ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 14 June 2021
Sec. Coronary Artery Disease
This article is part of the Research Topic Highlights in Coronary Artery Disease: 2021 View all 8 articles

Acute Mechanical Performance of Magmaris vs. DESolve Bioresorbable Scaffolds in a Real-World Scenario

\nNiklas F. Boeder
Niklas F. Boeder1*Oliver DrrOliver Dörr1Tim KoeppTim Koepp1Florian BlachutzikFlorian Blachutzik1Stephan AchenbachStephan Achenbach2Albrecht ElssserAlbrecht Elsässer3Christian W. Hamm,,Christian W. Hamm1,4,5Holger M. NefHolger M. Nef1
  • 1Medical Clinic I, University Hospital of Giessen, Giessen, Germany
  • 2Medical Clinic 2, University Hospital of Erlangen, Erlangen, Germany
  • 3Department of Cardiology, University Hospital of Oldenburg, Oldeburg, Germany
  • 4Department of Cardiology, Kerckhoff Heart and Thorax Center, Bad Nauheim, Germany
  • 5German Centre for Cardiovascular Research, RheinMain Chapter, Frankfurt am Main, Germany

Background: After the bioresorbable PLLA-based vascular scaffold (Absorb BVS) was taken from the market due to its high adverse event rates, a magnesium-based scaffold (Magmaris) was introduced.

Objective: To compare the acute performance of the sirolimus-eluting magnesium alloy Magmaris scaffold with that of the novolimus-eluting PLLA-based DESolve scaffold in terms of appropriate scaffold deployment using optical coherence tomography (OCT).

Methods and Results: Data from the final OCT pullback of 98 patients were included (19 Magmaris, 79 DESolve) and analyzed at 1-mm intervals. The following indices were calculated: mean and minimal area, residual area stenosis, incomplete strut apposition, tissue prolapse, eccentricity index, symmetry index, strut fracture, and edge dissection. OCT showed a minimum lumen area for Magmaris vs. DESolve of 6.6 ± 1.6 vs. 6.0 ± 1.9 (p = 0.06). Scaffolds with residual area stenosis >20% were predominantly seen in the DESolve group (15.8 vs. 46.8%; p = 0.01). The mean eccentricity index did differ significantly (0.74 ± 0.06 vs. 0.63 ± 0.09; p < 0.001). No fractures were observed for Magmaris scaffolds, but 15.2% were documented for DESolve BRS (p < 0.001). Incomplete scaffold apposition area was significantly higher in the DESolve group (0.01 ± 0.02 vs. 1.05 ± 2.32 mm2; p < 0.001).

Conclusion: This is the first study to compare the acute mechanical performance between Magmaris and DESolve in a real-world setting. The acute mechanical performance of Magmaris BRS seems to be superior to that of DESolve BRS, whereas OCT showed a good acute mechanical performance for both BRS in terms of generally accepted imaging criteria.

Introduction

As the latest innovation in coronary stent therapy, the bioresorbable scaffold (BRS) was rapidly embraced by the interventional cardiologist community due to its potential for long-term benefits and for overcoming limitations inherent to existing drug eluting metallic stents (DES) (1). The absence of permanent metal in the treated vessel wall was seen as an advantage regarding some of the issues that are still unresolved despite improvements in existing DES platforms. BRS were conceived with the concept of offering transient vessel support in that they dissolve after several years. Hereby, they allow and support the restoration of vasomotor function and at the same time facilitate future surgical revascularization (13). However, current BRS have been shown to have important limitations including lower radial strength, lower expansion capabilities, and higher rates of scaffold thrombosis when compared with DES (46). Ultimately, the most widely investigated poly-L-lactic acid (PLLA) BRS, the Absorb BVS (Abbott Vascular, Santa Clara, CA, USA), was taken from the market. More recently, a magnesium-based scaffold was introduced, the Magmaris (Biotronik AG, Bülach, Switzerland). Mechanical properties of the alloy are closer to permanent metallic DES, and this may be reflected in its improved properties over the PLLA scaffolds. In fact, Waksman et al. (7) were able to demonstrate that the Magmaris is significantly less thrombogenic compared with the Absorb BVS in an ex vivo porcine arteriovenous shunt model, and available outcome data are thus far promising (8). Hence, we sought to investigate the acute mechanical performance of Magmaris in comparison with the PLLA-based DESolve BRS (Elixir Medical Corporation, Sunnyvale, CA, USA) in patients who were treated in a real-world scenario.

