ORIGINAL RESEARCH article

Front. Sports Act. Living, 17 December 2020
Sec. Sports Science, Technology and Engineering
This article is part of the Research Topic Wearable Sensors for Remote Health Monitoring and Intelligent Disease Management View all 9 articles

The Validity, Reliability, and Sensitivity of a Smartphone-Based Seated Postural Control Assessment in Wheelchair Users: A Pilot Study

  • 1Motor Control Research Laboratory, Department of Kinesiology and Community Health, University of Illinois at Urbana Champaign, Urbana, IL, United States
  • 2Illinois Multiple Sclerosis Research Collaborative, University of Illinois at Urbana Champaign, Urbana, IL, United States
  • 3Disability Participation & Quality of Life Research Laboratory, Department of Kinesiology and Community Health, University of Illinois at Urbana Champaign, Urbana, IL, United States

Seated postural control is essential for wheelchair users to maintain proper position while performing activities of daily living. Clinical tests are commonly used to measure seated postural control, yet they are subjective and lack sensitivity. Lab-based measures are highly sensitive but are limited in scope and restricted to research settings. Establishing a valid, reliable, and accessible measurement tool of seated postural control is necessary for remote, objective assessments. Therefore, the purpose of this study was to examine the validity, reliability, and sensitivity of smartphone-based postural control assessments in wheelchair users. Eleven participants (age: 35.4 ± 17.9) completed two experimental visits 1-week apart consisting of three clinical tests: Trunk Control Test (TCT), Function in Sitting Test (FIST), and Tee-shirt Test, as well as, standardized instrumented balance tasks that manipulated vision (eyes open and closed), and trunk movement (functional reach and stability boundary). During these tasks, participants held a smartphone instrumented with a research-grade accelerometer to their chest. Maximum and root mean square (RMS) acceleration in the medial-lateral (ML) and anterior-posterior (AP) axes were derived. Participants were grouped into non-impaired and impaired postural groups based on FIST scores. Spearman rank-order correlations were conducted between the two devices' outcome measurements and between these measures and those of the clinical tests. Receiver operating characteristic (ROC) curves and the area under the curves (AUC) were determined to distinguish participants with and without impaired postural control. The reliability of outcome variables was assessed using inter-class correlations. Strong correlations between outputs derived from the smartphone and research-grade accelerometer were seen across balance tasks (ρ = −0.75–1.00; p ≤ 0.01). Numerous significant moderate correlations between clinical test outcomes and smartphone and research-grade RMS ML accelerometry were seen (ρ = −0.62 to 0.83 (p ≤ 0.044)]. On both devices, the AUC for ROC plots were significant for RMS ML sway during the eyes open task and functional stability boundary (p < 0.05). Reliability of smartphone accelerometry was comparable to the research-grade accelerometer and clinical tests. This pilot study illustrated that smartphone-based accelerometry may be able to provide a valid and reliable assessment of seated postural control and have the ability to distinguish between those with and without impaired postural control.

Introduction

It is currently estimated that there are ~65 million wheelchair users worldwide (Physiopedia Contributors, 2018). Of these, ~3.3 million reside in the United States of America, where researchers are expecting annual growth of new users due to the exponential growth of older adults (Karmarkar et al., 2011). Wheelchair users face numerous challenges to maintaining an active and engaged life, which can be exacerbated by impaired seated postural control. Seated postural control is the ability to maintain one's center of mass within stability boundaries while in a seated position, and is comprised of a complex interplay of sensory processing and motor outputs (Ivanenko and Gurfinkel, 2018; Barbado et al., 2019). Alterations to sensory or motor processing can result in a decline in seated postural control (Shin and Sosnoff, 2013), and jeopardize an individual's ability to safely perform activities of daily living (Rice et al., 2015). As such, improving seated postural control is a common goal of rehabilitation interventions (Williams and Vette, 2019). Consequently, objectively measuring seated postural control in wheelchair users is necessary to guide prevention and rehabilitative strategies.

There are numerous ways to measure seated postural control. Researchers have developed several clinical measures including, but not limited to, the Function in Sitting Test (Gorman et al., 2010; Sung et al., 2016; Abou et al., 2019), Trunk Control Test (TCT) (Quinzaños et al., 2014), and the Tee-shirt Test (Boswell-Ruys et al., 2009) to assess seated postural control. These clinical measures have few technological requirements but require clinical expertise to perform. There are also concerns that these measures are subjective and lack sensitivity (Nguyen et al., 2012). Researchers have also utilized three-dimensional motion capture techniques (Murans et al., 2011; Curtis et al., 2015), video-based measurements (Sánchez et al., 2017), post-urography (Murans et al., 2011; Shin and Sosnoff, 2013), and research-grade accelerometers (Kim et al., 2018) to assess seated postural control. These research lab-based measures are objective and sensitive to impairment but require relatively expensive technology, expertise, and are limited to lab-based settings. Establishing an objective valid and reliable measurement tool to understand and monitor seated postural control is warranted (Rice et al., 2015).

