Skip to main content

ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 10 April 2024
Sec. Cardiac Rhythmology

Comparison of infection and complication rates associated with transvenous vs. subcutaneous defibrillators in patients with stage 4 chronic kidney disease: a multicenter long-term retrospective follow-up

\r\nFabian Schiedat,
Fabian Schiedat1,2*Benjamin MeuterodtBenjamin Meuterodt3Magnus PrullMagnus Prull4Assem AweimerAssem Aweimer1Michael GotzmannMichael Gotzmann5Stephen O&#x;ConnorStephen O’Connor6Christian PeringsChristian Perings3Johannes KorthJohannes Korth7Thomas LawoThomas Lawo8Ibrahim El-Battrawy,Ibrahim El-Battrawy1,9Christoph HanefeldChristoph Hanefeld5Andreas Mügge,Andreas Mügge1,5Axel Kloppe,\r\nAxel Kloppe1,2
  • 1Department of Cardiology and Angiology, University Hospital Bergmannsheil Bochum of the Ruhr-University Bochum, Bochum, Germany
  • 2Department of Cardiology and Angiology at Marienhospital Gelsenkirchen, Academic Hospital of the Ruhr University Bochum, Gelsenkirchen, Germany
  • 3Department of Cardiology, Electrophysiology, Pneumology and Intensive Care Medicine, St. Marien-Hospital Luenen, Academic Hospital of the University Muenster, Luenen, Germany
  • 4Department of Cardiology, Augusta Hospital Bochum, Academic Hospital of the University Duisburg-Essen, Bochum, Germany
  • 5Department of Cardiology, Katholische Kliniken Bochum of the Ruhr University Bochum, Bochum, Germany
  • 6Department of Biomedical Engineering, City, University of London, London, United Kingdom
  • 7Department of Nephrology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
  • 8Department of Cardiology, Elisabeth Hospital Recklinghausen, Recklinghausen, Germany
  • 9Department of Molecular and Experimental Cardiology, Institut für Forschung und Lehre (IFL), Ruhr-University Bochum, Bochum, Germany

Background: Patients with progressive chronic kidney disease (CKD) are at higher risk of infections and complications from cardiac implantable electronic devices (CIED). In patients with a primary or secondary prophylactic indication, implantable cardiac defibrillators (ICD) can prevent sudden cardiac deaths (SCD). We retrospectively compared transvenous-ICD (TV-ICD) and intermuscularly implanted subcutaneous-ICD (S-ICD) associated infections and complication rates together with hospitalizations in recipients with stage 4 kidney disease.

Methods: We retrospectively analyzed 70 patients from six German centers with stage 4 CKD who received either a prophylactic TV-ICD with a single right ventricular lead, 49 patients, or a S-ICD, 21 patients. Follow-Ups (FU) were performed bi-annually.

Results: The TV-ICD patients were significantly older. This group had more patients with a history of atrial arrhythmias and more were prescribed anti-arrhythmic medication compared with the S-ICD group. There were no significant differences for other baseline characteristics. The median and interquartile range of FU durations were 55.2 (57.6–69.3) months. During FU, patients with a TV-ICD system experienced significantly more device associated infections (n = 8, 16.3% vs. n = 0; p < 0.05), device-associated complications (n = 13, 26.5% vs. n = 1, 4.8%; p < 0.05) and device associated hospitalizations (n = 10, 20.4% vs. n = 1, 4.8%; p < 0.05).

Conclusion: In this long-term FU of patients with stage 4 CKD and an indication for a prophylactic ICD, the S-ICD was associated with significantly fewer device associated infections, complications and hospitalizations compared with TV-ICDs.

Introduction

Implantable cardioverter defibrillators (ICD) are indicated for heart failure (NYHA II or III) with a reduced left-ventricular ejection fraction (LV-EF) ≤ 35% despite optimal medical treatment for more than 3 months, as well as patients who have recovered from a hemodynamic unstable ventricular arrhythmias without a reversible cause (1, 2). Patients are also required to have a life expectancy of more than one year with a good quality survival. Patients without the need for bradycardia pacing, anti-tachycardia pacing or cardiac resynchronization therapy (CRT) may receive either a subcutaneous implantable cardioverter defibrillator (S-ICD) or a transvenous implantable cardioverter defibrillator (TV-ICD) (1).

With progressive chronic kidney disease (CKD), patients are at an increased risk of cardiovascular mortality and morbidity, which still exists after correction for typical concomitant risk factors (3). A main trigger is believed to be an inflammatory reaction in CKD (4) which has been identified as an independent risk factor for cardiac implantable electronic device (CIED) infections and complications (5). Patients with a glomerular filtration rate (GFR) < 30 ml/min/1.73 m2 are a CKD subgroup exposed to a risk for infections and lead complications (6) whilst those with a GFR 15–29 ml/min/1.73 m2 for at least three months are defined as stage 4 CKD and at even higher risk (7).

