Skip to main content

ORIGINAL RESEARCH article

Front. Bioeng. Biotechnol., 13 March 2023
Sec. Nanobiotechnology
This article is part of the Research Topic Terahertz Technologies for Biosensing and Biomedical Analysis View all 11 articles

Terahertz time-domain attenuated total reflection spectroscopy integrated with a microfluidic chip

Ying Fu,Ying Fu1,2Tunan ChenTunan Chen3Ligang Chen,Ligang Chen1,2Yuansen Guo,Yuansen Guo1,2Zhongbo Yang,Zhongbo Yang1,2Ning MuNing Mu3Hua FengHua Feng3Mingkun Zhang,
Mingkun Zhang1,2*Huabin Wang,
Huabin Wang1,2*
  • 1Center of Super-Resolution Optics & Chongqing Engineering Research Center of High-Resolution and Three-Dimensional Dynamic Imaging Technology, Chongqing Institute of Green and Intelligent Technology, Chinese Academy of Sciences, Chongqing, China
  • 2Chongqing School, University of Chinese Academy of Sciences, Chongqing, China
  • 3Department of Neurosurgery and Key Laboratory of Neurotrauma, Southwest Hospital, Third Military Medical University (Army Medical University), Chongqing, China

The integration of a microfluidic chip into terahertz time-domain attenuated total reflection (THz TD-ATR) spectroscopy is highly demanded for the accurate measurement of aqueous samples. Hitherto, however little work has been reported on this regard. Here, we demonstrate a strategy of fabricating a polydimethylsiloxane microfluidic chip (M-chip) suitable for the measurement of aqueous samples, and investigate the effects of its configuration, particularly the cavity depth of the M-chip on THz spectra. By measuring pure water, we find that the Fresnel formulae of two-interface model should be applied to analyze the THz spectral data when the depth is smaller than 210 μm, but the Fresnel formula of one-interface model can be applied when the depth is no less than 210 μm. We further validate this by measuring physiological solution and protein solution. This work can help promote the application of THz TD-ATR spectroscopy in the study of aqueous biological samples.

1 Introduction

Terahertz (THz) wave normally refers to the electromagnetic radiation with a frequency ranging from 0.1 THz to 10.0 THz, which is located between microwave and infrared regions (Wang et al., 2020a; Li et al., 2020; Yang et al., 2020; Meng et al., 2023). In the recent 2 decades, THz biological studies have attracted much attention among scientific communities, mainly thanks to the broad-band, label-free, and bio-sensitive features of THz wave (Zaytsev et al., 2015; Zhou et al., 2016; Sun et al., 2018; Bai et al., 2020; Chen et al., 2020; Zhu et al., 2020; Shen J. et al., 2021; Shen S. et al., 2021; Zhang et al., 2022b; Tang et al., 2022; Zhang et al., 2023). THz time-domain attenuated total reflection (THz TD-ATR) spectroscopy is an advanced THz technique, and superior to many conventional THz techniques due to its remarkable sensitivity by taking the advantage of evanescent-field detection (Arikawa et al., 2008; Wang et al., 2019; Hu et al., 2021; Liao et al., 2022). Moreover, THz TD-ATR spectroscopy enables the minimization of THz absorption by water when measuring aqueous samples. As a result, THz TD-ATR spectroscopy is preferred by the researchers probing aqueous biological samples where high detection sensitivity and minimized water absorption are urgently required (Baxter and Guglietta, 2011; McIntosh et al., 2012; Qin et al., 2017; Peng et al., 2021). Employing this technique, various aqueous biological samples such as protein, DNA, amino acids, and saccharide have been investigated, which greatly promoted the development and application of THz TD-ATR spectroscopy in biological studies (Qin et al., 2013; Yang et al., 2016; Wang et al., 2020b; Tang et al., 2020).

THz TD-ATR spectroscopy includes two core parts, i.e., a THz TD spectroscopy system and an ATR prism incorporated into the THz optical path to allow in- and out-coupling of a THz beam. Normally, the prim is made of high-resistivity silicon, which is an almost perfectly transparent, non-dispersive and high-index material in the THz band. In a typical THz TD-ATR spectroscopy experiment, a sample is deposited onto the prism base, and the THz beam incident at a certain angle experiences total internal reflection at the prism-sample interface, producing an evanescent field attenuating exponentially inside the sample along the normal direction of the sample-prism boundary (Nagai et al., 2006; Peng et al., 2020). By detecting and analyzing the reflected detected THz beam, useful information on the measured sample is extracted. To obtain a reliable result, the sample thickness is required to be no less than the penetration depth of the evanescent field which is dependent on the THz frequency and the sample refractive index (Averett et al., 2008; Gotz et al., 2020; Mendoza-Galvan et al., 2021). To meet this requirement, a common practice in the study of aqueous sample is depositing a large amount of sample solution directly onto the ATR prism base or into an open liquid cell situated on the ATR prism base to guarantee the ATR reflection surface to be completely submerged and the thickness of the solution is high enough to prevent the penetration of THz evanescent field (Gu et al., 2019). Although this operation may work in some cases, it has at least two issues: 1) the inevitable solvent evaporation causes changes in the sample concentration, which compromises the accuracy of results; and 2) it needs relatively a large amount of sample, which is not allowed for rare or precious samples. The integration of a microfluidic chip (M-chip) into THz TD-ATR spectroscopy is a promising approach for tackling the above-mentioned issues, simultaneously (Karabudak, 2013). Although a few groups reported the application of M-chips in transmission mode THz spectroscopy (Xu et al., 2017; Gong et al., 2019), little work has been conducted to fabricate suitable M-chips for THz TD-ATR spectroscopy. Usually, M-chips used for transmission mode THz spectroscopy are not applicable for THz TD-ATR spectroscopy, and the mechanisms underlying the interactions between THz wave and M-chips are different for the 2 THz techniques.

