, which has obtained the patent NO. WO 2014/133153 A1 (Table 2). DNA amplification and real-time monitoring of the sample turbidity were conducted using the turbidimeter (LA-320C), for 40 min at 65°C. For qPCR assays, primers and amplification procedures were as described previously (Alonso et al., 2018). The primer sequence is targeting ACT1 gene (Table 2). The cycling protocol were: 5 min at 95°C, 39 cycles of denaturation during 15 s at 94°C, annealing at 55°C for 30 s, and elongation at 60°C for 15 s, with melt curve tested with gradual increase of 0.5°C from 55 to 95°C.

Application of C. albicans-MCDA-LFB Technology for Clinical Samples

Two Hundred and Forty human clinical sputum specimen were used to assess the applicability of our C. albicans-MCDA-LFB technology for clinical samples. By the gold-standard method, 87 (36.3%) of the 240 samples were identified as C. albicans-positive, while the remaining 153 samples were identified as C. albicans-negative, including samples with other fungi or bacteria growing, or microbe-free samples. The true positive rate, true negative rate, and consistency rate of results from C. albicans-MCDA-LFB, qPCR, or LAMP assays were computed based on the result from gold-standard method.

Statistical Method

True positive rate of the MCDA-LFB, qPCR, or LAMP assays for clinical sample detection was computed by determining the number of true positive/(number of true positive + number of false negative) × 100%. True negative rate of the MCDA-LFB, qPCR, or LAMP assays for clinical detection was computed by determining the number of true negative/(number of true negative + number of false positive) × 100%. Consistency rate of the MCDA-LFB, qPCR, or LAMP assays for clinical sample detection was computed by determining the (number of true positive + number of true negative)/number of all samples × 100%.

Results

Detection of C. albicans-MCDA Products

To identify the most effective of the three sets of primers for C. albicans, MCDA assays were performed at 62° for 40 min using DNA from pure cultures as template. The most effective set of primers was selected by inspection of the curves from real-time turbidity (LC320) for use as future experimental primers (data not shown). Then, to confirm the effectiveness of the MCDA primers we selected (Table 1; Figure 1), the amplification products were detected by two different methods, as mentioned in the Methods section. Of the two methods, results were observed to be positive when the template was DNA from C. albicans ATCC14053, but not with DNA from K. pneumoniae ATCC2146 or P. aeruginosa SGH-PA001 (negative controls), or from the blank control (Figure 2). Therefore, the C. albicans-MCDA primer set we selected in the current study was suitable for MCDA-LFB assay development.

FIGURE 2
www.frontiersin.org

Figure 2. Detection of C. albicans-MCDA products with the appropriate primers. (A) Using MG, amplification products in the C. albicans-MCDA assay were detected visually by a color change (sky blue for positive samples, colorless for negative samples); (B) 2% agarose gel electrophoresis of C. albicans-MCDA products. Lane M, DNA maker DL2000. Tube 1/Lane 1: amplification of C. albicans strain ATCC14053; Tube 2/Lane 2: amplification of Klebsiella pneumoniae strain ATCC2146 (negative control); Tube 3/Lane 3: amplification of Pseudomonas aeruginosa SGH-PA001 (negative control); Tube 4/Lane 4: double distilled water (blank control). MCDA, multiple cross displacement amplification; MG, Malachite Green; LFB, lateral flow biosensor.

Optimal Assay Temperature

To identify the optimal assay temperature, the MCDA assay was conducted from 60 to 65°C with 10 pg/reaction of C. albicans ATCC14053 DNA as template. All reactions were monitored by real-time turbidity (LC320). The kinetic data in Figure 3 show that 64°C was the optimum reaction temperature for the selected primers for detection of C. albicans by MCDA.

FIGURE 3
www.frontiersin.org

Figure 3. Optimal amplification temperature for C. albicans-MCDA assay. C. albicans-MCDA reactions were monitored by real-time measurement of turbidity. The corresponding curves are shown in the panels. K. pneumoniae and sterile double-distilled water were used as a negative control, and a blank control, respectively. Turbidity of >0.1 was considered a positive reaction. Six kinetic curves (A–F) were generated from 60 to 65°C, with 10 pg C. albicans DNA per reaction. MCDA, multiple cross displacement amplification.