A possible avenue for achieving objective accessible measures of seated postural control is through the utilization of mobile technology. Indeed, researchers have leveraged mobile health technology, specifically smartphone and tablet embedded sensors, to assess standing postural control (Roeing et al., 2017; Reyes et al., 2018). Recent work has shown that mobile technology is a valid (Cerrito et al., 2015; Hsieh et al., 2019) and reliable (Mellone et al., 2012; Cerrito et al., 2015) tool to provide objective assessments of standing balance, have a high level of usability (Hsieh et al., 2018) and are sensitive to impairment (Hsieh et al., 2019). Although promising, the validity, reliability, and sensitivity of smartphone-based seated postural control assessments in wheelchair users have not yet been investigated. Therefore, the purpose of the current pilot study is to determine the validity, reliability, and sensitivity of smartphone-based seated postural control assessments in adult wheelchair users, as an initial step in the remote monitoring of seated postural control. Based on previous standing balance research, we hypothesized that smartphone-based accelerometry can provide a valid and reliable measure of seated postural control and have the ability to distinguish between those with and without impaired postural control.

Materials and Methods

Participants

Eleven non-ambulatory adults (age: 35.4 ± 17.9 years; gender: 4 males, 7 females) were recruited from the local community to participate in the current study. To be eligible, individuals were required to be ≥18 years old, utilize a wheeled mobility device for their main form of mobility, manual dexterity sufficient to swipe on a smartphone, normal or corrected to normal hearing and vision, and able to read and speak English. Individuals were excluded from the study if they were unable to meet these criteria or if they were unable to sit upright for at least 1-h. The University of Illinois at Urbana-Champaign Institutional Review Board approved all procedures, and all participants provided written informed consent before engaging in research activities.

Research Protocol and Data Analyses

Participants completed two identical experimental sessions, one week apart. The first experimental session began by obtaining written informed consent and participant demographic information. At each session, participants completed three clinical tests that have been shown to provide a valid measure of seated posture control: the Function in Sitting Test (FIST) (Abou et al., 2019), Trunk Control Test (TCT) (Quinzaños et al., 2014), and Tee-shirt Test (Boswell-Ruys et al., 2009). An individual's maximum forward and lateral reach distances (cm) were recorded during the FIST and also used as a clinical outcome, as they are an indication of seated balance (Lynch et al., 1998; Boswell-Ruys et al., 2009; Kalron et al., 2016). Following these tests, participants completed a series of unsupported seated balance tasks on a flat surface while holding a smartphone (Samsung Galaxy S6, Samsung, Seoul, South Korea) with a research-grade accelerometer affixed to the back with double-sided tape in a standardized position and orientation (The Opal, APDM Wearable Technologies, Portland, OR) (see Figure 1). In this arrangement, there were no gaps between the accelerometer and smartphone. Pilot testing informed the standardized alignment, positioning, and ensured a rigid adhesion of the research-grade accelerometer to the smartphone. This methodological approach is based in part on previous research examining the use of smartphones to access standing postural control (Hsieh et al., 2019).

FIGURE 1
www.frontiersin.org

Figure 1. Illustration of two participants completing the seated postural control assessment. (A) represents a participant facing the side, (B) represents a participant facing forward, (C) depicts the research-grade accelerometer affixed to the back of the smartphone and the aligned axes orientation (ML, medial-lateral; AP, anterior-posterior), and (D) provides an example plot of accelerometry data from the smartphone and research-grade accelerometer, recorded during the Functional Stability Boundary Test.

Four seated balance tasks were completed in a standardized order that increased in difficulty: static sitting with eyes open (EO), static sitting with eyes closed (EC), functional reach (FR), and functional stability boundary (FSB). These tests were chosen because of their ability to provide insight into those with and without impaired postural control (Shin and Sosnoff, 2013).

