Skip to main content

ORIGINAL RESEARCH article

Front. Bioeng. Biotechnol., 26 June 2024
Sec. Biomaterials

Personalized compression therapeutic textiles: digital design, development, and biomechanical evaluation

Yu Shi,Yu Shi1,2Rong Liu,
Rong Liu1,2*Chongyang YeChongyang Ye1
  • 1School of Fashion and Textiles, The Hong Kong Polytechnic University, Hong Kong, Hong Kong SAR, China
  • 2Laboratory for Artificial Intelligence in Design, Hong Kong, Hong Kong SAR, China

Physical-based external compression medical modalities could provide sustainable interfacial pressure dosages for daily healthcare prophylaxis and clinic treatment of chronic venous disease (CVD). However, conventional ready-made compression therapeutic textiles (CTs) with improper morphologies and ill-fitting of pressure exertions frequently limit patient compliance in practical application. Therefore, the present study fabricated the personalized CTs for various subjects through the proposed comprehensive manufacturing system. The individual geometric dimensions and morphologic profiles of lower extremities were characterized according to three-dimensional (3D) body scanning and reverse engineering technologies. Through body anthropometric analysis and pressure optimization, the knitting yarn and machinery variables were determined as the digital design strategies for 3D seamless fabrication of CTs. Next, to visually simulate the generated pressure mappings of developed CTs, the subject-specific 3D finite element (FE) CT-leg modelings with high accuracy and acceptability (pressure prediction error ratio: 11.00% ± 7.78%) were established based on the constructed lower limb models and determined tissue stiffness. Moreover, through the actual in vivo trials, the prepared customized CTs efficiently (Sig. <0.05; ρ = 0.97) distributed the expected pressure requirements referring to the prescribed compression magnitudes (pressure error ratio: 10.08% ± 7.75%). Furthermore, the movement abilities and comfortable perceptions were evaluated subjectively for the ergonomic wearing comfort (EWC) assessments. Thus, this study promotes the precise pressure management and clinical efficacy for targeted users and leads an operable development approach for related medical biomaterials in compression therapy.

Introduction

Chronic venous disease (CVD) in biologically lower extremity caused by the valvular incompetence of venous system has been considered as the clinical spectrum ranging from asymptomatic symptoms to venous ulceration (Guo et al., 2019; Kumar et al., 2022). Recently, the prevalence of CVD globally increased from 7.5% to 73% varied by gender, aging, geographic area, and occupational factors (Saliba et al., 2020; Aslam et al., 2022; Gujja et al., 2022; Stücker and Rabe, 2022). Compression therapeutic textiles (CTs)—as the traditional intervention modalities—are extensively applied for the medical treatment and prevention of chronic leg ulcers in physical-based compression therapy (Stücker et al., 2020; Mestre et al., 2022; Xiang et al., 2022). By varying material stiffness, gradient external compressions distributed by elastic knitted CTs could reduce venous wall distension and promote venous hemodynamics from the distal to the proximal regions of lower extremities (Orhurhu et al., 2021; Raffetto et al., 2021). Therefore, CTs lead to beneficial therapeutic functions and support an available solution for CVD clinic treatment and daily healthcare prophylaxis.

In compression management, fabric geometric circumferential dimensions are the size selection criteria of CTs. The applied compression levels—standardized by the measured interfacial pressure dosages at the ankle (B) point—are determined through the medical diagnosis according to a patient’s clinical presentation (Nørregaard et al., 2014). Moreover, the ergonomic wearing comforts (EWCs)—relating to the physical and psychological explorations—primarily impact user compliance with CTs (Nørregaard et al., 2014). The relationships between the body shape, movement freedom, and comfortable perception are crucial for CT EWCs (Jin et al., 2015; Venkatraman et al., 2015). Nevertheless, for existing commercial CTs—due to the regional morphological diversity and measurement discrepancy—the recommended material morphologies and pressure prescriptions from various regions or countries frequently led to the wrong selection and inappropriate applications for different patient groups (Laxa et al., 2016; Jindal et al., 2020). The ill-fitting of CTs commonly increased the risks of threading difficulties, wrinkle, slippage, and discomfort in wearing perceptions (Winslow and Brosz, 2008; Wade et al., 2017), further preventing the compromised therapeutic treatment and exacerbated the unexpected tourniquet effects (i.e., skin break, ulcers, blister, and mobility difficulty) (Nelson and Bell, 2014; Buset et al., 2021; Schupke et al., 2014). The major causes of the poor adherence included the intrinsic disadvantages of the CTs, knowledge gaps in the treatment or management, few recommendations from medical staff, and other sociopsychological factors (Naci et al., 2020). Especially, the established morphology and invariable pressure supply of CTs limited the practical applications and simultaneously caused uncomfortable feelings (Bar et al., 2021). Therefore, personalized CTs with individual-fitted morphology require pressure generation, and subjective EWCs need to be developed to improve functional therapeutic efficacy and user compliance.

In the biodesign and development processes of CTs, three-dimensional (3D) body scanning and anthropometric characterization are frequently utilized for body measurements with high accuracy and reasonable reliability (Kuzmichev et al., 2019; Yan et al., 2020). For instance, through the scanned 3D images and introduced non-uniform rational B-splines method, Wang et al. (2023) proposed a parametric modeling approach to efficiently reconstruct the lower limbs for customized design of CTs. However, in previous studies, the leg circumferences were characterized as the major referring design strategies of user-oriented CTs (Wang et al., 2023). The irregular cross-sectional shapes and varied curvatures of patient bodies frequently led to uneven and insufficient pressure generations (Li et al., 2022). The anatomic structural diversities of targeted users still need to be investigated to improve the clinic efficiency of CTs. Then, to facilitate the fabrication of CTs with custom fits, previous studies have modified Laplace’s law to predict the pressure magnitudes by calculating the material property variables, such as the circumferential mechanical tensions (Hui and Ng, 2001; Barhoumi et al., 2020), stretched ratios (Leung et al., 2010; Siddique et al., 2020), and material dimensional morphologies (Macintyre et al., 2004), etc. However, the corresponding inputting fabric’s physical–mechanical properties could only be achieved by performing multiple knitting attempts and excessive experimental tests. Furthermore, the parametric fitting equations relating to the yarn and machinery knitting variables were established to effectively achieve controllable fabric morphology and pressure behaviors of CTs (Shi et al., 2023; Wang et al., 2022). Nevertheless, the established equations were obtained through the sample measurements using leg mannequins with circular sectional profiles. The input interfacial pressure dosages in digital equations need to be optimized based on the irregular anatomic features of biological bodies. Thus, the pressure fitness and precise management of CTs need to be achieved through individual biological characterization, optimization of pressure values, and digitalization of designed knitting variables.

For estimations of personalized CTs, objective compression performances and subjective EWC assessments are conducted comprehensively to evaluate the medical therapeutic functions and patient adherences. First, instrumental pressure sensors and finite element (FE) modeling were applied for compression magnitude regional measurement and mapping visualization, respectively (Ghorbani et al., 2019; Ghorbani et al., 2020). The physical in vivo testing could directly estimate the specific generated pressure values at each leg position. Conversely, 3D FE models could visually provide the biomechanical distributions and profiles along lower limbs. Second, the EWCs of CTs were commonly evaluated through the movement ability and comfort sensation by actual wearing trails (Teyeme et al., 2021). Through various designed protocols, the postexperiment comfort survey (Baek et al., 2020; Cheng et al., 2022), visual analog scale (VAS) (Ke and Zheng, 2023), and wearing sensation questionnaire (Barhoumi et al., 2022) were performed for the EWC estimations and subjective preferences. Therefore, compression assessment systems of CTs could not only facilitate the evaluation of evidence-based treatment efficacy but also provide scientific information for material property optimization.

