Evaluation of femoral head bone quality by Hounsfield units: A predictor of implant failure for intertrochanteric fractures after intramedullary nail fixation

\r\nJixing Fan,,&#x;Jixing Fan1,2,†Yang Lv,,&#x;Yang Lv1,2,†Xiangyu Xu,Xiangyu Xu1,2Fang Zhou,
Fang Zhou1,2*Zhishan Zhang,Zhishan Zhang1,2Yun Tian,Yun Tian1,2Hongquan Ji,Hongquan Ji1,2Yan Guo,Yan Guo1,2Zhongwei Yang,Zhongwei Yang1,2Guojin Hou,\r\nGuojin Hou1,2
  • 1Department of Orthopedics, Peking University Third Hospital, Beijing, China
  • 2Engineering Research Center of Bone and Joint Precision Medicine, Ministry of Education, Beijing, China

Purpose: The aim of present study is to evaluate the femoral head bone quality by Hounsfield units and its relationship to the occurrence of implant failure for intertrochanteric fractures after intramedullary nail fixation.

Methods: This retrospective study assessed 160 intertrochanteric fractures treated with intramedullary fixation. Patients with and without implant failure were divided into failure and control groups, respectively. The demographic information, femoral head Hounsfield unit (HU) value, the reduction quality, status of posteromedial support and position of the screw/blade were collected and compared. The logistic regression analyses were performed to evaluate risk factors of implant failure in intertrochanteric fractures after intramedullary nail fixation.

Results: Of the patients, 15 (9.38%) suffered from implant failure after intramedullary fixation. The mean HU value of femoral head was much lower in the failure group than the control group (133.25 ± 34.10 vs. 166.12 ± 42.68, p = 0.004). And the univariate analyses showed that A3 fracture and poor reduction quality were associated with implant failure (p < 0.05). After adjustment for confounding variables, the multivariable logistic regression analyzes showed that femoral head HU value (odds ratio [OR], 0.972; 95% CI, 0.952–0.993; p = 0.008) and poor reduction quality (OR, 7.614; 95% CI, 1.390–41.717; p = 0.019) were independent influencing factors for implant failure.

Conclusion: The femoral head HU value was significantly correlated with the incidence of implant failure and can be used as an independent factor to predict implant failure for intertrochanteric fractures after intramedullary fixation.

Introduction

Intertrochanteric fractures are very common in the elderly patients, which have substantial mortality, morbidity, and healthcare costs. The major reason is osteoporosis, which is characterized by low bone mass and destruction of the bone microarchitecture (1). With the increase of the aging population, osteoporosis has gradually become a global health problem that affects nearly 200 million people worldwide (2). Furthermore, older patients are reported to have a higher rate of osteoporosis than the general population (3). Consequently, the incidence of intertrochanteric fractures is increasing quickly over the past several decades owing to the absolute increase in the elderly population, and this trend is expected to continue (4).

Intramedullary nail fixation is a common procedure used in the surgical treatment of intertrochanteric fractures (5, 6). Implant failure is one of the main reasons for secondary surgical interventions after initial surgical treatment of intertrochanteric fractures. It has been investigated that intramedullary nails have a failure rate ranging from 3% to 22% (79). Although several risk factors have been investigated to be associated with the occurrence of implant failure, poor bone quality is one of the influencing factors which could increase the risk of early mechanical failure (10, 11). However, there is a relative scarcity of literature describing the relation between low bone mineral density (BMD) and implant failure in intertrochanteric fractures after intramedullary nail fixation.

BMD measured in Hounsfield units (HU) on computed tomography (CT) is used in the evaluation of BMD and can serve as a diagnostic tool for osteoporosis (12). Although it has been reported that there is a significant correlation between HU values and dual energy x-ray absorptiometry (DXA) measurements, few studies have been carried out on the application of femoral head HU values to the prediction of implant failure for intertrochanteric fractures after intramedullary nail fixation. In the present study, we retrospectively reviewed patients who received surgical intervention with intramedullary nail fixation for intertrochanteric fractures, and we investigated possible correlations between femoral head HU values and the incidence of implant failure in these patients.

