As a portable, radiation-free imaging modality, electrical impedance tomography (EIT) technology has shown promise in the bedside visual assessment of lung perfusion distribution in critically ill patients. The two main methods of EIT for assessing lung perfusion are the pulsatility and conductivity contrast (saline) bolus method. Increasing attention is being paid to the saline bolus EIT method in the evaluation of regional pulmonary perfusion in clinical practice. This study seeks to provide an overview of experimental and clinical studies with the aim of clarifying the progress made in the use of the saline bolus EIT method. Animal studies revealed that the saline bolus EIT method presented good consistency with single-photon emission CT (SPECT) in the evaluation of lung regional perfusion changes in various pathological conditions. Moreover, the saline bolus EIT method has been applied to assess the lung perfusion in a pulmonary embolism and the effect of positive end-expiratory pressure (PEEP) on regional ventilation/perfusion ratio (V/Q) and acute respiratory distress syndrome (ARDS) in several clinical studies. The implementation of saline boluses, data analyses, precision, and cutoff values varied among different studies, and a consensus must be reached regarding the clinical application of the saline bolus EIT method. Further study is required to validate the impact of the described saline bolus EIT method on decision-making, therapeutic management, and outcomes in critically ill patients.
Introduction
The evaluation of regional pulmonary perfusion is of great interest for cardiopulmonary management in the intensive care unit (ICU). For many years, researchers have been searching for imaging techniques suitable for the assessment of pulmonary blood flow. Recent years have witnessed substantial progress in this area, and several techniques for pulmonary perfusion assessment, including MRI, CT, PET, and single-photon emission CT (SPECT), have become available (Hopkins et al., 2012; McCollough et al., 2015; Bondesson et al., 2019; Kohli et al., 2019). However, these methods are rarely used in critically ill patients because of the non-bedside implementation, high risk of transfer, inconvenience, etc.
Electrical impedance tomography uses a single-ring external electrode to image changes in impedance within a transversal section of the lungs; it is a non-invasive, radiation-free imaging modality that enables the continuous bedside monitoring of ventilation. A multitude of studies has confirmed the feasibility and reliability of ventilation images obtained by electrical impedance tomography (EIT) to inform mechanical ventilation settings (PEEP titration, lung recruitment, etc.) in critically ill patients (Frerichs et al., 2017; Yoshida et al., 2019). Electrical impedance tomography is also able to assess lung perfusion distribution. There are two EIT methods used for assessing lung perfusion: the pulsatility and conductivity contrasting (saline) bolus methods. Recently, increasing attention has been given to the application of the saline bolus EIT method in the visual evaluation of regional pulmonary perfusion in severe patients. This study seeks to provide an overview of experimental and clinical studies with the aim of clarifying the progress made in the use of the saline bolus EIT method.
Pulsatility-Based EIT for Assessing Lung Perfusion
The pulsatility-based EIT method of measuring cardiac-related impedance variation, which has been widely investigated in both unwell and healthy subjects (Frerichs et al., 2009; Grant et al., 2011; Li et al., 2014), features non-invasive and real-time perfusion monitoring. In the last two decades, developments in EIT signal separating techniques involving frequency-domain filtering (Frerichs et al., 2009), ECG gating (Vonk Noordegraaf et al., 1998), respiratory pause (Fagerberg et al., 2009), and principal component analysis (PCA) (Deibele et al., 2008) have optimized pulsatile signal processing. We summarized the clinical studies using pulsatility-based EIT assessments for pulmonary perfusion in healthy volunteers and patients with pulmonary hypertension (PH) (Table 1).
The basic principle of the pulsatility method for estimating pulmonary perfusion is based on the measurement of pulsatile changes in pulmonary blood volume instead of real forward lung blood flow. Hence, many potential factors (such as changes in vascular tone, synchronous changes in air content, airway pressure, and distensibility of the small pulmonary vessels) could affect the accuracy of using cyclic pulsatile changes in regional impedance to reflect the proportion of lung blood flow in those regions. The validity of the pulsatility method for assessing perfusion has been questioned (Borges et al., 2012; Putensen et al., 2019). Studies have found that downstream pulmonary vascular resistance and the distensibility of small pulmonary vessels can impact EIT pulsatility data (Smit et al., 2004; Borges et al., 2012). Borges JB et al. found that pulsatile impedance increased within the collapsed lung region, which had reduced lung blood flow based on both the SPECT and saline bolus EIT methods (Borges et al., 2012). Although the pulsatility method could easily provide continuous information on lung perfusion, the conductivity contrast bolus injection method showed superiority in terms of feasibility and accuracy (Frerichs et al., 2002; Deibele et al., 2008; Borges et al., 2012).
