Skip to main content

ORIGINAL RESEARCH article

Front. Cardiovasc. Med. , 12 March 2025

Sec. Cardiac Rhythmology

Volume 12 - 2025 | https://doi.org/10.3389/fcvm.2025.1500196

Comparison of electrophysiological left bundle branch pacing characteristics in different bilateral electrode pacing vector configurations

  • 1Department of Cardiovascular Medicine, Ningbo No. 2 Hospital, Ningbo, Zhejiang, China
  • 2Cardiovascular Disease Clinical Medical Research Center of Ningbo, Ningbo, Zhejiang, China

Introduction: Left bundle branch pacing (LBBP) in bipolar pacing with anodal capture produces a more balanced ventricular activation than conventional unipolar pacing but need high pacing output. The present study aimed to compare the electrophysiological characteristics of LBBP in different bilateral electrode pacing vector configurations.

Methods: A total of 57 patients who met the criteria for left bundle branch (LBB) capture and underwent three bilateral electrode pacing vector configuration test were enrolled. The electrocardiogram (ECG) and electrogram (EGM) parameters were evaluated and other electrophysiological characteristics were analyzed using a three-electrode configuration test.

Results: Seven capture modes [right ventricular septal (RVS) + left ventricular septal (LVS) + LBB, RVS + LBB, LVS + LBB, RVS + LVS, RVS, LVS, and LBB] were utilized in the study. The thresholds of full fusion mode (RVS + LVS + LB) in Bilateral Cathodes and Ring Bipolar were all lower than that in Tip Bipolar (1.2 ± 0.5 V vs. 2.7 ± 1.0 V, P < 0.001; 1.6 ± 0.6 V vs. 2.7 ± 1.0 V, P < 0.001). Full fusion mode had the shortest P-QRS (116.9 ± 12.8 ms) and V1 RWPT (94.5 ± 12.3 ms), V6 RWPT remain the shortest (64.9 ± 9.7 ms).

Conclusion: Changing the bilateral electrode pacing vector configuration to Bilateral Cathodes and Ring Bipolar can reduce the full fusion mode capture threshold compared to conventional bipolar pacing.

Graphical Abstract
www.frontiersin.org

Graphical Abstract. The characteristics of electrophysiological in three bilateral electrode pacing vector configurations.

Introduction

Conventional right ventricular (RV) pacing is a standard treatment for patients with symptomatic bradyarrhythmia, but it is associated with the risk of heart failure due to electrical and mechanical dyssynchrony (1, 2). His bundle pacing (HBP) is a more physiological pacing modality with an improvement in exercise capacity and left ventricular (LV) ejection fraction (3, 4). However, it also has limitations, including longer procedure times and high and sometimes unstable pacing capture thresholds (5, 6). Recently, left bundle branch pacing (LBBP) has emerged as another conduction system pacing modality with a low and stable threshold compared to HBP (710). LBBP produces a fast LV activation, but the presence of a right bundle branch block (RBBB) or even incomplete RBBB morphology remains a concern for interventricular dyssynchrony (11).

Bipolar LBBP pacing with anodal and cathodal capture can simultaneously capture the left bundle branch (LBB), left ventricular septal (LVS) myocardium, and right ventricular septal (RVS) myocardium when the lead tip is placed in the left septal subendocardial region and the ring electrode is in contact with the right septal endocardium. This mode can shorten the time to RV activation and results in a more narrow Paced (P)-QRS complex than conventional unipolar LBBP pacing (12, 13). However, the LBBP with anodal capture usually required a high pacing output, and in bipolar pacing RVS capture via the anode needed much higher pacing outputs than that in unipolar pacing by cathode (12). We hypothesized that RVS thresholds might differ as a result of capture by either the cathode or anode. Consequently, we examined whether the threshold of anodal capture in LBBP could be adjusted by changing the bilateral electrode pacing vectors, such as using the tip electrode as the anode or both the tip and ring electrodes as the cathodes. The present study sought to compare the electrophysiological characteristics of LBBP in different bilateral electrode pacing vector configurations.

Methods

Patient population and implantation procedure

We prospectively enrolled consecutive patients referred for LBBP between March 2022 and November 2022 at the Ningbo No. 2 Hospital. The study protocol was approved by the hospital institutional review board (SL-KYSB-NBEY-2021-079-01), and all patients provided written informed consent.