Consequently, to improve pressure fitness and EWCs for physical compression therapeutic modalities, personalized CTs are developed based on individual functional requirements and subjective comfortable perceptions. Through 3D reconstruction and body characterization, the geometric and morphologic variations of lower limbs were obtained as the biodesign strategies for the determinations of digitalized knitting variables. After 3D advanced seamless fabrication, the pressure performances and EWCs of CTs were evaluated by the proposed biomechanical FE CT-leg systems and VAS assessments, respectively. The present study will provide operable guidance for the design and manufacturing of biomedical therapeutic materials with precise pressure management and enhanced wearing comforts.

Materials and methods

Subject recruitment

In the present study, three subjects (code: S1, S2, and S3) were voluntarily recruited at The Hong Kong Polytechnic University campus with various occupations and health conditions. Through the Clinical-Etiology-Anatomy Pathophysiology criteria and clinical presentation (Lurie and De, 2020), their prescribed pressure levels (S1-class I [18–21 mmHg]; S2-class II [23–32 mmHg]; and S3-class III [34–46 mmHg]) for physical-based compression therapy were referred to the Germany RAL-GZ 387 standard (Medical Compression Hosiery Quality Assurance; Figure 1A). The individual information and body mass index (BMI) of each subject were as follows: 1) S1: female, age: 51 years, height: 1.57 m, BMI: 25.6 kg/m2; 2) S2: male, age: 33 years, height: 1.80 m, BMI: 17.7 kg/m2; and 3) S3: female, age: 28 years, height: 1.68 m, BMI: 20.6 kg/m2.

Figure 1
www.frontiersin.org

Figure 1. Three-dimensional (3D) body scanning and morphological characterization. (A) Pressure classification and gradient distribution through the Germany RAL-GZ 387 standard. (B) 3D body scanning for data acquisition using handheld EinScan-Pro 2X PLUS Scanner. (C) Clinic measurement guidance of lower limbs in compression therapy. The reconstructed lower extremity models, plotted biological curves, and the corresponding curvature radii of studied cross-sectional slices for subjects (D) S1, (E) S2, and (F) S3, respectively.

3D reconstruction of lower extremity

To construct 3D lower limb models for body morphological characterization and biomechanical performance evaluation, the professional handheld EinScan-Pro 2X PLUS Scanner (Shining 3D Tech Co., Ltd. Hangzhou, China; scan precision: 0.4 mm; scan speed: 1.5 million dots/s) was applied for 2D image capturing (Figure 1B). During data acquisition process, participants were requested to stand steadily with torso separation and maintain a stable breathing state (Choi and Ashdown, 2011). Then, the initial Stereolithography (STL) files (right lower limbs) were imported into the Geomagic Studio 2014 (64 bit) software (Raindrop Geomagic, Research Triangle Park, NC, United States). The scanned derived cloud data was preliminarily processed for extra data elimination, noise reduction, and model hole repairment, respectively. Furthermore, the 3D entity leg models were reconstructed through reverse engineering technology using SpaceClaim Direct Modeler (SCDM; ANSYS, Pennsylvania, Pittsburgh, United States) and computer-aided design systems (Chockalingam et al., 2021). Then, the splattering points were obtained for further body characterization and data calculation.

Anthropometric characterization and optimization of pressure distribution

Through the reconstructed lower extremity models, the digitalized 2D coordinate data was processed by using the MATLAB R2016a system (MathWorks Inc., Natick, MA, United States). According to randomized clinical experiments (Byrne, 2001; Prandoni et al., 2004), the below-knee length of compression stockings (CSs) with improved wearing comforts could efficiently prevent vein thrombosis and thrombotic syndrome. Thus, through the measurement guidance (Figure 1C), four (B, B1, C, and D) anatomic sites were determined as the major studied positions for the biofabrication of CTs. As shown in Figures 1D, E, the leg cross-sectional, biological curves were plotted by the sequential scattering points ([xi, yi], i = 1,2, …, n) for geometric calculation and shape recognition. Furthermore, the curvature radii (r1, r2,., rn) and curvatures (1r1,1r2,...,1rn) of each slice were obtained by applying the Inline function, and dimensional circumferences (CirLeg) of the subject’s lower limbs were calculated through CirLeg=i=1nxi+1xi2+yi+1yi2 (Figure 2A). In our previous study (Shi et al., 2024), the insufficient pressure supply caused by biological irregular shape diversities was optimized by the redistribution development guidance. In detail, based on the characterized standard deviations of curvature (SDC) values of each slice, the profile irregularities of leg cross-sections could be classified by our clustered irregular (IR) levels. Then, referring to the recommended optimization strategies, the reshaped pressure values (PRS) of each leg position for all subjects were obtained to scientifically enhance the delivered dosages of CTs. Through the standardized compression levels and gradient residual pressure ratios (Figure 1A), the specific expected pressure magnitudes (PEX) were designed, as illustrated in Figure 2B. Therefore, the required pressure values (PRQ; PRQ = PEX + PRS) were obtained as the functional individual requirements for further knitting variable design and achievements of precise compression management of CTs. For instance, for the B position of subject S1, the calculated SDC value and corresponding IR level were 1.25 and 2, respectively. Thus, through the reshaped PRS of 5 mmHg and expected PEX of 20 mmHg, the practical PRQ dosage was optimized as 25 mmHg for further CS manufacturing.

Figure 2
www.frontiersin.org

Figure 2. Pressure design and CT fabrication. (A) Leg dimensional circumferences were characterized as the basic size design strategy for the development of personalized CTs. (B) Practical pressure magnitudes of each leg position for each subject were optimized through their expected and reshaped pressure dosages. (C) Through body morphological profile characterization and compression performance optimization, the personalized CTs with 1 × 1 knitted structure were fabricated through the designed yarn combinations (ground yarn and inlay yarn) and machinery variables by the digital computerized development system and 3D seamless medical knitting machine.