All tests except for the functional reach task were completed for 30 s. The functional reach task was not constrained by time. Two trials of each task were completed. During EO, EC, and FSB balance tasks, participants held the smartphone with their dominant hand against their sternum and in standardized orientation (see Figure 1). This standardized orientation was also used during the functional reach task except for the phone being held in the participant's non-dominated hand. Proper holding and placement of the smartphone were modeled for the participants by the research staff. Consistency of alignment (anterior-posterior and medial-lateral axes) between the smartphone and participant were visually monitored by trained research personnel within and between testing sessions. If participants did not hold the phone correctly, trials were removed from analyses. With findings from the current study informing the development of a self-administered mobile health application, the methods of holding the smartphone in place of fixing it to the individual are imperative when accounting for “real-life” use and error. The smartphone was sampled at an average rate of 200 Hz and the research-grade accelerometer was collected at 128 Hz.

A custom MATLAB script (MathWorks Inc., Natick, MA) was used to downsample and temporally and spatially align (with respect to axes and gravity) all accelerometry data. Data were downsampled to 100 Hz and filtered using a 4th-order lowpass Butterworth filter set to 10 Hz. The time-series used for analysis was 30 s. Maximum (MAX) and root mean squared (RMS) acceleration time-series from each device along the anterior-posterior (AP) and medial-lateral (ML) axes, as well as the 95% confidence ellipse area (CEA), were calculated. These measures are seen to be a valid assessment of postural stability (Ozinga et al., 2015) and sensitive enough to identify impairment in other populations (Galán-Mercant and Cuesta-Vargas, 2014; Ozinga et al., 2015; Hsieh et al., 2019).

Statistical Analysis

IBM Statistical Package for the Social Sciences (SPSS) for Windows, version 26 (IBM Corp, Armonk, NY) was used to complete all statistical analyses with statistical significance set at α = 0.05. All values, with exception of area under the curve, are reported as mean ± standard deviation. Extreme outliers, as defined as, any data values which lie more than 3.0 times the interquartile range were removed from the data set. Within the first and second sessions, extreme outliers made up 0.97 and 1.29% of the data, respectively. Once extreme outliers were removed, the two trials of each balance task (EO, EC, FR, and FSB) from a given session were averaged together. To assess the validity of the smartphone measurements, Spearman rank-order correlations between the smartphone and research-grade accelerometer were determined for all conditions. To understand the validity of these measurements in assessing seated postural control, Spearman rank-order correlations were completed between the smartphone and research-grade accelerometer to clinical measure outcomes. Correlation coefficients of 0.1 were considered small, 0.3 were considered moderate, and 0.5 were considered large (Cohen et al., 2003). The reliability of the smartphone, research-grade accelerometer, and clinical tests were measured by conducting interclass correlations (ICC) of their respective outcome variables from session 1 and session 2. Values <0.5 were considered poor, between 0.5 and 0.75 were considered moderate, between 0.75 and 0.9 were considered good, and >0.90 were indicative of excellent reliability (Koo and Li, 2016).

A median split of FIST scores was used to separate participants into two groups: those with and without impaired seated postural control. Once separated, independent sample T-tests were performed to identify potential differences in age and clinical test outcome measures during all balance conditions. To further understand the difference between the two groups, the effect sizes (Cohen's d) were calculated. Effect sizes were classified as small (d = 0.20), medium (d = 0.50), and large (d = 0.80) (Jacob, 1988). To determine the sensitivity of the smartphone and research-grade accelerometer, receiving operating characteristic (ROC) curves were constructed and the area under the curve (AUC) was calculated for RMS ML, RMS AP, and CEA to determine the classification accuracy of those with and without impaired seated postural control.

Results

Results from the research grade accelerometer indicated that MAX ML acceleration ranged from −6.92 to 5.47 m/s2 and had mean value of −0.68 ± 2.16 m/s2, and MAX AP acceleration ranged from 0.68 to 8.66 m/s2 and had mean value of 5.37 ± 2.09 m/s2. RMS ML acceleration ranged from 0.22 to 3.98 m/s2 and had mean value of 1.16 ± 0.86 m/s2, and RMS AP acceleration ranged from 0.24 to 7.47 m/s2 and had mean value of 4.13 ± 1.87 m/s2.

As for the smartphone, MAX ML acceleration ranged from −0.99 to 6.71 m/s2 and had mean value of 1.79 ± 1.64 m/s2, and MAX AP acceleration ranged from −8.28 to 7.46 m/s2 and had mean value of 2.11 ± 3.13 m/s2. RMS ML acceleration ranged from 0.16 to 4.16 m/s2 and had mean value of 1.18 ± 0.88 m/s2, and RMS AP acceleration ranged from 0.39 to 7.73 m/s2 and had mean value of 4.31 ± 1.90 m/s2.

