Acceptability of wearable inertial sensors, completeness of data, and day-to-day variability of everyday life motor activities in children and adolescents with neuromotor impairments

  • 1Swiss Children’s Rehab, University Children’s Hospital Zurich, Affoltern am Albis, Switzerland
  • 2Children’s Research Center, University Children’s Hospital of Zurich, University of Zurich, Zurich, Switzerland
  • 3Rehabilitation Engineering Laboratory, Department of Health Sciences and Technology, ETH Zurich, Zurich, Switzerland

Monitoring the patients' motor activities in a real-world setting would provide essential information on their functioning in daily life. In this study, we used wearable inertial sensors to monitor motor activities of children and adolescents with congenital and acquired brain injuries. We derived a set of clinically meaningful performance measures and addressed the following research questions: Is the target population willing to wear the sensors in their habitual environment? Which factors lead to missing data, and can we avoid them? How many measurement days are needed to obtain reliable estimates of the children's and adolescents' motor performance? The study participants wore our sensor system for seven consecutive days during waking hours. First, we derived the daily hand use of all participants, the duration of different body positions and the wheeling activity of individuals using a manual wheelchair, and walking-related measures in individuals being able to walk. Then, we analyzed the reasons for missing data and determined the reliability of the performance measures mentioned above. The large majority (41 of 43 participants) was willing to wear the sensor system for a week. However, forgetting to reattach the sensors after charging them overnight and taking them off during bathing and swimming was the main contributor to missing data. Consequently, improved battery life and waterproofness of the sensor technology are essential requirements for measurements in daily life. Besides, 5 of 11 performance measures showed significant differences between weekdays and weekend days. The reliability, measured with the intraclass correlation coefficient, ranged between 0.82 and 0.98. Seven measurement days were enough to obtain significantly higher reliability scores than the desired level of 0.8 for all but two performance measures. In children and adolescents with neuromotor impairments, we recommend monitoring everyday life motor activities on seven consecutive days. The target population accepted this measurement protocol, it covers school days and weekend days, and the number of measurement days is sufficient to obtain reliable estimates of motor performance.

Introduction

Children and adolescents with congenital or acquired brain injuries often have difficulties in executing everyday life motor activities, such as grasping a glass of water, transferring from a wheelchair to a car seat, or walking to school. They undertake intensive rehabilitation programs as in- or out-patients with an emphasis on fostering their functional independence in these activities. To monitor the children's progress over time and evaluate the effect of therapeutic interventions, usually, motor capacity (“what a child can do”) is measured in a standardized environment at the clinic. In the habitual environment outside of the clinic, however, motor performance (“what a child does do”) becomes much more important, and it remains unclear whether children can translate their improvements during rehabilitation into everyday life (13). Consequently, there is a need to assess motor performance to quantify what children and adolescents do in their habitual environment.

Self-report or proxy-report measures can be used to assess motor performance. However, these tools rely on the subjective perception of these children and adolescents, or their parents, and are prone to recall and proxy bias (4). Wearable inertial sensors overcome the limitation of subjectivity by enabling objective monitoring of motor activities in real-world settings (5). However, the most commonly used outcome measure to assess performance is activity counts, which quantifies the general level of physical activity rather than the type and quality of activities performed (6). Therefore, sophisticated algorithms are needed to derive activity-specific and clinically meaningful performance measures from data of wearable sensors.

We developed such an algorithm based on the findings of two preceding studies investigating the needs of pediatric rehabilitation (7, 8). The current algorithm determines functional hand use with wrist sensors; the duration of lying, sitting, and standing positions with a trunk and a thigh sensor; the distance and speed of self-propelled wheeling periods with a wrist and a wheel sensor; the duration, distance, and speed of walking periods, and the altitude change during stair climbing periods with a single ankle sensor; and discriminates between free and assisted walking with a sensor placed on walking aids. The algorithm can be applied in a modular way. Its outcome measures and the required sensor placement are depicted in Table 1. Then, we verified the validity of this algorithm in three subsequent studies. They showed sufficient criterion validity of the hand use measures (9), good to excellent activity classification accuracy except for stair climbing (10), and accurate gait speed estimations (11).

TABLE 1
www.frontiersin.org

Table 1. Study groups, body-worn sensor configurations, and performance measures.

