Comparison of Application Value of Different Radiation Dose Evaluation Methods in Evaluating Radiation Dose of Adult Thoracic and Abdominal CT Scan

\nJimin HeJimin He1Guanwei DongGuanwei Dong1Yi DengYi Deng2Jun HeJun He1ZhiGang XiuZhiGang Xiu1Fanzi Feng
Fanzi Feng1*
  • 1Department of Radiology, The First People's Hospital of Longquanyi District, Chengdu, China
  • 2Department of Rehabilitation, The First People's Hospital of Longquanyi District, Chengdu, China

Objective: To explore the differences among volumetric CT dose index (CTDIvol), body-specific dose assessment (SSDEED) based on effective diameter (ED), and SSDEWED based on water equivalent diameter (WED) in evaluating the radiation dose of adult thoracic and abdominal CT scanning.

Methods: From January 2021 to October 2021, enhanced chest CT scans of 100 patients and enhanced abdomen CT scans of another 100 patients were collected. According to the body mass index (BMI), they can be divided into groups A and D (BMI < 20 kg/m2), groups B and E (20 kg/m2 ≤ BMI ≤ 24.9 kg/m2), and groups C and F (BMI > 24.9 kg/m2). The CTDIvol, anteroposterior diameter (AP), and the left and rght diameter (LAT) of all the patients were recorded, and the ED, water equivalent diameter (WED), the conversion factor (fsize,ED), (fsize, WED), SSDEED, and SSDEWED were calculated. The differences were compared between the different groups.

Results: The AP, LAT, ED, and WED of groups B, E, C, and F were higher than those of groups A and D, and those of groups C and F were higher than those of groups B and E (P < 0.05). The fsize,ED and fsize, WED of groups B, E, C, and F are lower than those of groups A and D, and those of groups C and F are lower than those of groups B and E (P < 0.05). CTDIvol, SSDEED, and SSDEWED in groups B, E, C, and F are higher than those in groups A and D, and those in groups C and F are higher than those in groups B and E (p < 0.05). In the same group, patients with chest- and abdomen-enhanced have higher SSDEWED and SSDEED than CTDIvol, patients with chest-enhanced CT scans have higher SSDEWED than SSDEED, and patients with abdomen-enhanced CT scans have higher SSDEED than SSDEWED (P < 0.05).

Conclusion: CTDIvol and ED-based SSDEED underestimated the radiation dose of the subject exposed, where the patient was actually exposed to a greater dose. However, SSDEWED based on WED considers better the difference in patient size and attenuation characteristics, and can more accurately evaluate the radiation dose received by patients of different sizes during the chest and abdomen CT scan.

Introduction

With the rapid development of clinical diagnosis and treatment and radiology technology, the application and popularity of CT examination are constantly improving, and the ionizing radiation received by patients is also constantly increasing, which has aroused widespread concern about the potential cancer risk (1). The radiation diagnosed by CT is usually higher than that reported, so it is necessary to accurately evaluate and strictly control the radiation dose of CT. At present, the CT radiation dose index (CTDIvol) under the reference standard phantom is usually used to characterize the CT radiation dose clinically, and its value reflects the radiation dose output by the CT equipment but does not consider the patient's body shape factor. In the actual scanning process, different objects have different scanning diameters and attenuation coefficients (2). Therefore, it is inaccurate to evaluate the effective dose of CT in patients with CTDIvol index. The research of Kidoh et al. (3) shows that there is a strong correlation between the specific body dose assessment (SSDE) of patients and the average skin dose, which can more accurately estimate the error of radiation dose reduction. Based on the factors of the patient's body shape, American Medical Physics Association proposed to use effective diameter (ED) and water equivalent diameter (WED) to estimate the specific dose assessment (SSDE) based on the patient's body shape to make up and correct the influence of body shape on CTDIvol and other indicators (4). In this study, we compare the differences among CTDIvol, ED-based SSDEED, and WED-based SSDEWED in evaluating the radiation dose of CT scan in the chest and the abdomen of adults with different body mass index (BMI) and discuss the further application of different radiation dose evaluation methods in clinic to provide a reference for clinical research.