Specificity of the C. albicans-MCDA-LFB

When DNA from 37 fungi strains and 2 bacteria strains was tested by MCDA-LFB assay, only 13 C. albicans DNA showed positive results. Amplification products were not detected when the template was DNA from non-C. albicans strains (Figure 4). Therefore, the MCDA-LFB was highly specific for C. albicans detection.

FIGURE 4
www.frontiersin.org

Figure 4. Specificity of the MCDA-LFB detection assay for different strains. Biosensor 1, C. albicans ATCC14053; Biosensor 2–13, different C. albicans strains isolated from clinical samples; Biosensor 14–18, Candida tropicalis; Biosensor 19–22, Candida parapsilosis; Biosensor 23–27, Candida glabrata; Biosensor 28–30, Candida krusei; Biosensor 31–39, Candida dubiniensis, Candida rugosa, Candida guilliermondii, Cryptococcus neoformans, Cryptococcus gattii, Cryptococcus luteolus, Cryptococcus curvatus, Klebsiella pneumoniae, Pseudomonas aeruginosa.

Sensitivity of the C. albicans-MCDA Assay

Two Hundred femtogram of C. albicans template DNA yielded amplicons that could be detected colorimetrically, by LFB, by electrophoresis, or by turbidity (Figure 5). The results obtained using the LFB (Figure 5B) agreed completely with those by the other three methods. The MCDA assay was 2.5-fold more sensitive than LAMP for the detection of C. albicans by LFB, and was 5-fold more sensitive than qPCR for the detection of C. albicans by electrophoresis (Table 3).

FIGURE 5
www.frontiersin.org

Figure 5. Sensitivity of the C. albicans-MCDA assay. Four detection techniques, including a colorimetric indicator (MG) (A), lateral flow biosensor (B), gel electrophoresis (C), and real-time turbidity (D), were applied for analyzing amplification products. Serial dilutions (2 ng, 20 pg, 2 pg, 500 fg, 200 fg, 20 fg, and 2 fg) of target template DNA were subjected to standard MCDA reactions. Tubes (A)/Lanes (B)/Biosensor (C)/Turbidity signals (D) 1–7 represent the template DNA mass levels of 2 ng, 20 pg, 2 pg, 500 fg, 200 fg, 20 fg, and 2 fg per reaction, respectively; 8 represents a blank control. MG, malachite green; MCDA, multiple cross displacement amplification.

TABLE 3
www.frontiersin.org

Table 3. Limit of detection (LOD) and amplification time for detection of C. albicans using MCDA-LFB, LAMP, and qPCR.

Evaluation of Clinical Samples for MCDA-Based Identification of C. albicans

A total of 240 clinical sputum samples were assayed using C. albicans-MCDA-LFB and qPCR (Table 4). By the gold-standard method, 87 of the 240 clinical samples were positive for C. albicans. The MCDA-LFB assay results were in complete agreement. However, it was identified that there were 82 of 87 positive samples by LAMP, and 75 of 87 positive samples by qPCR assays. Thus, the true positive rate and true negative rate for C. albicans-MCDA-LFB assay were both 100%, while they were 94.3 and 100% for LAMP, 86.2 and 100% for qPCR, respectively.

TABLE 4
www.frontiersin.org

Table 4. Clinical application of C. albicans-MCDA-LFB assay in sputum samples.

Discussion

C. albicans, the most common human fungal pathogen causing diseases ranging from mucosal to systemic infections, is an important nosocomial, infectious pathogen. Culture based microscopic examination and biochemical identification is the gold-standard method for identification of C. albicans infections, but is labor-intensive and minimally effective. Thus, early and rapid identification of C. albicans in clinical samples of patients will help to control infection by this pathogen. To date, many techniques other than the culture based method have been developed to detect C. albicans in clinical samples (Maaroufi et al., 2003; Kasai et al., 2006; Zehm et al., 2012; Gunasekera et al., 2015; Vahidnia et al., 2015; Jiang et al., 2016; Noguchi et al., 2017). However, to the best of our knowledge, the present study is the first report of a MCDA-LFB assay developed to identify C. albicans, with a LOD for C. albicans as low as 200 fg of DNA from pure cultures.

