Invasive assessment of aortic stenosis in contemporary practice

\r\nJoo Brito,*João Brito1,2*Luís Raposo,Luís Raposo1,2Rui Campante Teles,Rui Campante Teles1,2
  • 1Cardiovascular Intervention Unit, Hospital de Santa Cruz, Centro Hospitalar de Lisboa Ocidental, Lisbon, Portugal
  • 2Interventional Cardiology Center, Hospital da Luz, Lisbon, Portugal

The authors review the current role of cardiac catheterization in the characterization of aortic stenosis, its main clinical applications, its pitfalls, and its additional value to the information provided by echocardiography. Discrepancies that may arise between these two modalities are discussed and further explained. Hemodynamic variables besides transvalvular pressure drop are described, and emphasis is given to an integrative approach to aortic stenosis assessment, that includes invasive and noninvasive evaluation.

Introduction

Aortic valve stenosis (AS) is increasingly prevalent in developed countries, and its etiologies include congenital, degenerative (the most common), and rheumatic disease (1). Degenerative AS is a complex process of progressive inflammation, fibrosis, and calcification, affecting an otherwise structurally normal valve at the macroscopic level, which eventually leads to leaflet restriction and related hemodynamic consequences (2), and constitutes the main indication for aortic valve intervention. Regardless of etiology, stenosis of the aortic valve causes obstruction of the blood flow from the left ventricle (LV) to the aorta, which generates a systolic flow-dependent pressure drop (ΔP, a more accurate term for the widely used gradient) across the valve and chronic overload of the LV. Understanding the hemodynamic principles behind AS assessment allows us to critically integrate all the information provided by noninvasive diagnostic modalities and to acknowledge the important role of invasive hemodynamic studies in this setting.

Essential anatomic and functional concepts underlying the measurements of aortic stenosis severity

In degenerative senile AS, calcification induces progressive leaflet immobility and obstruction, leading to a decrease in the aortic valve area (AVA). The narrowed AV orifice leads to the acceleration of blood through the valve, from a lower velocity in the LV outflow tract (LVOT) to the peak velocity at the vena contracta (VC) of the jet. The area between the free edges of the valve leaflets is the true anatomical measure of AVA and is known as the geometric orifice area (GOA). Although it can be measured by planimetry by using either computed tomography (CT) or echocardiogram (usually transesophageal echo), its assessment is challenging, namely due to dependence on image quality and difficulty to locating the exact plane of maximal leaflet opening in a tridimensional structure. The area of flow convergence at the VC is the echocardiographic-obtained AVA, i.e., the effective orifice area (EOA). The latter is smaller than the GOA and corresponds to the smallest measure of AVA. The pressure drop between the LVOT and the EOA is ΔPmax. This decrease in pressure just distally to the valve, in the proximal ascending aorta, is primarily driven by the spatial acceleration of the blood flow (3). As the bloodstream flows to the distal ascending aorta, its kinetic energy is partially converted back into potential energy, resulting in an increase in local pressure. This phenomenon, known as the pressure recovery effect (Figure 1A), has implications for measurements and their interpretation, as will be further discussed below.

FIGURE 1
www.frontiersin.org

Figure 1. Echocardiographic and invasive characterization of aortic stenosis. Panel (A) The geometric orifice area (GOA) is the true anatomical area of the aortic valve and the area of the flow jet at the vena contracta, which occurs downstream of the valve orifice, is the effective orifice area (EOA), and corresponds to the calculated AVA by the continuity equation. GOA is always larger than EOA (they will be equal if GOA has the same size as LVOT). The pressure difference between the LVOT and EOA is known as ΔPmax. The pressure difference between the ascending aorta and LVOT is ΔPnet, as it is recorded after the occurrence of pressure recovery, and corresponds to the measured pressure drop in the catheterization laboratory. In the presence of the pressure recovery phenomenon, ΔPmax is higher than ΔPnet, which partially explains the discrepancies between Doppler and invasive metrics. Panel (B) The shaded area represents the mean transaortic pressure drop (ΔPmean); peak-to-peak pressure drop (ΔPP–P) is the difference between the peak LV pressure and the peak aortic pressure at two different points in time; maximum instantaneous pressure drop (ΔPpeak) is the maximum recorded difference between the LV and aortic pressure at the same point in time. Ao indicates aortic pressure; Asc Ao, ascending aorta; LV, left ventricle pressure; LVOT, left ventricular outflow tract.