Data and Methods

General Information

Enhanced chest CT scans of 100 patients and enhanced abdomen CT scans of another 100 patients were collected from January 2021 to October 2021 in our hospital. Inclusion criteria: patients and families members' informed consent; complete clinical image data; clear image, which can meet the research requirements; no metal artifact affecting the radiation dose. Among 200 patients, there were 118 men and 82 women, 21–72 years of age with an average of (48.92 ± 7.24) years, and a body mass index (BMI) of (24.02±3.19) kg/m2. This study was approved by the Ethics Committee of our hospital, and the patients and their families provided informed consent.

Research Methods

GE 128-slice spiral CT scanner was used. The patient was placed in the supine position, feet moved forward, hands raised. The chest scanning was from the top of the lung to the bottom of the lung and abdominal scanning was from the top of the liver to the lower pole of both the kidneys. During the scan, the patient was told to hold his/her breath. The scanning parameters are: adopting automatic tube current modulation technology, the tube current is 80–370 mAs, the tube voltage is 120 kV, the detector collimation is 64 lli.625 mm, the screw pitch is 0.993, and the X-ray tube rotation time is 0.75 s. All scanned images were transmitted to the image storage and transmission system for measurement, and CTDIvol of all patients was recorded. The anteroposterior diameter (AP) and left-right diameter (LAT) of all the patients were measured (at the level of left renal vein trunk and nipple) using workstation measurement software, and ED = AP×ALT, conversion factor (fsize,ED) = a × eb×ED, and SSDEED = ffsize,ED × CTDIvol were calculated at the same time (5).

An elliptical ROI was selected, including the whole section (except the bed board), the average CT value and area (A) of ROI was recorded, and the WED = =(2)(CT1000+1)×A_, the conversion factor (fsize, WED) = a × eb×WED, and SSDEWED = fsize, WED×CTDIvol for each patient was calculated. In this examination, all subjects used a 16-cm phantom in the scanning except the scout, and the other four enhanced scans used a 32-cm standard phantom to obtain CTDIvol values (6).

A total of 100 patients with enhanced chest CT scanning and 100 patients with enhanced abdomen CT scanning were divided into groups according to BMI. Patients with enhanced chest CT scan were divided into 30 patients in group A (BMI <20 kg/m2), 36 patients in group B (20 kg/m2 ≤ BMI ≤ 24.9 kg/m2), and 34 patients in group C (BMI > 24.9 kg/m2). Patients with abdominal enhanced CT scan were divided into 31 patients in group D (BMI <20 kg/m2), 35 patients in group E (20 kg/m2 ≤ BMI ≤ 24.9 kg/m2), and 34 patients in group F (BMI > 24.9 kg/m2).

Statistical Methods

SPSS22.0 software was used for processing, experimental data were measured using mean standard deviation (±s), and one-way analysis of variance was used to compare the differences between the groups in AP, LAT, ED, f sizeED, WED, and f sizeWED, respectively. The differences of CTDIvol, SSDEED, and SSDEWED among different BMI groups were compared using the t-test. The test level is α=0.05, and the difference is statistically significant when P < 0.05.

Results

Comparison of AP, LAT, ED, fsize,ED, WED, and fsize, WED in Patients With Enhanced Chest CT Scan

The AP, LAT, ED, and WED of groups B and C are all higher than those of group A, and that of group C is higher than that of group B, with statistical significance (P < 0.05). The fsize,ED and fsize, WED of group B and C are lower than that of group A, and that of group C is lower than that of group B (P < 0.05), as shown in Table 1.

TABLE 1
www.frontiersin.org

Table 1. Comparison of AP, LAT, ED, fsize,ED, WED and fsize, WED in patients with enhanced chest CT scan (n,±s).

Comparison of CTDIvol, SSDEED, and SSDEWED in Patients With Enhanced Chest CT Scan

The values of CTDIvol, SSDEED, and SSDEWED in groups B and C are higher than those in group A, and those in group C are higher than those in group B, with statistical significance (P < 0.05). In the same group, SSDEED and SSDEWED were higher than CTDIvol; SSDEWED was higher than SSDEED; and the difference was statistically significant (p < 0.05), as shown in Table 2.