Biodesign and fabrication of personalized CTs

In compression development studies, synthetic yarns were commonly utilized to not only improve the elasticity and extensibility but also facilitate the strength and tenacity of knitted fabrics (Sathish et al., 2020). Therefore, the Nylon (Polyamide) double-covered Spandex (Polyurethane) yarn with linear density (D, Denier) of 40D/40D/40D (code: GY; yarn diameter: 0.27 mm) was technically used as the ground yarn, and the Nylon double-covered Spandex yarn with yarn count of 210D/40D/40D (code: Y1; yarn diameter: 0.41 mm), 280D/40D/40D (code: Y2; diameter: 0.42 mm), and 420D/40D/40D (code: Y3; yarn diameter: 0.47 mm) were adopted as the inlay yarns. Thus, in this study, three yarn combinations of GY/Y1, GY/Y2, and GY/Y3 were applied for the fabrication of personalized CTs with various material mechanical property requirements (Figure 2C). The circular seamless CTs were developed by the frequent-used 1 × 1 laid-in weft-knitted structure through a 3D LONATI LA-45 ME advanced medical knitting machine (Francesco Lonati, Brescia, Italy). Through existing experimental investigations (Machin et al., 2022), PYF feeding controller could adjust the feeding speed of inlay yarn materials and considerably influence the fabric circumferential morphologies. In addition, the sizing motor settings could control the loop lengths of knitted loops to vary fabric mechanical tensile behaviors (Shi et al., 2023). To effectively design the yarn and machinery variables, the quantitative relationships between the fabric circumferences (Cirf; cm) and pressure values (PRQ; mmHg) were fitted by our previous explorations through Eqs 1, 2, respectively (Shi et al., 2024). Thus, the specific applied yarn material combinations and machinery parameters for each customized CTs could be digitally determined. Moreover, the practical fabric course (circumferential) stretched ratios were referred to the Germany standard (from 15% to 120%), and inlay yarn diameter was used to replace the yarn linear density for simplifying the parametric calculations. Through the quantitative equations, the yarn-machinery knitting settings for each CTs were obtained as listed in Table 1.

Cirf=0.016×x1+0.002×x2+84.493×x333.201,(1)
PRQn=3.9030.02×x10.062×x2+213.072×x3+0.217×x4,(2)

where x1 is the PYF feeding velocity of the inlay yarn (range: 550–1,300 m/min), x2 is the sizing motor setting values of knitted loops (range: 500–850 mm), x3 is the inlay yarn diameter (mm), and x4 is the fabric course tensile ratio (%) during the wearing state.

Table 1
www.frontiersin.org

Table 1. Designed knitting settings of personalized CTs for recruited subjects

Property investigations of CTs and determination of tissue stiffness

For the instrumental measurements of physical characteristics, fabric loop stitch densities (LD) were observed through clear microscopic images from Leica M165 C electronic device (Wetzlar, Germany) during the relaxed state in accordance with the ASTM D1577-79 standard test method. In addition, the fabric Cirf, thickness (h), and mass density (MD) of tubular CS knitted samples were tested based on the standards of ASTM D3774, ASTM D1777, and ASTM D 3776/D 3776M-09a, respectively. Additionally, the mean errors of Cirf between the designed and tested values were calculated to evaluate the morphologic size fits of user-oriented customized CTs. For mechanical behaviors, the stress–strain curves, Young’s moduli (along the fabric course [Ec] and wale [Ew] directions) and Poisson’s ratios (v) were investigated by Instron 4411 uniaxial tension tester through ASTM D4964 standard. Shear moduli (G) of CTs were tested by the Kawabata (KES-FB3) pure shear testing system. To estimate the compression performance and pressure fitness of designed CTs, the practical interfacial pressure magnitudes (P) were measured at biological lower bodies by applying the Picopress@ pressure tester (Microlab Elettronica, Italy, precision: ±3 mmHg).

Through previous biological FE numerical models, the mechanical properties of soft tissue materials were frequently analyzed by Neo-Hookean model (Korhonen et al., 2011). In this study, the soft muscles—as the main compositions—were determined as the major studied tissue components of lower limbs (Ghorbani et al., 2020). Thus, the muscular tissues were assumed as the hyperelastic, isotropic, homogeneous, and incompressible materials. The constitutive functions are as follows (Chen et al., 2018):

W=C10I1¯3+D1J12;C10=Sm2;D1=Bm2,(3)

where W represents the strain energy density, I1¯ is the first strain variant, and J is the Jacobian determinant of the deformation gradient. C10 and D1 can be obtained under a linear elastic condition. Sm and Bm are the shear and bulk moduli, respectively. Additionally, for the incompressible materials, J value is 1.

Therefore, the subject individual tissue stiffness was measured using Aixplorer@ MultiWave ultrasound system (Supersonic Imagine, Aix-en-Provence, France). The tissue elasticities were physically tested using SuperLinear™ SL10-2 transducer array (element number: 192, bandwidth: 2–10 MHz). Then, the mechanical C10 values for subjects of S1, S2, and S3 (0.029 MPa, 0.020 MPa, and 0.031 MPa) were calculated by Eqs 4, 5.

Eleg=3ρvsw2,(4)
Sm=Eleg21+vleg,(5)

where Eleg and ρ are the tissue Young’s modulus and muscular density (1,000 kg/m3), vsw is the shear wave velocity (0–7.7 m/s), and vleg is the Poisson’s ratio (0.5; Dubois et al., 2018).

Construction and validation of subject-specific biomechanical modeling

3D FE CT-leg biomechanical modeling was established to visually evaluate the pressure performances of personalized CTs, as illustrated by Figure 3A. The geometric models of tubular CTs were constructed according to the actual designed fabric dimensions using ANSYS Workbench Design Modeler software (v19.2, ANSYS, Pennsylvania, Pittsburgh, United States). Then, the knitted CTs and lower limb models (established by Section 2.2) were simultaneously imported into ANSYS LS-DYNA explicit solver to dynamically simulate the wearing process. Commonly, compression-knitted textiles were assumed as the orthotropic elastic materials in mechanical simulations (Weeger et al., 2018). Thus, the corresponding values of MD, Ec, Ew, v, and G, as well as the determined subject-specific tissue properties (C10), were imported into the CT-leg system as the inputting material characteristics. In addition, the CS (shell element; mesh size: 8 mm) and lower limb (solid entity element; mesh size: 7 mm) were meshed by Quadrilateral Dominant and Tetrahedrons elements, respectively. To simulate the sliding process, the upper and bottom leg surfaces were fixed to facilitate the freedom of CS sliding. In addition, the interfacial contact condition was applied to the frictional non-linear contact with a coefficient of 0.2 (Askari and Andersen, 2021). The longitudinal displacements (S1: 34 cm; S2: 44 cm; and S3: 36 cm) of CTs were determined as the boundary conditions of a biomechanical system for pressure behavior visualization. Moreover, to estimate the acceptability of proposed FE models, the validation study was conducted to compare the simulated and tested p values by in vivo measurements.

Figure 3
www.frontiersin.org

Figure 3. Numerical simulation and EWC evaluation protocol. (A) The established 3D FE CT-leg systems—including the geometric models, mesh sizes, mesh element types, and boundary conditions—for pressure performance visualization of recruited subjects of S1, S2, and S3. (B) Designed wearing trial experimental protocol for estimations of the movement ability and wearing comfort. (C) Utilized VAS scoring for EWC evaluation of developed user-oriented CTs.

Experimental protocol and data analysis for EWC evaluation

To evaluate the subjective EWC of designed customized CTs, VAS assessments were applied to estimate the movement ability and comfortable sensation by wearing trials. Participants were instructed to wear the personalized CTs in a controlled standard environment (temperature of 25°C ± 1°C and relative humidity of 70% ± 5%; ASTM D1776-04) for 24 h to achieve the equilibrium status prior to use. In detail, three phases were included to simulate the practical application scenarios (Leung et al., 2021; Figure 3B): 1) phase I: 10-min test in a sitting position for the stabilization of CTs and acclimatization; 2) phase II: 10-min static standing for active recovery; and 3) phase III: 10-min walk at a speed of 5 km/h as normal activity in daily life. Then, the critical VAS scoring forms with ratings of 1–5 were required to be filled by the participants during each phase interval (5 min for each survey fulfillment) (Figure 3C). Moreover, the Pearson correlation tests were conducted to further investigate the relationship between the compression levels with the CT EWCs (Manfei et al., 2017). The collected ranking scores for EWC evaluation were analyzed by Statistical Package for the Social Sciences software (version 23.0, IBM Corporation, United States). Additionally, the level of significance was set at α = 0.05 (Amrherin et al., 2017).

