The application of a dual-lead locking screw could enhance the reduction and fixation stability of the proximal humerus fractures: a biomechanical evaluation

  • 1Department of BioMedical Engineering, School of BioMed Science, Daegu Catholic University, Gyoungbuk, Republic of Korea
  • 2Department of Orthopaedic Surgery, Seoul National University College of Medicine, Seoul National University Bundang Hospital, Seoul, Republic of Korea

Introduction: Bicortical screw fixation, which penetrates and fixes the near and far cortex of bone, has been conventionally used to achieve compressive fixation for fracture using screws. Open reduction and internal fixation using the locking plate are widely used for treating proximal humerus fractures. However, minimal contact between the bone and the locking plate can lead to an insufficient reduction. Theoretically, a dual-lead locking screw with different leads for the screw head and body could enhance the reduction and fixation stability of fragments in proximal humeral fractures without bicortical fixation, and achieve additional compression at the bone-plate-screw interface. This study assessed the insertion mechanics of the lead ratio of the dual-lead locking screw and its effect on the fixation stability of the proximal humerus fracture.

Methods: A Multi-Fix® locking plating system composed of 3.5 mm locking screws and a locking plate was used to make a locked plating for Sawbone bone blocks and fourth-generation composite humeri. Two different types of Sawbone bone blocks were used to simulate the osteoporotic (10 PCF) and normal cancellous (20 PCF) bones. The lead of the screw head thread (Lhead) was 0.8 mm, and that of the screw body (Lbody) was 0.8, 1.25, 1.6, 2.0, and 2.4 mm, whose lead ratios (Rlead=Lbody/Lhead) were 1.0, 1.56, 2.0, 2.5, and 3.0, respectively.

Results: The dual-lead locking screw elevated the compression between the locking plate and the bone. The elevation in the compression due to the dual-lead thread became weaker for the cancellous bone when the lead of the screw body was more than twice that of the screw head. The plate/humerus compression with strong bone quality withstood higher dual-lead-driven compression.

Discussion: A dual-lead locking screw of Lbody=1.25mm (Rlead=1.56) is recommended for maximum rotational stability for the locked humerus plating. The screws with over Lbody=1.6mm (Rlead=2) have no advantage in terms of the failure torque and maximum torsional deformation. Any locking dual-lead screw with a body thread lead of <1.6 mm (Rlead=2) can be used without the risk of bone crush when surgeons require additional compression to the locked cancellous bone plating.

1 Introduction

The proximal humerus fracture is one of the major osteoporotic fractures among older adults (14). Aging could decrease bone mineral density (BMD) and increase osteoporosis risk. Osteoporosis does not provoke clinical symptoms; however, low BMD increases the risk of osteoporotic fracture and hinders the appropriate reduction and stable fixation during surgery (5).

Open reduction and internal fixation using the locking plate is one of the widely used methods for treating proximal humerus fractures (6). In a bone healing process, the periosteal vessels contribute to endochondral ossification. Locking the screw head and the plate hole inhibits normal and transverse movements between the surfaces of the bone and plate (7, 8). The restriction in the normal movement of the locked plating is advantageous because it exerts no or minimal compression on the periosteal vessels. However, locked plating does not achieve bone traction along the transcortical direction by screw fixation (913). In contrast, the use of bicortical fixation is limited in proximal humerus fractures because the proximal screws for the locked plating are directed to the articular cartilage. Therefore, obtaining accurate anatomical reduction through compressive fixation using a locking plate is a technically demanding procedure.

A dual-lead locking screw was designed to overcome this limitation. Theoretically, the dual-lead screw could achieve additional reduction through the limited compression originating from the different lead ratios. Therefore, this study aimed to experimentally assess the insertion mechanics of the dual-lead locking screw and its effect on the fixation stability of proximal humerus fractures. Eventually, this study will provide surgeons with the biomechanical clue for the effective utilization of dual-lead locking screws when they want to apply additional anatomical reduction at performing a locked plating for the proximal humerus fracture.

2 Materials and methods

2.1 Test design

The experiments were designed in two main testing modes: screw insertion and fixation stability tests (Figure 1). The screw insertion tests assessed the intrinsic screwing features of the locking head screws, whereas the fixation stability tests evaluated the structural stability of the locked proximal humerus plating complex.

Figure 1
www.frontiersin.org

Figure 1. Scheme of tests. Four kinds of tests (T1–T4) were designed to assess the effect of the lead ratio on screw insertion mechanics and failure properties.

2.2 Ethics approval and consent to participate

This study does not include any data from individuals.

2.3 Materials

2.3.1 Screws and plates for the fixation of proximal humerus fractures

The tested fixation system for the proximal humerus fracture was a Multi-Fix® locking plating system (ORTHOTECH Co., Ltd, Republic of Korea). It is composed of a locking head and locking head screws that are made of titanium.

In screw mechanics, a lead is the linear travel per screw revolution, while a pitch is the distance between the crests of adjacent threads. The number of thread starts is the number of neighboring threads that simultaneously start screwing. Therefore, a lead is the multiplication of a pitch and the number of thread starts (Figure 2).