These validity studies were conducted in supervised experiments at the clinic to allow for the inclusion of criterion measures. However, in real-world settings, other factors such as the acceptance to wear the sensors, the completeness of data, and the naturally occurring day-to-day variability of motor activities must be considered. On the one hand, incomplete datasets could occur due to non-wearing time or technical issues of the sensor system leading to missing or biased estimates of the users' daily motor activities (12). On the other hand, a sufficient number of repeated measurement days is needed to capture the day-to-day variability of motor activities and obtain reliable estimates of the patients' overall activity levels. The literature suggests measuring performance over a week to incorporate variability between weekdays and differences between weekdays and weekends (1315). However, this has to be reevaluated in our newly developed performance measures. Moreover, the recommendation on how many measurement days are needed will depend not only on maximizing the reliability of the performance measures but also on the children's and adolescents' willingness to wear the sensors and minimizing the burden to their everyday lives.

Therefore, we aimed to determine our sensor system's acceptability, the completeness of data, and the performance measure's reliability in a real-word setting. Children and adolescents with neuromotor impairments wore our sensor system for seven consecutive days in their habitual environment, including school days and weekend days, and we addressed the following research questions: Are children and adolescents with neuromotor impairments willing to wear the sensors in daily life? Are there other issues leading to missing data or data with insufficient quality? How many measurement days are needed to derive reliable estimates of motor performance?

Materials and methods

This study was part of a larger ongoing study investigating the influence of contextual factors on translating rehabilitation progress into daily life. The local ethics committee approved the study protocol (BASEC-No.: 2020-00724).

Participants

We recruited school-aged children and adolescents with congenital or acquired injuries or illnesses of the central or peripheral nervous system. They fulfilled the following inclusion criteria: ability to wheel or walk for household distances; ability to transfer between a wheelchair and a chair over a standing position in individuals who used a wheelchair; living with the mother, father, or psychological parent during the whole measurement period; no wounds or other medical conditions that prevented sensor placement; cognitive abilities to understand and follow basic verbal instructions; and signed consent form.

The participants were allocated to two of three subgroups to minimize the number of body-worn sensors and derive clinically meaningful performance measures. All participants were part of the upper limb group in which we measured their daily hand use with wrist-worn sensors. Additionally, they were allocated either to the wheelchair group or the walking group based on their primary mobility at home. In the wheelchair group, we measured the duration they spent in different body positions and their wheeling activities with additional sensors on the trunk and the thigh, while walking-related performance measures were determined in the walking group with an additional sensor on the ankle. The three subgroups and the corresponding body-worn sensor configurations and performance measures are illustrated in Table 1.

Equipment and procedure

The study procedure comprised three parts. First, we determined the participants' motor abilities with motor assessments at the clinic to describe the study population's levels of motor impairment. Second, we monitored their motor performance with wearable inertial sensors for seven days. Afterwards, we asked the participants to rate the obtrusiveness of the sensor system. In in-patients, we conducted the motor assessments during the last week of their stay at the clinic and measured their motor performance two to four weeks after rehabilitation. We chose this time interval to allow for a habituation phase at home after the in-patient rehabilitation. In out-patients, the measurement of motor performance started directly after the motor assessments.

The motor assessments included the Melbourne Assessment 2 (MA2) to measure the quality of upper limb movements (16), and the Gross Motor Function Measure (GMFM) to determine the capacity of gross motor activities (17). Here, we performed dimensions B (sitting) and D (standing) in the wheelchair group and dimensions D and E (walking, running, and jumping) in the walking group.

To measure performance, participants were equipped with multiple ZurichMOVE sensor modules containing an accelerometer, a gyroscope, and an altimeter (18). One sensor weighs 18 g and its dimensions are 35 × 35 × 12 mm. The sensor's internal storage capacity and battery life allow for continuous recordings of roughly 72 h at a sampling frequency of 50 Hz. One of the authors demonstrated the placement of the sensors with corresponding hook-and-loop straps on the first day and provided accompanying instructions on usage and charging of the devices. During the measurement period, the participants handled the sensors themselves and were assisted by their parents when needed. All participants wore a sensor on each wrist. Those of the wheelchair group wore additional sensors on the trunk and the thigh, and we fixated a sensor on the spokes of their wheelchair. The straps of the trunk and the thigh had a silicone strip on the inside to prevent them from slipping down. Participants of the walking group wore a sensor on the ankle of their less-affected leg, and, if applicable, we fixated sensors on their walking aids. We instructed the participants to wear the sensors during the day and charge them overnight on a corresponding docking station. Besides, they needed to take off the sensors during bathing and swimming activities. They received a leaflet and instruction videos to ensure the proper replacement of the sensors (19, 20). Moreover, they were encouraged to journalize each non-wearing period. After seven days, the participants' parents were asked how obtrusive it was for their child to wear the sensors. They rated the obtrusiveness together with their child and had the choice between not obtrusive, little obtrusive, or very obtrusive.