Principles of the Conductivity Contrast Bolus EIT Method
The conductivity contrast bolus method, also named contrast-enhanced EIT or the saline bolus EIT method, was performed by introducing a bolus of hypertonic saline into the pulmonary circulation during a respiratory hold. During the breath-hold, tidal impedance was lacking; however, chest impedance was relatively constant. Hence, the change in impedance caused by the saline bolus reflects forward lung blood flow. The high variation in the regional impedance indicates more saline passing through the corresponding region, which also indicates high local lung perfusion. According to the typical first-pass kinetics manifested in indicator-enhanced CT and MRI (Konstas et al., 2009; Ohno et al., 2016), hypertonic saline following the blood flow from the right atrium through the pulmonary circulation is indicative of a significant decline in electrical impedance due to its high conductivity. The impedance–time curves produced by saline boluses were reconstructed to fit a gamma variate model and were quantitively analyzed based on the Fick principle and the accumulated mass of the indicator in every region of interest (Mullani and Gould, 1983).
In general, there were two methods used to analyze the indicator dilution, namely, the transfer model and the slope analysis. The slope analysis was validated against PET in animal studies (Borges et al., 2012; Bluth et al., 2019). Several recent publications on lung perfusion using EIT and saline bolus injections in human subjects adopted the slope analysis (He et al., 2020a,b; Mauri et al., 2020; Spinelli et al., 2021). A high slope in the lung regional time–impedance curves obtained following a saline injection indicates a high accumulation of saline, which also reflects a high lung blood perfusion.
More recently, Kircher et al. optimized an EIT-gamma function fitting based on a linear piecewise approximation of the drift model in a porcine acute respiratory distress syndrome (ARDS) model, simplifying the quantization algorithm for potential clinical applications (Kircher et al., 2021). Simultaneously, following the administration of a saline bolus, by calculating the SD of each pixel and plotting them at the relevant image position, a new type of EIT image presenting the exact position, into which the indicator is carried by the bloodstream and visualizing the temporal variation in perfusion distribution can be generated (Frerichs et al., 2002).
Implementation of the Saline Bolus Method
The experimental and clinical studies using the saline bolus EIT method to visually assess lung perfusion are summarized in Tables 2–4 (Frerichs et al., 2002; Borges et al., 2012; Nguyen et al., 2013, 2015; Reinius et al., 2015; Hentze et al., 2018; Bluth et al., 2019; Grassi et al., 2020; He et al., 2020a,b,c, 2021b; Mauri et al., 2020; Safaee Fakhr et al., 2020; Kircher et al., 2021; Spinelli et al., 2021; Yuan et al., 2021). The administration of saline boluses, data analyses, precision, and cutoff values differ among studies, and a consensus must be reached regarding the clinical application of the saline bolus EIT method.
Breath Holding
With the aim of reducing the influence of tidal impedance, breath-holding is an important part of saline bolus implementation. It remains controversial how the breath should be held and how it should be implemented.
How Long Does the Breath-Holding Maneuver Last?
The mean lung transit time, which is defined as the time required for blood flow transfer from the right ventricle to the left atrium, should be taken into consideration regards to the length of the breath-hold. However, the apnea time does not need to be as long as the pulmonary transit time (PTT) for lung perfusion (He et al., 2021a). Recent experimental explorations suggested that it took ~3–5 s from the injection for the saline to enter the lungs and ultimately concentrate (Hentze et al., 2018). An 8- to 20-s breath-hold maneuver was used in different experimental and clinical studies (Table 2). Breath-holding for a longer duration might demand neuromuscular blocking agents. In contrast, breath-holding for a shorter time might be more practicable at the bedside. Our experience showed that deep sedation (Richmond Agitation Sedation Scale at—4) was enough to obtain a breath hold of 8 s. Moreover, it should be considered that some conditions, such as extremely low/high cardiac outputs, have apparent abnormal pulmonary transit times. Further study is required to optimize the breath-holding time and ensure the effective acquisition of the entire impedance curve.