Figure 2
www.frontiersin.org

Figure 2. Screw structure. (A) The anatomy of a locking head screw, (B) Single-start and multi-start threads Lhead=shead×pheadLbody=sbody×pbody. p: pitch, i.e., the longitudinal distance between two adjacent thread mountains. s: the number of thread starts. L: lead, i.e., the linear travel per one screw revolution.

The head and body threads of all screws are single-start, with 3.5 mm of nominal screw diameter. The lead of the head thread (Lhead) was 0.8 mm and the leads of the body (Lbody) were 0.8 mm, 1.25 mm, 1.6 mm, 2.0 mm, and 2.4 mm. As given in Equation 1, the lead ratio (Rlead) of a locking screw was defined as the ratio of the thread lead of the screw body with respect to that of the screw head,

Rlead=Lbody/Lhead(1)

All the screws of Rlead1 are dual-lead screws. The screw having the same 0.8 mm lead for its head and body is a single-lead screw (Rlead=1). In contrast, other screws with different leads for their head and body are Rlead>1 and are referred to as “dual-lead screws” (Table 1). Consequently, one single-lead and four dual-lead screws were compared. Simplified plates with seven holes were made since the fixation stability test of the plated humerus uses a maximum of seven screws (Table 2).

Table 1
www.frontiersin.org

Table 1. Locking head screws.

Table 2
www.frontiersin.org

Table 2. Locking humerus plates.

2.3.2 Artificial bone blocks and composite humeri

Synthetic bone blocks and humeri were used for the screw insertion tests (plate/bone compression test [T1] and the bone crush test [T2]) (Table 3). The densities of the bone blocks (Sawbones, SKU1522-01, USA) were 10 and 20 pounds per cubic foot (PCF), corresponding to osteoporotic and normal cancellous bones, respectively (14, 15). The artificial humerus was a fourth-generation composite humerus (Sawbones, SKU3404, USA) with a mean density of 17 PCF (16).

Table 3
www.frontiersin.org

Table 3. Artificial bones (data from available at: https://www.matweb.com).

2.4 Screw insertion tests (plate/bone compression [T1] and the bone crush tests [T2])

The screw insertion tests aim to assess interactive mechanical responses between the locking head screws and the bone block. As in Figure 2, the screw insertion tests can be subdivided into the plate/bone compression test (T1) and the bone crush test (T2).

2.4.1 Plate/bone block compression test (T1)

The plate/boneblock compression test (T1) is an angular displacement control test. As the screw enters the bone block, the computer records the insertion torque and the plate's compression force against the bone block. An automatic screw insertion tester (Shinsung 88, SS-5511, Shinsung Co., Republic of Korea) was newly developed based on Mecmesin torque testing parts and software (Mecmesin, a part of PPT Group UK Ltd, United Kingdom) (Figure 3A). To measure the compressive force of the plate against the bone block, a load cell (UNCDW-500N, Unipulse Co., Japan) was settled on the cylindrical housing in the bone block. The compression exerted by the plate was measured as the test locking screw head entered the locking plate's screw hole (Figure 3B).

Figure 3
www.frontiersin.org

Figure 3. Testing setup for the plate/bone compression test. (A) An automatic screw insertion tester, (B) the loadcell settles in a cylindrical housing in a bone block and can measure the compression applied by the plate's contact against the bone block.

The holes and housing for the load cell were precisely machined for repeatable tests. The holes were machined to 2.8 mm in diameter and 35 mm in length, serving as pilot holes for the 3.5 mm screw. A 28 mm long screw was inserted as a stabilization anchor, and subsequently, 40 mm long screws were tested. Based on ASTM F543A2, the automatic screw tester rotated screws at 3 rpm with an axial compression of 1.1 kgf. Real-time load cell and torque data were recorded on the computer.

2.4.2 Bone crush test (T2)

Dual-lead locking screw plating will make a compression between the plate and the bone, and the bones contacting the screw's body can be crushed due to the compression. If the crush is enormous, bone healing to fill the crushed space will be prolonged. Therefore, understanding the extent of bone crush emanating from thread compression is crucial.

Here, 10 PCF and 20 PCF bone blocks were used (Table 3). Figure 4 depicts the measurement process of the crushed bone's width. First, two bone blocks were precisely machined into halves of pilot holes on one of the side walls (Figure 4A). Thereafter, two blocks were clamped such that the halves of their pilot holes matched perfectly constructed holes. The footprint created by only the screw insertion was excluded in observing the bone crush due to the dual-lead locked plating. Therefore, the footprints without the plate and the dual-lead locked plating were separately measured (Figures 4B,C, respectively). The screw insertion without the plate was performed while the insertion torque remained <0.1 N-m. Conversely, the screw insertion with the plate was manually performed until the screw/plate locking was complete because the automatic screw tester lacked sufficient space for the bone block and metal clampers. Figure 4D displays the footprint created by dual-lead locked plating of various lead ratios.