A SelectSecure™ lead (model 3830, 69 cm, Medtronic Inc., Minneapolis, MN, USA) and a fixed-curve sheath (C315 HIS, Medtronic, Minneapolis, MN) were used during the lead implant procedure. The continuous LBBP pacing technique was utilized as described previously (1416). Electrocardiograms (ECG) were obtained in patients with direct evidence of LBB capture as demonstrations of non-selective LBB to selective LBB capture or non-selective LBB to septal capture transition during unipolar pacing threshold testing and were included in the study to ensure data accuracy.

Patients who had already undergone LBBP and three bilateral electrode pacing vector configuration testing, which was not yet a conventional procedure before March 2022 (July 2021–March 2022), were retrospectively enrolled to increase the study sample size.

Multiple bilateral electrode pacing vector configuration testing

Three different bilateral electrode pacing vector configurations were tested during the threshold test (Figure 1): (1) tip electrode as the cathode and ring electrode as the anode (Tip Bipolar); (2) tip electrode as the anode and ring electrode as the cathode (Ring Bipolar); and (3) both tip and ring electrodes as the cathodes (Bilateral Cathodes). There were predictable and consistent ECG and electrogram (EGM) morphology changes during the threshold test of the three different bilateral electrode pacing vector configurations (Figures 2, 3).

Figure 1
www.frontiersin.org

Figure 1. Multiple bilateral electrode pacing vector configuration testing. (A) Schematic representation of three bilateral electrode pacing vector configurations. (B) The distribution of seven capture modes. “ + ” = anodal; “−“ = cathodal. The dotted line indicates electric current flow path.

Figure 2
www.frontiersin.org

Figure 2. A single patient with four capture mode transitions in three bilateral electrode pacing vector configurations. In Tip Bipolar, the output decreased to 2.0 V, RVS + LVS + LBB transferred to LVS + LBB, LB lead EGM was separated, and V1 RWPT was prolonged. The output further decreased to 0.4 V, LVS + LBB transferred to LBB, V1 changed to “M” type, V1 RWPT was further prolonged, LB lead EGM was further separated, and there was an isoelectric interval between pacing artifact and local ventricular electrogram. In Ring Bipolar, the output decreased to 1.1 V, RVS + LVS + LBB transferred to RVS + LBB, ECG remained the same, and LB lead EGM was separated. The output further decreased to 0.9 V and RVS + LBB transferred to LBB. Bilateral Cathodal transition was similar to that of Ring Bipolar.

Figure 3
www.frontiersin.org

Figure 3. A single patient with five capture mode transitions in three bilateral electrode pacing vector configurations. In Tip Bipolar, the output decreased to 2.5 V, RVS + LVS + LBB transferred to LVS + LBB, LB lead EGM was separated, and V1 RWPT was prolonged. The output further decreased to 1.0 V, LVS + LBB transferred to LVS, LB lead EGM remained similar, and V6 RWPT was prolonged. In Ring Bipolar, the output decreased to 2.4 V, RVS + LVS + LBB transferred to RVS + LVS, LB lead EGM remained similar, and V6 RWPT was prolonged. The output further decreased to 1.1 V, RVS + LVS transferred to RVS, LB lead EGM was separated, and V6 RWPT was further prolonged. Bilateral Cathodal transition was similar to that of Ring Bipolar.

Evaluation of ECG and EGM parameters

Our previous study distinguished four capture modes during a conventional bipolar pacing threshold test based on the changes in ECG and left bundle (LB) lead EGM (12). When the pacing output was reduced, LB lead EGM separation (a complex mixed wave changed to two connected individual waves) indicated one-sided ventricular myocardial loss capture. Losing the RVS capture prolonged the V1 stimulus–peak of the R wave (RWPT). Losing the LBB capture prolonged the V6 RWPT or resulted in an isoelectric interval between the pacing artifact and local ventricular EGM in selected LBB captures. Seven different capture modes can be distinguished based on this theory according to the changes in EGM and ECG during the threshold test of the three bilateral electrode pacing vector configurations (Supplementary Figure S1; Figures 2, 3). These were: (A) full fusion mode, capture RVS + LVS + LBB; (B) semi fusion mode, included three capture modes, RVS + LBB, LVS + LBB, and RVS + LVS; and (C) select capture mode, also included three capture modes, RVS, LVS, and LBB.