How Should Breath-Holding Be Performed?
Both end-expiratory and end-inspiratory holds were used during the saline bolus implementation for lung perfusion measurement. The end-expiratory hold had the following potential benefits: (1). little impact on venous return and circulation; (2). a distinct impedance change might be more easily caused by a saline bolus at end-expiration with a lower global impedance baseline. On the other hand, end-inspiratory holding with maximal inflation of the lungs can suppress respiratory drive and dilute carbon dioxide (CO2) levels, and it is considered to be more readily implemented. However, the higher airway pressure at end-inspiration may cause an impairment of global circulation and lung perfusion (He et al., 2021a). Further exploration is necessary to compare the clinical applicability of end-inspiration and end-expiration occlusion methods.
Saline Bolus Injection
Briefly, breath-holding was required during the saline bolus injection. If spontaneous breathing was present during the saline bolus administration, the impedance–time curve would be unfit for assessing lung perfusion. It is worth noting that the saline bolus must be injected manually as soon as possible (<2 s) through the central venous catheter (He H. W. et al., 2021).
Due to its high conductivity, hypertonic saline was used as an EIT contrast agent. The concentrations (from 3 to 20%) and volume (from 5 to 60 ml) of saline solution varied among different studies (shown in Tables 2, 3). Technically, a higher concentration of saline solution could improve the signal-to-noise ratio of EIT imaging but could introduce safety risks regarding potential pulmonary fluid transfer and a faster variation in hemodynamic parameters (Hellige et al., 2012). There has been no consensus on the frequency of the saline bolus injection over 1 day. However, caution should be paid to excessive saline infusion that could disturb electrolyte balance.
EIT Data Analysis
Functional ventilation maps were derived by averaging the tidal impedance variation images. Functional perfusion maps were calculated as the slope of the regional impedance–time curves after the saline bolus injection. Figure 1 illustrates the impedance–time curves and the corresponding ventilation/perfusion images following the injection of a saline bolus.
Figure 1. Illustrates impedance–time curves and corresponding ventilation/perfusion images during the saline bolus. The tidal impedance variation during normal tidal breathing before apnea was used to calculate the tidal ventilation image. The impedance–time curve caused by saline bolus during the apnea period was used for the perfusion image (adapted from our recent paper).
The perfusion value of pixel i Pi in the perfusion image was equal to –ai. Furthermore, the ventilated and perfused regions were defined as follows: region k was ventilated if
and, similarly, region g was perfused if
Moreover, regional ventilation/perfusion ratio (V/Q) matching could be obtained based on the combination of the ventilation and perfusion distribution images from EIT (He et al., 2020b,c). Three regions were identified based on the ventilation/perfusion patterns: regions that were only ventilated (RVs), regions that were only perfused (RPs), and regions that were both ventilated and perfused (RV + Ps). The following V/Q matching from the EIT-derived parameters was obtained:
Mauri and Spinelli et al. defined non-ventilated or non-perfused pixels when the Vpixel or Qpixel were less than or equal to 10% of the highest pixel-level impedance change in a similar method (Mauri et al., 2020). Subsequently, the percentage of the pixels with a V/Q mismatch was calculated to investigate the relationship between the mismatches and the prognoses of ARDS (Spinelli et al., 2021).
Recently, Perier et al. developed an analytical method for assessing lung perfusion and V/Q matching with EIT (Perier et al., 2020), including methods for determining the impedance changes caused by ventilation (ΔZV) and perfusion (ΔZQ), cardiac output, and alveolar ventilation volume. The authors assumed a 30% fixed anatomical dead space; the V/Q matching was calculated pixel by pixel, and then relevant parameters were defined:
Moreover, Nguyen et al. compared the different parameters of the averaged contrast dilution curve, including the peak value, maximum uptake, maximum washout, and area under the curve to estimate the relative distribution of pulmonary perfusion in an animal model of pulmonary embolism-like events (Nguyen et al., 2015). The authors found that the right lung-to-left lung perfusion ratios of the area under the curve and the peak value of the averaged contrast dilution curve are the most promising and reliable parameters in assessing pulmonary embolism (PE). Variation was found among different impedance–time curve analysis methods in two patients (He et al., 2021a), and further study is required to optimize the analysis methods.
