Influencing factors and improvement methods of coronary artery plaque evaluation in CT

\r\nYaqi GaoYaqi GaoYao PanYao PanChongfu Jia
\r\nChongfu Jia*
  • Department of Cardiovascular Radiology, The First Affiliated Hospital of Dalian Medical University, Dalian, China

Accurate evaluation of the nature and composition of coronary plaque involves clinical follow-up and prognosis. Coronary CT angiography is the most commonly non-invasive method for plaque evaluation, however, the qualitative and quantitative evaluation of plaque based on CT value is inaccurate, due to the influence of luminal attenuation, tube voltage, parameter setting and the subjectivity.

1 Introduction

Coronary heart disease (CHD) incidence has been on the rise year by year, with its mortality rate ranking first among cardiovascular diseases (1). The primary cause of cardiovascular events is the rupture of unstable coronary artery plaques followed by thrombus formation. Numerous postmortem and intravascular ultrasound studies have shown a close correlation between plaque nature and composition with stability, particularly emphasizing the heightened activity and instability of non-calcified plaques, especially those with high lipid content (2, 3). Moreover, the assessment of plaque burden and composition has a good predictive value for cardiovascular events, and can further evaluate clinical risk factors, reflect drug treatment and adverse clinical outcomes (46). Therefore, accurately assessing plaque nature and composition is crucial for risk stratification and prognosis evaluation in patients with CHD.

Intravascular imaging techniques such as optical coherence tomography (OCT) are considered the gold standard for assessing plaque nature and composition in vivo. However, their invasive, complexity of operation, and cost make them challenging for widespread clinical adoption (7). In recent years, coronary computed tomographic angiography (CCTA) has become a routine clinical tool for non-invasive evaluation of plaques (8). Nevertheless, challenges persist in accurately assessing coronary artery plaque nature and composition based on CT values due to factors such as luminal attenuation, tube voltage, parameter settings, and subjectivity, leading to significant overlap in CT values between fibrous and lipid-rich plaques (9). In this review, we summarize the methods and influencing factors of plaque assessment for CCTA, and provide the corresponding improvement methods.

1.1 Qualitative and quantitative assessment of plaques in CCTA

The qualitative assessment of plaques involves manually selecting regions of interest (ROIs) and determining plaque characteristics based on the measured CT values, as shown in Figure 1. However, unified standard for ROI selection are not clearly defined, including criteria such as the selection of representative slices, distance from the lumen, and ROI size. It has been reported that efforts should be made to avoid the lumen and vessel walls while encompassing the plaque as much as possible. Multiple ROIs are then measured, and their average is calculated (10). However, specific guidelines on how much to avoid are not clearly defined. Currently, to use a generalized HU-criterion is not yet possible as the reported HU values vary considerably (3, 1116), as shown in Table 1. Matsumoto et al. (3) conducted CCTA examinations on 77 patients with known or suspected coronary artery disease, using intravascular ultrasound (IVUS) as the gold standard. They found that using 45 HU as the cutoff value resulted in higher diagnostic accuracy for lipid plaques. In another study, Han et al. (12) identified 75 HU as the optimal cutoff value for accurate identification of lipid plaques, showing good consistency with histological regions rich in lipid plaques. This variability in cutoff values poses challenges for widespread clinical application.

Figure 1
www.frontiersin.org

Figure 1 Plaque recognition and CT value measurement in previous literature (10). (A) Non-calcified plaque was detected on proximal segment of LAD on curved reformation. (B) The measurement of qualitative.

Table 1
www.frontiersin.org

Table 1 Comparison of attenuation values used in selected studies.