Materials and methods

Source of patients

This retrospective study was performed at a level 1 trauma center over a 6-year period from January 1, 2012, to December 31, 2018. Patients who were surgically treated by the same trauma surgeon group in our department for intertrochanteric fractures were retrospectively included in current study. Included patients should meet the following criteria: (1) age 60 years or older; (2) acute intertrochanteric fracture (<2 weeks from injury); (3) low-energy injury, a low-energy injury was defined as an injury which patients would sustain while falling over slippery ground in a walking or sitting position; (4) underwent CT scans of the injured hip before surgery; (5) intramedullary nail (IMN) fixation; (6) a minimum follow-up period of 12 months or until the time of implant failure. Exclusion criteria were as follows: (1) pathological fracture (secondary to tumor or primary hyperparathyroidism); (2) delayed fracture; (3) stress fracture; (4) open fracture, or ipsilateral knee or ankle fractures; (5) periprosthetic fracture; (6) previous history of hip surgery; (7) presence of other circumstances that affect bone metabolism, such as long-term steroid use; and (8) femoral heads with any pathologies such as focal lytic or sclerotic lesions. The demographic information (including age, gender, side of injury, and body mass index [BMI]) were collected.

Measurement of femoral head HU value

A preoperative hip CT scan (Defnition, Siemens) was performed for all of the included patients. The tube voltage of all of the CT scans was set at 120 kV. An experienced trauma surgeon (JF) independently used a PACS system (GE Electrics, Fairfield, USA) to measure the femoral head HU value. In order to facilitate measurement and repetition, we selected a single axial femoral head level. The combination of the first slice of femoral head and the last slice of femoral neck was selected. A spherical region of interest (ROI) was drawn with the largest possible diameter excluding the cortical bone (Figure 1). The reasons for choosing this region are as followed: (I) this area is the main area through which tensile trabeculae and compressive trabeculae pass together, which are weakened in osteoporosis; (II) the spherical ROI can be easily measured; and (III) it is a well-known location to ensure that measurements are performed from the same point (13).

FIGURE 1
www.frontiersin.org

Figure 1. The methods of measuring computed tomography (CT) Hounsfield units (HU) values of the femoral head. A single axial femoral head level was selected. The level included the last slice of femoral neck–first slice of femoral head combination. A spherical region of interest (ROI) was drawn with the largest possible diameter excluding the cortical bone in femoral head and average HU value (181.5 HU) was noted. CT, computed tomography; ROI, region of interest.

Surgical technique

All operations were performed by five experienced orthopedic doctors. The type of anesthesia was either spinal anesthesia or general anesthesia. All patients were treated with the reduction and internal fixation. According to the manufacturer's protocol, intramedullary devices were implanted in routine surgical procedure. The number, model, and trade mark of the IM devices were summarized in Table 1. The patient received antibiotics and deep venous thrombosis prevention postoperatively. When the x-ray showed the appearance of fracture healing, partial weight-bearing was initiated, and total weight-bearing began after clinical fracture healing. The follow-up evaluation was performed at 1, 3, 6, 12 months after the surgery and yearly thereafter.

TABLE 1
www.frontiersin.org

Table 1. Utilization of implant types.

Outcome assessment

The reduction quality, status of posteromedial support and position of the screw/blade were used to evaluate the outcome by postoperative radiographs immediately after surgery.

The reduction quality was described as good, acceptable, or poor, according to the modified criteria of Baumgaertner et al. (14) and and Kim et al. (15), which was addressed by measuring alignment and displacement of main fragments on the AP and lateral view.

According to the extent of displacement of the posteromedial fragment, the status of the posteromedial support was defined as existence or loss. Good alignment and a displacement of less than the cortical thickness were defined as existence. Otherwise, the status of the lateral femoral wall was recorded as loss (16).