Materials and Methods

Scaffold Devices

The Magmaris BRS is a sirolimus-eluting magnesium alloy scaffold (9). The alloy degrades to magnesium hydroxide, magnesium phosphate, and amorphous calcium phosphate. Approximately 95% of the magnesium is resorbed at 12 months. Drug dose is 1.4 μm/mm2 and strut thickness is 150 μm. The maximum expandable diameter is 0.6 mm over the nominal diameter at 10 atm. There are two radiopaque tantalum markers at each end (10).

The DESolve BRS is a novolimus-eluting PLLA-based scaffold with a resorption time of 1–2 years. The design comprises sinusoidal hoops with nine peaks and valleys and three connecters per hoop. Strut thickness is 150 μm (11). It can be overexpanded up to 0.5 mm above the nominal diameter. Furthermore, the DESolve BRS is able to self-correct for minor malappositioning (12).

Patient Cohort

Consecutive patients undergoing percutaneous coronary intervention (PCI) with either Magmaris or DESolve BRS irrespective of their clinical presentation were enrolled in this retrospective study. The index procedure was between January 2014 and July 2017 at the University of Gießen.

All patients gave written informed consent. The investigation conforms to the principles outlined in the declaration of Helsinki and was approved by the local ethics committee.

Procedure and OCT Analysis

PCI was performed in accordance with standard clinical practice. The radial approach was favored, using a six French guiding catheter. If necessary, access was switched to femoral. Patients were administered unfractionated heparin at 70 U/kg body weight at the beginning of the procedure. Prior to lesion preparation patients were administered intracoronary nitroglycerine. Lesion preparation was initiated with pre-dilatation utilizing a non-compliant balloon that corresponded 1:1 to vessel size. The use of a debulking device was left to the operator's discretion. BRS was chosen to be sized 1:1 with respect to the vessel diameter. Its deployment was accomplished by slowly inflating the scaffold balloon (1 atm over 10 s, 2 atm over 10 s, and then 2 s per atm). The final pressure was maintained for 20 s. Post-dilatation was performed with a non-compliant balloon in accordance with the maximum expansion limits of the BRS.

Frequency domain optical coherence tomography (OCT) was performed using the Ilumen Optis system (St. Jude Medical, Inc., Minneapolis, MN, USA). Pullbacks were performed at 36 mm/s during contrast injection at a rate of 3–5 ml/s: after having placed the imaging catheter distally to the lesion, the pullback was recorded until either the guiding catheter was reached or the maximum pullback length was completed. A second sequential pullback was combined to image the whole lesion, if necessary. Data from the final pullback just before the end of the procedure were used for the offline analysis in this study.

OCT measurements were performed offline using the LightLab Imaging workstation (St. Jude Medical, Inc.). The pullback was divided into cross-sections at 1-mm intervals within the stented lesion and 5 mm proximally and distally to the scaffold (Figures 1, 2). The following quantitative parameters were determined manually: the percentage of incomplete strut apposition (ISA) at 1-mm intervals, calculated as a percentage of the total number of malapposed struts divided by the total number of struts; the ISA area; the tissue prolapse area, defined as the projection of tissue into the lumen between struts; residual area stenosis (RAS) calculated as [1 – minimum lumen area (MLA)/reference vessel area (RVA)]; the eccentricity index, computed as the ratio between the minimum and maximum diameters; the symmetry index, defined as the difference between maximum scaffold diameter and minimum scaffold diameter divided by the maximum scaffold diameter. If no meaningful value for proximal or distal RVA was obtained, the largest luminal cross-sectional area at the distal or proximal end of the scaffold was used. An edge dissection was defined as any disruption of the vessel luminal surface at the edges of the scaffold with a visible flap (>300 μm). A scaffold fracture was assumed if isolated struts were observed to be unopposed within the scaffold lumen or if struts were stacked and discontinuities were present.

