The two-dimensional (2D)-based left atrial appendage (LAA) occluder (LAAO) size determination by using transesophageal echocardiography (TEE) is limited by the structural complexity and wide anatomical variation of the LAA.
This study aimed to assess the accuracy of the LAAO size determination by implantation simulation by using a three-dimensional (3D)-printed model compared with the conventional method based on TEE.
We retrospectively reviewed patients with anatomically and physiologically properly implanted the Amplatzer Cardiac Plug and Amulet LAAO devices between January 2014 and December 2018 by using the final size of the implanted devices as a standard for size prediction accuracy. The use of 3D-printed model simulations in device sizing was compared with the conventional TEE-based method.
A total of 28 cases with the percutaneous LAA occlusion were reviewed. There was a minimal difference [−0.11 mm; 95% CI (−0.93, 0.72 mm);
The use of 3D-printed LA models in the LAAO size determination showed improvement in comparison with conventional 2D TEE method.