TABLE 2
www.frontiersin.org

Table 2. Comparison of CTDIvol, SSDEED, and SSDEWED in patients with enhanced chest CT scan (n, ±s).

Comparison of AP, LAT, ED, fsize,ED, WED, and fsize, WED in Patients With Abdominal CT Enhanced Scanning

The AP, LAT, ED, and WED of groups E and F are all higher than those of group D, and those of group F are higher than those of group E, with statistical significance (P < 0.05). The fsize,ED and fsize, WED of group E and F are lower than those of group D, and that of group F is lower than that of group E, with statistical significance (P < 0.05), as shown in Table 3.

TABLE 3
www.frontiersin.org

Table 3. Comparison of AP, LAT, ED, fsize,ED, WED, and fsize, WED in patients with abdominal CT enhanced scanning (n, ±s).

Comparison of CTDIvol, SSDEED, and SSDEWED in Patients With Abdominal CT Enhanced Scanning

The values of CTDIvol, SSDEED, and SSDEWED in groups E and F are higher than those in group D, and those in group F are higher than those in group E, with statistical significance (p < 0.05). In the same group, SSDEED and SSDEWED were higher than CTDIvol, SSDEED was higher than SSDEWED, and the difference was statistically significant (P < 0.05), as shown in Table 4.

TABLE 4
www.frontiersin.org

Table 4. Comparison of CTDIvol, SSDEED and SSDEWED in patients with abdominal CT enhanced scan (n, ±s).

Discussion

The area on the parallel lines along the axis (z) under the single-layer scanning dose distribution curve is denoted by CDTI. Due to some limitations in its measurement, CDTI100, CDTIW, and CTDIvol were subsequently exported. CTDIvol can be used to compare the radiation doses from different CT scanners. CTDIvol represents the radiation dose value of one-layer images along the rotation axis, which is the radiation dose output level calculated based on the standard phantom. However, it has nothing to do with the scanning length. It reflects the radiation dose level output of the equipment, rather than the radiation dose received by patients, and can not truly reflect the radiation dose assessment received by patients with different body types (7, 8). Therefore, when CTDIvol is used to evaluate the radiation dose received by patients, the problem of underestimating the radiation dose received by patients with low body weight will appear (9). CTDIvol is very sensitive to the changes of scanning parameters, such as tube voltage, tube current, X-ray tube rotation time, etc. For different human bodies, its scanning diameter is different and the radiation dose is different. The emergence of SSDE parameter solves this problem.

Body-specific dose estimation is a CT dose estimation value corrected by the patient's body shape. It is obtained by standardizing CTDIvol with f on the basis of CTDIvol. Considering f , a factor related to the patient's body shape, it can more accurately evaluate the actual radiation dose received by the patient (10). Australia, New Zealand, and other countries have suggested using SSDE in chest examination to establish the dose reference (11, 12). The results show that with the increase of BMI, AP, LAT, ED, and WED of different types of patients' chest and abdomen enhanced CT scans all increased to varying degrees, while fsize,ED, fsize, WED showed a downward trend. In this study, the standard phantom with a diameter of 32 cm was used, but the ED of abdominal CT scan in most patients was <30 cm, and only 6 patients had an ED that fluctuated in the range of 30–32 cm, which was obviously different from that of the standard phantom. This study also shows that CTDIvol is used to evaluate the radiation dose in enhanced CT scans of the chest and abdomen, which is obviously lower than that of SSDEED and SSDEWED, and there is a problem of underestimating the actual radiation dose.