FIGURE 1
www.frontiersin.org

Figure 1. (A) Distal reference vessel area (DRVA) = 3.77 mm2. (B) Asterisk indicating a distal edge dissection. (C) Cross section with minimum eccentricity index (minimum/maximum diameter) = (2.29/3.96 mm) = 0.57. (D) Proximal reference vessel area (DRVA) = 11.59 mm2. Reference vessel area (RVA) = (PRVA + DRVA)/2 = (11.59 + 3.37 mm2)/2 =7.48 mm2.

FIGURE 2
www.frontiersin.org

Figure 2. (A) Malapposition (lumen area – scaffold area) = (6.73 – 6.45 mm2) = 0.28 mm2. (B) Prolaps area (scaffold area – lumen area) (4.95 cm2 – 3.38 mm2) = l.57 mm2.

Quantitative coronary angiography (QCA) analysis was carried out with the help of offline QCA software (Medis Suite 3.2.36.2, Medis Medical Imaging, Leiden, The Netherlands). The following parameters were assessed during post-hoc analysis: reference vessel diameter (RVD) through automatic interpolation, minimum lumen diameter (MLD), percentage area stenosis (AS), and lesion length.

Statistics

Statistical analysis was performed using IBM SPSS Statistics (SPSS Statistics 23, IBM Deutschland GmbH, Ehningen, Germany). Continuous variables with normal distribution are expressed as mean and standard deviation; categorical variables are given as number and percent. Chi-square and Fisher's exact test were used for comparison of categorical variables, and Student's t-test or the Wilcoxon rank-sum test was applied for continuous variables. p < 0.05 were considered statistically significant.

Results

A total of 98 patients were enrolled in this study. Nineteen patients were implanted with a Magmaris BRS and 79 patients were treated with a DESolve BRS. More female patients were enrolled in the Magmaris BRS group (68.4 vs. 36.7%; p = 0.01). Patients in the DESolve BRS group had less frequently been treated by percutaneous intervention (57.9 vs. 26.6%; p = 0.002). Otherwise, patients did not differ significantly with respect to further baseline characteristics, including age (62.0 ± 8.1 vs. 61.7 ± 8.9; p = 0.81) and clinical presentation (Table 1).

TABLE 1
www.frontiersin.org

Table 1. Baseline characteristics.

Lesions were located in the right coronary artery for most of the patients. Overall, there was no difference with regard to the vessel that was treated between both groups (Table 2). Most of the lesions were of low complexity according to AHA/ACC lesion classification. QCA analysis showed no difference between the two groups with respect to reference vessel diameter (2.4 ± 0.7 vs. 2.5 ± 0.6 mm; p = 0.57), area stenosis (68 vs. 74.6%; p = 0.86), and lesion length (12.2 ± 4.5 vs. 10.4 ± 4.8 mm; p = 0.10).

TABLE 2
www.frontiersin.org

Table 2. Angiographic and QCA lesions characteristics.

Lesion preparation started with a thorough pre-dilatation in all cases that were treated with Magmaris, and in the majority of DESolve BRS cases (100 vs. 94.9%; p = 0.33). There was no difference in maximum length and diameter of the balloon used for pre- and post-dilatation (Table 3) or its maximum inflation pressure. The implanted scaffolds were 3.1 ± 0.4 vs. 3.1 ± 0.4 mm (p = 0.19) in mean diameter and 18.9 ± 4.0 vs. 19.7 ± 5.7 mm (p = 0.68) in mean length. The deployment pressure applied did not deviate between patients that were treated with Magmaris or DESolve BRS (15.1 ± 1.9 vs. 13.9 ± 2.6 atm; p = 0.07). Post-dilatation was performed with a non-compliant balloon in all cases.