Applications of Contrast EIT in Animal and Clinical Studies
The availability of contrast EIT presented in multiple experimental and clinical studies allowed the instant measurement of lung regional perfusion changes under various pathological conditions, such as PE, atelectasis, and ARDS (Frerichs et al., 2002; Borges et al., 2012; He et al., 2020a,c; Mauri et al., 2020; Kircher et al., 2021). A remarkable agreement between EIT and SPECT further proved the correctness and practicability of this emerging perfusion imaging method.
Animal Studies
We summarized eight animal studies on conductivity contrast bolus EIT, four of which compared EIT and SPECT perfusion results and presented significant consistency (Table 2).
In a regional lung artery occlusion model, Frerichs et al. were the first to demonstrate the feasibility of EIT imaging in detecting pulmonary perfusion impairments when a conductive contrast agent was used (Frerichs et al., 2002). A strong correlation was found in the paired measurements of the perfusion distribution by EIT and SPECT, with a mean difference of 0.6% and an SD of 2.9% (Borges et al., 2012). In addition, the redistributive pulmonary perfusion, as estimated by EIT, with a distinct decrease in perfusion within the injured lung compartment between the injured and healthy regions, was validated by a dynamic multidetector computed tomography (MDCT) in a study of animal models with induced regional sublobar ARDS with reasonable limits of agreement value of 8.74% (Kircher et al., 2021).
Clinical Studies
The following section summarizes nine clinical studies on contrast EIT, including four case reports and five clinical trials (Table 3).
It was demonstrated that EIT perfusion imaging is accurate and effective in diagnosing and monitoring the therapeutic response of PE at the bedside. The normal ventilation distribution with massive perfusion defects in the affected regions as determined by EIT imaging showed strong diagnostic efficiency in studies comprising subjects suffering multiple embolisms confirmed by CT pulmonary angiography (Grassi et al., 2020; He et al., 2020a,c; Safaee Fakhr et al., 2020; Yuan et al., 2021). He et al. found that patients with PE had significantly higher dead space% and lower V/Q match% than patients without PE in a prospective observational study (He et al., 2020c). A cutoff value of 30.37% for dead space% resulted in a sensitivity of 90.9% and a specificity of 98.6% for PE diagnoses (He et al., 2020c). Moreover, the same team utilized saline bolus EIT to assess the relationship of ventilation–perfusion matching with regional overdistention and recruitment induced by PEEP in patients with ARDS (He et al., 2020b).
Marui et al. found that EIT could detect imbalanced ventilation/perfusion matches based on ventilated areas with impaired perfusion in ARDS resulting from COVID-19 (C-ARDS) and predict the therapeutic outcomes of ARDS patients when considering a variety of EIT-based information (Mauri et al., 2020; Perier et al., 2020; Spinelli et al., 2021). Perier et al., using EIT, confirmed that prone positioning and increased PEEP resulted in better V/Q matches in patients with C-ARDS (Perier et al., 2020).
Prospects and Limitations
Saline bolus EIT can determine differences in regional perfusion resulting from perfusion distributions reflecting the diseases, therapeutic maneuvers (e.g., administration of vasoactive agents, and mechanical ventilation), and ventilation/perfusion maps of multiple pathophysiological phenomena. Therefore, it is possible for the contrast EIT method to have clinical applications in the future. First, saline bolus-based EIT might have the potential to identify various etiologies of acute respiratory failure (ARF). Recently, He et al. demonstrated that contrast EIT-derived regional ventilation and perfusion measurements were able to characterize three broad ARF etiologies, including pulmonary embolism-related disease (PED), diffuse lung involvement disease (DLD), and focal lung involvement disease (FLD) (He et al., 2021b). Second, for mechanically ventilated patients, increasing clinical evidence has shown that perfusion EIT could provide bedside guidance for tidal setting, PEEP titration, and prone ventilation and help physicians evaluate the prognosis of patients with ARDS. Third, this saline bolus-based EIT method featured with convenience and promptness might be able to satisfy the clinical use for monitoring therapeutic effects for severe patients, for example, to assess the effects of inhaled medicine to PH and the therapeutic effect of anticoagulation/thrombolytic therapy on PE.
