ORIGINAL RESEARCH article

Front. Robot. AI, 17 December 2024
Sec. Biomedical Robotics
This article is part of the Research Topic Robotics in the Performance, Safety and Learning of Surgery - What Next? View all articles

Launching Stealth AutoGuideTM robot for stereotactic biopsy procedures in a neurosurgical centre: learning curve and workflow optimization

Marcus BarthMarcus BarthEtienne HollEtienne HollFabian FlaschkaFabian FlaschkaSila KarakayaSila KarakayaVitus KrblerVitus KörblerMelanie PichlsbergerMelanie PichlsbergerStefan Wolfsberger
Stefan Wolfsberger*Alexander MickoAlexander Micko
  • Department of Neurosurgery, Medical University of Graz, Graz, Styria, Austria

Background: Accurate histological analysis is crucial for confirming intracerebral neoplasia due to the diverse array of potential diagnoses presented by imaging. In the realm of biopsy techniques, the use of robot-based systems is on the rise, primarily owing to their heightened targeting accuracy. The objective of this study was to elucidate the practicality, learning curve and workflow associated with robot-guided biopsies upon their introduction to a neurosurgical centre.

Materials and methods: In March 2022, the neurosurgical department at our institution adopted the Medtronic Stealth AutoGuide™ cranial robotic guidance platform, a miniaturized robotic guidance device designed for stereotactic procedures. Four experienced neurosurgeons underwent training with the Stealth AutoGuide™ system, after which 51 consecutive biopsies were performed. The evaluation encompassed entry and target accuracy, preoperative setup time, time required for the biopsy procedure, and overall operating time. Statistical analysis was conducted to identify any differences between the initial 26 and subsequent sets of 25 patients, with the aim of identifying changes in workflow.

Results: The study included all patients necessitating a diagnostic biopsy for intracerebral tumours, with only one patient excluded due to the inaccessibility of the intended target point caused by tumour calcification. Notably, there were no significant differences between the first 25 and last 26 patients in the median time from incision to the first biopsy (overall: 11.5 min, IQR 9.03–15.0), the procedure time (overall: 30.0 min, IQR 23.5–46.5), median accuracy at entry (overall: 2.05 mm, IQR 0.8–3.8), or target point (overall: 2.2 mm (IQR 1.6–3.7). However, a significant reduction in robot setup time was observed between the two groups, median 2.69 min versus 1.17 min, respectively (p ≤ 0.001).

Conclusion: The deployment of the robotic biopsy system, Stealth AutoGuide™, showed high feasibility, steep learning curve due to uncomplicated technical handling using our standardized operative workflow. Therefore, even in prone position a high diagnostic yield was achieved. Overall, the Stealth AutoGuide™ system facilitated biopsies in traditionally challenging regions with concise procedure time and surgeon-independent high accuracy.

Introduction

Advancements in medical technology have profoundly reshaped the landscape of neurosurgery, especially in the sphere of stereotactic intracranial biopsies. The confirmation of intracerebral tumours requires a precise histological analysis owing to the vast array of potential diagnoses that are derived from imaging. On the basis of pathologist’s diagnosis of these tissue samples, oncologists can tailor patient-specific targeted therapies, representing a cornerstone of contemporary cancer treatment practices.

One crucial consideration lies in the imperative requirement for precision to ensure the efficacy of these procedures. However, many biopsy needle procedures for obtaining tissue specimens are currently performed with limited accuracy, i.e., by using a mechanical arm. Such methods hold the potential for procedural setbacks, ranging from inconclusive tissue sampling to severe complications like cerebral haemorrhage resulting in subsequent neurological deficits. In context of stereotactic needle biopsies, inaccuracy is associated with the acquisition of non-diagnostic samples in up to 24% of stereotactic biopsy series (Dammers et al., 2010; Dammers et al., 2008; Zoeller et al., 2009) or non-representative tumour specimen in up to 64% of cases (Aker et al., 2005; Jackson et al., 2001; Muragaki et al., 2008). To address this limitation, serial biopsies may be conducted, albeit associated with increased risk of intracranial haemorrhages (Dickerman et al., 2005). These haemorrhages have been reported in 0.3%–59.8% of cases (Dammers et al., 2008; Field et al., 2001; Grossman et al., 2005) significantly contributing to the reported morbidity rates ranging from 0% to 16.1% associated with this procedure (Dammers et al., 2010; Dorward et al., 2002; Lunsford et al., 2008; Tilgner et al., 2005).

The advent of robotic assistance systems has opened up novel pathways for precision and efficiency in neurosurgical procedures (Kaushik et al., 2020). Among the diverse techniques of performing intracranial biopsies that have demonstrated feasibility, robot-based frameless solutions have emerged as a predominant trend recently (Minchev et al., 2020; Deboeuf et al., 2023; Dlaka et al., 2021; Wang et al., 2022; Wu et al., 2021).