TABLE 3
www.frontiersin.org

Table 3. Procedural characteristics.

Results of the analysis of the final OCT pullbacks are summarized in Table 4. A total of 1,948 cross-sections were analyzed. Final mean and maximum scaffold diameters were similar between the two groups (mean: 3.1 ± 0.3 vs. 3.1 ± 0.4 mm; p = 0.30; maximum: 3.4 ± 0.3 vs. 3.4 ± 0.5 mm; p = 0.61). Likewise, the mean scaffold area and mean lumen area did not differ significantly between the two groups (Table 4). There was a significant difference in minimum scaffold diameter (2.9 ± 0.2 vs. 2.7 ± 0.4 mm; p = 0.01). The RAS was determined to be 13.5% in patients treated with Magmaris and 16.6% after DESolve implantation (p = 0.23). Overall, there were more DESolve scaffolds that showed a RAS >20% than Magmaris scaffolds (15.8 vs. 46.8%, p = 0.01). Furthermore, there was greater eccentricity as revealed by the mean eccentricity index (0.89 ± 0.2 vs. 0.77 ± 0.1; p < 0.001) and less symmetry according to the symmetry index (0.31 ± 0.08 vs. 0.42 ± 0.09; p < 0.001) in patients treated with a DESolve BRS. Malapposition was less frequently observed in the Magmaris group than in the DESolve group (0.03 vs. 2.3%; p < 0.001). OCT showed dissections more frequently in the DESolve group, and strut fractures were observed only in this group (Table 4).

TABLE 4
www.frontiersin.org

Table 4. Optical coherence tomography findings.

No adverse event occurred within the ensuing post-procedural period. Of note, no scaffold thrombosis was documented.

Discussion

This is the first study to investigate the acute mechanical performance between the PLLA-based DESolve BRS and the magnesium-based Magmaris BRS by means of OCT in a real-world scenario. The principal findings are that OCT imaging criteria revealed a good acute mechanical performance for both BRS, but acute mechanical performance of the Magmaris BRS appeared to be superior to that of the DESolve BRS.

The patients enrolled in this study fulfilled the typical criteria for the implantation of BRS. They were relatively young and—especially in the DESolve BRS group—had a short history of coronary disease. Furthermore, the lesion complexity was predominantly simple and none of the cases involved a bifurcation or ostial lesion.

Imaging studies, which have predominantly used intravascular ultrasound (IVUS) to investigate and evaluate the acute and long-term clinical outcome in patients treated with either bare metal stents or first-generation DES, have found that a cross-sectional MLA <5.5 mm2 and/or an in-scaffold RAS >20% increases the risk of stent thrombosis. Patients treated in our cohort had a mean RAS of 13.5% in the Magmaris group and 16.5% in the DESolve group (p = 0.23). However, patients treated with the latter BRS more frequently had scaffolds with an RAS > 20% (15.8 vs. 46.8% of the respective group; p = 0.01). It should be noted that the periprocedural characteristics, particularly pre- and post-dilatation, did not significantly differ between groups. Moreover, there was only a tendency for the DESolve group to have a smaller MLA (6.6 ± 1.6 vs. 6.0 ± 1.9 mm2; p = 0.06), but this was not significant. As OCT tends to measure lower absolute areas than IVUS (13), the published reference values may not be entirely applicable for the assessment of adequate scaffold deployment; however, both groups showed values above the cut-offs.