In the present work, we fabricate a polydimethylsiloxane (PDMS) M-chip, integrate it into THz TD-ATR spectroscopy to form a THz TD-ATR microfluidic system, and investigate the configuration, particularly the influence of the depth of M-chip cavity on THz spectra. By measuring pure water, we find that the Fresnel formula of two-interface model should be used to extract aqueous sample properties when the cavity depth is smaller than 210 μm, but a simple Fresnel formula of one-interface model can be applied when the cavity depth is no less than 210 μm. This observation is further confirmed by measuring physiological solution and protein solution. Collectively, we demonstrate that the THz-TD ATR microfluidic system can serve as a useful platform for the study of aqueous biological samples.

2 Results and discussion

2.1 Configuration of microfluidic chips

PDMS is a popular material for fabricating microfluidic devices due to its advantages such as highly biocompatible, chemical inert, impermeable to water, inexpensive, flexible, and easy to be fabricated and bonded to other surfaces. PDMS M-chips were fabricated via injection molding (Sia and Whitesides, 2003) (Notes S1 and S2 of the Supplementary Material). Figure 1A is a three-dimensional (3D) schematic view of a fabricated M-chip, including an inlet, an outlet, a cavity with an elliptic cylinder shape, and two channels tangential to the wall of the cavity. The M-chip has upper wall and side wall, but no bottom (Figure 1B). This design ensures an aqueous sample in the cavity to contact with the ATR prism base without any barriers, enabling the sample to be directly probed with the evanescent filed generated by the total internal reflection occurring at the sample-prism interface (Figure 1C). Before a THz TD-ATR measurement, the M-chip was aligned carefully to make the cavity to fully cover the elliptical THz beam spot, and securely attached and sealed on the prism base by applying an even loading. A syringe or pump was employed to infuse a sample solution through a soft tube into the inlet by a pressure-driven mechanism, and the solution passed through the channel connected with the inlet to fill the cavity. The excess sample solution and air bubbles were drained out through the outlet.

FIGURE 1
www.frontiersin.org

FIGURE 1. Configuration of a M-chip. (A) A 3D schematic view of a M-chip, in which the inlet “1” and outlet “2” holes, channels “3”, and cavity “4” are connected. The diameter of the holes is 1 mm. The cavity is in a shape of elliptic cylinder, with a major axis length of 33 mm, and a minor axis length of 16 mm. The distances between the holes and the chip outside edges are 3.5 mm along the major axis direction and 4.5 mm along the minor axis direction of the cavity, respectively. (B) Side view of a M-chip situated on an ATR prism. The depth of channels is the same as the cavity depth (d), which is a variable. The depth of the holes is the summation of the cavity depth and the thickness (2 cm) of the upper wall. (C) A 3D schematic view of the ATR apparatus integrated with a M-chip. A p-polarized THz beam is incident horizontally to the left slanted side of the ATR prism, an isosceles triangulated column, then refracted and incident to the prism base at which the attenuated total internal reflection occurs. The emergent beam is detected for analysis.

2.2 Analytical models for THz TD-ATR spectroscopy integrated with a M-chip

For proper application of the THz TD-ATR spectroscopy integrated with a PDMS M-chip, namely, the THz TD-ATR microfluidics system, we have to carefully consider the possible influence of the configuration of the M-chip on THz spectra. It is a non-trivial work to extract the true THz spectra of an aqueous sample filled in the cavity if the side wall of the M-chip interacts with the evanescent field. Fortunately, for a certain THz TD system, the cross section of the evanescent filed at the sample-prism interface can be estimated. In our case, the diameter of the THz beam is about 12 mm, yielding an elliptical beam spot on the prism base with major axis length of ∼31 mm (12/sin2 38.4°) and minor axis length of ∼12 mm. Thus, we designed a M-chip with a cavity cross section larger than that of the evanescent field. With proper alignment of the M-chip, we ensured only the cavity region overlaid on the evanescent field, by which the influence of the side wall was avoided (Figure 2A). Although the effects of the side wall can be eliminated with this design, the influence of the upper wall of the M-chip has to be taken into account, i.e., the THz reflection occurred at the interface formed by the sample and PDMS upper wall needs to be considered, in order to extract reliable THz spectral data of the sample measured by the THz TD-ATR microfluidics system.