The tip-apex distance (TAD) was used to access the position of the screw/blade in the femoral head. The TAD was determined by measuring the distance from the tip of the helical blade to the apex of the femoral head on both anteroposterior (AP) and lateral radiographs (14). The amount of radiographic magnification was determined precisely by the known diameter of the helical blade.

In this study, helical blade cut out, perforation and back out were defined as the implant failure. The perforation of the helical blade through the superior cortex of the femoral head or neck was defined as the blade cut out. The penetration of the helical blade from the femoral head into the surrounding soft tissues and hip joint was defined as the blade perforation (Figure 2).

FIGURE 2
www.frontiersin.org

Figure 2. Types of implant failure. (A) Screw/blade cut out; (B) screw/blade perforation; (C) screw/blade back out.

Statistical analysis

In this study, the SPSS 21.0 software was used for statistical analysis (SPSS, Chicago, Illinois, USA). For quantitative data, the normal distribution was evaluated by the one-sample Kolmogorov–Smirnov test. Continuous variables were compared by the independent sample Student t-test or the Mann–Whitney test as appropriate. For qualitative data, the χ2 test was used. Risk factors of implant failure in intertrochanteric fractures after intramedullary nail fixation were evaluated by the multivariate logistic regression analysis when the factor's p value was <0.1. The likelihood ratio backward test was conducted to find the best-fit model by selecting variables one by one. A p value <0.05 was considered statistically significant, and all tests were two sided.

Results

In the present study, a total of 160 patients were included in the final analysis. The patient's age ranged from 60 years to 97 years, and the mean age was 77.37 (±8.81) years. Among these patients, 15 patients (9.38%) suffered from implant failure. The mean failure time was 5.7 months (ranging from 1 to 27 months). Of these 15 patients with implant failure, cut out occurred in six patients, blade perforation occurred in two patients, and blade back out occurred in seven patients.

In the present study, patients were divided into control group and failure group according to patients with and without implant failure. The comparison of general characteristics between the control group and failure group was shown in Table 2. The results showed that there were significant differences between control group and failure group in AO classification, femoral head HU value and reduction quality (p < 0.05). Specially, the mean HU value of femoral head was much lower in the failure group than the control group (133.25 ± 34.10 vs. 166.12 ± 42.68, p = 0.004). However, the age, gender, injury side, BMI, TAD, posteromedial support, and NSA showed no significant differences between control group and failure group (p > 0.05).

TABLE 2
www.frontiersin.org

Table 2. The comparison of patient characteristics between control group and failure group.

The results of the adjusted multivariate logistic regression analyses were presented in Table 3. After adjustment for other risk factors, the results showed that femoral head HU value (odds ratio [OR], 0.972; 95% CI, 0.952–0.993; p = 0.008) and poor reduction quality (OR, 7.614; 95% CI, 1.390–41.717; p = 0.019) were independent influencing factors for implant failure whereas the AO classification of intertrochanteric fracture was not associated with implant failure (p > 0.05).

TABLE 3
www.frontiersin.org

Table 3. Multivariable logistic regression analysis for risk factors associated with implant failure.

To identify better predictors for implant failure, the receiver operating curve (ROC) analysis was performed, and area under the curve (AUC) values were calculated for femoral head HU. The femoral head HU had a good AUC value (0.724). The cut-off value of femoral head HU was 124.85, with a sensitivity of 84.8% and specificity of 53.3%.

Discussion

The DXA is the most commonly used parameter for the evaluation of BMD. However, the DXA had some shortcomings inherent to the imaging modality itself, such as an inability to both estimate true volumetric bone density as well as discriminate between cortical and trabecular bone mas (17, 18). The usage of CT, particularly quantitative computed tomography (QCT), has been investigated in evaluating the osteoporosis (19, 20). In addition, the QCT has been shown to be effective in distinguishing cortical bone from trabecular bone; however, this technique relies on the use of phantoms for calibration during acquisition (21). Furthermore, due to the high cost of equipment, the use of QCT software requires strict training. At present, it is difficult to apply this technology in many clinical environments (12). A HU is a dimensionless unit unique to CT examination. BMD evaluation based on HU value is correlated with the standard BMD value and physical strength of the bone, which has been widely reported (22). Furthermore, the femoral head HU obtained from routine CT could be used to evaluate regional bone quality of femoral head (13). Thus, theoretically, the CT HU value is an effective predictor of BMD-related complications after surgical treatment of intertrochanteric fractures. However, limited studies have investigated its use in assessing the femoral head bone quality in intertrochanteric fractures.

