ORIGINAL RESEARCH article

Front. Bioeng. Biotechnol., 03 December 2021
Sec. Synthetic Biology
This article is part of the Research Topic Molecular Engineering of Sensory Mechanisms in Bacteria for Biosensing Technologies and Novel Tools for Microbial Engineering View all 7 articles

Highly Sensitive Whole-Cell Biosensor for Cadmium Detection Based on a Negative Feedback Circuit

Guangbao ZhangGuangbao Zhang1Shuting HuShuting Hu1Xiaoqiang Jia,,
Xiaoqiang Jia1,2,3*
  • 1Department of Biochemical Engineering, School of Chemical Engineering and Technology, Tianjin University, Tianjin, China
  • 2Frontier Science Center for Synthetic Biology and Key Laboratory of Systems Bioengineering (MOE), School of Chemical Engineering and Technology, Tianjin University, Tianjin, China
  • 3Collaborative Innovation Center of Chemical Science and Engineering (Tianjin), Tianjin, China

Although many whole-cell biosensors (WCBs) for the detection of Cd2+ have been developed over the years, most lack sensitivity and specificity. In this paper, we developed a Cd2+ WCB with a negative feedback amplifier in P. putida KT2440. Based on the slope of the linear detection curve as a measure of sensitivity, WCB with negative feedback amplifier greatly increased the output signal of the reporter mCherry, resulting in 33% greater sensitivity than in an equivalent WCB without the negative feedback circuit. Moreover, WCB with negative feedback amplifier exhibited increased Cd2+ tolerance and a lower detection limit of 0.1 nM, a remarkable 400-fold improvement compared to the WCB without the negative feedback circuit, which is significantly below the World Health Organization standard of 27 nM (0.003 mg/L) for cadmium in drinking water. Due to the superior amplification of the output signal, WCB with negative feedback amplifier can provide a detectable signal in a much shorter time, and a fast response is highly preferable for real field applications. In addition, the WCB with negative feedback amplifier showed an unusually high specificity for Cd2+ compared to other metal ions, giving signals with other metals that were between 17.6 and 41.4 times weaker than with Cd2+. In summary, the negative feedback amplifier WCB designed in this work meets the requirements of Cd2+ detection with very high sensitivity and specificity, which also demonstrates that genetic negative feedback amplifiers are excellent tools for improving the performance of WCBs.

1 Introduction

Cadmium (Cd) is a heavy metal without a physiological function and considerable toxicity (Sinicropi et al., 2010; Friberg et al., 2019). Over the past century, many different forms of cadmium exposure have been identified, among which groundwater cadmium pollution is a severe global health problem (Rahimzadeh et al., 2017). The continuing cadmium contamination is related to its application in industry as a corrosive agent as well as its use in polyvinyl chloride (PVC) products, pigments, and Ni-Cd batteries (Genchi et al., 2020). Long-term exposure to cadmium can cause a variety of diseases, including cancer (Saha et al., 2016; Zhang and Reynolds, 2019). Cadmium pollution in the environment is a serious international problem, affecting the health of hundreds of millions of people worldwide (Elonheimo et al., 2021). Due to its toxicity and harm to human health, cadmium detection requires low-cost, sensitive environmental monitoring equipment.

The World Health Organization (WHO) has set the safe allowable level of cadmium in drinking water at 27 nM (0.003 mg/L). The EPA recommends 1 × 10−3 mg/kg/day and 5 × 10−4 mg/kg/day as reference doses for human exposure to Cd2+ in food and water. However, the cadmium concentrations released into water from industrial activities are much higher than the prescribed limits, causing environmental contamination and increasing hazards to human health (WHO, 2011; Ashizawa et al., 2012). At present, atomic absorption spectrophotometry, dithizone spectrophotometry, atomic fluorescence spectrometry, and anodic dissolution voltammetry are the globally approved techniques for measuring cadmium concentrations (Ure et al., 1978; Iu et al., 1979; Albert et al., 1992; Hou et al., 2001; Feng and Liu, 2010; Rosolina et al., 2015; Santos et al., 2015; Kumar et al., 2017; Zhou et al., 2018). However, the pre-processing and analysis of materials using these procedures frequently requires complicated and expensive instruments and experienced specialists, making them challenging to utilize in the field. Accordingly, biosensors based on enzymes, antibodies, and microbial cells have sparked interest for detecting cadmium in drinking water (Van Der Meer and Belkin, 2010; Bereza-Malcolm et al., 2015).

In recent years, whole-cell biosensors (WCBs) have received increasing attention as a precise and sensitive way of detecting hazardous heavy metal ions. Heavy metal resistance operon regulatory elements (such as transcriptional regulators and their cognate promoters) are frequently connected to reporter genes that generate a measurable output (such as fluorescence, luminescence, or enzyme analysis), so that the reporter’s signal intensity is proportional to the concentration of the heavy metal to be detected. A relatively well-studied cadmium-resistance operon was found in P. putida 06909, which contains CadR (a transcriptional regulator) and CadA (a P-type ATPase metal efflux pump) (Lee et al., 2001). In the absence of Cd2+, CadR binds to the CadR binding site within the Pcad promoter, which prevents RNA polymerase from recognizing the promoter Pcad bidirectional and blocks the transcription of cadR and cadA. In the presence of Cd2+, CadR binds to Cd2+ and alters the local structure of the promoter, stimulating cad gene transcription and eliminating cadmium from the cell (Ma et al., 2009; Summers, 2009; Liu et al., 2019). CadR promoters and regulators have been utilized to construct cadmium WCBs in a variety of microbial hosts (Sørensen et al., 2006; Van Der Meer and Belkin, 2010; Su et al., 2011). However, when used for cadmium detection below the WHO limit, poor sensitivity and specificity are serious concerns (Sayut et al., 2006; Close et al., 2012; Kim et al., 2016; Bereza-Malcolm et al., 2017).

In this study, we used P. putida KT2440, which naturally contains the cad operon, as a host to construct cadmium WCBs with improved performance through genetic circuit engineering techniques. The sensitivity of the biosensor is closely related to the relative concentration of the receptor (CadR) and ligand (Cd2+). Accordingly, the detection limit can be increased by adjusting their respective concentrations in the cell. A high-gain transcription amplifier based on the tetracycline repressor (TetR) negative feedback was designed as the amplification module to adjust the intracellular concentration of CadR, thereby increasing the expression of the reporter protein MCherry in the output module to increase its dynamic. Negative feedback loops are prevalent in nature, and negative feedback mediation mechanisms are widely recognized (Savageau, 1974). Such loops have been utilized to increase the sensitivity of WCBs to a range of analytes, including antibiotics, heavy metals, and other contaminants (Bertram and Hillen, 2008; Wang et al., 2015a; Zhang et al., 2020). This work introduces a negative feedback loop that incorporates TetR self-regulating elements into cadmium WCBs to improve the sensitivity and specificity of WCBs for the first time. This paper presents a comparison of designs with and without negative feedback amplifiers, which offers valuable insights for efficient combined circuit forms of functional and modulation modules in microbial sensors based on negative feedback amplifiers and the future development of WCBs.