Data collection

Baseline patient characteristics and indications for pacing were recorded in addition to echocardiographic data and baseline QRS duration. Implantation and threshold test procedures were recorded on a digital electrophysiological system (Abbott Laboratories, Chicago, IL). The band-pass filter for the pacing lead was set to: “High Pass-200 Hz\Low Pass-500 Hz”. To ensure high precision, the measurements were performed using all 12 surface ECG leads recorded simultaneously, digital calipers, fast sweep speed (200–600 mm/s), and appropriate signal augmentation. At least three QRS complexes were measured and their values were averaged. P-QRS duration was calculated based on the pacing stimulus and the final QRS component in any of the 12 ECG leads. The presence of an LBB potential and the LBB potential-final QRS component duration (LB-QRS) were also recorded.

The characteristics of various changes in ECG and EGM morphology in three different bilateral electrode pacing vector configurations and the thresholds of seven capture modes were documented. The RWPT in surface leads V6 and V1 (paced ECG morphology of V1 lead in RVS capture mode presented as QS type, so it had no V1 RWPT in RVS capture mode) and P-QRS of seven capture modes were determined.

Statistical analysis

Continuous variables were reported as mean ± standard deviation. Categorical variables were expressed as percentages. Paired comparisons were made using the student t-test if the data were normally distributed. ANOVA was used for comparing more than two groups in repeated measurements, while a least significant difference (LSD) post hoc test was used for two-group comparisons. A p-value of <0.05 was considered statistically significant. Statistical analysis was performed using IBM SPSS Statistics for Macintosh (version 26.0, IBM Corp, Armonk, NY, USA).

Results

Population

A total of 72 patients underwent the LBBP procedure from March 2022 to November 2022. Of these, 56 cases had clear dynamic ECG maneuvers during the unipolar pacing threshold test, and 53 patients underwent three different bilateral electrode pacing vector configuration testing. Four patients who have undergone three different bilateral electrode pacing vector configuration testing before March 2022 were retrospectively enrolled. Thus, a total of 57 patients were enrolled in the study. The primary clinical and procedure-related characteristics of the included patients are presented in Tables 1, 2. The mean age was 71.3 ± 10.1 years and 49.1% of patients were men. Indications for pacing included atrioventricular block 73.7%, sick sinus syndrome 24.6%, and heart failure 1.7%. The unipolar pacing threshold test showed the S-LB threshold captured by the tip electrode to be 0.5 ± 0.2 V, the LVS threshold captured by the tip electrode to be 0.6 ± 0.1 V, and the RVS threshold captured by the Ring electrode to be 0.9 ± 0.5 V.

Table 1
www.frontiersin.org

Table 1. Patient characteristics (n = 57).

Table 2
www.frontiersin.org

Table 2. Pacing and procedure-related characteristics (n = 57).

Transitions of seven capture modes during three bilateral electrode pacing vector configuration testing

The distribution of seven capture modes during three different bilateral electrode pacing vector configuration testing is shown in Figure 1. The transition of seven capture modes during three different bilateral electrode pacing vector configuration testing is shown in Figures 2, 3.

The full fusion mode (RVS + LVS + LBB) was achieved in Tip Bipolar, Ring Bipolar, and Bilateral Cathodes with a higher pacing output in all 57 patients. LBB and LVS were captured using the tip side and RVS was captured using the ring side, which presented as “QS, Qr, or QR” pattern in lead V1, a complex mixed wave in LB lead EGM, short V1\V6 RWPT, and narrow P-QRS.

There are two types of transition in the Tip Bipolar configuration (conventional bipolar pacing) that were previously described (12). Therefore, the detailed changes in ECG and EGM are not discussed in this paper. A loss of RVS capture resulted in LBB + LVS capture mode in all 57 patients as the output decreased. LBB + LVS change to selected LBB capture in 42 patients and LBB + LVS change to LVS capture in 15 patients occurred as the output was reduced further.

