AUTHOR=Lepillier Antoine , Maggio Ruggero , De Sanctis Valerio , Malacrida Maurizio , Stabile Giuseppe , Zakine Cyril , Champ-Rigot Laure , Anselmino Matteo , Segreti Luca , Dell’Era Gabriele , Garnier Fabien , Mascia Giuseppe , Pandozi Claudio , Dello Russo Antonio , Scaglione Marco , Cosaro Giuseppe , Ferraro Anna , Paziaud Olivier , Maglia Giampiero , Solimene Francesco TITLE=Insight into contact force local impedance technology for predicting effective pulmonary vein isolation JOURNAL=Frontiers in Cardiovascular Medicine VOLUME=10 YEAR=2023 URL=https://www.frontiersin.org/journals/cardiovascular-medicine/articles/10.3389/fcvm.2023.1169037 DOI=10.3389/fcvm.2023.1169037 ISSN=2297-055X ABSTRACT=Background

Highly localized impedance (LI) measurements during atrial fibrillation (AF) ablation have the potential to act as a reliable predictor of the durability of the lesions created.

Objective

We aimed to collect data on the procedural parameters affecting LI-guided ablation in a large multicenter registry.

Methods

A total of 212 consecutive patients enrolled in the CHARISMA registry and undergoing their first pulmonary vein (PV) isolation for paroxysmal and persistent AF were included.

Results

In all, 13,891 radiofrequency (RF) applications of ≥3 s duration were assessed. The first-pass PV isolation rate was 93.3%. A total of 80 PV gaps were detected. At successful ablation spots, baseline LI and absolute LI drop were larger than at PV gap spots (161.4 ± 19 Ω vs. 153.0 ± 13 Ω, p < 0.0001 for baseline LI; 22.1 ± 9 Ω vs. 14.4 ± 5 Ω, p < 0.0001 for LI drop). On the basis of Receiver operating characteristic curve analysis, the ideal LI drop, which predicted successful ablation, was >21 Ω at anterior sites and >18 Ω at posterior sites. There was a non-linear association between the magnitude of LI drop and contact-force (CF) (r = 0.14, 95% CI: 0.13–0.16, p < 0.0001) whereas both CF and LI drop were inversely related with delivery time (DT) (−0.22, −0.23 to −0.20, p < 0.0001 for CF; −0.27, −0.29 to −0.26, p < 0.0001 for LI drop).

Conclusion

An LI drop >21 Ω at anterior sites and >18 Ω at posterior sites predicts successful ablation. A higher CF was associated with an increased likelihood of ideal LI drop. The combination of good CF and adequate LI drop allows a significant reduction in RF DT.