2 Materials and Methods

2.1 Bacterial Strains, Reagents, and Culture Conditions

The intended WCBs were constructed and characterized in P. putida KT2440. LB broth (10 g/L peptone, 5 g/L NaCl, 5 g/L yeast extract) was used to culture the cells, and 50 mg/L kanamycin (Kan) was added where required. The same medium was used to make solid plates, with 1.5 percent (wt/vol) agar added. Unless stated otherwise, all experiments were carried out at 30°C.

The PCR reagents, restriction endonucleases, and the Basic Seamless Cloning and Assembly Kit were purchased from TransGen Biotech (China). CdCl2, Pb(NO3)2, ZnCl2, CuCl2, and NaAsO2 were purchased from Shandong Western Chemical Industry Co. Ltd., China. Anhydrotetracycline hydrochloride (aTc, purity ≥ 98%) was purchased from Aladdin (China). PCR primer synthesis and sequencing were performed by Genewiz (China).

2.2 Design and Construction of WCBs

The cadR coding sequence and the Pcad promoter that controls its expression were amplified from the genome of P. putida KT2440 (GenBank: AE015451.2). The plasmid pAM1 served as a template for the amplification of the mcherry gene, codon optimized for P. putida. The genome of Escherichia coli strain DH5α (NCBI Reference Sequence: NZ CP026085.1) served as a template for the amplification of tetR, which was then codon-optimized for P. putida. The TetR repressible promoter Plteto1 was synthesized following the BioBrick standard (http://biobricks.org). A terminator (BBa_B0015) was synthesized based on the iGEM Registry (http://parts.igem.org/Catalog) and was used to terminate gene transcription in all cases. All genes were synthesized and optimized by Genewiz (China) (Supplementary Table S1). Cadmium WCB gene circuits were constructed in the broad-host-range shuttle plasmid pBBR1MCS-2 (Supplementary Figures S1, S2), and PCR/gel electrophoresis and Sanger sequencing were used to confirm each of the genetic circuits.

2.3 Growth Curves of the WCB Strains

The growth curves of the constructed WCBs were examined at different Cd2+ concentrations to understand the harmful effects of Cd2+ on the engineered strains. A CdCl2 solution was added to the shake flasks to achieve a final concentration of 0, 0.1, 1, 10, 100, 200, 300, 400, 500, or 600 μM, and aTC solution was added to the shake flask to achieve a final concentration of 400 ng/ml. They were incubated at 30°C and 220 rpm. Before the addition of Cd2+, samples were taken every hour. When the OD600 value reached 0.6–0.8, the addition of Cd2+ was started, after which the samples were taken every 2 hours, and the OD600 was measured using a UV spectrophotometer.

2.4 Measurement of the Response Time of WCBs

In terms of practical use, the response time of WCBs is critical. When the bacterial solution’s OD600 reached 0.6–0.8, CdCl2 solution was added to the test tubes to achieve final concentrations of 0, 0.01, 0.02, 0.04, 0.05, 0.1, 1, or 10 μM, and aTC solution was added to a final concentration of 400 ng/ml. The test tubes were incubated at 30°C and 220 rpm. Every 2 h, samples comprising 200 μL of the bacterial culture were transferred to a 96-well plate, and the RFU and OD600 of the samples were determined using a microplate reader.

2.5 Dose-Dependent Response to Cadmium

The whole-cell sensor’s fluorescence response was studied at various Cd2+ concentrations. Final concentrations of 0, 0.00001, 0.0001, 0.001, 0.01, 0.02, 0.04, 0.05, 0.1, 1, or 10 μM Cd2+ and 400 ng/ml aTC solution were added when the OD600 of the bacterial solution reached 0.6–0.8, followed by incubation at 30°C and 220 rpm. After 8 h, 200 µL of the bacterial solution was transferred to a 96-well plate, and the RFU and OD600 values were determined using a microplate reader.

2.6 Specificity of the WCBs

The specificity for cadmium, in addition to the sensitivity and output signal intensity of WCBs in reaction to cadmium, is an essential element in the evaluation of WCBs. Cd2+, Pd2+, Zn2+, Cu2+, and As3+ solutions with final concentrations of 0.01, 0.1, 1, or 10 μM and a final concentration of 400 ng/ml aTC were added to a bacterial solution with an OD600 of 0.6–0.8 and incubated in shake flasks at 30°C and 220 rpm. After 8 h, 200 μL of the bacterial solution were pipetted to a 96-well plate, and the RFU and OD600 values were measured using a microplate reader.

2.7 Fluorescence Measurement

Fluorescence intensity was measured using a microplate reader (infinite 200Pro; TECAN, Switzerland), with 200 μL of the sample per well of a 96-well microtiter plate. The optical density was measured at 600 nm and the fluorescence intensity was measured using excitation/emission wavelengths of 580/610 nm, respectively. All tests were carried out in triplicates, with the unmodified P. putida KT2440 strain as a negative control (Jia et al., 2019).

The formula FIR = AFU/BFU was used to calculate the fluorescence induction rates (FIRs). The relative fluorescence value (RFU) was divided by the sample absorbance to obtain the fluorescence value (AFU). The fluorescence value of the control (BFU) was defined as the relative fluorescence value (RFU) of the unmodified P. putida KT2440 strain (negative control) divided by its absorbance. The control assay fluorescence value (BFU) was set to 1.0, and the other samples were normalized to the control assay fluorescence value (BFU).

2.8 Important Indicators of WCBs

The response time of the biosensor for Cd2+ was determined as the time at which the fluorescence response changed substantially (p < 0.05) at low Cd2+ concentrations (≤0.05 µM). The detection limit of the biosensor was established as the Cd2+ concentration that induced a significant change in fluorescence response (p < 0.05) (Liao et al., 2006; Hou et al., 2015; Bereza-Malcolm et al., 2017). The sensitivity of the biosensor was determined by the slope of the linear detection curve of the fluorescence response to Cd2+ concentration (Pola-López et al., 2018; Guo et al., 2019). The specificity of the biosensor was determined by the lowest ratio of the biosensor’s fluorescence response to Cd2+ at 0.1 µM to the fluorescence response to other heavy metal ions at 10 µM.

2.9 Analyses of Spiked Sample

To remove impurities and bacteria, the initial river water samples (collected from Longfeng River, Wuqing District, Tianjin) were filtered using a 0.22-μm pore-size membrane. To assess whether additional ions or complexes in the river water have an effect on the biosensor, LB medium was prepared directly from river water samples with 50 mg/L kanamycin for bacterial culture. Various concentrations of cadmium (0.01, 0.027, 0.045, 0.050 μM, all samples in triplicate) were added to 10 ml aliquots of bacterial culture in the test tubes and incubated until the OD600 of the bacterial culture reached 0.6–0.8. At the same time, the medium for bacterial growth was produced using ultrapure water, and a standard curve was created by adding a series of cadmium concentrations (0.01, 0.02, 0.04, 0.05 μM) (Supplementary Figure S3). To replicate a field test setting, the trials were carried out at room temperature. The standard curve equation was then used to calculate the concentration of cadmium ions in river water samples. Similarly, based on the 6 h time-dependent fluorescence response, we tested the sensor’s detection performance at medium and high concentrations of cadmium. Various concentrations of cadmium (5, 10, 50, 100, and 200 μM, all samples in triplicate) were added to 10 ml aliquots of bacterial culture in the test tubes and incubated until the OD600 of the bacterial culture reached 0.6–0.8.

