Skip to main content

CASE REPORT article

Front. Cardiovasc. Med., 10 August 2023
Sec. Structural Interventional Cardiology
This article is part of the Research Topic Case Reports in Structural Interventional Cardiology: 2023 View all 9 articles

Case report: A complex case of valve-in-valve TAVI and left bundle branch pacing for severe aortic regurgitation with partially corrected type A aortic dissection and low ejection fraction

  • 1Department of Cardiology, University Medical Center Ljubljana, Slovenia
  • 2Department of Cardiovascular Surgery, University Medical Center Ljubljana, Slovenia
  • 3Medical Faculty, University of Ljubljana, Slovenia

Background: Aortic regurgitation is a major concern following transcatheter aortic valve implantation (TAVI), as even low-grade regurgitation is associated with increased mortality. This is of particular concern to patients with pre-existing aortic disease who are at increased risk of TAVI valve slippage. Furthermore, conduction system disturbances after TAVI, namely left bundle branch block (LBBB), may have an additional detrimental effect on cardiac function.

Case presentation: This report documents a successful treatment strategy in a frail patient with a bicuspid aortic valve and aortic disease after valve-sparing surgical repair in 1998, who subsequently developed aortic stenosis and underwent TAVI with an Evolut R self-expanding aortic valve. The progression of aortic disease, aortic root dilatation, and leaflet degeneration over the following years caused aortic regurgitation of the self-expanding aortic valve, resulting in left ventricular dilatation and heart failure along with LBBB and left ventricular (LV) mechanical dyssynchrony. Diagnostic workup of the patient showed persistence of the aneurysm distal to the graft with a dissection spanning the ascending aorta, arch, and terminating proximal to the aortic isthmus. After consideration by the cardiac team, a balloon-expandable valve was chosen for a valve-in-valve (ViV) procedure to provide sufficient radial force to expand the existing valve and correct the regurgitation. Due to the anatomy, a J-wire and pigtail catheter were successfully used for a safe approach and placement of the valve. Following the procedure, intermittent complete atrioventricular block was observed in addition to the pre-existing left bundle branch block, necessitating resynchronization pacing. Due to anatomical considerations, ease of placement, and the expected good level of resynchronization due to the proximal block, we opted for left bundle branch pacing, which showed improvement in left ventricular dyssynchrony and LV function at follow-up.

Conclusion: Valve-in-valve implantation of a balloon-expandable Myval TAVI device to treat aortic regurgitation caused by slippage and right leaflet disfunction of slef valve is feasible in challenging anatomical scenarios. Left bundle branch pacing is a viable alternative to correct mechanical dyssynchrony in complex patients with LBBB and anatomical challenges necessitating resynchronization.

Introduction

Aortic regurgitation following transcatheter aortic valve implantation (TAVI) is an important concern, as even low-grade paravalvular leak (PVL) is associated with increased mortality (1). This is especially important in patients with pre-existing aortic disease, who are at inherent risk for PVL and TAVI valve slippage due to unfavorable characteristics such as larger-diameter annuli, complex valve shapes, and progressive aortopathy (2). TAVI can also lead to conduction system disturbances, namely left bundle branch block (LBBB), which has detrimental effects on cardiac function due to impaired ventricular mechanics and cardiac remodeling (3). We outline a complex case of an elderly frail patient with partially corrected aortic disease and persistent aortic dissection, dilatative heart failure, and aortic regurgitation on a previously implanted self-expanding transcatheter heart valve, along with mechanical dyssynchrony due to left bundle branch block.

