Knee sleeves improve gait symmetry during fast walking in older adults

  • Health and Medical Research Institute, National Institute of Advanced Industrial Science and Technology (AIST), Takamatsu, Kagawa, Japan

Knee sleeves are commonly used to address knee-related concerns, particularly in older individuals. Although previous studies have demonstrated their efficacy in improving gait and functional outcomes in knees with pathological conditions, the effectiveness of knee sleeves for improving gait characteristics in healthy older adults remains unclear. The harmonic ratio (HR), an index for assessing gait symmetry commonly used to discriminate between individuals with different functional levels, can be used to detect alterations in gait characteristics. This study investigated the effects of knee sleeves on gait symmetry in healthy older adults. Sixteen healthy community-dwelling older adults walked barefoot with and without knee sleeves at normal and fast speeds. Gait symmetry indices (HR and improved HR [iHR]) and spatiotemporal gait parameters were compared under different conditions. A significant interaction between knee condition and walking speed was observed for mean iHR in the anteroposterior direction (p = 0.006). A significant simple main effect of knee condition was found during fast walking, with a larger iHR with knee sleeves than without (p = 0.002). In the condition without knee sleeves, the iHR was significantly lower during fast walking than during normal walking (p = 0.035). Furthermore, a significant main effect of knee condition was observed for the variability of iHR in the anteroposterior direction, with a smaller variability when walking with knee sleeves than when walking without (p = 0.006). These results suggest that knee sleeves may enhance gait symmetry along the anteroposterior direction, particularly during fast walking, where symmetry disruption is more likely than walking at a comfortable pace. A significant reduction in gait symmetry variability also suggests a stabilizing effect on gait dynamics. These findings provide the first evidence supporting the efficacy of knee sleeves for improving gait symmetry. The use of knee sleeves could be a valuable option for restoring disrupted gait symmetry during fast walking, with potential implications for reducing the risk of falls.

1 Introduction

Knee sleeves are commonly used by individuals experiencing knee-related issues, particularly older individuals, in various contexts, including daily activities and clinical settings. Previous studies have demonstrated the effectiveness of knee sleeves in alleviating knee pain (Bryk et al., 2011; Cudejko et al., 2018; Mohd Sharif et al., 2019; Dzidotor et al., 2023) and influencing gait parameters (Schween et al., 2015; Mohd Sharif et al., 2019). However, most observed enhancements in gait and functional aspects appear to be limited to knees with pathological conditions, such as knee osteoarthritis (Bryk et al., 2011; Collins et al., 2011; Collins et al., 2014; Schween et al., 2015; Mohd Sharif et al., 2019) and anterior cruciate ligament reconstruction (Sole et al., 2022). Consequently, the biomechanical effect of knee sleeves during gait in healthy populations remains poorly understood (Ko et al., 2017). Current research appears to lack sufficient quantitative evidence to substantiate the efficacy of knee sleeves with respect to gait characteristics, particularly in healthy older adults.

The harmonic ratio (HR), a quantitative assessment metric for evaluating the quality of walking, has been widely used to discriminate between populations of different ages and during different functions (Menz et al., 2003a; Lowry et al., 2012; Doi et al., 2013; 2015; Bisi et al., 2014; Brodie et al., 2015; Bisi and Stagni, 2016; Howcroft et al., 2016; Pau et al., 2017; 2020; Iijima et al., 2019; Wada et al., 2019; Leban et al., 2020; Conway et al., 2021; Castiglia et al., 2022; Igarashi et al., 2023). The HR index is calculated based on the harmonic content of the acceleration signals, which are generally acquired at a lower trunk level through spectral analysis. It assesses the symmetry of the body’s acceleration pattern within a gait cycle (a stride) (Pasciuto et al., 2017). Since a stride consists of two steps, the acceleration pattern should inherently repeat in multiples of two within any given stride. Consequently, acceleration patterns in both the anteroposterior (AP) and vertical directions are biphasic for each stride, primarily dominated by the second harmonic (Menz et al., 2003b). On the other hand, mediolateral (ML) acceleration differs from AP or vertical accelerations, as the fundamental ML pattern is limb-dependent (left or right) and thus repeats only once per stride. Therefore, ML accelerations are primarily dominated by odd rather than even harmonics (Menz et al., 2003b). Hence, a perfectly symmetrical gait exclusively results in even harmonics along the AP and vertical directions, whereas only odd harmonics are present in the ML direction (Roche et al., 2013). HR is defined as the ratio between the sum of the amplitudes of these harmonics, which are considered intrinsic to gait, and the extrinsic harmonics that lead to deviations from the ideal gait (Roche et al., 2013; Pasciuto et al., 2017). Using this index, previous studies have reported various factors affecting gait symmetry, such as age (Lowry et al., 2012; Bisi et al., 2014; Bisi and Stagni, 2016; Leban et al., 2020), fall risk (Menz et al., 2003a; Doi et al., 2013; Brodie et al., 2015; Howcroft et al., 2016), cognitive impairment (Pau et al., 2020), white matter lesions (Doi et al., 2015), total hip arthroplasty (Wada et al., 2019), knee pain and disability in knee osteoarthritis (Iijima et al., 2019), multiple sclerosis (Pau et al., 2017), malnutrition (Misu et al., 2017), and the effectiveness of rehabilitation in Parkinson’s disease (Conway et al., 2021; Castiglia et al., 2022) or stroke (Igarashi et al., 2023). These studies have clearly demonstrated that the HR index is a sensitive measure for identifying individuals with functional limitations or pathological conditions, suggesting that gait symmetry is altered in these populations. However, to the best of our knowledge, no study has investigated the effects of knee sleeves on gait symmetry in healthy individuals. Whether the use of knee sleeves alters gait symmetry, even in healthy older adults, remains unclear.