The S-ICD has been associated with lower rates for lead related complications during long-term follow-up (FU) while demonstrating comparable safety and efficacy compared to TV-ICDs (8). S-ICD device-related infections are less frequent as there are no leads on or in the heart. If they do occur, S-ICD infections can be handled conservatively without the need for urgent removal as is the case with TV-ICDs, where transvenous lead extraction (TLE) is associated with high mortality and morbidity (9, 10). Very limited long-term data on infection and complication rates in patients with progressive CKD receiving an ICD have been published. There is however no comparative data in this compromised cohort of CKD patients between TV-ICD and S-ICD recipients. The aim of the present study was to compare long-term device associated safety and outcomes in patients with stage 4 CKD who received either an TV-ICD or S-ICD with a single right ventricular lead.

Methods

Out of the population of patients undergoing TV-ICD or S-ICD implantation, those with stage 4 CKD, GFR 15–29 ml/min/1.73 m2 for at least 3 months, were included in this retrospective analysis from six experienced centers in Germany (University Hospital Bergmannsheil Bochum, University Hospital Katholische Kliniken St. Josef Bochum, Marienhospital Gelsenkirchen, Augusta Hospital Bochum, Marien-Hospital Luenen and Elisabeth Hospital Recklinghausen). Patients on hemodialysis were excluded from this study as hemodialysis is an independent risk factor for device infection and complication (11). Patients with stage 5 CKD without dialysis were excluded as well, as there is a very high likelihood for hemodialysis during FU compared to stage 4 CKD (12).

The devices were implanted between 2012 and 2020. We obtained informed consent from all patients. The study protocol conforms to the ethical guidelines of the 1964 Declaration of Helsinki and its later amendments. It was approved by the local ethics committees of the Ruhr-University, Bochum as the leading ethics committee (Register 22-7593-BR). All patients had indications for a cardiac defibrillator implantation following the guidelines and a life-expectancy >1 year (1, 2).

Implantation procedure

Implantation was performed under local anesthesia combined with deep sedation. Intravenous antibiotic prophylaxis as a single dose was given prior to the procedure with 2 grams of cefazolin or alternatively 1 gram of vancomycin. The decision to implant either an TV-ICD or S-ICD was done according to operators' discretion.

For TV-ICD implantation, the pulse generator was placed in a left-sided, pre-pectoral, sub-fascial position. The lead was placed via the cephalic, axillary or subclavian vein, in a mid-septal right ventricular position and sutures were applied twice at the sleeve and once at the generator. An inter-muscular approach was used for the S-ICD implantation (13). The device (Emblem S-ICD, Model 209, Boston Scientific, Marlborough, MA, USA or SQ-RX 1010, Cameron Health, San Clemente, CA, USA) was placed in a left postero-lateral position between the anterior surface of serratus anterior and the posterior surface of the latissimus dorsi over the left fifth or sixth rib in an intermuscular space between the mid and posterior axillary lines. The S-ICD electrode was implanted left parasternally via a two- incision technique and sutures were applied at all incision sites. The device was placed in the pocket and sutures were applied at both sides of the generator. Induced arrhythmia conversion testing was performed once the electrode and pulse generator were in their respective positions prior to closure. Induction of ventricular fibrillation was facilitated by 50 Hz stimulation burst between the shock coil and pulse generator and terminated by a 65-Joule shock thereby giving a 15-Joule safety margin. In case of non-conversion of the first induction, the position of the electrode and pulse generator were checked with fluoroscopy and repositioned, if necessary. Testing was then repeated with the option for reversing shock polarity. Complications that were life-threatening, with permanent adverse sequelae or resulted in death were defined as major complication. All other procedure related complications were defined as minor complications.

Patient follow-up

All patients were seen prior to discharge for wound control as well as device interrogation and programming. All three S-ICD sensing vectors were tested in a supine and standing positions. Patients were followed up bi-annually in the outpatient department. Device interrogation, signs of device associated complications and infection, as well as hospitalizations, were documented. Where a patient failed to attend their FU, they and/or their family physician were contacted to confirm whether they were alive. Complications, hospitalizations and other events were thereby ascertained and documented. If hemodialysis was required during FU, the FU period for these patients ended with the day of the first hemodialysis treatment.

Data collection

Data has been collected continuously but analyzed retrospectively and audited with hospitals' clinical information system.

Statistical analysis

All statistical analyses were performed using IBM SPSS Statistics Version 28.0.0 on Mac. Categorial variables were expressed as frequencies and percentages (normal distribution) or median and interquartile range (non-normal distribution). Continuous variables were stated as mean ± standard deviation. Assessment of descriptive statistics and baseline characteristics was done using linear regression. If necessary, Chi-quadrat test was used for normally distributed, non-normally distributed and binary data with linear trends. Time to primary endpoints were performed using Kaplan–Meier analyses and compared with the log-rank test.