Invasive and echocardiographic assessment of aortic stenosis severity

Current guidelines define severe AS as an AVA < 1.0 cm2 or indexed AVA (iAVA) < 0.6 cm2/m2, mean transvalvular pressure drop (ΔPmean) ≥ 40 mmHg, and/or peak transaortic velocity ≥ 4 m/s assessed by Doppler echocardiography (4). Indeed, AS is accurately diagnosed in a significant proportion of patients by Doppler echocardiographic assessment and this is mandatory to guarantee that only suitable patients are referred to valve intervention, considering that a faulty evaluation may prevent a patient from receiving the recommended treatment. In the past, invasive hemodynamic studies were critical for understanding the physiology and pathophysiology of valvular heart disease, but this role was downgraded with the advent of echocardiography, a noninvasive modality. Currently, cardiac catheterization for hemodynamic evaluation of AS is only recommended to accommodate any perceived inconsistencies between clinical and echocardiographic data or if non-invasive imaging is inconclusive (4).

It should be noted that a fundamental difference between these two techniques is that cardiac catheterization can directly measure actual pressure and pressure drops (ΔP), whereas Doppler ultrasound measures velocities that are converted into ΔP by applying the modified (and oversimplified) Bernoulli equation:ΔP=4v2, where v is the peak velocity measured by continuous Doppler through the LVOT and the aortic valve, in m/s. AVA can then be estimated from the velocities across the aortic valve and LVOT using the continuity equation:

A V A = L V O T a r e a × L V O T V T I A V V T I

where VTI is velocity time integral, measured by pulsed Doppler.

Although Doppler echocardiography has been established as the gold standard for assessing AS severity, it should be emphasized that echo parameters were initially derived as surrogates of invasive measurements and that there are important pitfalls that may jeopardize their accuracy. Echocardiography is highly operator-dependent, and image quality may be occasionally mediocre; a lack of alignment between the Doppler beam and the direction of the aortic jet can result in underestimation of the pressure drop and, on the other hand, the pressure drop may be overestimated in severe anemia or conditions associated with high output; AVA calculation relies on the accurate measurement of LVOT diameter, which is challenging and prone to intraobserver and interobserver variability (ranging from 5 to 8%) (5). Since the square of the radius is used to derive the area of the LVOT in the continuity equation, a small measurement error causes a significant error in AVA. Moreover, the LVOT shape is elliptical rather than circular in most of the patients, which may further result in underestimation or overestimation of echo-derived areas (5). Finally, the use of the simplified Bernoulli formulation may introduce a variable source of error, as further discussed in the following section.

For the assessment of AS severity in the cardiac catheterization laboratory, it is essential to accurately measure both the transvalvular pressure drop and cardiac output (CO; flow). ΔP can be obtained by simultaneous measurement of LV and ascending aorta pressures, either by using two arterial accesses, dual lumen fluid-filled catheters, multitransducer micromanometer catheters, or common pressure wires (PWs) (6). Care must be taken with the potential damping of aortic pressure with double-lumen catheters, which may falsely increase the pressure drop. Also, the cross-sectional area of a catheter crossing the aortic valve may increase the measured pressure drop, especially in very tight stenoses, and there is in vitro evidence that catheter geometry may produce significant measurement bias in both the peak pressure and the waveform shape (7). In our experience, 4-to-5 French catheters (pigtail or multipurpose shapes with side holes) will be adequate for most cases. The use of a PW in the LV for pressure measurement further obviates these issues, however, at the expense of a higher procedural cost and the possible need for post-procedure analysis, as some polygraphs will not co-register both PW and fluid-filled signals simultaneously. Non-disposable multitransducer micromanometer catheters are very accurate but costly and are less often used in clinical laboratories (6). Catheterization should allow for the measurement of both ΔPP–P, i.e., the peak-to-peak systolic pressure drop (the difference between the peak LV pressure and the peak aortic pressure), and ΔPmean, i.e., the invasive mean pressure drop (average of instantaneous pressure drops over the ejection period) (Figure 1B). It is important to emphasize that single-catheter pullback curves from the LV to the aorta provide an approximation of the peak-to-peak systolic pressure drop, which is not a physiological measurement since it is obtained at two different points in time and, as such, may be imprecise for diagnostic purposes. The mean pressure drop should be used for diagnosis and measured from at least 3 consecutive beats in patients with sinus rhythm or 8–10 consecutive beats when a rhythm is irregular (8). CO is usually assessed invasively by the Fick method or thermodilution. The Fick method is the gold standard and requires the measurement of real-time oxygen consumption using dedicated equipment, which can be time-consuming and unpractical in the catheterization laboratory. Alternatively, oxygen consumption may be estimated from gender- and age-specific nomograms (indirect Fick method), which constitutes a potential source of error, as the impact of disease states is not accounted for. When thermodilution is used, inaccuracy may result from severe tricuspid regurgitation, cardiac shunts, very low output states, and highly irregular rhythms (8).