The purpose of this study was to examine the effects of knee sleeves on gait symmetry in healthy older adults, both at a comfortable pace and at a faster pace, to challenge motor and postural control. We hypothesized that the application of knee sleeves would improve gait symmetry, particularly under challenging fast walking conditions.

2 Methods

2.1 Participants

Sixteen community-dwelling healthy older adults participated (eight males and eight females, age: 71.8 ± 3.4 years, height: 1.59 ± 0.09 m, body mass: 55.5 ± 8.5 kg). Participants were excluded if they required walking aids, had undergone surgery for trauma or orthopedic conditions, had neurological disorders, or were professional athletes. All participants reported no history of neurological pathology, head trauma, cerebrovascular accident, vestibular dysfunction, or visual impairment uncorrectable by lenses. This study adhered to the principles of the Declaration of Helsinki. The study protocol was approved by the Ethics Committee of AIST, Japan (IRB number: hi2023-534). All participants provided written informed consent before participation.

2.2 Instrumentation and data collection

An inertial measurement unit (IMU) (XsensDOT; Movella Inc., Henderson, NV, United States) was placed on the sacrum and dorsum of each foot (Figure 1A). Acceleration data were collected at 120 Hz. The sacral acceleration was used to calculate the index of gait symmetry, whereas the angular velocities of the feet were used to determine the timing of the heel-strike instant.

Figure 1
www.frontiersin.org

Figure 1. The placements of inertial measurement units (IMUs) and reflective markers. (A) Three IMUs were attached to the sacrum and dorsum of the feet. (B) Fifty-seven reflective markers were attached to various landmarks on the body (see Supplementary Material S1 for abbreviations). This figure was created using OpenSim (Delp et al., 2007; Seth et al., 2018).

Fifty-seven reflective markers were placed on each participant’s bony landmarks (Kobayashi et al., 2019) (Figure 1B). Three-dimensional marker trajectories were collected using a 10-camera motion capture system (MAC3D; Motion Analysis Corp., Santa Rosa, CA, United States) at 200 Hz. Trajectories were smoothed using a fourth-order Butterworth filter with an 8-Hz cut-off frequency (Chang et al., 2008; Schmitz et al., 2014). The center of the hip joint was estimated using the method described by Bell et al. (1987), Bell et al. (1990). The centers of the knee and ankle joints were defined as the midpoints of the femoral epicondyles and malleoli, respectively (Inai et al., 2021). Marker trajectories and joint centers were used to calculate the gait parameters.

2.3 Protocol

Participants walked barefoot back and forth along a 15-m straight pathway five times at two different walking speeds (Normal [Norm] and Fast), both with and without knee sleeves (KS and CO). They were instructed to walk at their preferred, comfortable speed in the Norm speed condition. For the Fast speed condition, they were asked to walk as fast as possible. They wore knee sleeves (Actcyc Walk [ACT-80], Kagawa Seamless Inc., Kagawa, Japan; Figure 2) on both sides during the KS condition and walked without wearing them during the CO condition. A knee sleeve with a standard wraparound design without a hinge was selected, considering the need for regular use by the participants. Given the potential impact on subsequent comfortable walking speeds and patterns when performing fast walking first (Zhang et al., 2022), the Norm speed condition was followed by the Fast speed condition. The order of the CO and KS conditions was counterbalanced among the participants.

Figure 2
www.frontiersin.org

Figure 2. Knee sleeves used in this study. Each participant wore knee sleeves on both knees.

2.4 Data analysis

2.4.1 Gait symmetry

The HR (no unit) and improved HR (iHR; unit: %) were calculated using frequency analysis to evaluate the symmetry of the acceleration pattern in the AP, ML, and V directions (Roche et al., 2013; Pasciuto et al., 2017). The sacral IMU accelerations were used for the calculations. Since filtering can potentially cause harmonic distortion (Pasciuto et al., 2017), the raw signals were left unfiltered. Higher values in the HR and iHR indices indicate greater symmetry. The harmonic contents of the acceleration signal, categorized as “in phase” (intrinsic harmonics), were compared to those categorized as “out of phase” (extrinsic harmonics). Since a perfectly symmetrical gait exhibits even harmonics along the AP and V directions, whereas only odd harmonics are present in the ML direction, we considered even harmonics as intrinsic for the AP and V accelerations, and odd harmonics as intrinsic for the ML acceleration (Roche et al., 2013; Pasciuto et al., 2017). The HR was then calculated by dividing the sum of the amplitudes of the intrinsic harmonics by the sum of the amplitudes of the extrinsic harmonics. Specifically, a perfectly symmetrical gait would yield only even harmonics in the AP and V directions and only odd harmonics in the ML direction as “in phase” (intrinsic harmonics) (Roche et al., 2013; Pasciuto et al., 2017). The iHR is an improved index that demonstrates greater robustness and reliability than the traditional HR (Pasciuto et al., 2017). It was proposed to address the following limitations of the HR index: (1) its significant variability in highly symmetrical gait due to the minor contribution of extrinsic harmonics (in the denominator), (2) its unintuitive interpretation, and (3) its lack of mathematical rigor. Derived from the power of the harmonics involved and expressed as a percentage, the iHR allows for a more intuitive understanding and interpretation. The formulas for calculating HR and iHR are as follows (Pasciuto et al., 2017):

HR=j=1kAIjj=1kAEj,
iHR=j=1kAIj2j=1kAIj2+j=1kAEj2100,

where k represents the number of intrinsic (or extrinsic) harmonics (k = 10). Therefore, the total number of considered harmonics is 20. AIj and AEj indicate the amplitudes of the j th intrinsic and extrinsic harmonic, respectively.