Suwannasom et al. (14) found that BRS in general are more frequently associated with asymmetric and eccentric morphology compared with DES. Though, analysis of the final pullbacks shows a significant difference in post-procedural geometry between the two studied device types. In our study of the two BRS devices, the mean eccentricity index after implantation of Magmaris was significantly higher than that of DESolve (0.89 ± 0.2 vs. 0.77 ± 0.10; p <0.001). This finding is supported by the data of Abellás-Sequeiros et al. (15) and is of clinical relevance, as Suwannasom et al. were able to demonstrate that post-procedural asymmetry was independently associated with device-oriented composite endpoints. Therefore, our data suggest a more favorable acute result for the Magmaris group, although both groups met the imaging criteria that were evaluated in the MUSIC trial (16). Here, first-generation DES with an eccentricity index of 0.7 were associated with favorable angiographic results at the 6-month follow-up. The asymmetry can also be assessed by the symmetry index, with a value near zero indicating a symmetric structure throughout the entire length (17). Thus, the acute results after Magmaris implantation also appear to be superior regarding symmetry, showing less bending stiffness and greater flexibility.

Recently, malapposition was identified as a predictor of late and very late scaffold thrombosis in patients treated with BRS (18). One explanation is that malapposed struts may disrupt the laminar flow and activate platelets due to high shear stress (19), thereby contributing to the multifactorial etiology of scaffold thrombosis. Accordingly, event rates were reduced if the operator made use of intravascular imaging modalities to guide the implantation process of DES (20, 21). The prospective, multi-center PRAGUE 19 BRS study revealed 1.1% malapposed struts, much less than the 3.5% observed in the Absorb Cohort B study (22, 23). The rate of incomplete apposition in our cohort was negligible in the Magmaris group and 2.3% in the DESolve group (p < 0.001). It has previously been shown that post-dilatation to reduce malapposition can be done safely in patients undergoing BRS implantation (24); however, incomplete apposition of struts remained a problem in our DESolve group even after post-dilatation. Although the implantation and post-dilatation pressures did not differ between the groups, only patients treated with DESolve BRS showed fractures (0 vs. 15.2%; p = 0.07). This difference further underlines our impression that the PLLA-based platform is more prone to mechanical problems than the magnesium scaffold and suggests that the Magmaris has a greater expansion capacity and radial force. One can also speculate that underlying coronary plaque composition, morphology, and burden may have influenced this result. However, the dependency of expansion and eccentricity seems to be primarily a problem associated with the Absorb BVS (25) rather than the DESolve BRS (26). Barkholt et al. (27) performed a bench evaluation of the mechanical properties of the Magmaris BRS and compared them with those of DES and PLLA-based BRS. They demonstrated that the Magmaris was more resistant to strut fracture than the Absorb BVS. It had a larger crossing profile and similar radial and longitudinal strengths. While recoil after deployment was greater with Magmaris, all devices had similar diameters 120 min after 3.5-mm post-dilatation. Therefore, further randomized clinical trials are required to evaluate and identify the scaffolds that ultimately achieve the goal of matching the performance of the DES and overcoming their long-term limitations.

Limitations

There are several limitations associated with this study. Its registry nature, with retrospective collection of patient data, has inherent limitations and the evidence provided should be seen as hypothesis generating. Furthermore, although the protocols used for lesion preparation, scaffold deployment, and post-dilatation were the same for all operators contributing to this study, their variable application due to the operator's decision may have affected the final acute mechanical result and cannot be excluded. Characteristic of a scaffold may not only be defined by its intrinsic properties but also lesion preparation and underlying morphological aspects. Therefore, it has to be taken into consideration that lesion preparation and post-dilatation numerically that differed between the two groups while not reaching statistical significance may have had an impact on acute mechanical result that cannot be excluded. Furthermore, the sample size of the study was small with few diabetic patients. Treated lesions affected the RCA in most cases while overall, there was no difference between the two groups regarding the distribution of the target vessel. Furthermore, the follow-up was limited to the post-procedural period, and it must be evaluated whether the acute findings translate into an improved long term clinical outcome.