Finally, AVA can be calculated from ΔP and CO using the Gorlin equation (9):

A V A ( c m 2 ) = CO ( l / min ) / [ HR ( bpm ) x SEP ( mSec ) ] 44.3 x Δ P ( m m H g )

where SEP is the systolic ejection period, K = 44.3 (empirical constant), and ΔP is the mean pressure drop. It must be noted that this equation has several inherent limitations (mainly stemming from the fact that it has not been primarily derived for the aortic valve) and that accuracy may be lower in patients with bradycardia, tachycardia, aortic regurgitation, or low output states (10).

Thorough invasive evaluation of a patient with AS is multiparametric. It must include measurement of transvalvular pressure drop, CO, and calculation of AVA, but also an appraisal of left ventricular contractility and peripheral vascular resistance. In addition, other indexes can be used to arbitrate inconsistency. Aortic valve resistance can be easily calculated using the same essential parameters and has been suggested to be less flow-dependent than the Gorlin-derived AVA (11). In the end, critical interpretation and integration of all the obtained values are mandatory for a correct diagnosis.

Questions have been raised regarding the risk of embolic stroke resulting from aortic root manipulation and retrograde aortic valve crossing, with one study showing a high frequency of magnetic resonance imaging (MRI) defects in this context (12). However, a subsequent study failed to corroborate these findings (13). One further study showed that the time required for crossing the aortic valve was the most important independent predictor of silent cerebral infarction (14). In this context, the increasing number of transcatheter aortic valve interventions (TAVI) has brought technical and device improvements which have reduced the procedure time and increased its efficacy and safety (15). Besides this potential issue, one should also notice other known complications of cardiac catheterization, such as local vascular injury, bleeding complications, and also the exposure of both patients and operators to ionizing radiation. Taking all these aspects into consideration, the decision for invasive assessment of AS should be judiciously made, when a gain of diagnostic ability is expected, as unnecessary cardiac surgery and TAVI are themselves associated with a risk of neurologic and other systemic complications (16, 17). It is wise to recommend that this procedure should be performed by experienced operators.

Discrepancies between echocardiographic and catheterization findings

There is evidence that the correlation between noninvasive and invasive AS assessments is weaker than previously reported (18, 19). This observation should be highlighted and critically appraised since the values used in guidelines to define severe AS are derived from outcome studies using invasive hemodynamics, whereas echo values are recommended to evaluate AS (20). As previously discussed, several sources of error exist in both echocardiographic and invasive evaluation that can contribute to this discrepancy, starting with the essential assumptions inherent to both techniques. Transvalvular pressure drops derived from catheterization are lower than echo-derived values, and traditionally, this observation has been mainly explained by the pressure recovery phenomenon (21, 22). While Doppler echocardiography measures ΔP from the velocity obtained at the VC, catheterization directly measures the pressure drop between LVOT and the ascending aorta, after the conversion of some kinetic energy back into potential energy (ΔPnet) (Figure 1A). The degree of pressure recovery depends on many factors, including the ratio of actual AVA/ascending aorta area, with more pressure recovery typically observed in patients with larger valve orifice and smaller ascending aorta (22). Therefore, in the presence of significant pressure recovery, invasive pressure drops are lower and the estimated AVA is higher than the corresponding echocardiographic values. However, the exact anatomic point where the pressure is fully recovered is not known and differs from subject to subject, which potentially introduces a source of error in invasive estimations.