Figure 4
www.frontiersin.org

Figure 4. The measurement process of crushed bone length. (A) Two bone blocks with half pilot holes on its wall, (B) the screw insertion into the pilot holes formed by clamping corresponding blocks, (C) the screw insertion through the locking plate and then into the pilot holes formed by clamping corresponding blocks, (D) footprint formed by dual-lead locking head screws of various lead ratios, and (E) the calculation of the bone crush width.

The footprints were observed in detail using the NiKon Tool Measuring Microscope MM-800/l (NiKon Co., Japan). The microscope images were imported to CAD software (SolidWorks Co., Dassault Systèmes, France), and the footprint created by the screw thread was measured using the scale bar (Figure 4E). Moreover, the thread footprint can vary depending on the location; therefore, the footprint at the center of the whole screw footprint area was measured. The footprint width (l) was defined as the longitudinal length of the central thread footprint.

Any screw thread traveling through the bone will leave its footprint. The true bone crush emanating from dual-lead locked plating should eliminate the footprint created by the plating-free invasion of the screw body thread. Consequently, the bone crush width (C) was calculated as the width remaining after subtracting the footprint width of the screw insertion (lscrew) from the footprint width of dual-lead locked plating (lplate+screw) as illustrated below:

C=lplate+screwlscrew(2)

where lscrew is the thread footprint width formed by screw insertion into a bone block without plating, and lplate+screw is the thread footprint width formed by screw insertion through the locking plate and a bone block.

2.5 Fixation stability tests (plate/humerus compression [T3] and torsional failure load [T4] tests)

The structural stability of the proximal humerus fracture fixed with a locked plating was evaluated by plate/humerus compression test (T3) and torsional failure load test (T4). Unlike the screw insertion tests (T1 and T2), these fixation stability tests use the composite humeri (Table 2) to evaluate the locked proximal plating fixation considering anatomical topology and structural deformity.

2.5.1 Plate/humerus compression test (T3)

The composite humerus was precisely machined to simulate a proximal humeral fracture of a 5 mm gap (17). Furthermore, the screw pilot holes and load cell housing were machined (Figure 5A). A fixation mold was machined using a 5-axis computer numerical control machine to rigidly hold the humerus to mimic the medial anatomical contour of the proximal humerus. Thus, once the humerus fracture compartments have been fixed on the mold, they typically maintain the alignment of the previous specimens (Figure 5B). Figure 5C illustrates that distal and proximal fixation screws were inserted sequentially, keeping the loadcell subjected to a compression <25 N. We chose the 25 N compression at which fixation screws could complete locked plating and the plate's under surfaces engaged with the fixation screws could make contact with the humerus. Finally, the plate/humerus compression was measured sequentially at every locking of the anterior and posterior test screws completed (Figure 5D).

Figure 5
www.frontiersin.org

Figure 5. The process of the plate/humerus compression test. (A) Screw pilot holes and the loadcell housing were precisely machined, (B) the humerus was mounted on the fixation mold for repeatable humerus alignment, (C) a scheme representing the locations and installation sequence of fixation screws and a simulated fracture, and (D) real testing scene.

2.5.2 Torsional failure test (T4)

The torsional stability of the locked proximal humerus plating was evaluated by measuring failure torque. The specimens used for T3 were utilized for this test. The loadcells from the specimens used for T3 were removed by releasing the four proximal screws and re-inserting them until the screw heads were completely locked. The specimen's distal end was fixed in the plastic pipe using Kirschner wires and plastered to be clamped with the rotational metal cylinder (Figure 6). The dynamic tester 810E (Testresources Co., USA) was used for the mechanical test. A customized rotational jig converted the vertical linear movement of the dynamic tester into a torsional movement of the metal cylinder clamping the distal end (Figures 6A,B). The loading head of the dynamic tester moved at a downward rate of 5 N/s, corresponding to a torsion rate of 0.167 N-m/s (18, 19). Due to the torque, the distal humerus rotates externally while the humeral head is fixed at the end of the rotation jig.

Figure 6
www.frontiersin.org

Figure 6. The fixation torsional stability test. (A) The mechanical tester, customized rack and pinion system enable the converting of the linear movement into an axial torsion. The force rate of 5 N/s was a converted torque rate of 0.167 N-m/sec, (B) the humerus fixation failed at the maximum internal torque, and (C) the failed specimen due to the torsion.

3 Results

3.1 Plate/bone block compression test (T1)

Figure 7A reveals a locked plating with the bone block and the mechanical response according to the screw insertion process. As the automatic screw tester began to rotate the 1.25 mm screw (lead ratio; 1.65 = 1.25/0.8) diameter, the screw body entered the pilot hole, and the insertion torque slightly increased because the self-tapping flutes engaged the pilot hole entrance. The insertion torque and plate/bone compression remained <0.5 N-m and 0 N, respectively, until the screw head met the locking plate. As long as only the screw body was inserted into the bone block, the insertion torque did not increase since the screw needed to construct a thread passage on the wall of the pilot hole rather than drilling the opposing bone volume. Immediately the screw head contacted the plate, the screw-head and plate locking started, and the inertion torque and plate/bone compression increased rapidly. When the plate/screw locking was completed, the insertion torque reached its maximum. At this maximum torque, the value of the plate/bone compression was defined as the maximum compression, as it marginally increased and slightly decreased again to the level at the maximum torque.