Conclusions

This is the first study that compares the acute mechanical performance of two different BRS that included Magmaris in a real-world setting. The Magmaris BRS seems to be a promising alternative to the DESolve BRS, especially after the bioresorbable PLLA-based Absorb BVS was taken from the market due to its high adverse event rates. The acute mechanical performance of the Magmaris BRS seems to be superior to the DESolve BRS, whereas OCT showed a good acute mechanical performance for both BRS in terms of generally accepted imaging criteria.

Data Availability Statement

The datasets presented in this article are not readily available because of privacy reasons. Requests to access the datasets should be directed to the corresponding author.

Ethics Statement

The studies involving human participants were reviewed and approved by University of Gießen. The patients/participants provided their written informed consent to participate in this study.

Author Contributions

NB and HN designed the study and drafted the manuscript. NB, TK, FB, SA, AE, and CH contributed the data. All authors were involved in critically revising the manuscript.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Acknowledgments

We are grateful to Elizabeth Martinson, Ph.D., of the KHFI Editorial Office, for providing editorial assistance.

References

1. Serruys PW, Garcia-Garcia HM, Onuma Y. From metallic cages to transient bioresorbable scaffolds: change in paradigm of coronary revascularization in the upcoming decade? Eur Heart J. (2012) 33:16–25. doi: 10.1093/eurheartj/ehr384

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Roura G, Homs S, Ferreiro JL, Gomez-Lara J, Romaguera R, Teruel L, et al. Preserved endothelial vasomotor function after everolimus-eluting stent implantation. EuroIntervention. (2015) 11:643–9. doi: 10.4244/EIJY14M07_09

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Costopoulos C, Naganuma T, Latib A, Colombo A. Looking into the future with bioresorbable vascular scaffolds. Expert Rev Cardiovasc Ther. (2013) 11:1407–16. doi: 10.1586/14779072.2013.839200

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Cassese S, Byrne RA, Jüni P, Wykrzykowska JJ, Puricel S, Ndrepepa G, et al. Midterm clinical outcomes with everolimus-eluting bioresorbable scaffolds versus everolimus-eluting metallic stents for percutaneous coronary interventions: a meta-analysis of randomised trials. EuroIntervention. (2018) 13:1565–73. doi: 10.4244/EIJ-D-17-00492

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Wykrzykowska JJ, Kraak RP, Hofma SH, van der Schaaf RJ, Arkenbout EK, IJsselmuiden AJ, et al. Bioresorbable scaffolds versus metallic stents in routine pci. N Engl J Med. (2017) 376:2319–28. doi: 10.1056/NEJMoa1614954

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Capodanno D, Gori T, Nef H, Latib A, Mehilli J, Lesiak M, et al. Percutaneous coronary intervention with everolimus-eluting bioresorbable vascular scaffolds in routine clinical practice: early and midterm outcomes from the european multicentre ghost-eu registry. EuroIntervention. (2015) 10:1144–53. doi: 10.4244/EIJY14M07_11

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Waksman R, Lipinski MJ, Acampado E, Cheng Q, Adams L, Torii S, et al. Comparison of acute thrombogenicity for metallic and polymeric bioabsorbable scaffolds: Magmaris versus absorb in a porcine arteriovenous shunt model. Circ Cardiovasc Interv. (2017) 10:e004762. doi: 10.1161/CIRCINTERVENTIONS.116.004762

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Haude M, Ince H, Kische S, Abizaid A, Tölg R, Alves Lemos P, et al. Safety and clinical performance of a drug eluting absorbable metal scaffold in the treatment of subjects with de novo lesions in native coronary arteries: pooled 12-month outcomes of biosolve-ii and biosolve-iii. Catheter Cardiovasc Interv. (2018) 92:E502–e11. doi: 10.1002/ccd.27680

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Biotronik AG. Magmaris, Product Brochure. 426444/F/Oct_2019 (2019).