In contrast to the qualitative assessment of plaque characteristics, the quantitative measurement of plaque components involves semi-automatically delineating the entire plaque using quantitative analysis software, as shown in Figure 2. Component volumes are obtained based on pre-set thresholds. Although this eliminates the subjectivity of manually selecting ROIs, there is still inconsistency in the definition of component thresholds. Most studies define −30 to 30 HU, 30–130 HU, 131–350 HU, and >350 HU as necrotic core, fibrofatty, fibrous, and calcified components, respectively (17). Some studies define −30 to 75 HU and 76–130 HU as lipid and fibrofatty components, respectively (18). Takx et al. (19) define fibrous plaque as 70–129 HU, not fibrofatty. Therefore, the cutoff values for distinguishing lipid and fibrous components remain a contentious issue. Furthermore, most studies define 131–350 HU as fibrous components, which significantly differs from the qualitative cutoff value for fibrous plaques. This definition contradicts even the definition in calcium scoring, where >130 HU is considered calcified plaque. Takahashi et al. (16) demonstrated that the fibrous component load measured by CCTA was poorly correlated with the gold standard virtual IVUS, with a correlation coefficient of only 0.18. This discrepancy may be attributed to the influence of partial volume effects in the vessel lumen, leading to the erroneous classification of 131–350 HU as fibrous plaques, which may not necessarily represent true fibrous tissue.

Figure 2
www.frontiersin.org

Figure 2 (A) Multiplanar reconstruction of the left anterior descending coronary artery (LAD) showing the whole coronary artery lesion, ranging from the proximal to distal lesion marker (green markers). Fibrotic tissue was labeled in dark yellow, Fibro-fatty tissue in light green and necrotic core was labeled in blue. (B) A histogram of the distribution of CT attenuation within the plaque. (C) Plaque volumes were derived for the quantitative analysis software based on pre-set thresholds.

1.2 Factors influencing plaque evaluation

There are numerous factors affecting the assessment of coronary artery plaques based on CT, primarily including partial volume effect of lumen, imaging parameters, and subjective factors.

Firstly, luminal attenuation is the most significant influencing factor. This is due to the limited spatial resolution of CT equipment. Studies indicate a significant positive correlation between lipid and fibrous plaques and luminal HU, with correlation coefficients of 0.49 and 0.32, respectively. In contrast, calcified plaques show no correlation with luminal HU (r = 0.04) (11). Takagi et al. (20) further demonstrated that as luminal attenuation increases, the proportion of calcified components also increases, while the proportions of fibrous plaques, fibrous lipid, and necrotic core decrease. Kristanto et al. (21) using a phantom model, investigated the impact of luminal attenuation on plaques and found that within a 2-pixel radius outside the lumen, the influence followed an exponential decay pattern: y = Aeλx + c (where A, λ, x, and c represent amplitude, coefficient, distance from the luminal boundary, and constant, respectively). They recommended selecting ROIs at a distance of at least 0.8 mm from the lumen to minimize the partial volume effect, providing insights for accurate plaque measurement. However, this approach has limitations in reflecting the overall plaque nature and volume of individual components, especially for small-volume plaques.

Secondly, low tube voltage is a primary means of reducing radiation dose in CCTA. However, due to the weak penetration ability of low-energy x-rays, it significantly influences plaque assessment based on CT values (22). Wang et al. (15) compared the effects of conventional and low tube voltage on the HU of non-calcified plaques using a phantom model. The results showed that the HU of low tube voltage plaques were significantly higher than those at conventional voltage. Different tube voltages require different cutoff HU for distinguishing lipid and fibrous components. For example, the volumes of both components measured at 80 kV (82 HU) and 120 kV (60 HU) showed good correlation (r = 0.841, r = 0.879). The study indicated two pathways through which tube voltage influences plaques: the direct pathway driven by x-ray energy and the indirect pathway by altering luminal attenuation, reducing tube voltage would decrease the Compton effect, raising the HU of the iodine-containing contrast agent (23). Matsumoto et al. (11) compared the effects of 100 kV and 120 kV on plaque components, finding that lower tube voltage led to a significant increase in the HU of fibrous and lipid components, by approximately 2.2 HU and 5.3 HU, respectively. However, after correcting for the indirect effect of luminal attenuation, 100 kV had almost no direct impact on non-calcified plaques. Takagi et al. (20) found that compared to 120 kV, 100 kV increased calcium volume (2.4%) but decreased fibrofatty (4.8%) and necrotic core (1.5%) volumes. This change was primarily induced by the indirect pathway. Besides the indirect pathway, the authors further discovered that even lower tube voltage (80 kV) directly affected fibrofatty and necrotic core components, increasing their volumes by 6.2% and 0.9%, respectively. However, both 100 kV and 80 kV had a direct impact on calcified components, possibly due to the higher likelihood of photoelectric interaction in dense tissue at lower x-ray energies, with the maximum variation occurring in high-density calcium.