FIGURE 2
www.frontiersin.org

FIGURE 2. Schematic illustration of the interaction mechanism between the THz evanescent field and sample and M-chip. (A) THz evanescent field (in orange color) occurs at the prism-substance interface, penetrates the cavity, and enters the upper wall of the M-chip. The angle of ATR prism slant side is 38.4°. The substance in the cavity of the M-chip can be either air (i.e., reference) or aqueous sample. A prism-substance interface and a substance-PDMS upper-wall interface are formed. As the side wall of the M-chip does not influence the analysis in our case, it is not shown for simplicity. (B) r12 and r23 are the Fresnel’s reflection coefficients of prism-substance interface and the substance-PDMS upper-wall interface, respectively. εw, ε, εSi denote the complex dielectric constants of the PDMS, substance in the cavity, and silicon prism, respectively. Ein and Eout are the incident and emergent beams, respectively, and r is the ultimate Fresnel’s reflection coefficient of the two-interface system. θ is the incident angle of the THz beam in the prism.

The sample complex dielectric constant is key for analyzing the sample properties. For our THz TD-ATR microfluidic system, the Fresnel’s formulae for two-interface model are used to extract the complex dielectric constant of the sample for general cases (Møller et al., 2007),

r12=εSi1εSi/εsin2θεcosθεSi1εSi/εsin2θ+εcosθ(1)
r23=ε1εSi/εwsin2θεw1εSi/εsin2θε1εSi/εwsin2θ+εw1εSi/εsin2θ(2)

where r12 and r23 are the Fresnel’s reflection coefficients of prism-substance interface and the substance-PDMS upper-wall interface, respectively; εw, ε, and εSi are the complex dielectric constants of the PDMS, substance filled in the cavity and silicon prism, respectively; and θ is the incident angle of the THz beam in the prism.

For a cavity with a depth of d and an incident wave with a wavelength of λ, the ultimate Fresnel’s reflection coefficient (r) that contains the information of the substance in the cavity and PDMS upper-wall due to their interactions with the evanescent wave is calculated by (McIntyre and Aspnes, 1971)

r=r12+r23expi4πdλεεSisin2θ1+r12r23expi4πdλεεSisin2θ(3)

It needs to point out that if the evanescent field cannot penetrate the cavity to reach the upper wall of the M-chip, r23 is zero, thus r12 is equal to r. In this case, the two-interface model is degenerated to the one-interface model.

In the experiments, the M-chip without and with an aqueous sample was measured in tandem to obtain the information of the reference (i.e., air) and the sample, respectively. The above three equations are applicable to both the reference and the sample. By substituting Eqs 1, 2 into Eq. 3, the ultimate Fresnel’s reflection coefficients of the reference (rref) and the sample (rsam) were obtained, respectively. Furthermore, the relationship between the ultimate reflected electromagnetic field amplitudes of the sample (Esam) and reference (Eref) from the prism-substance interface and their corresponding reflection coefficients can be described by Eq. 4,

EsamEref=rsamrref(4)

In the above equations, εSi (3.42), d; λ, θ (51.6°) and the dielectric constant of air (1) are all known, and εw was measured using the THz TD spectroscopy in transmission mode (Supplementary Figure S3). With the above information, the complex dielectric constant of the measured sample can be calculated.

THz absorption coefficient (α) is also often used to describe the properties of samples, which is calculated according to the following equations (Jepsen et al., 2011; Zhang et al., 2022a),

εν=ενiεν(5)
εν=n2νk2ν(6)
εν=2nνkν(7)
αν=4πνkνc(8)

where ε and ε are the real and imaginary part of ε, respectively; ν is the frequency; n and k are the refractive index and extinction coefficient, respectively; and c is the speed of light in vacuum.

2.3 Dielectric measurement of pure water using the THz TD-ATR microfluidic system

Pure water is commonly used as the solvent for many biological samples. Therefore, we measured the THz spectra of pure water (Milli-Q water, 18.2 MΩ cm) using our fabricated M-chips with various cavity depths. By applying the two-interface model, we can see that the complex dielectric constants of pure water measured in the M-chips of various depths are highly consistent with each other (Figures 3A,B). They are also highly consistent with those measured without using the M-chips, for pure bulk water through which the THz evanescent field cannot penetrate. The results tell us that the two-interface model is accurately enough to be used to extract the THz properties of pure water in the cavity of M-Chips.

FIGURE 3
www.frontiersin.org

FIGURE 3. THz spectra of pure water measured using the THz TD-ATR microfluidic system. (A–B) Real and imaginary parts of the complex dielectric constants of pure water filled in the M-chips with various depths (90, 130, 170, and 210 μm) and the infinite thickness relative to the penetration depth of the evanescent waves, denoted by “∞”, which were extracted by the two-interface model. (C–D) These parameters were extracted by the one-interface model. The semitransparent homochromic shadows around the lines represent the standard deviation (SD) of three independent measurements.