Moreover, the very use of the modified Bernoulli equation for the noninvasive assessment of ΔP provides an additional and important explanation for these discrepancies. It relies on two assumptions: (1) the pressure drop is entirely due to spatial acceleration of blood flow, neglecting the impact of unsteady and viscous components, and (2) the blood flow is considered a single streamline, which neglects the velocity distribution across the aortic valve plane (3). While it is known that the first assumption is indeed accurate as the spatial acceleration of blood is the dominant pressure component, there is evidence that the use of a single peak velocity value to the detriment of a complete velocity profile results in consistent overestimation of transvalvular pressure drop and is a source of uncontrolled variability (3, 23).

A comparison of the most important features of these two modalities for AS severity assessment is summarized in Table 1.

TABLE 1
www.frontiersin.org

Table 1. Comparison of cardiac catheterization and Doppler echocardiography for AS evaluation.

Role of cardiac catheterization in low-flow states

Low-flow, low-gradient aortic valve stenosis

A subset of patients with severe AS presents with low CO, ΔPmean< 40 mmHg, and reduced LV ejection fraction. The challenge in this setting is to ensure that the small, calculated AVA is due to true severe AS or “pseudo-aortic stenosis”. In the latter, the aortic valve has the moderate disease, but the leaflet opening is insufficient due to a weak LV, with reduced inotropy. As the Gorlin formula is flow-dependent, the severity of AS may be overestimated in this situation. Dobutamine infusion (whether during catheterization or echocardiography) is the gold standard to differentiate true AS from pseudostenosis, as it induces an increase in inotropy and, consequently, increases CO. In true AS, ΔPmean rises to ≥ 40 mmHg. If CO normalizes but ΔPmean remains low (< 30 mmHg), with an increase in AVA, then pseudostenosis is present. Some patients will not be able to increase CO due to a lack of contractile reserve (defined as an increase in stroke volume <20%). These subjects have indeterminate AS and will require the integration of clinical, imaging, and laboratory parameters for a comprehensive evaluation, bearing in mind that the long-term prognosis after valve intervention is poorer in this group of patients (24).

Paradoxical low-flow, low-gradient aortic valve stenosis

This group of patients presents with iAVA < 0.6 cm2/m2, ΔPmean< 40 mmHg, LVEF > 50%, and indexed stroke volume (SVI) < 35 ml/m2. Despite the preserved ejection fraction, the low-flow state may generally be explained by higher LV filling pressures, reduced systolic longitudinal myocardial shortening, and increased afterload on the LV through decreased systemic arterial compliance and increased systemic vascular resistance causing higher vascular impedance. This pattern may lead to an underestimation of AS severity and prevent appropriate valve intervention. Cardiac catheterization has an important role in the evaluation of this entity when noninvasive metrics are inconclusive or if there is a discrepancy between clinical and echocardiographic findings. One approach in this setting is to evaluate the global LV hemodynamic burden by determining the valvulo-arterial impedance (Zva) by the following equation (25):

Z v a = SBP + Δ Pmean SVI

where SBP is systolic blood pressure and SVI is indexed stroke volume.

Patients with paradoxical low-flow, low-gradient AS tend to have Zva ≥ 5.5 mmHg/ml/m2, and these values are associated with a worse prognosis (26). In this setting, a nitroprusside challenge may be further helpful to assess the fixed component of total left ventricular afterload, unmask true aortic stenosis (27), and ideally predict symptomatic response to hemodynamic relief of aortic stenosis.

Clinical application of invasive assessment of Aortic valve stenosis

To illustrate the previously outlined concepts, we present the case of an 81-year-old overweight woman with a history of chronic obstructive lung disease, deep vein and pulmonary thromboembolism, and reactive depression, who was evaluated in the outpatient clinic with complaints of fatigue and effort dyspnea (New York Heart Association class II). The echocardiogram showed preserved ejection fraction (LVEF 57%) and calcified AS, with a mean pressure drop of 26 mmHg, AVA 0.36 cm2/m2, and SVI 26 ml/m2. These findings were consistent with the diagnosis of paradoxical low-flow, low-gradient AS. However, doubts persisted regarding the severity of AS, and the presence of significant comorbidities suggested that the functional limitation might be due to other causes. Further investigation by CT scan revealed an aortic valve calcium score of 569 AU, which suggested that severe AS was less likely (28). An invasive hemodynamic assessment was performed to reconcile these discrepancies: ΔPmean was 20 mmHg, CO was 5.1 L/min, and SVI was 32.1 ml/m2. AVA estimated by the Gorlin equation was 1.16 cm2. Zva was not elevated (4.4 mmHg.ml–1.m2) as would be expected in paradoxical low-flow low-gradient AS and the intrinsic valve resistance was < 120 dynes.s.cm–5. The usefulness of this latter variable remains controversial, although one study postulates that in low-flow, low-gradient aortic stenosis, such a value identifies pseudo stenosis, while the values of > 180 dynes.s.cm–5 identify truly severe AS (11). Integration of all these invasive parameters allowed us to exclude severe AS and reclassify it as moderate.

