AUTHOR=Pastapur Aishwarya , Pescatore Nicole A. , Shah Nirav , Kheterpal Sachin , Nallamothu Brahmajee K. , Golbus Jessica R. TITLE=Evaluation of atrial fibrillation using wearable device signals and home blood pressure data in the Michigan Predictive Activity & Clinical Trajectories in Health (MIPACT) Study: A Subgroup Analysis (MIPACT-AFib) JOURNAL=Frontiers in Cardiovascular Medicine VOLUME=10 YEAR=2023 URL=https://www.frontiersin.org/journals/cardiovascular-medicine/articles/10.3389/fcvm.2023.1243574 DOI=10.3389/fcvm.2023.1243574 ISSN=2297-055X ABSTRACT=Background

The rising adoption of wearable technology increases the potential to identify arrhythmias. However, specificity of these notifications is poorly defined and may cause anxiety and unnecessary resource utilization. Herein, we report results of a follow-up screening protocol for incident atrial fibrillation/flutter (AF) within a large observational digital health study.

Methods

The MIPACT Study enrolled 6,765 adult patients who were provided an Apple Watch and blood pressure (BP) monitors. From March to July 2019, participants were asked to contact the study team for any irregular heart rate (HR) notification. They were assessed using structured questionnaires and asked to provide 6 Apple Watch EKGs. Those with arrhythmias or non-diagnostic EKGs were sent 7-day monitors. The EHR was reviewed after 3 years to determine if participants developed arrhythmias.

Results

86 participants received notifications and met inclusion criteria. Mean age was 50.5 (SD 16.9) years, and 46 (53.3%) were female. Of 76 participants assessed by the study team, 32 (42.1%) reported anxiety surrounding notifications. Of 59 participants who sent at least 1 EKG, 52 (88.1%) were in sinus rhythm, 3 (5.1%) AF, 2 (3.4%) indeterminate, and 2 (3.4%) sinus bradycardia. Cardiac monitor demonstrated AF in 2 of 3 participants with AF on Apple Watch EKGs. 2 contacted their PCPs and were diagnosed with AF. In total, 5 cases of AF were diagnosed with 1 additional case identified during EHR review.

Conclusion

Wearable devices produce alarms that can frequently be anxiety provoking. Research is needed to determine the implications of these alarms and appropriate follow-up.