Data analysis

We removed non-wearing periods based on the participants' journals and visual inspections of the sensor data. Measurement days with non-wearing periods resulting in less than ten hours of data were considered invalid and were not analyzed (21). Saturdays and Sundays without non-wearing periods resulting in less than ten hours of data were kept since we assumed the participants were sleeping in. We summarized the numbers and reasons for missing or invalid measurement days. These reasons were divided into (a) concerning all sensors at once or (b) concerning a single sensor only. Eventually, we determined the performance measures for each valid measurement day as follows:

Upper limb group

The functional hand uses of the more and less affected sides were estimated with functional activity counts. Conventional activity counts were determined per second (22) but limited to periods with functional forearm elevations to minimize bias from walking and wheeling activities (9). These counts were summed to derive the daily hand use, and the use ratio was calculated by dividing the counts of the more affected hand by the counts of the less affected hand.

Wheelchair group

First, lying, sitting, and standing positions were classified with the orientation of the thigh and trunk sensors (10). Then, we derived the time spent in each position per day based on these classifications.

Wheeling periods were detected with the sensor on the wheel and by applying predefined rules to the gyroscope data of the wheel sensor (23). Subsequently, these wheeling periods were classified as active or passive wheeling with the orientation of the wrist sensor of the dominant hand (10). First, the wheeling speed was determined by multiplying the angular rate with the radius of the wheel. Then, active wheeling speed and distance were calculated by averaging and integrating the wheeling speed during all active wheeling periods, respectively. Moreover, the ratio between active wheeling and total wheeling distance was determined.

Walking group

Specific characteristics of the ankle's gyroscope signal were used to identify walking periods (10). These periods were further classified as level walking or stair climbing based on the altimeter of the ankle sensor (10). In level walking periods, the walking speed and distance were determined by segmenting the data into individual gait cycles and deriving each stride length and stride time (11, 24). Besides, level walking periods were separated into free and assisted walking based on the walking aid's acceleration signal (10). Eventually, we determined the duration, distance, and mean speed of all level walking periods, the ratio between assisted and total walking duration, and the altitude change while going up- and downstairs.

Statistical analysis of the day-to-day variability

We fitted a linear mixed-effects model to each performance measure Y using maximum likelihood:

Yij=μ+αi+βj+εij,

where αi is the fixed effect of weekday i and βj is the random effect of participant j. We assumed that the random effects and the residuals are normally distributed as

βji.i.d.N(0,σparticipants2),εiji.i.d.N(0,σresiduals2).

Mixed models were chosen, since they allow for the inclusion of incomplete datasets. Participants without any valid measurement day were excluded from this analysis. Then, we conducted an F-test with the fixed effects to determine whether the performance measures differed significantly between weekdays. Post-hoc analyses were done by pairwise comparisons between the estimated marginal means of each weekday, and p-values were adjusted for multiple comparisons with Tukey's method. Afterward, we calculated relative reliability with the intra-class correlation coefficient (ICC) as follows (25):

ICC(3,k)=σparticipants2σparticipants2+(σresiduals2/k)

where k is the number of measurement days. Initially, we set k = 7 to reflect an ICC of the average performance measure of seven measurement days. Confidence intervals of the ICC scores were estimated with bootstrapping. Then, we determined the minimum number of required measurement days kICC>0.8 to obtain an ICC score of which the confidence interval is above 0.8. This value reflects acceptable reliability (26). Finally, we determined absolute reliability with the smallest detectable change (SDC) (27):

SDC=1.962σresiduals7

For interpretability, SDC can be expressed as a percentage value, the SDC%, which was defined as follows (28):

SDC%=SDCgrandmean100.