For the extensive application of the saline bolus EIT lung perfusion method in the ICU, more clinical researches and experimental studies are required to confirm the accuracy and validity of saline contrast EIT. Importantly, there is a strong need for a consensus on the cutoff value to determine pulmonary perfusion parameters, like the one for ventilated/perfused units in assessing V/Q mismatch (He et al., 2020b,c; Mauri et al., 2020; Spinelli et al., 2021). Although there has been little evidence showing the feasibility of EIT perfusion imaging in patients with spontaneous breathing, there is a great chance that breath-holding maneuvers adjustive to awake patients like inspiratory holds using typical entry standards are next in line. Furthermore, to achieve the continuous monitoring of pulmonary blood flow, it is foreseeable that the combination of contrast EIT with pulsatility-based EIT for dynamic assessments will be explored in the future.
However, the EIT analysis has drawbacks in providing detailed morphological information to accurately identify the vertical location of perfusion deficits attributed to restrained spatial resolution. Hence, the reconstruction algorithm warrants further optimization to provide robust evaluations in clinical applications. Additionally, using a single-ring external electrode, the EIT method is only allowed to monitor one cylindrical lung part at one time, instead of providing an overall lung image like CT and MRI. Moreover, auxiliary evaluations or data processing procedures to address the conductivity variations are necessary to accommodate background conductivity changes when a conductive contrast agent is used (Bluth et al., 2019). Finally, despite the validation in intubated subjects, the feasibility of perfusion EIT in spontaneous respiratory patients is yet to be fully understood. Since the effectiveness of the breath-holding maneuver may vary due to the individual variations in conscious patients, contrast EIT focusing on applications in non-mechanically ventilated patients demands future discussion before being used in clinical practice.
Conclusions
Using a hypertonic saline solution based on first-pass kinetics, contrast EIT imaging generating dilution impedance waveforms is reliable for determining lung perfusion changes reflecting atelectasis, spatial heterogeneity, and artery embolism in a regional manner. With the ability to monitor regional lung perfusion and produce V/Q mapping, saline bolus EIT is promising and meaningful for patients in ICUs. Further study is required to validate the impact of the described saline bolus EIT method on decision-making, therapeutic management, and outcomes in critically ill patients.
Author Contributions
MX, HH, and YL contributed to conception and design of the review. MX and HH searched and organized the database. MX wrote the first draft of the manuscript. HH and YL wrote and refined several important sections of the manuscript. All authors contributed to manuscript revision, read, and approved the submitted version.
Funding
This study was supported by the CAMS Innovation Fund for Medical Sciences (No. CAMS Innovation Fund for Medical Sciences (No. 2020-I2M-C&T-B-042), Capital's Funds for Health Improvement and Research (No. 2020-2-40111), Excellence Program of Beijing Clinical Key Specialty for Critical Care Medicine (2020), and Beijing Municipal Science and Technology Commission (Grant No. Z201100005520051).
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher's Note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
Bluth, T., Kiss, T., Kircher, M., Braune, A., Bozsak, C., Huhle, R., et al. (2019). Measurement of relative lung perfusion with electrical impedance and positron emission tomography: an experimental comparative study in pigs. Br. J. Anaesth. 123, 246–254. doi: 10.1016/j.bja.2019.04.056