In the Ring Bipolar configuration mode, only 14 patients changed to LVS + LBB capture as the output was reduced. A total of 33 patients changed to RVS + LBB capture (EGM separated and ECG remained the same as full fusion mode), and 10 patients changed to LVS + RVS capture (V6 RWPT suddenly prolonged and EGM remained almost the same as full fusion mode). Furthermore, 39 patients changed to selected LBB capture, three patients changed to LVS capture, and 15 patients changed to RVS capture (V6 RWPT further prolonged) when the output was reduced further.

In the Bilateral Cathodes mode, a similar transition of capture modes as that in the Ring Bipolar configuration mode was observed during the threshold test. A total of 27 patients changed to LVS + LBB capture, 20 patients changed to RVS + LBB capture, and 10 patients changed to RVS + LVS capture. As the output was reduced further, 39 patients changed to LBB capture, 10 patients changed to LVS capture, and eight patients changed to RVS capture.

Ventricular sensing, impedance, and threshold of seven capture modes in three bilateral electrode pacing vector configurations

Bilateral Cathodes mode had the lowest ventricular sensing and impedance (7.0 ± 4.3 mV; 421.4 ± 83.5 Ω; Figure 4). Full fusion mode thresholds in Bilateral Cathodes and Ring Bipolar were all lower than that in Tip Bipolar (1.2 ± 0.5 V vs. 2.7 ± 1.0 V, P < 0.001; 1.6 ± 0.6 V vs. 2.7 ± 1.0 V, P < 0.001). The threshold of LVS + LBB capture mode in Ring Bipolar mode was much higher than that in Tip Bipolar (1.1 ± 0.5 V vs. 0.7 ± 0.2 V, P < 0.001). The select LBB and LVS capture threshold in Ring Bipolar mode was much higher than that in Tip Bipolar (0.7 ± 0.2 V vs. 0.4 ± 0.2 V, P < 0.001; 1.0 ± 0.2 V vs. 0.6 ± 0.2 V, P < 0.001). There was no patient with RVS + LBB and LVS + RVS capture mode in Tip Bipolar.

Figure 4
www.frontiersin.org

Figure 4. Ventricular sensing, impedance, and seven capture mode thresholds in three bilateral electrode pacing vector configurations.

QRS in seven capture modes

Full fusion mode had a similar P-QRS compared to that in RVS + LBB mode (116.9 ± 12.8 ms vs. 122.1 ± 13.5 ms), but was shorter compared to those of other modes (Figure 5). The P-QRS of select RVS capture mode was the longest (143.1 ± 20.9 ms).

Figure 5
www.frontiersin.org

Figure 5. V6 RWPT, V1 RWPT, and P-QRS in seven capture modes.

V6 RWPT in seven capture modes

Full fusion mode, LVS + LBB, RVS + LBB, and select LBB capture mode all had a short V6 RWPT and there were no significant differences among them (64.9 ± 9.7 ms vs. 65.7 ± 9.1 ms vs. 68.6 ± 12.5 ms vs. 62.8 ± 6.9 ms; Figure 5). Select RVS capture mode demonstrated a longer V6 RWPT than select LVS capture mode (100.1 ± 13.9 ms vs. 81.6 ± 10.1 ms, P < 0.001).

V1 RWPT in seven capture modes

Full fusion mode and RVS + LBB capture mode all showed the shortest V1 RWPT (94.5 ± 12.3 ms vs. 94.0 ± 23.6 ms; Figure 5). LVS capture and select LBB capture mode had the longest V1 RWPT (114.4 ± 16.8 ms vs. 113.9 ± 16.7 ms).

Full fusion mode in patients with native narrow and wide QRS

Full fusion mode was compared in subgroups of patients with a native narrow and wide QRS. A total of 40 patients had a native narrow QRS, and 17 patients had a native wide QRS (Figure 6). The P-QRS duration of full fusion mode was almost as short as the LB-QRS duration in intrinsic rhythm in patients with native narrow QRS (118.5 ± 10.8 ms vs. 112.8 ± 17.5 ms, P = 0.1). In patients with native wide QRS, the P-QRS duration of full fusion mode showed a statistically significant difference to the LB-QRS duration in intrinsic rhythm (120.1 ± 14.5 ms vs. 149.6 ± 19.7 ms, P < 0.05).