To estimate the sample size, we used the method from Walter et al. (29) With an expected ICC score of 0.9, we would need 24 participants in each subgroup to reject the null hypothesis (ICC = 0.8) with a Type I error of 5%.

We determined the performance measures in MATLAB R2018b (MathWorks, Natick, Massachusetts, USA) and conducted the statistical analysis in R 4.1.2. (R Core Team, Vienna, Austria).

Results

The upper limb group consisted of 43 participants. Eleven of those used a manual wheelchair (wheelchair group), and 31 walked for household distances (walking group). One participant was only recruited for the upper limb group. In the walking group, six participants used a walker, two used crutches, and two used both devices in daily life. The remaining 21 participants walked freely. The participants' demographics, motor abilities, and obtrusiveness ratings are listed in Table 2.

TABLE 2
www.frontiersin.org

Table 2. Participants’ demographics, motor abilities and obtrusiveness rating.

Most participants rated wearing the sensors as not obtrusive and all but one as not or little obtrusive. Besides, 14% of the measurement days were missing or invalid because of reasons concerning all the sensors at once. Additionally, specific performance measures could not be determined because the data of single sensors were missing. The rate of missing measurement days depended on the sensor position and ranged between 0% and 26%. Taking reasons concerning all sensors and single sensors together, the rate of missing values ranged between 11% and 36%. Specific numbers and reasons for missing values are shown in Table 3. Forgetting to put on the sensors in the morning, forgetting to reattach them after showering or bathing, and prolonged swimming activities resulted in 20 invalid measurement days with less than ten hours of data. Two children refused to wear the sensors after one and two days, respectively, resulting in 11 missing measurement days. Families often forgot to charge the sensors on assistive devices or did not replace them in the morning, which resulted in 20 missing measurement days. The sensor on the thigh slipped down to the shank on 9 measurement days, resulting in confusion between sitting and standing positions and thus in invalid datasets.

TABLE 3
www.frontiersin.org

Table 3. Numbers and reasons for missing measurement days divided into concerning all sensors at once or a single sensor only.

The results of the day-to-day variability, including the F-test, the ICC, and the SDC are shown in Table 4. Descriptive statistics of the wearing time, the performance measures, and the pairwise comparison between weekdays are illustrated in Supplementary file S1. Participants without valid measurement days were excluded from this analysis explaining the altered number of participants and missing values in Table 4 compared to Tables 2, 3. One participant had to be excluded from the upper limb group, one participant from the wheelchair group, while two participants had to be excluded from the walking group. Five performance measures showed significant differences between weekdays. Participants were less active on Saturday and Sunday than on school days. This trend was observed in all performance measures related to the duration or amount of a motor activity except for stair climbing. The measurement of speed and ratios did not significantly differ between weekdays. The ICC (3,7) ranged between 0.82 and 0.98. Upper limb-, standing-, and walking-related performance measures revealed higher ICC scores and would require one to two measurement days to obtain reliable outcomes. The remaining performance measures would require 5–8 measurement days. The SDC% ranged between 16% and 98%, with lower values for upper limb and speed-related measures, and for the duration of sitting and walking activities.

TABLE 4
www.frontiersin.org

Table 4. Day-to-day variability of each performance measure.

Discussion

This study investigated the acceptability of wearable inertial sensors to monitor everyday life motor activities, the completeness of data, and the reliability of motor performance measures in children and adolescents with neuromotor impairments.

Acceptability and completeness of data

The large majority of children and adolescents was willing to wear the sensors for a week and perceived them as only minimally affecting their everyday life motor activities. This is in line with previous studies investigating the acceptability of wrist-worn sensors in typically developing children (30), and a waist-worn sensor in children with cerebral palsy (31). Refusing to wear the sensors resulted only in 4% of missing values in our study. However, there were other issues leading to missing or invalid measurement days. Insufficient wearing time, loose thigh sensors, and issues related to charging and replacing the sensors were the main reasons for missing measurement days. Our sensors' waterproofness allowed for short showers but not for bathing and swimming activities. The latter resulted in prolonged non-wearing periods, especially because of forgetting to reattach the sensors afterwards. Hence, improving the sensors' waterproofness would decrease the rate of invalid measurement days. Still, non-wearing periods can probably not be prevented completely, and future studies should implement strategies to impute missing sensor data (12). The thigh sensor could be firmly attached with adhesive tape, but studies using this approach had similar rates of missing values (32, 33). The battery life of wearable sensors needs to be improved to avoid having to charge the sensors overnight and allow 24 h-measurements over a week. Our algorithm relies on gyroscope data and Bluetooth communication between sensors to derive valid estimates of motor performance. However, these technologies have a high energy consumption and currently prevent longer measurements than two to three days.