In current literature, the overall implant failure rate was 9.38%. Furthermore, we found that patients in the failure group had a significantly lower mean HU value of femoral head than those in the control group. Significant differences were also found in AO classification and reduction quality between control group and failure group. In addition, after adjustment for confounding variables, the multivariate logistic regression analysis showed that a lower femoral head HU value and poor reduction quality were independent influencing factors for implant failure whereas the AO classification of intertrochanteric fracture was not associated with the implant failure.

With regard to the stability of reconstructions after intertrochanteric fracture reduction and fixation, five influencing factors had been summarized by Kaufer et al. (5), that is, bone quality (osteoporosis), the choice of implant, the quality of reduction, the placement of the implant in femoral head, and the fragment geometry. Growing biomechanical evidence had suggested that osteoporosis was an important risk factor for mechanical failure in intertrochanteric fractures (23, 24). In this study, we found that the femoral head HU value was significantly lower in the failure group than that in the control group. Furthermore, multivariate logistic regression analysis showed that a lower femoral head HU value was an independent risk factor of implant failure. The ROC analysis showed that the cut-off value of femoral head HU was 124.85. This meant that the femoral head HU value below 124.85 was associated with increased risk of implant failure. With an AUC value of 0.724, the femoral head HU value was deemed as a reliable predictor of implant failure. The possible reason might be that poor bone quality could challenge the anchorage of the load carrier in the femoral head. Therefore, cement-augmented screws or arthroplasty (joint replacement) might be an alternative for the surgical treatment of intertrochanteric fractures with poor bone quality.

A previous study had developed a novel method based on voxel-based morphometry for estimating nail-tract bone density in intertrochanteric fracture patients treated with intramedullary nails by 3D image analysis software (25). However, the screw pathway's CT HU value could not be measured directly with routine PACS on axial images of the femoral head. If we only focus on the CT HU value of the screw's pathway, the measurement might be complex and could not be applied in routine clinical practice. Thus, we chose the routine method to measure the preoperative CT HU value of the femoral head to predict implant failure, which was convenient and simple.

It was well known that the fracture geometry had a significant impact on the outcome of surgical treatment for intertrochanteric fractures. Classification of intertrochanteric fracture served as a guideline for treatment and helped to predict the result or provided a reasonable estimation of the likely outcome (26). In some previous reports, relatively high complication rates were noted in patients with unstable intertrochanteric fractures (27). In the present study, patients with A3 fracture tended to have greater risk of implant failure compared with A1 fracture (OR, 6.250; 95% CI, 1.083–36.076; p = 0.040) in the univariate analyses. After adjusting confounding factors, we found that patients with the A3 fracture remained a nearly four-fold greater risk compared with A1 fracture in the multivariate analyses; however, no statistically significant difference was observed. This result was consistent with a previous study and implied that the fracture geometry could influence the stability of intertrochanteric fractures; this result was necessary to be verified in the future.

In the present study, for intertrochanteric fractures after intramedullary nail fixation, we demonstrated that a poor reduction quality was an independent risk factor for the occurrence of implant failure (OR, 7.614; 95% CI, 1.390–41.717; p = 0.019). This was consistent with previous studies, which revealed that fracture reduction quality was one of the paramount factors in maintaining the stability of the intertrochanteric fracture after surgical treatment (6, 28). The possible reason might be that cortical continuity could contribute to the ability of the cortex to resist collapse; whereas, Mal-angulation, in particular varus mal-angulation, might increase bone-implant stresses and risk of collapse (29, 30). The reduction quality had been described as good, acceptable, or poor according to the alignment and displacement of main fragments (14, 15). Previous studies had shown that good reduction quality could reduce the risk of mechanical failure in intertrochanteric fractures whereas poor reduction quality was associated with the occurrence of implant failure (7, 31). Therefore, great emphasis should be paid on the fracture reduction quality in the future.