Figure 7
www.frontiersin.org

Figure 7. The plate/boneblock compression test results. For each condition, 10–12 successful test results were obtained: (A) The real-time record of mechanical response according to the screw insertion process of a locked plating with the bone block. As the automatic screw tester began to rotate the 1.25 mm screw (lead ratio; 1.65 = 1.25/0.8) diameter, the screw body entered the pilot hole, and the insertion torque slightly increased because the self-tapping flutes engaged the pilot hole entrance, (B) The maximum insertion torque measured at the complete locked plating, (C) The maximum plate/bone block compression.

Seeing the maximum insertion torque, only Rlead=1.56 cases had a higher value than the single lead screw (Rlead=1). Other screws of Rlead>1.56 demonstrated lower insertion torque than the single lead screw (Figure 7B). At the 10 PCF bone block, the maximum insertion torques were 2.8 ± 2.5, 3.3 ± 2.2, 0.4 ± 0.1, 0.7 ± 1.3, and 0.3 ± 0.1 N-m for Rlead=1.0,1.56,2.0,2.5,and3.0, respectively. At the 20 PCF bone block, the maximum torques were 3.9 ± 2.4, 4.3 ± 0.9, 3.5 ± 1.9, 2.5 ± 1.5, and 1.29 ± 0.2 N-m for Rlead=1, 1.56,2,2.5,and3, respectively. Typically, the highest value was observed for Rlead=1.56 at 10 and 20 PCF and other dual-lead screws revealed lower values than the single lead screw.

As the locking mechanism works, the locked plating induces plate/bone block compression due to the dual-lead structure (Figure 7C). At the 10 PCF bone block, the mean maximum plate/bone block compressions were 1.8 ± 2.7, 218.9 ± 89.6, 197.1 ± 38.2, 146.1 ± 36.1 N, and 114.4 ± 36.1 N for Rlead=1.0,1.56,2.0, 2.5,and3.0, respectively. The highest compression was observed for Rlead=1.56, and it decreased to 50% of that of Rlead=1.56 with increased Rlead. At the 20 PCF bone block, the mean maximum plate/bone block compressions were 11.1 ± 21.9, 413.3 ± 125.9, 410.9 ± 122.3, 302.8 ± 152.2, and 313.8 ± 133.2 N for Rlead=1.0,1.56,2.0,2.5,and3.0, respectively. Similar to 10 PCF, the highest value was observed for Rlead=1.56; however, it was very close to that for Rlead=2. As Rlead rose to 2 and 2.5, the compression decreased by approximately 25% of that of Rlead=1.56. The standard deviation of 20 PCF cases was larger than that of 10 PCF cases.

3.2 Bone crush test (T2)

The footprint width due to the screw insertion without the plate was <0.2 mm, except for the insertion of Rlead=3.0 screw into the 10 PCF bone block (Figure 8A). This exception might result from the abrupt large stepping of the screw body thread of Rlead=3 into the osteoporotic bone. The footprint width of the locked plating increased significantly for the screws over Rlead=2.5. The increase was more prominent in the locked plating of Rlead=3.

Figure 8
www.frontiersin.org

Figure 8. The bone crush test results. For each condition, a single successful test result was collected. (A) Thread footprint widths made using only the screw insertion and locking screw/plate plating, respectively. (B) The bone crush width calculated using Equation 2.

The bone crush due to locked plating was not large for the cases of Rlead=1.0,1.56,or2.0 (Figure 8B). However, it increased with Rlead>2, which was more prominent in the cases of Rlead=2.5and3 and Rlead=3 for 10 and 20 PCF, respectively.

3.3 Plate/humerus compression test (T3)

Locked plating to the proximal humerus induces plate/humerus compression. At the completion of the locked plating, the plate/humerus compression was measured by the compression sensor installed under the plate. The plate/humerus test was performed only for Rlead=1.0,1.56,2.0,and2.5 cases because the plate/bone block compression did not differ for Rlead=3 cases compared with Rlead=2.5 cases and the bone crush width was too large for Rlead=3.0.

The insertion of a Rlead=1.0 test screw induced the compression of 39–138 N (88.93 ± 43) higher than the values observed during the plate/bone block insertion tests because the fixation screws were inserted before the test screws (Figure 9). The anterior test screw demonstrated an increase in the compression for Rlead=1.56and2.0, and almost no change was observed for Rlead=2.5. The posterior test screw compression increased with Rlead.

Figure 9
www.frontiersin.org

Figure 9. The plate/humerus compression, due to the sequential locked plating of anterior and posterior screws. For each condition, six humerus specimens were tested.