Conclusion and future perspectives

Currently, invasive hemodynamic evaluation of AS is indicated to clarify inconsistencies between clinical and echocardiographic findings, or when those findings are not conclusive. Although AS severity assessment relies mostly on the echocardiographic evaluation, one should be aware that the obtained metrics often differ from invasive parameters. While catheterization allows for direct measurement of pressure drop, Doppler echocardiography measures velocities that are converted into pressure drops. This is an essential and distinguishing feature between these two modalities that must be accounted for when interpreting the whole clinical picture. It is critical to understand the hemodynamic concepts behind AS evaluation to identify potential inconsistencies in diagnosis and the subsets of patients that benefit from an integrated approach that includes cardiac catheterization.

Also, although AS is a valve disease, looking exclusively at the valve may be deceiving and it should be noted that understanding the coupling between the valve and the LV is equally essential. Thus, investigation of the extent of myocardial fibrosis (by using cardiovascular magnetic resonance (CMR), echocardiography, or quantification of brain natriuretic peptide levels) may be useful in determining the prognostic impact of AS and potential valve intervention (29, 30). Recent advances in CMR have also proven to provide more precise and accurate values of pressure drop by addressing the limitations of the simplified Bernoulli formulation (3, 31). Interestingly, imaging modalities such as 4D flow CMR have the advantage of acquiring three-dimensional blood velocity vector fields, which have been validated against gold-standard techniques, and potentially overcome a previously discussed limitation of Doppler echocardiography (31). Unanswered questions, whether 4D flow CMR-derived pressure computations correlate accurately with transduced pressure data and whether these measures prove to have a strong prognostic impact, are a field of promising current and future research.

Finally, AS has increasingly become a disease of the elderly and is likely accompanied by multiple comorbidities including LV dysfunction, coronary artery disease, lung disease, and frailty. As such, their symptoms might arise from causes other than aortic stenosis, and establishing this link is often complicated and ambiguous. While cardiac catheterization has a role in clarifying AS severity, a noninvasive test such as a cardiopulmonary exercise test may help to define the symptomatic status and the functional capacity of such patients.

In the current era of expanding TAVI, we have observed a trend for an increased referral of AS patients with different degrees of severity and different hemodynamic states. The role of cardiac catheterization in the accurate hemodynamic characterization of the disease should be considered at a lower threshold, and this integrated, multimodality approach should become the cornerstone of patients’ evaluation for treatment decisions and the best counseling.

Author contributions

JB wrote the first draft of the manuscript. JB, LR, and RT wrote sections of the manuscript. All authors contributed to manuscript revision, read, and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Nkomo VT, Gardin JM, Skelton TN, Gottdiener JS, Scott CG, Enriquez-Sarano M. Burden of valvular heart diseases: a population-based study. Lancet. (2006) 368:1005–11. doi: 10.1016/S0140-6736(06)69208-8

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Rajamannan NM, Subramaniam M, Rickard D, Stock SR, Donovan J, Springett M, et al. Human aortic valve calcification is associated with an osteoblast phenotype. Circulation. (2003) 107:2181–4. doi: 10.1161/01.CIR.0000070591.21548.69

CrossRef Full Text | Google Scholar

3. Donati F, Myerson S, Bissell MM, Smith NP, Neubauer S, Monaghan MJ, et al. Beyond Bernoulli: improving the accuracy and precision of noninvasive estimation of peak pressure drops. Circ Cardiovasc Imaging. (2017) 10:e005207. doi: 10.1161/CIRCIMAGING.116.005207

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J, et al. 2021 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J. (2022) 43:561–632. doi: 10.1093/eurheartj/ehab395