Case description

Presentation

A 79-year-old female patient was referred to our center due to severe aortic insufficiency after a previously implanted transcatheter aortic valve (TAVI). The patient's complex medical history included a bicuspid aortic valve with an aneurysm of the ascending aorta, which resulted in an acute type A aortic dissection in 1998. Following the dissection, emergency aortic valve reconstruction and a valve-sparing aortic root procedure (David procedure) were performed. In 2013, aortic stenosis and heart failure (HF) were diagnosed, both of which deteriorated during subsequent visits to the outpatient clinic. A Bentall procedure was a suboptimal therapeutic option due to the high procedural risk and the possibility of complications. TAVI was chosen as the intervention of choice in 2016, with planned oversizing due to the expected dilatation of the aortic root. An Evolut R 29 mm self-expanding valve was successfully inserted via a transfemoral approach with good expansion but low placement and minimal paravalvular leak seen after the procedure. Left bundle branch block (LBBB) developed after the first TAVI procedure. The patient was admitted again to a regional hospital shortly before referral to our center in 2021 due to worsening HF symptoms. Echocardiography confirmed a severely reduced ejection fraction (EF 20%) with concomitant severe aortic valve insufficiency resulting from a severe paravalvular leak. The patient was transferred to our center for further evaluation of possible treatment options (patient characteristics available in Supplementary S1, Timeline in Figure 1).

FIGURE 1
www.frontiersin.org

Figure 1. Timeline of care.

Examination

BMI 20.44 kg/m2, BSA 1.38 m2, afebrile, pulse rate of 75 bpm, respiratory rate of 21/min, and blood pressure of 112/65 mmHg. Blood oxygen saturation was normal without supplemental oxygen therapy. Lower extremity pretibial edema was present.

Diagnostic assessment

Laboratory tests

NT-pro BNP of 17,098 ng/L, creatinine 59 umol/L.

The ECG showed sinus rhythm with a left bundle branch block with a QRS width of 230 ms (Supplementary Figure S2).

Echocardiography revealed an enlarged left ventricle (EDD 6.4 cm, EDV 305 ml, EDVI 221 ml/m2) with mild wall hypertrophy (IVS 1.1 cm, inf-lat 1.4 cm) with severely reduced ejection fraction, a normal stroke volume (LVEF 22%, SV 70 ml), and severe valvular and paravalvular insufficiency of the Evolut R biological TAVI valve with mild mitral regurgitation. The central valvular insufficiency was caused by the degeneration of the right leaflet, which led to a diagnostic work-up and exclusion of infective endocarditis of the artificial valve—this was ruled out with transesophageal ultrasound (TEE), cardiac magnetic resonance imaging (MRI), and positron emission tomography (PET-CT) scans, in addition to serial blood cultures, which remained sterile. Echocardiographic signs of LV mechanical dyssynchrony were present. These findings were also confirmed by MRI. A CT of the thoracic and abdominal aorta revealed persistence of the ascending aortic aneurysm distal to the graft placement with dissection spanning the ascending aorta and aortic arch and terminating proximal to the aortic isthmus with involvement of the brachiocephalic trunk and progression of the aortic root dilatation compared to previous CT scans (Figure 2, Supplementary Figures S3, S4). Severe tortuosity of the iliac arteries was evident upon 3D reconstruction. Coronary angiography showed no significant atherosclerotic lesions. Infective endocarditis was excluded as the cause of the valve degeneration.

FIGURE 2
www.frontiersin.org

Figure 2. CT angiography following TAVI protocol: (A) EvolutR TAVI valve with graft repair and aortic aneurysm with type A dissection of the ascending aorta and proximal arch. Severe tortousity of the iliac arteries. (B,C) Type A aortic dissection in the transverse plane.