Apart from luminal attenuation and tube voltage, the accuracy of plaque evaluation is also influenced by scanning parameters, post-processing methods, and subjectivity. Different CT machine models may result in different cutoff HU for plaque assessment (24). The HU of plaques decrease with an increase in detector rows and reach their highest point at maximum collimation width, possibly due to higher spatial resolution reducing partial volume effect. Additionally, using different reconstruction parameters, such as slice thickness and kernel, affects plaque HU. Cademartiri et al. (25) comparing four different convolution kernels (b30f, b36f, b46f, and b60f), found that using sharper kernels increased spatial resolution and image noise, raising the HU of calcified plaques (740.5, 758.5, 785.7, and 1,145.8 HU) and decreasing non-calcified plaque HU (20.8, 14.2, 14.0, and 3.2 HU). Thicker slice thickness also decreases spatial resolution, increasing plaque HU (26). Moreover, human interpretation is prone to fatigue and subjectivity. In many CTA studies, plaque CT values may be influenced by the selection, position, and size of ROIs (27).

1.3 Improvement methods

In addressing the aforementioned influencing factors, multiple researchers emphasize that achieving similar luminal attenuation by unifying imaging parameters may be the most ideal solution, especially in the context of serial CCTA examinations within the same patient cohort (11, 20). However, the complete standardization of parameters to eliminate variations posed by CT machine models, reconstruction methods, physician experience, and individual patient differences is challenging. In light of this, various studies have proposed adaptive approaches, dual-energy techniques, dose adjustments of contrast agents, and artificial intelligence (AI) methods as potential solutions (3, 11, 2838).

Firstly, to mitigate the impact on luminal attenuation, de Knegt et al. (28) proposed an adaptive method by adjusting plaque assessment thresholds based on luminal CT values. Results indicated that adaptive thresholds more accurately assessed plaques, whereas conventional CT values underestimated the volume of fibrous and fibrofatty plaques and overestimated the volume of necrotic cores and calcifications. Another study utilized the average luminal CT values of plaques in the proximal and distal regions for calibration, revealing an overlap in the conventional CT values distributions of lipid and fibrous plaques (−4.5 to 74.6 vs. 61.9–149.7). However, standardized CT values (plaque/luminal) ratios eliminated this overlap (−0.012 to 0.147 vs. 0.153–0.394) (11). Nonetheless, luminal CT values gradually decrease with increasing stenosis severity and decreasing diameter, leading to a significant reduction in luminal CT values in severely stenosed regions (29). Therefore, even with adjustments for scan-to-scan variations in luminal attenuation, applying the same thresholds in stenotic and non-stenotic regions may still be inadequate. Shin et al. (30) proposed two novel adaptive thresholds: scan-adaptive threshold, using the same threshold at each cross-section, and position-adaptive threshold, applying different CT thresholds at each cross-section (i.e., <50% * corresponding cross-sectional luminal CT value for non-calcified plaques, >110% * corresponding luminal CT value for calcified plaques). The results indicated that scan-adaptive thresholds overestimated plaque volume, while position-adaptive thresholds exhibited higher accuracy.