In contrast, the complex dielectric constants are not necessarily consistent with each other when the one-interface model was applied to analyze the data. Although the complex dielectric constant of pure water measured in the cavity with a depth of 210 μm is consistent with that of pure bulk water, the parameters extracted for the pure water in the cavities with depths of 90, 130, and 170 μm deviate from those of pure bulk water at the lower THz frequency range (Figures 3C,D). These observations indicate that the evanescent field penetrated the pure water with a thickness less than 210 μm, and reached the upper wall of the PDMS M-chip. Obviously, it is inappropriate to apply the one-interface model when the cavity depth is less than 210 μm because the one interface model is only suitable for the situation where the evanescent field does not penetrate the pure water. The influence of the cavity depth on the detected THz wave was further explored by analyzing the depth-related term contained in Eq. 3 for the two-interface model, as shown below,

Φ=r23expi4πdλεεSisin2θ(9)

This term is associated with the cavity depth when the evanescent wave reaches the upper wall of the PDMS M-chip. However, this term is zero, so the two-interface is degenerated to the one-interface model when the evanescent wave cannot penetrate through the aqueous sample in the cavity to reach the upper wall of the PDMS M-chip.

Experimentally, indeed, we observed that both the real and imaginary parts of Φ show a cavity-depth and frequency dependent character, particularly for the lower frequency range (Figures 4A,B). Interestingly, it is found that both the real and imaginary parts of Φ for the pure water with a thickness of 210 μm are nearly the same as those of the pure bulk water for the frequency is higher than 0.3 THz. The result verifies that the two-interface model can be degenerated to the one-interface model when the pure water thickness is no less than 210 μm, and that the two-interface model should be applied for pure water with the thickness less than 210 μm. These observations well explain the phenomena observed for the dielectric complex constants in Figure 3. For a commonly used THz TD-ATR spectroscopy system, the THz signal below 0.3 THz is very noisy and often not included in the data analysis. Thus, we recommend to use a M-chip with a cavity depth no less than 210 μm in THz TD-ATR microfluidic system, by which one can employ the simple one-interface model to analyze the data with enough reliability while avoid the complexity of the two-interface model. Specifically, the one-interface model is also applicable for M-chips with cavity depths of 90 μm and above when only the frequency beyond ∼1.0 THz is considered in data analysis (Figure 4).

FIGURE 4
www.frontiersin.org

FIGURE 4. The real parts and imaginary parts of the depth-related term. (A) Real parts of the depth-related term Φ. (B) Imaginary parts of Φ. The cavity depth (pure water thickness) includes 90, 130, 170, and 210 μm “∞” indicates the thickness of pure bulk water, for which either the real part or the imaginary part of Φ is zero for the frequency range investigated.

2.4 Dielectric measurement of physiological solution using the THz TD-ATR microfluidic system

From the above results we know that the THz TD-ATR spectroscopy integrated with a M-chip with a cavity depth no less than 210 μm can be reliably applied to measure pure water by simply using the one-interface model in the data analysis. To test the wide applicability of the THz TD-ATR system of the above configuration, we measured the THz spectra of phosphate buffered saline (PBS) solution, which is a physiological solution widely used to dissolve biological samples. It is evident that the extracted complex dielectric constants obtained by the one-interface model are well superimposed on those retrieved by the two-interface model (Figure 5).

FIGURE 5
www.frontiersin.org

FIGURE 5. THz spectra of PBS measured by the THz TD-ATR microfluidic system. Real parts (A) and imaginary parts (B) of the complex dielectric constants of PBS were extracted by employing the one-interface model and the two-interface model, respectively. Translucent areas around each line indicate the standard deviation from three independent tests.

2.5 THz absorption measurement of protein solutions using the THz TD-ATR microfluidic system

One of our major purposes is to integrate a suitable M-chip into THz TD-ATR spectroscopy to study biological samples. Thus, we further measured lactic dehydrogenase (LDH) solutions with different concentrations using the PDMS M-chip with a cavity depth of 210 μm, and extracted their THz absorption coefficients which are widely used to assess the character of biological solutions. The THz coefficients were obtained by both the two-interface model and the one-interface model. For example, the THz coefficient of LDH solution with a concentration of 60 mg/mL plotted against a wide-band frequency was shown in Figure 6A. It is clear that for this M-chip configuration, the one-interface model is as accurate as the two-interface model for the measurement of LDH solutions. In addition, we also checked the THz absorption coefficients against the LDH concentrations at certain frequencies (e.g., 0.3, 0.4, and 0.5 THz; Figure 6B). Again, it confirms that there are no observable differences when applying the two models to analyze protein solutions measured by our THz TD-ATR microfluidic system. The THz absorption coefficient decreases with the increase of LDH concentration observed here is consistent with previous studies (Bye et al., 2014), which is very likely due to the solute (LDH) induced dilution and static depolarization effects (Penkov et al., 2017).

FIGURE 6
www.frontiersin.org

FIGURE 6. LDH solutions measured using the THz TD-ATR microfluidic system. (A) THz absorption coefficients of LDH solution (60 mg/mL) are plotted against the THz frequency, which were extracted using the two-interface model and one-interface model, respectively. (B) THz absorption coefficients of LDH solutions at specific THz frequencies are plotted against the LDH concentrations. The solution with a LDH concentration of 0 mg/mL is PBS solution. The error bars for the three independent experiments are too small to be clearly observed.