This calculation involved 20 harmonics and 20 strides (four strides per straight-path round trip) (Pasciuto et al., 2017). We calculated the mean and standard deviation (SD) of the HR and iHR in the AP, ML, and vertical directions using the 20 strides for each participant. These values were used to represent the mean and within-subject variability. To define a gait cycle (one stride), the heel-strike timing was identified by detecting the zero-crossing point using the foot angular velocity measured by the IMU in the sagittal plane (Jung et al., 2021; Shi et al., 2023). We utilized our original codes written in Scilab 6.1.1 (Dassault Systèmes, Vélizy-Villacoublay, France) for data analysis, including the extraction of harmonics and calculation of the HR and iHR.

2.4.2 Gait parameters

Spatiotemporal gait parameters (walking speed, peak walking velocity and acceleration, step length, stride length, walk ratio, cadence, step time, stride time, stance time, swing time, and minimum toe clearance) and joint angles (peak hip flexion and extension angles, peak knee flexion angle, peak ankle dorsiflexion, and plantarflexion angles, and their range of motion) were calculated based on three-dimensional motion capture data.

Walking speed was calculated using stride time and AP displacement of the whole-body center of mass (COM). Using body parameters, the COM was calculated as the weighted sum of 15 body segments, including the head and neck, trunk, pelvis, upper arms, forearms, hands, thighs, shanks, and feet (Robertson et al., 2008). Walking velocity and acceleration were calculated as the first and second derivatives of AP COM displacement, respectively. Step and stride lengths were calculated as the AP distances between the heel markers from the first to the second heel strike of the contralateral and ipsilateral legs, respectively. Both parameters were normalized to body height. The walk ratio was calculated by dividing the step length by the cadence (Lindemann et al., 2014; Peng et al., 2020). Cadence was calculated using stride time (i.e., 60 × 2/stride time). The stride, stance, and swing times were determined using the heel-strike and toe-off timings. Heel-strike and toe-off timings were detected using midfoot vertical velocity (O’Connor et al., 2007). The hip, knee, and ankle joint angles were computed using the local coordinate systems of the pelvis, thigh, shank, and foot in the ipsilateral leg (Inai et al., 2023). The sagittal plane range of motion for these joints during the gait cycle was also calculated. Minimum toe clearance was defined as the minimum value of the toe marker in the vertical direction during 30%–90% of the swing phase (Yamagata et al., 2019). It was then normalized to body height (Ullauri et al., 2019; Inai et al., 2023).

2.5 Statistical analysis

Two-way repeated-measures analysis of variance (ANOVA) with speed (Norm and Fast) and condition (CO and KS) was used to examine the main and interaction effects. When an interaction was confirmed, a paired t-test was used to determine simple main effects. The index of effect size (d for pairwise comparison; partial eta-squared, ηp2 for ANOVA) was reported as p-values. Small, medium, and large effects for Cohen’s d were defined as 0.2–0.5, 0.5–0.8, and >0.8, respectively (Cohen, 1992). Small, medium, and large effects for ηp2 were defined as 0.01–0.06, 0.06–0.14, and >0.14, respectively (Cohen, 1992). Statistical analyses were performed using R language 4.3.0 (R Core Team, Vienna, Austria). Statistical significance was set at p < 0.05 for all comparisons.

3 Results

3.1 Gait symmetry

Figure 3 shows mean HR and iHR. A significant interaction between speed and condition was observed for the mean iHR in the AP direction (p = 0.006, ηp2 = 0.40). A simple main effect of condition was found for Fast speed. A pairwise comparison revealed that iHR was significantly larger in the KS condition compared with the CO condition (p = 0.002, d = 0.53; CO: 88.9% ± 4.8% vs KS: 91.2% ± 3.6%). A simple main effect of speed was also observed under the CO condition. A pairwise comparison revealed that iHR was significantly smaller in the Fast speed condition compared to the Norm speed condition (p = 0.035, d = 0.52; Norm: 91.2% ± 3.9% vs Fast: 88.9% ± 4.8%). No significant interactions were observed for the other gait symmetry indices.

Figure 3
www.frontiersin.org

Figure 3. Means of harmonic ratio (HR) and improved HR (iHR). AP: antero-posterior; ML: medio-lateral; V: vertical. * Significant interaction (p < 0.05). ‡ Significant main effect of speed (p < 0.05).

Significant main effects of speed were observed for mean HR and iHR in the ML direction (p = 0.019, ηp2 = 0.32, and p = 0.038, ηp2 = 0.26, respectively). The HR and iHR were both significantly larger in the Fast speed condition compared to the Norm speed condition (Norm: 2.4 ± 0.7 vs Fast: 2.6 ± 0.6 and Norm: 80.7% ± 7.2% vs Fast: 83.3% ± 7.8%, respectively).

Figure 4 shows the within-subject variability of the HR and iHR. No significant interactions were observed for any of the gait symmetry indices. A significant main effect of condition was observed for the variability of iHR in the AP direction (p = 0.006, ηp2 = 0.41). The variability was smaller in the KS condition compared with the CO condition (CO: 6.7% ± 3.1% vs KS: 5.9% ± 2.6%).

Figure 4
www.frontiersin.org

Figure 4. Variabilities in the harmonic ratio (HR) and improved HR (iHR). AP: antero-posterior; ML: medio-lateral; V: vertical. † Significant main effect of condition (p < 0.05).