Procedure

Due to the frailty of the patient and the high surgical risk, the cardiac team's consensus was to correct the valve regurgitation with a valve-in-valve TAVI using a Myval 26 mm (Meril, Gujarat, India) valve due to the appropriate size, the radial force achieved by balloon-expandable valves, and the shorter stent frame height of the valve, reducing the neo-skirt. Careful procedural planning was performed in order to assess the anatomical considerations of the slipped Evolut R valve and its relationship to the aortic root. A bifemoral approach was obtained by positioning a 16F transcatheter aortic valve introducer in the right femoral artery and a 6F introducer for a pigtail catheter in the left femoral artery. Due to the severe iliac artery tortuosity, a nitinol hydrophilic guide wire covered with polyurethane Radifocus ™guidewire M (Terumo, Tokyo, Japan) with RJ 4.0 6F support was used. A distal part “S-reshaped” J soft wire-mounted pigtail catheter was used to safely cross the dissected segment (Supplementary Figure S5). The aortic valve was then crossed with a straight tip soft wire using an Amplatz left (AL) 1 catheter (Launcher, Medtronic, Minnesota, USA). The soft wire was exchanged for an extra-stiff 0.0035” Lunderquist wire (Cook Medical, Indiana, USA). We then proceeded with the implantation of a 26 mm Myval valve, which was placed without predilation of the existing Evolut R (Supplementary Figure S6). The Myval valve was positioned at the level of the natural bicuspid aortic valve ring (10/90, outflow/aortic bulbus). Control selective coronary angiography showed good patency of the coronary arteries, and control aortography showed good alignment of the prosthetic valve with no residual aortic regurgitation (Figure 3).

FIGURE 3
www.frontiersin.org

Figure 3. (A) Myval 26 mm positioning no predilation. Good angiographic result with no regurgitation. The position of the valve was at the level of the natural annulus. We avoided overextension of the neo-skirt. (B,C) Unobstructed coronary ostia with good patency.

Pacemaker implantation

Following valve-in-valve TAVI, intermittent complete AV block was recorded. As LBBB was already present after the first TAVI procedure in conjunction with severely reduced ejection fraction, resynchronization therapy was indicated. As an alternative to biventricular pacing, we opted for the LBBAP technique. A right atrial lead (Capsurefix MRI Surescan 52 cm, Medtronic) was temporarily inserted into the right ventricle (RV) to prevent potential asystole due to mechanical injury of the right bundle branch during LBBP lead positioning. As previously described (1418), a 3,830 SelectSecure 69 cm lead (Medtronic) and a C315His (Medtronic) catheter were used for transseptal lead insertion. Selective left bundle branch capture with correction of the left bundle branch block was achieved (Supplementary Figures S5, S6). The right atrial lead was then positioned in the RA, and the atrioventricular delay was optimized to achieve the shortest QRS duration (Figure 4).

FIGURE 4
www.frontiersin.org

Figure 4. (A) Postoperative x-ray of the dual-chamber pacemaker with the atrial lead positioned in the right atrial appendage and the ventricular lead positioned transeptally for left bundle branch pacing (LBBP). (B) (Left) Initial QRS complex with left bundle branch block morphology. (Right) Final QRS duration after atrio- ventricular delay optimization.

Post-procedure echocardiography

A post-procedural echocardiography did not show immediate improvement in left ventricular function with a residual dilatation of the left ventricle (EDV 333 ml, EDVI 247 ml/m2), severely reduced ejection fraction (EF 17%) with a normal stroke volume (SV 74 ml), elevated LV filling pressures, and moderate post-capillary pulmonary hypertension. The TAVI valve was well expanded with minimal trace paravalvular leakage. A Doppler ultrasound of the bifemoral puncture sites was performed before discharge due to a murmur, which did not reveal signs of fistulas or pseudoaneurysms.

Follow-up

Follow-up in the cardiology clinic revealed functional improvement in the patient with some residual exertional dyspnea but minimal signs of heart failure (NYHA class II). Follow-up echocardiography was also performed and showed improvement in LV function with an increase in ejection fraction (3D EF 27% as compared to 17% before discharge, Supplementary S7) with some minimal residual signs of mechanical dyssynchrony and normal filling pressures of the left ventricle, minimal paravalvular and central aortic regurgitation, and mild tricuspid regurgitation adjacent to the ventricular pacemaker lead with normal pulmonary pressures. Laboratory markers of HF also improved, with an NT-proBNP of 4,089 ng/L.