Results and discussion

Physical–mechanical properties and morphologic fits of developed personalized CTs

Through the experimental sample testing, the physical–mechanical properties of each studied stocking part of developed personalized CTs were listed in Table 2. First, accepted for fabric h, the physicla LD and MD of knitted CTs varied related to the applied yarn combinations and machinery parameters (Liu et al., 2013). Generally, fabric LD was decreased from the B to D stocking parts varied by the formed loop sizes through the adjustment of loop length settings (Figure 4). Moreover, for different compression levels of CS shells, the increased inlay yarn diameters positively influenced the fabric MD. Second, as shown in Figure 5A, the mean error ratio of fabric Cirf between the designed and fabricated CTs was approximately 3.87% ± 3.70%. Thus, the stocking dimensions could achieve the basic size fitness for each subject (Wang et al., 2022). Third, for the mechanical tensile behaviors, through the physical mechanisms of Laplace’s law, the generated compression magnitudes were fundamentally determined by the textile tensions and Young’s moduli along the fabric circumferential stretched directions (Shi et al., 2024). Therefore, the measured Ec values varied for material mechanical requirements caused by the designed loop length settings and applied yarn combinations. In detail, as illustrated in Figure 5B, the knitted samples (yarn combination: GY/Y3) for compression of class III could generate high material tensions under the identical uniaxial stretched ratios. Thus, the applied inlay yarn thickness could positively impact the material mechanical stiffness for the required pressure generations.

Table 2
www.frontiersin.org

Table 2. Measured fabric physical–mechanical properties of fabricated CTs

Figure 4
www.frontiersin.org

Figure 4. Microscopic investigations of designed CTs. Images of developed knitted CS samples for subjects (A) S1, (B) S2, and (C) S3, respectively.

Figure 5
www.frontiersin.org

Figure 5. Fabric circumferential characteristics and tensile properties. (A) Fabric morphology comparisons between the designed and fabricated personalized CTs. (B) Mechanical tensile behaviors of knitted CS samples (stocking part of B) with various yarn combinations along the fabric course (circumferential) stretching direction.

During the wearing process, stretched CTs were provided sustainable tensions for interfacial pressure generations (Fun et al., 2011). The corresponding fabric tensile ratios were determined by the dimensional diversities between CTs and applied biological bodies. For ready-made commercial CTs, the fabric Cirf of each part was designed by various manufacturers for extensive user groups. Nevertheless, the provided size tables and recommended selections could not simultaneously conform to the leg morphological variables. The oversized CTs could not generate the required pressure dosages along the low extremities. Conversely, the undersized CTs may lead to excessive compressions, unexpected tourniquet effects, and wearing discomforts (Lozo et al., 2021). Thus, in this study, based on the 3D-body scanning and reverse engineering technologies, biological geometric circumferences and morphologic profiles of each subject were characterized by the obtained anthropometric data. Through the calculated CirLeg values and standardized stretch ratios (15%–120%), Cirf of each part for user-oriented CTs was achieved through designed yarn-machinery knitting settings. Therefore, the customized CTs based on individual leg geometric parameters could facilitate the achievement of pressure precision and patient compliance (Yang et al., 2023).

Validation of proposed subject-specific FE CT-leg system

To validate the prediction precision and acceptability of the proposed subject-specific FE models, the measured and simulated pressure values of each CT were compared, as shown in Figure 6. For example, the tested pressure magnitudes at the leg positions of B, B1, C, and D for subject S1 were approximately 19.49 ± 3.27 mmHg, 18.75 ± 2.18 mmHg, 19.91 ± 2.85 mmHg, and 14.42 ± 3.59 mmHg, respectively (Figure 6A). Similarly, the simulated pressures were approximately 22.84 ± 5.41 mmHg, 23.67 ± 6.42 mmHg, 16.55 ± 7.71 mmHg, and 12.41 ± 4.48 mmHg, respectively. For the customized CTs prepared for subject S3, the prediction error ratios for studied lower limb positions of B, B1, C, and D were 8.43%, 4.73%, 1.42%, and 1.68%, respectively (Figure 6C). Through calculations, the averaged error ratio of established FE CT-leg modeling was approximately 11.00% ± 7.78%. Furthermore, Pearson test results showed that the measured results had significant correlations (Sig. <0.05, ρ = 0.96) with the simulated data (Figure 6D). Thus, the constructed FE mechanical modeling with subject morphologic characteristics and biomechanical properties could be accurately utilized for the pressure prediction and mapping visualization of CTs.

Figure 6
www.frontiersin.org

Figure 6. Validation results of FE systems. Experimental measurement results and simulated data by FE CT-leg systems for subjects (A) S1, (B) S2, and (C) S3, respectively. (D) Pressure comparisons and data correlation between the tested and simulated values of prepared customized CTs.

Previously, pressure evaluations were commonly performed by instrumental sensor devices through in vivo testing. However, for biological bodies—due to the individual anatomic structural characteristics—the irregular interfacial body surfaces and sectional curvatures could lead to uneven and insufficient pressure generations. Thus, the obtained regional magnitudes could not reflect the practical pressure profiles distributed by applied CTs. Moreover, in relevant compression simulation models—the pressure performances—biomechanical transmission behaviors of internal tissue stress and venous hemodynamics could be simulated by constructed 3D modeling. Nevertheless, their geometric body models were established by excessive scanned slices through magnetic resonance imaging or computed tomography scanning (CT). For our proposed FE system, through the 3D-body scanning and reconstructed lower extremity models, the visualization of pressure performances could be efficiently achieved. Additionally, the predicted errors of existing models ranged from 6.0% to 21% (Ghorbani et al., 2019; Ghorbani et al., 2021). Therefore, the proposed FE CT-leg system could be applied for pressure performance evaluation and material property optimization with high prediction accuracy and simulation efficiency.

Pressure performance evaluation of developed personalized CTs

Through the proposed 3D FE modeling, the pressure mappings exerted by developed CTs along lower extremities of each subject were visualized, as shown in Figures 7A–C. Generally, the gradient pressure distributions were regressive from the distal to the proximal regions generated by customized CTs through the development processes. Furthermore, to evaluate the pressure fitness of CTs, the expected compression exertions (PEX) and practical pressure values (P) were compared, as shown in Figure 8. In detail, referring to the Germany standard (Figure 1A), subject S1 expected compressions (class I) ranged from 18 mmHg to 21 mmHg and the tested pressure dosage at the leg B position was 19.49 ± 3.27 mmHg (Figure 8A). For subjects S2 (Figure 8B) and S3 (Figure 8C), the delivered interfacial pressures were 23.59 ± 3.51 mmHg and 35.84 ± 4.80 mmHg, respectively. Moreover, through correlation analysis, the tested P data had good agreements (Sig. <0.05, ρ = 0.97) with the designed PEX values (Figure 8D), and the mean pressure error ratio was 10.08% ± 7.75%. Therefore, the developed personalized CTs could achieve the pressure fitness and medical therapeutic functions for applied users with various requirements.