3.2 Gait parameters

Table 1 shows the mean gait parameters of the participants (see Supplementary Material S2 for details). No significant interactions were observed for any parameters. Significant main effects of the condition were observed for the mean swing time, peak hip flexion angle, range of motion of the hip and knee joints, and minimum toe clearance (p ≤ 0.024). Significant main effects of speed were observed for the mean walking speed, peak walking speed and acceleration, step and stride lengths, cadence, step and stride times, stance time, swing time, percentages of stance phase and swing phase, and peak hip flexion and ankle plantar flexion angles (p ≤ 0.020).

Table 1
www.frontiersin.org

Table 1. Gait parameters.

Table 2 shows the variability in the gait parameters (see Supplementary Material S3 for details). A significant interaction between the condition and variability of the range of motion of the ankle joint was observed (p = 0.046, ηp2 = 0.24). However, no simple main effect was observed in the post-analysis. No significant interactions were observed for any other gait parameters. Significant main effects of the condition were observed for variabilities in peak hip flexion and peak ankle plantar flexion angles (p ≤ 0.032). Significant effects of speed were observed for variabilities in walking speed, peak walking speed and acceleration, stride length, cadence, percentages of stance and swing phases, peak hip flexion angle, and range of motion of the knee joint (p ≤ 0.044).

Table 2
www.frontiersin.org

Table 2. Variabilities in gait parameters.

4 Discussion

This study examined the effects of knee sleeves on gait symmetry. Gait symmetry indices (HR and iHR) and spatiotemporal gait parameters were compared during walking with and without knee sleeves (KS and CO, respectively) under normal and fast walking speed conditions. A significant interaction between knee condition and walking speed was observed for mean iHR in the AP direction. A significant simple main effect of knee condition was found during fast walking, revealing a significantly larger iHR with knee sleeves than without knee sleeves. In the condition without knee sleeves, iHR was significantly lower during fast walking than during normal walking. Furthermore, a significant main effect of knee condition was observed for the variability in iHR in the AP direction, indicating a significantly smaller variability when walking with knee sleeves than without. These results suggest that knee sleeves may enhance gait symmetry, particularly in the AP direction during faster-paced walking, where symmetry might be disrupted compared to walking at a comfortable pace.

Without the use of knee sleeves, the iHR in the AP direction significantly decreased during fast walking compared with normal walking, indicating a reduction in gait symmetry. This finding aligns with previous studies that demonstrated a lower HR along the AP axis at increased walking speeds (Menz et al., 2003b; Latt et al., 2008). This reduction in symmetry index may be associated with an elevated risk of falls during fast walking in older adults. Previous studies have reported an increased likelihood of falls when walking at a faster speed (≥1.3 m/s) (Quach et al., 2011). Similarly, it has been reported that the HR in the AP direction is the highest at the preferred speed (approximately 1.2 m/s) and decreases as the speed increases (Latt et al., 2008). Furthermore, a recognized association exists between gait symmetry and fall risk, with a low HR along the AP axis indicating a higher risk of falls (Menz et al., 2003a; Doi et al., 2015). Walking at a pace faster than one’s comfortable speed appears to challenge the postural control system, potentially disrupting gait symmetry in the AP direction and subsequently increasing the risk of falls.

Based on our results, knee sleeves may be a viable approach to enhance gait symmetry during fast walking. The application of knee sleeves resulted in a significant increase in the iHR in the AP direction during fast walking, reaching a level comparable to that during walking at normal speed. Additionally, a significant decrease in variability in the AP direction was observed with the knee sleeves. These findings suggest that the gait symmetry is enhanced and stabilized when walking faster with knee sleeves. Compression applied to the thigh and shank by the knee sleeves may have contributed to these improvements. Although the underlying mechanisms have not been fully elucidated, previous studies have reported an enhanced joint position sense using knee sleeves (Birmingham et al., 1998; Herrington et al., 2005; Mohd Sharif et al., 2017; Dzidotor et al., 2023), elastic bandages (Perlau et al., 1995), and knee braces (McNair et al., 1996). The heightened sensory input from tactile stimulation as a result of tightness may be a key factor in producing these effects (Mohd Sharif et al., 2017; Dzidotor et al., 2023). Therefore, knee sleeves may induce proprioceptive changes, subsequently modulating gait dynamics and improving the disrupted gait symmetry during fast walking.

The constraint imposed by the knee sleeves may also have contributed to stabilizing the gait dynamics. Walking with knee sleeves resulted in a significant reduction in the range of motion of the knee and hip joints, independent of walking speed. The variability in both peak hip flexion and peak ankle plantar flexion angles also decreased with the knee sleeves. These alterations can be attributed to the compressive effect imposed on the knees by the sleeves, which restricts knee joint movement and influences other joint motions. This restriction seems to have reduced the range and variability of joint motion, which in turn resulted in the improvement of disrupted symmetry. Taken together, knee sleeves may offer a valuable approach for enhancing gait symmetry, potentially reducing the risk of falls during fast walking, possibly owing to their proprioceptive and/or constraining effects on the knee joint.

This study had some limitations. First, the participants were exclusively healthy older adults. Gait kinematics in older adults differ from those in other groups, such as healthy young adults (Chehab et al., 2017; Schloemer et al., 2017) and older adults who have experienced falls (Kobayashi et al., 2014; Inai et al., 2023). As the effect of knee sleeves on gait symmetry in other populations remains unknown, further research is required to investigate their impact across diverse demographics. Second, only one type of knee sleeve was used. It is crucial to recognize that different designs, materials, and tightness characteristics may influence gait symmetry and parameters. Therefore, future studies should explore whether various types of knee sleeves have different effects on these variables. Finally, HR and iHR are global measures and, therefore, cannot explain in which phase of the gait cycle the deviations from symmetry occur or why they happen. Further studies investigating time-series three-dimensional kinematics and kinetics are warranted to elucidate the underlying mechanisms affecting gait symmetry.