Results

Patient population and implant procedure

A total of 70 patients from six German centers with stage 4 CKD and indication for ICD implantation were included in this study. All procedures were performed by six experienced operators, one at each center. During the same period of time, a total of 2,757 prophylactic TV-ICDs with single right ventricular lead and S-ICD implantations were performed at the six participating centers. Of the 70 patients with stage 4 CKD in this study, 49 patients received a TV-ICD with a single right ventricular, single-coil lead, and 21 patients received a S-ICD. The indication was primary prophylactic in 73.5% of the TV-ICD and 85.7% of the S-ICD population (p = 0.267). The TV-ICD population was significantly older (73.0 (68.5–78.0) years vs. 59.0 (50.0–67.5) years; p < 0.01) than the S-ICD population. In addition, the TV-ICD cohort had a significantly higher burden of atrial fibrillation and received more anti-arrhythmic medication and cardiac glycosides. There were no significant differences in other baseline characteristics. All baseline characteristics are shown in Table 1. All S-ICD implantation procedures were successful with a 15-Joule safety margin being established from induced arrhythmia conversion testing, with no patients requiring an electrode position change or retest. Implant duration was not significantly different between defibrillator types.

Table 1
www.frontiersin.org

Table 1. Baseline and procedure characteristics.

There were no major complications during the intra-operative or the hospitalization period for implantation in either group. Minor perioperative complications occurred in 6 (12.2%) and 1 (4.7%) patients (p < 0.01) in the respective TV-ICD and S-ICD groups. There was no significant difference between groups for patients requiring an intervention for these complications (TV-ICD: n = 1, 2.0% vs. S-ICD: n = 0, p = 0.689). In the TV-ICD group, 1 (2.0%) patient experienced a minor vascular complications, 2 (4.1%) patients a pocket hematoma, 2 (4.1%) patients a pneumothorax and 1 (2.0%) patient a lead related complication due to undersensing within 2 weeks of implantation. Only the latter patient required an intervention, an RV lead revision. All other patients were managed conservatively without sequelae. In the S-ICD group, 1 (4.8%) patient had a pocket hematoma before discharge. This was handled conservatively without the need for a pocket revision.

All implant procedure characteristics and complications are shown in Table 1.

Follow-up

Duration of FU was 55.1 (48.2–64.4) months in the TV-ICD group and 55.4 (46.8–75.0) months in the S-ICD group (p = 0.790). During FU, CIED associated infections occurred significantly more often in patients in the TV-ICD group (TV-ICD: n = 8, 16.3% vs. S-ICD: n = 0; p < 0.05). This is illustrated in Figure 1. Time to device infection in the TV-ICD group was 29.2 ± 10.6 months with all cases requiring TLE and generator removal.

Figure 1
www.frontiersin.org

Figure 1. Kaplan–Meier graph illustrating the freedom from device associated infections during follow-up. TV-ICD, transvenous ICD; S-ICD, subcutaneous ICD.

CIED-associated complications without infection again occurred significantly more frequently in the TV-ICD group (TV-ICD: n = 13, 26.5% vs. S-ICD: n = 1, 4.8%, p < 0.05) as illustrated in Figure 2. The complications in the TV-ICD group were lead fracture with inadequate therapy delivery due to oversensing (n = 3), loss of right ventricular sensing <3 mV (n = 2), loss of capture (n = 2), drop in impedance below 250 Ohm (n = 1), increase in impedance above 1,500 Ohm (n = 2) and other non-infectious pocket or generator related complications (n = 3). Lead related complications in the TV-ICD group therefore occurred in 10 (20.4%) patients compared with 1 (4.8%) patient in the S-ICD group due to generator dislodgement during FU which required repositioning. The difference in lead related complications was significant (p < 0.05).

Figure 2
www.frontiersin.org

Figure 2. Kaplan–Meier graph illustrating the freedom from device associated complication during follow-up. TV-ICD, transvenous ICD; S-ICD, subcutaneous ICD.

Overall device associated hospitalizations occurred significantly more often in the TV-ICD group (TV-ICD: n = 18, 36.7% vs. S-ICD: n = 1, 4.8%; p < 0.001). There was no difference between groups for documented ventricular arrhythmias, ventricular arrhythmia episodes per patient or for appropriate or inappropriate shocks during FU. Anti-tachycardia pacing (ATP) was delivered to 9 (18.3%) of patients in the TV-ICD group but is not available from the S-ICD. During FU, no patient required a change of system for a bradycardia pacing in the S-ICD group or for cardiac resynchronization in either group. Outcome results did not differ between primary and secondary prophylactic indications for both cohorts. Hemodialysis was required in 3 (6.1%) and 2 (9.5%) patients in the respective TV-ICD and S-ICD groups without a significant difference (p = 0.318). All variables monitored during FU are shown in Table 2.