A notable innovation in this realm is the Medtronic Stealth AutoGuide™ cranial robotic guidance platform (Medtronic, Minneapolis, MN, United States) which has emerged as a promising tool for guiding frameless biopsies with high accuracy and straightforward handling. The Stealth AutoGuide™ system is a modular guidance system for surgical invasive tools, offering precise trajectory alignment based on predefined navigation data. It consists of two independent guidance modules, each capable of moving in two dimensions, mounted on top of one another to enable precise positioning of the biopsy needle. This configuration allows for any trajectory within a 40 × 40 mm range and up to 30° of angulation, achieving an internal precision of less than 0.1 mm. The robot is designed to automatically acquire and stabilize the desired trajectory, enabling the surgeon to operate hands-free without the need for manual adjustments. The main advantages over competing surgical systems, such as the ROSA One® Brain system (Zimmer Biomet, Jacksonville, FL, United States) and the Renishaw Neuromate® stereotactic robot (Renishaw plc, Wotton-under-Edge, United Kingdom), include its miniaturized design, ease of use, and comparatively lower cost. For further technical details, see Minchev et al., 2021 (Minchev et al., 2021).

In this retrospective study, we aim to evaluate the learning curve and workflow associated with the implementation of the Medtronic Stealth AutoGuide™ in a neurosurgical centre over a 20-month period spanning from 2022 to 2024. Robotic assistance provides the potential to transform stereotactic biopsy procedures, potentially leading to reduced procedure times and opportunities for standardization.

Methods

Study design

This retrospective, single centre study employed a one-armed design to analyse data spanning 20 months following the introduction of the Medtronic Stealth AutoGuide™ from 2022 to 2024. The study aimed to evaluate the learning curve and workflow associated with robot-guided biopsies upon their introduction to a tertiary neurosurgical centre. Eligible participants included all adults aged 18 or older, who underwent a stereotactic frameless biopsy using the Stealth AutoGuide™ robot device. Indications for biopsies were determined by board certified neurosurgeons following decisions made in interdisciplinary tumour board meetings. No patients were excluded following the analysis of the database. This study received approval from the institutional ethics committee (IRB number: 36-236-23/24) and was done in accordance with the principles of the Declaration of Helsinki. Informed consent was obtained from all participants.

Data acquisition

Data acquisition was conducted within a fixed protocol for consecutive cases: Preoperatively, patients underwent cCT (SOMATOM® Force, Siemens Healthineers; non tilted axial scan, soft tissue window, slice thickness 1 mm) and cMRI scans (SIEMENS MAGNETOM 3.0T XT Numaris, Siemens Healthineers; T1 with gadolinium contrast enhancement, slice thickness 1 mm, TR 20.0 ms, TE 4.92 ms, flip angle 25°, voxel size 1 ×1 × 1 mm) with isovoxel sequences, suitable for neuronavigation (Mert et al., 2015). Utilizing the T1-weighted, gadolinium-enhanced MRI scan, stereotactic trajectories were generated using the Medtronic StealthStation™ S8 surgical navigation system (Medtronic, Minneapolis, MN, United States) based on manual plans (Marcus et al., 2020). Parameters such as bone thickness, distance from dura to target, and entry angle were measured on preoperative imaging. In the operating theatre, the patients’ skull was fixated using a Mayfield® Cranial Stabilization System (INTEGRA LifeSciences, Princeton, NJ, United States). Subsequently, optical navigation was employed, and the Medtronic Stealth AutoGuide™ was attached to the Mayfield® skull clamp. The robot required positioning within a 40 × 40 mm area and maximum angulation of ±30° to approach the trajectory (Minchev et al., 2021). The setup time (duration from handover of the robot from the OR assistant to the surgeon until the start of skin disinfection and covering), biopsy time (duration from skin incision to the first biopsy) and procedure time (duration from skin incision to complete closure) were recorded. In cases where glioma or lymphoma have been suspected, the biopsy specimen was checked for 5-ALA positivity under 405-nm-wavelength blue light (Hadjipanayis et al., 2015; Stummer et al., 2006). Within 2 h postoperatively, patients underwent CT scan as recommended (Riche et al., 2022), to anticipate procedure-related haemorrhage before transferring the patient back to ward. To assess accuracy of entry and target points as well as deviation of trajectory angle, these CT sequences were merged with the preoperative neuronavigation plan. The entry point error (EPE, distance between planned and real entry point) was measured at the centre of the burr hole, the target point error (TPE, distance between planned and real target point) was determined at the centre of the air bubble within the biopsy area. Trajectory angle error (AE) was measured as degrees of deviation to the planned trajectory.

Patient-to-image registration

A skin-surface registration was employed in all cases according to a standard protocol (Woerdeman et al., 2007).

In prone position, we employed a strict registration workflow to minimize registration error and potential lack of diagnostic yield: In supine position, an approximation of the entry point is pre-marked using anatomical landmarks and correlations with the 3D model. Then, the patient is turned face down, the head is rotated and secured in the Mayfield® clamp so that the previously marked area becomes the highest point for visualization. For improved accessibility of the face area, the upper body is elevated without tilted neck (legs are elevated in opposite direction) so that the patient’s face is pointing slightly upwards. The camera arm is positioned on the left or right side of the operating table depending on head rotation. The navigation camera is pointed towards the patient’s face. Surface registration is conducted using the Medtronic optical pointing device marking facial aspects first and then continuing on the skull until a value of less than 1.5 mm median root mean square error (RMSE, systematic registration error of the robot) is achieved.