Validity

Spearman rank-order correlations between the smartphone and research-grade accelerometer outcome variables revealed numerous significant relations. Maximum acceleration along the ML (EO and EC) (p ≤ 0.01) and AP (EO, EC, and FR) (p ≤ 0.01) axes were significantly correlated between devices (see Table 1). Measures of RMS acceleration and CEA yielded strong, significant correlations between the two devices (p ≤ 0.011), except for ML acceleration during the eyes closed balance task (p = 0.124) (see Table 1).

TABLE 1
www.frontiersin.org

Table 1. Presents the correlations (Rho) of maximum (MAX) and root mean square (RMS) acceleration as derived through smartphone and research-grade accelerometry.

Spearman rank-order correlations were also completed between clinical test outcomes and smartphone and research-grade accelerometer outcome variables during the seated balance tasks (i.e., EO, EC, FR, and FSB). Overall, numerous significant correlations were seen between the clinical tests and RMS accelerometry on both devices in the ML axis (see Supplementary Material). Smartphone and research-grade RMS ML accelerometry yielded strong, positive, significant correlations with the clinical tests. Rho values from smartphone accelerometry ranged from 0.642 (p = 0.033) to 0.807 (p ≤ 0.01) (see Table 2), while research-grade accelerometry ranged from 0.615 (p = 0.044) to 0.826 (p ≤ 0.01). The TCT was the only clinical test to not display a significant association with RMS ML acceleration (see Table 2).

TABLE 2
www.frontiersin.org

Table 2. Presents the correlations (Rho) between clinical test outcomes and root mean square (RMS) acceleration as derived through smartphone and research-grade accelerometry.

Reliability

Reliability was seen across all clinical measurements (p ≤ 0.005), except the TCT (p = 0.077) (see Table 3). As for accelerometry, 55% of the smartphone and 70% of research-grade accelerometer outcome variables were found to be reliable (p < 0.05) (see Table 4). The smartphone was the most reliable across outcome variables during the EC balance test, while the research-grade accelerometer was equally reliable during the EC and FSB tests (see Table 4).

TABLE 3
www.frontiersin.org

Table 3. Interclass correlations (ICC) between clinical test outcomes during session 1 and session 2.

TABLE 4
www.frontiersin.org

Table 4. Interclass correlations (ICC) between maximum (MAX) acceleration, root mean squared (RMS) acceleration, and confidence ellipse area (CEA) as recorded through smartphone and research-grade accelerometry during session 1 and session 2.

Sensitivity

To determine sensitivity, the 11 participants were separated into two groups, those with (n = 5) and without (n = 6) impaired seated postural control (see Table 5). Per design, group differences were observed in the FIST (p = 0.009) as well as the TCT performance (p = 0.023). The effect sizes ranged from small to large (d: −0.40 to −2.59).

TABLE 5
www.frontiersin.org

Table 5. Participant demographic information.

To distinguish individuals with and without impaired seated postural control, ROC curves were constructed, and AUC was calculated for smartphone and research-grade accelerometry RMS ML, RMS AP, and CEA (see Table 6). The AUC for smartphone RMS ML ranged from 0.433 ± 0.188 to 0.933 ± 0.078, RMS AP ranged from 0.500 ± 0.186 to 0.667 ± 0.174, and CEA ranged from 0.467 ± 0.209 to 0.800 ± 0.144 (values are mean ± SE). Research-grade accelerometry yielded similar findings with the AUC for RMS ML ranging from 0.767 ± 0.149 to 0.900 ± 0.104, RMS AP ranging from 0.500 ± 0.186 to 0.600 ± 0.181, and CEA ranging from 0.367 ± 0.182 to 0.833 ± 0.128 (values are mean ± SE). The AUC was statistically significant for smartphone and research-grade accelerometry RMS ML sway during the EO (p = 0.045, both) and FSB (p = 0.018 and 0.028, respectively) balance tasks (see Table 6 and Figure 2). The AUC was also statistically significant for research-grade accelerometry RMS ML sway during the EC balance task (p = 0.045).

TABLE 6
www.frontiersin.org

Table 6. Receiver operating characteristic curve statistical outcomes.

FIGURE 2
www.frontiersin.org

Figure 2. Receiver operating characteristic (ROC) curves for root mean squared smartphone (A) and research-grade accelerometer (B) acceleration in the medial-lateral (ML) direction during all balance tasks.