Table 2
www.frontiersin.org

Table 2. Results from the follow-up.

Discussion

The aim of the present study was to investigate whether there was a difference in TV-ICD and S-ICD associated infections, complications and hospitalizations in patients with stage 4 CKD who received a prophylactic ICD.

Patients with progressive CKD exhibit an increased risk for heart failure, ventricular arrhythmias and sudden cardiac death (SCD) (3). This group of patients has been excluded from most studies demonstrating a reduced mortality for ICD compared with drug therapy (1, 14). Patients with progressive CKD (GFR <30 ml/min/1.73 m2) were excluded from the S-ICD Investigational Device Exemption Trial (15). Unsurprisingly, there are no specific recommendations in the ICD therapy guidelines for patients with progressive CKD.

The role of ICDs in this seriously compromised group of patients remains unclear as large studies failed to show a survival benefit from implanting ICDs in patients with CKD (16). Higher CIED infection rates are assumed to be a trigger for higher hospitalization rates in patients with CKD who received an ICD compared with control groups with CKD who did not receive an ICD (17). We therefore postulated that in a similar group of compromised patients, there would be a lower CIED infection and device related hospitalization rate in S-ICD recipients as there are no lead on or in the heart.

Baseline demographics are comparable to a cohort study of 5,877 patients with CKD and heart failure reported by Bansal et al. (16). However, the number of patients with coronary artery disease, history of myocardial infarction and ischemic cardiomyopathy is higher in our cohort whilst the age in our S-ICD group was significantly younger. Implanting physicians are more likely to choose an S-ICD in younger patients, as they expect fewer complications in patients with CKD due to preservation of the vasculature. In addition, there was a higher likelihood in younger patients for a further progression of the CKD with the need for hemodialysis (HD). The feasibility of creating an arteriovenous fistula in patients requiring HD can be complicated by the presence of a CIED in the vasculature. Venous occlusion or central venous stenosis has been reported in 11%–36% patients with transvenous CIED (18).

The age difference between our TV-ICD and S-ICD groups could be an explanation for a higher rate of atrial arrhythmia in the TV-ICD group and hence a higher proportion of patients receiving anti-arrhythmic medication and cardiac glycosides. With a higher incidence of atrial arrhythmias and anti-arrhythmic medication there can be concerns regarding a higher drug accumulation due to lower renal clearance (1921). This could be an important issue regarding morbidity and mortality data in future investigations. The significant age difference in our study between both groups could favor the beneficial results for the S-ICD, however data from a TV-ICD meta-analysis and a S-ICD study suggest that device associated complications are not associated with higher age (22, 23).

Implantation duration of our groups was comparable with other large studies comparing TV-ICD and S-ICD therapies (24). Patients with progressive CKD experience a significantly higher proportion of intraoperative complications, especially bleeding, vascular injury and in-hospital mortality according to an analysis of a nationwide database from the United States of 40,075 cases with ICD implantation (25). Minor procedural complications occurred significantly more often in our TV-ICD cohort. The procedural complication rate of 4.8% in our S-ICD cohort was comparable to a 3.8% procedural complication rate of 429 S-ICD procedures reported by Knops et al. (24). The rate of procedural complications associated with TV-ICD implantation was higher in our study (12.2%) compared with a rate of 4.7% in 423 procedures reported by Knops et al. (24). A high rate of oral anti-coagulation in our TV-ICD cohort and a high rate of complications in patients with CKD receiving a TV-ICD are explanations (16). However, no difference between our groups for procedural complications requiring surgical or other interventions has been reported.

Patients with progressive CKD are at an exponentially increased risk for death with decreasing kidney function and increasing cardiovascular mortality being the leading cause of death, rather than renal failure (3, 26). Progressive CKD is associated with a wide range of cardiac and non-cardiac features providing a substrate for vulnerable myocardium and increased risk of arrhythmias and consequently SCD (27, 28). The incidence of ventricular arrhythmias together with adequate and inadequate therapy in both ICD groups, and their lack of significance, is comparable with a large cohort reported by Russo et al. (29). Device lifetime infection rate is 2%–3% for TV-ICD systems in a general population according to the Danish ICD registry (30). CKD is an independent risk factor for device infections (31). In a general population, CIED infections occur significantly more often in TV-ICD patients compared with S-ICD patients (29). If infections with S-ICD systems occur, these can frequently be handled conservatively with antibiotics without the need for removal, whereas TV-ICD systems need to be explanted as soon as possible, often with complicated TLE procedures (15, 32). Our data is in line with these reports as there were no infections in the intermuscularly implanted S-ICD group and a significantly higher proportion of 16.3% CIED infections in the TV-ICD group. The high prevalence of device associated infections in the TV-ICD group can be explained by the progressive CKD. In S-ICD patients with progressive CKD, even those with HD, do not have an increased incidence of CIED infections (33, 34). If however a device removal of an S-ICD would be required, the procedure is safe and easy to perform (35).