Operative workflow

An ideal intraoperative workflow was developed at our department during the time of this study to optimise precision and reduce complications: Figures 1, 2

1. Starting with prepositioning of the robot in alignment with the trajectory, the entry point is marked on the skin.

2. The robot is then moved away from surgical field using manual joystick mode, so that the operating field at the entry point can be accessed. A skin incision is placed depending on the preferred burr hole width (12 mm for 3.5 mm drill, 25 mm for 7.5 mm drill) with subsequent cauterization and display of the bone surface.

3. The robot is automatically repositioned acquiring trajectory and target alignment of ≤0.1 mm and the guidance tube is then inserted and fixed to the bone.

4. A burr hole is placed by drilling with pre-set depth according to the bone thickness as measured on preoperative CT scans. Advancing in 1 mm steps reduces the risk of unwillingly opening the dura and is conducted until elastic resistance is palpated with the drill.

5. The biopsy-needle reduction tube is inserted.

6. Dura and pia are cauterized and pierced using a neuroendoscopic bipolar (Aesculap® Minop® Bipolar Fork Electrode 2.1 mm).

7. The trajectory is locked to navigate the depth advancement of the biopsy needle.

8. The biopsy needle is inserted and up to four specimen samples are withdrawn from the quadrants of the target.

9. Finally, the needle is removed.

10. Skin closure is performed as usual.

Figure 1
www.frontiersin.org

Figure 1. (A) - Robot setup, Mayfield clamp on patient head; (B) - Intraoperative robot positioning for biopsy; (C) - Burr hole using the 3.2 mm drill; (D) - 5-ALA fluorescence obviates the need for intraoperative histopathology decreasing procedure time.

Figure 2
www.frontiersin.org

Figure 2. Intraoperative workflow chart depicting the robot’s tasks (left, grey) and the surgeon’s tasks (right, blue).

5-ALA, frozen section, histopathology

5-ALA has been shown to be highly sensitive for detection of suspected lymphoma and glioblastoma (Kiesel et al., 2018). Therefore, an intraoperative frozen section sample was not performed in 33 of 51 cases (64.7%) where the specimen showed high 5-ALA positivity under 405 nm light examined using the Zeiss Kinevo® 900 microscope (Carl Zeiss AG, Oberkochen, Germany). However, for other pathologies and at the surgeon’s preference, additional frozen section analysis was conducted. This supplementary step extended the procedure time, albeit ensuring comprehensive diagnostic evaluation.

Trained surgeons

In total, four experienced neurosurgeons underwent training with the handling of the Stealth AutoGuide™. Throughout the entire investigation period, neither the setup nor the procedure underwent alterations. Preliminary differences were only apparent in the size of the applied burr hole, as the used drill provided either a 3.5 mm or a 7.5 mm tip diameter, based on surgeon preference.

Statistical analysis

Categorical data are depicted as numbers and relative frequencies, while nonparametric data are presented as median and interquartile range (IQR). Contingency tables underwent evaluation utilizing the Fisher exact test. Group comparisons of independent samples were conducted using the non-parametric Mann-Whitney-U test. Additionally, the Kruskal-Wallis H test was employed to discern differences among the four involved neurosurgeons. A two-sided p-value <0.05 was considered statistically significant. For statistical analyses, SPSS® version 29.0 software (IBM Corp., Armonk, NY, United States) was used.

Results

Patients’ characteristics

The application of the Stealth AutoGuide™ robot was feasibly in all 51 consecutive patients. Histology was conclusive in all cases. Within the study cohort, Glioblastoma multiforme (WHO Grade IV) (31/51, 60.8%) was predominant, followed by Astrocytoma (Pilocytic Astrocytoma WHO Grade I, Diffuse Astrocytoma WHO Grade II, Anaplastic Astrocytoma WHO Grade III) (5/51, 9.8%), Lymphoma (Diffuse large B-cell lymphoma) (9/51, 17.6%) and Gliosis (2/51, 7.8%). Median lesion volume was 15.2 cm³ (IQR 5.55–29.25 cm³). Most patients underwent surgery in supine position (40/51, 78.4%), median head rotation from sagittal plane was 7.5° (IQR 0.0°–37.5°) and the mainly utilised drill width was 3.2 mm in 70.6% of the cases. In patients that received 5-ALA (38/51, 74.5%) preoperatively, specimen showed a strong fluorescence in 68.4%. A frozen section was performed in 35.3% of the cases. Table 1.

Table 1
www.frontiersin.org

Table 1. Patient characteristics.