Google Scholar

10. Fajadet J, Haude M, Joner M, Koolen J, Lee M, Tölg R, et al. Magmaris preliminary recommendation upon commercial launch: a consensus from the expert panel on 14 april 2016. Eurointervention. (2016) 12:828–33. doi: 10.4244/EIJV12I7A137

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Ormiston JA, Webber B, Ubod B, Darremont O, Webster MW. An independent bench comparison of two bioresorbable drug-eluting coronary scaffolds (absorb and desolve) with a durable metallic drug-eluting stent (ml8/xpedition). Eurointervention. (2015) 11:60–7. doi: 10.4244/EIJY15M02_03

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Wiebe J, Bauer T, Dörr O, Möllmann H, Hamm CW, Nef HM. Implantation of a novolimus-eluting bioresorbable scaffold with a strut thickness of 100 μm showing evidence of self-correction. Eurointervention. (2015) 11:204. doi: 10.4244/EIJY15M02_01

CrossRef Full Text | Google Scholar

13. Gonzalo N, Serruys PW, García-García HM, van Soest G, Okamura T, Ligthart J, et al. Quantitative ex vivo and in vivo comparison of lumen dimensions measured by optical coherence tomography and intravascular ultrasound in human coronary arteries. Rev Esp Cardiol. (2009) 62:615–24. doi: 10.1016/S1885-5857(09)72225-X

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Suwannasom P, Sotomi Y, Ishibashi Y, Cavalcante R, Albuquerque FN, Macaya C, et al. The impact of post-procedural asymmetry, expansion, and eccentricity of bioresorbable everolimus-eluting scaffold and metallic everolimus-eluting stent on clinical outcomes in the absorb ii trial. JACC Cardiovasc Interv. (2016) 9:1231–42. doi: 10.1016/j.jcin.2016.03.027

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Abellás-Sequeiros RA, Ocaranza-Sanchez R, Galvaõ-Braga C, Marques J, Gonzalez-Juanatey C. Magnesium versus poly-l-lactic acid bioresorbable scaffolds: in vivo optical coherence tomography comparison of mechanical performance. Arch Cardiol Mex. (2020) 90:8–15. doi: 10.24875/ACME.M20000090

PubMed Abstract | CrossRef Full Text | Google Scholar

16. de Jaegere P, Mudra H, Figulla H, Almagor Y, Doucet S, Penn I, et al. Intravascular ultrasound-guided optimized stent deployment. Immediate and 6 months clinical and angiographic results from the multicenter ultrasound stenting in coronaries study (music study). Eur Heart J. (1998) 19:1214–23. doi: 10.1053/euhj.1998.1012

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Brugaletta S, Gomez-Lara J, Diletti R, Farooq V, van Geuns RJ, de Bruyne B, et al. Comparison of in vivo eccentricity and symmetry indices between metallic stents and bioresorbable vascular scaffolds: insights from the absorb and spirit trials. Catheter Cardiovasc Interv. (2012) 79:219–28. doi: 10.1002/ccd.22996

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Boeder NF, Weissner M, Blachutzik F, Ullrich H, Anadol R, Tröbs M, et al. Incidental finding of strut malapposition is a predictor of late and very late thrombosis in coronary bioresorbable scaffolds. J Clin Med. (2019) 8:580. doi: 10.3390/jcm8050580

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Gomez-Lara J, Radu M, Brugaletta S, Farooq V, Diletti R, Onuma Y, et al. Serial analysis of the malapposed and uncovered struts of the new generation of everolimus-eluting bioresorbable scaffold with optical coherence tomography. JACC Cardiovasc Interv. (2011) 4:992–1001. doi: 10.1016/j.jcin.2011.03.020