Discussion

Understanding the validity, reliability, and sensitivity of smartphone-based seated postural control during various balance tasks is critical in the efforts of providing wheelchair users with an objective and accessible tool to measure seated postural control. Within the current study, smartphone-based measures of seated postural control were found to be valid, have reliability that was on par or greater than the clinical tests, and capable of discriminating between individuals with and without impaired seated postural control. Collectively, the observations provide preliminary evidence that smartphone-based accelerometry is suitable for objectively measuring seated postural control in adult wheelchair users.

Due to the strong significant correlations between outputs derived from the smartphone and research-grade accelerometer, the current investigation indicates that the smartphone-based accelerometry is potentially as valid as a research-grade accelerometer. This is in line with recent studies which illustrated that smartphone accelerometry provided a valid measure of standing postural stability when compared against research-grade equipment (Cerrito et al., 2015; Hsieh et al., 2019).

Quantifying seated postural control has been a topic of scientific interest, utilizing a wide array of technology (e.g., three-dimensional motion capture, video-based measures, and force plate measures; Murans et al., 2011; Shin and Sosnoff, 2013; Curtis et al., 2015; Sánchez et al., 2017). Accelerometry has been used to evaluate the movement of transfers in adult wheelchair users (Barbareschi et al., 2018), yet limited work has utilized this technology to quantify seated postural control, resulting in limited recommendations concerning how to best quantify the acceleration signal. Research focusing on standing balance has recommended the use of RMS as the “best” measure (Kosse et al., 2015; Ozinga et al., 2015; Hsieh et al., 2019). Consistent with these recommendations, the current investigation observed numerous strong correlations between smartphone and research-grade accelerometry when RMS quantified the signal. Few significant relationships were observed between MAX measures of acceleration. This is in agreement with the discrepancy in MAX accelerometry ranges presented from each device. Similar findings were also seen when assessing the relationship between smartphone accelerometry and clinical measure outcomes. Collectively this supports the notion that RMS of acceleration is a valid measure of seated postural control and should be incorporated into future study designs investigating accelerometry-based seated postural control.

Along with identifying the validity of accelerometry-based movement and balance tasks, past investigations have shown this form of technology to be reliable (Mellone et al., 2012; Cerrito et al., 2015; Kosse et al., 2015; Silsupadol et al., 2017; Douma et al., 2018). Consistent with previous research, findings from the current study indicate that the RMS of smartphone acceleration is as reliable as that of a research-grade accelerometer.

Clinical tests, particularly the FIST and TCT, have been reported as reliable measures of seated postural control in clinical populations (Quinzaños et al., 2014; Abou et al., 2019). The current results confirm the reliability of the FIST, Tee-shirt Test, Forward Reach, and Lateral Reach. The level of reliability of smartphone-based accelerometry was on par with or greater than those of the clinical tests. Such observations further support the notion that smartphone technology is a reliable and objective measurement of seated postural control for wheelchair users.

In order to provide meaningful results, smartphone technology must have the sensitivity to differentiate between those with varying degrees of postural control. In the past, smartphone accelerometry has been able to do discriminate between standing postural control in frailty (frail/non-frail) (Galán-Mercant and Cuesta-Vargas, 2014) and fall risk (Hsieh et al., 2019) within older adults. Within the current study, measurements from both devices during the easiest (EO) and most challenging (FSB) balance tasks were able to identify participants with and without impaired postural control, specifically in the ML direction. This is further supported by the numerous strong, positive, significant correlations between RMS ML smartphone and research-grade accelerometry and clinical measure outcomes. Although ML postural control has been implicated in impairment in standing balance (Sosnoff et al., 2011), we believe this is one of the first investigations to highlight this in seated postural control. Recent work also supports this observation by providing evidence that those with impaired seated postural control exhibit greater decrements in their lateral (ML) reach than forward (AP) reach (Abou et al., 2019). These collective findings indicate that smartphone technology may have the sensitivity to identify those with and without impaired seated postural control and that postural instability within wheelchair users may be rooted in mediolateral instability.

Conclusion

To better understand seated postural control, and monitor changes over time, we must establish an objective and sensitive measurement tool. To our knowledge, this is the first investigation examining the validity, reliability, and sensitivity of smartphone-based accelerometry as a tool to quantify seated postural control in adult wheelchair users. Results from this study illustrated that smartphone technology may be able to provide a valid and reliable assessment of seated postural control and have the ability to distinguish between those with and without impaired postural control—especially in the ML plane. Given the ubiquitous nature of smartphones in society, there is great potential for mobile technology to provide quick, easily accessible, and objective remote monitoring of seated postural control in adult wheelchair users.