Accuracy and patient positioning

Evaluation of preoperative imaging revealed a median bone thickness of 7.5 mm (IQR 6.0–9.0 mm), median distance from dura to target of 36.6 mm (IQR 30.9–51.0 mm) and an angulation of the trajectory from the perpendicular axis of 8.5° (IQR 3.1°–16.0°). The median RMSE as systematic deviation provided by Medtronic StealthStation™ S8 after registration, was 1.25 mm (IQR 1.1–1.5 mm). Overall, accuracy measurements yielded an entry point error (EPE) of 2.05 mm (IQR 0.8–3.8 mm), a target point error (TPE) of 2.2 mm (IQR 1.6–3.5 mm) and a deviation from the axis of the planned trajectory of 3.0° (IQR 1.0°–5.0°) (Table 2).

Table 2
www.frontiersin.org

Table 2. Accuracy and workflow, differences between groups.

Ten patients were placed in prone position, as targets in these cases were located in the parietal, occipital lobes or within the posterior fossa to minimize entry angle and distance from dura to target. In comparison with patients in supine position, RSME (1.2 mm vs. 1.3 mm, p = 1.0) showed no difference between positioning, opposing recent research (Czabanka and Misch, 2016) where surface registration could not be performed with the same accuracy in prone position. However, the angle at entry error (3.0° vs. 1.0°, p = 0.019) was found significantly lower in patients in prone position. Both, accuracy and workflow parameters were not significantly different between positionings, with entry and target point error as well as setup time being lower in prone position, and biopsy time being lower in supine positioned patients (Table 3; Figure 3).

Table 3
www.frontiersin.org

Table 3. Comparison of supine and prone position.

Figure 3
www.frontiersin.org

Figure 3. 3D Bone reconstruction showing the planned (blue) and real (green) trajectories for biopsies in prone position.

Workflow and learning curve

Workflow measurements revealed a median setup time of 1.5 min (IQR 1.08–2.69 min), median biopsy time was 11.5 min (IQR 9.03–15.0 min) and median procedure time was 30.0 min (IQR 23.5–46.5 min). For further analysis patients were divided into two groups depended on the date of surgery. Group 1 being the first half of the patient cohort (n = 26) and Group 2 the second half (n = 25). The groups showed a congruent matching of baseline characteristics (Table 1). Regarding accuracy, entry and target point error non-significantly decreased over time (EPE 2.1 mm–2.0 mm, p = 0.472; TPE 2.65 mm–1.9 mm, p = 0.453). Analysis of the workflow revealed a significant decline in setup time (2.69 min–1.17 min, p < 0.001). However, no significant changes in biopsy and procedure time have been found. For both, accuracy and workflow analysis, no significant differences between the four trained surgeons could be found. (Figure 4; Table 2).

Figure 4
www.frontiersin.org

Figure 4. Differences in accuracy between surgeons.

Complications

Intraoperative bleeding occurred in one case (2%) during tumour sampling where the target was located close to the pineal gland and histopathological analysis revealed a pineal parenchymal tumour of intermediate differentiation. Postoperative cranial CT scan showed no sign of mass effect, but intraventricular blood accumulation. Therefore, an external ventricular drainage was placed, and the patient was transferred to ICU for 1 day. The patient could be released without clinical or neurological deficit on postop day 21, without the need for a ventriculoperitoneal shunt.

Discussion

Evaluating the learning curve and understanding workflow dynamics of this technology is paramount for optimizing patient outcomes and integrating robotic systems seamlessly into neurosurgical practice. We examined the initial 51 stereotactic biopsy cases conducted at an institutional neurosurgical department utilizing the Medtronic Stealth AutoGuide™ system, with a focus on evaluating workflow efficiency and the learning curve effect. A comparative analysis between a sample of 26 primarily treated patients and 25 consecutively treated patients revealed a significant improvement in the time required to set up the robot. No significant differences were observed either among the treating surgeons or between supine and prone positioning.

Accuracy

Key outcome criteria regarding accuracy slightly exceeded previously reported values (Minchev et al., 2019), with a median entry point error of 2.05 mm (IQR 0.8–3.8 mm) and a median target point error of 2.2 mm (IQR 1.6–3.5 mm) across the entire sample of 51 patients. A predominant contributing factor to these results has been observed by a RMSE of 1.25 mm due to the routine use of surface registration. Other studies employed screw-based registration (Minchev et al., 2017) which we deemed less favourable due to the invasive nature of this procedure. Nevertheless, we would recommend using screws for improved registration accuracy for small lesions (volume <1 cm³) in highly eloquent areas (such as the brainstem) which were not cases in study.

Our method to measure accuracy relied on entrapped air detection on postoperative CT scans approximately 2 h postoperative to measure entry and target deviation, which might have contributed to higher entry and target point error. Alternative imaging methods such as postoperative MRI scans were not adopted at our department due to the cost-benefit trade off. Nonetheless, the median accuracy remained within an acceptable range (Zrinzo, 2012).