3 Results

3.1 Design and Construction of Biosensor

The Cd2+ WCBs without (CM) and with (TCM) negative feedback were constructed in P. putida KT2440, as illustrated in Figure 1. The biosensor CM is based on the cadRA operon of Pseudomonas putida with its intrinsic Cd2+ regulation mechanism. The Plteto1 promoter (constitutively ON and repressed by TetR.), detection element and terminator BBa_B0015 (Plteto1-cadR-T), and cadmium-inducible promoter (Pcad) gene were coupled to the reporter element red fluorescent protein gene mcherry to construct a gene circuit (Plteto1-cadR-T-Pcad-mcherry-T) (Figure 1A). The intensity of the output signal of the red fluorescent protein mCherry is proportional to the concentration of the inducer Cd2+. As shown in Figure 1B, based on the biosensor CM, the TetR (inhibited by the addition of tetracycline or its analog, aTc), which regulates the intensity of the specific promoter Plteto1, was added to construct the biosensor TCM for transcriptional control by regulating the concentration of the intracellular receptor (cadR) (Wang et al., 2015b). When aTc is not present, TetR binds to the specific promoter Plteto1 and blocks the expression of the downstream genes. When aTc is added, it reduces the intensity of TetR inhibition at the specific promoter Plteto1. The resulting TetR continues to reduce the intensity of promoter Plteto1, keeping the intracellular CadR protein at a low level, forming a negative feedback loop that acts as a negative feedback amplifier for Cd2+ WCBs, enhancing the expression of the biosensor’s mCherry output signal. (Wang et al., 2015b). When aTc is added, it reduces the intensity of TetR inhibition at the specific promoter Plteto1. The resulting TetR continues to reduce the intensity of promoter Plteto1, keeping the intracellular CadR protein at a low level, forming a negative feedback loop that acts as a negative feedback amplifier for Cd2+ WCBs, enhancing the expression of the biosensor’s mCherry output signal.

FIGURE 1
www.frontiersin.org

FIGURE 1. Schematic diagram of a Cd2+ WCB without (A) and with negative feedback (B). (A) A typical Cd2+ WCB CM consists of the promoter Plteto1, the regulator cadR, cadR-regulated promoter Pcad, and the reporter gene mcherry. (B) The negative feedback WCB TCM additionally incorporates a negative feedback amplifier that uses the tetracycline repressor protein TetR to regulate the specific promoter Plteto1.

3.2 Growth Curves of Whole-Cell Sensor Strains

The growth curves of the CM and TCM sensor strains were evaluated at different doses of cadmium ions to better understand the harmful effects of cadmium on the designed strains. Samples were taken every hour, and Cd2+ was added when the OD600 value reached 0.6–0.8, after which samples were taken every 2 hours, and the OD600 was monitored for 24 h. After around 4 h of incubation, the CM and TCM biosensor strains both entered the logarithmic phase, as shown in Figure 2. The growth of cells was gradually inhibited with the increase of Cd2+ concentration. At Cd2+ concentrations below 10 μM, the growth of both sensor strains was largely unaffected. Cell growth was inhibited to a lesser extent at Cd2+ concentrations of 100–200 μM, and significantly inhibited above 300 μM. From the comparison of the growth curves, it can be concluded that the time to enter the stable phase of the biosensor TCM was delayed at low Cd2+ concentrations, and the time in the stable phase was prolonged at high Cd2+ concentrations. This indicates that the addition of a negative feedback amplifier improved the tolerance of the biosensor to Cd2+. Considering the inhibition of cell growth by Cd2+, 0–10 µM Cd2+ was selected for subsequent experiments.

FIGURE 2
www.frontiersin.org

FIGURE 2. Growth curves of the two WCB sensor strains at various Cd2+ concentrations. (A) Without negative feedback, WCB CM; (B) With negative feedback, WCB TCM.

3.3 Time-dependent Response

In practical applications, the response time of WCBs is a crucial issue to consider. Furthermore, an important factor when using genetic amplifiers in WCBs is base-level expression, stemming either from the sensor or amplification modules. The resulting expression loop can be self-reinforcing, resulting in a significant number of false-positive signals over time. After adding 0, 0.01, 0.02, 0.04, 0.05, 0.1, 1 or10 µM Cd2+, the time-dependent responses of the two WCBs were evaluated for 10 h. As shown in Figure 3, both WCBs showed a slight increase in background signal when incubated without Cd2+ for longer than 6 h. After 10 h, the fluorescence signal of the negative feedback biosensor TCM increased by about 2.1-fold, while that of the non-positive feedback biosensor CM increased by about 3.25-fold, and the biosensor containing the negative feedback circuit exhibited a reduction of background leakage to some extent. Thus, basal level expression of biosensors with and without negative feedback amplifiers was comparable, with no false-positive signals arising from potential leaky expression.

FIGURE 3
www.frontiersin.org

FIGURE 3. Time-dependent response of Cd2+ biosensors without (A) and with (B) negative feedback. The CM and TCM biosensor cells were grown in medium containing 0, 0.00001, 0.0001, 0.001, 0.01, 0.02, 0.04, 0.05, 0.1, 1, or 10 μM Cd2+. The error bars indicate the standard deviation of triplicate experiments. *p < 0.05; **p < 0.01; ***p < 0.001.

The fluorescence response of the biosensor gradually increased with time at different Cd2+ concentrations (Supplementary Figure S4). After adding Cd2+ at concentrations of 0–10 μM and incubating for 8 h, the fluorescence response of the biosensor CM to Cd2+ with a concentration of 0.04 μM changed significantly, and its response time was 8 h. TCM produced a significant fluorescence response to 0.05 μM Cd2+ at 4 h. Thus, the response time was reduced by 4 h. After 8 h of incubation, TCM produced a significantly higher fluorescence response to 0.01 μM Cd2+ than CM, which only produced a significant change at 0.04 μM Cd2+. This indicates that the addition of the negative feedback amplifier can significantly reduce the response time of the biosensor. In addition, the negative feedback biosensor TCM responded faster to Cd2+ with a much higher signal than the biosensor CM without negative feedback. The fluorescence intensity of TCM was about 3.91 times higher than that of CM after adding Cd2+ at a concentration of 0.1 µM for 8 h, and after 4 h of exposure to 10 μM Cd2+, its output signal was 5.81 times higher than that of CM. In order to compare the sensitivity, detection limit, and specificity between the sensors TCM and CM under the same condition, we set the incubation time of the sensor TCM the same to the sensor CM, that is 8 h, in the Concentration-dependent fluorescence response and Specificity tests. The minimum detection limit of the sensor TCM is 0.05 μM at 4 h, which does not meet the cadmium detection standard of 0.027 μM in drinking water specified by the WHO; while the minimum detection limit of the sensor TCM is 0.01 μM at 6 h, which does meet the requirement for cadmium determination. Therefore, in Analyses of spiked sample, we stipulated that the response time of the sensor TCM is 6 h.