In addition, dual-energy CT (DECT) can optimize image quality based on characteristic energy levels, distinguishing tissue types by absorbing high and low-energy x-rays. DECT provides additional parameters such as iodine maps, effective atomic number maps, and virtual monoenergetic images, yielding more diagnostic information. Recent research suggests that non-calcified plaques can be accurately identified using effective atomic number maps (32). Although multi-energy CT has advantages in plaque characterization, it faces limitations related to the complexity of plaque composition and the small size of coronary artery plaques, posing challenges for quantitative analysis due to partial volume effects (39).

Secondly, addressing the impact of tube voltage, adjusting the corresponding contrast agent dose can alleviate changes inluminal attenuation and improve the qualitative and quantitative assessment of plaques. Yin et al. (33) found that CCTA images and luminal attenuation obtained with 100 kV combined with lower concentration iodine contrast (270 mg iodine/ml) were similar to those obtained with 120 kV combined with higher concentration contrast (370 mg iodine/ml). Jia et al. (34) discovered that images obtained with 70 kVp combined with 30 ml of lower concentration contrast (300 mg iodine/ml) were comparable in quality to those obtained with 100/120 kVp combined with 65–75 ml of the same concentration contrast. These studies suggest that low-dose scans based on low tube voltage can reduce radiation dose while avoiding excessive luminal attenuation. However, this approach is associated with higher image noise and lower contrast-to-noise ratio, impacting accurate plaque quantification and characterization. Iterative reconstruction (IR) combined with low tube voltage scans is beneficial for maintaining diagnostic image quality while reducing radiation dose. Studies indicate that deep learning-based image reconstruction algorithms can reduce radiation dose by over 40% and improve image quality by 62%, without affecting stenosis severity, plaque composition, and quantitative assessment (35). However, there is limited literature on directly addressing the impact of low tube voltage, necessitating further research.

To mitigate subjective influences, studies have found that histogram analysis and AI-based algorithms can overcome the subjectivity of ROI selection by analyzing the entire plaque. Matsumoto et al. (3) utilized histogram analysis with IVUS as the gold standard, revealing that the area under the curve for diagnosing lipid-rich plaques based on the proportion of CT values ≤30 HU was significantly higher (0.9) than that for conventional CT values (0.3), with sensitivity and specificity reaching 95% and 80%, respectively. While histogram analysis eliminates the subjective impact of ROI selection, it does not account for objective influences such as luminal attenuation, resulting in varying threshold values across different studies (36). Additionally, AI-based processing algorithms can achieve image reconstruction, segmentation, measurement, and feature extraction with high accuracy and low reader variability, potentially reducing subjective influences while mitigating objective influences such as luminal attenuation (36, 37). Kolossváry et al. (38) processed CT images of 44 coronary artery plaques, extracting radiomic features (935 in total). The results indicated that, using NaF 18-positron emission tomography, intravascular ultrasound, and optical coherence tomography as gold standards, the diagnostic value of radiomic features in identifying vulnerable plaques was significantly superior to conventional CT parameters, with increased areas under the curve of 0.22, 0.13, and 0.14, respectively. However, whether radiomic features can classify plaque components and their capabilities remain unreported and warrant further investigation.

2 Conclusion

While coronary computed tomography angiography (CCTA) has become a common non-invasive imaging method for evaluating plaques, it can not only reliably evaluate luminal stenosis and its functional significance, but also accurately evaluate the morphology and composition of plaque and identify high-risk plaque, it is of great significance to guide the clinical management of patients with coronary heart disease, accurate evaluation of plaques based on CCTA still faces numerous challenges due to factors such as luminal attenuation, parameter settings, and subjectivity. The use of standardized scanning methods, self-adaption, dual-energy techniques, and artificial intelligence (AI) holds promise in addressing these challenges. Particularly, AI has demonstrated significant advantages in enhancing image quality and deeply exploring plaque characteristics, showing considerable potential for clinical applications. However, further research is needed to address the direct impact of improving low tube voltage.