Learning curve

Assessing the learning curve, we compared the first half (26 cases, Group 1) to the later biopsy cases (25 cases, Group 2). A trend toward slight improvement in accuracy measurements was noted, particularly in case of the target point error which decreased from 2.65 mm to 1.9 mm (p = 0.453), achieving a decline below the desirable threshold of less than 2 mm for stereotactic interventions. In all cases, a conclusive histologic diagnosis was obtained, reaffirming the technique’s feasibility as demonstrated in prior research (Minchev et al., 2017). Studies conducted during the development of the Stealth AutoGuide™ robot (Minchev et al., 2021) involved only a few pre-trained neurosurgeons and primarily utilized the prototype iSYS1 robot under technical supervision. In contrast, our investigation followed a more field research-oriented approach, examining the use of the Stealth AutoGuide™ system under routine clinical application. Hence, we posit that the data elucidated in this study more accurately reflect the authentic outcomes when newly integrating such robotic systems into a neurosurgical department.

Positioning

The analysis of challenging stereotactic biopsies performed in prone position did not reveal significant deviants from supine position. Several authors observed that “Neuronavigation system precision is more accurate in supine position compared to prone position due to improved visibility of face and convexity surface for the navigation camera” (Czabanka and Misch, 2016) and “[…] stereotactic procedures should be performed with the patient in the identical position during imaging and intervention.” (Rohlfing et al., 2003). We indeed could show that process standardisation as well as the application of a user-independent robotic device can minimize difficulties of prone positioning and path the way for performing biopsies of intricate target locations.

Workflow

Workflow measurements demonstrated a significant reduction in setup time, from 2.7 min to 1.2 min (p < 0.001), indicating enhanced efficiency in placing the robot on the Mayfield® clamp and positioning its workspace within reach for the entry point. Additionally, the individual improvement in initial robot positioning may have contributed to the slight trend of increased accuracy in Group 2. Recent RCTs underline that frameless procedures are superior to frame-based ones regarding the duration while they report procedure times of 79.1 min (Georgiopoulos et al., 2018), 59.0 min (Bradac et al., 2017) and 42.0 min (Gempt et al., 2012), respectively. Meanwhile, our biopsy and procedure times remained constant at 11.5 min and 30.0 min, respectively, suggesting these durations are among the shortest achievable with the applied technique. As frozen section analysis was conducted in 18 cases included in the analysis of procedure time, transportation, processing, and analysis of the specimen were probably prolonging factors independent of the actual biopsy procedure.

Comparative analysis of the four participating surgeons supported the concept that a robot-based biopsy technique is easily standardized and implemented across different users regardless of their level of training. No statistically significant differences were observed regarding the individual surgeons’ performance concerning outcome parameters. A prospective study comparing two different biopsy techniques, including 526 biopsies conducted over 12 years, revealed a cutoff within the learning curve concerning operating time. On average, the 12 participating surgeons required 10 cases for robot-based procedures with the Zimmer Biomet ROSA ONE® and four cases for procedures using the stereotactic mechanical arm Brainlab VarioGuide® to reduce the procedure time by approximately 15 min (Mallereau et al., 2022). The primary distinction from our technique is the minimalistic composition of the Stealth AutoGuide™ robot, which might explain its simpler adoption and the absence of inter- and intra-individual differences in biopsy and procedure time. However, future follow-up studies may identify factors contributing to improved results, given the considerable variance in sample sizes among the four surgeons.

Limitations

The present study is subject to limitations arising from its retrospective design and the relatively modest sample size. Consequently, regression analysis could not be applied in a manner that is empirically sensible. Accuracy measurements are lacking behind due to postop confirmation of biopsy targets via CT scan instead of MR scans applied in former studies (Minchev et al., 2021).

Nonetheless, our study aimed to assess the initial impacts of implementing a new robot-based procedure and delineate the associated learning curve. However, to validate our findings, follow-up studies employing a prospective patient cohort are imperative and will provide a more comprehensive understanding of the evaluated outcomes. Expanding the pool of surgeons and incorporating users from different institutions would yield more generalizable results. This approach could help demonstrate that the device’s learning curve and effectiveness are not influenced by local factors but remain consistent across different settings and varying skill levels.

Conclusion

The deployment of the robotic biopsy system, Stealth AutoGuide™, showed excellent feasibility, steep learning curve due to uncomplicated technical handling using our standardized operative workflow. Therefore, even in prone position a high diagnostic yield was achieved. Overall, the Stealth AutoGuide™ system facilitated biopsies in traditionally challenging regions with concise procedure time and surgeon-independent high accuracy.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by Ethikkommission Medizinische Universität Graz. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin because patients underwent standard procedures and gave informed consent for data collection, but due to retrospective data analysis further informed consent was not required.

Author contributions

MB: Writing–original draft, Writing–review and editing. EH: Conceptualization, Data curation, Investigation, Project administration, Writing–review and editing. FF: Data curation, Investigation, Writing–review and editing. SK: Data curation, Investigation, Writing–review and editing. VK: Data curation, Investigation, Writing–review and editing. MP: Data curation, Project administration, Writing–review and editing. SW: Writing–original draft, Writing–review and editing, Conceptualization, Data curation, Funding acquisition, Investigation, Methodology, Supervision. AM: Data curation, Formal Analysis, Investigation, Methodology, Project administration, Supervision, Visualization, Writing–review and editing.