The main advantage of this study was that it was the first study to evaluate the femoral head HU value on CT and its relationship to implant failure for intertrochanteric fractures after intramedullary nail fixation. Nevertheless, several limitations existed in this study. First, this was a retrospective study. Second, the sample size in this study was relatively small. Third, the ROIs for HU measurements were selected manually, which might raise concerns about the repeatability of these results. However, it had been proven that HU measurements had excellent interobserver and intraobserver reliability (32). Fourth, with the calibration of the device, the HU value could change. This was the fundamental limitation of the studies performed by HU. In further studies, it might be necessary to repeat the study using different devices and software.

Conclusion

In conclusion, we found that the HU value was significantly correlated with the incidence of implant failure and can be used as an independent factor to predict implant failure for intertrochanteric fractures after intramedullary fixation. Therefore, CT scan was a convenient and simple tool to evaluate femoral head bone quality. In addition, poor reduction quality could increase the risk of implant failure for intertrochanteric fractures after intramedullary fixation. Thus, we should pay attention to the reduction quality in the future.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by institutional ethical review board of Peking University Third Hospital, Beijing, China. The patients/participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

FZ designed the study. JF and YL were responsible for the data collection and analysis and wrote the manuscript. ZZ, HJ, YT, and YG were responsible for the data collection. ZY, GH, and XX were responsible of data analysis and interpretation. All authors contributed to the article and approved the submitted version.

Funding

This study was supported by the National Key R&D Program of China (grant no. 2018YFF0301102), National Natural Science Foundation of China (No. 81702127), and Peking University Third Hospital (grant no. Y62419-06).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Mears SC, Kates SL. A guide to improving the care of patients with fragility fractures, edition 2. Geriatr Orthop Surg Rehabil. (2015) 6:58–120. doi: 10.1177/2151458515572697

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Cotts KG, Cifu AS. Treatment of osteoporosis. JAMA. (2018) 319:1040–1. doi: 10.1001/jama.2017.21995

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Hernlund E, Svedbom A, Ivergard M, Compston J, Cooper C, Stenmark J, et al. Osteoporosis in the European Union: medical management, epidemiology and economic burden. A report prepared in collaboration with the International Osteoporosis Foundation (IOF) and the European Federation of Pharmaceutical Industry Associations (EFPIA). Arch Osteoporos. (2013) 8:136. doi: 10.1007/s11657-013-0136-1

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Burge R, Dawson-Hughes B, Solomon DH, Wong JB, King A, Tosteson A. Incidence and economic burden of osteoporosis-related fractures in the United States, 2005–2025. J Bone Miner Res. (2007) 22:465–75. doi: 10.1359/jbmr.061113

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Kaufer H. Mechanics of the treatment of hip injuries. Clin Orthop Relat Res. (1980) 146:53–61.7371269

PubMed Abstract | Google Scholar

6. Chang SM, Hou ZY, Hu SJ, Du SC. Intertrochanteric femur fracture treatment in Asia: what we know and what the world can learn. Orthop Clin North Am. (2020) 51:189–205. doi: 10.1016/j.ocl.2019.11.011

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Zhang W, Antony XR, Decruz J, Chen YD, Park DH. Risk factors for mechanical failure of intertrochanteric fractures after fixation with proximal femoral nail antirotation (PFNA II): a study in a southeast Asian population. Arch Orthop Trauma Surg. (2020) 141(4)569–75. doi: 10.1007/s00402-020-03399-2

CrossRef Full Text | Google Scholar

8. Hoffmann MF, Khoriaty JD, Sietsema DL, Jones CB. Outcome of intramedullary nailing treatment for intertrochanteric femoral fractures. J Orthop Surg Res. (2019) 14:360. doi: 10.1186/s13018-019-1431-3

