Leadless endocardial left ventricular (LV) pacing resynchronization therapy is a novel solution for patients with heart failure (HF) in whom conventional cardiac resynchronization therapy (CRT) failed.
PubMed and the Cochrane Library were searched for relevant cohort studies. Clinical outcomes of interest such as ejection fraction (EF), QRS duration (QRSd), and left ventricular end-systolic volume (LVESV) were extracted and analyzed.
Five studies involving 175 HF patients for WiSE CRT were included, and patients were followed-up for 6 months. The implanted success rate ranged from 76.5 to 100%. WiSE CRT resulted in significantly narrower QRSd [mean difference (MD): −38.21 ms, 95% confidence interval (CI): −44.36 to −32.07,
Evidence from current studies suggests that leadless endocardial LV pacing resynchronization is effective for HF patients who failed conventional CRT or needed a device upgrade, and it may be an interesting rescue therapy.