Author contributions

YG: Writing – original draft. YP: Writing – original draft. CJ: Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Yamagishi M, Tamaki N, Akasaka T, Ikeda T, Ueshima K, Uemura S, et al. JCS 2018 guideline on diagnosis of chronic coronary heart diseases. Circ J. (2021) 85(4):402–572. doi: 10.1253/circj.CJ-19-1131

PubMed Abstract | Crossref Full Text | Google Scholar

2. Osborn EA, Houstis NE, Jaffer FA. Synergizing light and machine learning to comprehensively reveal coronary plaque composition. JACC Basic Transl Sci. (2021) 6(12):961–3. doi: 10.1016/j.jacbts.2021.10.016

PubMed Abstract | Crossref Full Text | Google Scholar

3. Matsumoto H, Watanabe S, Kyo E, Tsuji T, Ando Y, Otaki Y, et al. Standardized volumetric plaque quantification and characterization from coronary CT angiography: a head-to-head comparison with invasive intravascular ultrasound. Eur Radiol. (2019) 29(11):6129–39. doi: 10.1007/s00330-019-06219-3

PubMed Abstract | Crossref Full Text | Google Scholar

4. Puchner SB, Liu T, Mayrhofer T, Truong QA, Lee H, Fleg JL, et al. High-risk plaque detected on coronary CT angiography predicts acute coronary syndromes independent of significant stenosis in acute chest pain: results from the ROMICAT-II trial. J Am Coll Cardiol. (2014) 64(7):684–92. doi: 10.1016/j.jacc.2014.05.039

PubMed Abstract | Crossref Full Text | Google Scholar

5. Williams MC, Kwiecinski J, Doris M, McElhinney P, D’Souza MS, Cadet S, et al. Low-attenuation noncalcified plaque on coronary computed tomography angiography predicts myocardial infarction: results from the multicenter SCOT-HEART trial (Scottish computed tomography of the HEART). Circulation. (2020) 141(18):1452–62. doi: 10.1161/CIRCULATIONAHA.119.044720

PubMed Abstract | Crossref Full Text | Google Scholar

6. Yamaura H, Otsuka K, Ishikawa H, Shirasawa K, Fukuda D, Kasayuki N. Determinants of non-calcified low-attenuation coronary plaque burden in patients without known coronary artery disease: a coronary CT angiography study. Front Cardiovasc Med. (2022) 9:824470. doi: 10.3389/fcvm.2022.824470

PubMed Abstract | Crossref Full Text | Google Scholar

7. Theofilis P, Sagris M, Antonopoulos AS, Oikonomou E, Tsioufis K, Tousoulis D. Non-invasive modalities in the assessment of vulnerable coronary atherosclerotic plaques. Tomography. (2022) 8(4):1742–58. doi: 10.3390/tomography8040147

PubMed Abstract | Crossref Full Text | Google Scholar

8. Antonopoulos AS, Angelopoulos A, Tsioufis K, Antoniades C, Tousoulis D. Cardiovascular risk stratification by coronary computed tomography angiography imaging: current state-of-the-art. Eur J Prev Cardiol. (2022) 29(4):608–24. doi: 10.1093/eurjpc/zwab067

PubMed Abstract | Crossref Full Text | Google Scholar

9. Shaw LJ, Blankstein R, Bax JJ, Ferencik M, Bittencourt MS, Min JK, et al. Society of cardiovascular computed tomography/north American society of cardiovascular imaging expert consensus document on coronary CT imaging of atherosclerotic plaque. J Cardiovasc Comput Tomogr. (2021) 15(2):93–109. doi: 10.1016/j.jcct.2020.11.002

PubMed Abstract | Crossref Full Text | Google Scholar

10. Li T, Tang T, Yang L, Zhang X, Li X, Luo C. Coronary CT angiography with knowledge-based iterative model reconstruction for assessing coronary arteries and non-calcified predominant plaques. Korean J Radiol. (2019) 20(5):729–38. doi: 10.3348/kjr.2018.0435