The SSDE effectively makes up for the deficiency of CTDIvol in body shape difference and tissue attenuation. Based on the method of ED evaluation, it is assumed that the patient's body cross-section is elliptical, and the internal components are all water, and then the circle diameter F equal to the elliptical area is used to correct. However, it is not suitable for this changeable and irregular geometric shape of the human body, and the radiation dose will be underestimated when the tissue density of the CT scan is quite different from that of water (13, 14). The method of calculating radiation dose based on WED can consider the size and X-ray attenuation factors of different parts of the patient's chest and abdomen. It is closely related to X-ray imaging and is suitable for irregular and uneven tissues of the human body (15, 16). However, there are few SSDE based on WED. This study discusses the application of SSDE based on WED calculation in the adult thorax and abdomen by comparing the differences of three different body-specific dose assessments in the body.

The results show that with the increase of BMI, the values of SSDEED and SSDEWED gradually increase. The SSDEWED of patients with enhanced CT scan in the chest is higher than that of SSDEED, while that of patients with enhanced CT scan in the abdomen is lower than that of SSDEED. Since the air content in the chest is obviously lower than that in the water model and the overall attenuation in the chest area is obviously lower than that in the water model, the average CT value corresponding to X-ray attenuation in vivo is negative, while the density of abdomen tissues is roughly the same as that in the water model, the overall attenuation is consistent with that in the water model, and the average CT value corresponding to X-ray attenuation is positive. With the increase of BMI, the difference between SSDEED and SSDEWED also gradually increases (17).

There are still some limitations in this research. This study involves only adult patients. Some studies based on ED SSDEED show that it is meaningful for infants and young children (18). At the same time, the number of cases distributed in different BMI ranges is relatively small, which needs to be further discussed in the follow-up study.

To sum up, CTDIvol and ED-based SSDEED underestimated the radiation dose to which the subject was exposed, and the patient was actually exposed to a greater dose. However, SSDEWED based on WED better considers the difference in patient size and attenuation characteristics, and can more accurately evaluate the radiation dose received by patients of different sizes during the chest and abdomen CT scan.

Data Availability Statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author/s.

Ethics Statement

This study was approved by the Medical Ethics Committee of the First People's Hospital of Longquanyi District. The patients/participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author Contributions

JiH is responsible for the writing of the paper. GD is responsible for the design of the study. YD is responsible for the inclusion of cases. JuH is responsible for the evaluation of the results. ZX is responsible for the statistics of the data. FF is responsible for the guidance of the entire study. All authors contributed to the article and approved the submitted version.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Garba I, Zarb F, McEntee MF, Fabri SG. Computed tomography diagnostic reference levels for adult brain, chest and abdominal examinations: a systematic review. Radiography (Lond). (2021) 27:673–81. doi: 10.1016/j.radi.2020.08.011

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Sun J, Li H, Li J, Yu T, Li M, Zhou Z, et al. Improving the image quality of pediatric chest CT angiography with low radiation dose and contrast volume using deep learning image reconstruction. Quant Imaging Med Surg. (2021) 11:3051–8. doi: 10.21037/qims-20-1158

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Kidoh M, Utsunomiya D, Oda S, Funama Y, Yuki H, Nakaura T, et al. Validity of the size-specific dose estimate in adults undergoing coronary CT angiography: comparison with the volume CT dose index. Int J Cardiovasc Imaging. (2015) 2:205–11. doi: 10.1007/s10554-015-0782-z

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Haba T, Koyama S, Kinomura Y, Ida Y, Kobayashi M. Influence of 320-detector-row volume scanning and AAPM report 111 CT dosimetry metrics on size-specific dose estimate: a Monte Carlo study. Australas Phys Eng Sci Med. (2016) 39:697–703. doi: 10.1007/s13246-016-0465-7

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Li B, Behrman RH. Comment on the “report of AAPM TG 204: size-specific dose estimates (SSDE) in pediatric and adult body CT examinations” [report of AAPM TG 204, 2011]. Med Phys. (2012) 39:4613–4. doi: 10.1118/1.4725756

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Gabusi M, Riccardi L, Aliberti C, Vio S, Paiusco M. Radiation dose in chest CT: assessment of size-specific dose estimates based on water-equivalent correction. Phys Med. (2016) 32:393–7. doi: 10.1016/j.ejmp.2015.12.008