In conclusion, this study examined gait symmetry indices and spatiotemporal gait parameters during walking with and without knee sleeves at different walking speeds. Our results indicate that the use of knee sleeves enhances gait symmetry along the AP direction, particularly during fast walking, where symmetry disruption is more likely than when walking at a comfortable pace. Additionally, knee sleeves were found to reduce the variability in gait symmetry across different walking speeds, suggesting a stabilizing effect on gait dynamics. To the best of our knowledge, this is the first study to provide evidence supporting the efficacy of knee sleeves in improving gait symmetry. The use of knee sleeves could be a valuable option for restoring disrupted gait symmetry during fast walking, with potential implications for reducing the risk of falls.

Data availability statement

The datasets presented in this article are not readily available because our IRB approval does not include data sharing. Requests to access the datasets should be directed to TI, takuma.inai@aist.go.jp.

Ethics statement

The studies involving humans were approved by the Ethics Committee of National Institute of Advanced Industrial Science and Technology (AIST). The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

TI: Conceptualization, Data curation, Formal Analysis, Funding acquisition, Investigation, Methodology, Resources, Validation, Visualization, Writing–original draft, Writing–review and editing. SK: Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Validation, Visualization, Writing–original draft, Writing–review and editing. WT: Conceptualization, Funding acquisition, Investigation, Methodology, Resources, Validation, Visualization, Writing–original draft, Writing–review and editing. MF: Conceptualization, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Validation, Visualization, Writing–original draft, Writing–review and editing.

Funding

The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. This work was supported by the collaborative research fund provided by Kagawa Seamless Inc. through the Kagawa Prefecture AI advanced technology research and development support project subsidy. The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article, or the decision to submit it for publication.

Acknowledgments

The authors thank Yuko Okazaki, Yumi Nakagawa, and Hiromi Chujo for their technical assistance with the experiments.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The authors declare that this study received funding from Kagawa Seamless Inc. through the Kagawa Prefecture AI advanced technology research and development support project subsidy. The authors disclose that knee sleeves were provided free of charge by Kagawa Seamless Inc. for the purpose of this study.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fbioe.2024.1394314/full#supplementary-material

References

Bell, A. L., Brand, R. A., and Pedersen, D. R. (1987). Prediction of hip joint center location from external landmarks. J. Biomech. 20, 913. doi:10.1016/0021-9290(87)90226-0

CrossRef Full Text | Google Scholar

Bell, A. L., Pedersen, D. R., and Brand, R. A. (1990). A comparison of the accuracy of several hip center location prediction methods. J. Biomech. 23, 617–621. doi:10.1016/0021-9290(90)90054-7

PubMed Abstract | CrossRef Full Text | Google Scholar

Birmingham, T. B., Kramer, J. F., Inglis, J. T., Mooney, C. A., Murray, L. J., Fowler, P. J., et al. (1998). Effect of a neoprene sleeve on knee joint position sense during sitting open kinetic chain and supine closed kinetic chain tests. Am. J. Sports Med. 26, 562–566. doi:10.1177/03635465980260041601

PubMed Abstract | CrossRef Full Text | Google Scholar

Bisi, M. C., Riva, F., and Stagni, R. (2014). Measures of gait stability: performance on adults and toddlers at the beginning of independent walking. J. Neuroeng Rehabil. 11, 131. doi:10.1186/1743-0003-11-131

PubMed Abstract | CrossRef Full Text | Google Scholar

Bisi, M. C., and Stagni, R. (2016). Complexity of human gait pattern at different ages assessed using multiscale entropy: from development to decline. Gait Posture 47, 37–42. doi:10.1016/j.gaitpost.2016.04.001

PubMed Abstract | CrossRef Full Text | Google Scholar

Brodie, M. A. D., Menz, H. B., Smith, S. T., Delbaere, K., and Lord, S. R. (2015). Good lateral harmonic stability combined with adequate gait speed is required for low fall risk in older people. Gerontology 61, 69–78. doi:10.1159/000362836

PubMed Abstract | CrossRef Full Text | Google Scholar

Bryk, F. F., de Jesus, J. F., Fukuda, T. Y., Moreira, E. G., Marcondes, F. B., and dos Santos, M. G. (2011). Efeito imediato da utilização da joelheira elástica em indivíduos com osteoartrite. Rev. Bras. Reumatol. 51, 440–446. doi:10.1590/s0482-50042011000500004

PubMed Abstract | CrossRef Full Text | Google Scholar

Castiglia, S. F., Trabassi, D., De Icco, R., Tatarelli, A., Avenali, M., Corrado, M., et al. (2022). Harmonic ratio is the most responsive trunk-acceleration derived gait index to rehabilitation in people with Parkinson’s disease at moderate disease stages. Gait Posture 97, 152–158. doi:10.1016/j.gaitpost.2022.07.235

PubMed Abstract | CrossRef Full Text | Google Scholar

Chang, R., Van Emmerik, R., and Hamill, J. (2008). Quantifying rearfoot-forefoot coordination in human walking. J. Biomech. 41, 3101–3105. doi:10.1016/j.jbiomech.2008.07.024

PubMed Abstract | CrossRef Full Text | Google Scholar

Chehab, E. F., Andriacchi, T. P., and Favre, J. (2017). Speed, age, sex, and body mass index provide a rigorous basis for comparing the kinematic and kinetic profiles of the lower extremity during walking. J. Biomech. 58, 11–20. doi:10.1016/j.jbiomech.2017.04.014

PubMed Abstract | CrossRef Full Text | Google Scholar

Cohen, J. (1992). A power primer. Psychol. Bull. 112, 155–159. doi:10.1037//0033-2909.112.1.155