Figure 7
www.frontiersin.org

Figure 7. FE simulated compression results. Pressure profile mappings obtained by FE CT-leg modeling systems for subjects (A) S1, (B) S2, and (3) S3, respectively.

Figure 8
www.frontiersin.org

Figure 8. Pressure performance evaluation results. Pressure distribution evaluations of the developed customized CTs for subjects (A) S1, (B) S2, and (3) S3, respectively. (D) Data comparison between the user expected and actual tested interfacial pressure dosages of CTs.

Moreover, in previous studies (Gao et al., 2015), customized CTs were developed according to the measured CirLeg values and expected pressure levels. The pressure error ratios between the designed and practical tested values were approximately 5%–52% (Wang et al., 2023). However, in development process, their referred quantitative relationships between the yarn-machinery variable with pressure values were fitted through the leg mannequin measurement with ideal circular cross-sectional profiles. Thus, in the present study, to promote pressure fitness, the insufficient compression exertions caused by individual irregular leg shapes were scientifically improved based on the reshaped biodesign guidance. In addition, through the performance evaluated results, the compression levels and gradient distributions of each developed CS conform to the Germany standard and specific pressure prescription. Thus, the proposed biodesign and development strategies could be utilized for manufacturing CTs with extensive personal requirements in healthcare, medical, and rehabilitation domains (Aydin et al., 2019).

EWC evaluation of developed personalized CTs

The subjective VAS scoring results through the subject wearing trials for EWC assessments of fabricated CTs are shown in Figure 9. Generally, the developed customized CTs for each subject had acceptable movement abilities and wearing comforts in practical application (Kırcı et al., 2021). The elastic knitted CTs for subject S1 had flexible mobility and comfortable perception during the sitting, standing, and walking states because of relatively lower stocking compression of class I (Figures 9A, C). Thus, the fabric stretched tensions for lower levels of pressure generation positively enhanced EWCs of therapeutic compression biomaterials. Through correlation analysis, the generated increasing compression classes significantly decreased (Sig. <0.05, ρ = −0.92) the CT EWCs. Furthermore, through applied scenario comparisons, during the dynamic processes of muscular activity, fabric mechanical tensile stresses varied by leg morphological variations led to an uncomfortable feeling and moveable limitation. Therefore, compared with the active walking states, the sitting and standing postures could subjectively benefit the movement ability and wearing comfort, respectively (Figures 9B, D).

Figure 9
www.frontiersin.org

Figure 9. Subjective assessment results of EWC. (A) Movement ability and (B) comfortable perception assessments of the developed customized CTs for EWC evaluations. (C) Movement ability and (D) comfortable perception comparisons during various applied body postures.

The material design, structural development, and fabrication technique crucially contributed to the CT EWCs. In the present study, the selections of double-covered yarn materials improved the fabric flexibility based on its excellent mechanical elastic recovery properties. For applied yarn materials, the increased yarn diameters provided more fabric stress during the stretching process. Thus, CTs with yarn combinations of GY/Y3 generated high pressure performances along the low limbs. Sequentially, the larger fabric tensions had negative impacts on practical CT EWC. Moreover, for the designed 1 × 1 laid-in loop pattern, the combinations of the ground and inlay yarn components could provide the basic fabric dimensional stabilities during the wearing stretching process. For the machinery parameter design, the stocking dimensional profiles and gradient mechanical material stiffness were achieved by varying the inlay yarn feeding speeds and loop lengths. Therefore, the proper material morphology, functional pressure distribution requirements, and allowing freedom of body movement simultaneously promoted EWCs in practical application. In our future work, natural and biodegradable fibers will be utilized as the yarn compositions to enhance the comfortable perceptions and moisture management for improving wearing comforts.

Furthermore, for prescribing the clinical efficacy of compression therapy, the static stiffness index of CTs reflects the ability to counter the muscle expansion during contraction. Conversely, the dynamic stiffness index indicates the pressure pulsation capability in dynamic wear (Partsch, 2005; Partsch et al., 2006). Through previous studies (Liu et al., 2020; Wang et al., 2023) on the development processes, only the morphological characteristics of leg circumferences were measured as CT design strategies. The individual curvature variations frequently led to inaccuracy or insufficient pressure generations of fabricated CTs. For the evaluation of CTs, only regional interfacial pressure magnitudes along the lower extremities were tested through instrumental sensor devices, lacking pressure mapping profiles and practical wearing perceptions. Thus, the practical functional performances of designed custom fit CTs would be estimated during various human activities in further study.

Conclusion

The present study fabricated the personalized therapeutic CTs for the needed users with individual morphologic and compression requirements. Through reverse engineering technology and body anthropometric characterization, the anatomic biological profiles and geometric variables were obtained as the basic design strategies of material morphology and pressure optimization. The applied knitting yarn combinations and machinery parameters were digitally obtained for the 3D seamless fabrication. Then, through the reconstructed geometric lower extremity models and determined tissue stiffness, the subject-specific 3D FE CT-leg modelings with high prediction accuracy (error ratio: 11.00% ± 7.78%) were established for the biomechanical evaluation of pressure performances. Based on the practical wearing trials, the interfacial compressions of developed customized CTs had good agreements with the expected data (pressure error ratio: 10.08% ± 7.75%). Moreover, through subjective VAS assessments, the CT EWCs had significant (Sig. <0.05, ρ = −0.92) correlations with the exerted pressure levels. Thus, through the proposed digital design, development, and evaluation systems, this study provided a comprehensive manufacturing approach for the therapeutic biomedical materials and promotes the precise pressure management for textile-based compression therapy.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by Human Subjects Ethics Subcommittee of The Hong Kong Polytechnic University. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s), and minor(s)’ legal guardian/next of kin, for the publication of any potentially identifiable images or data included in this article.

Author contributions

YS: Conceptualization, Investigation, Formal analysis, Validation, Methodology, Software, Data analysis, writing–original draft. RL: Conceptualization, Supervision, Resources, Project administration, Funding acquisition, Methodology, writing–review and editing. CY: Conceptualization, Formal analysis, Methodology, Data analysis, Software, Technical support, Validation, writing–review and editing.

Funding

The authors declare that financial support was received for the research, authorship, and/or publication of this article. The authors would like to thank Laboratory for Artificial Intelligence in Design through project RP1-5, Innovation and Technology Fund (ITF), Hong Kong Special Administrative Region, General Research Fund of University Grants Committee through project PolyU252153/18E, and Public Sector Trial Scheme of ITF-The Hong Kong Research Institute of Textiles and Apparel (ITF-HKRITA) through project ITT/010/23TI to support this study.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, publisher, editors, and reviewers. Any product that may be evaluated in this article—or claim that may be made by its manufacturer—is not guaranteed or endorsed by the publisher.