Day-to-day variability of performance measures

Children and adolescents were less active on weekends compared to school days. This confirms the findings of comparable studies (3436), and underpins the need to measure performance on weekend days and weekdays to capture the children's and adolescents' motor activities comprehensively (37).

The ICC and SDC values of our study correspond to average performance measures of seven repeated measurement days, which allows for dividing the variance of the residuals by seven (see Chapter 2.4). With this, all ICC scores exceeded the desired value of 0.8, indicating sufficient reliability to discriminate patients with different levels of motor performance. However, the SDC values seem to be rather large, implying a large between-day variability of motor performance. Consequently, we recommend measuring performance over a week, even though the ICC scores of some performance measures suggest that fewer measurement days would be sufficient. The heterogeneous study population with various diagnoses and different levels of motor impairment might have led to a large between-subject variability which explains the high ICC scores despite day-to-day variability of motor performance.

Hand use and walking-related measures revealed higher relative and absolute reliability than the wheelchair group and stair climbing measures. The SDC% of the different body positions' durations were comparable to previous findings in children with cerebral palsy (34). We suggest measuring body positions with a 24 h protocol to improve reliability. Otherwise, the lying and sitting durations depend more on the waking and non-wearing time than on the patients' actual performance, which could explain the large day-to-day variability in these measures. The wheeling-related performance measures were less reliable than in adults, for whom only four measurement days are sufficient to obtain reliable outcomes (38). An explanation could be the smaller between-participant variability of wheeling activity in children compared to adults, which leads to smaller ICC scores. Thus, more measurement days are needed to obtain reliable outcomes in children, which underpins the need to measure performance over a week in children and adolescents with neuromotor impairments. In contrast, walking-related measures were more reliable than previous findings in children with cerebral palsy (34, 39). However, the comparability is limited due to differences in the study population and the number of measurement days. The altitude change during stair climbing periods revealed the lowest reliability coefficients in this study. An explanation could be the limited accuracy to detect stair climbing periods (10), which adds a source of error to the day-to-day variability.

Study limitations

This study has three main limitations. First, the sample sizes of the wheelchair group and ambulatory children using walking aids were smaller than the desired value of the sample size calculation. This explains the larger confidence intervals in these groups, and the estimated numbers of required measurement days to obtain reliable outcomes might be too high. Still, the null hypothesis was rejected in six of the seven corresponding performance measures, indicating sufficient power in these subgroups. Another limitation could be the heterogeneity of the study population. The findings of this study might depend on age or the underlying diagnoses. However, this has to be shown in larger studies with sufficient participants in each subgroup. Last, the variability between days is composed of actual differences in movement behavior and measurement error of the sensors and the algorithm. These sources of variability cannot be disentangled with the chosen study design. Therefore, improvements in sensor technology and the underlying algorithm might enhance the reliability of performance measures. However, this needs to be investigated in future studies.

Conclusion

In children and adolescents with neuromotor impairments, we recommend monitoring everyday life motor activities on seven consecutive days. The target population accepted this measurement protocol, it covers school days and weekend days, and the number of measurement days is sufficient to obtain reliable estimates of motor performance. However, the battery life of the chosen sensor technology should last for a week, too. This would decrease the non-wearing time during waking hours, in which the sensors ran out of battery or the users forgot to reattach the sensors after charging them. Moreover, a sufficiently long battery life would allow for 24 h-measurements and a comprehensive view of the patients' daily activities.

Contribution to the field

Pediatric rehabilitation aims to foster functional independence in everyday life motor activities of children and adolescents with congenital and acquired injuries and illnesses. However, we currently lack the tools to assess their improvements in daily life. Therefore, we developed a wearable sensor device and a new algorithm deriving meaningful estimates of our patients' motor activities. In preceding studies, we have already shown the clinical relevance and the validity of our new technology. The findings of this study allow for a scientifically sound recommendation that motor activities need to be monitored over a week to capture the day-to-day variability of motor activities in pediatric patients. Moreover, we found a high acceptance of children and adolescents to wear the sensors in daily life. These findings are essential prerequisites to apply our new device in clinical practice and future research improving the functionality of our patients in everyday life motor activities.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by Kantonale Ethikkommission Zürich. Written informed consent to participate in this study was provided by the participants’ legal guardian/next of kin.