PubMed Abstract | CrossRef Full Text | Google Scholar

Collins, A., Blackburn, J. T., Olcott, C., Yu, B., and Weinhold, P. (2011). The impact of stochastic resonance electrical stimulation and knee sleeve on impulsive loading and muscle co-contraction during gait in knee osteoarthritis. Clin. Biomech. (Bristol, Avon) 26, 853–858. doi:10.1016/j.clinbiomech.2011.04.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Collins, A., Blackburn, T., Olcott, C., Jordan, J. M., Yu, B., and Weinhold, P. (2014). A kinetic and kinematic analysis of the effect of stochastic resonance electrical stimulation and knee sleeve during gait in osteoarthritis of the knee. J. Appl. Biomech. 30, 104–112. doi:10.1123/jab.2012-0257

PubMed Abstract | CrossRef Full Text | Google Scholar

Conway, Z. J., Silburn, P. A., Perera, T., O’Maley, K., and Cole, M. H. (2021). Low-frequency STN-DBS provides acute gait improvements in Parkinson’s disease: a double-blinded randomised cross-over feasibility trial. J. Neuroeng Rehabil. 18, 125. doi:10.1186/s12984-021-00921-4

PubMed Abstract | CrossRef Full Text | Google Scholar

Cudejko, T., van der Esch, M., van der Leeden, M., Roorda, L. D., Pallari, J., Bennell, K. L., et al. (2018). Effect of soft braces on pain and physical function in patients with knee osteoarthritis: systematic review with meta-analyses. Arch. Phys. Med. Rehabil. 99, 153–163. doi:10.1016/j.apmr.2017.04.029

PubMed Abstract | CrossRef Full Text | Google Scholar

Delp, S. L., Anderson, F. C., Arnold, A. S., Loan, P., Habib, A., John, C. T., et al. (2007). Opensim: open-source software to create and analyze dynamic simulations of movement. IEEE Trans. Biomed. Eng. 54, 1940–1950. doi:10.1109/TBME.2007.901024

PubMed Abstract | CrossRef Full Text | Google Scholar

Doi, T., Hirata, S., Ono, R., Tsutsumimoto, K., Misu, S., and Ando, H. (2013). The harmonic ratio of trunk acceleration predicts falling among older people: results of a 1-year prospective study. J. Neuroeng Rehabil. 10, 7. doi:10.1186/1743-0003-10-7

PubMed Abstract | CrossRef Full Text | Google Scholar

Doi, T., Shimada, H., Makizako, H., Tsutsumimoto, K., Hotta, R., Nakakubo, S., et al. (2015). Effects of white matter lesions on trunk stability during dual-task walking among older adults with mild cognitive impairment. Age (Dordr) 37, 120. doi:10.1007/s11357-015-9858-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Dzidotor, G. K., Moorhead, J. B., Ude, C. C., Ogueri, K. S., Ghosh, D., and Laurencin, C. T. (2023). Functions and effectiveness of unloader, patellofemoral, and knee sleeve orthoses: a review. Regen. Eng. Transl. Med. doi:10.1007/s40883-023-00313-1

CrossRef Full Text | Google Scholar

Herrington, L., Simmonds, C., and Hatcher, J. (2005). The effect of a neoprene sleeve on knee joint position sense. Res. Sports Med. 13, 37–46. doi:10.1080/15438620590922077

PubMed Abstract | CrossRef Full Text | Google Scholar

Howcroft, J., Kofman, J., Lemaire, E. D., and McIlroy, W. E. (2016). Analysis of dual-task elderly gait in fallers and non-fallers using wearable sensors. J. Biomech. 49, 992–1001. doi:10.1016/j.jbiomech.2016.01.015

PubMed Abstract | CrossRef Full Text | Google Scholar

Igarashi, T., Tani, Y., Takeda, R., and Asakura, T. (2023). Relationship between gait regularity and harmony, and gait speed at discharge in inpatients with subacute stroke. J. Phys. Ther. Sci. 35, 40–45. doi:10.1589/jpts.35.40

PubMed Abstract | CrossRef Full Text | Google Scholar

Iijima, H., Eguchi, R., Aoyama, T., and Takahashi, M. (2019). Trunk movement asymmetry associated with pain, disability, and quadriceps strength asymmetry in individuals with knee osteoarthritis: a cross-sectional study. Osteoarthr. Cartil. 27, 248–256. doi:10.1016/j.joca.2018.10.012

PubMed Abstract | CrossRef Full Text | Google Scholar

Inai, T., Kobayashi, Y., Huang, C., Fujita, K., Fujimoto, M., Nihey, F., et al. (2023). Identification of characteristics of foot position and angle during swing phase in fallers using principal component analysis. Front. Bioeng. Biotechnol. 11, 1117884. doi:10.3389/fbioe.2023.1117884

PubMed Abstract | CrossRef Full Text | Google Scholar

Inai, T., Takabayashi, T., Edama, M., and Kubo, M. (2021). Effects of step length and cadence on hip moment impulse in the frontal plane during the stance phase. PeerJ 9, e11870. doi:10.7717/peerj.11870

PubMed Abstract | CrossRef Full Text | Google Scholar

Jung, D., Kim, J., Kim, M., Won, C. W., and Mun, K.-R. (2021). Frailty assessment using temporal gait characteristics and a long short-term memory network. IEEE J. Biomed. Health Inf. 25, 3649–3658. doi:10.1109/JBHI.2021.3067931

CrossRef Full Text | Google Scholar

Ko, C.-Y., Chang, Y., Jeong, B., Kang, S., Ryu, J., and Kim, G. (2017). Effects of knee sleeves on coordination of lower-limb segments in healthy adults during level walking and one-leg hopping. PeerJ 5, e3340. doi:10.7717/peerj.3340

PubMed Abstract | CrossRef Full Text | Google Scholar

Kobayashi, Y., Hida, N., Nakajima, K., Fujimoto, M., and Mochimaru, M. (2019). AIST gait database 2019. Available at: https://unit.aist.go.jp/harc/ExPART/GDB2019.html.