Discussion

Structural valve degeneration and paravalvular leaks are not uncommon after TAVI. According to the literature, redo TAVI for the treatment of acute post-procedural or late paravalvular regurgitation is associated with favorable clinical and echocardiographic outcomes (4). In our patient, the deterioration was most likely due to a combination of progression of the native aortic disease with dilatation of the aortic root that resulted in TAVI valve slippage and degeneration of one of the valve leaflets, resulting in central valve regurgitation. After ruling out infective endocarditis and after consideration by the cardiac team, a ViV TAVI procedure using a balloon-expandable device was selected due to the higher radial expansion force and the hope of also expanding the underlying stent frame of the Evolut R valve and correcting the paravalvular leak. The Myval balloon-expandable valve was chosen for its design—the small cell lower part provides higher radial strength, and the small size and low skirt reduce the chances of coronary artery occlusion and minimize neo-skirt formation. Along with this, the external PET skirt provides good sealing for any paravalvular leak (5). The case was challenging for several reasons. First, the tortuosity of the iliac arteries proved to be a challenge for peripheral access. For this reason, a Lunderquist® Extra Stiff wire was used for support; due to its linear stiffness characteristics (6), the Lunderquist® wire is often the preferred wire for large endograft delivery (7) and is used when greater support is required to deliver the TAVI device due to aortic tortuosity (8). Second, the presence of an aortic dissection spanning the entire aortic arch required a non-traumatic approach that would lower the risk of rupture. For this, a mounted pigtail catheter was used to safely cross the dissected segment using a technique used for thoracic endovascular aortic repair (TEVAR) of dissection (9). The case also highlights the strengths and advantages of using a balloon-expandable valve-in-valve TAVI. This is one of the few published attempts at a Myval valve-in-valve TAVI procedure for aortic valve insufficiency and proves that the use of a Myval valve for this kind of procedure is feasible. The valve was positioned according to valve implantation recommendations for bicuspid valve implantation. We targeted the natural annulus of the bicuspid valve in addition to the closed-cell portion of the Evolut R valve. Another important aspect of this case is the choice of pacing modality. According to data, 3.8%–20% of patients develop atrioventricular heart block, necessitating pacemaker implantation after TAVI (10). Standard RV pacing is associated with worsening HF and increased mortality (11). TAVI patients with permanent pacemakers experience more postoperative HF admissions. There was also a trend toward increased mortality; this was especially notable in patients with >40% RV pacing (12). Furthermore, standard RV pacing after TAVI was associated with negative effects on LVEF (13). Left bundle branch pacing (LBBP) promises to be a more physiological pacing modality compared to standard RV pacing for bradycardia and resynchronization indications (14, 15). In a recent study of LBB pacing following prosthetic valve implantation, the procedure proved to be feasible (16). In addition, LBBP pacing is associated with higher implant success rates and more stable pacing parameters compared to the other physiologic pacing modality, His bundle pacing (17). Although not yet validated by large randomized controlled trials, LBB pacing is proving to be a simpler alternative to other pacing modalities, including biventricular pacing (18). Our patient developed intermittent complete AV block following the valve-in-valve TAVI procedure in addition to the previously known LBBB. According to guidelines, patients with HF and LBBB require a cardiac resynchronization therapy (CRT) device with an implantable cardioverter-defibrillator (ICD). The decision to opt for LBBP instead of standard biventricular pacing was based on several factors. First, the patient was very frail and had a very low BMI, which precluded the use of devices with bulkier batteries. Second, LBBB was likely to be induced by TAVI; therefore, LBBP could be performed for resynchronization as the proximal block was expected (19), using only one lead and the smallest possible dual-chamber pacemaker. Third, correction of LBBB with transeptal lead positioning to reach the conduction system resulted in complete QRS normalization and acute improvement of mechanical dyssynchrony, with further improvement of EF during follow-up.

Patient perspectives and conclusion

This complex case report highlights the feasibility of using a Myval transcatheter prosthetic valve for redo TAVI to correct paravalvular regurgitation and valve degeneration in a patient with pre-existing aortic disease. In addition, LBBP can be used as an alternative resynchronization pacing modality following TAVI in frail patients with a low BMI and concerns about device size.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by [(infoInput0)]. The studies were conducted in accordance with the local legislation and institutional requirements. The ethics committee/institutional review board waived the requirement of written informed consent for participation from the participants or the participants’ legal guardians/next of kin because [(infoInput1)]. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

PM: writing—original draft preparation and editing. MB: Both TAVI procedures, conceptualization, editing and final review, supervision. DŽ: LBBP procedure, draft review and editing, supervision. KS: second TAVI procedure and draft editing. PH: draft preparation, LV: editing and review, TK: surgery. All authors contributed to the article and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2023.1206811/full#supplementary-material

Supplementary Video S1
Left ventricular function before ViV TAVI and LBBP.