3.4 Concentration-Dependent Fluorescence Response

Next, the amplification impact of the negative feedback loop was analyzed at different initial Cd2+ concentrations. After 8 h of exposure to Cd2+ at concentrations ranging from 0 to 10 μM at 30°C, the two WCBs were compared. Both WCBs exhibited a dose-dependent pattern with a positive relationship between fluorescence intensity and Cd2+ concentration (Figure 4). At the same time, linear fitting was performed on the fluorescence response data measured at different Cd2+ concentrations (Supplementary Figure S5), and the linear response range of the biosensor CM was 0.04–0.1 µM, with a sensitivity of 509.09. By contrast, the linear response range of the biosensor TCM was 0.0001–0.05 µM, and its sensitivity was 678.07. Hence, the latter was 33% more sensitive. The output signal of the WCB with negative feedback was significantly amplified at Cd2+ concentrations of 0.00001–10 μM, and was 1.31–3.71 times higher than the output signal of the WCB without negative feedback at the same concentrations. There was detectable MCherry expression when Cd2+ was added at a concentration as low as 0.1 nM (p < 0.05) for the WCB with negative feedback and 40 nM (p < 0.05) for the one without negative feedback. After significance analysis, the detection limit of the biosensor CM at 8 h was 40 nM, and the detection limit of the biosensor TCM was 0.1 nM, which is a 400-fold increase. Notably, the detection limit of the biosensor TCM is significantly lower than the WHO standard for Cd2+ detection in drinking water (27 nM). These findings demonstrate that a WCB with a negative feedback amplifier exhibits significantly improved fluorescence intensity, detection range, and sensitivity.

FIGURE 4
www.frontiersin.org

FIGURE 4. Dose-dependent fluorescence response of the biosensor CM without negative feedback and biosensor TCM with negative feedback. The error bars represent the standard deviations from triplicate experiments. *p < 0.05; **p < 0.01; ***p < 0.001.

3.5 Specificity of the Constructed WCBs

The specificity for cadmium, in addition to the sensitivity and intensity of the output signal, is an important factor in the evaluation of WCBs. The mCherry fluorescence of the two WCBs was evaluated after 8 h of incubation in the presence of CdCl2, Pb(NO3)2, ZnCl2, CuCl2, or NaAsO2 at final concentrations of 0.01, 0.1, 1, or 10 µM. As shown from Figure 5, the response of the TCM biosensor to other metals was negligible at 1 and 10 μM, with single intensities less than 0.51 or 1.19% compared to Cd2+ at the same concentration. At 0.01 µM and 0.1 µM, the signal from other metals was 2.27–13.54% compared to Cd2+. By contrast, the fluorescence response of the biosensor CM to other metals was approximately 31–65% at 0.01 and 0.1 µM, 10–20% at 1 μM and 16–25% at 10 µM. The cadmium output signal of the WCB was therefore considerably improved by using a negative feedback amplifier. The specificity for cadmium ions was also considerably improved, which is noteworthy given that no other design has been documented to improve WCB specificity using negative feedback circuit engineering.

FIGURE 5
www.frontiersin.org

FIGURE 5. The Cd2+ specificity of the biosensor CM without negative feedback (A) and the negative feedback amplifier biosensor TCM (B). The fluorescence intensity of the biosensor was measured by incubating the cells for 8 h in the presence of 0.01, 0.1, 1, or 10 µM Cd2+, respectively. The error bars represent standard deviations from triplicate experiments.

3.6 Analyses of Spiked Sample

Natural river water containing complex components may influence the specificity and vitality of bacterial biosensors, resulting in incorrect results in some circumstances. Furthermore, complex components in natural water may interfere with the target inducer by rendering it inaccessible to bacteria. The performance of the TCM biosensor was therefore evaluated using a series of artificially cadmium-contaminated river water samples. Inductively coupled plasma-mass spectrometry (ICP-MS) analysis determined that the cadmium concentration was 0.00054 μM in the original river water samples. The Cd2+ calibration curve of the TCM biosensor was established after 6 h of induction at final Cd2+ concentrations 0.01, 0.02, 0.04, and 0.05 μM. Cadmium concentrations in spiked cadmium-contaminated river water samples were estimated using equation Y = 18.10713 + 294.92341 × X. As can be seen in Table 1, the recovery rates of cadmium determined by the TCM biosensor ranged from 98.08 to 101.11%, which corresponded well to the actual concentrations of the samples. However, high concentration of heavy metal ions may exceed the tolerance of the sensor and then affect its detection ability. Therefore, we also tested the performance of this WCB at higher Cd2+ concentrations such as 5, 10, 50, 100, and 200 μM, with the recovery rates of 97.18, 97.65, 102.74, 104.32, and 95.45%, respectively (Supplementary Table S2). These findings suggest that the biosensor TCM can achieve quantitative detection of cadmium in a low concentration range and has a high level of interference resistance and stability for detecting medium and high concentration cadmium in river water samples.

TABLE 1
www.frontiersin.org

TABLE 1. Accuracy and reliability of the constructed TCM biosensor in the measurement of cadmium in real river water samples.

4 Discussion

The ability of WCBs to detect harmful heavy metals and metalloids (such as cadmium, lead, mercury and arsenic) in the environment has been extensively studied (Wang et al., 2013; Wan et al., 2019; He et al., 2021). Despite the fact that many WCBs have been developed and used for metal ion detection, most WCBs cannot be used for environmental monitoring because they do not meet the high sensitivity and specificity requirements (Hansen and Sørensen, 2001; Baronian, 2004; Guo et al., 2019). Pola-López et al. created a novel vector in which phage T7 RNA polymerase serves as an amplifier and green fluorescent protein (GFP) serves as a reporter protein (Pola-López et al., 2018). The biosensor they designed has a detection range of 5–140 μg/L, and concentrations of the metalloid As(III) below the WHO limit were successfully detected. However, in contrast to our design, the amplifier did not improve the performance of the arsenic biosensor, and a biosensor without the amplifier was not used as a control to evaluate the amplification effect.

In order to improve the output signal and boost the sensitivity of WCBs, many studies incorporated genetic amplifier circuits (Nistala et al., 2010; Jia et al., 2018). However, negative feedback loops have a different role in enhancing loop sensitivity depending on the genetic context and regulatory factors present. Savageau made the early assertion that “negative feedback regulation is more stable than open-loop or positive feedback regulation” (Savageau, 1974). TetR negative feedback loops have been used to regulate a variety of protein-coding genes to increase output signals or improve sensitivity. However, their use for Cd2+ detection has not been reported (Bertram and Hillen, 2008; Wang et al., 2015c). Our study shows that the TetR negative feedback circuit coupled with the cadR regulation circuit not only improves the sensitivity of the biosensor but also improves the selectivity of the biosensor for Cd2+.