3.4 Torsional failure test (T4)

At the point where an abrupt decrease in torque, the failure torque and rotation were identified. We concluded that the locked plating with Rlead=2.5 screw is not useful because the plate/humerus compression for Rlead=2.5 slightly changed from Rlead=2.0 and the bone crush was immense. Therefore, this torsional failure test was performed only for Rlead=1,1.56, and 2.0.

For each Rlead, the six specimens used for the plate/humerus compression test were utilized for this torsional failure test. The specimens showing bone breaks outside the plating or breaks of the screw at low torsion were not considered: one specimen of Rlead=1.0 and two of Rlead=2.0. Consequently, results from the first successful four specimens each of Rlead=1.0,1.56,and2.0 were collected. The mean maximum failure torque was 26.2 ± 1.8, 27.5 ± 6.6, and 23.0 ± 4.9 N-m for Rlead=1.0,1.56,and2.0, respectively. The failure rotation for Rlead=1.56 was 82.8° ± 22.2° which was 52% and 30% higher relative to those of Rlead=1.0and2.0, respectively (Figure 10A). The failure rotation for Rlead=1.56 was 82.8° ± 22.2° which was 52% and 30% higher relative to those of Rlead=1and2, respectively. Torsional stiffness was lower for Rlead>1 and lowest at Rlead=1.56 (Figure 10B).

Figure 10
www.frontiersin.org

Figure 10. The fixation torsional failure test results. (A) Failure torque and rotation, (B) Torsional failure stiffness = (failure torque)/(failure rotation).

4 Discussion

This study assessed the biomechanics of the dual-lead locking screw and the fixation stability for various screw lead ratios. During the screw insertion tests (T1 and T2), we observed plate/bone block interactive mechanics when the dual-lead screws were inserted into locking plates and bone blocks. Furthermore, the fixation stability tests of the locked plating of the proximal humerus fracture (T3 and T4) demonstrated the effect of the dual-lead screws.

Accurate reduction and firm fixation are critical to healing after a fracture. However, achieving a firm fixation in elderly patients with osteoporosis remains challenging (4, 2022). Several kinds of screw compression mechanisms have been utilized for sufficient anatomical reduction and stable fixation of fractured fragments (23). Combinations of different diameters, lagging lengths, or leads create various interfragmentary screw compression. An increasing thread diameter can induce interfragmentary compression because the larger proximal diameter struggles to advance into the narrower distal screw diameter. Stryker TwinFix (Stryker, Kalamazoo, MI, USA) has identical thread pitch and lead but different proximal and distal diameters. In addition, a longer lagging length (i.e., non-threaded length) induces more interfragmentary compression since the compressed bone length is proportional to the lagging length. A varying screw lead design can draw a gradual compression between fracture fragments as the screw advances. Acumed Acutrak 2 Mini (Acumed, Hillsboro, OR, USA) used a continuously varying lead screw through the whole thread. In contrast, KLS Martin HBS 2 Midi (KLS Martin, Tuttlingen, Germany) adopted the dual-lead screw whose proximal and distal leads differ. The dual-lead screw is a screw with different head and body leads. Owing to the different stride lengths due to different proximal and distal leads, a compression exists between the plate and bone fragments as the head's external thread screws into the metal plate's internal thread. These additional anatomical reductions were achieved using the dual-lead screw only.

The dual-lead locked plating is the application of additional plate/bone compression utilizing the combined effect of the high stability of the locked plating fixation and the limited compression owing to the dual-lead screw. Compared to conventional screw fixation, the locked dual-lead screw plating can only apply minimal compression because the locking screw head cannot advance any further when its last male thread is locked to the plate's female thread. The findings are new experimentally-proven biomechanical clues for the effective utilization of dual-lead locking screws for surgeons to be able to parametrically adjust plate-bone compression without sacrificing fixation stability or periosteal blood circulation.

Several key findings on the relationship between the lead ratio of the dual-lead locking screw and fixation stability have been identified. First, the dual-lead locking screw elevates the compression between the locking plate and the bone. Second, for the cancellous bone, the elevation in the compression due to the dual-lead thread becomes weaker if the screw body lead is twice larger than the screw head lead. Third, for the cortical/cancellous composite humerus, the elevation in the compression due to the dual-lead thread increases with the screw body lead. Hence, the plate/bone compression varies between the cancellous bone block and the composite humerus.

Thus, the reason the compression trend along the lead ratio differs for the bone block and the humerus is crucial. The bone hardness corresponds with the bone crush intensity. The densities of the tested bone blocks were 10 and 20 PCF with strengths of 2.2 and 8.4 MPa, respectively, whereas the cortical layer of the composite humerus was 100 PCF with 157 MPa strength (Table 1). The cancellous bone block will be crushed more than the composite humerus, and the plate/bone block compression will be released proportionally to the crush. In contrast, the cortical bone layer of the composite humerus directly contacting the plate is strong and will be minimally crushed, resisting the thrusting force without the force release. The deformation caused by the crush can be elastic or plastic. The elastic deformation comes first and continues to resist the applied force and disappears when the force is removed. Conversely, the permanent and unrecoverable plastic deformation that follows elastic deformation cannot resist the applied load. Therefore, we concluded that the plate/humerus compression with strong bone quality withstands higher dual-lead-driven compression.