Supplementary Video S2
Left ventricular function after ViV TAVI and LBBP.

Supplementary Video S3
Aortic regurgitation prior to ViV TAVI.

References

1. Laakso T, Laine M, Moriyama N, Dahlbacka S, Airaksinen J, Virtanen M, et al. Impact of paravalvular regurgitation on the mid-term outcome after transcatheter and surgical aortic valve replacement. Eur J Cardiothorac Surg. (2020) 58(6):1145–52. doi: 10.1093/ejcts/ezaa254

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Vincent F, Ternacle J, Denimal T, Shen M, Redfors B, Delhaye C, et al. Transcatheter aortic valve replacement in bicuspid aortic valve stenosis. Circulation. (2021) 143(10):1043–61. doi: 10.1161/CIRCULATIONAHA.120.048048

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Vernooy K, Verbeek XA, Peschar M, Crijns HJ, Arts T, Cornelussen RN, et al. Left bundle branch block induces ventricular remodelling and functional septal hypoperfusion. Eur Heart J. (2005) 26(1):91–8. doi: 10.1093/eurheartj/ehi008

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Barbanti M, Webb JG, Tamburino C, Van Mieghem NM, Makkar RR, Piazza N, et al. Outcomes of redo transcatheter aortic valve replacement for the treatment of postprocedural and late occurrence of paravalvular regurgitation and transcatheter valve failure. Circ Cardiovasc Interv. (2016) 9(9):e003930. doi: 10.1161/CIRCINTERVENTIONS.116.003930

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Sengottuvelu G, Kumar V, Seth A. The myval transcatheter heart valve system for the treatment of severe aortic stenosis—current evidence and future directions. Heart Int. (2020) 14(2):86–91. doi: 10.17925/HI.2020.14.2.86

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Chaudhuri A, Heim F, Chakfe N. Are all wires created the same? A quality assurance study of the stiffness of wires typically employed during endovascular surgery using tension dynamometry. EJVES Vasc Forum. (2021) 52:20–4. doi: 10.1016/j.ejvsvf.2021.06.006

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Gokani VJ, Fishwick N, Choke E, Bown MJ, Collaboration, Monarc. ‘Trial of stiff guidewire’: a useful adjunct to determining suitability for endovascular aneurysm repair. Europ J Vasc Endovasc Surg. (2012) 44:527. doi: 10.1016/j.ejvs.2012.08.012

CrossRef Full Text | Google Scholar

8. Buzzatti N, Mangieri A, Cota L, Ruparelia N, Romano V, Alfieri O, et al. Use of double stiff wire allows successful transfemoral transcatheter aortic valve implantation through extreme thoracic aorta tortuosity. Circ Cardiovasc Interv. (2015) 8(4):e002331. doi: 10.1161/circinterventions.114.002331

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Yuan X, Castro Verdes M, Capoccia M, Rosendahl U, Nienaber CA. Urgent endovascular aortic dissection repair in a marfan patient during COVID-19 pandemic. Clin Case Rep. (2021) 9(8):e04634. doi: 10.1002/ccr3.4634

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Fadahunsi OO, Olowoyeye A, Ukaigwe A, Li Z, Vora AN, Vemulapalli S, et al. Incidence, predictors, and outcomes of per- manent pacemaker implantation following transcatheter aortic valve replacement: analysis from the U.S. society of thoracic surgeons/ American college of cardiology TVT registry. JACC Cardiovasc Interv. (2016) 9:2189–99. doi: 10.1016/j.jcin.2016.07.026