Limitations and Future Directions

Limitations of the current study include limited sample size, albeit with a diverse range of seated postural control, the use of a single smartphone and research-grade accelerometer, and the potential introduction of motion artifact with the use of a handheld accelerometer. Future research should incorporate a larger sample to assess the relationship between clinical scores and accelerometry, further investigate the reliability of accelerometry based seated postural control assessments and the feasibility of leveraging this form of technology for remote assessment, and examine the difference between a sensor worn on the body and a handheld sensor. Along with this, researchers need to develop a health application interface to provide this type of assessment and determine its usability, validity, and reliability of results, responsiveness to interventions (i.e., sensitivity to changes in seated balance), and home use acceptance.

Data Availability Statement

The datasets generated for this study are available on request to the corresponding author.

Ethics Statement

The studies involving human participants were reviewed and approved by the University of Illinois at Urbana-Champaign Office for the Protection of Research Subjects. The patients/participants provided their written informed consent to participate in this study.

Author Contributions

LR and JS: conceptualization and funding acquisition. MF and LA: data collection. MF: organized database, software coding, and statistical analysis. JS: provided resources and project administration. MF and JS: writing—original draft. All authors contributed to the methodology, manuscript revision, and read and approved the submitted version.

Funding

This work was supported by the National Institute on Disability, Independent Living, and Rehabilitation Research (NIDILRR) under Grant ##90REGE0006-01-00. MF was funded by MS Run the US Scholarship.

Conflict of Interest

JS has partial ownership in Sosnoff Technologies, LLC a company that may be affected by the research reported in the enclosed paper. This conflict of interest is managed by a plan approved by the University of Illinois at Urbana-Champaign.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fspor.2020.540930/full#supplementary-material

References

Abou, L., Sung, J., Sosnoff, J. J., and Rice, L. A. (2019). Reliability and validity of the function in sitting test among non-ambulatory individuals with spinal cord injury. J. Spinal Cord Med. 18:1–8. doi: 10.1080/10790268.2019.1605749

PubMed Abstract | CrossRef Full Text | Google Scholar

Barbado, D., Reina, R., Roldan, A., McCulloch, K., Campayo-Piernas, M., and Vera-Garcia, F. J. (2019). How much trunk control is affected in adults with moderate-to-severe cerebral palsy? J. Biomech. 82, 368–374. doi: 10.1016/j.jbiomech.2018.11.009

PubMed Abstract | CrossRef Full Text | Google Scholar

Barbareschi, G., Holloway, C., Bianchi-Berthouze, N., Sonenblum, S., and Sprigle, S. (2018). Use of a low-cost, chest-mounted accelerometer to evaluate transfer skills of wheelchair users during everyday activities: observational study. JMIR Rehabil. Assist. Technol. 5:e11748. doi: 10.2196/11748

PubMed Abstract | CrossRef Full Text | Google Scholar

Boswell-Ruys, C. L., Sturnieks, D. L., Harvey, L. A., Sherrington, C., Middleton, J. W., and Lord, S. R. (2009). Validity and reliability of assessment tools for measuring unsupported sitting in people with a spinal cord injury. Arch. Phys. Med. Rehabil. 90, 1571–1577. doi: 10.1016/j.apmr.2009.02.016

PubMed Abstract | CrossRef Full Text | Google Scholar

Cerrito, A., Bichsel, L., Radlinger, L., and Schmid, S. (2015). Reliability and validity of a smartphone-based application for the quantification of the sit-to-stand movement in healthy seniors. Gait Posture 41, 409–413. doi: 10.1016/j.gaitpost.2014.11.001

PubMed Abstract | CrossRef Full Text | Google Scholar

Cohen, P., Cohen, S. G., and West, L. S., and Aiken, J. (2003). Applied Multiple Regression/Correlation Analysis for the Behavioral Sciences. 2nd Edn. New York, NY: Lawrence Erlbaum Associates Inc.

Google Scholar

Curtis, D. J., Hansen, L., Luun, M., Løberg, R., Woollacott, M., Saavedra, S., et al. (2015). Measuring postural sway in sitting: a new segmental approach. J. Mot. Behav. 47, 427–435. doi: 10.1080/00222895.2014.1003782

CrossRef Full Text | Google Scholar

Douma, J. A. J., Verheul, H. M. W., and Buffart, L. M. (2018). Feasibility, validity and reliability of objective smartphone measurements of physical activity and fitness in patients with cancer. BMC Cancer 18:1052. doi: 10.1186/s12885-018-4983-4

CrossRef Full Text | Google Scholar

Galán-Mercant, A., and Cuesta-Vargas, A. I. (2014). Mobile romberg test assessment (mRomberg). BMC Res. Notes 7:640. doi: 10.1186/1756-0500-7-640