Google Scholar

Kobayashi, Y., Hobara, H., Matsushita, S., and Mochimaru, M. (2014). Key joint kinematic characteristics of the gait of fallers identified by principal component analysis. J. Biomech. 47, 2424–2429. doi:10.1016/j.jbiomech.2014.04.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Latt, M. D., Menz, H. B., Fung, V. S., and Lord, S. R. (2008). Walking speed, cadence and step length are selected to optimize the stability of head and pelvis accelerations. Exp. Brain Res. 184, 201–209. doi:10.1007/s00221-007-1094-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Leban, B., Cimolin, V., Porta, M., Arippa, F., Pilloni, G., Galli, M., et al. (2020). Age-related changes in smoothness of gait of healthy children and early adolescents. J. Mot. Behav. 52, 694–702. doi:10.1080/00222895.2019.1680949

PubMed Abstract | CrossRef Full Text | Google Scholar

Lindemann, U., Oksa, J., Skelton, D. A., Beyer, N., Klenk, J., Zscheile, J., et al. (2014). Effect of cold indoor environment on physical performance of older women living in the community. Age Ageing 43, 571–575. doi:10.1093/ageing/afu057

PubMed Abstract | CrossRef Full Text | Google Scholar

Lowry, K. A., Lokenvitz, N., and Smiley-Oyen, A. L. (2012). Age- and speed-related differences in harmonic ratios during walking. Gait Posture 35, 272–276. doi:10.1016/j.gaitpost.2011.09.019

PubMed Abstract | CrossRef Full Text | Google Scholar

McNair, P. J., Stanley, S. N., and Strauss, G. R. (1996). Knee bracing: effects on proprioception. Arch. Phys. Med. Rehabil. 77, 287–289. doi:10.1016/s0003-9993(96)90114-8

PubMed Abstract | CrossRef Full Text | Google Scholar

Menz, H. B., Lord, S. R., and Fitzpatrick, R. C. (2003a). Acceleration patterns of the head and pelvis when walking are associated with risk of falling in community-dwelling older people. J. Gerontol. A Biol. Sci. Med. Sci. 58, M446–M452. doi:10.1093/gerona/58.5.m446

PubMed Abstract | CrossRef Full Text | Google Scholar

Menz, H. B., Lord, S. R., and Fitzpatrick, R. C. (2003b). Acceleration patterns of the head and pelvis when walking on level and irregular surfaces. Gait Posture 18, 35–46. doi:10.1016/s0966-6362(02)00159-5

PubMed Abstract | CrossRef Full Text | Google Scholar

Misu, S., Asai, T., Doi, T., Sawa, R., Ueda, Y., Saito, T., et al. (2017). Association between gait abnormality and malnutrition in a community-dwelling elderly population. Geriatr. Gerontol. Int. 17, 1155–1160. doi:10.1111/ggi.12839

PubMed Abstract | CrossRef Full Text | Google Scholar

Mohd Sharif, N. A., Goh, S.-L., Usman, J., and Wan Safwani, W. K. Z. (2017). Biomechanical and functional efficacy of knee sleeves: a literature review. Phys. Ther. Sport 28, 44–52. doi:10.1016/j.ptsp.2017.05.001

PubMed Abstract | CrossRef Full Text | Google Scholar

Mohd Sharif, N. A., Usman, J., Wan Safwani, W. K. Z., Siew Li, G., Abdul Karim, S., Mohamed, N. A., et al. (2019). Effects of simple knee sleeves on pain and knee adduction moment in early unilateral knee osteoarthritis. Proc. Inst. Mech. Eng. H. 233, 1132–1140. doi:10.1177/0954411919874614

PubMed Abstract | CrossRef Full Text | Google Scholar

O’Connor, C. M., Thorpe, S. K., O’Malley, M. J., and Vaughan, C. L. (2007). Automatic detection of gait events using kinematic data. Gait Posture 25, 469–474. doi:10.1016/j.gaitpost.2006.05.016

PubMed Abstract | CrossRef Full Text | Google Scholar

Pasciuto, I., Bergamini, E., Iosa, M., Vannozzi, G., and Cappozzo, A. (2017). Overcoming the limitations of the Harmonic Ratio for the reliable assessment of gait symmetry. J. Biomech. 53, 84–89. doi:10.1016/j.jbiomech.2017.01.005

PubMed Abstract | CrossRef Full Text | Google Scholar

Pau, M., Mandaresu, S., Pilloni, G., Porta, M., Coghe, G., Marrosu, M. G., et al. (2017). Smoothness of gait detects early alterations of walking in persons with multiple sclerosis without disability. Gait Posture 58, 307–309. doi:10.1016/j.gaitpost.2017.08.023

PubMed Abstract | CrossRef Full Text | Google Scholar

Pau, M., Mulas, I., Putzu, V., Asoni, G., Viale, D., Mameli, I., et al. (2020). Smoothness of gait in healthy and cognitively impaired individuals: a study on Italian elderly using wearable inertial sensor. Sensors (Basel) 20, 3577. doi:10.3390/s20123577

PubMed Abstract | CrossRef Full Text | Google Scholar

Peng, H.-T., Lin, C.-H., Kuo, Y.-C., and Song, C.-Y. (2020). Effects of arch support insoles on single- and dual-task gait performance among community-dwelling older adults. Clin. Interv. Aging 15, 1325–1332. doi:10.2147/CIA.S254474