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Bank AJ, Gage RM, Burns KV. Right ventricular pacing, mechanical dyssynchrony, and heart failure. J Cardiovasc Transl Res. (2012) 5(2):219–31. doi: 10.1007/s12265-011-9341-8

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Nadeem F, Tsushima T, Ladas TP, Thomas RB, Patel SM, Saric P, et al. Impact of right ventricular pacing in patients who underwent implantation of permanent pacemaker after transcatheter aortic valve implantation. Am J Cardiol. (2018) 122(10):1712–17.30266255

PubMed Abstract | Google Scholar

13. Urena M, Webb JG, Tamburino C, Muñoz-García AJ, Cheema A, Dager AE, et al. Permanent pacemaker implantation after transcatheter aortic valve implantation: impact on late clinical outcomes and left ventricular function. Circulation. (2014) 129:1233–43. doi: 10.1161/CIRCULATIONAHA.113.005479

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Vijayaraman P, Ponnusamy S, Cano Ó, Sharma PS, Naperkowski A, Subsposh FA, et al. Left bundle branch area pacing for cardiac resynchronization therapy: results from the international LBBAP collaborative study group. JACC Clin Electrophysiol. (2021) 7(2):135–47. doi: 10.1016/j.jacep.2020.08.015

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Liu P, Wang Q, Sun H, Qin X, Zheng Q. Left bundle branch pacing: current knowledge and future prospects. Front Cardiovasc Med. (2021) 8:630399. doi: 10.3389/fcvm.2021.630399

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Guo J, Li L, Xiao G, Huang X, Li Q, Wang Y, et al. Feasibility and stability of left bundle branch pacing in patients after prosthetic valve implantation. Clin Cardiol. (2020) 43(10):1110–8. doi: 10.1002/clc.23413

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Niu H-X, Liu X, Gu M, Chen X, Cai C, Cai M, et al. Conduction system pacing for post transcatheter aortic valve replacement patients: comparison with right ventricular pacing. Front Cardiovasc Med. (2021) 8:772548. doi: 10.3389/fcvm.2021.772548

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Ravi V, Hanifin JL, Larsen T, Huang HD, Trohman RG, Sharma PS. Pros and cons of left bundle branch pacing: a single-center experience. Circ Arrhythm Electrophysiol. (2020) 13(12):e008874. doi: 10.1161/CIRCEP.120.008874

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Calle S, Coeman M, Demolder A, Philipsen T, Kayaert P, De Buyzere M, et al. Aortic valve implantation-induced conduction block as a framework towards a uniform electrocardiographic definition of left bundle branch block. Neth Heart J. (2021) 29(12):643–53. doi: 10.1007/s12471-021-01565-8

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: transcatheter aortic valve implantation (TAVI), valve in valve transcatheter aortic valve implantation, valve in valve implantation, aortic regurgitation (AR), heart failure, cardiac resychronisation therapy, left bundle area pacing, left bundle branch pacing (LBBP)

Citation: Mihailovič PM, Žižek D, Vitez L, Holc P, Klokočovnik T and Bunc M (2023) Case report: A complex case of valve-in-valve TAVI and left bundle branch pacing for severe aortic regurgitation with partially corrected type A aortic dissection and low ejection fraction. Front. Cardiovasc. Med. 10:1206811. doi: 10.3389/fcvm.2023.1206811

Received: 16 April 2023; Accepted: 21 July 2023;
Published: 10 August 2023.

Edited by:

Valeria Cammalleri, Campus Bio-Medico University Hospital, Italy

Reviewed by:

Longfu Jiang, University of Chinese Academy of Sciences, China
Fortunato Iacovelli, Azienda Ospedaliero Universitaria Consorziale Policlinico di Bari, Italy
Lukas Stolz, Ludwig Maximilian University of Munich, Germany
Andrea Buono, Istituto Clinico Sant'Ambrogio, Italy

© 2023 Mihailovič, Žižek, Vitez, Holc, Klokočovnik and Bunc. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Matjaž Bunc mbuncek@yahoo.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.