CrossRef Full Text | Google Scholar

Gorman, S. L., Radtka, S., Melnick, M. E., Abrams, G. M., and Byl, N. N. (2010). Development and validation of the function in sitting test in adults with acute stroke. J. Neurol. Phys. Ther. 34, 150–160. doi: 10.1097/NPT.0b013e3181f0065f

PubMed Abstract | CrossRef Full Text | Google Scholar

Hsieh, K. L., Fanning, J. T., Rogers, W. A., Wood, T. A., and Sosnoff, J. J. (2018). A fall risk mhealth app for older adults: development and usability study. JMIR Aging 1:e11569. doi: 10.2196/11569

PubMed Abstract | CrossRef Full Text | Google Scholar

Hsieh, K. L., Roach, K. L., Wajda, D. A., and Sosnoff, J. J. (2019). Smartphone technology can measure postural stability and discriminate fall risk in older adults. Gait Posture 67, 160–165. doi: 10.1016/j.gaitpost.2018.10.005

PubMed Abstract | CrossRef Full Text | Google Scholar

Ivanenko, Y., and Gurfinkel, V. S. (2018). Human postural control. Front. Neurosci. 12:171. doi: 10.3389/fnins.2018.00171

PubMed Abstract | CrossRef Full Text | Google Scholar

Jacob, C. (1988). Statistical Power Analysis for the Behavioral Sciences. 2nd Edn. New York, NY: Lawrence Erlbaum Associates.

Google Scholar

Kalron, A., Fonkatz, I., Frid, L., Baransi, H., and Achiron, A. (2016). The effect of balance training on postural control in people with multiple sclerosis using the CAREN virtual reality system: a pilot randomized controlled trial. J. Neuroeng. Rehabil. 13:13. doi: 10.1186/s12984-016-0124-y

PubMed Abstract | CrossRef Full Text | Google Scholar

Karmarkar, A. M., Dicianno, B. E., Cooper, R., Collins, D. M., Matthews, J. T., Koontz, A., et al. (2011). Demographic profile of older adults using wheeled mobility devices. J. Aging Res. 2011:560358. doi: 10.4061/2011/560358

PubMed Abstract | CrossRef Full Text | Google Scholar

Kim, D.-H., An, D.-H., and Yoo, W.-G. (2018). Changes in trunk sway and impairment during sitting and standing in children with cerebral palsy. Technol. Health Care 26, 761–768. doi: 10.3233/THC-181301

PubMed Abstract | CrossRef Full Text | Google Scholar

Koo, T. K., and Li, M. Y. (2016). A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J. Chiropr. Med. 15, 155–163. doi: 10.1016/j.jcm.2016.02.012

PubMed Abstract | CrossRef Full Text | Google Scholar

Kosse, N. M., Caljouw, S., Vervoort, D., Vuillerme, N., and Lamoth, C. J. C. (2015). Validity and reliability of gait and postural control analysis using the tri-axial accelerometer of the ipod touch. Ann. Biomed. Eng. 43, 1935–1946. doi: 10.1007/s10439-014-1232-0

PubMed Abstract | CrossRef Full Text | Google Scholar

Lynch, S. M., Leahy, P., and Barker, S. P. (1998). Reliability of measurements obtained with a modified functional reach test in subjects with spinal cord injury. Phys. Ther. 78, 128–133. doi: 10.1093/ptj/78.2.128

PubMed Abstract | CrossRef Full Text | Google Scholar

Mellone, S., Tacconi, C., and Chiari, L. (2012). Validity of a smartphone-based instrumented timed up and go. Gait Posture 36, 163–165. doi: 10.1016/j.gaitpost.2012.02.006

PubMed Abstract | CrossRef Full Text | Google Scholar

Murans, G., Gutierrez-Farewik, E. M., and Saraste, H. (2011). Kinematic and kinetic analysis of static sitting of patients with neuropathic spine deformity. Gait Posture 34, 533–538. doi: 10.1016/j.gaitpost.2011.07.009

PubMed Abstract | CrossRef Full Text | Google Scholar

Nguyen, U.-S. D. T., Kiel, D. P., Li, W., Galica, A. M., Kang, H. G., Casey, V. A., et al. (2012). Correlations of clinical and laboratory measures of balance in older men and women. Arthritis Care Res. 64, 1895–1902. doi: 10.1002/acr.21783

PubMed Abstract | CrossRef Full Text | Google Scholar

Ozinga, S. J., Machado, A. G., Miller Koop, M., Rosenfeldt, A. B., and Alberts, J. L. (2015). Objective assessment of postural stability in parkinson's disease using mobile technology. Mov. Disord. 30, 1214–1221. doi: 10.1002/mds.26214

PubMed Abstract | CrossRef Full Text | Google Scholar

Physiopedia Contributors (2018). Wheelchair Users. Available online at: https://www.physio-pedia.com/index.php?title=Wheelchair_Users&oldid=199167 (accessed April 20, 2020).