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Witzenbichler B, Maehara A, Weisz G, Neumann FJ, Rinaldi MJ, Metzger DC, et al. Relationship between intravascular ultrasound guidance and clinical outcomes after drug-eluting stents: the assessment of dual antiplatelet therapy with drug-eluting stents (adapt-des) study. Circulation. (2014) 129:463–70. doi: 10.1161/CIRCULATIONAHA.113.003942

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Prati F, Romagnoli E, Burzotta F, Limbruno U, Gatto L, La Manna A, et al. Clinical impact of oct findings during pci: the cli-opci ii study. JACC Cardiovasc Imaging. (2015) 8:1297–305. doi: 10.1016/j.jcmg.2015.08.013

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Kočka V, Malý M, Toušek P, Buděšínský T, Lisa L, Prodanov P, et al. Bioresorbable vascular scaffolds in acute st-segment elevation myocardial infarction: a prospective multicentre study 'prague 19'. Eur Heart J. (2014) 35:787–94. doi: 10.1093/eurheartj/eht545

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Serruys PW, Ormiston JA, Onuma Y, Regar E, Gonzalo N, Garcia-Garcia HM, et al. A bioabsorbable everolimus-eluting coronary stent system (absorb) 2-year outcomes and results from multiple imaging methods. Lancet. (2009) 373:897–9. doi: 10.1016/S0140-6736(09)60325-1

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Blachutzik F, Boeder N, Wiebe J, Mattesini A, Dörr O, Most A, et al. Post-dilatation after implantation of bioresorbable everolimus- and novolimus-eluting scaffolds: an observational optical coherence tomography study of acute mechanical effects. Clin Res Cardiol. (2017) 106:271–79. doi: 10.1007/s00392-016-1048-z

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Shaw E, Allahwala UK, Cockburn JA, Hansen TCE, Mazhar J, Figtree GA, et al. The effect of coronary artery plaque composition, morphology and burden on absorb bioresorbable vascular scaffold expansion and eccentricity - a detailed analysis with optical coherence tomography. Int J Cardiol. (2015) 184:230–6. doi: 10.1016/j.ijcard.2015.01.033

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Boeder NF, Dörr O, Bauer T, Elsässer A, Möllmann H, Achenbach S, et al. Effect of plaque composition, morphology, and burden on desolve novolimus-eluting bioresorbable vascular scaffold expansion and eccentricity - an optical coherence tomography analysis. Cardiovasc Revasc Med. (2019) 20:480–4. doi: 10.1016/j.carrev.2018.07.030

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Barkholt T, Webber B, Holm NR, Ormiston JA. Mechanical properties of the drug-eluting bioresorbable magnesium scaffold compared with polymeric scaffolds and a permanent metallic drug-eluting stent. Catheter Cardiovasc Interv. (2020) 96:E674–E82. doi: 10.1002/ccd.28545

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: bioresorbable scaffold, optical coherance tomography, Magmaris sirolimus-eluting bioresorbable scaffold, coronary heart disease, coronary imaging

Citation: Boeder NF, Dörr O, Koepp T, Blachutzik F, Achenbach S, Elsässer A, Hamm CW and Nef HM (2021) Acute Mechanical Performance of Magmaris vs. DESolve Bioresorbable Scaffolds in a Real-World Scenario. Front. Cardiovasc. Med. 8:696287. doi: 10.3389/fcvm.2021.696287

Received: 16 April 2021; Accepted: 21 May 2021;
Published: 14 June 2021.

Edited by:

Turgay Celik, VM Medical Park Ankara (Kecioren), Turkey

Reviewed by:

Jean Jacques Goy, Fribourg Cantonal Hospital, Switzerland
Rajiv Rampat, William Harvey Hospital, United Kingdom

Copyright © 2021 Boeder, Dörr, Koepp, Blachutzik, Achenbach, Elsässer, Hamm and Nef. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Niklas F. Boeder, bmlrbGFzLmJvZWRlciYjeDAwMDQwO2lubmVyZS5tZWQudW5pLWdpZXNzZW4uZGU=

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.