Compared to qPCR, the developed MCDA-LFB method is less dependent on equipment and more sensitive. As an isothermal amplification assay, MCDA only requires a simple water bath or heater. Moreover, there were 5 pairs of primers engaged on the target-specific sequence in our MCDA method, which make it more effective than the normal PCR method. In the present study, the PCR method had less true positive rate than MCDA when both were applied for C. albicans detection in clinical samples. We also compared MCDA with other isothermal amplification assays, such as LAMP, for C. albicans detection. The primers for LAMP were obtained from Noguchi's report (Noguchi et al., 2017). By their method, the LOD to detect C. albicans was 1 pg. Here, we reported the LOD was 500 fg for LAMP to detect C. albicans, while it was 200 fg for MCDA. These results showed that MCDA was 2.5-fold more sensitive than LAMP in our study, in accordance with previous studies (Wang et al., 2015). On the other hand, the positive reactions could be obtained in as rapidly as 25 min in real-time turbidity measurements. These results indicated that the MCDA is a sensitive and rapid alternative to other molecular techniques for detection of trace amounts of C. albicans.

In the present study, the objective, rapid, and simple gold-nanoparticle-based LFB was also introduced to “read” the amplification products. LFB could detect amplification products labeled with FITC and biotin, through the appearance of red lines on the LFB strips within 2 min. All of the C. albicans-positive products from MCDA were successfully identified using the LFB, which indicates that the LFBs are relatively more objective and reliable than turbidimetric and colorimetric indicators [9]. On the other hand, although the gel electrophoresis method could distinguish specific amplification from non-specific amplification, it is laborious and requires specific, dedicated equipment. Considering the high workload involved in diagnostic examinations in clinical laboratories, LFB represents a promising candidate to detect the amplification products of either normal PCR or isothermal amplification approaches.

Importantly, we also successfully applied the developed C. albicans MCDA-LFB assay to clinical sputum samples. When compared with the time-consuming and labor-intensive gold-standard method, the new assay is a rapid and easy-to-perform method, with 100% true positive rate and 100% true negative rate. In particular, its true positive rate is higher than that of the commonly used qPCR (100 vs. 86.2%). The entire C. albicans-MCDA-LFB procedure, including sample processing, could be completed within 2 h. This is much faster than the culture-based method. Our study suggests that the MCDA-LFB assay could be used for fast C. albicans detection in clinical sputum samples. In the near future, we will focus on other clinical sample types, e.g., urine, blood, and throat swabs, to test the clinical application of the C. albicans-MCDA-LFB assay in diverse samples.

There are also several possible limitations to our study, which should be addressed. First, the developed MCDA-LFB assay is a qualitative determination for C. albicans, as we could not quantify the amount of pathogen in the sample, whereas the culture-based method is also a qualitative determination for pathogen detection. Quantification of the amount of microbe in samples may be the direction of pathogen detection research. This is because if it is possible to quantify the amount of microbe in samples, we could monitor and evaluate the therapeutic effect of antibiotics. Second, for isothermal amplification methods with high concentrations of primers, carry-over contamination is a great concern for genomic template-based identification assays. A possibility to lower carryover contamination is to control the amplification time, which has been suggested by Wang et al. (2018). In this study, the amplification time was also limited to 40 min to reduce false-positive results. Moreover, we strictly controlled the laboratory environment to reduce the production of aerosols in experimental processes. It was recently reported that carry-over contamination could be removed by adding Antarctic thermolabile uracil-DNA-glycosylase (UNG) into the reaction system before amplification with the mixture of dTTP and dUTP instead of dTTP alone (Wang et al., 2017, 2018). Although UNG was not added to the reaction system in this study, the detection results were the same as the gold-standard methods in the evaluation of clinical samples for MCDA-based identification of C. albicans, indicating that false positive rates have been controlled in our laboratory. Certainly, it would be a good choice to further reduce carry-over contamination through adding Antarctic thermolabile UNG into the reaction system before MCDA assay performance.