3 Conclusion

In this work, we demonstrated a simple and low-cost method of fabricating a M-chip for the measurement of a liquid phase sample with the THz TD-ATR. With the integration of the M-chip into the THz TD-ATR, the sample volume could be reduced to an unprecedentedly small volume as less as 40 μL. We also rationalized the application of theoretical models for analyzing the data obtained by the THz TD-ATR microfluidic system, which is critical for the proper application of the system. By measuring pure water, we found that the one-interface model can be safely applied in the data analysis when the cavity depth of M-chips is no less than 210 μm, but the two-interface model should be applied when the depth is less than 210 μm. Based on these results, we propose to use PDMS M-chips with a cavity depth of at least 210 μm in a THz TD-ATR microfluidic system when measuring readily available or cheap aqueous sample solutions to avoid the complexity of using the two-interface model in data analysis. Instead, PDMS M-chips with a cavity depth smaller than 210 μm and the two-interface model are suggested to be applied when measuring precious or expensive aqueous sample solutions in order to save the samples while keep the experimental reliability. With such a system, we successfully detected physiological (PBS) and protein (LDH) solutions. We also note that the theories shown here also hold for other solutions as far as the solutions are inert to PDMS, and that the threshold value of the cavity depth for applying the one-interface model may change, dependent on the performance of the THz TD-ATR and the refractive index of the measured solution (Liu et al., 2017). Collectively, the work presented here is helpful for researchers who exploit THz TD-ATR spectroscopy to investigate aqueous biological samples in a reliable and economical way.

4 Methods

4.1 Chemicals

Lactic dehydrogenase powder was purchased from Beijing Solarbio Science & Technology Co., Ltd. (Beijing, China), and dissolved in phosphate buffer saline (PBS, pH 7.4, Shanghai Branch of Thermo Fisher Scientific Inc. Shanghai, China) and diluted to desired concentrations with PBS. The prepolymer and the crosslinker of PDMS were ordered from the Shanghai Branch of Dow Corning Corp. (Shanghai, China).

4.2 Experimental setup

The THz-TD ATR spectroscopy system was established by incorporating an ATR apparatus (BATOP GmbH, Jena, Germany) into the optical path of a commercial THz-TDS system (Tera K15, Menlo Systems GmbH, Münich, Germany). A femtosecond laser with a center wavelength of 1,560 nm, a repetition rate of 100 MHz, and a pulse width less than 90 fs was split into the pump beam and the probe beam. The THz radiation was emitted from a biased photoconductive antenna irradiated by the pump beam, and detected by another photoconductive antenna with the aid of the probe beam. Before experiments, the microfluidic chip fabricated by us was integrated into the system for the measurement of aqueous samples. The whole system was located in a clean room maintained at a temperature of 22°C ± 1.0°C. In experiments, the apparatus including the THz emitter, ATR prism integrated with microfluidic chip and THz detector were enclosed in a sealed container filled with pure nitrogen and kept the humidity to less than 4% to minimize the vapor absorption of THz wave. A p-polarized pulsed THz beam was horizontally incident on the left slanted side of the prism, and the refracted beam was detected from the right side of the prism. The ATR prism and PDMS chip can be re-used after cleaning by ethanol and distilled water in order. Either a milliliter medical syringe (Minkang Medical Materials Co., Ltd. Changsha, China) or a pressure-based flow controller (MFCS-EZ, Fluigent, Paris, France) was used to infuse sample solution into an M-chip.

4.3 Data analysis

Time-domain spectra were measured and converted to frequency-domain data by the fast Fourier transform algorithm, from which the electric field amplitude at different frequencies can be obtained. Then the complex dielectric constant and absorption coefficient can be calculated according to Eqs 18. The data for each sample represents the average value obtained from three independent measurements.

4.4 Fabrication of the microfluidic chip

Microfluidic chips with different cavity depths were fabricated by using a standard protocol, and the details were described in the Supplementary Material.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.

Author contributions

MZ and HW conceived the academic idea, HW guided the experiments, YF, TC, LC, NM, YG, and MZ performed the experiments, YF, MZ, ZY, and HW analyzed the data, YF completed the drawing work, TC contributed with helpful discussions, YF and HW wrote the manuscript with revisions by all.

Funding

This work is supported by the National Natural Science Foundation of China (62175238), the National Key Research and Development Program of China (2021YFA1301503), the University of Chinese Academy of Sciences Supported Program for Tackling Key Problems in Science and Technology (E029610601), Natural Science Foundation of Chongqing (cstc2019jcyj-msxmX0051), and Youth Innovation Promotion Association CAS (2022390).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fbioe.2023.1143443/full#supplementary-material

References

Arikawa, T., Nagai, M., and Tanaka, K. (2008). Characterizing hydration state in solution using terahertz time-domain attenuated total reflection spectroscopy. Chem. Phys. Lett. 457 (1), 12–17. doi:10.1016/j.cplett.2008.03.062