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Geller JA, Saifi C, Morrison TA, Macaulay W. Tip-apex distance of intramedullary devices as a predictor of cut-out failure in the treatment of peritrochanteric elderly hip fractures. Int Orthop. (2010) 34:719–22. doi: 10.1007/s00264-009-0837-7

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Lobo-Escolar A, Joven E, Iglesias D, Herrera A. Predictive factors for cutting-out in femoral intramedullary nailing. Injury. (2010) 41:1312–6. doi: 10.1016/j.injury.2010.08.009

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Hsueh KK, Fang CK, Chen CM, Su YP, Wu HF, Chiu FY. Risk factors in cutout of sliding hip screw in intertrochanteric fractures: an evaluation of 937 patients. Int Orthop. (2010) 34:1273–6. doi: 10.1007/s00264-009-0866-2

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Anderson PA, Polly DW, Binkley NC, Pickhardt PJ. Clinical use of opportunistic computed tomography screening for osteoporosis. J Bone Joint Surg Am. (2018) 100:2073–81. doi: 10.2106/JBJS.17.01376

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Kilinc RM, Acan AE, Turk G, Kilinc CY, Yeniceri IO. Evaluation of femoral head bone quality by Hounsfield units: a comparison with dual-energy x-ray absorptiometry. Acta Radiol. (2021) 63(7):933–41.:200276715. doi: 10.1177/02841851211021035

CrossRef Full Text | Google Scholar

14. Baumgaertner MR, Solberg BD. Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br. (1997) 79:969–71. doi: 10.1302/0301-620x.79b6.7949

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Kim Y, Dheep K, Lee J, Yoon YC, Shon WY, Oh CW, et al. Hook leverage technique for reduction of intertrochanteric fracture. Injury. (2014) 45:1006–10. doi: 10.1016/j.injury.2014.02.007

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Hao Y, Zhang Z, Zhou F, Ji H, Tian Y, Guo Y, et al. Risk factors for implant failure in reverse oblique and transverse intertrochanteric fractures treated with proximal femoral nail antirotation (PFNA). J Orthop Surg Res. (2019) 14:350. doi: 10.1186/s13018-019-1414-4

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Pisani P, Renna MD, Conversano F, Casciaro E, Muratore M, Quarta E, et al. Screening and early diagnosis of osteoporosis through x-ray and ultrasound based techniques. World J Radiol. (2013) 5:398–410. doi: 10.4329/wjr.v5.i11.398

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Seo SH, Lee J, Park IH. Efficacy of dual energy x-ray absorptiometry for evaluation of biomechanical properties: bone mineral density and actual bone strength. J Bone Metab. (2014) 21:205–12. doi: 10.11005/jbm.2014.21.3.205

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Yaprak G, Gemici C, Seseogullari OO, Karabag IS, Cini N. CT derived Hounsfield unit: an easy way to determine osteoporosis and radiation related fracture risk in irradiated patients. Front Oncol. (2020) 10:742. doi: 10.3389/fonc.2020.00742

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Leonhardt Y, May P, Gordijenko O, Koeppen-Ursic VA, Brandhorst H, Zimmer C, et al. Opportunistic QCT bone mineral density measurements predicting osteoporotic fractures: a use case in a prospective clinical cohort. Front Endocrinol. (2020) 11:586352. doi: 10.3389/fendo.2020.586352

CrossRef Full Text | Google Scholar

21. Engelke K, Adams JE, Armbrecht G, Augat P, Bogado CE, Bouxsein ML, et al. Clinical use of quantitative computed tomography and peripheral quantitative computed tomography in the management of osteoporosis in adults: the 2007 ISCD official positions. J Clin Densitom. (2008) 11:123–62. doi: 10.1016/j.jocd.2007.12.010

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am. (2011) 93:1057–63. doi: 10.2106/JBJS.J.00160