Patients with a TV-ICD system and CKD are not only exposed to a high risk of CIED infection, but also other device associated complications. Jukema et al. reported an overall TV-ICD associated complication rate in 80 (31.3%) patients with end-stage CKD, of which 4 (5.0%) patients required ICD removal for bacteremia during a median FU of 6.8 ± SD years (36). A hospitalization rate for device-associated complications of 36.7% in our TV-ICD groups seems reasonable for a high risk population considering a pooled complication-rate, excluding inappropriate therapy, of 9.1% in a large TV-ICD meta-analysis by Ezzat et al. during an average FU of 17.9 months and an annual rate of 12.0% for complications requiring surgical intervention in prospective registry reported by Hawkins et al. of 3410 ICD recipients in a general population during a median FU of 34 months (22, 37). A meta-analysis of 2,387 patients by Fong et al. reported that compared with TV-ICD patients, S-ICD patients experienced significantly fewer (RR: 0.14, 95% CI: 0.07–0.29, p < 0.0001) lead related complications (8). These findings are supported by our data showing significantly fewer CIED-associated complications, especially lead related complications, in the S-ICD group during a long-term FU.

Limitations

Even though six centers participated, the sample size remains small. A population with a GFR between 15 and 29 ml/min/1.73 m2, an ICD indication as well as an expected good quality survival of more than one year is rare.

This study represents the only comparative analysis, albeit retrospective, of TV-ICD and S-ICD recipients with stage 4 CKD. The study was neither powered nor designed to investigate mortality or overall hospitalization rates. Due to small sample size we were only able to perform basic statistical analysis without multivariate analysis and adjustments. A large prospective, appropriately powered, randomized trial is required to verify the results presented in this study and draw further conclusions. Overall mortality and hospitalization in each of the ICD groups should be examined in the proposed trial as well as a third group of patients with the same baseline characteristics who do not receive an ICD.

Conclusion

During a long-term follow-up of patients with stage 4 chronic kidney disease at high risk for ICD related complications, the intermuscularly implanted S-ICD has significantly fewer device-associated infections and complications compared with TV-ICDs.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by Ethics Committee of the Ruhr University Bochum. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

FS: Conceptualization, Data curation, Formal Analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing. BM: Conceptualization, Data curation, Methodology, Validation, Writing – review & editing. MP: Data curation, Formal Analysis, Software, Writing – review & editing. AA: Conceptualization, Data curation, Validation, Visualization, Writing – original draft. MG: Data curation, Investigation, Software, Writing – review & editing. SO: Data curation, Formal Analysis, Supervision, Writing – original draft. CP: Data curation, Investigation, Validation, Writing – review & editing. JK: Formal Analysis, Software, Validation, Visualization, Writing – review & editing. TL: Data curation, Formal Analysis, Investigation, Methodology, Writing – review & editing. IE: Conceptualization, Formal Analysis, Project administration, Writing – review & editing. CH: Data curation, Software, Validation, Visualization, Writing – review & editing. AM: Formal Analysis, Methodology, Project administration, Validation, Writing – original draft. AK: Data curation, Formal Analysis, Project administration, Validation, Writing – original draft.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Conflict of interest

FS Received speakers' honoraria from Boston Scientific, Abbott and Impulse Dynamics. MP received speakers' honoraria from Boston Scientific. MG received speakers' honoraria from Abbott, Bristol-Myers Squibb, Novartis, and Pfizer and is an advisor for Lilly and Boehringer Ingelheim. SO is a former employee of Cameron Health and Boston Scientific. AM received speakers' honoraria from Pfizer Pharma, Novartis Pharma and Bristol Myers Squib Pharma. AK received speakers' honoraria from Boston Scientific, Abbott, Medtronic and Impulse Dynamics.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Zeppenfeld K, Tfelt-Hansen J, De Riva M, Winkel BG, Behr ER, Blom NA, et al. 2022 ESC guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Eur Heart J. (2022) 43(40):3997–4126. doi: 10.1093/eurheartj/ehac262

PubMed Abstract | Crossref Full Text | Google Scholar

2. McDonagh TA, Metra M, Adamo M, Gardner RS, Baumbach A, Böhm M, et al. 2021 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur Heart J. (2021) 42(36):3599–726. doi: 10.1093/eurheartj/ehab368