Figure 6
www.frontiersin.org

Figure 6. Comparison of P-QRS of full fusion mode and LB-QRS in intrinsic rhythm in patients with native narrow and wide QRS after LBBP.

Discussion

The main findings of the present study were as follows: (1) Three bilateral electrode pacing vector configurations can all achieve full fusion mode but with different capture thresholds. Bilateral Cathodes had the lowest threshold in full fusion mode, and Tip Bipolar had the highest threshold in full fusion mode. (2) Seven capture modes were observed in Ring Bipolar and Bilateral Cathodes during threshold testing, including full fusion mode (RVS + LVSP + LBB), semi fusion mode (LVS + LBB, RVS + LBB, and RVS + LVS), and select capture mode (LBB, LVS, and RVS). (3) Full fusion mode had the shortest P-QRS and V1 RWPT, the V6 RWPT remain the shortest. It achieved a short P-QRS in both narrow and wide native QRS types.

Different full fusion mode thresholds in three bilateral electrode pacing vector configurations

The present study results showed that Bilateral Cathodes had the lowest threshold in full fusion mode (1.2 ± 0.5 V), while Tip Bipolar had the highest (2.7 ± 1.0 V). This indicates that the full fusion mode threshold can be reduced by changing the pacing vector configuration. Cathodal stimulation of the cardiac septal myocardium occurs as a negative charge is applied to the myocardial–lead interface, which can direct cell depolarization in the region under the electrode and usually has a low capture threshold. On the contrary, anodal stimulation results in the accumulation of negatively charged ions at the myocardial–lead interface, which hyperpolarizes adjacent cells, so that the ability to trigger an action potential is paradoxical. However, anodal capture is routinely observed in cardiac pacing and usually requires a high pacing output, which can be explained by the bidomain cardiac tissue model that proposes to induce virtual cathodes at sites distant from the electrode to initiate depolarization (1719). Therefore, the myocardium capture threshold depends on whether it is stimulated by the anode or cathode. Since the RVS capture threshold was higher than that of LVS and LBB in unipolar cathodal stimulation (Table 2), and the myocardial RVS capture by anodal stimulation in Tip Bipolar resulted in a further decrease in RVS myocardial capture threshold, the full fusion mode capture threshold was the highest. In Bilateral Cathodes configuration, all of the LBB\LVS and RVS were captured by cathodal stimulation, so the full fusion mode capture threshold was the lowest. In Ring Bipolar configuration, the LVS and LBB capture threshold increased as they were captured by anodal stimulation, but the full fusion mode capture threshold was still lower than that in Tip Bipolar configuration. The LVS and LBB capture threshold was lower than that of RVS in unipolar cathodal. In addition, the full fusion mode threshold was much more affected by the Tip Bipolar configuration than the Ring Bipolar configuration, likely because the tip electrodes had different surrounding ventricular myocardial areas with completely different morphology from those of the ring electrodes.

Three new capture modes in ring bipolar and bilateral cathodes configurations

Only four capture modes were observed in conventional bipolar pacing mode (Tip Bipolar configuration), including full fusion mode (LVS + RVS + LBB), LVS + LBB, LBB, and LVS. Three new capture modes in Ring Bipolar and Bilateral Cathodes configurations were RVS + LBB, RVS + LVS, and RVS. It was speculated that LVS and LBB were captured by anodal stimulation in Ring Bipolar configuration, and the LVS and LBB capture threshold was increased and close to that of the RVS threshold. When the LVS capture threshold was higher than that of RVS and LBB, it changed to RVS + LBB capture mode. When the LBB capture threshold was higher than that of LVS and RVS, it changed to RVS + LVS capture mode. When the RVS capture threshold was lowest, it changed to a select RVS capture while further reducing the pacing output. In Bilateral cathodal configuration, the RVS, LVS, and LBB thresholds were also similar as they were all stimulated by the cathode. Thus, similar capture modes can be observed in Ring Bipolar configuration. On the contrary, RVS was captured by anodal stimulation in Tip Bipolar configuration, so the threshold was further increased and was much higher than that of LVS and LBB. This explained why only the LVS + LBB capture mode exists in Tip Bipolar configuration.