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Baumgartner H, Hung J, Bermejo J, Chambers JB, Evangelista A, Griffin BP, et al. Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice. J Am Soc Echocardiogr. (2009) 22:1–23; quiz101–2. doi: 10.1016/j.echo.2008.11.029

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Kern MJ. The Cardiac Catheterization Handbook. 5th ed. Philadelphia, PA: Saunders (2011). p. 79

Google Scholar

7. de Vecchi A, Clough RE, Gaddum NR, Rutten MC, Lamata P, Schaeffter T, et al. Catheter-induced errors in pressure measurements in vessels: an in-vitro and numerical study. IEEE Trans Biomed Eng. (2014) 61:1844–50. doi: 10.1109/TBME.2014.2308594

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Nishimura RA, Carabello BA. Hemodynamics in the cardiac catheterization laboratory of the 21st century. Circulation. (2012) 125:2138–50. doi: 10.1161/CIRCULATIONAHA.111.060319

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Gorlin R, Gorlin SG. Hydraulic formula for calculation of the area of the stenotic mitral valve, other cardiac valves, and central circulatory shunts. I. Am Heart J. (1951) 41:1–29. doi: 10.1016/0002-8703(51)90002-6

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Baim DS. Grossman’s Cardiac Catheterisation, Angiography, and Intervention. 7th ed. Philadelphia, PA: Lippincot Williams & Wilkins (2006).

Google Scholar

11. Baumgartner H, Hung J, Bermejo J, Chambers JB, Edvardsen T, Goldstein S, et al. Recommendations on the echocardiographic assessment of aortic valve stenosis: a focused update from the European Association of cardiovascular imaging and the American society of echocardiography. J Am Soc Echocardiogr. (2017) 30:372–92. doi: 10.1016/j.echo.2017.02.009

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Omran H, Schmidt H, Hackenbroch M, Illien S, Bernhardt P, von der Recke G, et al. Silent and apparent cerebral embolism after retrograde catheterisation of the aortic valve in valvular stenosis: a prospective, randomised study. Lancet. (2003) 361:1241–6. doi: 10.1016/S0140-6736(03)12978-9

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Hamon M, Gomes S, Oppenheim C, Morello R, Sabatier R, Lognoné T, et al. Cerebral microembolism during cardiac catheterization and risk of acute brain injury: a prospective diffusion-weighted magnetic resonance imaging study. Stroke. (2006) 37:2035–8. doi: 10.1161/01.STR.0000231641.55843.49

CrossRef Full Text | Google Scholar

14. Ropers D, Arnold M, Mundkowski D, Ludwig J, Achenbach S, Daniel W, et al. Prevalence of silent cerebral microembolism after cardiac catheterization of patients with high-grade aortic stenosis. J Am Coll Cardiol. (2010) 55(Suppl. 10):A149.E1400.

Google Scholar

15. Schoels W, Mahmoud MS, Kullmer M, Dia M. A safe and simple technique for crossing stenotic aortic valves. Clin Res Cardiol. (2021) 110:377–81. doi: 10.1007/s00392-020-01744-4

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Jilaihawi H, Chakravarty T, Weiss RE, Fontana GP, Forrester J, Makkar RR. Meta-analysis of complications in aortic valve replacement: comparison of Medtronic-Corevalve, Edwards-Sapien and surgical aortic valve replacement in 8,536 patients. Catheter Cardiovasc Interv. (2012) 80:128–38. doi: 10.1002/ccd.23368

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Messé SR, Acker MA, Kasner SE, Fanning M, Giovannetti T, Ratcliffe SJ, et al. Stroke after aortic valve surgery: results from a prospective cohort. Circulation. (2014) 129:2253–61. doi: 10.1161/CIRCULATIONAHA.113.005084

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Yang CS, Marshall ES, Fanari Z, Kostal MJ, West JT, Kolm P, et al. Discrepancies between direct catheter and echocardiography-based values in aortic stenosis. Catheter Cardiovasc Interv. (2016) 87:488–97. doi: 10.1002/ccd.26033

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Gertz ZM, Raina A, O’Donnell W, McCauley BD, Shellenberger C, Kolansky DM, et al. Comparison of invasive and noninvasive assessment of aortic stenosis severity in the elderly. Circ Cardiovasc Interv. (2012) 5:406–14. doi: 10.1161/CIRCINTERVENTIONS.111.967836