CrossRef Full Text | Google Scholar

Averett, L. A., Griffiths, P. R., and Nishikida, K. (2008). Effective path length in attenuated total reflection spectroscopy. Anal. Chem. 80 (8), 3045–3049. doi:10.1021/ac7025892

PubMed Abstract | CrossRef Full Text | Google Scholar

Bai, Z., Liu, Y., Kong, R., Nie, T., Wen, L., Li, H., et al. (2020). Near-field terahertz sensing of Hela cells and pseudomonas based on monolithic integrated metamaterials with a spintronic terahertz emitter. ACS Appl. Mater. Interfaces 12 (32), 35895–35902. doi:10.1021/acsami.0c08543

PubMed Abstract | CrossRef Full Text | Google Scholar

Baxter, J. B., and Guglietta, G. W. (2011). Terahertz spectroscopy. Anal. Chem. 83 (12), 4342–4368. doi:10.1021/ac200907z

PubMed Abstract | CrossRef Full Text | Google Scholar

Bye, J. W., Meliga, S., Ferachou, D., Cinque, G., Zeitler, J. A., and Falconer, R. J. (2014). Analysis of the hydration water around bovine serum albumin using terahertz coherent synchrotron radiation. J. Phys. Chem. A 118 (1), 83–88. doi:10.1021/jp407410g

PubMed Abstract | CrossRef Full Text | Google Scholar

Chen, L., Ren, G., Liu, L., Guo, P., Wang, E., Zhou, L., et al. (2020). Terahertz signatures of hydrate formation in alkali halide solutions. J. Phys. Chem. Lett. 11 (17), 7146–7152. doi:10.1021/acs.jpclett.0c02046

PubMed Abstract | CrossRef Full Text | Google Scholar

Gong, A., Qiu, Y., Chen, X., Zhao, Z., Xia, L., and Shao, Y. (2019). Biomedical applications of terahertz technology. Appl. Spectrosc. Rev. 55 (5), 418–438. doi:10.1080/05704928.2019.1670202

CrossRef Full Text | Google Scholar

Gotz, A., Nikzad-Langerodi, R., Staedler, Y., Bellaire, A., and Saukel, J. (2020). Apparent penetration depth in attenuated total reflection Fourier-transform infrared (ATR-FTIR) spectroscopy of Allium cepa L. epidermis and cuticle. Spectrochim. Acta A 224, 117460. doi:10.1016/j.saa.2019.117460

PubMed Abstract | CrossRef Full Text | Google Scholar

Gu, Y., Li, S., Xu, Y., Han, J., Gu, M., Cai, Z., et al. (2019). The effect of magnetic field on the hydration of cation in solution revealed by THz spectroscopy and MDs. Colloid Surf. a-physicochem. Eng. Asp. 582, 123822. doi:10.1016/j.colsurfa.2019.123822

CrossRef Full Text | Google Scholar

Hu, M., Tang, M., Wang, H., Zhang, M., Zhu, S., Yang, Z., et al. (2021). Terahertz, infrared and Raman absorption spectra of tyrosine enantiomers and racemic compound. Spectrochim. Acta A 254, 119611. doi:10.1016/j.saa.2021.119611

CrossRef Full Text | Google Scholar

Jepsen, P. U., Cooke, D. G., and Koch, M. (2011). Terahertz spectroscopy and imaging - modern techniques and applications. Laser Photonics Rev. 5 (1), 124–166. doi:10.1002/lpor.201000011

CrossRef Full Text | Google Scholar

Karabudak, E. (2013). Micromachined silicon attenuated total reflectance infrared spectroscopy: An emerging detection method in micro/nanofluidics. Electrophoresis 35 (2-3), 236–244. doi:10.1002/elps.201300248

PubMed Abstract | CrossRef Full Text | Google Scholar

Li, Z., Yan, S., Zang, Z., Geng, G., Yang, Z., Li, J., et al. (2020). Single cell imaging with near-field terahertz scanning microscopy. Cell Prolif. 53 (4), e12788. doi:10.1111/cpr.12788

PubMed Abstract | CrossRef Full Text | Google Scholar

Liao, Y., Zhang, M., Tang, M., Chen, L., Li, X., Liu, Z., et al. (2022). Label-free study on the effect of a bioactive constituent on glioma cells in vitro using terahertz ATR spectroscopy. Biomed. Opt. Exp. 13 (4), 2380–2392. doi:10.1364/boe.452952

CrossRef Full Text | Google Scholar

Liu, H., Wang, Y., Xu, D., Wu, L., Yan, C., Yan, D., et al. (2017). High-sensitivity attenuated total internal reflection continuous-wave terahertz imaging. J. Phys. D. Appl. Phys. 50 (37), 375103. doi:10.1088/1361-6463/aa7d9a

CrossRef Full Text | Google Scholar

McIntosh, A. I., Yang, B., Goldup, S. M., Watkinson, M., and Donnan, R. S. (2012). Terahertz spectroscopy: A powerful new tool for the chemical sciences? Chem. Soc. Rev. 41 (6), 2072–2082. doi:10.1039/c1cs15277g