PubMed Abstract | Crossref Full Text | Google Scholar

3. Jankowski J, Floege J, Fliser D, Böhm M, Marx N. Cardiovascular disease in chronic kidney disease: pathophysiological insights and therapeutic options. Circulation. (2021) 143(11):1157–72. doi: 10.1161/CIRCULATIONAHA.120.050686

PubMed Abstract | Crossref Full Text | Google Scholar

4. Zoccali C, Vanholder R, Massy ZA, Ortiz A, Sarafidis P, Dekker FW, et al. The systemic nature of CKD. Nat Rev Nephrol. (2017) 13(6):344–58. doi: 10.1038/nrneph.2017.52

PubMed Abstract | Crossref Full Text | Google Scholar

5. Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. EP Eur. (2015) 17(5):767–77. doi: 10.1093/europace/euv053

Crossref Full Text | Google Scholar

6. Buiten MS, De Bie MK, Van Der Heijden AC, Rotmans JI, Bootsma M, Marc Groeneveld JH, et al. Chronic kidney disease and implantable cardioverter defibrillator related complications: 16 years of experience. J Cardiovasc Electrophysiol. (2014) 25(9):998–1004. doi: 10.1111/jce.12435

PubMed Abstract | Crossref Full Text | Google Scholar

7. Chapter 1: definition and classification of CKD. Kidney Int Suppl. (2013) 3(1):19–62. doi: 10.1038/kisup.2012.64

Crossref Full Text | Google Scholar

8. Fong KY, Ng CJR, Wang Y, Yeo C, Tan VH. Subcutaneous versus transvenous implantable defibrillator therapy: a systematic review and meta-analysis of randomized trials and propensity score–matched studies. J Am Heart Assoc. (2022) 11(11):e024756. doi: 10.1161/JAHA.121.024756

PubMed Abstract | Crossref Full Text | Google Scholar

9. Knops RE, Pepplinkhuizen S, Delnoy PPHM, Boersma LVA, Kuschyk J, El-Chami MF, et al. Device-related complications in subcutaneous versus transvenous ICD: a secondary analysis of the PRAETORIAN trial. Eur Heart J. (2022) 43(47):4872–83. doi: 10.1093/eurheartj/ehac496

PubMed Abstract | Crossref Full Text | Google Scholar

10. Mehta VS, Elliott MK, Sidhu BS, Gould J, Kemp T, Vergani V, et al. Long-term survival following transvenous lead extraction: importance of indication and comorbidities. Heart Rhythm. (2021) 18(9):1566–76. doi: 10.1016/j.hrthm.2021.05.007

PubMed Abstract | Crossref Full Text | Google Scholar

11. Asif A, Salman L, Lopera G, Haqqie SS, Carrillo R. Transvenous cardiac implantable electronic devices and hemodialysis catheters: recommendations to curtail a potentially lethal combination. Semin Dial. (2012) 25(5):582–6. doi: 10.1111/j.1525-139X.2012.01053.x

PubMed Abstract | Crossref Full Text | Google Scholar

12. Lin CM, Yang MC, Hwang SJ, Sung JM. Progression of stages 3b–5 chronic kidney disease—preliminary results of Taiwan national Pre-ESRD disease management program in southern Taiwan. J Formos Med Assoc. (2013) 112(12):773–82. doi: 10.1016/j.jfma.2013.10.021

PubMed Abstract | Crossref Full Text | Google Scholar

13. Winter J, Siekiera M, Shin DI, Meyer C, Kröpil P, Clahsen H, et al. Intermuscular technique for implantation of the subcutaneous implantable cardioverter defibrillator: long-term performance and complications. EP Eur. (2017) 19(12):2036–41. doi: 10.1093/europace/euw297

Crossref Full Text | Google Scholar

14. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al. Amiodarone or an implantable cardioverter–defibrillator for congestive heart failure. N Engl J Med. (2005) 352(3):225–37. doi: 10.1056/NEJMoa043399

PubMed Abstract | Crossref Full Text | Google Scholar

15. Weiss R, Knight BP, Gold MR, Leon AR, Herre JM, Hood M, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. (2013) 128(9):944–53. doi: 10.1161/CIRCULATIONAHA.113.003042

PubMed Abstract | Crossref Full Text | Google Scholar

16. Bansal N, Szpiro A, Reynolds K, Smith DH, Magid DJ, Gurwitz JH, et al. Long-term outcomes associated with implantable cardioverter defibrillator in adults with chronic kidney disease. JAMA Intern Med. (2018) 178(3):390–8. doi: 10.1001/jamainternmed.2017.8462