P-QRS\V6 RWPT\V1 RWPT of full fusion mode

The present study compared the P-QRS, V6 RWPT, and V1 RWPT of seven capture modes. Full fusion mode had the shortest P-QRS, V6 RWPT, and V1 RWPT. The P-QRS duration of full fusion mode was compared to LB-QRS duration in patients with a native narrow or wide QRS. The results showed that the full fusion mode can achieve a short P-QRS in patients with a native wide QRS. Full fusion mode significantly improved interventricular dyssynchrony and RV depolarization durations compared to other capture modes in LBBP.

Although, we also found there was a finding about bilateral bundle branch area pacing has been reported (20), which was similar to full fusion mode. But the LBB capture criteria was used in that study can not precisely confirm LBB capture as the direct evidence of LBB capture had not been reported at that time. The average V6 RWPT was about 82–84 ms in the study of bilateral bundle branch area pacing, which was longer than it was reported in our study (65–68 ms) and other study recently (64–69 ms) (21), it indicated there may had patients only capture LVS and without LBB capture. Full fusion mode was recognized by dynamic ECG and EGM maneuvers which can precisely confirm capture myocardial components in our study. So full fusion mode should be different from bilateral bundle branch area pacing.

Clinical implications

The capture threshold is high in conventional bipolar pacing mode in order to achieve full fusion mode. The present study showed that changing the bilateral electrode pacing vector configuration can reduce the full fusion mode capture threshold, especially in Bilateral Cathode and Ring Bipolar configurations. The CRT device already has multiple electrode pacing vector configurations. If this function can be applied in a normal pacemaker device, more patients can benefit from full fusion mode as it can achieve a more physiological ventricular depolarization than conventional LBBP can.

Limitations

Although it was demonstrated that changing the bilateral electrode pacing vector configuration can reduce the full fusion mode capture threshold, whether it can be used in a clinical setting remains unknown. The seven capture mode thresholds in three bilateral electrode pacing vector configurations were acute pacing parameters, long-term follow-up was necessary to confirm the threshold stable. Pacemaker manufacturers might need to make technical improvements to accommodate this technique. In addition, full fusion mode is a physiological pacing mode based on cardiac electrophysiological data. Whether it will result in a better hemodynamic response or less mechanical dyssynchrony in a clinical setting remains a question to be addressed. Clinical trials are necessary to further investigate the efficacy of this method. Changes in the LB lead EGM can help to distinguish RVS + LBB capture mode from full fusion mode since paced ECG morphology results were similar. There may be some difficulty in designating separate EGM in a few patients in Bilateral Cathodal mode as the EGM amplitude was low and may cause some deviation. Finally, the present study was performed at a single center and had a small sample size, which may cause bias in the statistical results.

Conclusion

The present study showed that changing the bilateral electrode pacing vector configuration to Bilateral Cathodes and Ring Bipolar configurations can reduce the full fusion mode capture threshold compared to conventional bipolar pacing. There were seven capture modes in Ring Bipolar and Bilateral Cathodes during threshold testing, including full fusion mode (RVS + LVSP + LBB), semi fusion mode (LVS + LBB, RVS + LBB, and RVS + LVS), and select capture mode (LBB, LVS, and RVS). The full fusion mode had the shortest P-QRS and V1 RWPT, the V6 RWPT remain the shortest.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by Ningbo No. 2 Hospital institutional review board (SL-KYSB-NBEY-2021-079-01). The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

HW: Writing – original draft. LJ: Writing – review & editing. JS: Writing – review & editing. LZ: Writing – review & editing.

Funding

The author(s) declare financial support was received for the research and/or publication of this article. This work was supported by the Ningbo Health Branding Subject Fund [grant number PPXK2018-01], Medical Scientific Research Foundation of Zhejiang Province, China [grant number 2022KY1137], Zhu Xiu Shan Talent Project of Ningbo No. 2 Hospital, China [grant number 2021HMYQ05] and Zhu Xiu Shan Talent Project of Ningbo No. 2 Hospital, China [grant number 2023HMYQ30].

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2025.1500196/full#supplementary-material

Supplementary Figure 1 | The changes in EGM and ECG during the threshold test of the three bilateral electrode pacing vector configurations.