Author contributions

FR and RL contributed to conception and design of the study. FR and SH recruited participants, collected the data, and performed the statistical analysis. FR wrote the first draft of the manuscript. All authors contributed to the article and approved the submitted version.

Funding

This study was supported by the Walter Muggli Fund of the ACCENTUS Foundation, Zurich, Switzerland; the Children's Research Center of the University Children's Hospital of Zurich, Switzerland; the Anna Mueller Grocholski Foundation, Zurich, Switzerland; and the Vontobel Foundation, Zurich, Switzerland. The funding sources were not involved in designing the study; in collecting, analyzing, and interpreting data; and in writing the report. Open access funding provided by ETH Zurich.

Acknowledgments

We want to thank Nadja Delliehausen for recruiting participants and collecting data. Additionally, we thank all children and adolescents and their families for participating in this research project.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fresc.2022.923328/full#supplementary-material.

References

1. Holsbeeke L, Ketelaar M, Schoemaker MM, Gorter JW. Capacity, capability, and performance: different constructs or three of a kind? Arch Phys Med Rehabil. (2009) 90(5):849–55. doi: 10.1016/j.apmr.2008.11.015

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Smits D-W, Gorter JW, van Schie PE, Dallmeijer AJ, Ketelaar M. How do changes in motor capacity, motor capability, and motor performance relate in children and adolescents with cerebral palsy? Arch Phys Med Rehabil. (2014) 95(8):1577–84. doi: 10.1016/j.apmr.2014.04.013

PubMed Abstract | CrossRef Full Text | Google Scholar

3. World Health Organization. Towards a Common Language for Functioning, Disability and Health ICF. (2002) [cited 2017 Mar 21]. Available from: http://www.who.int/classifications/icf/icfbeginnersguide.pdf

4. Clanchy KM, Tweedy SM, Boyd R. Measurement of habitual physical activity performance in adolescents with cerebral palsy: a systematic review. Dev Med Child Neurol. (2011) 53(6):499–505. doi: 10.1111/j.1469-8749.2010.03910.x

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Lang CE, Barth J, Holleran CL, Konrad JD, Bland MD. Implementation of wearable sensing technology for movement: pushing forward into the routine physical rehabilitation care field. Sensors. (2020) 20(20):5744. doi: 10.3390/s20205744

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Rachele JN, McPhail SM, Washington TL, Cuddihy TF. Practical physical activity measurement in youth: a review of contemporary approaches. World J Pediatr. (2012) 8(3):207–16. doi: 10.1007/s12519-012-0359-z

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Rast FM, Labruyère R. ICF Mobility and self-care goals of children in inpatient rehabilitation. Dev Med Child Neurol. (2020) 62(4):483–8. doi: 10.1111/dmcn.14471

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Rast FM, Labruyère R. Clinical relevance of sensor-based outcomes to monitor everyday life motor activities of children and adolescents with neuromotor impairments: a survey with health professionals. Front Rehabil Sci. (2022) In Press. doi: 10.3389/fresc.2022.865701

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Rast FM, Labruyère R. Concurrent validity of different sensor-based measures: activity counts do not reflect functional hand use in children and adolescents with upper limb impairments. Arch Phys Med Rehabil. (2022). In Press. doi: 10.1016/j.apmr.2022.03.021

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Rast FM, Jucker F, Labruyère R. Accuracy of sensor-based classification of clinically relevant motor activities in daily life of children with mobility impairments. medRxiv. (2022) 22278307; doi: 10.1101/2022.08.01.22278307

CrossRef Full Text | Google Scholar

11. Rast FM, Aschwanden S, Werner C, Labruyère R. Accuracy and comparison of sensor-based gait speed estimations under standardized and daily life conditions in children undergoing rehabilitation. J NeuroEngineering Rehabil. (2022) In Press. doi: 10.1186/s12984-022-01079-3

CrossRef Full Text | Google Scholar

12. Stephens S, Beyene J, Tremblay MS, Faulkner G, Pullnayegum E, Feldman BM. Strategies for dealing with missing accelerometer data. Rheum Dis Clin N Am. (2018) 44(2):317–26. doi: 10.1016/j.rdc.2018.01.012