We think the next three quantities should be regarded clinically more important: the plate/bone compression, the bone crush width, and the torsional fixation stability. For insightful and intuitive investigation of the clinical application of the outcomes, we combined the three quantities and made curve fits of the quantities after normalizing the maximum value of each measurement type (Figure 11).

Figure 11
www.frontiersin.org

Figure 11. Normalized second-order polynomial curve fits of clinically important results. (A) The plate/boneblock compression and bone crush width. (B) The plate/humerus compression, torsional failure torque, and rotation of the locked proximal humerus plating. The quantities were normalized using the maximum of each test domain (T1, T2, T3, and T4). The lead of the screw head was 0.8 mm for all the screws.

Figure 11A displays the curve fits of the plate/bone block compression and bone crush width. The compression was high; nevertheless, the bone remained small until the lead of the screw body was 1.6 mm. Thus, when surgeons require additional compression to the locked cancellous bone plating without severely damaging the bone, they can use any locking dual-lead screw whose body's thread lead is <1.6 mm (Rlead=2). Figure 11B illustrates the integration of the curve fits of the plate/humerus compression, torsional failure torque, and rotation of the locked proximal humerus plating. Notably, the screws with over Lbody=1.6mm (Rlead=2) have no advantage in terms of failure torque and maximum torsional deformation compared with the single lead locking screw (Rlead=1). However, when surgeons require maximum rotational stability for the locked humerus plating, the dual-lead locking screw of Lbody=1.25mm (Rlead=1.56) may be considered. Even though its use is not significantly advantageous in terms of failure torque, the fixation using the dual-lead locking screw of Rlead=1.56 would withstand approximately 35% more rotational deformation compared to fixation using single-lead locked screw plating.

The natural proximal humerus consists of abundant cancellous bone and relatively thinner cortical bone, and osteoporosis could reduce the cortical thickness of the proximal humerus, thereby increasing its fragility (24). To prevent the increase of fracture comminution in the osteoporotic fracture of the proximal humerus, direct reduction of fracture fragments using forceps or clamps should be avoided, and this pathoanatomical characteristic makes the accurate reduction of fracture difficult. Clinically, stay sutures over the rotator cuffs are frequently used to reduce the fracture fragments and maintain fixation by neutralizing the traction force of rotator cuffs (25); however, in cases of concurrent rotator cuff tears with proximal humerus fractures, the use of stay sutures might be limited. Furthermore, locking plate fixation would provide stronger biomechanical fixation than suture (26). Loss of reduction following locking plate fixation in proximal humerus fractures could increase the reoperation risk (27); therefore, dual-lead locked screw fixation might be helpful to achieve and maintain accurate reduction and ultimately improve the clinical outcomes after osteosynthesis surgery for proximal humerus fractures.

Our study has some limitations. The tested composite humeri were the replica of normal bone density humerus, and the test did not include those with osteoporosis. Osteoporotic proximal humerus fractures may exhibit outcomes similar to those of cancellous bone block tests. Dual-lead locked plating will be a good study to solve the fixation of osteoporotic proximal humerus. Additionally, this study tested only artificial humeri and bone blocks. The cadaveric human bones will provide more realistic results. However, the uniformity in shape and density of the artificial humerus is beneficial, unlike those of a cadaveric humerus. This study investigated the effect of the dual-lead screw without considering variations in anatomical structure or local bone quality. Finally, the maximum or minimum specific values or the trend according to the screw lead ratio may differ for the locked plating for other human body parts because other bone fixations use various screw fixation implants and have different bone quality.

In conclusion, this study demonstrated the plate/bone block interactive mechanics when the dual-lead screws were inserted into locking plates and bones. Additionally, the dual-lead locked plating can apply additional anatomical reduction through limited compression owing to the dual-lead screw of 3.5 mm diameter. Moreover, any locking dual-lead screw with a body thread lead of <1.6 mm (Rlead=2) can be utilized without the risk of bone crush. If pursuing safe fixation stability of the osteoporotic or normal proximal humerus, the dual-lead locking screw with Lbody=1.25mm when Lhead=0.8 is recommended. Our study findings provide surgeons with a biomechanical clue for the effective utilization of dual-lead locking screws when applying additional anatomical reduction during locked plating for proximal humerus fractures.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding authors.

Ethics statement

We read and agreed to the copyright and publication policy of Frontiers in Surgery.

Author contributions

EL: Data curation, Methodology, Writing – original draft, Resources. HJ: Formal Analysis, Investigation, Resources, Validation, Writing – review & editing. YL: Data curation, Methodology, Writing – original draft, Conceptualization, Investigation, Supervision, Writing – review & editing. JO: Conceptualization, Funding acquisition, Investigation, Project administration, Resources, Supervision, Writing – review & editing.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article.

This research was supported by a grant from the Seoul National University Bundang Hospital research fund (No. 13-2017-008) to JO & YL The locking plating system used in the biomechanical test was provided by its manufacturers (ORTHOTECH Co., Ltd, Republic of Korea). The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article or the decision to submit it for publication.