References

Amrhein, V., Korner-Nievergelt, F., and Roth, T. J. P. (2017). The earth is flat (p> 0.05): significance thresholds and the crisis of unreplicable research. PeerJ 5, e3544. doi:10.7717/peerj.3544

PubMed Abstract | CrossRef Full Text | Google Scholar

Askari, E., and Andersen, M. S. (2021). On the effect of friction on tibiofemoral joint kinematics. Appl. Sci. 11 (16), 7516. doi:10.3390/app11167516

CrossRef Full Text | Google Scholar

Aslam, M. R., Muhammad Asif, H., Ahmad, K., Jabbar, S., Hayee, A., Sagheer, M. S., et al. (2022). Global impact and contributing factors in varicose vein disease development. SAGE Open Med. 10, 205031212211189. doi:10.1177/20503121221118992

CrossRef Full Text | Google Scholar

Aydin, A., Atiç, R., Aydin, Z. S., Alemdar, C., Karakoç, M., and Nas, K. (2019). Effects of the use of conventional versus computer-aided design/computer-aided manufacturing sockets on clinical characteristics and quality of life of transfemoral amputees. J. Clin. Anal. Med. 10 (1), 67–71. doi:10.4328/JCAM.5845

CrossRef Full Text | Google Scholar

Baek, H. J., Kim, H. S., Ahn, M., Cho, H., and Ahn, S. J. (2020). Ergonomic issues in brain-computer interface technologies: current status, challenges, and future direction. Comput. Intell. Neurosci. 2020, 4876391–4876392. doi:10.1155/2020/4876397

PubMed Abstract | CrossRef Full Text | Google Scholar

Bar, L., Brandis, S., and Marks, D. J. P. (2021). Improving adherence to wearing compression stockings for chronic venous insufficiency and venous leg ulcers: a scoping review. Patient Prefer. adherence 15, 2085–2102. doi:10.2147/PPA.S323766

PubMed Abstract | CrossRef Full Text | Google Scholar

Barhoumi, H., Marzougui, S., and Abdessalem, S. B. (2020). Clothing pressure modeling using the modified Laplace’s law. Text. Res. J. 38 (2), 134–147. doi:10.1177/0887302X19880270

CrossRef Full Text | Google Scholar

Barhoumi, H., Marzougui, S., and Abdessalem, S. B. (2022). A novel design approach and ergonomic evaluation of Class I compression legging. Int. J. Cloth. Sci. Technol. 34 (2), 273–284. doi:10.1108/IJCST-11-2020-0179

CrossRef Full Text | Google Scholar

Buset, C. S., Fleischer, J., Kluge, R., Graf, N. T., Mosti, G., Partsch, H., et al. (2021). Compression stocking with 100% donning and doffing success: an open label randomised controlled trial. Eur. J. Vasc. Endovascular Surg. 61 (1), 137–144. doi:10.1016/j.ejvs.2020.09.027

CrossRef Full Text | Google Scholar

Byrne, B. J. H. (2001). Deep vein thrombosis prophylaxis: the effectiveness and implications of using below-knee or thigh-length graduated compression stockings. Heart and Lung 30 (4), 277–284. doi:10.1067/mhl.2001.116009

PubMed Abstract | CrossRef Full Text | Google Scholar

Chen, Y., Guo, W., Yang, P., Zhao, J., Guo, Z., Dong, L., et al. (2018). Constitutive modeling of neo-Hookean materials with spherical voids in finite deformation. Extreme Mech. Lett. 24, 47–57. doi:10.1016/j.eml.2018.08.007

CrossRef Full Text | Google Scholar

Cheng, Z., Wu, X., Kuzmichev, V., and Adolphe, D. J. S. (2022). The influence of major ergonomic factors on the demand for underwear in the highly educated male group. Sustainability 14 (19), 12464. doi:10.3390/su141912464

CrossRef Full Text | Google Scholar

Chockalingam, K., Jawahar, N., Muralidharan, N., and Balachandar, K. J. (2021). Development of comfort fit lower limb prosthesis by reverse engineering and rapid prototyping methods and validated with gait analysis. Int. J. Biomed. Eng. Technol. 35 (4), 362–381. doi:10.1504/IJBET.2021.114814

CrossRef Full Text | Google Scholar

Choi, S., and Ashdown, S. P. (2011). 3D body scan analysis of dimensional change in lower body measurements for active body positions. Text. Res. J. 81 (1), 81–93. doi:10.1177/0040517510377822

CrossRef Full Text | Google Scholar

Dubois, G. J., Bachasson, D., Lacourpaille, L., Benveniste, O., and Hogrel, J. (2018). Local texture anisotropy as an estimate of muscle quality in ultrasound imaging. Ultrasound Med. Biol. 44 (5), 1133–1140. doi:10.1016/j.ultrasmedbio.2017.12.017

PubMed Abstract | CrossRef Full Text | Google Scholar

Gao, Z., Liu, T., and Shi, C. (2015). The research of Medical Compression Stockings equipment based on 3D scanner. Pap. Present. A. T. 2015 Int. Conf. Appl. Sci. Eng. Innovation. doi:10.2991/asei-15.2015.381

CrossRef Full Text | Google Scholar

Ghorbani, E., Hasani, H., and Jafari Nedoushan, R. J. (2019). Finite element modelling the mechanical performance of pressure garments produced from elastic weft knitted fabrics. J. Text. Inst. 110 (5), 724–731. doi:10.1080/00405000.2018.1513772

CrossRef Full Text | Google Scholar

Ghorbani, E., Hasani, H., Nedoushan, R. J., and Jamshidi, N. J. F. (2020). Finite element modeling of the compression garments structural effect on the pressure applied to leg. Polymers-Basel 21, 636–645. doi:10.1007/s12221-020-9542-3

CrossRef Full Text | Google Scholar

Gujja, K., Kayiti, T., and Sanina, C. (2022) Chronic venous insufficiency. Interventional cardiology: Principles and practice, 835–843. doi:10.1002/9781119697367.ch87

CrossRef Full Text | Google Scholar

Guo, X., Cui, H., and Wen, X. (2019). Treatment of varicose veins of lower extremity: a literature review. Int. J. Clin. Exp. Med. 12 (3), 2142–2150.

Google Scholar

Hui, C. L., and Ng, S. (2001). Model to predict interfacial pressures in multilayer elastic fabric tubes. Text. Res. J. 71 (8), 683–687. doi:10.1177/004051750107100806

CrossRef Full Text | Google Scholar

Jin, Z., Chen, D., and Yang, Y. (2015) “Research of effect of ergonomics on athletic shoes and costume design project,” in 2015 international conference on education technology, management and humanities science (ETMHS 2015). doi:10.2991/etmhs-15.2015.50

CrossRef Full Text | Google Scholar

Jindal, R., Uhl, J. F., and Benigni, J. (2020). Sizing of medical below-knee compression stockings in an Indian population: a major risk factor for non-compliance. Phlebology 35 (2), 110–114. doi:10.1177/0268355519854611

PubMed Abstract | CrossRef Full Text | Google Scholar

Ke, Y., and Zheng, Q. J. (2023). Development of mining workwear with high ergonomic performance fabrics based on the modular design concept. Int. J. Occup. Saf. Ergonomics 29 (2), 547–554. doi:10.1080/10803548.2022.2057691

CrossRef Full Text | Google Scholar

Kırcı, F., Karamanlargil, E., Duru, S. C., Nergis, B., and Candan, C. J. (2021). Comfort properties of medical compression stockings from biodesigned and cotton fibers. Polymers-Basel 22 (10), 2929–2936. doi:10.1007/s12221-021-0615-8

CrossRef Full Text | Google Scholar

Konarska, M., Soltynski, K., Sudol-Szopinska, I., and Chojnacka, A. (2007). Comparative evaluation of clothing thermal insulation measured on a thermal manikin and on volunteers. Fibres Text. East. Eur. 15 (2), 73.