CrossRef Full Text | Google Scholar

13. Sonenblum SE, Sprigle S, Lopez RA. Manual wheelchair use: bouts of mobility in everyday life. Rehabil Res Pract. (2012) 2012:753165. doi: 10.1155/2012/753165

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Dollman J, Okely AD, Hardy L, Timperio A, Salmon J, Hills AP. A hitchhiker’s guide to assessing young people’s physical activity: deciding what method to use. J Sci Med Sport. (2009) 12(5):518–25. doi: 10.1016/j.jsams.2008.09.007

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Clemes SA, Biddle SJH. The use of pedometers for monitoring physical activity in children and adolescents: measurement considerations. J Phys Act Health. (2013) 10(2):249–62. doi: 10.1123/jpah.10.2.249

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Randall M, Imms C, Carey LM, Pallant JF. Rasch analysis of the melbourne assessment of unilateral upper limb function. Dev Med Child Neurol. (2014) 56(7):665–72. doi: 10.1111/dmcn.12391

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Avery LM, Russell DJ, Raina PS, Walter SD, Rosenbaum PL. Rasch analysis of the gross motor function measure: validating the assumptions of the Rasch model to create an interval-level measure. Arch Phys Med Rehabil. (2003) 84(5):697–705. doi: 10.1016/S0003-9993(02)04896-7

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Popp WL, Schneider S, Bär J, Bösch P, Spengler CM, Gassert R, et al. Wearable sensors in ambulatory individuals with a spinal cord injury: from energy expenditure estimation to activity recommendations. Front Neurol. (2019) 10:1092. doi: 10.3389/fneur.2019.01092

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Rast F. Instruktionsvideo Fussgänger. YouTube (2020) [cited 2022 Mar 16]. Available from: https://www.youtube.com/watch?v=DXf3(2VLOqI

20. Rast F. Instruktionsvideo Rollstuhl. YouTube (2020) [cited 2022 Mar 16]. Available from: https://www.youtube.com/watch?v=S7GrVwy5dn0

21. Rich C, Geraci M, Griffiths L, Sera F, Dezateux C, Cortina-Borja M. Quality control methods in accelerometer data processing: defining minimum wear time. Dasgupta K, editor. PLoS ONE. (2013) 8(6):e67206. doi: 10.1371/journal.pone.0067206

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Brønd JC, Andersen LB, Arvidsson D. Generating ActiGraph counts from raw acceleration recorded by an alternative monitor. Med Sci Sports Exerc. (2017) 49(11):2351–60. doi: 10.1249/MSS.0000000000001344

CrossRef Full Text | Google Scholar

23. Popp WL, Brogioli M, Leuenberger K, Albisser U, Frotzler A, Curt A, et al. A novel algorithm for detecting active propulsion in wheelchair users following spinal cord injury. Med Eng Phys. (2016) 38(3):267–74. doi: 10.1016/j.medengphy.2015.12.011

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Werner C, Awai Easthope C, Curt A, Demkó L. Towards a mobile gait analysis for patients with a spinal cord injury: a robust algorithm validated for slow walking speeds. Sensors. (2021) 21(21). doi: 10.3390/s21217381

CrossRef Full Text | Google Scholar

25. Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. (1979) 86(2):420–8. doi: 10.1037/0033-2909.86.2.420

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Trost SG, Mciver KL, Pate RR. Conducting accelerometer-based activity assessments in field-based research. Med Sci Sports Exerc. (2005) 37(Suppl.):S531–43. doi: 10.1249/01.mss.0000185657.86065.98

PubMed Abstract | CrossRef Full Text | Google Scholar

27. de Vet HCW, Terwee CB, Knol DL, Bouter LM. When to use agreement versus reliability measures. J Clin Epidemiol. (2006) 59(10):1033–9. doi: 10.1016/j.jclinepi.2005.10.015

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Flansbjer U-B, Holmbäck AM, Downham D, Patten C, Lexell J. Reliability of gait performance tests in men and women with hemiparesis after stroke. J Rehabil Med. (2005) 37(2):75–82. doi: 10.1080/16501970410017215