PubMed Abstract | Crossref Full Text | Google Scholar

11. Matsumoto H, Watanabe S, Kyo E, Tsuji T, Ando Y, Otaki Y, et al. Effect of tube potential and luminal contrast attenuation on atherosclerotic plaque attenuation by coronary CT angiography: in vivo comparison with intravascular ultrasound. J Cardiovasc Comput Tomogr. (2019) 13(4):219–25. doi: 10.1016/j.jcct.2019.02.004

PubMed Abstract | Crossref Full Text | Google Scholar

12. Han D, Torii S, Yahagi K, Lin FY, Lee JH, Rizvi A, et al. Quantitative measurement of lipid rich plaque by coronary computed tomography angiography: a correlation of histology in sudden cardiac death. Atherosclerosis. (2018) 275:426–33. doi: 10.1016/j.atherosclerosis.2018.05.024

PubMed Abstract | Crossref Full Text | Google Scholar

13. Yamaki T, Kawasaki M, Jang IK, Raffel OC, Ishihara Y, Okubo M, et al. Comparison between integrated backscatter intravascular ultrasound and 64-slice multi-detector row computed tomography for tissue characterization and volumetric assessment of coronary plaques. Cardiovasc Ultrasound. (2012) 10:33. doi: 10.1186/1476-7120-10-33

PubMed Abstract | Crossref Full Text | Google Scholar

14. You S, Sun JS, Park SY, Baek Y, Kang DK. Relationship between indexed epicardial fat volume and coronary plaque volume assessed by cardiac multidetector CT. Medicine (Baltimore). (2016) 95:e4164. doi: 10.1097/MD.0000000000004164

PubMed Abstract | Crossref Full Text | Google Scholar

15. Wang C, Liao Y, Chen H, Zhen X, Li J, Xu Y, et al. Influence of tube potential on quantitative coronary plaque analyses by low radiation dose computed tomography: a phantom study. Int J Cardiovasc Imaging. (2018) 34(8):1315–22. doi: 10.1007/s10554-018-1344-y

PubMed Abstract | Crossref Full Text | Google Scholar

16. Takahashi S, Kawasaki M, Miyata S, Suzuki K, Yamaura M, Ido T, et al. Feasibility of tissue characterization of coronary plaques using 320-detector row computed tomography: comparison with integrated backscatter intravascular ultrasound. Heart Vessels. (2016) 31(1):29–37. doi: 10.1007/s00380-014-0577-5

PubMed Abstract | Crossref Full Text | Google Scholar

17. Bax AM, Lin FY, van Rosendael AR, Ma X, Lu Y, van den Hoogen IJ, et al. Marked variation in atherosclerotic plaque progression between the major epicardial coronary arteries. Eur Heart J Cardiovasc Imaging. (2022) 23(11):1482–91. doi: 10.1093/ehjci/jeac044

PubMed Abstract | Crossref Full Text | Google Scholar

18. Kuneman JH, van Rosendael SE, van der Bijl P, van Rosendael AR, Kitslaar PH, Reiber JHC, et al. Pericoronary adipose tissue attenuation in patients with acute coronary syndrome versus stable coronary artery disease. Circ Cardiovasc Imaging. (2023) 16(2):e014672. doi: 10.1161/CIRCIMAGING.122.014672

PubMed Abstract | Crossref Full Text | Google Scholar

19. Takx RA, Willemink MJ, Nathoe HM, Schilham AM, Budde RP, de Jong PA, et al. The effect of iterative reconstruction on quantitative computed tomography assessment of coronary plaque composition. Int J Cardiovas Imaging. (2014) 30:155–63. doi: 10.1007/s10554-013-0293-8

Crossref Full Text | Google Scholar

20. Takagi H, Leipsic JA, Indraratna P, Gulsin G, Khasanova E, Tzimas G, et al. Association of tube voltage with plaque composition on coronary CT angiography: results from PARADIGM registry. JACC Cardiovasc Imaging. (2021) 14(12):2429–40. doi: 10.1016/j.jcmg.2021.07.011