Bondesson, D., Schneider, M. J., Gaass, T., Kühn, B., Bauman, G., Dietrich, O., et al. (2019). Nonuniform Fourier-decomposition MRI for ventilation- and perfusion-weighted imaging of the lung. Magn. Reson. Med. 82, 1312–1321. doi: 10.1002/mrm.27803
Borges, J. B., Suarez-Sipmann, F., Bohm, S. H., Tusman, G., Melo, A., Maripuu, E., et al. (2012). Regional lung perfusion estimated by electrical impedance tomography in a piglet model of lung collapse. J. Appl. Physiol. 112, 225–236. doi: 10.1152/japplphysiol.01090.2010
Carlisle, H. R., Armstrong, R. K., Davis, P. G., Schibler, A., Frerichs, I., and Tingay, D. G. (2010). Regional distribution of blood volume within the preterm infant thorax during synchronised mechanical ventilation. Intens. Care Med. 36, 2101–2108. doi: 10.1007/s00134-010-2049-4
Deibele, J. M., Luepschen, H., and Leonhardt, S. (2008). Dynamic separation of pulmonary and cardiac changes in electrical impedance tomography. Physiol. Meas. 29, S1–S14. doi: 10.1088/0967-3334/29/6/S01
Fagerberg, A., Stenqvist, O., and Åneman, A. (2009). Monitoring pulmonary perfusion by electrical impedance tomography: an evaluation in a pig model: EIT and pulmonary perfusion. Acta Anaesthesiol. Scand. 53, 152–158. doi: 10.1111/j.1399-6576.2008.01847.x
Frerichs, I., Amato, M. B. P., van Kaam, A. H., Tingay, D. G., Zhao, Z., Grychtol, B., et al. (2017). Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group. Thorax 72, 83–93. doi: 10.1136/thoraxjnl-2016-208357
Frerichs, I., Hinz, J., Herrmann, P., Weisser, G., Hahn, G., Quintel, M., et al. (2002). Regional lung perfusion as determined by electrical impedance tomography in comparison with electron beam CT imaging. IEEE Trans. Med. Imaging 21, 646–652. doi: 10.1109/TMI.2002.800585
Frerichs, I., Pulletz, S., Elke, G., Reifferscheid, F., Schädler, D., Scholz, J., et al. (2009). Assessment of changes in distribution of lung perfusion by electrical impedance tomography. Respiration 77, 282–291. doi: 10.1159/000193994
Grant, C. A., Pham, T., Hough, J., Riedel, T., Stocker, C., and Schibler, A. (2011). Measurement of ventilation and cardiac related impedance changes with electrical impedance tomography. Crit. Care 15:R37. doi: 10.1186/cc9985
Grassi, L. G., Santiago, R., Florio, G., and Berra, L. (2020). Bedside evaluation of pulmonary embolism by electrical impedance tomography. Anesthesiology 132, 896–896. doi: 10.1097/ALN.0000000000003059
He, H., Chi, Y., Long, Y., Yuan, S., Frerichs, I., Möller, K., et al. (2020b). Influence of overdistension/recruitment induced by high positive end-expiratory pressure on ventilation–perfusion matching assessed by electrical impedance tomography with saline bolus. Crit. Care 24:586. doi: 10.1186/s13054-020-03301-x
He, H., Chi, Y., Long, Y., Yuan, S., Zhang, R., Frerichs, I., et al. (2020c). Bedside evaluation of pulmonary embolism by saline contrast electrical impedance tomography method: a prospective observational study. Am. J. Respir. Crit. Care Med. 202, 1464–1468. doi: 10.1164/rccm.202005-1780LE
He, H., Chi, Y., Long, Y., Yuan, S., Zhang, R., Yang, Y., et al. (2021b). Three broad classifications of acute respiratory failure etiologies based on regional ventilation and perfusion by electrical impedance tomography: a hypothesis-generating study. Ann. Intensive Care 11:134. doi: 10.1186/s13613-021-00921-6
He, H., Long, Y., Chi, Y., Yuan, S., and Zhao, Z. (2021a). Reply to wang and zhong: bedside evaluation of pulmonary embolism by saline contrast–enhanced electrical impedance tomography: considerations for future research. Am. J. Respir. Crit. Care Med. 203, 395–397. doi: 10.1164/rccm.202010-3768LE
He, H., Long, Y., Frerichs, I., and Zhao, Z. (2020a). Detection of acute pulmonary embolism by electrical impedance tomography and saline bolus injection. Am. J. Respir. Crit. Care Med. 202, 881–882. doi: 10.1164/rccm.202003-0554IM