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Walter SD, Eliasziw M, Donner A. Sample size and optimal designs for reliability studies. Stat Med. (1998) 17(1):101–10. doi: 10.1002/(SICI)1097-0258(19980115)17:1%3C101::AID-SIM727%3E3.0.CO;2-E

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Mackintosh KA, Chappel SE, Salmon J, Timperio A, Ball K, Brown H, et al. Parental perspectives of a wearable activity tracker for children younger than 13 years: acceptability and usability study. JMIR MHealth UHealth. (2019) 7(11):e13858. doi: 10.2196/13858

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Wiedmann I, Grassi M, Duran I, Lavrador R, Alberg E, Daumer M, et al. Accelerometric gait analysis devices in children—will they accept them? Results from the AVAPed study. Front Pediatr. (2021) 8:574443. doi: 10.3389/fped.2020.574443

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Schneller MB, Bentsen P, Nielsen G, Brønd JC, Ried-Larsen M, Mygind E, et al. Measuring children’s physical activity: compliance using skin-taped accelerometers. Med Sci Sports Exerc. (2017) 49(6):1261–9. doi: 10.1249/MSS.0000000000001222

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Duncan S, Stewart T, Mackay L, Neville J, Narayanan A, Walker C, et al. Wear-time compliance with a dual-accelerometer system for capturing 24-h behavioural profiles in children and adults. Int J Environ Res Public Health. (2018) 15(7):1296. doi: 10.3390/ijerph15071296

CrossRef Full Text | Google Scholar

34. Gerber CN, Carcreff L, Paraschiv-Ionescu A, Armand S, Newman CJ. Reliability of single-day walking performance and physical activity measures using inertial sensors in children with cerebral palsy. Ann Phys Rehabil Med. (2019) 64(3):101250. doi: 10.1016/j.rehab.2019.02.003

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Van Wely L, Becher JG, Balemans ACJ, Dallmeijer AJ. Ambulatory activity of children with cerebral palsy: which characteristics are important? Activity of children with CP. Dev Med Child Neurol. (2012) 54(5):436–42. doi: 10.1111/j.1469-8749.2012.04251.x

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Bloemen MAT, van den Berg-Emons RJG, Tuijt M, Nooijen CFJ, Takken T, Backx FJG, et al. Physical activity in wheelchair-using youth with spina bifida: an observational study. J Neuroeng Rehabil. (2019) 16(1):9. doi: 10.1186/s12984-018-0464-x

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Mâsse LC, Fuemmeler BF, Anderson CB, Matthews CE, Trost SG, Catellier DJ, et al. Accelerometer data reduction: a comparison of four reduction algorithms on select outcome variables. Med Sci Sports Exerc. (2005) 37(Suppl.):S544–54. doi: 10.1249/01.mss.0000185674.09066.8a

CrossRef Full Text | Google Scholar

38. Schneider S, Popp WL, Brogioli M, Albisser U, Demkó L, Debecker I, et al. Reliability of wearable-sensor-derived measures of physical activity in wheelchair-dependent spinal cord injured patients. Front Neurol. (2018) 9:1039. doi: 10.3389/fneur.2018.01039

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Ishikawa S, Kang M, Bjornson KF, Song K. Reliably measuring ambulatory activity levels of children and adolescents with cerebral palsy. Arch Phys Med Rehabil. (2013) 94(1):132–7. doi: 10.1016/j.apmr.2012.07.027

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: rehabilitation, children with disabilities, activities of daily living, accelerometry, patient compliance, reproducibility of results

Citation: Rast FM, Herren S and Labruyère R (2022) Acceptability of wearable inertial sensors, completeness of data, and day-to-day variability of everyday life motor activities in children and adolescents with neuromotor impairments. Front. Rehabilit. Sci. 3:923328. doi: 10.3389/fresc.2022.923328

Received: 19 April 2022; Accepted: 1 September 2022;
Published: 9 December 2022.

Edited by:

Josep M. Font-Llagunes, Universitat Politecnica de Catalunya, Spain

Reviewed by:

Roger Gallart Agut, Spanish National Research Council (CSIC), Spain
Serena Maggioni, Hocoma (Switzerland), Switzerland

© 2022 Rast, Herren and Labruyère. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Fabian Marcel Rast fabian.rast@hest.ethz.ch

Specialty Section: This article was submitted to Rehabilitation in Neurological Conditions, a section of the journal Frontiers in Rehabilitation Sciences

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.