Google Scholar

Quinzaños, J., Villa, A. R., Flores, A. A., and Pérez, R. (2014). Proposal and validation of a clinical trunk control test in individuals with spinal cord injury. Spinal Cord 52, 449–454. doi: 10.1038/sc.2014.34

PubMed Abstract | CrossRef Full Text | Google Scholar

Reyes, A., Qin, P., and Brown, C. A. (2018). A standardized review of smartphone applications to promote balance for older adults. Disabil. Rehabil. 40, 690–696. doi: 10.1080/09638288.2016.1250124

PubMed Abstract | CrossRef Full Text | Google Scholar

Rice, L. A., Ousley, C., and Sosnoff, J. J. (2015). A systematic review of risk factors associated with accidental falls, outcome measuresand interventions to manage fall risk in non-ambulatory adults. Disabil. Rehabil. 37, 1697–1705. doi: 10.3109/09638288.2014.976718

PubMed Abstract | CrossRef Full Text | Google Scholar

Roeing, K. L., Hsieh, K. L., and Sosnoff, J. J. (2017). A systematic review of balance and fall risk assessments with mobile phone technology. Arch. Gerontol. Geriatr. 73, 222–226. doi: 10.1016/j.archger.2017.08.002

PubMed Abstract | CrossRef Full Text | Google Scholar

Sánchez, M. B., Loram, I., Darby, J., Holmes, P., and Butler, P. B. (2017). A video based method to quantify posture of the head and trunk in sitting. Gait Posture 51, 181–187. doi: 10.1016/j.gaitpost.2016.10.012

PubMed Abstract | CrossRef Full Text | Google Scholar

Shin, S., and Sosnoff, J. J. (2013). Spinal cord injury and time to instability in seated posture. Arch. Phys. Med. Rehabil. 94, 1615–1620. doi: 10.1016/j.apmr.2013.02.008

PubMed Abstract | CrossRef Full Text | Google Scholar

Silsupadol, P., Teja, K., and Lugade, V. (2017). Reliability and validity of a smartphone-based assessment of gait parameters across walking speed and smartphone locations: body, bag, belt, hand, and pocket. Gait Posture 58, 516–522. doi: 10.1016/j.gaitpost.2017.09.030

PubMed Abstract | CrossRef Full Text | Google Scholar

Sosnoff, J. J., Socie, M. J., Boes, M. K., Sandroff, B. M., Pula, J. H., Suh, Y., et al. (2011). Mobility, balance and falls in persons with multiple sclerosis. PLoS ONE 6:e28021. doi: 10.1371/journal.pone.0028021

PubMed Abstract | CrossRef Full Text | Google Scholar

Sung, J., Ousley, C. M., Shen, S., Isaacs, Z. J. K., Sosnoff, J. J., and Rice, L. A. (2016). Reliability and validity of the function in sitting test in nonambulatory individuals with multiple sclerosis. Int. J. Rehabil. Res. 39, 308–312. doi: 10.1097/MRR.0000000000000188

PubMed Abstract | CrossRef Full Text | Google Scholar

Williams, A. D., and Vette, A. H. (2019). A vibrotactile feedback device for seated balance assessment and training. J. Vis. Exp. 20. doi: 10.3791/58611

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: seated postural control, smartphone technology, wheelchair users, clinical tests, measurement tool

Citation: Frechette ML, Abou L, Rice LA and Sosnoff JJ (2020) The Validity, Reliability, and Sensitivity of a Smartphone-Based Seated Postural Control Assessment in Wheelchair Users: A Pilot Study. Front. Sports Act. Living 2:540930. doi: 10.3389/fspor.2020.540930

Received: 06 March 2020; Accepted: 10 September 2020;
Published: 17 December 2020.

Edited by:

Andrea Cereatti, Politecnico di Torino, Italy

Reviewed by:

Sandra Saavedra, University of Hartford, United States
Alberto Botter, Politecnico di Torino, Italy

Copyright © 2020 Frechette, Abou, Rice and Sosnoff. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Jacob J. Sosnoff, jsosnoff@illinois.edu

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.