Acknowledgments

This study was conducted solely for research purposes and the locking plating system used in the biomechanical test was provided by its manufacturers (ORTHOTECH Co., Ltd, Republic of Korea). This paper was not affected by the support received and is solely the responsibility of the authors.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Abbreviations

BMD, bone mineral density; T1, plate/bone block compression test; T2, bone crush test; T3, plate/humerus compression test; T4, torsional failure load test.

References

1. Shin S-J, Sohn H-S, Do N-H, Kang S-S, Baek K-Y. Minimally invasive plate osteosynthesis of proximal, middle and distal humerus fractures. J Korean Orthop Assoc. (2010) 45(6):448–55. doi: 10.4055/jkoa.2010.45.6.448

Crossref Full Text | Google Scholar

2. Volgas DA, Stannard JP, Alonso JE. Nonunions of the humerus. Clin Orthop Relat Res. (2004) 419:46–50. doi: 10.1097/00003086-200402000-00008

Crossref Full Text | Google Scholar

3. Röderer G, Scola A, Schmölz W, Gebhard F, Windolf M, Hofmann-Fliri L. Biomechanical in vitro assessment of screw augmentation in locked plating of proximal humerus fractures. Injury. (2013) 44(10):1327–32. doi: 10.1016/j.injury.2013.05.008

Crossref Full Text | Google Scholar

4. Padegimas EM, Zmistowski B, Lawrence C, Palmquist A, Nicholson TA, Namdari S. Defining optimal calcar screw positioning in proximal humerus fracture fixation. J Shoulder Elbow Surg. (2017) 26(11):1931–7. doi: 10.1016/j.jse.2017.05.003

PubMed Abstract | Crossref Full Text | Google Scholar

5. Infante A, Rodríguez CI. Osteogenesis and aging: lessons from mesenchymal stem cells. Stem Cell Res Ther. (2018) 9(1):244. doi: 10.1186/s13287-018-0995-x

PubMed Abstract | Crossref Full Text | Google Scholar

6. Wang G, Mao Z, Zhang L, Zhang L, Zhao Y, Yin P, et al. Meta-analysis of locking plate versus intramedullary nail for treatment of proximal humeral fractures. J Orthop Surg Res. (2015) 10(1):1–9. doi: 10.1186/s13018-015-0338-x

PubMed Abstract | Crossref Full Text | Google Scholar

7. Fulkerson E, Egol KA, Kubiak EN, Liporace F, Kummer FJ, Koval KJ. Fixation of diaphyseal fractures with a segmental defect: a biomechanical comparison of locked and conventional plating techniques. J Trauma Acute Care Surg. (2006) 60(4):830–5. doi: 10.1097/01.ta.0000195462.53525.0c

Crossref Full Text | Google Scholar

8. Björkenheim J-M, Pajarinen J, Savolainen V. Internal fixation of proximal humeral fractures with a locking compression plate A retrospective evaluation of 72 patients followed for a minimum of 1 year. Acta Orthop Scand. (2004) 75(6):741–5. doi: 10.1080/00016470410004120

Crossref Full Text | Google Scholar

9. Cronier P, Pietu G, Dujardin C, Bigorre N, Ducellier F, Gerard R. The concept of locking plates. Proceedings of Soo Annual Meeting, Rennes; June 2009 (2010). p. S17–36

10. Park J-Y, Yoo J-H. Selection of plate in internal fixation of fractures: locking plate and compression plate. J Korean Fract Soc. (2013) 26(1):92–102. doi: 10.12671/jkfs.2013.26.1.92

Crossref Full Text | Google Scholar

11. Kim J-H, Lee Y-S, Ahn T-K, Choi J-P. Comparison of radiologic and clinical results between locking compression plate and unlocked plate in proximal humerus fractures. Clin Shoulder Elbow. (2008) 11(2):143–9. doi: 10.5397/CiSE.2008.11.2.143

Crossref Full Text | Google Scholar

12. Kang HJ, Hwang BY, Lee JJ, Shin KH, Hahn SB, Kim SJ. Surgical treatment of pathologic humeral fracture. J Korean Fract Soc. (2010) 23(2):187–93. doi: 10.12671/jkfs.2010.23.2.187

Crossref Full Text | Google Scholar

13. Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ. Biomechanics of locked plates and screws. J Orthop Trauma. (2004) 18(8):488–93. doi: 10.1097/00005131-200409000-00003

PubMed Abstract | Crossref Full Text | Google Scholar

14. Chapman JR, Harrington RM, Lee KM, Anderson PA, Tencer AF, Kowalski D. Factors affecting the pullout strength of cancellous bone screws. J Biomech Eng. (1996) 118(3):391–8. doi: 10.1115/1.2796022

PubMed Abstract | Crossref Full Text | Google Scholar

15. Ramaswamy R, Evans S, Kosashvili Y. Holding power of variable pitch screws in osteoporotic, osteopenic and normal bone: are all screws created equal? Injury. (2010) 41(2):179–83. doi: 10.1016/j.injury.2009.08.015