One problem with adding amplifiers to WCBs is the high noise level and the false positive signal amplified due to basal expression and promoter leakage. While this leakage can be negligible in some application contexts, it can cause problems in other situations. For example, 1) at high basal expression, sensitive enzyme-based colorimetric outputs can rapidly saturate, limiting titrimetric analysis (Wackwitz et al., 2008); 2) reducing the dynamic range of the downstream output reporter expression (Nielsen et al., 2016); and 3) causing non-severe side effects in biosensor outputs with therapeutic killing capabilities. In addition, negative feedback plays an essential role in achieving bistable gene expression, traditionally referred to as high/low or ON/OFF. In these scenarios, the initial concentration of the inducer can determine the gene expression level of the biosensor, and the initial input level of the inducer can induce a switch in the gene expression of the biosensor between two steady states. Single-cell studies revealed that the activity of the TetR protein determined whether the negative feedback amplifier enhanced cell noise relative to the non-negative feedback control. Both the negative feedback and non-negative feedback systems have comparable noise levels. At greater inducer concentrations, however, the increase in noise may be attributed to an increasing system burden caused by slower cell growth rates. In our work, to compare the basal background of the biosensors CM and TCM, we measured the time-dependent response of the two biosensors at different Cd2+ concentrations. We found that the basic expression levels of the biosensors with and without negative feedback amplifiers were comparable. False-positive signals generated by potential amplification of leaky expression were not observed, and the biosensor TCM with a negative feedback circuit reduced background leakage to a certain extent.

Sensitivity and specificity are the primary indicators of microbial biosensor performance, with sensitivity being related to the detection limit as well as the resolution, and specificity being the ratio of the signal intensity of the response to the target detection to that of other ions. In 2015, Hou et al. constructed the sensor pcadCluc and pzntRluc for Cd2+ and Pb2+ with detection limits below 0.1 and 0.05 μM, respectively, and the sensor parsRluc for Cd2+ with a detection limit of 5.0 μM and less than 0.1 μM for As3+ (Hou et al., 2015). In 2015, Kumar et al. developed a microarray-based synthetic system biosensor, which reduced the reaction volume to 200 μL and the detection limit for cadmium was 0.045 μM (Kumar et al., 2017). None of these sensors meet the WHO standard of 0.027 μM for cadmium in drinking water. Since the negative feedback loop can amplify the output signal at a low concentration of Cd2+ and improve the sensitivity, the Cd2+ biosensor with the negative feedback amplifier had a wider detection range and lower detection limit. The constructed negative feedback biosensor TCM was able to detect cadmium down to 0.0001 µM and was more sensitive than other biosensors (Supplementary Table S3) constructed to date (Sayut et al., 2006; Close et al., 2012). The detection limit of the negative feedback biosensor TCM was 400-fold higher than that of the negative feedback-free biosensor CM.

In 2017, Biosensors and Bioelectronics published a paper on the functional study of a composite microbial cadmium biosensor based on different genera of Gram-negative bacteria develop by Franks’ team at La Trobe University, Australia. They expressed a multiplex cadmium microbial sensor in different Gram-negative bacteria. However, the detection limits of Cd2+ did not meet the Chinese standard of cadmium concentration in drinking water (45 nM or 0.005 mg/L), and the multiple cadmium microbial biosensor responded to several other heavy metals (e.g., similar concentrations of arsenic, mercury, and lead) (Bereza-Malcolm et al., 2017). In 2018, the sensor developed by the Bereza-Malcolm group expressed in different strains, exhibited better selectivity in the detection of arsenic ions, but could not completely overcome the interference problem of other heavy metal ions, e.g., the response signal to arsenic ions at 100 μg/L was only three times higher than that of other metal ions (Bereza-Malcolm et al., 2018). In addition to increasing the biosensor’s sensitivity, the specificity of the negative feedback amplification system for Cd2+ was also improved, since it significantly amplified the response of the fluorescence signal to Cd2+, while the response to other metals was only slightly increased. As a result, it exhibited an outstanding fluorescence difference in response to CdCl2 and other metals in general. Many cadmium WCBs have specificity issues (Sayut et al., 2006; Close et al., 2012; Segall-Shapiro et al., 2014; Kim et al., 2016; Yoon et al., 2016; Bereza-Malcolm et al., 2017). In general, protein engineering can enhance specificity by changing the interplay between regulatory factors and induced metal ions. Surprisingly, the negative feedback loop added to cadmium WCBs selectively increased the fluorescence output signal in response to cadmium and other metals. The fluorescence response of the sensor at a Cd2+ concentration of 0.1 µM is 34.2 times, 17.6 times, 34.6 times, and 41.4 times that of 10 μM Pd2+, Zn2+, Cu2+, As3+, respectively (Figure 5B). At 0.01, 0.1, 1, 10 µM concentration gradients, the signal produced by Cd2+ was between 7 and 197 times stronger than the fluorescence output signals of other metals, allowing the accurate detection of cadmium in the presence of other metals. This may be because the specific promoter Plteto1 regulates the expression level of cadR gene. At the same time, aTc molecules diffuse into the cytoplasm and bind to TetR, somewhat alleviating its repression of the specific promoter Plteto1, which results in the expression of the metal-binding protein CadR at an appropriately lower concentration in the biosensor TCM compared with the biosensor CM without the negative feedback amplifier. The special metal-binding sites of cadmium-responsive transcription factor CadR protein determine their different binding abilities to certain metals. Therefore, when cadmium is present simultaneously with other metal ions, the biosensor CM without negative feedback amplifier will bind metal ions with a weaker binding capacity to CadR to a large extent, after preferentially binding cadmium due to the higher intracellular CadR concentration. On contrary, the biosensor TCM containing negative feedback amplifier expresses the CadR at an appropriately low concentration, which can highly bind to cadmium and increase the specificity of the sensor.

The difference between the actual measurement results and the initial concentration of the sample is a worrying problem in practical applications, because when a water sample is added to the medium, the analyte will be diluted. Even if the initial concentration is within the detection limit of the sensor, a high dilution ratio may cause the sample to be undetectable. One method is to make Luria-Bertani (LB) broth with sample water and use it to inoculate WCBs. Since bacteria, like many other WCBs (Hansen and Sørensen, 2001), do not require special conditions for rapid growth and reproduction, we chose bacteria as the host in this study. Bacteria can not only survive well at room temperature, but they also have relatively low requirements in terms of pH and humidity. Bacterial WCBs are relatively more robust because the expression of reporter proteins is dependent on cell growth. The Cd2+ concentration determined by the TCM biosensor closely matched the actual concentration of the sample, showing strong resistance to interference and good stability. Although the current biosensor cells require 6 h of incubation to produce sufficient fluorescence output for the actual assay, they can be further optimized to reduce the processing time if needed. For example, the cell density can be optimized to increase the fluorescence level so that the output can quickly reach the visual detection threshold, and the output based on a fast enzymatic reaction (for example, NanoLuc (Cevenini et al., 2018) or LacZα peptide (Ma et al., 2018) can accelerate the detection response. Compared with the incubation time of WCBs under other real sample detection conditions (Supplementary Table S4), the sensor TCM with negative feedback amplifier can achieve the quantitative detection of cadmium in the low concentration range of 0.01–0.05 µM (which can achieve WHO drinking water cadmium detection standard of 0.027 μM) in 6 h. This research demonstrates the modulation element’s sensitivity in Cd2+ detection, as well as the role of negative feedback in improving the sensitivity. In order to avoid problems caused by using living cells as hosts, the same regulatory elements can also be used in the design of cell-free Cd2+ biosensors.