In conclusion, we developed a MCDA-LFB assay for rapid and effective detection of C. albicans. The LOD of the new assay for C. albicans detection from pure cultures was as low as 200 fg. In comparison to other molecular methods (PCR, LAMP) for detection of C. albicans, the MCDA-LFB method is more sensitive, time-saving, and does not require expensive equipment or diagnostic technology. When using the MCDA-LFB assay to detect infected clinical samples, the results were consistent with those of the gold-standard method. This suggests the possibility for further clinical application of our C. albicans-MCDA-LFB assay.

Ethics Statement

This study was approved by the Ethics Committee of Shougang Hospital, and conducted according to the regulations of the Ministry of Health, China. The protocol was approved by the Ethics Committee of Shougang Hospital. Patients who provided sputum samples gave written informed consent, in accordance with the Declaration of Helsinki.

Disclosure

SH, FaZ, LN, LY, JN, FeZ, NG, XZ, LW, and CW have filed for a patent from the State Intellectual Property Office of the People's Republic of China, which includes the sequences and the MCDA-LFB method used for detecting C. albicans in this manuscript [application number CN 201811580513.9].

Author Contributions

FaZ, JN, and SH conceived and designed the experiments; FaZ, LY, CW, JW, NG, XZ, and LW performed the experiments; FeZ, JN, LN, NG, XZ, and LW analyzed the data; FaZ, LN, FaZ, and SH contributed reagents, materials, analysis tools; FaZ and LN wrote the manuscript; LN, LY, and SH reviewed the manuscript.

Funding

This work was supported by the Education Department of Hainan Province [grant number: Hnky2019ZD-27], the Beijing Natural Science Foundation [grant number: 7192240], the National Natural Science Foundation of China [grant number: 81760376] and the Peking University Shougang Hospital [grant numbers: 2017-Yuan-LC-08, SGYYZ201701].

Conflict of Interest Statement

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcimb.2019.00102/full#supplementary-material

Table S1. Three sets of MCDA primers for C. albicans detection.

Abbreviations

MCDA, Multiple cross displacement amplification; LFB, Lateral flow biosensors; LAMP, Loop-Mediated Isothermal Amplification; LOD, Limit of Detection; ITS II, Internal Transcribed Spacer II; ATCC, American Type Culture Collection; MG, Malachite Green; SGH, Shougang Hospital; FITC, Fluorescein Isothiocyanate; UNG, Uracil-DNA-Glycosylase.

References

Alonso, G. C., Pavarina, A. C., Sousa, T. V., and Klein, M. I. (2018). A quest to find good primers for gene expression analysis of Candida albicans from clinical samples. J. Microbiol. Methods 147, 1–13. doi: 10.1016/j.mimet.2018.02.010

PubMed Abstract | CrossRef Full Text | Google Scholar

Ang, G. Y., Yu, C. Y., Chan, K. G., Singh, K. K., and Chan Yean, Y. (2015). Development of a dry-reagent-based nucleic acid-sensing platform by coupling thermostabilised LATE-PCR assay to an oligonucleotide-modified lateral flow biosensor. J. Microbiol. Methods 118, 99–105. doi: 10.1016/j.mimet.2015.08.024

PubMed Abstract | CrossRef Full Text | Google Scholar

Chen, Y., Cheng, N., Xu, Y., Huang, K., Luo, Y., and Xu, W. (2016). Point-of-care and visual detection of P. aeruginosa and its toxin genes by multiple LAMP and lateral flow nucleic acid biosensor. Biosens Bioelectron 81, 317–323. doi: 10.1016/j.bios.2016.03.006