He, H. W., Long, Y., Chi, Y., Yuan, S. Y., Zhou, X., Su, L. X., et al. (2021). [Technology specification of bedside hypertonic saline-contrast electrical impedance tomography of lung perfusion and clinical application]. Zhonghua Yi Xue Za Zhi 101, 1097–1101. doi: 10.3760/cma.j.cn112137-20200926-02723
Hellige, N. C., Hahn, G., and Hellige, G. (2012). Comment on Borges et al. “Regional lung perfusion estimated by electrical impedance tomography in a piglet model of lung collapse”. J. Appl. Physiol. 112, 2127–2127. doi: 10.1152/japplphysiol.00286.2012
Hentze, B., Muders, T., Luepschen, H., Maripuu, E., Hedenstierna, G., Putensen, C., et al. (2018). Regional lung ventilation and perfusion by electrical impedance tomography compared to single-photon emission computed tomography. Physiol. Meas. 39:065004. doi: 10.1088/1361-6579/aac7ae
Hopkins, S. R., Wielpütz, M. O., and Kauczor, H.-U. (2012). Imaging lung perfusion. J. Appl. Physiol. 113, 328–339. doi: 10.1152/japplphysiol.00320.2012
Hovnanian, A. L. D., Costa, E. L. V., Hoette, S., Fernandes, C. J. C. S., Jardim, C. V. P., Dias, B. V., et al. (2021). Electrical impedance tomography in pulmonary arterial hypertension. PLoS ONE 16:e0248214. doi: 10.1371/journal.pone.0248214
Kircher, M., Elke, G., Stender, B., Hernandez Mesa, M., Schuderer, F., Dossel, O., et al. (2021). Regional lung perfusion analysis in experimental ARDS by electrical impedance and computed tomography. IEEE Trans. Med. Imaging 40, 251–261. doi: 10.1109/TMI.2020.3025080
Kohli, P., Kelly, V. J., Kehl, E. G., Rodriguez-Lopez, J., Hibbert, K. A., Kone, M., et al. (2019). Perfusion imaging distinguishes exercise pulmonary arterial hypertension at rest. Am. J. Respir. Crit. Care Med. 199, 1438–1441. doi: 10.1164/rccm.201810-1899LE
Konstas, A. A., Goldmakher, G. V., Lee, T.-Y., and Lev, M. H. (2009). Theoretic basis and technical implementations of CT perfusion in acute ischemic stroke, part 1: theoretic basis. AJNR Am. J. Neuroradiol. 30, 662–668. doi: 10.3174/ajnr.A1487
Li, Y., Tesselaar, E., Borges, J. B., Böhm, S. H., Sjöberg, F., and Janerot-Sjöberg, B. (2014). Hyperoxia affects the regional pulmonary ventilation/perfusion ratio: an electrical impedance tomography study: Hyperoxic pulmonary V/Q mismatch. Acta Anaesthesiol. Scand. 58, 716–725. doi: 10.1111/aas.12323
Mauri, T., Spinelli, E., Scotti, E., Colussi, G., Basile, M. C., Crotti, S., et al. (2020). Potential for lung recruitment and ventilation-perfusion mismatch in patients with the acute respiratory distress syndrome from coronavirus disease 2019*. Crit. Care Med. 48, 1129–1134. doi: 10.1097/CCM.0000000000004386
McCollough, C. H., Leng, S., Yu, L., and Fletcher, J. G. (2015). Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology 276, 637–653. doi: 10.1148/radiol.2015142631
Mullani, N. A., and Gould, K. L. (1983). First-pass measurements of regional blood flow with external detectors. J. Nucl. Med. 24:5.
Nguyen, D. T., Bhaskaran, A., Chik, W., Barry, M. A., Pouliopoulos, J., Kosobrodov, R., et al. (2015). Perfusion redistribution after a pulmonary-embolism-like event with contrast enhanced EIT. Physiol. Meas. 36, 1297–1309. doi: 10.1088/0967-3334/36/6/1297
Nguyen, D. T., Kosobrodov, R., Barry, M. A., Chik, W., Pouliopoulos, J., Oh, T. I., et al. (2013). Preliminary results on different impedance contrast agents for pulmonary perfusion imaging with electrical impedance tomography. J. Phys. Conf. Ser. 434:012079. doi: 10.1088/1742-6596/434/1/012079
Ohno, Y., Koyama, H., Lee, H. Y., Miura, S., Yoshikawa, T., and Sugimura, K. (2016). Contrast-enhanced CT- and MRI-based perfusion assessment for pulmonary diseases: basics and clinical applications. Diagn. Interv. Radiol. 22:407. doi: 10.5152/dir.2016.16123
Perier, F., Tuffet, S., Maraffi, T., Alcala, G., Victor, M., Haudebourg, A.-F., et al. (2020). Effect of positive end-expiratory pressure and proning on ventilation and perfusion in COVID-19 acute respiratory distress syndrome. Am. J. Respir. Crit. Care Med. 202, 1713–1717. doi: 10.1164/rccm.202008-3058LE