PubMed Abstract | CrossRef Full Text | Google Scholar

McIntyre, J. D. E., and Aspnes, D. E. (1971). Differential reflection spectroscopy of very thin surface films. Surf. Sci. 24 (2), 417–434. doi:10.1016/0039-6028(71)90272-X

CrossRef Full Text | Google Scholar

Mendoza-Galvan, A., Mendez-Lara, J. G., Mauricio-Sanchez, R. A., Jarrendahl, K., and Arwin, H. (2021). Effective absorption coefficient and effective thickness in attenuated total reflection spectroscopy. Opt. Lett. 46 (4), 872–875. doi:10.1364/ol.418277

PubMed Abstract | CrossRef Full Text | Google Scholar

Meng, Q., Wang, X., Zhang, B., Qian, S., Peng, B., Zhou, H., et al. (2023). Magnetic induced terahertz modulation characteristics based on ferromagnetic nematic liquid crystals. Spectrochim. Acta A 289, 122232. doi:10.1016/j.saa.2022.122232

CrossRef Full Text | Google Scholar

Møller, U., Merbold, H., Folkenberg, J. R., and Jepsen, P. U. (2007). Determination of alcohol concentration in aqueous solutions and food analysis using reflection terahertz time-domain spectroscopy. OSA Tech. Dig. Ser. (CD) 2007, MB2. doi:10.1364/OTST.2007.MB2

CrossRef Full Text | Google Scholar

Nagai, M., Yada, H., Arikawa, T., and Tanaka, K. (2006). Terahertz time-domain attenuated total reflection spectroscopy in water and biological solution. Int. J. Infrared. Milli. Waves 27 (4), 505–515. doi:10.1007/s10762-006-9098-3

CrossRef Full Text | Google Scholar

Peng, W., Chen, S., Kong, D., Zhou, X., Lu, X., and Chang, C. (2021). Grade diagnosis of human glioma using Fourier transform infrared microscopy and artificial neural network. Spectrochim. Acta A 260, 119946. doi:10.1016/j.saa.2021.119946

CrossRef Full Text | Google Scholar

Peng, Y., Shi, C., Zhu, Y., Gu, M., and Zhuang, S. (2020). Terahertz spectroscopy in biomedical field: A review on signal-to-noise ratio improvement. PhotoniX 1 (1), 12. doi:10.1186/s43074-020-00011-z

CrossRef Full Text | Google Scholar

Penkov, N., Yashin, V., Fesenko, E., Manokhin, A., and Fesenko, E. (2017). A study of the effect of a protein on the structure of water in solution using terahertz time-domain spectroscopy. Appl. Spectrosc. 72 (2), 257–267. doi:10.1177/0003702817735551

PubMed Abstract | CrossRef Full Text | Google Scholar

Qin, J., Xie, L., and Ying, Y. (2017). Rapid analysis of tetracycline hydrochloride solution by attenuated total reflection terahertz time-domain spectroscopy. Food Chem. 224, 262–269. doi:10.1016/j.foodchem.2016.12.064

PubMed Abstract | CrossRef Full Text | Google Scholar

Qin, J., Ying, Y., and Xie, L. (2013). The detection of agricultural products and food using terahertz spectroscopy: A review. Appl. Spectrosc. Rev. 48 (6), 439–457. doi:10.1080/05704928.2012.745418

CrossRef Full Text | Google Scholar

Shen, J., Zhu, Z., Zhang, Z., Guo, C., Zhang, J., Ren, G., et al. (2021a). Ultra-broadband terahertz fingerprint spectrum of melatonin with vibrational mode analysis. Spectrochim. Acta A 247, 119141. doi:10.1016/j.saa.2020.119141

PubMed Abstract | CrossRef Full Text | Google Scholar

Shen, S., Liu, X., Shen, Y., Qu, J., Pickwell-MacPherson, E., Wei, X., et al. (2021b). Recent advances in the development of materials for terahertz metamaterial sensing. Adv. Opt. Mater. 10 (1), 2101008. doi:10.1002/adom.202101008

CrossRef Full Text | Google Scholar

Sia, S. K., and Whitesides, G. M. (2003). Microfluidic devices fabricated in poly (dimethylsiloxane) for biological studies. Electrophoresis 24 (21), 3563–3576. doi:10.1002/elps.200305584

PubMed Abstract | CrossRef Full Text | Google Scholar

Sun, Y., Du, P., Lu, X., Xie, P., Qian, Z., Fan, S., et al. (2018). Quantitative characterization of bovine serum albumin thin-films using terahertz spectroscopy and machine learning methods. Biomed. Opt. Exp. 9 (7), 2917–2929. doi:10.1364/boe.9.002917

PubMed Abstract | CrossRef Full Text | Google Scholar

Tang, C., Wang, Y., Cheng, J., Chang, C., Hu, J., and Lu, J. (2022). Probing terahertz dynamics of multidomain protein in cell-like confinement. Spectrochim. Acta A 275, 121173. doi:10.1016/j.saa.2022.121173