Funding

The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. This work was supported by Medtronic Plc. Medtronic has supported this study financially and with the provision of equipment through the External Research Program (ERP-2022-13,037), and was not involved in the study design, collection, analysis, and interpretation of the data.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Abbreviations

IRB, Institutional Review Board; TR, Repetition Time; TE, Time to Echo; 5-ALA, 5-Aminolevulinic acid; EPE, Entry Point Error; TPE, Target Point Error; AE, Angle Error; RMSE, Root Mean Square Error; IQR, Interquartile Range; ICU, Intensive Care Unit; RCT, Randomised Controlled Trial.

References

Aker, F. V., Hakan, T., Karadereler, S., and Erkan, M. (2005). Accuracy and diagnostic yield of stereotactic biopsy in the diagnosis of brain masses: comparison of results of biopsy and resected surgical specimens. Neuropathology official J. Jpn. Soc. Neuropathology 25, 207–213. doi:10.1111/j.1440-1789.2005.00634.x

PubMed Abstract | CrossRef Full Text | Google Scholar

Bradac, O., Steklacova, A., Nebrenska, K., Vrana, J., Lacy, P. de, and Benes, V. (2017). Accuracy of VarioGuide frameless stereotactic system against frame-based stereotaxy: prospective, randomized, single-center study. World Neurosurg. 104, 831–840. doi:10.1016/j.wneu.2017.04.104

PubMed Abstract | CrossRef Full Text | Google Scholar

Czabanka, M., and Misch, M.: Brain biopsy of a suspected cerebellar lymphoma, 2016. doi:10.24296/jomi/139

CrossRef Full Text | Google Scholar

Dammers, R., Haitsma, I. K., Schouten, J. W., Kros, J. M., Avezaat, C. J. J., and Vincent, AJPE (2008). Safety and efficacy of frameless and frame-based intracranial biopsy techniques. Acta Neurochir. 150, 23–29. doi:10.1007/s00701-007-1473-x

PubMed Abstract | CrossRef Full Text | Google Scholar

Dammers, R., Schouten, J. W., Haitsma, I. K., Vincent, AJPE, Kros, J. M., and Dirven, C. M. F. (2010). Towards improving the safety and diagnostic yield of stereotactic biopsy in a single centre. Acta Neurochir. 152, 1915–1921. doi:10.1007/s00701-010-0752-0

PubMed Abstract | CrossRef Full Text | Google Scholar

Deboeuf, L., Moiraghi, A., Debacker, C., Peeters, S. M., Simboli, G. A., Roux, A., et al. (2023). Feasibility and accuracy of robot-assisted, stereotactic biopsy using 3-dimensional intraoperative imaging and frameless registration tool. Neurosurgery 92, 803–811. doi:10.1227/neu.0000000000002294

PubMed Abstract | CrossRef Full Text | Google Scholar

Dickerman, R. D., Mittler, M. A., and Morgan, J. T. (2005). Stereotactic brain biopsies and operative complications: technique to further decrease risks. Acta Neurochir. 147, 911–912. doi:10.1007/s00701-005-0566-7

PubMed Abstract | CrossRef Full Text | Google Scholar

Dlaka, D., Švaco, M., Chudy, D., Jerbić, B., Šekoranja, B., Šuligoj, F., et al. (2021). Frameless stereotactic brain biopsy: a prospective study on robot-assisted brain biopsies performed on 32 patients by using the RONNA G4 system. Int. J. Med. robotics + computer assisted surgery MRCAS 17, e2245. doi:10.1002/rcs.2245

CrossRef Full Text | Google Scholar

Dorward, N. L., Paleologos, T. S., Alberti, O., and Thomas, D. G. T. (2002). The advantages of frameless stereotactic biopsy over frame-based biopsy. British journal of neurosurgery 16, 110–118. doi:10.1080/02688690220131705

PubMed Abstract | CrossRef Full Text | Google Scholar

Field, M., Witham, T. F., Flickinger, J. C., Kondziolka, D., and Lunsford, L. D. (2001). Comprehensive assessment of hemorrhage risks and outcomes after stereotactic brain biopsy. Journal of neurosurgery 94, 545–551. doi:10.3171/jns.2001.94.4.0545

PubMed Abstract | CrossRef Full Text | Google Scholar

Gempt, J., Buchmann, N., Ryang, Y.-M., Krieg, S., Kreutzer, J., Meyer, B., et al. (2012). Frameless image-guided stereotaxy with real-time visual feedback for brain biopsy. Acta neurochirurgica 154, 1663–1667. doi:10.1007/s00701-012-1425-y

PubMed Abstract | CrossRef Full Text | Google Scholar

Georgiopoulos, M., Ellul, J., Chroni, E., and Constantoyannis, C. (2018). Efficacy, safety, and duration of a frameless fiducial-less brain biopsy versus frame-based stereotactic biopsy: a prospective randomized study. Journal of neurological surgery. Part A, Central European neurosurgery 79, 031–038. doi:10.1055/s-0037-1602697