Google Scholar

Korhonen, R. K., and Saarakkala, S. (2011). Biomechanics and modeling of skeletal soft tissues. Theor. Biomech. 6, 113–132. doi:10.5772/19975

CrossRef Full Text | Google Scholar

Kumar, P., Khan, I. A., Das, A., and Shah, H. J. (2022). Chronic venous disease. Part 1: pathophysiology and clinical features. Clin. Exp. Dermatol. 47 (7), 1228–1239. doi:10.1111/ced.15143

PubMed Abstract | CrossRef Full Text | Google Scholar

Kuzmichev, V. E., Tislenko, I. V., and Adolphe, D. C. (2019). Virtual design of knitted compression garments based on bodyscanning technology and the three-dimensional-to-two-dimensional approach. Text. Res. J. 89 (12), 2456–2475. doi:10.1177/0040517518792722

CrossRef Full Text | Google Scholar

Laxa, M., John, S., Mathew, J., Kopecky, C., Campbell, A., Gumban, D., et al. (2016). One size does not fit all: nurse education for properly fitted anti-embolism compression stockings. J. PeriAnesthesia Nurs. 31 (4), e39. doi:10.1016/j.jopan.2016.04.091

CrossRef Full Text | Google Scholar

Leung, K., Shin, K., Han, F., and Jiao, J. J. (2021). Ergonomic mastectomy bra design: effect on core body temperature and thermal comfort performance. Appl. Ergon. 90, 103249. doi:10.1016/j.apergo.2020.103249

PubMed Abstract | CrossRef Full Text | Google Scholar

Leung, W., Yuen, D., Ng, S. P., and Shi, S. Q. (2010). Pressure prediction model for compression garment design. J. burn care and Res. 31 (5), 716–727. doi:10.1097/BCR.0b013e3181eebea0

PubMed Abstract | CrossRef Full Text | Google Scholar

Li, Q., Sun, G., Chen, Y., Chen, X., Shen, Y., Xie, H., et al. (2022). Fabricated leg mannequin for the pressure measurement of compression stockings. Text. Res. J. 92 (19-20), 3500–3510. doi:10.1177/00405175221083216

CrossRef Full Text | Google Scholar

Liu, R., Lao, T. T., and Wang, S. (2013). Impact of weft laid-in structural knitting design on fabric tension behavior and interfacial pressure performance of circular knits. J. Eng. Fibers Fabr. 8 (4), 155892501300800. doi:10.1177/155892501300800404

CrossRef Full Text | Google Scholar

Liu, R., Xu, B., and Ye, C. (2020). “Biodigital design and functional visualization of multi-class personalized compression textiles for ergonomic fit,” in Advances in design for inclusion: proceedings of the AHFE 2019 international conference on design for inclusion and the AHFE 2019 international conference on human factors for Apparel and textile engineering, 24–28. doi:10.1007/978-3-030-20444-0_51

CrossRef Full Text | Google Scholar

Lozo, M., Lovričević, I., Pavlović, Ž., and Vrljičak, Z. J. (2021). Designing compression of preventive compression stockings. J. Eng. Fibers Fabr. 16, 155892502110604. doi:10.1177/15589250211060406

CrossRef Full Text | Google Scholar

Lurie, F., and De Maeseneer, M. G. (2020). The 2020 update of the CEAP classification: what is new? Eur. J. Vasc. Endovascular Surg. 59 (6), 859–860. doi:10.1016/j.ejvs.2020.04.020

PubMed Abstract | CrossRef Full Text | Google Scholar

Machin, M., Thapar, A., and Davies, A. H. (2022). “Compression stockings,” in In approach to lower limb oedema (Springer), 159–178. doi:10.1007/978-981-16-6206-5_12

CrossRef Full Text | Google Scholar

Macintyre, L., Baird, M., and Weedall, P. (2004). The study of pressure delivery for hypertrophic scar treatment. Int. J. Cloth. Sci. Technol. 16 (1/2), 173–183. doi:10.1108/09556220410520450

CrossRef Full Text | Google Scholar

Manfei, X., Fralick, D., Zheng, J. Z., Wang, B., Changyong, F. J., and Feng, C. (2017). The differences and similarities between two-sample t-test and paired t-test. Shanghai archives psychiatry 29 (3), 184–188. doi:10.11919/j.issn.1002-0829.217070

CrossRef Full Text | Google Scholar

Mestre, S., Triboulet, J., Demattei, C., Veye, F., Nou, M., Pérez-Martin, A., et al. (2022). Acute effects of graduated and progressive compression stockings on leg vein cross-sectional area and viscoelasticity in patients with chronic venous disease. J. Vasc. Surg. Venous Lymphatic Disord. 10 (1), 186–195.e25. doi:10.1016/j.jvsv.2021.03.021

CrossRef Full Text | Google Scholar

Naci, B., Ozyilmaz, S., Aygutalp, N., Demir, R., Baltaci, G., and Yigit, Z. J. (2020). Effects of Kinesio Taping and compression stockings on pain, edema, functional capacity and quality of life in patients with chronic venous disease: a randomized controlled trial. Clin. Rehabil. 34 (6), 783–793. doi:10.1177/0269215520916851

PubMed Abstract | CrossRef Full Text | Google Scholar

Nelson, E. A., and Bell-Syer, S. E. (2014). Compression for preventing recurrence of venous ulcers. Cochrane Database Syst. Rev. 2014 (9), CD002303. doi:10.1002/14651858.CD002303.pub3

PubMed Abstract | CrossRef Full Text | Google Scholar

Nørregaard, S., Bermark, S., and Gottrup, F. J. (2014). Do ready-made compression stockings fit the anatomy of the venous leg ulcer patient? J. wound care 23 (3), 128–135. doi:10.12968/jowc.2014.23.3.128

PubMed Abstract | CrossRef Full Text | Google Scholar

Orhurhu, V., Chu, R., Xie, K., Kamanyi, G. N., Salisu, B., Salisu-Orhurhu, M., et al. (2021). Management of lower extremity pain from chronic venous insufficiency: a comprehensive review. Cardiol. Ther. 10, 111–140. doi:10.1007/s40119-021-00213-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Partsch, H. (2005). The static stiffness index: a simple method to assess the elastic property of compression material in vivo. Dermatol. Surg. 31 (6), 625–630. doi:10.1111/j.1524-4725.2005.31604

PubMed Abstract | CrossRef Full Text | Google Scholar

Partsch, H., Clark, M., Bassez, S., Benigni, J. P., Becker, F., Blazek, V., et al. (2006). Measurement of lower leg compression in vivo: recommendations for the performance of measurements of interface pressure and stiffness. Dermatol. Surg. 32 (2), 224–233. doi:10.1111/j.1524-4725.2006.32039.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Prandoni, P., Lensing, A. W., Prins, M. H., Frulla, M., Marchiori, A., Bernardi, E., et al. (2004). Below-knee elastic compression stockings to prevent the post-thrombotic syndrome: a randomized, controlled trial. Ann. Intern. Med. 141 (4), 249–256. doi:10.7326/0003-4819-141-4-200408170-00004

PubMed Abstract | CrossRef Full Text | Google Scholar

Raffetto, J. D., and Khalil, R. A. (2021). Mechanisms of lower extremity vein dysfunction in chronic venous disease and implications in management of varicose veins. Vessel plus 5, 36. doi:10.20517/2574-1209.2021.16