CrossRef Full Text | Google Scholar

Tang, M., Zhang, M., Xia, L., Yang, Z., Yan, S., Wang, H., et al. (2020). Detection of single-base mutation of DNA oligonucleotides with different lengths by terahertz attenuated total reflection microfluidic cell. Biomed. Opt. Exp. 11 (9), 5362–5372. doi:10.1364/boe.400487

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Cui, Z., Zhang, X., Zhang, X., Hu, H., Chen, S., et al. (2020a). Excitation of surface plasmon resonance onmultiwalled carbon nanotube metasurfaces for pesticide sensors. ACS Appl. Mater. Interfaces 12 (46), 52082–52088. doi:10.1021/acsami.0c10943

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Jiang, Z., Xu, D., Chen, T., Chen, B., Wang, S., et al. (2019). Study of the dielectric characteristics of living glial-like cells using terahertz ATR spectroscopy. Biomed. Opt. Exp. 10 (10), 5351–5361. doi:10.1364/boe.10.005351

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, Y., Wang, G., Xu, D., Jiang, B., Ge, M., Wu, L., et al. (2020b). Terahertz spectroscopic diagnosis of early blast-induced traumatic brain injury in rats. Biomed. Opt. Exp. 11 (8), 4085–4098. doi:10.1364/boe.395432

PubMed Abstract | CrossRef Full Text | Google Scholar

Xu, W., Xie, L., and Ying, Y. (2017). Mechanisms and applications of terahertz metamaterial sensing: A review. Nanoscale 9 (37), 13864–13878. doi:10.1039/c7nr03824k

PubMed Abstract | CrossRef Full Text | Google Scholar

Yang, X., Zhao, X., Yang, K., Liu, Y., Liu, Y., Fu, W., et al. (2016). Biomedical applications of terahertz spectroscopy and imaging. Trends Biotechnol. 34 (10), 810–824. doi:10.1016/j.tibtech.2016.04.008

PubMed Abstract | CrossRef Full Text | Google Scholar

Yang, Z., Tang, D., Hu, J., Tang, M., Zhang, M., Cui, H., et al. (2020). Near-field nanoscopic terahertz imaging of single proteins. Small 17 (3), 2005814. doi:10.1002/smll.202005814

PubMed Abstract | CrossRef Full Text | Google Scholar

Zaytsev, K. I., Kudrin, K. G., Karasik, V. E., Reshetov, I. V., and Yurchenko, S. O. (2015). In vivo terahertz spectroscopy of pigmentary skin nevi: Pilot study of non-invasive early diagnosis of dysplasia. Appl. Phys. Lett. 106 (5), 053702. doi:10.1063/1.4907350

CrossRef Full Text | Google Scholar

Zhang, G., Wang, Y., Qian, J., Wang, Y., Li, X., and Lü, J. (2023). Terahertz refractive phenotype of living cells. Front. Bioeng. Biotechnol. 10, 1105249. doi:10.3389/fbioe.2022.1105249

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, Z., Li, Y., Xiang, Z., Huang, Y., Wang, R., and Chang, C. (2022a). Dielectric dispersion characteristics of the phospholipid bilayer with subnanometer resolution from terahertz to mid-infrared. Front. Bioeng. Biotechnol. 10, 984880. doi:10.3389/fbioe.2022.984880

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, Z., Zhang, T., Fan, F., Ji, Y., and Chang, S. (2022b). Terahertz polarization sensing of bovine serum albumin proteolysis on curved flexible metasurface. Sens. Actuators A Phys. 338, 113499. doi:10.1016/j.sna.2022.113499

CrossRef Full Text | Google Scholar

Zhou, S., Valchev, D. G., Dinovitser, A., Chappell, J. M., Iqbal, A., Ng, B. W. H., et al. (2016). Terahertz signal classification based on geometric algebra. IEEE Trans. Terahertz Sci. Technol. 6 (6), 793–802. doi:10.1109/tthz.2016.2610759

CrossRef Full Text | Google Scholar

Zhu, Z., Zhang, J., Song, Y., Chang, C., Ren, G., Shen, J., et al. (2020). Broadband terahertz signatures and vibrations of dopamine. Analyst 145 (18), 6006–6013. doi:10.1039/d0an00771d

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: terahertz, attenuated total reflection, microfluidic chip, evanescent field, lactate dehydrogenase

Citation: Fu Y, Chen T, Chen L, Guo Y, Yang Z, Mu N, Feng H, Zhang M and Wang H (2023) Terahertz time-domain attenuated total reflection spectroscopy integrated with a microfluidic chip. Front. Bioeng. Biotechnol. 11:1143443. doi: 10.3389/fbioe.2023.1143443

Received: 13 January 2023; Accepted: 02 March 2023;
Published: 13 March 2023.

Edited by:

Junhong Lü, Shanghai Advanced Research Institute (CAS), China

Reviewed by:

Annalisa D'Arco, National Institute of Nuclear Physics of Rome, Italy
Hongwei Zhao, Shanghai Advanced Research Institute (CAS), China

Copyright © 2023 Fu, Chen, Chen, Guo, Yang, Mu, Feng, Zhang and Wang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Mingkun Zhang, zhangmk@cigit.ac.cn; Huabin Wang, wanghuabin@cigit.ac.cn

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.