PubMed Abstract | Crossref Full Text | Google Scholar

17. Singh SM, Wang X, Austin PC, Parekh RS, Lee DS. Prophylactic defibrillators in patients with severe chronic kidney disease. JAMA Intern Med. (2014) 174(6):995. doi: 10.1001/jamainternmed.2014.1208

PubMed Abstract | Crossref Full Text | Google Scholar

18. Nowak K, Kusztal M. Cardiac implantable electronic devices in hemodialysis and chronic kidney disease patients—an experience-based narrative review. J Clin Med. (2021) 10(8):1745. doi: 10.3390/jcm10081745

PubMed Abstract | Crossref Full Text | Google Scholar

19. Turakhia MP, Blankestijn PJ, Carrero JJ, Clase CM, Deo R, Herzog CA, et al. Chronic kidney disease and arrhythmias: conclusions from a kidney disease: improving global outcomes (KDIGO) controversies conference. Eur Heart J. (2018) 39(24):2314–25. doi: 10.1093/eurheartj/ehy060

PubMed Abstract | Crossref Full Text | Google Scholar

20. Goto S, Angchaisuksiri P, Bassand J, Camm AJ, Dominguez H, Illingworth L, et al. Management and 1-year outcomes of patients with newly diagnosed atrial fibrillation and chronic kidney disease: results from the prospective GARFIELD-AF registry. J Am Heart Assoc. (2019) 8(3):e010510. doi: 10.1161/JAHA.118.010510

PubMed Abstract | Crossref Full Text | Google Scholar

21. Vamos M, Oldgren J, Nam GB, Lip GYH, Calkins H, Zhu J, et al. Dronedarone vs. placebo in patients with atrial fibrillation or atrial flutter across a range of renal function: a post hoc analysis of the ATHENA trial. Eur Heart J Cardiovasc Pharmacother. (2022) 8(4):363–71. doi: 10.1093/ehjcvp/pvab090

PubMed Abstract | Crossref Full Text | Google Scholar

22. Ezzat VA, Lee V, Ahsan S, Chow AW, Segal O, Rowland E, et al. A systematic review of ICD complications in randomised controlled trials versus registries: is our ‘real-world’ data an underestimation? Open Heart. (2015) 2(1):e000198. doi: 10.1136/openhrt-2014-000198

PubMed Abstract | Crossref Full Text | Google Scholar

23. Schiavone M, Gasperetti A, Gulletta S, Steffel J, Kaiser L, Mitacchione G, et al. Age-related differences and associated outcomes of S-ICD: insights from a large, European, multicenter, real-world registry. EP Eur. (2022) 24(Suppl 1):euac053-446. doi: 10.1093/europace/euac053.446

Crossref Full Text | Google Scholar

24. Knops RE, Olde Nordkamp LRA, Delnoy PPHM, Boersma LVA, Kuschyk J, El-Chami MF, et al. Subcutaneous or transvenous defibrillator therapy. N Engl J Med. (2020) 383(6):526–36. doi: 10.1056/NEJMoa1915932

PubMed Abstract | Crossref Full Text | Google Scholar

25. Ayoub K, Fry E, Marji M, Masri A, Hesselson A, Ellison K. Implantable cardioverter-defibrillators with end stage renal disease: nationwide inpatient sample database results. Pacing Clin Electrophysiol. (2022) 45(1):124–31. doi: 10.1111/pace.14411

PubMed Abstract | Crossref Full Text | Google Scholar

26. Tonelli M, Wiebe N, Culleton B, House A, Rabbat C, Fok M, et al. Chronic kidney disease and mortality risk: a systematic review. J Am Soc Nephrol. (2006) 17(7):2034–47. doi: 10.1681/ASN.2005101085

PubMed Abstract | Crossref Full Text | Google Scholar

27. Whitman IR, Feldman HI, Deo R. CKD and sudden cardiac death: epidemiology, mechanisms, and therapeutic approaches. J Am Soc Nephrol. (2012) 23(12):1929–39. doi: 10.1681/ASN.2012010037

PubMed Abstract | Crossref Full Text | Google Scholar

28. Pickup LC, Law JP, Townend JN, Ferro CJ. Sudden cardiac death in chronic renal disease: aetiology and risk reduction strategies. Nephrol Dial Transplant. (2021) 36(8):1386–8. doi: 10.1093/ndt/gfz232

PubMed Abstract | Crossref Full Text | Google Scholar

29. Russo V, Rago A, Ruggiero V, Cavaliere F, Bianchi V, Ammendola E, et al. Device-related complications and inappropriate therapies among subcutaneous vs. transvenous implantable defibrillator recipients: insight monaldi rhythm registry. Front Cardiovasc Med. (2022) 9:879918. doi: 10.3389/fcvm.2022.879918

