The application of three-dimensional printed patient-specific drilling templates for expansive open-door laminoplasty: A single-center, prospective randomized controlled study

\r\nKangkang Huang,&#x;Kangkang Huang1,†Xuelin Pan,&#x;Xuelin Pan2,†Yuting WenYuting Wen2Beiyu WangBeiyu Wang1Chen DingChen Ding1Tingkui WuTingkui Wu1Xin Rong
Xin Rong1*Hao Liu
\r\nHao Liu1*
  • 1Department of Orthopedics, West China Hospital, Sichuan University, Chengdu, China
  • 2Department of Radiology, West China Hospital, Sichuan University, Chengdu, China

Background: The choice of trough position in Expansive open-door laminoplasty (EOLP) mostly relied on bony landmarks and surgeons’ experience. The present study was to validate the efficacy of the three-dimensional (3D) printed patient-specific drilling templates with the function of locating and depth control for EOLP.

Materials and Methods: A single-center, prospective randomized controlled study was conducted on the patients who underwent unilateral EOLP from August 2019 to December 2020. The 3D printed patient-specific drilling template was fabricated and used in the template group. All the EOLP were performed by a senior surgeon and a junior surgeon. The clinical outcomes and radiographic results were evaluated.

Results: A total of 37 patients who completed the 12-month follow-up were analyzed. The clinical outcomes were significantly improved after surgery in both groups (P < 0.05). The visual analogue scale (VAS) scores were significantly lower in the template group at 12 months postoperatively (P < 0.05). The anteroposterior diameter, Pavlov's ratio and Open angle were all higher in the template group than those in the control group at 3 days and 12 months postoperatively (P < 0.05). The satisfaction of the trough position on both sides and incomplete fracture rate on the hinge side were higher in the template group based on the CT scans taken 3 days after surgery (P < 0.05). To the junior surgeon, the satisfaction and the incomplete fracture rate were significantly higher in the template group compared with those in the control group (P < 0.05).

Conclusion: The application of 3D printed patient-specific drilling templates with the function of locating and depth control for EOLP could improve the outcome of neck pain relief and expand the decompression. It can also improve the satisfaction of the trough position on the open-door side and the hinge side and decrease the complete fracture rate on the hinge side, especially for the junior surgeon.

Introduction

Expansive open-door laminoplasty (EOLP) has been widely performed for patients with cervical spondylotic myelopathy caused by multilevel disc herniation (MDH), congenital cervical stenosis (CCS) or ossification of the posterior longitudinal ligament (OPLL). Since Hirabayashi et al. (1) first described the EOLP in 1983, many authors have demonstrated that this technique is effective for neurological improvement in patients with spondylotic myelopathy (2, 3). The trough preparation is one of the most technic demanding procedures of the EOLP, requiring ideal trough position and precise bone removal. Nowadays, the choice of the trough position mostly relied on the bony landmarks and the surgeons' experience. However, complications such as complete fracture of the hinge side and insufficient decompression due to unsatisfactory trough position or excessive bone removal were reported range from 5.2%–57.6% in the literatures (46). Our previous study reported the safety and efficacy of the three-dimensional (3D) printed templates with the function of trough locating in the unilateral EOLP (79). We found that the templates could decrease the incidence of complete fracture on the hinge side through accurate locating. However, the depth of the cutting path still depended on the surgeon's experience. We subsequently modified the templates by adding the function of depth control. The trough position and cutting depth for both the hinge side and the open-door side were designed in the computer software based on the patients' computed tomography (CT) scans. The purpose of this study was to prospectively and randomly validate the efficacy of the modified 3D printed templates. The material of the template was acrylate resin which had been approved safety in patients by the United States Food and Drug Administration.