PubMed Abstract | CrossRef Full Text | Google Scholar

Grossman, R., Sadetzki, S., Spiegelmann, R., and Ram, Z. (2005). Haemorrhagic complications and the incidence of asymptomatic bleeding associated with stereotactic brain biopsies. Acta neurochirurgica 147, 627–631. doi:10.1007/s00701-005-0495-5

PubMed Abstract | CrossRef Full Text | Google Scholar

Hadjipanayis, C. G., Widhalm, G., and Stummer, W. (2015). What is the surgical benefit of utilizing 5-aminolevulinic acid for fluorescence-guided surgery of malignant gliomas? Neurosurgery 77, 663–673. doi:10.1227/NEU.0000000000000929

PubMed Abstract | CrossRef Full Text | Google Scholar

Jackson, R. J., Fuller, G. N., Abi-Said, D., Lang, F. F., Gokaslan, Z. L., Shi, W. M., et al. (2001). Limitations of stereotactic biopsy in the initial management of gliomas. Neuro-oncology 3, 193–200. doi:10.1093/neuonc/3.3.193

PubMed Abstract | CrossRef Full Text | Google Scholar

Kaushik, A., Dwarakanath, T. A., Bhutani, G., and Srinivas, D. (2020). Robot-based autonomous neuroregistration and neuronavigation: implementation and case studies. World neurosurgery 134, e256–e271. doi:10.1016/j.wneu.2019.10.041

PubMed Abstract | CrossRef Full Text | Google Scholar

Kiesel, B., Millesi, M., Woehrer, A., Furtner, J., Bavand, A., Roetzer, T., et al. (2018). 5-ALA-induced fluorescence as a marker for diagnostic tissue in stereotactic biopsies of intracranial lymphomas: experience in 41 patients. Neurosurgical focus 44, E7. doi:10.3171/2018.3.FOCUS1859

PubMed Abstract | CrossRef Full Text | Google Scholar

Lunsford, L. D., Niranjan, A., Khan, A. A., and Kondziolka, D. (2008). Establishing a benchmark for complications using frame-based stereotactic surgery. Stereotactic and functional neurosurgery 86, 278–287. doi:10.1159/000147636

PubMed Abstract | CrossRef Full Text | Google Scholar

Mallereau, C.-H., Chibbaro, S., Ganau, M., Benmekhbi, M., Cebula, H., Dannhoff, G., et al. (2022). Pushing the boundaries of accuracy and reliability during stereotactic procedures: a prospective study on 526 biopsies comparing the frameless robotic and Image-Guided Surgery systems. Journal of clinical neuroscience official journal of the Neurosurgical Society of Australasia 95, 203–212. doi:10.1016/j.jocn.2021.11.034

PubMed Abstract | CrossRef Full Text | Google Scholar

Marcus, H. J., Vakharia, V. N., Sparks, R., Rodionov, R., Kitchen, N., McEvoy, A. W., et al. (2020). Computer-Assisted versus manual planning for stereotactic brain biopsy: a retrospective comparative pilot study. Operative neurosurgery (Hagerstown, Md.) 18, 417–422. doi:10.1093/ons/opz177

PubMed Abstract | CrossRef Full Text | Google Scholar

Mert, A., Kiesel, B., Wöhrer, A., Martínez-Moreno, M., Minchev, G., Furtner, J., et al. (2015). Introduction of a standardized multimodality image protocol for navigation-guided surgery of suspected low-grade gliomas. Neurosurgical focus 38, E4. doi:10.3171/2014.10.FOCUS14597

PubMed Abstract | CrossRef Full Text | Google Scholar

Minchev, G., Kronreif, G., Martínez-Moreno, M., Dorfer, C., Micko, A., Mert, A., et al. (2017). A novel miniature robotic guidance device for stereotactic neurosurgical interventions: preliminary experience with the iSYS1 robot. Journal of neurosurgery 126, 985–996. doi:10.3171/2016.1.JNS152005

PubMed Abstract | CrossRef Full Text | Google Scholar

Minchev, G., Kronreif, G., Ptacek, W., Dorfer, C., Micko, A., Maschke, S., et al. (2019). A novel robot-guided minimally invasive technique for brain tumor biopsies. Journal of neurosurgery 132, 150–158. doi:10.3171/2018.8.JNS182096

PubMed Abstract | CrossRef Full Text | Google Scholar

Minchev, G., Kronreif, G., Ptacek, W., Kettenbach, J., Micko, A., Wurzer, A., et al. (2020). Frameless stereotactic brain biopsies: comparison of minimally invasive robot-guided and manual arm-based technique. Operative neurosurgery (Hagerstown, Md.) 19, 292–301. doi:10.1093/ons/opaa123