PubMed Abstract | Crossref Full Text | Google Scholar

30. Olsen T, Jørgensen OD, Nielsen JC, Thøgersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete danish device-cohort (1982–2018). Eur Heart J. (2019) 40(23):1862–9. doi: 10.1093/eurheartj/ehz316

PubMed Abstract | Crossref Full Text | Google Scholar

31. Lekkerkerker JC, Van Nieuwkoop C, Trines SA, Van Der Bom JG, Bernards A, Van De Velde ET, et al. Risk factors and time delay associated with cardiac device infections: leiden device registry. Heart. (2009) 95(9):715–20. doi: 10.1136/hrt.2008.151985

PubMed Abstract | Crossref Full Text | Google Scholar

32. Lee JZ, Majmundar M, Kumar A, Thakkar S, Patel HP, Sorajja D, et al. Impact of timing of transvenous lead removal on outcomes in infected cardiac implantable electronic devices. Heart Rhythm. (2022) 19(5):768–75. doi: 10.1016/j.hrthm.2021.12.023

PubMed Abstract | Crossref Full Text | Google Scholar

33. El-Chami MF, Burke MC, Herre JM, Shah MH, Sadhu A, Niebauer MJ, et al. Outcomes of subcutaneous implantable cardioverter-defibrillator in dialysis patients: results from the S-ICD post-approval study. Heart Rhythm. (2020) 17(9):1566–74. doi: 10.1016/j.hrthm.2020.04.036

PubMed Abstract | Crossref Full Text | Google Scholar

34. Kloppe A, Winter J, Prull M, Aweimer A, El-Battrawy I, Hanefeld C, et al. Subcutaneous cardioverter defibrillator implanted intermuscularly in patients with end-stage renal disease requiring hemodialysis: 5-year follow-up. J Interv Card Electrophysiol. (2024). doi: 10.1007/s10840-024-01767-1. [Epub ahead of print].38383674

PubMed Abstract | Crossref Full Text | Google Scholar

35. De Filippo P, Migliore F, Palmisano P, Nigro G, Ziacchi M, Rordorf R, et al. Procedure, management, and outcome of subcutaneous implantable cardioverter–defibrillator extraction in clinical practice. Europace. (2023) 25(6):euad158. doi: 10.1093/europace/euad158

PubMed Abstract | Crossref Full Text | Google Scholar

36. Jukema JW, Timal RJ, Rotmans JI, Hensen LCR, Buiten MS, De Bie MK, et al. Prophylactic use of implantable cardioverter-defibrillators in the prevention of sudden cardiac death in dialysis patients: the prospective, randomized, controlled ICD2 trial. Circulation. (2019) 139(23):2628–38. doi: 10.1161/CIRCULATIONAHA.119.039818

PubMed Abstract | Crossref Full Text | Google Scholar

37. Hawkins NM, Grubisic M, Andrade JG, Huang F, Ding L, Gao M, et al. Long-term complications, reoperations and survival following cardioverter-defibrillator implant. Heart. (2018) 104(3):237–43. doi: 10.1136/heartjnl-2017-311638

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: sudden cardiac death (SCD), implantable cardiac defibrillator (ICD), S-ICD, device infection, device complication, chronic kidney disease

Citation: Schiedat F, Meuterodt B, Prull M, Aweimer A, Gotzmann M, O’Connor S, Perings C, Korth J, Lawo T, El-Battrawy I, Hanefeld C, Mügge A and Kloppe A (2024) Comparison of infection and complication rates associated with transvenous vs. subcutaneous defibrillators in patients with stage 4 chronic kidney disease: a multicenter long-term retrospective follow-up. Front. Cardiovasc. Med. 11:1397138. doi: 10.3389/fcvm.2024.1397138

Received: 6 March 2024; Accepted: 27 March 2024;
Published: 10 April 2024.

Edited by:

Mate Vamos, University of Szeged, Hungary

Reviewed by:

Roberto Rordorf, San Matteo Hospital Foundation (IRCCS), Italy
Helmut Ulrich Klein, University of Rochester, United States

© 2024 Schiedat, Meuterodt, Prull, Aweimer, Gotzmann, O’Connor, Perings, Korth, Lawo, El-Battrawy, Hanefeld, Mügge and Kloppe. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Fabian Schiedat fschiedat@aol.com

Abbreviations BMI, body mass index; CIED, cardiac implantable electronic device; CKD, chronic kidney disease; CRP, C-reactive proteins; CRT, cardiac resynchronization therapy; GFR, glomerular filtration rate; HD, hemodialysis; HF, heart failure; ICD, implantable cardioverter defibrillator; LV-EF, left ventricular ejection fraction; NYHA, New York Heart Association; SCD, sudden cardiac death; S-ICD, subcutaneous implantable cardioverter defibrillator; TLE, transvenous lead extraction; TV-ICD, transvenous implantable cardioverter defibrillator.

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.