References

1. Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman RA, Lee KL, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. (2003) 107(23):2932–7. doi: 10.1161/01.CIR.0000072769.17295.B1

PubMed Abstract | Crossref Full Text | Google Scholar

2. Nielsen JC, Kristensen L, Andersen HR, Mortensen PT, Pedersen OL, Pedersen AK. A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome: echocardiographic and clinical outcome. J Am Coll Cardiol. (2003) 42(4):614–23. doi: 10.1016/s0735-1097(03)00757-5

PubMed Abstract | Crossref Full Text | Google Scholar

3. Lustgarten DL, Crespo EM, Arkhipova-Jenkins I, Lobel R, Winget J, Koehler J, et al. His-bundle pacing versus biventricular pacing in cardiac resynchronization therapy patients: a crossover design comparison. Heart Rhythm. (2015) 12(7):1548–57. doi: 10.1016/j.hrthm.2015.03.048

PubMed Abstract | Crossref Full Text | Google Scholar

4. Huang W, Su L, Wu S, Xu L, Xiao F, Zhou X, et al. Long-term outcomes of his bundle pacing in patients with heart failure with left bundle branch block. Heart. (2019) 105(2):137–43. doi: 10.1136/heartjnl-2018-313415

PubMed Abstract | Crossref Full Text | Google Scholar

5. Subzposh FA, Vijayaraman P. Long-term results of his bundle pacing. Card Electrophysiol Clin. (2018) 10(3):537–42. doi: 10.1016/j.ccep.2018.05.011

PubMed Abstract | Crossref Full Text | Google Scholar

6. Barba-Pichardo R, Moriña-Vázquez P, Fernández-Gómez JM, Venegas-Gamero J, Herrera-Carranza M. Permanent his-bundle pacing: seeking physiological ventricular pacing. Europace. (2010) 12(4):527–33. doi: 10.1093/europace/euq038

PubMed Abstract | Crossref Full Text | Google Scholar

7. Huang W, Su L, Wu S, Xu L, Xiao F, Zhou X, et al. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the conduction block. Can J Cardiol. (2017) 33(12):1736.e1–3. doi: 10.1016/j.cjca.2017.09.013

PubMed Abstract | Crossref Full Text | Google Scholar

8. Vijayaraman P, Subzposh FA, Naperkowski A, Panikkath R, John K, Mascarenhas V, et al. Prospective evaluation of feasibility and electrophysiologic and echocardiographic characteristics of left bundle branch area pacing. Heart Rhythm. (2019) 16(12):1774–82. doi: 10.1016/j.hrthm.2019.05.011

PubMed Abstract | Crossref Full Text | Google Scholar

9. Chen K, Li Y, Dai Y, Sun Q, Luo B, Li C, et al. Comparison of electrocardiogram characteristics and pacing parameters between left bundle branch pacing and right ventricular pacing in patients receiving pacemaker therapy. Europace. (2019) 21(4):673–80. doi: 10.1093/europace/euy252

PubMed Abstract | Crossref Full Text | Google Scholar

10. Su L, Wang S, Wu S, Xu L, Huang Z, Chen X, et al. Long-term safety and feasibility of left bundle branch pacing in a large single-center study. Circ Arrhythm Electrophysiol. (2021) 14(2):e009261. doi: 10.1161/CIRCEP.120.009261

PubMed Abstract | Crossref Full Text | Google Scholar

11. Curila K, Jurak P, Jastrzebski M, Prinzen F, Waldauf P, Halamek J, et al. Left bundle branch pacing compared to left ventricular septal myocardial pacing increases interventricular dyssynchrony but accelerates left ventricular lateral wall depolarization. Heart Rhythm. (2021) 18(8):1281–9. doi: 10.1016/j.hrthm.2021.04.025

PubMed Abstract | Crossref Full Text | Google Scholar

12. Wu H, Jiang L, Shen J, Zhang L, Zhong J, Zhuo S. Electrophysiological characteristics and possible mechanism of bipolar pacing in left bundle branch pacing. Heart Rhythm. (2022) 19(12):2019–26. doi: 10.1016/j.hrthm.2022.06.022

PubMed Abstract | Crossref Full Text | Google Scholar

13. Curila K, Jurak P, Prinzen F, Jastrzebski M, Waldauf P, Halamek J, et al. Bipolar anodal septal pacing with direct LBB capture preserves physiological ventricular activation better than unipolar left bundle branch pacing. Front Cardiovasc Med. (2023) 10:1140988. doi: 10.3389/fcvm.2023.1140988