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Stoffel KK, Leys T, Damen N, Nicholls RL, Kuster MS. A new technique for cement augmentation of the sliding hip screw in proximal femur fractures. Clin Biomech. (2008) 23:45–51. doi: 10.1016/j.clinbiomech.2007.08.014

CrossRef Full Text | Google Scholar

24. Fensky F, Nuchtern JV, Kolb JP, Huber S, Rupprecht M, Jauch SY, et al. Cement augmentation of the proximal femoral nail antirotation for the treatment of osteoporotic pertrochanteric fractures–a biomechanical cadaver study. Injury. (2013) 44:802–7. doi: 10.1016/j.injury.2013.03.003

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Zhang R, Wang L, Lin Y, Yang M, Guo Z, Xia W, et al. A novel method for estimating nail-tract bone density for intertrochanteric fractures. J Orthop Translat. (2019) 18:40–7. doi: 10.1016/j.jot.2018.11.005

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Shen J, Hu F, Zhang L, Tang P, Bi Z. Preoperative classification assessment reliability and influence on the length of intertrochanteric fracture operations. Int Orthop. (2013) 37:681–7. doi: 10.1007/s00264-012-1748-6

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Kregor PJ, Obremskey WT, Kreder HJ, Swiontkowski MF. Unstable pertrochanteric femoral fractures. J Orthop Trauma. (2014) 28(Suppl 8):S25–8. doi: 10.1097/BOT.0000000000000187

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Fan J, Xu X, Zhou F, Zhang Z, Tian Y, Ji H, et al. Risk factors for implant failure of intertrochanteric fractures with lateral femoral wall fracture after intramedullary nail fixation. Injury. (2021) 52(11):3397–403. doi: 10.1016/j.injury.2021.07.025

CrossRef Full Text | Google Scholar

29. Marmor M, Liddle K, Buckley J, Matityahu A. Effect of varus and valgus alignment on implant loading after proximal femur fracture fixation. Eur J Orthop Surg Traumatol. (2016) 26:379–83. doi: 10.1007/s00590-016-1746-2

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Socci AR, Casemyr NE, Leslie MP, Baumgaertner MR. Implant options for the treatment of intertrochanteric fractures of the hip: rationale, evidence, and recommendations. Bone Joint J. (2017) 99-B:128–33. doi: 10.1302/0301-620X.99B1.BJJ-2016-0134.R1

PubMed Abstract | CrossRef Full Text | Google Scholar

31. De Bruijn K, den Hartog D, Tuinebreijer W, Roukema G. Reliability of predictors for screw cutout in intertrochanteric hip fractures. J Bone Joint Surg Am. (2012) 94:1266–72. doi: 10.2106/JBJS.K.00357

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Pompe E, de Jong PA, de Jong WU, Takx RA, Eikendal AL, Willemink MJ, et al. Inter-observer and inter-examination variability of manual vertebral bone attenuation measurements on computed tomography. Eur Radiol. (2016) 26:3046–53. doi: 10.1007/s00330-015-4145-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: intertrochanteric fracture, Hounsfield units, bone quality, implant failure, intramedullary nail

Citation: Fan J, Lv Y, Xu X, Zhou F, Zhang Z, Tian Y, Ji H, Guo Y, Yang Z and Hou G (2023) Evaluation of femoral head bone quality by Hounsfield units: A predictor of implant failure for intertrochanteric fractures after intramedullary nail fixation. Front. Surg. 9:816742. doi: 10.3389/fsurg.2022.816742

Received: 17 November 2021; Accepted: 21 October 2022;
Published: 6 January 2023.

Edited by:

Chenchen Yan, Huazhong University of Science and Technology, China

Reviewed by:

Junlin Zhou, Capital Medical University, China
Peifu Tang, People's Liberation Army General Hospital, China

© 2023 Fan, Lv, Xu, Zhou, Zhang, Tian, Ji, Guo, Yang and Hou. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Fang Zhou zhouf@bjmu.edu.cn

These authors have contributed equally to this work and share first authorship.

Specialty Section: This article was submitted to Orthopedic Surgery, a section of the journal Frontiers in Surgery

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.