5 Conclusion

Overall, in terms of reaction time, sensitivity, and specificity, our findings reveal that the negative feedback biosensor TCM outperforms the non-negative feedback biosensor CM. This biosensor with a negative feedback amplifier is capable of detecting cadmium ions below WHO and FAO standards. It has the potential to be used as a field detection tool for regular monitoring of cadmium levels in drinking water. Our research emphasizes the significance of genetic circuit engineering in enhancing the performance of WCBs and provides new ideas for the development of additional biosensors. Negative feedback loops are a workable strategy to construct and improve WCBs for the detection of other heavy metals or pollutants, even though their influence on negative feedback circuits may vary due to genetic background.

Data Availability Statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Author Contributions

All authors listed have made a substantial, direct, and intellectual contribution to the work and approved it for publication. GZ performed all experiments with the help of SH, analyzed the data, and wrote the manuscript. XJ and GZ conceived the idea, designed the experiments, and modified the manuscript.

Funding

The authors wish to acknowledge the financial support provided by the National Key Research and Development Program of China (Project No. 2018YFA0902100) and the National Natural Science Foundation of China (No. 22178262, No. 21576197).

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fbioe.2021.799781/full#supplementary-material

References

Albert, L., Atherton, J., Elias, R., El-Sebae, A., and Temmink, J. (1992). Cadmium: Environ Aspects. Geneva, Switzerland: World Health Organization.

Google Scholar

Ashizawa, A., Faroon, O., Ingerman, L., Jenkins, K., Tucker, P., and Wright, S. (2012). Toxicological Profile for Cadmium. Atlanta, Georgia: Agency for Toxic Substances and Disease Registry (US).

Google Scholar

Baronian, K. H. R. (2004). The Use of Yeast and Moulds as Sensing Elements in Biosensors. Biosens. Bioelectron. 19, 953–962. doi:10.1016/j.bios.2003.09.010

PubMed Abstract | CrossRef Full Text | Google Scholar

Bereza-Malcolm, L., Aracic, S., Kannan, R., Mann, G., and Franks, A. E. (2017). Functional Characterization of Gram-Negative Bacteria from Different Genera as Multiplex Cadmium Biosensors. Biosens. Bioelectron. 94, 380–387. doi:10.1016/j.bios.2017.03.029

PubMed Abstract | CrossRef Full Text | Google Scholar

Bereza-Malcolm, L., Aracic, S., Mann, G., and Franks, A. E. (2018). The Development and Analyses of Several Gram-Negative Arsenic Biosensors Using a Synthetic Biology Approach. Sensors Actuators B: Chem. 256, 117–125. doi:10.1016/j.snb.2017.10.068

CrossRef Full Text | Google Scholar

Bereza-Malcolm, L. T., Mann, G., and Franks, A. E. (2015). Environmental Sensing of Heavy Metals through Whole Cell Microbial Biosensors: A Synthetic Biology Approach. ACS Synth. Biol. 4, 535–546. doi:10.1021/sb500286r

PubMed Abstract | CrossRef Full Text | Google Scholar

Bertram, R., and Hillen, W. (2008). The Application of Tet Repressor in Prokaryotic Gene Regulation and Expression. Microb. Biotechnol. 1, 2–16. doi:10.1111/j.1751-7915.2007.00001.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Cevenini, L., Lopreside, A., Calabretta, M. M., D’Elia, M., Simoni, P., Michelini, E., et al. (2018). A Novel Bioluminescent NanoLuc Yeast-Estrogen Screen Biosensor (nanoYES) with a Compact Wireless Camera for Effect-Based Detection of Endocrine-Disrupting Chemicals. Anal. Bioanal. Chem. 410, 1237–1246. doi:10.1007/s00216-017-0661-7

PubMed Abstract | CrossRef Full Text | Google Scholar

Close, D., Xu, T., Smartt, A., Rogers, A., Crossley, R., Price, S., et al. (2012). The Evolution of the Bacterial Luciferase Gene Cassette (Lux) as a Real-Time Bioreporter. Sensors 12, 732–752. doi:10.3390/s120100732

PubMed Abstract | CrossRef Full Text | Google Scholar

Elonheimo, H., Lange, R., Tolonen, H., and Kolossa-Gehring, M. (2021). Environmental Substances Associated with Osteoporosis-A Scoping Review. Ijerph 18, 738. doi:10.3390/ijerph18020738

CrossRef Full Text | Google Scholar

Feng, L., and Liu, J. (2010). Solid Sampling Graphite Fibre Felt Electrothermal Atomic Fluorescence Spectrometry with Tungsten Coil Atomic Trap for the Determination of Cadmium in Food Samples. J. Anal. Spectrom. 25, 1072–1078. doi:10.1039/B924270H

CrossRef Full Text | Google Scholar

Friberg, L. T., Elinder, G. G., Kjellstrom, T., and Nordberg, G. F. (2019). Cadmium and Health: A Toxicological and Epidemiological Appraisal: Volume 2: Effects and Response. Boca Raton, FL: CRC Press.

Google Scholar

Genchi, G., Carocci, A., Lauria, G., Sinicropi, M. S., Catalano, A., and health, p. (2020). Nickel: Human Health and Environmental Toxicology. Ijerph 17, 679. doi:10.3390/ijerph17030679

PubMed Abstract | CrossRef Full Text | Google Scholar

Guo, M., Du, R., Xie, Z., He, X., Huang, K., Luo, Y., et al. (2019). Using the Promoters of MerR Family Proteins as "rheostats" to Engineer Whole-Cell Heavy Metal Biosensors with Adjustable Sensitivity. J. Biol. Eng. 13, 1–9. doi:10.1186/s13036-019-0202-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Hansen, L. H., and S rensen, S. J. (2001). The Use of Whole-Cell Biosensors to Detect and Quantify Compounds or Conditions Affecting Biological Systems. Microb. Ecol. 42, 483–494. doi:10.1007/s00248-001-0025-9

PubMed Abstract | CrossRef Full Text | Google Scholar

He, M.-Y., Lin, Y.-J., Kao, Y.-L., Kuo, P., Grauffel, C., Lim, C., et al. (2021). Sensitive and Specific Cadmium Biosensor Developed by Reconfiguring Metal Transport and Leveraging Natural Gene Repositories. ACS Sens. 6, 995–1002. doi:10.1021/acssensors.0c02204