Materials and methods

This was a single-center, prospective randomized controlled study started from August 2019. The study protocol was approved by the Ethics Committee of our Hospital. The inclusion criteria were: (1) the age was older than 18 years old; (2) cervical spondylosis due to MDH, CCS or OPLL refractory to conservative treatments for at least 6 weeks; (3) the diagnosis was confirmed by imaging (CT, magnetic resonance imaging, or radiography); (4) the K line was (+); (5) the surgical plan was the unilateral EOLP from C3 to C7 with Centerpiece mini-plate (Centerpiece TM Plate Fixation System, Medtronic Sofamor Danek, USA); and (6) the patient agreed to participate in this study and gave the written informed consent. The exclusion criteria were: (1) cervical trauma, deformity, primary tumor, cervical metastasis or infection; (2) prior cervical spine surgery; (3) mental illness. The patient was rejected if the surgical plan was changed. The target sample size was 20 patients in the template group and 20 in the control group. The timepoints for evaluation were pre-operation, 3 days postoperatively and 12 months postoperatively.

The patients were randomly assigned to the template group and control group through a random number generator. Patients with odd numbers were attributed to the template group and even numbers were assigned to the control group. In the template group, the EOLP was performed with the aid of the 3D printed patient-specific drilling templates. The treatment was blinded to the radiologists who were responsible for the image taking and parameters measurement.

All the EOLP were performed by two surgeons. One was a senior surgeon with a surgical experience of more than 20 years. The other one was a junior surgeon with a surgical experience of fewer than 5 years.

EOLP template fabrication

For each patient of the template group, the preoperative CT scan was imported into the commercially available software Mimics 17.0 (Materialize, Belgium) for the 3D reconstruction. Each vertebra from C3 to C7 was reconstructed and exported as “.stl” file. The “.stl” files were processed using the 3-Matic 9.0 software (Materialize, Belgium) for surgical simulation and template designing. Two cutting paths, slightly wider than the diameter of the bur used on the hinge side and the piezosurgery used on the open-door side, were simulated at the junction between the lamina and the lateral mass. The width and depth of the cutting path on the hinge side were based on the length and diameter of the bur (Figures 1A,B). The width and depth of the cutting path on the open-door side were based on the length and thickness of the piezosurgery (Figures 1C,D). These two cutting paths were also simulated to be 5 to 10 degrees deviated from the midline, according to the surgeon's preference. Next, part of the posterior surface of the lamina was extracted to create the body of the EOLP template, which was then subtracted by the two cutting paths. Thereafter, the EOLP template was exported as “.stl” file, which was processed by the 3D printer to fabricate the actual EOLP template in acrylate resin by stereolithography. Then, the templates were thoroughly washed and vacuum dried (Figure 1E). The 3D printed templates were sterilized with ethylene oxide and stocked in plastic bags.

FIGURE 1
www.frontiersin.org

Figure 1. Template fabrication. (A,B) the measurement of the bur. (C,D) the measurement of the piezosurgery. (E) the 3D printed patient-specific drilling templates.

Surgical technique

The procedure was carried out as described previously (10). Briefly, after general anesthesia, the patient was put in the prone position with the head fixed by the Mayfield holder. A midline incision was made to expose the spinous process and bilateral lamina from C3 to C7. The supraspinous ligament and interspinous ligament were then cut before the spinous processes were amputated. Thereafter, in the template group, the template was firmly pressed to the lamina. On the hinge side, the bur was placed through the path of the 3D-printed template and the trough was made by sliding the bur along the path. On the open side, the lamina was completely cut through the cutting path with piezosurgery (Figure 2). In the control group, the hinge side and open side were cut based on the bony landmarks and the surgeons' experience. Then the lamina was opened and carefully fixed with the Centerpiece mini-plate. The Centerpiece mini-plate was anchored to the lamina and lateral mass by two 5 mm screws and two 7 mm screws, respectively.

FIGURE 2
www.frontiersin.org

Figure 2. The application of the 3D printed patient-specific drilling templates during operation.

Clinical evaluation

The neurological function was assessed by the Japanese Orthopedic Association (JOA) scores. Visual analogue scale (VAS) scores were used to evaluate neck pain intensity. The JOA and VAS scores were evaluated before surgery and at 12 months postoperatively. The occurrence of complications, including wound infection, cerebrospinal fluid (CSF) leakage, axial symptoms, and C5 palsy, were recorded and dealt with accordingly.