PubMed Abstract | CrossRef Full Text | Google Scholar

Gow, N. R., and Yadav, B. (2017). Microbe profile: Candida albicans: a shape-changing, opportunistic pathogenic fungus of humans. Microbiology 163, 1145–1147. doi: 10.1099/mic.0.000499

PubMed Abstract | CrossRef Full Text | Google Scholar

Gunasekera, M., Narine, M., Ashton, M., and Esfandiari, J. (2015). Development of a dual path platform (DPP(R)) immunoassay for rapid detection of Candida albicans in human whole blood and serum. J. Immunol. Methods 424, 7–13. doi: 10.1016/j.jim.2015.04.014

PubMed Abstract | CrossRef Full Text | Google Scholar

Jiang, X., Dong, D., Bian, L., Zou, D., He, X., Ao, D., et al. (2016). Rapid Detection of Candida albicans by polymerase spiral reaction assay in clinical blood samples. Front. Microbiol. 7:916. doi: 10.3389/fmicb.2016.00916

PubMed Abstract | CrossRef Full Text | Google Scholar

Kasai, M., Francesconi, A., Petraitiene, R., Petraitis, V., Kelaher, A. M., Kim, H. S., et al. (2006). Use of quantitative real-time PCR to study the kinetics of extracellular DNA released from Candida albicans, with implications for diagnosis of invasive Candidiasis. J. Clin. Microbiol. 44, 143–150. doi: 10.1128/JCM.44.1.143-150.2006

PubMed Abstract | CrossRef Full Text | Google Scholar

Maaroufi, Y., Heymans, C., De Bruyne, J. M., Duchateau, V., Rodriguez-Villalobos, H., Aoun, M., et al. (2003). Rapid detection of Candida albicans in clinical blood samples by using a TaqMan-based PCR assay. J. Clin. Microbiol. 41, 3293–3298. doi: 10.1128/jcm.41.7.3293-3298.2003

PubMed Abstract | CrossRef Full Text | Google Scholar

Martins, N., Ferreira, I. C., Barros, L., Silva, S., and Henriques, M. (2014). Candidiasis: predisposing factors, prevention, diagnosis and alternative treatment. Mycopathologia 177, 223–240. doi: 10.1007/s11046-014-9749-1

PubMed Abstract | CrossRef Full Text | Google Scholar

Noguchi, H., Iwase, T., Omagari, D., Asano, M., Nakamura, R., Ueki, K., et al. (2017). Rapid detection of Candida albicans in oral exfoliative cytology samples by loop-mediated isothermal amplification. J. Oral Sci. 59, 541–547. doi: 10.2334/josnusd.16-0717

PubMed Abstract | CrossRef Full Text | Google Scholar

Pfaller, M. A., and Diekema, D. J. (2007). Epidemiology of invasive candidiasis: a persistent public health problem. Clin. Microbiol. Rev. 20, 133–163. doi: 10.1128/CMR.00029-06

PubMed Abstract | CrossRef Full Text | Google Scholar

Schlecht, L. M., Peters, B. M., Krom, B. P., Freiberg, J. A., Hansch, G. M., Filler, S. G., et al. (2015). Systemic Staphylococcus aureus infection mediated by Candida albicans hyphal invasion of mucosal tissue. Microbiology 161, 168–181. doi: 10.1099/mic.0.083485-0

PubMed Abstract | CrossRef Full Text | Google Scholar

Toubanaki, D. K., Athanasiou, E., and Karagouni, E. (2016). Gold nanoparticle-based lateral flow biosensor for rapid visual detection of Leishmania-specific DNA amplification products. J. Microbiol. Methods 127, 51–58. doi: 10.1016/j.mimet.2016.05.027

PubMed Abstract | CrossRef Full Text | Google Scholar

Turenne, C. Y., Sanche, S. E., Hoban, D. J., Karlowsky, J. A., and Kabani, A. M. (1999). Rapid identification of fungi by using the ITS2 genetic region and an automated fluorescent capillary electrophoresis system. J. Clin. Microbiol. 37, 1846–1851.