PubMed Abstract | Crossref Full Text | Google Scholar

21. Kristanto W, Tuncay V, Vliegenthart R, van Ooijen PM, Oudkerk M. Correction of lumen contrast-enhancement influence on non-calcified coronary atherosclerotic plaque quantification on CT. Int J Cardiovasc Imaging. (2015) 31(2):429–36. doi: 10.1007/s10554-014-0554-1

PubMed Abstract | Crossref Full Text | Google Scholar

22. Stocker TJ, Leipsic J, Hadamitzky M, Chen MY, Rubinshtein R, Deseive S, et al. Application of low tube potentials in CCTA: results from the PROTECTION VI study. JACC Cardiovasc Imaging. (2020) 13(2 Pt 1):425–34. doi: 10.1016/j.jcmg.2019.03.030

PubMed Abstract | Crossref Full Text | Google Scholar

23. McCollough CH, Leng S, Yu L, Fletcher JG. Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology. (2015) 276(3):637–53. doi: 10.1148/radiol.2015142631

PubMed Abstract | Crossref Full Text | Google Scholar

24. Saremi F, Achenbach S. Coronary plaque characterization using CT. AJR Am J Roentgenol. (2015) 204(3):W249–60. doi: 10.2214/AJR.14.13760

PubMed Abstract | Crossref Full Text | Google Scholar

25. Cademartiri F, La Grutta L, Runza G, Palumbo A, Maffei E, Mollet NR, et al. Influence of convolution filtering on coronary plaque attenuation values: observations in an ex vivo model of multislice computed tomography coronary angiography. Eur Radiol. (2007) 17(7):1842–9. doi: 10.1007/s00330-006-0548-z

PubMed Abstract | Crossref Full Text | Google Scholar

26. Achenbach S, Boehmer K, Pflederer T, Ropers D, Seltmann M, Lell M, et al. Influence of slice thickness and reconstruction kernel on the computed tomographic attenuation of coronary atherosclerotic plaque. J Cardiovasc Comput Tomogr. (2010) 4(2):110–5. doi: 10.1016/j.jcct.2010.01.013

PubMed Abstract | Crossref Full Text | Google Scholar

27. Kigka VI, Sakellarios A, Kyriakidis S, Rigas G, Athanasiou L, Siogkas P, et al. A three-dimensional quantification of calcified and non-calcified plaques in coronary arteries based on computed tomography coronary angiography images: comparison with expert’s annotations and virtual histology intravascular ultrasound. Comput Biol Med. (2019) 113:103409. doi: 10.1016/j.compbiomed.2019.103409

PubMed Abstract | Crossref Full Text | Google Scholar

28. de Knegt MC, Haugen M, Jensen AK, Linde JJ, Kühl JT, Hove JD, et al. Coronary plaque composition assessed by cardiac computed tomography using adaptive hounsfield unit thresholds. Clin Imaging. (2019) 57:7–14. doi: 10.1016/j.clinimag.2019.04.014

PubMed Abstract | Crossref Full Text | Google Scholar

29. Choi JH, Min JK, Labounty TM, Lin FY, Mendoza DD, Shin DH, et al. Intracoronary transluminal attenuation gradient in coronary CT angiography for determining coronary artery stenosis. JACC Cardiovasc Imaging. (2011) 4(11):1149–57. doi: 10.1016/j.jcmg.2011.09.006

PubMed Abstract | Crossref Full Text | Google Scholar

30. Shin CI, Park SJ, Kim JH, Yoon YE, Park EA, Koo BK, et al. Coronary artery lumen segmentation using location-adaptive threshold in coronary computed tomographic angiography: a proof-of-concept. Korean J Radiol. (2021) 22(5):688–96. doi: 10.3348/kjr.2020.0296

PubMed Abstract | Crossref Full Text | Google Scholar

31. Tatsugami F, Higaki T, Nakamura Y, Honda Y, Awai K. Dual-energy CT: minimal essentials for radiologists. Jpn J Radiol. (2022) 40(6):547–59. doi: 10.1007/s11604-021-01233-2