Radiography study

The C2–C7 Cobb angle and the C2–C7 range of motion (ROM) were measured using the Cobb method on the x-ray films before surgery and at 12 months postoperatively (11). The anteroposterior diameter of the canal and the Pavlov's ratio were determined on the lateral view of x-ray films before surgery, at 3 days postoperatively and 12 months postoperatively. The open angle was assessed using the axial view of CT scans taken 3 days postoperatively and 12 months postoperatively. The fracture type on the hinge side and the satisfaction of the trough position on the hinge side and open side were assessed using the axial view of CT scans taken 3 days postoperatively. It was deemed incomplete fracture if the ventral cortex was continuous. Otherwise, it was deemed complete fracture (6). The position of the hinge side and open-door side were evaluated in the axial view of CT scans before surgery and 3 days postoperatively. The satisfactory position was simulated at the axial view of CT scans before surgery in both groups. The final position was marked at the axial view of CT scans 3 days postoperatively. The deviation of the final position from the simulated position was recorded (Figure 3). If the deviation was less than 2 mm, it was recorded as satisfactory. Otherwise, it was recorded as unsatisfactory. The radiological evaluation was performed by two independent spine surgeons, and the mean values were used for statistical analysis.

FIGURE 3
www.frontiersin.org

Figure 3. The method for the evaluation of the position of the hinge side and open-door side. (A) the satisfactory position was simulated at the transverse view of CT scans before surgery (the yellow lines). (B) the final position was marked at the transverse view of CT scans 3 days after surgery (the red lines). (C) A and B were overlapped to evaluate the deviation of the final position from the simulated position.

Statistical analysis

Statistical analysis was performed using SPSS software (version 19.0, SPSS Inc., IL, USA). The findings are presented as the mean values ± standard deviation (SD) or counts, as indicated. The paired Student's t-test was performed to test the difference before and after surgery within the groups. The independent Student's t-test and chi-square test were performed for analyzing the differences between the two groups. A value of P < 0.05 was considered statistically significant. A chi-square test or Fisher's exact test was used to analyze categorical data between the two groups.

Results

Patient information

The study started in August 2019 and ended in December 2020. A total of 40 patients were recruited. One patient in the template group and 2 patients in the control group were excluded because of the loss of follow-up. Other patients were all followed up for 12 months. The baseline demographic information for the two groups, including age, sex, body mass index (BMI), diagnosis, operative time, blood loss, size of mini-plate, preoperative VAS of neck and preoperative JOA were evaluated, and no significant differences were observed between the characteristics in the template group and the control group (P > 0.05). The occurrence rate of complications, including CSF leakage, axial symptoms and C5 palsy was a little higher in the control group, but no significant differences were observed (P > 0.05). All the complications were recovered within 1 month after surgery. No wound infection was observed in both groups. The details are described as follows (Table 1).

TABLE 1
www.frontiersin.org

Table 1. Basic demographic information for the two groups.

Clinical results

In the template group, The VAS scores decreased significantly from 4.26 ± 1.59 preoperatively to 1.32 ± 0.86 at 12 months postoperatively (P < 0.05). The JOA scores increased significantly from 8.26 ± 2.16 preoperatively to 15.00 ± 1.73 at 12 months postoperatively (P < 0.05). In the control group, the VAS scores decreased significantly from 4.50 ± 1.29 to 2.06 ± 1.21, and the JOA scores increased significantly from 8.50 ± 2.09 to 14.61 ± 1.33 (P < 0.05). The significant difference of VAS scores between the two groups was observed at 12 months postoperatively (P < 0.05). The details are described as follows (Table 2).

TABLE 2
www.frontiersin.org

Table 2. Comparison of the clinical outcomes between the two groups.

Radiographic results

The C2–C7 Cobb angle and the C2–C7 ROM were decreased with no significant difference at 12 months postoperatively compared with that before surgery, and no significant difference was observed between the two groups (P > 0.05). The anteroposterior diameter and Pavlov's ratio increased significantly after surgery and remained well during the follow-up in both groups (P < 0.05). Besides, the data increased more in the template group compared with that in the control group with significantly differences at 3 days postoperatively and 12 months postoperatively (P < 0.05). The Open angle was also significantly higher in the template group compared with that in the control group at 3 days postoperatively and 12 months postoperatively (P < 0.05). The details are described as follows (Table 3).