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Pibarot P, Dumesnil JG. Improving assessment of aortic stenosis. J Am Coll Cardiol. (2012) 60:169–80. doi: 10.1016/j.jacc.2011.11.078

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Saikrishnan N, Kumar G, Sawaya FJ, Lerakis S, Yoganathan AP. Accurate assessment of aortic stenosis: a review of diagnostic modalities and hemodynamics. Circulation. (2014) 129:244–53. doi: 10.1161/CIRCULATIONAHA.113.002310

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Abbas AE, Pibarot P. Hemodynamic characterization of aortic stenosis states. Catheter Cardiovasc Interv. (2019) 93:1002–23. doi: 10.1002/ccd.28146

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Dockerill C, Gill H, Fernandes JF, Nio AQX, Rajani R, Lamata P. Blood speckle imaging compared with conventional Doppler ultrasound for transvalvular pressure drop estimation in an aortic flow phantom. Cardiovasc Ultrasound. (2022) 20:18. doi: 10.1186/s12947-022-00286-1

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Burwash IG. Low-flow, low-gradient aortic stenosis: from evaluation to treatment. Curr Opin Cardiol. (2007) 22:84–91. doi: 10.1097/HCO.0b013e32801466f5

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Briand M, Dumesnil JG, Kadem L, Tongue AG, Rieu R, Garcia D, et al. Reduced systemic arterial compliance impacts significantly on left ventricular afterload and function in aortic stenosis: implications for diagnosis and treatment. J Am Coll Cardiol. (2005) 46:291–8. doi: 10.1016/j.jacc.2004.10.081

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Dumesnil JG, Pibarot P, Carabello B. Paradoxical low flow and/or low gradient severe aortic stenosis despite preserved left ventricular ejection fraction: implications for diagnosis and treatment. Eur Heart J. (2010) 31:281–9. doi: 10.1093/eurheartj/ehp361

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Alkhouli M, Eleid MF, Nishimura RA, Rihal CS. The role of invasive hemodynamics in guiding contemporary transcatheter valvular interventions. JACC Cardiovasc Interv. (2021) 14:2531–44. doi: 10.1016/j.jcin.2021.08.072

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Mascherbauer J, Schima H, Rosenhek R, Czerny M, Maurer G, Baumgartner H. Value and limitations of aortic valve resistance with particular consideration of low flow-low gradient aortic stenosis: an in vitro study. Eur Heart J. (2004) 25:787–93. doi: 10.1016/j.ehj.2004.02.011

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Weidemann F, Herrmann S, Störk S, Niemann M, Frantz S, Lange V, et al. Impact of myocardial fibrosis in patients with symptomatic severe aortic stenosis. Circulation. (2009) 120:577–84. doi: 10.1161/CIRCULATIONAHA.108.847772

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Lancellotti P, Moonen M, Magne J, O’Connor K, Cosyns B, Attena E, et al. Prognostic effect of long-axis left ventricular dysfunction and B-type natriuretic peptide levels in asymptomatic aortic stenosis. Am J Cardiol. (2010) 105:383–8. doi: 10.1016/j.amjcard.2009.09.043

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Gill H, Fernandes J, Chehab O, Prendergast B, Redwood S, Chiribiri A, et al. Evaluation of aortic stenosis: from Bernoulli and Doppler to Navier-Stokes. Trends Cardiovasc Med. (2021). doi: 10.1016/j.tcm.2021.12.003

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: aortic stenosis, invasive assessment, echocardiography/catheterization discrepancies, pressure drop, low-flow, integrative approach

Citation: Brito J, Raposo L and Teles RC (2022) Invasive assessment of aortic stenosis in contemporary practice. Front. Cardiovasc. Med. 9:1007139. doi: 10.3389/fcvm.2022.1007139

Received: 30 July 2022; Accepted: 08 November 2022;
Published: 01 December 2022.

Edited by:

Giuseppe Tarantini, University of Padua, Italy

Reviewed by:

Pablo Lamata, King’s College London, United Kingdom
Antonio Miceli, Istituto Clinico Sant’Ambrogio, Italy

Copyright © 2022 Brito, Raposo and Teles. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: João Brito, jdbrito@netcabo.pt

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.