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Demb J, Chu P, Yu S, Whitebird R, Solberg L, Miglioretti DL, et al. Analysis of computed tomography radiation doses used for lung cancer screening scans. JAMA Intern Med. (2019) 179:1650–7. doi: 10.1001/jamainternmed.2019.3893

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Konishi T, Shibutani T, Okuda K, Yoneyama H, Moribe R, Onoguchi M, et al. Metal artifact reduction for improving quantitative SPECT/CT imaging. Ann Nucl Med. (2021) 35:291–8. doi: 10.1007/s12149-020-01560-w

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Özsoykal I, Yurt A, Akgüngör K. Size-specific dose estimates in chest, abdomen, and pelvis CT examinations of pediatric patients. Diagn Interv Radiol. (2018) 24:243–8. doi: 10.5152/dir.2018.17450

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Anam C, Haryanto F, Widita R, Arif I, Dougherty G, McLean D. The impact of patient table on size-specific dose estimate (SSDE). Australas Phys Eng Sci Med. (2017) 40:153–8. doi: 10.1007/s13246-016-0497-z

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Jackson D, Atkin K, Bettenay F, Clark J, Ditchfield MR, Grimm JE, et al. Paediatric CT dose: a multicentre audit of subspecialty practice in Australia and New Zealand. Eur Radiol. (2015) 25:3109–22. doi: 10.1007/s00330-015-3727-y

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Hwang JY, Choi YH, Yoon HM Ryu YJ, Shin HJ, Kim HG, et al. Establishment of local diagnostic reference levels of pediatric abdominopelvic and chest CT examinations based on the body weight and size in Korea. Korean J Radiol. (2021) 22:1172–84. doi: 10.3348/kjr.2020.0890

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Bashier EH, Suliman II. Multi-slice CT examinations of adult patients at Sudanese hospitals: radiation exposure based on size-specific dose estimates (SSDE). Radiol Med. (2018) 123:424–310. doi: 10.1007/s11547-018-0859-6

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Kidoh M, Utsunomiya D, Oda S, Nakaura T, Funama Y, Yuki H, et al. Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE). Eur Radiol. (2017) 27:2267–74. doi: 10.1007/s00330-016-4618-6

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Xu J, He X, Xiao H, Xu J. Comparative study of volume computed tomography dose index and size-specific dose estimate head in computed tomography examination for adult patients based on the mode of automatic tube current modulation. Med Sci Monit. (2019) 25:71–6. doi: 10.12659/MSM.913927

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Bostani M, McMillan K, Lu P, Kim GH, Cody D, Arbique G, et al. Estimating organ doses from tube current modulated CT examinations using a generalized linear model. Med Phys. (2017) 44:1500–13. doi: 10.1002/mp.12119

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Fujii K, Nomura K, Muramatsu Y, Goto T, Obara S, Ota H, et al. Correlation analysis of organ doses determined by Monte Carlo simulation with dose metrics for patients undergoing chest-abdomen-pelvis CT examinations. Phys Med. (2020) 77:1–9. doi: 10.1016/j.ejmp.2020.07.016

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Imai R, Miyazaki O, Horiuchi T, Kurosawa H, Nosaka S. Local diagnostic reference level based on size-specific dose estimates: assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children's hospital. Pediatr Radiol. (2015) 45:345–53. doi: 10.1007/s00247-014-3189-4

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: volume CT dose index, effective diameter, equivalent diameter of water, body specific dose assessment, CT scanning

Citation: He J, Dong G, Deng Y, He J, Xiu Z and Feng F (2022) Comparison of Application Value of Different Radiation Dose Evaluation Methods in Evaluating Radiation Dose of Adult Thoracic and Abdominal CT Scan. Front. Surg. 9:860968. doi: 10.3389/fsurg.2022.860968

Received: 24 January 2022; Accepted: 22 February 2022;
Published: 25 March 2022.

Edited by:

Songwen Tan, Central South University, China

Reviewed by:

Wenjie Song, Fourth Military Medical University, China
Feng Qin, Shanghai Jiao Tong University School of Medicine, China

Copyright © 2022 He, Dong, Deng, He, Xiu and Feng. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Fanzi Feng, 47991493@qq.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.