TABLE 3
www.frontiersin.org

Table 3. Comparison of the clinical outcomes between the two groups.

A total of 185 lamina were analyzed. Based on the CT scans taken 3 days after surgery, 37 out of 90 (41.11%) were completely fractured on the hinge side in the control group, which was significantly higher compared with that in the template group (P < 0.05) (Table 4). The satisfaction of the trough position on both sides was 87.37% (166/190) in the template group, which was significantly higher than that in the control group (P < 0.05) (Table 5).

TABLE 4
www.frontiersin.org

Table 4. Comparison of the fracture type between the two groups.

TABLE 5
www.frontiersin.org

Table 5. Comparison of the satisfaction of the trough position on both sides between the two groups.

Analysis in senior and junior surgeons

Nine patients in the template group and 9 patients in the control group were performed by the senior surgeon, and 10 patients in the template group and 9 patients in the control group were performed by the junior surgeon. The incomplete fracture rate and satisfaction of the trough position on both sides had no significant difference between the two groups when the EOLP was performed by the senior surgeon (P > 0.05). However, to the junior surgeon, the incomplete fracture rate and satisfaction were significantly higher in the template group compared with that in the control group (P < 0.05). The details are described as follows (Tables 6, 7).

TABLE 6
www.frontiersin.org

Table 6. Comparison of the fracture type between the two surgeons.

TABLE 7
www.frontiersin.org

Table 7. Comparison of the satisfaction of the trough position on both sides between the two surgeons.

Discussion

The making of the open-door side cutting path and hinge side trough was the most technically demanding procedure during the EOLP surgery. The proper position and depth of both sides were reported to be related to the clinical outcomes and complications (12, 13). Ideally, the open-door side cutting path should be made at the junction between the lamina and the lateral mass with the cutting off of the ventral cortex appropriately. The hinge side trough should be made at the junction between the lamina and the lateral mass with the dorsal cortex and cancellous bone removed, keeping the ventral cortex intact. However, achieving such a target is technically demanding. Accurately locating the position and appropriately cutting the bone could decrease the incidence of hinge side complete fracture, achieve sufficient decompression and ensure the safety of the surgery (1416).

The complete fracture of the lamina on the hinge side compromised the stability of the lamina (17). Under the push and pull effect of the posterior cervical soft tissue, the floppy lamina could be displaced. The displacement of the lamina into the spinal canal could compress the spinal cord or encroach the nerve root, causing deterioration of the neurological functions. We previously reported the incidence of complete fracture to be 56.8%, among which 3.68% of the lamina displaced into the canal (6). Although no spinal disorders were detected in these patients during the follow-up, longer observation was still needed.

The 3D printing technique has been recently applied in surgery (18, 19). This technique has been tested both in vitro and clinically for posterior cervical surgery, especially for the pedicle screw placement (20, 21). In these cases, the entry point and trajectory of the screws were planned before surgery, and the templates were modeled in the computer software. The templates can be designed for a specific patient according to his or her anatomical features (7). Thus, it is especially useful for the surgeries performed on the site of anatomical abnormality or variation. In our study, the 3D printed patient-specific drilling template could help the surgeon locate the trough position and control the depth more properly.

With the aid of the 3D printed templates with the function of locating and depth control, the patients could achieve a better relief of the neck pain. That may attribute to the accurate slotting on the hinge side which achieved a high fusion rate at 12 months postoperatively. Besides, the anteroposterior diameter, Pavlov's ratio and Open angle were all increased in the template group, which meant the decompression was more expansive with the aid of the templates. The satisfaction of the trough position on the open-door side and hinge side and the incomplete fracture rate on the hinger side were higher in the template group, which meant the templates used in the template group could improve the accuracy of locating and depth control. This phenomenon may not be outstanding to the senior surgeon because of his experienced technique. But to the junior surgeon, the effect was remarkable.

There is a limitation in our study. This is a single-center, prospective randomized controlled study with a relatively small sample size. In the future, a multi-center prospective randomized controlled study with a larger sample size should be performed to further validate the results of our study.