PubMed Abstract | CrossRef Full Text | Google Scholar

Hou, J., Chen, G., and Wang, Z. (2001). Simultaneous Determination of Trace Cadmium and Arsenic in Grains Using Atomic Absorption Spectrophotometry. Guang Pu Xue Yu Guang Pu Fen Xi 21, 387–390. doi:10.1016/S0924-2031(02)00007-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Hou, Q., Ma, A., Wang, T., Lin, J., Wang, H., Du, B., et al. (2015). Detection of Bioavailable Cadmium, lead, and Arsenic in Polluted Soil by Tailored Multiple Escherichia coli Whole-Cell Sensor Set. Anal. Bioanal. Chem. 407, 6865–6871. doi:10.1007/s00216-015-8830-z

PubMed Abstract | CrossRef Full Text | Google Scholar

Iu, K. L., Pulford, I. D., and Duncan, H. J. (1979). Determination of Cadmium, Cobalt, Copper, Nickel and lead in Soil Extracts by Dithizone Extraction and Atomic Absorption Spectrometry with Electrothermal Atomization. Analytica Chim. Acta 106, 319–324. doi:10.1016/S0003-2670(01)85015-7

CrossRef Full Text | Google Scholar

Jia, X., Bu, R., Zhao, T., and Wu, K. (2019). Sensitive and Specific Whole-Cell Biosensor for Arsenic Detection. Appl. Environ. Microbiol. 85. doi:10.1128/AEM.00694-19

PubMed Abstract | CrossRef Full Text | Google Scholar

Jia, X., Zhao, T., Liu, Y., Bu, R., and Wu, K. (2018). Gene Circuit Engineering to Improve the Performance of a Whole-Cell lead Biosensor. FEMS Microbiol. Lett. 365, fny157. doi:10.1093/femsle/fny157

PubMed Abstract | CrossRef Full Text | Google Scholar

Jin, Y., Chen, E., Chen, C., Zhang, X., and Chen, L. (2006). Standards for Drinking Water Quality (GB-5749-2006). Beijing, China: Beijing Ministry of Health of the People’s Republic of China.

Google Scholar

Kim, H. J., Lim, J. W., Jeong, H., Lee, S.-J., Lee, D.-W., Kim, T., et al. (2016). Development of a Highly Specific and Sensitive Cadmium and lead Microbial Biosensor Using Synthetic CadC-T7 Genetic Circuitry. Biosens. Bioelectron. 79, 701–708. doi:10.1016/j.bios.2015.12.101

PubMed Abstract | CrossRef Full Text | Google Scholar

Kumar, S., Verma, N., and Singh, A. K. (2017). Development of Cadmium Specific Recombinant Biosensor and its Application in Milk Samples. Sensors Actuators B: Chem. 240, 248–254. doi:10.1016/j.snb.2016.08.160

CrossRef Full Text | Google Scholar

Lee, S.-W., Glickmann, E., and Cooksey, D. A. (2001). Chromosomal Locus for Cadmium Resistance in Pseudomonas Putida Consisting of a Cadmium-Transporting ATPase and a MerR Family Response Regulator. Appl. Environ. Microbiol. 67, 1437–1444. doi:10.1128/AEM.67.4.1437-1444.2001

PubMed Abstract | CrossRef Full Text | Google Scholar

Liao, V. H.-C., Chien, M.-T., Tseng, Y.-Y., and Ou, K.-L. (2006). Assessment of Heavy Metal Bioavailability in Contaminated Sediments and Soils Using green Fluorescent Protein-Based Bacterial Biosensors. Environ. Pollut. 142, 17–23. doi:10.1016/j.envpol.2005.09.021

PubMed Abstract | CrossRef Full Text | Google Scholar

Liu, X., Hu, Q., Yang, J., Huang, S., Wei, T., Chen, W., et al. (2019). Selective Cadmium Regulation Mediated by a Cooperative Binding Mechanism in CadR. Proc. Natl. Acad. Sci. USA 116, 20398–20403. doi:10.1073/pnas.1908610116

PubMed Abstract | CrossRef Full Text | Google Scholar

Ma, D., Shen, L., Wu, K., Diehnelt, C. W., and Green, A. A. (2018). Low-cost Detection of Norovirus Using Paper-Based Cell-free Systems and Synbody-Based Viral Enrichment. Synth. Biol. 3, ysy018. doi:10.1093/synbio/ysy018

PubMed Abstract | CrossRef Full Text | Google Scholar

Ma, Z., Jacobsen, F. E., and Giedroc, D. P. (2009). Coordination Chemistry of Bacterial Metal Transport and Sensing. Chem. Rev. 109, 4644–4681. doi:10.1021/cr900077w

PubMed Abstract | CrossRef Full Text | Google Scholar

Nielsen, A. A. K., Der, B. S., Shin, J., Vaidyanathan, P., Paralanov, V., Strychalski, E. A., et al. (2016). Genetic Circuit Design Automation. Science 352, aac7341. doi:10.1126/science.aac7341

PubMed Abstract | CrossRef Full Text | Google Scholar

Nistala, G. J., Wu, K., Rao, C. V., and Bhalerao, K. D. (2010). A Modular Positive Feedback-Based Gene Amplifier. J. Biol. Eng. 4, 1–8. doi:10.1186/1754-1611-4-4

CrossRef Full Text | Google Scholar

Pola-López, L. A., Camas-Anzueto, J. L., Martínez-Antonio, A., Luján-Hidalgo, M. C., Anzueto-Sánchez, G., Ruíz-Valdiviezo, V. M., et al. (2018). Novel Arsenic Biosensor "POLA" Obtained by a Genetically Modified E. coli Bioreporter Cell. Sensors Actuators B: Chem. 254, 1061–1068. doi:10.1016/j.snb.2017.08.006

CrossRef Full Text | Google Scholar

Rafati Rahimzadeh, M., Rafati Rahimzadeh, M., Kazemi, S., and Moghadamnia, A. A. (2017). Cadmium Toxicity and Treatment: An Update. Caspian J. Intern. Med. 8, 135–145. doi:10.22088/cjim.8.3.135

CrossRef Full Text | Google Scholar

Rosolina, S. M., Chambers, J. Q., Lee, C. W., and Xue, Z.-L. (2015). Direct Determination of Cadmium and lead in Pharmaceutical Ingredients Using Anodic Stripping Voltammetry in Aqueous and DMSO/water Solutions. Analytica Chim. Acta 893, 25–33. doi:10.1016/j.aca.2015.07.010

CrossRef Full Text | Google Scholar

Saha, N., Mollah, M. Z. I., Alam, M. F., and Safiur Rahman, M. (2016). Seasonal Investigation of Heavy Metals in marine Fishes Captured from the Bay of Bengal and the Implications for Human Health Risk Assessment. Food control 70, 110–118. doi:10.1016/j.foodcont.2016.05.040

CrossRef Full Text | Google Scholar

Santos, I. C., Mesquita, R. B. R., and Rangel, A. O. S. S. (2015). Micro Solid Phase Spectrophotometry in a Sequential Injection Lab-On-Valve Platform for Cadmium, Zinc, and Copper Determination in Freshwaters. Analytica Chim. Acta 891, 171–178. doi:10.1016/j.aca.2015.08.021