Proença, M., Braun, F., Lemay, M., Solà, J., Adler, A., Riedel, T., et al. (2020). Non-invasive pulmonary artery pressure estimation by electrical impedance tomography in a controlled hypoxemia study in healthy subjects. Sci. Rep. 10:21462. doi: 10.1038/s41598-020-78535-4
Proença, M., Braun, F., Solà, J., Lemay, M., Adler, A., Thiran, J.-P., et al. (2016). Non-invasive monitoring of pulmonary artery pressure from timing information by EIT: experimental evaluation during induced hypoxia. Physiol. Meas. 34, 713–721. doi: 10.1088/0967-3334/37/6/713
Putensen, C., Hentze, B., Muenster, S., and Muders, T. (2019). Electrical impedance tomography for cardio-pulmonary monitoring. J. Clin. Med. 8:1176. doi: 10.3390/jcm8081176
Reinius, H., Borges, J. B., Fredén, F., Jideus, L., Camargo, E. D. L. B., Amato, M. B. P., et al. (2015). Real-time ventilation and perfusion distributions by electrical impedance tomography during one-lung ventilation with capnothorax: EIT during OLV/Capnothorax. Acta Anaesthesiol. Scand. 59, 354–368. doi: 10.1111/aas.12455
Safaee Fakhr, B., Araujo Morais, C. C., De Santis Santiago, R. R., Di Fenza, R., Gibson, L. E., Restrepo, P. A., et al. (2020). Bedside monitoring of lung perfusion by electrical impedance tomography in the time of COVID-19. BJA Br. J. Anaesth. 125, e434–e436. doi: 10.1016/j.bja.2020.08.001
Smit, H. J., Vonk Noordegraaf, A., Marcus, J. T., Boonstra, A., de Vries, P. M., and Postmus, P. E. (2004). Determinants of pulmonary perfusion measured by electrical impedance tomography. Eur. J. Appl. Physiol. 92, 45–49. doi: 10.1007/s00421-004-1043-3
Smit, H. J., Vonk Noordegraaf, A., Roeleveld, R. J., Bronzwaer, J. G., Postmus, P. E., de Vries, P. M., et al. (2002). Epoprostenol-induced pulmonary vasodilatation in patients with pulmonary hypertension measured by electrical impedance tomography. Physiol. Meas. 23, 237–43. doi: 10.1088/0967-3334/23/1/324
Smit, H. J., Vonk-Noordegraaf, A., Boonstra, A., de Vries, P. M., and Postmus, P. E. (2006). Assessment of the pulmonary volume pulse in idiopathic pulmonary arterial hypertension by means of electrical impedance tomography. Respiration 73, 597–602. doi: 10.1159/000088694
Smit, H. J., Vonk-Noordegraaf, A., Marcus, J. T., van der Weijden, S., Postmus, P. E., de Vries, P. M., et al. (2003). Pulmonary vascular responses to hypoxia and hypoxia in healthy volunteers and COPD patients measured by electrical impedance tomography. Chest 123, 1803–1839. doi: 10.1378/chest.123.6.1803
Spinelli, E., Kircher, M., Stender, B., Ottaviani, I., Basile, M. C., Marongiu, I., et al. (2021). Unmatched ventilation and perfusion measured by electrical impedance tomography predicts the outcome of ARDS. Crit. Care 25:192. doi: 10.1186/s13054-021-03615-4
Vonk Noordegraaf, A., Kunst, P. W. A., Janse, A., Marcus, J. T., Postmus, P. E., Faes, T. J. C., et al. (1998). Pulmonary perfusion measured by means of electrical impedance tomography. Physiol. Meas. 19, 263–273. doi: 10.1088/0967-3334/19/2/013
Yoshida, T., Piraino, T., Lima, C. A. S., Kavanagh, B. P., Amato, M. B. P., and Brochard, L. (2019). Regional ventilation displayed by electrical impedance tomography as an incentive to decrease positive end-expiratory pressure. Am. J. Respir. Crit. Care Med. 200, 933–937. doi: 10.1164/rccm.201904-0797LE
Keywords: lung perfusion, regional ventilation/perfusion match, electrical impedance tomography, pulmonary embolism, acute respiratory distress syndrome
Citation: Xu M, He H and Long Y (2021) Lung Perfusion Assessment by Bedside Electrical Impedance Tomography in Critically Ill Patients. Front. Physiol. 12:748724. doi: 10.3389/fphys.2021.748724
Received: 28 July 2021; Accepted: 13 September 2021;
Published: 13 October 2021.
Edited by:
Ling Sang, First Affiliated Hospital of Guangzhou Medical University, ChinaCopyright © 2021 Xu, He and Long. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.