PubMed Abstract | Crossref Full Text | Google Scholar

16. Zheng Y-F, Zhou J-L, Wang X-H, Shan L, Liu Y. Biomechanical study of the fixation strength of anteromedial plating for humeral shaft fractures. Chin Med J. (2016) 129(15):1850. doi: 10.4103/0366-6999.186645

PubMed Abstract | Crossref Full Text | Google Scholar

17. Clavert P, Hatzidakis A, Boileau P. Anatomical and biomechanical evaluation of an intramedullary nail for fractures of proximal humerus fractures based on tuberosity fixation. Clin Biomech. (2016) 32:108–12. doi: 10.1016/j.clinbiomech.2015.12.005

Crossref Full Text | Google Scholar

18. Schumer RA, Muckley KL, Markert RJ, Prayson MJ, Heflin J, Konstantakos EK, et al. Biomechanical comparison of a proximal humeral locking plate using two methods of head fixation. J Shoulder Elbow Surg. (2010) 19(4):495–501. doi: 10.1016/j.jse.2009.11.003

PubMed Abstract | Crossref Full Text | Google Scholar

19. Foruria AM, Carrascal MT, Revilla C, Munuera L, Sanchez-Sotelo J. Proximal humerus fracture rotational stability after fixation using a locking plate or a fixed-angle locked nail: the role of implant stiffness. Clin Biomech. (2010) 25(4):307–11. doi: 10.1016/j.clinbiomech.2010.01.009

Crossref Full Text | Google Scholar

20. Oh JH, Kim YH. The current concepts in the treatment of proximal humerus fracture. J Korean Fract Soc. (2012) 25(1):94–104. doi: 10.12671/jkfs.2012.25.1.94

Crossref Full Text | Google Scholar

21. Jabran A, Peach C, Ren L. Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review. Biomed Eng Online. (2018) 17(1):47. doi: 10.1186/s12938-018-0479-3

PubMed Abstract | Crossref Full Text | Google Scholar

22. Cornell CN. Internal fracture fixation in patients with osteoporosis. J Am Acad Orthop Surg. (2003) 11(2):109–19. doi: 10.5435/00124635-200303000-00005

PubMed Abstract | Crossref Full Text | Google Scholar

23. Gruszka D, Herr R, Hely H, Hofmann P, Klitscher D, Hofmann A, et al. Impact of different screw designs on durability of fracture fixation: in vitro study with cyclic loading of scaphoid bones. PloS One. (2016) 11(1):e0145949. doi: 10.1371/journal.pone.0145949

PubMed Abstract | Crossref Full Text | Google Scholar

24. Mather J, MacDermid JC, Faber KJ, Athwal GS. Proximal humerus cortical bone thickness correlates with bone mineral density and can clinically rule out osteoporosis. J Shoulder Elbow Surg. (2013) 22(6):732–8. doi: 10.1016/j.jse.2012.08.018

PubMed Abstract | Crossref Full Text | Google Scholar

25. Baker HP, Gutbrod J, Strelzow JA, Maassen NH, Shi L. Management of proximal humerus fractures in adults—a scoping review. J Clin Med. (2022) 11(20):6140. doi: 10.3390/jcm11206140

PubMed Abstract | Crossref Full Text | Google Scholar

26. Gaudelli C, Ménard J, Mutch J, Laflamme G-Y, Petit Y, Rouleau DM. Locking plate fixation provides superior fixation of humerus split type greater tuberosity fractures than tension bands and double row suture bridges. Clin Biomech. (2014) 29(9):1003–8. doi: 10.1016/j.clinbiomech.2014.08.013

Crossref Full Text | Google Scholar

27. Bouliane M, Silveira A, AlEidan A, Heinrichs L, Kang SH, Sheps DM, et al. Factors associated with maintaining reduction following locking plate fixation of proximal humerus fractures: a population-based retrospective cohort study. JSES Intl. (2020) 4(4):724–9. doi: 10.1016/j.jseint.2020.07.022

Crossref Full Text | Google Scholar

Keywords: proximal humerus, locked plating, dual-lead, lead ratio, compression, bone crush, insertion torque, stability

Citation: Lee E, Jeong HJ, Lee YS and Oh JH (2024) The application of a dual-lead locking screw could enhance the reduction and fixation stability of the proximal humerus fractures: a biomechanical evaluation. Front. Surg. 11:1333670. doi: 10.3389/fsurg.2024.1333670

Received: 5 November 2023; Accepted: 4 March 2024;
Published: 22 March 2024.

Edited by:

Tae Soo Bae, Jungwon University, Republic of Korea

Reviewed by:

Dohyung Lim, Sejong University, Republic of Korea
Jaeseok Ahn, Yeungnam University, Republic of Korea

© 2024 Lee, Jeong, Lee and Oh. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Yeon Soo Lee YmlvbWVjaGFuaWNzLnlzbGVlQGdtYWlsLmNvbQ== Joo Han Oh b2hqaDFAc251LmFjLmty

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.