CrossRef Full Text | Google Scholar

Savageau, M. A. (1974). Comparison of Classical and Autogenous Systems of Regulation in Inducible Operons. Nature 252, 546–549. doi:10.1038/252546a0

PubMed Abstract | CrossRef Full Text | Google Scholar

Sayut, D. J., Niu, Y., and Sun, L. (2006). Construction and Engineering of Positive Feedback Loops. ACS Chem. Biol. 1, 692–696. doi:10.1021/cb6004245

PubMed Abstract | CrossRef Full Text | Google Scholar

Segall‐Shapiro, T. H., Meyer, A. J., Ellington, A. D., Sontag, E. D., and Voigt, C. A. (2014). A 'resource Allocator' for Transcription Based on a Highly Fragmented T7 RNA Polymerase. Mol. Syst. Biol. 10, 742. doi:10.15252/msb.20145299

PubMed Abstract | CrossRef Full Text | Google Scholar

Sinicropi, M. S., Amantea, D., Caruso, A., and Saturnino, C. (2010). Chemical and Biological Properties of Toxic Metals and Use of Chelating Agents for the Pharmacological Treatment of Metal Poisoning. Arch. Toxicol. 84, 501–520. doi:10.1007/s00204-010-0544-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Sørensen, S. J., Burmølle, M., and Hansen, L. H. (2006). Making Bio-Sense of Toxicity: New Developments in Whole-Cell Biosensors. Curr. Opin. Biotechnol. 17, 11–16. doi:10.1016/j.copbio.2005.12.007

PubMed Abstract | CrossRef Full Text | Google Scholar

Su, L., Jia, W., Hou, C., and Lei, Y. (2011). Microbial Biosensors: a Review. Biosens. Bioelectron. 26, 1788–1799. doi:10.1016/j.bios.2010.09.005

PubMed Abstract | CrossRef Full Text | Google Scholar

Summers, A. O. (2009). Damage Control: Regulating Defenses against Toxic Metals and Metalloids. Curr. Opin. Microbiol. 12, 138–144. doi:10.1016/j.mib.2009.02.003

PubMed Abstract | CrossRef Full Text | Google Scholar

Ure, A. M., Hernendez-Artiga, M. P., and Mitchell, M. C. (1978). A Carbon-Rod Atomizer for the Determination of Cadmium and lead in Plant Materials and Soil Extracts. Analytica Chim. Acta 96, 37–43. doi:10.1016/S0003-2670(01)93393-8

CrossRef Full Text | Google Scholar

Van Der Meer, J. R., and Belkin, S. (2010). Where Microbiology Meets Microengineering: Design and Applications of Reporter Bacteria. Nat. Rev. Microbiol. 8, 511–522. doi:10.1038/nrmicro2392

PubMed Abstract | CrossRef Full Text | Google Scholar

Wackwitz, A., Harms, H., Chatzinotas, A., Breuer, U., Vogne, C., and Van Der Meer, J. R. (2008). Internal Arsenite Bioassay Calibration Using Multiple Bioreporter Cell Lines. Microb. Biotechnol. 1, 149–157. doi:10.1111/j.1751-7915.2007.00011.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Wan, X., Volpetti, F., Petrova, E., French, C., Maerkl, S. J., and Wang, B. (2019). Cascaded Amplifying Circuits Enable Ultrasensitive Cellular Sensors for Toxic Metals. Nat. Chem. Biol. 15, 540–548. doi:10.1038/s41589-019-0244-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, B., Barahona, M., and Buck, M. (2013). A Modular Cell-Based Biosensor Using Engineered Genetic Logic Circuits to Detect and Integrate Multiple Environmental Signals. Biosens. Bioelectron. 40, 368–376. doi:10.1016/j.bios.2012.08.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, B., Barahona, M., and Buck, M. (2015a). Amplification of Small Molecule-Inducible Gene Expression via Tuning of Intracellular Receptor Densities. Nucleic Acids Res. 43, 1955–1964. doi:10.1093/nar/gku1388

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, B., Barahona, M., and Buck, M. (2015b). Amplification of Small Molecule-Inducible Gene Expression via Tuning of Intracellular Receptor Densities. Nucleic Acids Res. 43, 1955–1964. doi:10.1093/nar/gku1388

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, B., Barahona, M., and Buck, M. (2015c). Amplification of Small Molecule-Inducible Gene Expression via Tuning of Intracellular Receptor Densities. Nucleic Acids Res. 43, 1955–1964. doi:10.1093/nar/gku1388

PubMed Abstract | CrossRef Full Text | Google Scholar

Who, G. (2011). Guidelines for Drinking-Water Quality. World Health Organ. AIDS Tech. Bull. 216, 303–304. doi:10.1016/0035-9203(84)90051-8

CrossRef Full Text | Google Scholar

Yoon, Y., Kim, S., Chae, Y., Kim, S. W., Kang, Y., An, G., et al. (2016). Simultaneous Detection of Bioavailable Arsenic and Cadmium in Contaminated Soils Using Dual-Sensing Bioreporters. Appl. Microbiol. Biotechnol. 100, 3713–3722. doi:10.1007/s00253-016-7338-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, H., and Reynolds, M. (2019). Cadmium Exposure in Living Organisms: A Short Review. Sci. Total Environ. 678, 761–767. doi:10.1016/j.scitotenv.2019.04.395

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, J., Liang, Q., Xu, Z., Cui, M., Zhang, Q., Abreu, S., et al. (2020). The Inhibition of Antibiotic Production in Streptomyces Coelicolor Over-expressing the TetR Regulator SCO3201 Is Correlated with Changes in the Lipidome of the Strain. Front. Microbiol. 11, 1399. doi:10.3389/fmicb.2020.01399

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhou, M., Wu, Y., Zhang, J., Zhang, Y., Chen, X., Ye, J., et al. (2019). Development and Collaborative Study of a Diluted Acid Mild Extraction Method for Determination of Cadmium in Grain by Graphite Furnace Atomic Absorption Spectrometry. Anal. Sci. 35, 283–287. doi:10.2116/analsci.18P321

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: cadmium detection, negative feedback amplifier, sensitivity, specificity, whole-cell biosensor

Citation: Zhang G, Hu S and Jia X (2021) Highly Sensitive Whole-Cell Biosensor for Cadmium Detection Based on a Negative Feedback Circuit. Front. Bioeng. Biotechnol. 9:799781. doi: 10.3389/fbioe.2021.799781

Received: 22 October 2021; Accepted: 17 November 2021;
Published: 03 December 2021.

Edited by:

Rodolfo García-Contreras, National Autonomous University of Mexico, Mexico

Reviewed by:

Bongiwe Silwana, Cape Peninsula University of Technology, South Africa
Jyoti Chawla, Manav Rachna International Institute of Research and Studies (MRIIRS), India
Antonia Lopreside, University of Bologna, Italy

Copyright © 2021 Zhang, Hu and Jia. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Xiaoqiang Jia, xqjia@tju.edu.cn

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.