PubMed Abstract | CrossRef Full Text | Google Scholar

Minchev, G., Wurzer, A., Ptacek, W., Kronreif, G., Micko, A., Dorfer, C., et al. (2021). Development of a miniaturized robotic guidance device for stereotactic neurosurgery. Journal of neurosurgery 137, 479–488. doi:10.3171/2021.9.JNS21794

PubMed Abstract | CrossRef Full Text | Google Scholar

Muragaki, Y., Chernov, M., Maruyama, T., Ochiai, T., Taira, T., Kubo, O., et al. (2008). Low-grade glioma on stereotactic biopsy: how often is the diagnosis accurate? Minimally invasive neurosurgery. MIN 51, 275–279. doi:10.1055/s-0028-1082322

PubMed Abstract | CrossRef Full Text | Google Scholar

Riche, M., Marijon, P., Amelot, A., Bielle, F., Mokhtari, K., Chambrun, M. P. de, et al. (2022). Severity, timeline, and management of complications after stereotactic brain biopsy. Journal of neurosurgery 136, 867–876. doi:10.3171/2021.3.JNS21134

PubMed Abstract | CrossRef Full Text | Google Scholar

Rohlfing, T., Maurer, C. R., Dean, D., and Maciunas, R. J. (2003). Effect of changing patient position from supine to prone on the accuracy of a Brown-Roberts-Wells stereotactic head frame system. Neurosurgery 52, 610–618. doi:10.1227/01.neu.0000048727.65969.36

PubMed Abstract | CrossRef Full Text | Google Scholar

Stummer, W., Pichlmeier, U., Meinel, T., Wiestler, O. D., Zanella, F., and Reulen, H.-J. (2006). Fluorescence-guided surgery with 5-aminolevulinic acid for resection of malignant glioma: a randomised controlled multicentre phase III trial. The Lancet. Oncology 7, 392–401. doi:10.1016/S1470-2045(06)70665-9

PubMed Abstract | CrossRef Full Text | Google Scholar

Tilgner, J., Herr, M., Ostertag, C., and Volk, B. (2005). Validation of intraoperative diagnoses using smear preparations from stereotactic brain biopsies: intraoperative versus final diagnosis--influence of clinical factors. Neurosurgery 56, 257–265. doi:10.1227/01.neu.0000148899.39020.87

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, M., Zhang, Y., Shi, W., Zhu, R., Li, H., and Zhao, R. (2022). Frameless robot-assisted stereotactic biopsy: an effective and minimally invasive technique for pediatric diffuse intrinsic pontine gliomas. Journal of neuro-oncology 160, 107–114. doi:10.1007/s11060-022-04122-4

PubMed Abstract | CrossRef Full Text | Google Scholar

Woerdeman, P. A., Willems, P. W. A., Noordmans, H. J., Tulleken, C. A. F., and van der Sprenkel, J. W. B. (2007). Application accuracy in frameless image-guided neurosurgery: a comparison study of three patient-to-image registration methods. Journal of neurosurgery 106, 1012–1016. doi:10.3171/jns.2007.106.6.1012

PubMed Abstract | CrossRef Full Text | Google Scholar

Wu, S., Wang, J., Gao, P., Liu, W., Hu, F., Jiang, W., et al. (2021). A comparison of the efficacy, safety, and duration of frame-based and Remebot robot-assisted frameless stereotactic biopsy. British journal of neurosurgery 35, 319–323. doi:10.1080/02688697.2020.1812519

PubMed Abstract | CrossRef Full Text | Google Scholar

Zoeller, G. K., Benveniste, R. J., Landy, H., Morcos, J. J., and Jagid, J. (2009). Outcomes and management strategies after nondiagnostic stereotactic biopsies of brain lesions. Stereotactic and functional neurosurgery 87, 174–181. doi:10.1159/000222661

PubMed Abstract | CrossRef Full Text | Google Scholar

Zrinzo, L. (2012). Pitfalls in precision stereotactic surgery. Surgical neurology international 3, S53–S61. doi:10.4103/2152-7806.91612

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: autoguide, stereotactic biopsy, learning curve, workflow, prone position, accuracy

Citation: Barth M, Holl E, Flaschka F, Karakaya S, Körbler V, Pichlsberger M, Wolfsberger S and Micko A (2024) Launching Stealth AutoGuideTM robot for stereotactic biopsy procedures in a neurosurgical centre: learning curve and workflow optimization. Front. Robot. AI 11:1437568. doi: 10.3389/frobt.2024.1437568

Received: 23 May 2024; Accepted: 02 December 2024;
Published: 17 December 2024.

Edited by:

Garnette Sutherland, University of Calgary, Canada

Reviewed by:

Andrej Steno, Faculty of Medicine of Comenius University and University Hospital Bratislava, Slovakia
Marina Carbone, University of Pisa, Italy

Copyright © 2024 Barth, Holl, Flaschka, Karakaya, Körbler, Pichlsberger, Wolfsberger and Micko. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Stefan Wolfsberger, c3RlZmFuLndvbGZzYmVyZ2VyQG1lZHVuaWdyYXouYXQ=

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.