PubMed Abstract | Google Scholar

Vahidnia, A., Bekers, W., Bliekendaal, H., and Spaargaren, J. (2015). High throughput multiplex-PCR for direct detection and diagnosis of dermatophyte species, Candida albicans and Candida parapsilosis in clinical specimen. J. Microbiol. Methods 113, 38–40. doi: 10.1016/j.mimet.2015.03.024

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Li, H., Li, D., Li, K., Wang, Y., Xu, J., et al. (2016a). Multiple cross displacement amplification combined with gold nanoparticle-based lateral flow biosensor for detection of Vibrio parahaemolyticus. Front. Microbiol. 7:2047. doi: 10.3389/fmicb.2016.02047

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Li, H., Wang, Y., Xu, H., Xu, J., and Ye, C. (2018). Antarctic thermolabile uracil-DNA-glycosylase-supplemented multiple cross displacement amplification using a label-based nanoparticle lateral flow biosensor for the simultaneous detection of nucleic acid sequences and elimination of carryover contamination. Nano Res. 11, 2632–2647. doi: 10.1007/s12274-017-1893-z

CrossRef Full Text | Google Scholar

Wang, Y., Liu, D., Deng, J., Wang, Y., Xu, J., and Ye, C. (2017). Loop-mediated isothermal amplification using self-avoiding molecular recognition systems and antarctic thermal sensitive uracil-DNA-glycosylase for detection of nucleic acid with prevention of carryover contamination. Anal. Chim. Acta 996, 74–87. doi: 10.1016/j.aca.2017.10.022

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Wang, Y., Ma, A. J., Li, D. X., Luo, L. J., Liu, D. X., et al. (2015). Rapid and sensitive isothermal detection of nucleic-acid sequence by multiple cross displacement amplification. Sci. Rep. 5:11902. doi: 10.1038/srep11902

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Wang, Y., Xu, J., and Ye, C. (2016b). Development of multiple cross displacement amplification label-based gold nanoparticles lateral flow biosensor for detection of shigella spp. Front. Microbiol. 7:1834. doi: 10.3389/fmicb.2016.01834

PubMed Abstract | CrossRef Full Text | Google Scholar

Zehm, S., Schweinitz, S., Wurzner, R., Colvin, H. P., and Rieder, J. (2012). Detection of Candida albicans by mass spectrometric fingerprinting. Curr. Microbiol. 64, 271–275. doi: 10.1007/s00284-011-0064-5

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhao, F., Niu, L., Nong, J., Wang, C., Wang, J., Liu, Y., et al. (2018). Rapid and sensitive detection of Pseudomonas aeruginosa using multiple cross displacement amplification and gold nanoparticle-based lateral flow biosensor visualization. FEMS Microbiol. Lett. 365, 1–6. doi: 10.1093/femsle/fny147

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: Candida albicans, opportunistic pathogenic yeast, multiple cross displacement amplification, gold nanoparticle-based lateral flow biosensor, loop-mediated isothermal amplification, qPCR, clinical samples

Citation: Zhao F, Niu L, Yan L, Nong J, Wang C, Wang J, Gao N, Zhu X, Wu L, Zheng F and Hu S (2019) Establishment and Application of Multiple Cross Displacement Amplification Coupled With Lateral Flow Biosensor (MCDA-LFB) for Visual and Rapid Detection of Candida albicans in Clinical Samples. Front. Cell. Infect. Microbiol. 9:102. doi: 10.3389/fcimb.2019.00102

Received: 16 November 2018; Accepted: 25 March 2019;
Published: 16 April 2019.

Edited by:

Yanan Zhao, Center for Discovery and Innovation, Hackensack Meridian Health, United States

Reviewed by:

Lukasz Kozubowski, Clemson University, United States
Yi Wang, Beijing Children's Hospital, Capital Medical University, China

Copyright © 2019 Zhao, Niu, Yan, Nong, Wang, Wang, Gao, Zhu, Wu, Zheng and Hu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Shoukui Hu, shoukuihu@163.com

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.