PubMed Abstract | CrossRef Full Text | Google Scholar

Reich-Schupke, S., Feldhaus, F., Altmeyer, P., Mumme, A., and Stücker, M. J. (2014). Efficacy and comfort of medical compression stockings with low and moderate pressure six weeks after vein surgery. Phlebology 29 (6), 358–366. doi:10.1177/0268355513484142

PubMed Abstract | CrossRef Full Text | Google Scholar

Saliba Júnior, O. A., Rollo, H. A., Saliba, O., and Sobreira, M. L. (2020). Graduated compression stockings effects on chronic venous disease signs and symptoms during pregnancy. Phlebology 35 (1), 46–55. doi:10.1177/0268355519846740

PubMed Abstract | CrossRef Full Text | Google Scholar

Sathish Babu, B., Senthil Kumar, P., and Senthil Kumar, M. J. (2020). Effect of yarn type on moisture transfer characteristics of double-face knitted fabrics for active sportswear. J. Ind. Text. 49 (8), 1078–1099. doi:10.1177/1528083718805717

CrossRef Full Text | Google Scholar

Sau-Fun, N., Chi-Leung, H., and Lai-Fan, W. J. (2011). Development of medical garments and apparel for the elderly and the disabled. Text. Prog. 43 (4), 235–285. doi:10.1080/00405167.2011.573240

CrossRef Full Text | Google Scholar

Shi, Y., Liu, R., and Lv, J. (2023). Effects of knitting variables for pressure controlling of tubular compression fabrics. Int. J. Mater. Text. Eng. 17 (7), 90–94.

Google Scholar

Shi, Y., Liu, R., Lv, J., and Ye, C. (2024). Biomedical therapeutic compression textiles: physical-mechanical property analysis to precise pressure management. J. Mech. Behav. Biomed. Mater. 106392, 106392. doi:10.1016/j.jmbbm.2024.106392

CrossRef Full Text | Google Scholar

Shi, Y., Ye, C., and Liu, R. (2024). A novel optimization approach for bio-design of therapeutic compression stockings with pressure fit. Comput. Biol. Med. 168, 107768. doi:10.1016/j.compbiomed.2023.107768

PubMed Abstract | CrossRef Full Text | Google Scholar

Siddique, H. F., Mazari, A. A., Havelka, A., and Kus, Z. J. F. (2020). Performance characterization and pressure prediction of compression socks. Polymers-Basel 21, 657–670. doi:10.1007/s12221-020-9420-z

CrossRef Full Text | Google Scholar

Stücker, M., Danneil, O., Dörler, M., Hoffmann, M., Kröger, E., and Reich-Schupke, S. J. (2020). Safety of a compression stocking for patients with chronic venous insufficiency (CVI) and peripheral artery disease (PAD). JDDG J. der Deutschen Dermatologischen Gesellschaft 18 (3), 207–213. doi:10.1111/ddg.14042

CrossRef Full Text | Google Scholar

Stücker, M., and Rabe, E. J. (2022). Medical compression stockings for chronic venous diseases and lymphedema: scientific evidence and results of a patient survey on quality of care. Dermatologie 73 (9), 708–717. doi:10.1007/s00105-022-05007-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Teyeme, Y., Malengier, B., Tesfaye, T., Ciesielska-Wrobel, I., Haji Musa, A. B., and Van Langenhove, L. J. A. (2021). A review of contemporary techniques for measuring ergonomic wear comfort of protective and sport clothing. Autex Res. J. 21 (1), 32–44. doi:10.2478/aut-2019-0076

CrossRef Full Text | Google Scholar

Venkatraman, P., and Tyler, D. (2015). Smart materials for sportswear. Mater. Technol. Sportsw. Perform. Appar. 195, 153–170. doi:10.1201/b19359-7

CrossRef Full Text | Google Scholar

Wade, R., Paton, F., and Woolacott, N. J. (2017). Systematic review of patient preference and adherence to the correct use of graduated compression stockings to prevent deep vein thrombosis in surgical patients. J. Adv. Nurs. 73 (2), 336–348. doi:10.1111/jan.13148

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, X., Wu, Z., Xiong, Y., Li, Q., and Tao, X. (2023). Fast NURBS-based parametric modeling of human calves with high-accuracy for personalized design of graduated compression stockings. Comput. Methods Programs Biomed. 229, 107292. doi:10.1016/j.cmpb.2022.107292

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., and Gu, L. (2022). Predictability of pressure characterization of medical compression stockings (MCSs) directly based on knitting parameters. Polymers-Basel 23 (2), 527–536. doi:10.1007/s12221-022-3427-6

CrossRef Full Text | Google Scholar

Wang, Y., and Gu, L. (2023). Patient-specific medical compression stockings (MCSs) development based on mathematic model and non-contact 3D body scanning. J. Text. Inst. 114 (8), 1157–1166. doi:10.1080/00405000.2022.2111644

CrossRef Full Text | Google Scholar

Weeger, O., Sakhaei, A. H., Tan, Y. Y., Quek, Y. H., Lee, T. L., Yeung, S. K., et al. (2018). Nonlinear multi-scale modelling, simulation and validation of 3D knitted textiles. Appl. Compos. Mater. 25, 797–810. doi:10.1007/s10443-018-9702-4

CrossRef Full Text | Google Scholar

Winslow, E. H., and Brosz, D. L. (2008). Graduated compression stockings in hospitalized postoperative patients: correctness of usage and size. AJN Am. J. Nurs. 108 (9), 40–50. doi:10.1097/01.NAJ.0000334973.82359.11

PubMed Abstract | CrossRef Full Text | Google Scholar

Xiang, Y., Zhou, Q., Wu, Z., and Gou, J. (2022). Chronic venous insufficiency in a selected nurse population: a cross-sectional study. Angiology 75, 29–35. doi:10.1177/00033197221130571

PubMed Abstract | CrossRef Full Text | Google Scholar

Yan, S., Wirta, J., and Kämäräinen, J. (2020). Anthropometric clothing measurements from 3D body scans. Mach. Vis. Appl. 31 (1-2), 7. doi:10.1007/s00138-019-01054-4

CrossRef Full Text | Google Scholar

Yang, W. T., Xiong, Y., Wang, S. X., Ren, H. L., Gong, C., Jin, Z. Y., et al. (2023). A randomized controlled trial of standard vs customized graduated elastic compression stockings in patients with chronic venous disease. J. Vasc. Surg. Venous Lymphatic Disord. 12, 101678. doi:10.1016/j.jvsv.2023.08.017

CrossRef Full Text | Google Scholar

Keywords: therapeutic biomaterials, biofabrication, compression supply, biomechanical modeling, performance evaluation

Citation: Shi Y, Liu R and Ye C (2024) Personalized compression therapeutic textiles: digital design, development, and biomechanical evaluation. Front. Bioeng. Biotechnol. 12:1405576. doi: 10.3389/fbioe.2024.1405576

Received: 23 March 2024; Accepted: 23 May 2024;
Published: 26 June 2024.

Edited by:

Qiang Wei, Sichuan University, China

Reviewed by:

Jean Francois Uhl, Université Paris Cité, France
Kaizheng Liu, Chinese Academy of Sciences, China

Copyright © 2024 Shi, Liu and Ye. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Rong Liu, rong.liu@polyu.edu.hk

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.