PubMed Abstract | Crossref Full Text | Google Scholar

14. Shen J, Jiang L, Cai X, Wu H, Pan L. Left bundle branch pacing guided by continuous pacing technique that can monitor electrocardiograms and electrograms in real time: a technical report. Can J Cardiol. (2022) 38(8):1315–7. doi: 10.1016/j.cjca.2022.03.003

PubMed Abstract | Crossref Full Text | Google Scholar

15. Wu H, Jiang L, Shen J. Recording an isoelectric interval as an endpoint of left bundle branch pacing with continuous paced intracardiac electrogram monitoring. Kardiol Pol. (2022) 80(6):664–71. doi: 10.33963/KP.a2022.0094

PubMed Abstract | Crossref Full Text | Google Scholar

16. Shen J, Jiang L, Wu H, Li H, Zhang L, Zhong J, et al. High-pass filter settings and the role and mechanism of discrete ventricular electrograms in left bundle branch pacing. Front Cardiovasc Med. (2023) 9:1059172. doi: 10.3389/fcvm.2022.1059172

PubMed Abstract | Crossref Full Text | Google Scholar

17. Lloyd MS, Heeke S, Lerakis S, Langberg JJ. Reverse polarity pacing: the hemodynamic benefit of anodal currents at lead tips for cardiac resynchronization therapy. J Cardiovasc Electrophysiol. (2007) 18(11):1167–71. doi: 10.1111/j.1540-8167.2007.00943.x

PubMed Abstract | Crossref Full Text | Google Scholar

18. Ranjan R, Chiamvimonvat N, Thakor NV, Tomaselli GF, Marban E. Mechanism of anode break stimulation in the heart. Biophys J. (1998) 74(4):1850–63. doi: 10.1016/S0006-3495(98)77895-6

PubMed Abstract | Crossref Full Text | Google Scholar

19. Roth BJ, Wikswo JP Jr. Electrical stimulation of cardiac tissue: a bidomain model with active membrane properties. IEEE Trans Biomed Eng. (1994) 41(3):232–40. doi: 10.1109/10.284941

PubMed Abstract | Crossref Full Text | Google Scholar

20. Lin J, Chen K, Dai Y, Sun Q, Li Y, Jiang Y, et al. Bilateral bundle branch area pacing to achieve physiological conduction system activation. Circ Arrhythm Electrophysiol. (2020) 13(8):e008267. doi: 10.1161/CIRCEP.119.008267

PubMed Abstract | Crossref Full Text | Google Scholar

21. Chen K, Liu X, Hou X, Qiu Y, Lin J, Dai Y, et al. Computed tomography imaging-identified location and electrocardiographic characteristics of left bundle branch area pacing in bradycardia patients. J Cardiovasc Electrophysiol. (2022) 33(6):1244–54. doi: 10.1111/jce.15493

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: bipolar pacing, conduction system pacing, capture threshold, electrode configuration, left bundle branch pacing

Citation: Wu H, Jiang L, Shen J and Zhang L (2025) Comparison of electrophysiological left bundle branch pacing characteristics in different bilateral electrode pacing vector configurations. Front. Cardiovasc. Med. 12:1500196. doi: 10.3389/fcvm.2025.1500196

Received: 22 September 2024; Accepted: 19 February 2025;
Published: 12 March 2025.

Edited by:

Marco Schiavone, Monzino Cardiology Center (IRCCS), Italy

Reviewed by:

Gabriel Cismaru, University of Medicine and Pharmacy Iuliu Hatieganu, Romania
Richard Gary Trohman, Rush University, United States
Uğur Canpolat, Hacettepe University, Türkiye

Copyright: © 2025 Wu, Jiang, Shen and Zhang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Longfu Jiang, bG9uZ2Z1amlhbmdod2FtZWlAMTYzLmNvbQ==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Research integrity at Frontiers

Man ultramarathon runner in the mountains he trains at sunset

95% of researchers rate our articles as excellent or good

Learn more about the work of our research integrity team to safeguard the quality of each article we publish.


Find out more