Conclusion

The application of 3D printed patient-specific drilling templates with the function of locating and depth control for EOLP could improve the outcome of neck pain relief and expand the decompression. It can also improve the satisfaction of the trough position on the open-door side and the hinge side and decrease the complete fracture rate on the hinge side, especially for the junior surgeon.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by the medical ethics committee of West China Hospital of Sichuan University (No. 2019-599). The patients/participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

All authors contributed to the study conception and design. Material preparation, data collection and analysis were performed by KH, HL, XR, BW, CD, TW. The radiological examinations and radiographic measurements were carried by XP and YW. The first draft of the manuscript was written by KH and XR and all authors commented on previous versions of the manuscript. All authors contributed to the article and approved the submitted version.

Funding

This study was supported by the Sichuan Province Science and Technology Program of China (Grant No. 2018SZ0176) and the National Natural Science Foundation of China (NSFC) (Grant No. 81902190 and Grant No. 82172522).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Abbreviations

EOLP, expansive open-door laminoplasty; MDH, multilevel disc herniation (MDH); CCS, congenital cervical stenosis (CCS); OPLL, ossification of the posterior longitudinal ligament; 3D, three-dimensional; CT, computed tomography; JOA, Japanese orthopedic association; VAS, visual analog scale; ROM, range of motion; CSF, cerebrospinal fluid; SD, standard deviation; BMI, body mass index.

References

1. Hirabayashi K, Watanabe K, Wakano K, Suzuki N, Satomi K, Ishii Y. Expansive open-door laminoplasty for cervical spinal stenotic myelopathy. Spine. (1983) 8(7):693–9. doi: 10.1097/00007632-198310000-00003

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Satomi K, Nishu Y, Kohno T, Hirabayashi K. Long-term follow-up-studies of open-door expansive laminoplasty for cervical stenotic myelopathy. Spine. (1994) 19(5):507–10. doi: 10.1097/00007632-199403000-00003

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Chiba K, Ogawa Y, Ishii K, Takaishi H, Nakamura M, Maruiwa H, et al. Long-term results of expansive open-door laminoplasty for cervical myelopathy—average 14-year follow-up study. Spine. (2006) 31(26):2998–3005. doi: 10.1097/01.brs.0000250307.78987.6b

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Li H, Dai LY. A systematic review of complications in cervical spine surgery for ossification of the posterior longitudinal ligament. Spine J. (2011) 11(11):1049–57. doi: 10.1016/j.spinee.2011.09.008

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Seichi A, Hoshino Y, Kimura A, Nakahara S, Watanabe M, Kato T, et al. Neurological complications of cervical laminoplasty for patients with ossification of the posterior longitudinal ligament-a multi-institutional retrospective study. Spine. (2011) 36(15):E998–E1003. doi: 10.1097/BRS.0b013e3181fda7fa

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Chen H, Liu H, Zou L, Li T, Gong Q, Song YM, et al. Effect of Mini-plate fixation on hinge fracture and bony fusion in unilateral open-door cervical expansive laminoplasty. Clin Spine Surg. (2016) 29(6):E288–E95. doi: 10.1097/BSD.0000000000000131

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Huang KK, Rong X, Liu H, Wang BY, Chen H, Hong Y. Safety and efficacy of three-dimensional printed patient-specific drilling templates for expansive open-door laminoplasty. Chin Med J. (2019) 132(21):2641–3. doi: 10.1097/cm9.0000000000000476

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Rong X, Wang BY, Deng YX, Liu H. Three-dimensional printed patient-specific drilling templates for expansive open-door laminoplasty. Chin Med J. (2016) 129(21):2644–6. doi: 10.4103/0366-6999.192786

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Rong X, Wang BY, Chen H, Ding C, Deng YX, Ma LP, et al. Use of rapid prototyping drill template for the expansive open door laminoplasty: a cadaveric study. Clin Neurol Neurosurg. (2016) 150:13–7. doi: 10.1016/j.clineuro.2016.08.013

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Chen H, Li HB, Deng YX, Rong X, Gong Q, Li T, et al. Optimal area of lateral mass mini-screws implanted in plated cervical laminoplasty: a radiography anatomy study. Eur Spine J. (2017) 26(4):1140–8. doi: 10.1007/s00586-016-4785-9