PubMed Abstract | CrossRef Full Text | Google Scholar

Perlau, R., Frank, C., and Fick, G. (1995). The effect of elastic bandages on human knee proprioception in the uninjured population. Am. J. Sports Med. 23, 251–255. doi:10.1177/036354659502300221

PubMed Abstract | CrossRef Full Text | Google Scholar

Quach, L., Galica, A. M., Jones, R. N., Procter-Gray, E., Manor, B., Hannan, M. T., et al. (2011). The nonlinear relationship between gait speed and falls: the maintenance of balance, independent living, intellect, and zest in the elderly of Boston study: THE NON-linear relationship between gait speed and falls. J. Am. Geriatr. Soc. 59, 1069–1073. doi:10.1111/j.1532-5415.2011.03408.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Robertson, D. G. E., Caldwell, G. E., Hamill, J., Kamen, G., and Whittlesey, S. N. (2008). Research Methods in Biomechanics. 2nd Edn. Champaign,ill: Human Kinetics.

Google Scholar

Roche, J. L., Lowry, K. A., Vanswearingen, J. M., Brach, J. S., and Redfern, M. S. (2013). Harmonic Ratios: a quantification of step to step symmetry. J. Biomech. 46, 828–831. doi:10.1016/j.jbiomech.2012.12.008

PubMed Abstract | CrossRef Full Text | Google Scholar

Schloemer, S. A., Thompson, J. A., Silder, A., Thelen, D. G., and Siston, R. A. (2017). Age-related differences in gait kinematics, kinetics, and muscle function: a principal component analysis. Ann. Biomed. Eng. 45, 695–710. doi:10.1007/s10439-016-1713-4

PubMed Abstract | CrossRef Full Text | Google Scholar

Schmitz, A., Ye, M., Shapiro, R., Yang, R., and Noehren, B. (2014). Accuracy and repeatability of joint angles measured using a single camera markerless motion capture system. J. Biomech. 47, 587–591. doi:10.1016/j.jbiomech.2013.11.031

PubMed Abstract | CrossRef Full Text | Google Scholar

Schween, R., Gehring, D., and Gollhofer, A. (2015). Immediate effects of an elastic knee sleeve on frontal plane gait biomechanics in knee osteoarthritis. PLoS One 10, e0115782. doi:10.1371/journal.pone.0115782

PubMed Abstract | CrossRef Full Text | Google Scholar

Seth, A., Hicks, J. L., Uchida, T. K., Habib, A., Dembia, C. L., Dunne, J. J., et al. (2018). OpenSim: simulating musculoskeletal dynamics and neuromuscular control to study human and animal movement. PLoS Comput. Biol. 14, e1006223. doi:10.1371/journal.pcbi.1006223

PubMed Abstract | CrossRef Full Text | Google Scholar

Shi, X., Wang, Z., Zhao, H., Qiu, S., Liu, R., Lin, F., et al. (2023). Threshold-free phase segmentation and zero velocity detection for gait analysis using foot-mounted inertial sensors. IEE Trans. Hum. Mach. Syst. 53, 176–186. doi:10.1109/THMS.2022.3228515

CrossRef Full Text | Google Scholar

Sole, G., Lamb, P., Pataky, T., Pathak, A., Klima, S., Navarre, P., et al. (2022). Immediate and six-week effects of wearing a knee sleeve following anterior cruciate ligament reconstruction on knee kinematics and kinetics: a cross-over laboratory and randomised clinical trial. BMC Musculoskelet. Disord. 23, 560. doi:10.1186/s12891-022-05488-2

PubMed Abstract | CrossRef Full Text | Google Scholar

Ullauri, J. B., Akiyama, Y., Okamoto, S., and Yamada, Y. (2019). Technique to reduce the minimum toe clearance of young adults during walking to simulate the risk of tripping of the elderly. PLoS One 14, e0217336. doi:10.1371/journal.pone.0217336

PubMed Abstract | CrossRef Full Text | Google Scholar

Wada, O., Asai, T., Hiyama, Y., Nitta, S., and Mizuno, K. (2019). Gait variability in women with hip osteoarthritis before and after total hip replacement: a prospective cohort study. Am. J. Phys. Med. Rehabil. 98, 866–871. doi:10.1097/PHM.0000000000001206

PubMed Abstract | CrossRef Full Text | Google Scholar

Yamagata, M., Tateuchi, H., Shimizu, I., and Ichihashi, N. (2019). The effects of fall history on kinematic synergy during walking. J. Biomech. 82, 204–210. doi:10.1016/j.jbiomech.2018.10.032

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, L., Yan, Y., Liu, G., Han, B., Fei, J., and Zhang, Y. (2022). Effect of fatigue on kinematics, kinetics and muscle activities of lower limbs during gait. Proc. Inst. Mech. Eng. H. 236, 1365–1374. doi:10.1177/09544119221112516

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: walking, harmonic ratio, acceleration, inertial measurement unit, motion capture

Citation: Inai T, Kudo S, Tsuchida W and Fujimoto M (2024) Knee sleeves improve gait symmetry during fast walking in older adults. Front. Bioeng. Biotechnol. 12:1394314. doi: 10.3389/fbioe.2024.1394314

Received: 01 March 2024; Accepted: 19 June 2024;
Published: 17 July 2024.

Edited by:

Naomichi Ogihara, The University of Tokyo, Japan

Reviewed by:

SungHyek Kim, Tokoha University, Japan
Prasanna Sritharan, The University of Melbourne, Australia

Copyright © 2024 Inai, Kudo, Tsuchida and Fujimoto. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Masahiro Fujimoto, masahiro-fujimoto@aist.go.jp

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.