PubMed Abstract | Crossref Full Text | Google Scholar

32. Nakajima S, Ito H, Mitsuhashi T, Kubo Y, Matsui K, Tanaka I, et al. Clinical application of effective atomic number for classifying non-calcified coronary plaques by dual-energy computed tomography. Atherosclerosis. (2017) 261:138–43. doi: 10.1016/j.atherosclerosis.2017.03.025

PubMed Abstract | Crossref Full Text | Google Scholar

33. Yin WH, Lu B, Gao JB, Li PL, Sun K, Wu ZF, et al. Effect of reduced x-ray tube voltage, low iodine concentration contrast medium, and sinogram-affirmed iterative reconstruction on image quality and radiation dose at coronary CT angiography: results of the prospective multicenter REALISE trial. J Cardiovasc Comput Tomogr. (2015) 9(3):215–24. doi: 10.1016/j.jcct.2015.01.010

PubMed Abstract | Crossref Full Text | Google Scholar

34. Jia CF, Zhong J, Meng XY, Sun XX, Yang ZQ, Zou YJ, et al. Image quality and diagnostic value of ultra low-voltage, ultra low-contrast coronary CT angiography. Eur Radiol. (2019) 29(7):3678–85. doi: 10.1007/s00330-019-06111-0

PubMed Abstract | Crossref Full Text | Google Scholar

35. Benz DC, Ersözlü S, Mojon FLA, Messerli M, Mitulla AK, Ciancone D, et al. Radiation dose reduction with deep-learning image reconstruction for coronary computed tomography angiography. Eur Radiol. (2022) 32(4):2620–8. doi: 10.1007/s00330-021-08367-x

PubMed Abstract | Crossref Full Text | Google Scholar

36. Yin WH, Zhang Y, Li XN, Wang HY, An YQ, Sun Y, et al. In vivo detection of lipid-core plaques by coronary CT angiography: a head-to-head comparison with histologic findings. Korean J Radiol. (2020) 21(2):210–7. doi: 10.3348/kjr.2019.0557

PubMed Abstract | Crossref Full Text | Google Scholar

37. Sheahan M, Ma X, Paik D, Obuchowski NA, St Pierre S, Newman WP 3rd, et al. Atherosclerotic plaque tissue: noninvasive quantitative assessment of characteristics with software-aided measurements from conventional CT angiography. Radiology. (2018) 286:622–31. doi: 10.1148/radiol.2017170127

PubMed Abstract | Crossref Full Text | Google Scholar

38. Kolossváry M, Park J, Bang JI, Zhang J, Lee JM, Paeng JC, et al. Identification of invasive and radionuclide imaging markers of coronary plaque vulnerability using radiomic analysis of coronary computed tomography angiography. Eur Heart J Cardiovasc Imaging. (2019) 20(11):1250–8. doi: 10.1093/ehjci/jez033

Crossref Full Text | Google Scholar

39. Ding H, Wang C, Malkasian S, Johnson T, Molloi S. Characterization of arterial plaque composition with dual energy computed tomography: a simulation study. Int J Cardiovasc Imaging. (2021) 37(1):331–41. doi: 10.1007/s10554-020-01961-y

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: coronary, plaque, computed tomography, angiography, evaluation

Citation: Gao Y, Pan Y and Jia C (2024) Influencing factors and improvement methods of coronary artery plaque evaluation in CT. Front. Cardiovasc. Med. 11:1395350. doi: 10.3389/fcvm.2024.1395350

Received: 3 March 2024; Accepted: 14 May 2024;
Published: 25 June 2024.

Edited by:

Sebastian Kelle, German Heart Center Berlin, Germany

Reviewed by:

Chiara Martini, University Hospital of Parma, Italy
Kenichiro Otsuka, Osaka City University, Japan

© 2024 Gao, Pan and Jia. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Chongfu Jia, chongfujia@sina.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.