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Takeshita K, Murakami M, Kobayashi A, Nakamura C. Relationship between cervical curvature index (Ishihara) and cervical spine angle (C2–7). J Orthop Sci. (2001) 6(3):223–6. doi: 10.1007/s007760100038

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Lee JH, Chough CK. Risk factors for hinge fracture associated with surgery following cervical open-door laminoplasty. Korean J Neurotrauma. (2018) 14(2):118–22. doi: 10.13004/kjnt.2018.14.2.118

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Hur JW, Park YK, Kim BJ, Moon HJ, Kim JH. Risk factors for delayed hinge fracture after plate-augmented cervical open-door laminoplasty. J Korean Neurosurg Soc. (2016) 59(4):368–73. doi: 10.3340/jkns.2016.59.4.368

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Fujimura Y, Nishi Y, Chiba K, Nakamura M, Hirabayashi K. Multiple regression analysis of the factors influencing the results of expansive open-door laminoplasty for cervical myelopathy due to ossification of the posterior longitudinal ligament. Arch Orthop Trauma Surg. (1998) 117(8):471–4. doi: 10.1007/s004020050296

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Tetreault L, Ibrahim A, Cote P, Singh A, Fehlings MG. A systematic review of clinical and surgical predictors of complications following surgery for degenerative cervical myelopathy. J Neurosurg-Spine. (2016) 24(1):77–99. doi: 10.3171/2015.3.spine14971

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Cho SK, Kim JS, Overley SC, Merrill RK. Cervical laminoplasty: indications, surgical considerations, and clinical outcomes. J Am Acad Orthop Surg. (2018) 26(7):E142–E52. doi: 10.5435/jaaos-d-16-00242

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Park YK, Lee DY, Hur JW, Moon HJ. Delayed hinge fracture after plate-augmented, cervical open-door laminoplasty and its clinical significance. Spine J. (2014) 14(7):1205–13. doi: 10.1016/j.spinee.2013.08.012

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Waran V, Narayanan V, Karuppiah R, Owen SLF, Aziz T. Utility of multimaterial 3d printers in creating models with pathological entities to enhance the training experience of neurosurgeons technical note. J Neurosurg. (2014) 120(2):489–92. doi: 10.3171/2013.11.jns131066

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Kusaka M, Sugimoto M, Fukami N, Sasaki H, Takenaka M, Anraku T, et al. Initial experience with a tailor-made simulation and navigation program using a 3-D printer model of kidney transplantation surgery. Transplant Proc. (2015) 47(3):596–9. doi: 10.1016/j.transproceed.2014.12.045

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Martelli N, Serrano C, van den Brink H, Pineau J, Prognon P, Borget I, et al. Advantages and disadvantages of 3-dimensional printing in surgery: a systematic review. Surgery. (2016) 159(6):1485–500. doi: 10.1016/j.surg.2015.12.017

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Tack P, Victor J, Gemmel P, Annemans L. 3d-Printing Techniques in a medical setting: a systematic literature review. Biomed Eng Online. (2016) 15:21. doi: 10.1186/s12938-016-0236-4

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: laminoplasty, three-dimensional printing, template, hinge, cervical spondylosis

Citation: Huang K, Pan X, Wen Y, Wang B, Ding C, Wu T, Rong X and Liu H (2023) The application of three-dimensional printed patient-specific drilling templates for expansive open-door laminoplasty: A single-center, prospective randomized controlled study. Front. Surg. 9:1084804. doi: 10.3389/fsurg.2022.1084804

Received: 31 October 2022; Accepted: 19 December 2023;
Published: 9 January 2023.

Edited by:

Aikeremujiang Muheremu, Sixth Affiliated Hospital of Xinjiang Medical University, China

Reviewed by:

Xuexiao Ma, The Affiliated Hospital of Qingdao University, China
Pei Li, Third Affiliated Hospital of Chongqing Medical University, China

© 2023 Huang, Pan, Wen, Wang, Ding, Wu, Rong and Liu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Hao Liu dr.liuhao6304@hotmail.com Xin Rong rongxin@scu.edu.cn

These authors have contributed equally to this work and share first authorship

Specialty Section: This article was submitted to Orthopedic Surgery, a section of the journal Frontiers in Surgery

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.